You are on page 1of 7

Changes in Automated Complete Blood Cell Count and

Differential Leukocyte Count Results Induced by


Storage of Blood at Room Temperature
Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD; Rolland Schwarting, MD

● Objective.—To delineate changes that occur in various the count was within or above the normal range), and
parameters of automated complete blood cell count (CBC) platelet count was stable for at least 4 days (up to 7 days,
and differential leukocyte count (differential) on prolonged if the count was within or above the normal range). The
storage of blood at room temperature. mean corpuscular volume, mean platelet volume, hemat-
Design.—A CBC and an automated differential were per- ocrit, and red blood cell distribution width each increased,
formed on the Coulter Gen.S on 40 K3 (tripotassium and the mean corpuscular hemoglobin concentration de-
ethylenediamine-tetraacetate) EDTA-anticoagulated blood creased from day 2 onward. Among the differential param-
specimens once daily, specimen volume permitting, for 3 eters, the relative percentages and absolute numbers of
to 7 days. Specimens were kept at room temperature neutrophils, lymphocytes, and eosinophils tended to in-
throughout the study. The results were tabulated using a
crease, whereas those of monocytes trended downward
personal computer with Excel software. Percent change or
absolute difference from the initial value for each param- over time. Limited data on basophils did not reveal an ap-
eter for each subsequent day of the study period was cal- preciable change.
culated. Conclusion.—Blood specimens stored at room temper-
Results.—Among the CBC parameters, hemoglobin, red ature for more than 1 day (up to 3 days or possibly longer)
blood cell count, and mean corpuscular hemoglobin were were found to be acceptable with some limitations for CBC
stable for the duration of the study (7 days), white blood but not for the differential.
cell count was stable for at least 3 days (up to 7 days, if (Arch Pathol Lab Med. 2002;126:336–342)

