You are on page 1of 6

| |

Received: 25 November 2018    Revised: 2 December 2019    Accepted: 13 December 2019

DOI: 10.1111/ijlh.13151

LETTER TO THE EDITOR

Establishment of common reference intervals for hematology


parameters in adults, measured in a multicenter study on the
Sysmex XN-series analyzer

Dear Editors, Over a period of 12  weeks, venipunctures were performed by


Clinical laboratories are responsible for the reference intervals re- skilled laboratory technicians or nurses using standard K2-EDTA col-
ported on their laboratory reports, as specified by the International lection tubes (Sarstedt, Becton Dickinson, Terumo Europe, Greiner).
Organization for Standardization standard 15  189. Although re- Samples were stored in an upright position at room temperature be-
quired by the directive on in vitro diagnostic medical devices in the fore analysis and were analyzed within 2 hours.
European Union (98/79/EC), manufacturers’ information about the Complete blood count (CBC) including white blood cell (WBC)
reference intervals is often very limited, is frequently based on pre- differentiation and reticulocyte count were analyzed on an XN an-
vious generation analyzers and/or is not transferable to the patient alyzer by trained laboratory technicians. Platelet counts obtained
population that is served. This emphasizes the need for laboratories by the impedance method were used in this study. All XN analyzers
to establish their own local reference intervals. However, produc- were calibrated and maintained according to the manufacturer's
ing reference intervals is expensive and a heavy task for most labo- instructions, but were not adjusted to each another, thereby mim-
ratories. Therefore, the production of common reference intervals icking at most daily practice. IPU software versions 00-13 or 00-15
for homogeneous populations shared by all laboratories within one were used and every laboratory performed multilevel IQC accord-
country or region has attracted a lot of interest. In this study, we ing to local practice. As requested by Belgian government rule,
calculated multicentric reference intervals for hematology parame- all laboratories participated in the Belgian National EQA program
ters on the XN hematology analyzer (Sysmex). Although this class of (Sciensano) for CBC, WBC differentiation and reticulocyte count.
analyzers has already been launched in 2011, there are few studies Results of the EQA rounds (n = 2) that were sent during the study
to report reference intervals for the reportable parameters on this period were reviewed, with no Q-score results exceeding the max-
analyzer.1,2 To the best of our knowledge, no specific multicentric imum allowable deviation (as defined by Sciensano), with Q = (re-
reference intervals for the XN analyzer have been published so far. ported value  −  assigned value)/assigned value and the assigned
The direct, a priori sampling technique was used to select the value  =  median of all Belgian XN instruments that participated in
reference individuals. Clear exclusion criteria were defined in a ques- the EQA (n = 20 to 26).
tionnaire. Volunteers (20 to 60  years old) were selected based on All statistical analyses were performed using Medcalc software
their subjective healthy status, with female individuals being not version 15.6.1 (Medcalc, Mariakerke). The width of the 90% CI of
pregnant or breastfeeding. Biochemical tests to confirm the healthy the lower/upper limit was evaluated against the recommendation
status (CRP, liver enzymes…) were not performed. Volunteers had made in Clinical Laboratory Standards Institute (CLSI) document
not been hospitalized or seriously ill in the previous 3 months and EP28-A3C, 2010, that is being less than 0.2 times the width of the
did not use chronic medication (excluding oral contraceptives and reference interval.3
statins). Fasting prior to blood drawing was not required. Three hundred and one reference individuals were included in
Reference individuals from ten hospital laboratories (H. the study (female: 202; male: 99; median age: 41  years). After the
Hart Hospital Leuven, AZ Sint-Dimpna Hospital Geel, AZ Klina Tukey outlier rejection, the Kruskall-Wallis test was performed and
Brasschaat, AZ Zeno, AZ Delta Roeselare, AZ Groeninge Kortrijk, AZ revealed significant differences (P  <  .05) between some laborato-
Sint-Lucas Brugge, Imelda Hospital Bonheiden, AZ Alma Eeklo, and ries for a subset of parameters. However, exclusion of results from
Ghent University Hospital) using a XN hematology analyzer in daily these laboratories did not lead to statistically different reference in-
practice and operating in the narrow geographical area of Flanders, tervals (data not shown). Moreover, as age and gender distribution
Belgium (in a radius of maximum 75 km), were enrolled in this study. also differed among the cohorts recruited by the participating lab-
Protocol, informed consent form and participant information sheet oratories, no laboratories were excluded. Instead, imbalances in age
were approved by the central and participating hospital ethic com- and sex distributions were adjusted (for age  ±  4  years) by random
mittees (Belgian registration number B670201317912). Written ap- deletion of 102 female results. The characteristics of this cohort are
proval was obtained from all study participants. shown in Table S1. Due to quality issues for reticulocyte count in