C linical laboratory professionals are familiar with the


scenario whereby a blood specimen is collected on a
Friday but not delivered to the laboratory for processing
CBC and the automated differential, is scarce, particularly
in the recent literature. This study delineates changes that
occur in various parameters of automated CBC and dif-
until the following Monday or even later. When such a ferential during storage of blood at room temperature for
specimen arrives at the laboratory, the staff needs to de- several days.
cide (1) whether to accept or reject it; (2) if accepted,
whether to perform all of the ordered tests or only those MATERIALS AND METHODS
deemed appropriate based on the age of the specimen; A total of 40 K3 (tripotassium ethylenediamine-tetraacetate)
and (3) what comments, if any, should be appended to the EDTA-anticoagulated blood specimens were processed through
reported results regarding their reliability or unreliability. the Coulter Gen.S (Beckman Coulter, Miami, Fla) for CBC and
Such decision making requires laboratorians to be familiar differential once daily, specimen volume permitting, for 3 to 7
with changes known to occur in blood specimens during days. Specimens were selected from the routine laboratory work-
load over several days to represent a range of normal and ab-
storage. Manufacturers of automated analyzers and pub-
normal CBC and differential values, as indicated in Table 1.
lished literature often cite that blood specimens, kept at Among the 40 specimens, 11 were leukopenic (white cell count,
either room temperature or at 48C (refrigerated) for up to ,4.0 3 103/mL), 16 had leukocytosis (white cell count, .11.0 3
24 hours, generally yield reliable results for complete 103/mL), 15 were thrombocytopenic (platelet count, ,140 3 103/
blood cell count (CBC) and automated differential leuko- mL), and 11 had thrombocytosis (platelet count, .400 3 103/mL).
cyte count (differential).1–4 However, specific information All but 4 specimens were anemic, with hemoglobin levels less
concerning the suitability or unsuitability of specimens than 12.5 g/dL. Specimens were kept at room temperature
older than 1 day for various laboratory tests, including the throughout the study. The Coulter Gen.S uses impedance tech-
nology to generate a 10-part CBC and volume (impedance), con-
ductivity, and light scatter measurements, commonly referred to
Accepted for publication October 19, 2001. as VCS technology, to generate a 5-part differential.1 Software
From the Department of Pathology, Anatomy, and Cell Biology, versions 1C and 2A were used during this evaluation. Calibration
Jefferson Medical College, Thomas Jefferson University, Philadel- and quality control of the analyzer were performed using S-cal
phia, Pa. and 3 levels of 5C control (both from Beckman Coulter), respec-
Reprints: Gene L. Gulati, PhD, 307 Pavilion Bldg, Thomas Jefferson tively, according to the manufacturer’s instructions.1 The 10-part
University Hospital, 125 S 11th St, Philadelphia, PA 19107 (e-mail: CBC consisted of white blood cell count (WBC), red blood cell
gene.gulati@mail.tju.edu). count (RBC), hemoglobin, hematocrit, mean corpuscular volume
336 Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al
Table 1. Range of Initial Automated CBC and some of the values obtained initially and/or on subsequent days
Differential Results of Specimens Included in were 0. Ultimately, mean percent change or difference and SD
the Study* were determined, also using Excel software, for each parameter
for each day. Adobe Illustrator, version 8.0, was used to prepare
Parameter Range the figures.
CBC
WBC, 103/mL 0.3–67.9
RESULTS
RBC, 106/mL 1.91–5.43 Changes observed in the CBC results of blood speci-
Hemoglobin, g/dL 6.2–16.4 mens stored at room temperature are summarized in Ta-
Hematocrit, % 17.9–47.3
MCV, mm3 68–110
ble 2 and illustrated in Figure 1. For each CBC parameter,
MCH, pg 21.5–37.4 mean percent change from the value on day 0 is plotted
MCHC, g/dL 31.3–36.2 along with its 95% confidence interval (CI) (62 SDs)
RDW, % 12.3–28.7 against time in days after collection of blood (Figure 1).
Platelet count, 103/mL 14–850 The mean percent change in the WBC was less than 1
MPV, mm3 6.2–10.4 for the first 3 days but fluctuated between 25.3 and 16.3
Differential from days 4 to 7. The 95% CIs ranged between 212.2%
Neutrophils, % 23.2–94.9 and 113.8% for the first 3 days and between 230% and
Lymphocytes, % 1.5–69.2 142% from days 4 to 7.
Monocytes, % 0.1–29.0 The RBC, hemoglobin, and MCH were relatively stable,
Eosinophils, % 0.0–9.3
Basophils, % 0.0–1.7
each with a mean percent change of equal to or less than
Neutrophils, No. 0.1–62.6 1.1 and a combined 95% CI in the range of 23.7% to
Lymphocytes, No. 0.1–6.2 15.1%.
Monocytes, No. 0.0–2.1 The MCV, hematocrit, and RDW each increased over
Eosinophils, No. 0.0–1.5 time, revealing maximum mean percent changes on day 7
Basophils, No. 0.0–1.1 of 10.4, 10.3, and 15.8, respectively. The respective 95%
* CBC indicates complete blood cell count; WBC, white blood cell CIs for MCV, hematocrit, and RDW ranged from 21.8%
count; RBC, red blood cell count; MCV, mean corpuscular volume; to 114.8%, 22.8% to 116.5%, and 25.4 % to 125.7%
MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular he-
moglobin concentration; RDW, red blood cell distribution width; and
throughout the 7-day period.
MPV, mean platelet volume. The MCHC trended downward over time, with a max-
imum mean percent change of 29.4 on day 7 and 95% CIs
of 213.9% to 13.2% for the 7-day period.
(MCV), mean corpuscular hemoglobin (MCH), mean corpuscular The mean percent change in the platelet count fluctu-
hemoglobin concentration (MCHC), red cell distribution width ated between 20.9 and 14.3 for the first 4 days and
(RDW), platelet count, and mean platelet volume (MPV). The 5- ranged from 16.1 to 16.6 for the subsequent 3 days. Sim-
part differential consisted of relative percentages and absolute ilarly, the 95% CIs ranged from 223.8% to 130.2% for the
numbers of neutrophils, lymphocytes, monocytes, eosinophils, initial 4 days and from 242.5% to 154.8% for the subse-
and basophils.
quent 3 days. The platelet count of 2 of the 40 specimens
The CBC and differential results were tabulated using a per-
sonal computer with Excel software. Percent change from the day dropped progressively (up to 37% of the value on day 0)
0 value for each parameter for each subsequent day of the study from day 3 onward, and on blood smear review was found
period was calculated. For percentages and numbers of eosino- to be associated with platelet clumping. Only a slight in-
phils, basophils, and monocytes, absolute difference rather than crease in mean percent change and either no change or
percent change from the value on day 0 was calculated because minimal decrease in SD was noted on each day when the

Table 2. Mean Percent Changes in CBC Parameters Induced by Storage of Blood at Room Temperature*
Day 0
(n 5 Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7
Parameter 40) (n 5 40†) (n 5 40) (n 5 37) (n 5 15‡) (n 5 24) (n 5 16) (n 5 12)
WBC 0.0 20.9 (5.5) 20.1 (5.7) 0.8 (6.5) 25.3 (5.6) 4.6 (17.3) 6.3 (17.8) 1.5 (9.9)
RBC 0.00 20.34 (1.19) 20.61 (1.08) 20.11 (1.61) 0.14 (1.93) 0.13 (1.04) 0.11 (1.11) 0.07 (1.26)
Hemoglobin 0.0 20.4 (1.2) 20.7 (1.0) 20.6 (0.9) 1.0 (1.2) 20.3 (0.9) 0.2 (0.8) 0.2 (0.9)
Hematocrit 0.0 1.1 (2.0) 3.4 (2.2) 6.0 (2.5) 8.4 (2.7) 8.6 (2.5) 9.9 (2.7) 10.3 (3.1)
MCV 0.0 1.5 (1.7) 4.0 (2.1) 6.2 (2.4) 8.3 (2.2) 8.5 (2.1) 9.6 (2.2) 10.4 (2.2)
RDW 0.0 4.3 (3.2) 8.0 (6.6) 9.5 (7.4) 7.9 (4.8) 12.1 (5.2) 14.2 (4.1) 15.8 (5.0)
MCH 0.0 0.0 (1.5) 0.0 (1.3) 20.4 (1.7) 1.1 (2.0) 20.4 (1.0) 0.1 (1.0) 0.0 (1.1)
MCHC 0.0 21.5 (2.3) 23.9 (2.1) 26.1 (2.3) 26.6 (2.2) 28.2 (2.2) 28.8 (2.2) 29.4 (2.3)
Platelet count 0.0 20.2 (9.0) 4.3 (12.9) 2.6 (13.2) 20.9 (10.5) 6.6 (23.6) 6.3 (21.1) 6.1 (24.3)
PLTa 0.0 20.1 (9.2) 4.6 (13.2) 3.5 (13.0) 0.3 (9.7) 10.0 (21.4) 9.3 (18.3) 10.1 (21.1)
MPV 0.0 5.4 (6.9) 10.8 (10.6) 11.1 (11.0) 12.0 (12.0) 12.5 (12.3) 6.3 (6.3) 8.2 (8.2)
* Data are presented as mean (SD). CBC indicates complete blood cell count; WBC, white blood cell count; RBC, red blood cell count; MCV,
mean corpuscular volume; RDW, red blood cell distribution width; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin
concentration; PLTa , platelet data after excluding 2 specimens that revealed a drop in platelet count associated with platelet clumping; and MPV,
mean platelet volume.
† Number of specimens on day 1 was 39 for platelet count and 38 for MPV because for unknown reasons the analyzer voted out the platelet
count on 1 specimen and did not generate the MPV result on 2 specimens.
‡ Total number of specimens on day 4 is lower than that on day 5 because a run on the Coulter Gen.S was inadvertently missed for a batch of
specimens on day 4.

Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al 337
Figure 1. Mean (62 SDs) changes observed
in complete blood cell count result on storage
of blood at room temperature. Mean percent
change from the value on day 0 and the 95%
confidence intervals are plotted against time
in days after collection of specimen. The
number of specimens is indicated on top of
the 95% confidence interval bars for each
day. The number of specimens on day 1 was
39 for platelet count (PLT) and 38 for mean
platelet volume (MPV) because for unknown
reasons the analyzer voted out the PLT on 1
specimen and did not generate the MPV re-
sult on 2 specimens. WBC indicates white
blood cell count; RBC, red blood cell count;
MCV, mean corpuscular volume; RDW, red
blood cell distribution width; MCH, mean
corpuscular hemoglobin; MCHC, mean cor-
puscular hemoglobin concentration; and
PLTa, platelet data after excluding 2 speci-
mens that revealed a drop in platelet count
associated with platelet clumping.

platelet count data of these 2 specimens were excluded with the changes observed for specimens with normal or
from final calculations (PLTa in Table 2 and Figure 1). high initial respective counts (Table 3). Furthermore, the
The MPV initially increased steadily, reaching a maxi- platelet counts of thrombocytopenic specimens tended to
mum mean percent change of 112.5 on day 5. On days 6 increase over time, whereas the opposite was true for
and 7, the mean percent changes were 16.3 and 18.2, specimens with high initial count (Figure 2, B). Similarly,
respectively. The 95% CIs for MPV ranged between the WBC of leukopenic specimens tended to increase from
212.0% to 137.1% throughout the 7-day period. day 5 onward, whereas the opposite was true, although
When changes observed in the CBC parameters were minimally so, from day 4 onward for specimens with high
categorized based on whether the initial (day 0) value was initial counts (Figure 2, A). Changes observed in the re-
low (below the lowest normal value), normal (within the maining CBC parameters (data not shown) did not show
normal range), or high (above the highest normal value), any particular relationship to the levels of initial values.
the number of specimens analyzed per day decreased in Changes observed in the automated differential results
each group to between 4 and 16, but certain trends became of blood specimens stored at room temperature are shown
evident among the groups. Leukopenic and thrombocy- in Table 4 and Figure 3.
topenic specimens revealed the greatest day-to-day mean A minimal initial decrease in the percentage of neutro-
percent changes and the widest day-to-day SDs compared phils on days 1 and 2 and in the number of neutrophils
338 Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al
Table 3. Subcategorization of Changes in WBC and Platelet Count Induced by Storage of Blood at Room Temperature*
Parameter Day 0 Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7
WBC
Low (0.3–3.8 3 103/mL)
Mean (SD) 0.0 22.7 (9.7) 0.9 (11.0) 3.7 (11.8) 26.4 (7.9) 12.7 (29.1) 26.0 (28.6) ND
No. of specimens 11 11 11 10 6 8 4 2
Normal (5.5–10.6 3 103/mL)
Mean (SD) 0.0 20.2 (2.2) 20.3 (1.9) 20.9 (1.9) 25.5 (4.8) 0.9 (4.2) 1.1 (6.2) 0.4 (7.9)
No. of specimens 13 13 13 13 5 8 6 6
High (11.5–67.9 3 103/mL)
Mean (SD) 0.0 20.2 (2.8) 20.6 (2.6) 0.2 (2.9) 23.2 (1.4) 0.0 (2.0) 21.5 (2.9) 23.9 (4.8)
No. of specimens 16 16 16 14 4 8 6 4
Platelet count
Low (14–135 3 103/mL)
Mean (SD) 0.0 1.8 (14.3) 10.4 (17.8) 9.8 (17.7) 8.2 (15.2) 27.9 (21.4) 21.4 (15.4) 27.5 (13.4)
No. of specimens 15 14 15 13 4 9 7 5
Normal (152–371 3 103/mL)
Mean (SD) 0.0 0.4 (2.8) 3.0 (6.2) 20.1 (9.2) 24.3 (8.1) 26.8 (15.7) 26.7 (17.4) 27.9 (19.2)
No. of specimens 14 14 14 13 6 11 7 6
Normala (152–371 3 103/mL)
Mean (SD) 0.0 20.2 (2.6) 3.9 (6.3) 2.4 (7.7) 21.5 (5.0) 21.3 (11.0) 21.8 (12.9) 22.1 (14.3)
No. of specimens 12 12 12 11 5 9 6 5
High (403–850 3 103/mL)
Mean (SD) 0.0 22.3 (4.7) 22.4 (7.4) 22.8 (6.3) 24.1 (4.5) 24.7 (5.9) ND ND
No. of specimens 11 11 11 11 5 4 1 1
* WBC indicates white blood cell count; Normala, normal range platelet data after excluding 2 specimens that revealed a drop in platelet count
associated with platelet clumping; and ND, not determined due to inadequate number of specimens.