Int J Lab Hematol. 2020;00:1–6. wileyonlinelibrary.com/journal/ijlh© 2020 John Wiley & Sons Ltd     1 |
|
2       LETTER TO THE EDITOR

one laboratory, data for reticulocytes from that laboratory were ex- values were found for 16 out of 24 parameters both for the fe-
cluded from reference value calculation. male and male cohort versus the total cohort (upper and/or lower
Testing for normality was done with the D’Agostino-Pearson reference value calculated for the gender-specific cohort was not
test for normal distribution. Since normality was not accepted within the 90% confidence limit of the corresponding reference
for all parameters, it was decided to calculate reference intervals value for all pooled data). Decisions whether reference intervals
with the nonparametric method, according to the CLSI document for men and women could be pooled were based on calculating
EP28-A3c. 3 95% nonparametric double-sided reference intervals z-scores, as described by Harris & Boyd. 3 Based on the calculated
(2.5%-97.5%) and corresponding 90% confidence intervals (CI) z-scores, gender-specific reference intervals should be reported
for the lower and upper limits were calculated for all pooled data for 10 parameters (ie, WBC, RBC, Hb, HCT, MCHC, PLT, NEU abs,
(males and females; n = 199) (Table 1). In addition, reference values RET abs, NEU%, and MONO%). This is in line with the results of
were calculated with gender stratification (Table 2). Calculations other recent studies.4,5
for the reference intervals for men (n = 99) and women (n = 100) According to the CLSI document EP28-A3c, transference check
separately were performed with the robust method, which is of the newly calculated reference intervals was performed using
recommended by the CLSI document EP28-A3c for sample sizes samples of presumably healthy individuals (24 men, 24 women with
smaller than 120. 3 The width of the 90% confidence intervals was exclusion criteria: not pregnant and no medication, except oral con-
lower in the gender-specific groups than in the pooled population. traceptives) obtained from an independent hospital not involved
The 0.2 width criterium was reached for most parameters and in the multicentric study (AZ Sint-Maarten, Mechelen, Flanders,
the threshold of 0.3 times the width of the reference interval was Belgium).3 Instead of the Tukey exclusion, the latent abnormal value
never exceeded. Statistically significant differences in reference exclusion (LAVE) methodology 6 was used to exclude volunteers that

TA B L E 1   Reference intervals for all


Test (n) Units Low 90% CI low High 90% CI high
pooled data (n = 199; both males and
WBC (191) 109/L 4.3 3.67-4.69 9.64 9.03-10.16 females) for 24 hematological parameters
RBC (199) 1012/L 3.93 3.73-4.07 5.62 5.51-5.73 on the XN-1000 automated blood cell
analyzer, established according to CLSI
HGB (199) g/L 123 118-123 168 167-173
guideline EP28-A3
HCT (199) L/L 0.362 0.353-0.372 0.508 0.487-0.520
MCV (197) fL 83.9 83.1-84.5 98.0 96.6-98.4
MCH (197) pg 27.7 27.5-28.0 32.8 32.5-33.2
MCHC (194) g/L 317 313-318 354 350-356
RDW (186) % 11.5 11.3-11.7 13.9 13.7-14.3
9
PLT (194) 10 /L 163 150-174 347 323-357
NEU abs (193) 109/L 1.93 1.17-2.17 5.87 5.56-6.17
9
LYMPH abs 10 /L 1.23 1.07-1.30 3.42 3.23-3.57
(195)
MONO abs 109/L 0.26 0.21-0.31 0.78 0.75-0.82
(192)
EOS abs (190) 109/L 0.03 0.01-0.04 0.37 0.34-0.41
9
BASO abs (190) 10 /L 0.02 0.01-0.02 0.08 0.07-0.08
IG abs (189) 109/L 0.01 0.00-0.01 0.05 0.05-0.06
NEU % (196) % 39.2 37.1-41.8 71.5 69.1-73.0
LYMPH% (197) % 18.9 16.4-21.1 47.1 44.4-51.9
MONO% (187) % 4.8 3.8-5.3 11.5 11.2-11.9
EOS% (190) % 0.4 0.2-0.5 5.9 5.5-6.3
BASO% (195) % 0.2 0.2-0.3 1.4 1.2-1.5
IG% (192) % 0.0 0.0-0.1 0.8 0.7-0.9
RET % (174) % 0.64 0.45-0.73 2.0 1.8-2.2
RET abs (172) 109/L 30.0 21.4-34.4 93.0 87.9-96.7
RET-He (176) pg 28.3 27.8-29.6 36.1 35.5-37.2

Abbreviations: abs, absolute white blood cell differential count; CI low/high, 90% Confidence
Interval of the lower/upper reference limit; Low/High, lower/upper reference limit; n, number of
samples after the Tukey outlier exclusion.
LETTER TO THE EDITOR |
      3

had more than one parameter outside the 90% limits of the gen- with previously published data 8 and may reflect the difference in
der-specific reference interval, with the outlying results not being Hb levels between males and females and respective erythroid re-
derived from the same measurement (as is the case for calculated generation rates.
parameters). This resulted in exclusion of three females and three This study showed some limitations, such as the limited number
males. For acceptance of results, no more than 10% of the obtained of volunteers. However, as the width of the 90% CI of the lower/
values for the different parameters may fall out of the 95% reference upper limit resulted in a tolerable uncertainty in the gender-spe-
intervals. This criterion was met for all parameters for gender-spe- cific reference intervals, we believe adding more volunteers would
cific as well as pooled reference values. not have improved our reference intervals in a significant way.
Our calculated reference intervals for the XN analyzer often Our calculated reference limits are based on an aged-matched and
differed in a statistically significant way from the reference val- gender-matched pooled dataset of all laboratories. This was done
ues calculated by Park et al1 and Arbiol-Roca et al2 for the XN to achieve a maximal number of participants in every subgroup.
analyzer and the reference values proposed by Sysmex7 (Table 3). We worked with volunteers that were feeling subjectively healthy,
However, these differences were not considered clinically signifi- without an objective measure of their health status. This might be a
cant. For reticulocyte count however, we found higher results for limitation although Wu et al showed in an extensive study popula-
XN compared to XE. Moreover, the reticulocyte reference range tion the association between self-rated health and objective health
was shown to be lower in females than males, which is consistent status, exemplified by the prevalence of diseases, abnormalities in

TA B L E 2   Reference intervals for 24 hematological parameters on the XN-1000 automated blood cell analyzer, calculated separately for
males (n = 99), and females (n = 100), established with the CLSI robust method

Male Female

Test Units n Low 90% CI low High 90% CI high n Low 90% CI low High 90% CI high

WBC 109/L 94 3.64* 3.29-4.01 8.46* 8.04-8.82 96 3.80 3.35-4.19 9.39 8.94-9.84
12
RBC 10 /L 99 4.36* 4.26-4.47 5.78* 5.68-5.87 99 3.81 3.73-3.91 5.13* 5.03-5.22
HGB g/L 97 135* 133-138 170 167-172 99 117* 115-120 151* 149-154
HCT L/L 99 0.399* 0.391-0.407 0.510 0.502-0.518 100 0.354 0.347-0.362 0.461* 0.454-0.468
MCV fL 97 83.7 82.8-84.6 95.8* 94.9-96.6 100 83.3 82.3-84.3 98.5* 97.4-99.6
MCH pg 98 27.8 27.5-28.2 32.6 32.3-32.9 100 27.1* 26.7-27.5 33.1 32.7-33.5
MCHC g/L 97 316 313-319 355 352-358 98 313 310-315 348* 345-350
RDW % 90 11.4 11.3-11.6 13.6* 13.4-13.7 99 11.3 11.2-11.5 14.0 13.8-14.2
PLT 109/L 99 143* 130-157 325 312-338 99 175* 163-186 343 329-356
9
NEU abs 10 /L 94 1.53 1.30-1.77 4.90* 4.62-5.18 95 1.66 1.36-1.97 5.92 5.60-6.23
LYMPH 109/L 97 0.87* 0.73-1.02 3.12* 2.91-3.31 96 0.94* 0.83-1.11 3.25 3.08-3.48
abs
MONO 109/L 97 0.23 0.19-0.27 0.82 0.77-0.87 97 0.25 0.22-0.28 0.74* 0.70-0.77
abs
EOS abs 109/L 93 0.00 0.00-0.00 0.33* 0.30-0.37 95 0.00* 0.00-0.00 0.30* 0.26-0.33
BASO abs 109/L 98 0.00* 0.00-0.00 0.08 0.075-0.088 96 0.01 0.00-0.01 0.07 0.07-0.08
9
IG abs 10 /L 94 0.01 n.a. 0.05 n.a. 97 0.00 n.a. 0.06 n.a.
NEU% % 99 38.3 36.2-40.4 69.0* 66.7-71.2 97 41.6 39.2-44.1 73.7* 71.4-75.8
LYMPH% % 99 19.7 17.8-21.6 47.1 45.2-48.9 97 17.1 15.0-19.2 46.4 44.0-48.7
MONO% % 96 5.2 4.7-5.6 11.7 11.1-12.3 95 4.2 3.7-4.7 10.3* 9.8-10.7
EOS% % 93 0.0* 0.0-0.0 5.1* 4.6-5.7 98 0.0* 0.0-0.0 5.2* 4.5-5.7
BASO% % 96 0.0* 0.0-0.2 1.3 1.2-1.4 97 0.1* 0.0-0.2 1.1* 1.0-1.2
IG% % 94 0.0 n.a. 0.7 n.a. 98 0.0 n.a. 0.7 n.a.
RET% % 83 0.52 0.41-0.62 1.86 1.75-1.97 90 0.45 0.36-0.55 1.78* 1.66-1.90
RET abs 10 9/L 83 23.4 18.3-28.8 95.3* 88.5-101.5 91 16.0* 11.7-21.4 82.0* 75.8-89.0
RET-HE pg 82 30.1* 29.6-30.5 36.1 35.6-36.5 93 28.3 27.8-29.0 36.6 35.9-37.1