on day 1 was followed by a variable degree of increase in solute difference from the value on day 0 for the number
both on subsequent days. A maximum mean percent of basophils remained unchanged at 0 throughout the
change in percentage of neutrophils of 121.4 was noted study. The respective 95% CIs for the mean difference in
on day 4 and in number of neutrophils of 124.5 on day the percentage and number of basophils ranged from
6. The 95% CIs during the 7-day period ranged from 21.4 to 11.5 and 20.4 to 10.3.
269.9% to 190.8% for the percentage of neutrophils and
from 2119.2% to 1168.1% for the number of neutrophils. COMMENT
The widest 95% CI was noted on day 6 for both the per- Storage of blood at room temperature for up to 7 days
centage and number of neutrophils. caused changes in CBC and automated differential results.
The percentage and number of lymphocytes tended to However, the changes observed in some CBC parameters
increase over time, and the mean percent change fluctu- may be considered clinically insignificant.
ated between 129.1 and 197.8 for the former and between The maximum change of 4% in RBC, 3.4% in hemoglo-
13.3 and 181.3 for the latter. The respective CIs for the bin, and 5.1% in MCH carries little, if any, clinical signif-
percentage and number of lymphocytes throughout the 7- icance. Moreover, the narrow 95% CIs for each of these 3
day period ranged from 2782.3% to 1978% and 2207.5% parameters clearly demonstrate their stability at room
to 1355.2%. temperature for at least up to 7 days. The MCV, and con-
The percentage and number of monocytes tended to de- sequently the hematocrit and RDW, increased over time,
crease over time. The mean percent change fluctuated be- at an apparently faster rate during the first 4 days when
tween 285.4 and 10.4 for the former, and the mean ab- compared with that during the subsequent 3 days. The
solute difference fluctuated between 20.6 and 20.1 for the increase in MCV reflects red blood cell swelling at room
latter. The respective 95% CIs for the percentage and num- temperature. The degree of change in MCV, hematocrit,
ber of monocytes throughout the 7-day period ranged and RDW observed from day 2 onward may be considered
from 2406.7% to 1407.5% and 22.9 to 12.8, respectively. less than desirable, particularly when the results are bor-
The mean absolute difference from the value on day 0 derline normal or abnormal. A continuous increase in he-
for the percentage and number of eosinophils increased matocrit with a relatively stable hemoglobin level led to a
steadily with time, with a maximum difference of 17.9 for steady decline in MCHC over time. A potential change of
the former and 10.6 for the latter, both on day 7. The 5% or greater in MCHC, as noted from day 2 onward, is
respective 95% CIs for the percentage and number of eo- also less than desirable. In general, the WBC appeared
sinophils ranged from 225.6 to 140.4 and 21.1 to 12.0. stable for the first 3 days. However, the degree of change
Basophils represented the smallest fraction (0.0% to in the WBC of specimens with a normal or high count
1.7%) of all WBC and revealed minimal or no change with observed throughout the study may be considered ac-
time in both the percentage and number. The maximum ceptable for clinical utility. In contrast, the changes in the
mean absolute difference from the value on day 0 was 0.2 WBC of leukopenic specimens, as observed from day 5
on day 4 for the percentage of basophils. The mean ab- onward in particular, are less than desirable. Like the
Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al 339
Figure 2. Subcategorization of changes ob-
served in (A) white blood cell count (WBC)
and (B) platelet count (PLT) on storage of
specimens at room temperature, based on the
initial (day 0) result. Mean (62 SDs) percent
change from the value on day 0 and the 95%
confidence intervals are plotted against time
in days after collection of specimen. The
number of specimens is indicated on top of
the 95% confidence interval bars for each
day. Data for low WBC on day 7 and for high
PLT on days 6 and 7 were excluded due to
inadequate number of specimens. PLT (Nor-
mala) indicates normal range platelet data af-
ter excluding 2 specimens that revealed a
drop in platelet count associated with platelet
clumping.

Table 4. Percent Changes or Differences in Automated Differential Results Induced by Storage of Blood at Room
Temperature*
Day 0 Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7
Parameter (n 5 40) (n 5 40) (n 5 39) (n 5 37) (n 5 15†) (n 5 24) (n 5 16) (n 5 12)
Neutrophils, % 0.0 21.1 (12.6) 20.6 (26.8) 5.6 (32.9) 21.4 (33.4) 12.3 (39.2) 10.5 (40.2) 1.1 (24.3)
Lymphocytes, % 0.0 34.8 (150.6) 39.3 (121.5) 97.8 (440.1) 62.2 (251.3) 29.1 (93.9) 65.4 (137.2) 73.1 (104.9)
Monocytes, % 0.0 220.7 (51.4) 0.4 (203.6) 256.4 (47.7) 285.4 (14.6) 250.6 (164.7) 265.6 (73.4) 278.5 (23.8)
Eosinophils, %‡ 0.0 0.2 (1.1) 1.6 (8.3) 2.0 (12.9) 4.0 (13.7) 7.4 (16.5) 4.7 (7.3) 7.9 (9.7)
Basophils, %‡ 0.0 0.1 (0.7) 20.1 (0.5) 20.1 (0.5) 20.2 (0.3) 20.1 (0.2) 0.0 (0.3) 0.0 (0.4)
Neutrophils, No. 0.0 20.4 (11.2) 3.5 (21.3) 6.7 (33.3) 16.0 (31.4) 20.1 (55.7) 24.5 (71.8) 4.8 (40.2)
Lymphocytes, No. 0.0 12.6 (32.0) 38.3 (110.2) 27.6 (58.8) 3.3 (61.2) 45.2 (93.9) 73.9 (140.7) 81.3 (89.6)
Monocytes,
No. 3 103/mL‡ 0.0 20.2 (0.3) 20.1 (1.4) 20.5 (0.4) 20.5 (0.3) 20.6 (0.5) 20.6 (0.6) 20.6 (0.6)
Eosinophils,
No. 3 103/mL‡ 0.0 0.0 (0.2) 0.0 (0.3) 0.1 (0.9) 0.1 (0.3) 0.3 (0.7) 0.3 (0.4) 0.6 (0.7)
Basophils,
No. 3 103/mL‡ 0.0 0.0 (0.2) 0.0 (0.2) 0.0 (0.2) 0.0 (0.0) 0.0 (0.0) 0.0 (0.0) 0.0 (0.0)
* Data are presented as mean (SD).
† Total number of specimens on day 4 is lower than that on day 5 because a run on the Coulter Gen.S was inadvertently missed for a batch of
specimens on day 4.
‡ Absolute difference.