Abbreviations: abs, absolute white blood cell differential count; CI low/high, 90% Confidence Interval of the lower/upper reference limit; Low/High,
lower/upper reference limit; n.a., not applicable; n, number of samples after the Tukey outlier exclusion.
*Statistically significant difference with reference interval from all pooled data.
|
4       LETTER TO THE EDITOR

TA B L E 3   Overview of calculated reference intervals and comparison with published reference intervals for other Sysmex hematology
analyzers

Arbiol- Sysmex
Pekelharing et Park et al Roca et al cooperation
Parameter Units Low CI Low High CI High al (XE-series) (XN-series) (XN-series) (XN-series)

WBC 109/L 4.3 3.67-4.69 9.64 9.03-10.16 M: 3.91-10.90 3.89-9.23 3.9-9.5 3.37*-8.38*
F: 4.49-12.68
RBC (Male) 1012/L 4.36 4.26-4.47 5.78 5.68-5.87 4.44-5.61 4.24*-5.65* 4.3-5.6* 4.29-5.70
RBC (Female) 1012/L 3.81 3.73-3.91 5.13 5.03-5.22 3.92-5.08 3.66*-4.76* 3.9-5.1 3.72*-5.06
HGB (Male) g/L 135* 133-138 170 167-172 135-169 131.9*-168.7 137.4-164.7* 133-166*
HGB (Female) g/L 117 115-120 151 149-154 119-146 118.5-142.0* 120.0-146.8* 110*-147*
HCT (Male) L/L 0.399* 0.391-0.407 0.510 0.502-0.518 0.400-0.494 0.3906-0.5127 0.40-0.50* 0.413*-0.521*
HCT (Female) L/L 0.354 0.347-0.362 0.461 0.454-0.468 0.366-0.440 0.3426*-0.4345* 0.36-0.45* 0.352-0.467
MCV fL 83.9 83.1-84.5 98.0 96.6-98.4 M: 81.8-95.5 86.50*-101.79* 83.6-97.0 86.7*-102.3*
F: 82.9-98.0
MCH pg 27.7 27.5-28.0 32.8 32.5-33.2 27.0-32.3 27.23*-33.60* 27*-32 27.1*-32.4*
MCHC g/L 317 313-318 354 350-356 M: 324-350 M: 314.5-347.4* 314-349a 297*-331*
F: 318-347 F: 305.9*-337.6*
RDW % 11.5 11.3-11.7 13.9 13.7-14.3 M: 12.0-13.6 11.40-15.29* 11.6-14.3 12.2*-14.8*
F: 12.1-14.3
PLT 109/L 163 150-174 347 323-357 M: 166-308 M: 146.6*-420.8* M: 149*-303* 172-378*
F: 173-390 F: 157.8-389.0* F: 153-368*
NEU abs 109/L 1.93 1.17-2.17 5.87 5.56-6.17 M: 1.8-6.98 1.78-6.04 1.5-5.7 1.49-4.67*
F: 2.1-8.89
LYMPH abs 109/L 1.23 1.07-1.30 3.42 3.23-3.57 1.26-3.35 1.24-3.05* 1.3-3.4 1.05*-2.88*
MONO abs 109/L 0.26 0.21-0.31 0.78 0.75-0.82 M: 0.29-0.95 M: 0.29-0.72* 0.31-0.92* 0.21-0.61*
F: 0.25-0.84 F: 0.24-0.72*
EOS abs 109/L 0.03 0.01-0.04 0.37 0.34-0.41 M: 0.03-0.59 M: 0.04-0.58* 0.03-0.39 0.03-0.28*
F: 0.01-0.40 F: 0.01-0.59*
BASO abs 109/L 0.02 0.01-0.02 0.08 0.07-0.08 0.01-0.07 0.01-0.09* 0.01-0.09* 0.02-0.07
IG abs 109/L 0.01 0.00-0.01 0.05 0.05-0.06 0.00-0.06 0.00-0.04* NR 0.00-0.00*
NEU% % 39.2 37.1-41.8 71.5 69.1-73.0 M: 41.0-70.7 40.8-70.4 37.1-68.4* 39.8-70.5
F: 41.9-74.3
LYM% % 18.9 16.4-21.1 47.1 44.4-51.9 M: 19.1-47.9 20.1-46.8 21-50 23.1*-49.9