340 Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al
Figure 3. Changes observed in automated
differential results on storage of blood at
room temperature. Mean (62 SDs) percent
change or difference, as indicated, from the
value on day 0 and the 95% confidence in-
tervals are plotted against time in days after
collection of specimen. The number of spec-
imens is indicated on top of the 95% confi-
dence interval bars for each day.

WBC, the platelet count in general appeared stable for the storage of blood at room temperature agree with those of
first 4 days. However, the degree of change in the platelet Cohle and associates5 published 2 decades ago. This agree-
count observed throughout the study in specimens with ment is noted despite the difference in the total number
the platelet count within or above the normal range may (12 to 40 vs 10) and types of specimens (normal plus ab-
be considered acceptable for clinical utility. A less than normal vs only normal) used in the 2 studies. The auto-
desirable degree of change was observed only in throm- mated analyzer used by Cohle et al5 to obtain CBC results
bocytopenic specimens, particularly from day 5 onward. was Coulter S-Plus, also an impedance technology–based
The MPV behaved like the MCV for the initial 5 days, cell counter. These findings suggest that clinically usable
showing a continuous increase, with a relatively faster rate results may be obtained for some CBC parameters from
of change during the first 2 days when compared with blood specimens stored at room temperature for up to 7
that during the subsequent 3 days. A mean percent change days. Specifically, the RBC, hemoglobin, and MCH were
in the MPV of 10 or greater, as noted from day 2 onward, stable for the maximum study period of 7 days. The WBC
is also less than desirable. and platelet count were fairly stable for 4 days or even
Our observations of changes in CBC results induced by longer (up to 7 days) if the counts were within or above
Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al 341
the normal range. Relatively wide fluctuations (large SDs) after 1 day of storage at room temperature are also cor-
in the WBC and platelet count of specimens with counts roborated by the published reports of Wood et al,6 who
below the normal range are understandable and attribut- analyzed 21 specimens from healthy volunteers and 113
able, at least partly, to lower degree of precision known to patient specimens using Cell Dyn-3500, which performs
be associated with low counts when compared with nor- differential analysis by measuring laser light scattering
mal or high counts by automated or manual methods. Ad- properties of cells at various angles.
ditionally, the small number of specimens from day 4 on- To the best of our knowledge, there are no other pub-
ward could have contributed to wider fluctuations noted lished reports in the English language pertaining to the
during the same period for specimens with counts below study of changes in the automated differential induced by
or within the normal range. A progressive increase in prolonged storage of blood at room temperature. Based
MPV, MCV, hematocrit, and RDW and a decrease in on the findings of this study and the cited published re-
MCHC from day 2 onward make the results of these pa- ports, it can be surmised that even after only 24 hours a
rameters suspect in terms of clinical utility, especially in specimen may yield unreliable differential results when
cases where the results of these parameters are borderline using the VCS technology. Storage of blood at 48C has
normal or abnormal. been shown to prolong the specimen stability by several
Storage of blood at room temperature caused changes days for many CBC parameters3,4,6,7 and by a couple of
in virtually all of the parameters of automated differential. days for some differential parameters.2,6 Refrigeration is
Beginning on day 1, 2, or 3, the relative percentages and not a solution, however, for specimens that arrive at the
absolute numbers of neutrophils, lymphocytes, and eosin- laboratory after being stored at room temperature for a
ophils tended to increase, whereas the opposite was true few days.
for the relative percentage and absolute number of mono- CONCLUSION
cytes. The increase in the percentage of neutrophils began Although it is preferable to perform a CBC and differ-
on day 3, number of neutrophils on day 2, percentage and ential on a blood specimen as soon as possible after col-
number of lymphocytes on day 1, percentage of eosino- lection, our study suggests that clinically reliable results
phils on day 1, and number of eosinophils on day 3. The may be obtained for some CBC parameters, but not for
percentage and number of monocytes trended downward the differential, from specimens older than 1 day, when
from day 1 onward. The relative percentage and absolute analyzed on the Coulter Gen.S. Specifically, hemoglobin,
number of basophils, although difficult to reliably evaluate RBC, and MCH are stable for at least 7 days, WBC for at
because they represented a very small fraction, if any, of least 3 days, and platelets for at least 4 days of storage at
total leukocytes, trended either downward or showed no room temperature. Based on these findings, laboratories
appreciable change, particularly in the absolute number using VCS technology for CBC and differential may de-
throughout the study period. These changes in white cide to accept specimens older than 24 hours for analysis
blood cell differential are most likely the sum effect of the and report only the results of the stable CBC parameters.
loss of individual cell characteristics specifically measured Under these circumstances, a comment such as ‘‘unable to
by the analyzer and the cellular degeneration that is obtain reliable results due to the age of the specimen’’ or
known to occur as the cell ages. ‘‘specimen too old to obtain reliable results’’ could replace
An increase in lymphocytes and a decrease in mono- the results of unstable parameters. Alternatively, one may
cytes after 24 and 48 hours of storage of blood at room choose to report the results of all CBC parameters, but
temperature have also been reported previously by Warn- with a comment such as ‘‘specimen pp days old, results
er and Reardon,2 who analyzed 25 specimens from may be affected’’ appended to the result(s) of one or more
healthy subjects using Coulter STKS, which, like the Coul- unstable parameters.
ter Gen.S, uses VCS technology for the differential count. We are grateful to the hematology laboratory staff for technical
Compared with an appreciable decrease in neutrophils help, Medical Media Graphics staff for preparing the illustrations,
noted at the same period by these investigators, we ob- and John O’Connor, PhD, of Department of Health Policy for
served only a small or rather negligible decrease in per- helpful hints in data analysis.
centage of neutrophils on days 1 and 2 and in number of References
neutrophils on day 1, beyond which both the percentage 1. Coulter Gen.S System Manuals: Operator’s Guide. Miami, Fla: Coulter Corp;
November 1996.
and number of neutrophils increased, although to a vari- 2. Warner BA, Reardon DM. A field evaluation of the Coulter-STKS. Am J Clin
able degree, for the remainder of the study period. Unlike Pathol. 1991;95:207–217.
our observations of an increase in the percentage and 3. Brittin GM, Brecher G, Johnson CA, Elashoff RM. Stability of blood in com-
monly used anticoagulants. Am J Clin Pathol. 1969;52:690–694.
number of eosinophils and a decrease in the percentage 4. Lampasso JA. Changes in hematologic values induced by storage of ethy-
and number of basophils, Warner and Reardon2 noted a lenediaminetetra-acetate human blood for varying periods of time. Am J Clin
decrease in the percentage of eosinophils and an increase Pathol. 1968;49:443–447.
5. Cohle SD, Saleem A, Makkaoui DE. Effects of storage of blood on stability
in the percentage of basophils after 24 and 48 hours of of hematologic parameters. Am J Clin Pathol. 1981;76:67–69.
storage at room temperature. Differences in the sample 6. Wood BL, Andrews J, Miller S, Sabath DE. Refrigerated storage improves
size and sample type between the 2 studies may at least the stability of the complete blood cell count and automated differential. He-
matopathology. 1999;112:687–695.
partly account for the different findings. The observations 7. Lawrence ACK, Bevington JM, Young M. Storage of blood and the mean
on changes in neutrophils, lymphocytes, and monocytes corpuscular volume. J Clin Pathol. 1975;28:345–349.

342 Arch Pathol Lab Med—Vol 126, March 2002 Automated CBC and Differential Result Changes—Gulati et al

You might also like