F: 18.3-45.7
MON% % 4.8 3.8-5.3 11.5 11.2-11.9 M: 5.2-15.2 M: 4.91-10.38* 5.1-11.2 4.3-10.0*
F: 4.2-11.8 F: 4.83-11.15*
EOS% % 0.4 0.2-0.5 5.9 5.5-6.3 M: 0.6-7.6 M: 0.73*-8.86* 0.4-6.6* 0.6*-5.4*
F: 0.2-5.3 F: 0.24-10.24*
BASO% % 0.2 0.2-0.3 1.4 1.2-1.5 M: 0.1-1.2 0.2-1.5 0.2-1.3 0.3-1.4
F: 0.1-1.0
IG% % 0.0 0.0-0.1 0.8 0.7-0.9   0.00-0.50* NR 0.16*-0.61
RET% % 0.64 0.45-0.73 2.0 1.8-2.2 0.43-1.36 0.99*-2.11a 0.68-1.86 0.82*-2.25*
RET abs 109/L 30.0 21.4-34.4 93.0 87.9-96.7 M: 23.0-70.1 M: 40*-120*a 34-102* 36.3*-105.7*
F: 17.0-63.8 F: 40-110*a
RET-HE pg 28.3 27.8 −29.6 36.1 35.5-37.2 32.1-38.8 M: 31,61*-36,28* NR 30.3*-36.0
F: 28,36-35,68

Abbreviations: Age range, 41-60 y; CI low/high, Confidence interval of the lower/upper reference limit; F, female; Low/ High, lower/ upper reference
limit; M, male; NR, not reported.
a
Corrected value based on personal communication with author.
*Statistically significant difference with reference interval calculated in this manuscript.
LETTER TO THE EDITOR |
      5

4
laboratory test results (including some hematology parameters), and Department of Laboratory Medicine, Imelda Hospital,
9
health-related factors. Bonheiden, Belgium
5
This study focused on the XN hematology analyzer, the successor Department of Laboratory Medicine, Heilig Hart Ziekenhuis
of the XE analyzer series. Schoorl et al conducted a multicenter verifi- Leuven, Leuven, Belgium
6
cation study of the XN analyzer compared to the XE-2100 analyzer and Department of Laboratory Medicine, Algemeen Ziekenhuis
found an acceptable correlation between both analyzers.10 Recently, a Delta, Roeselare, Belgium
7
sigma-metric analysis was carried out using the data from the Schoorl Department of Laboratory Medicine, Algemeen Ziekenhuis Sint-
publication.11 They concluded that the high bias between the XN and Lucas, Brugge, Belgium
8
XE analyzer for some parameters exceeded the total allowable error, Department of Laboratory Medicine, Algemeen Ziekenhuis
leading to negative Sigma-metrics. In our study, this discrepancy be- Klina, Brasschaat, Belgium
9
tween both analyzers was most accentuated for the reference range Department of Laboratory Medicine, Algemeen Ziekenhuis
of the reticulocyte count (Table 3). Laboratories should be aware of Groeninge, Kortrijk, Belgium
10
the differences between both generation analyzers. Department of Laboratory Medicine, Algemeen Ziekenhuis
In conclusion, our study provides reference intervals that could Zeno, Knokke-Heist, Belgium
11
serve as a guide for clinical laboratories that use the Sysmex XN an- Department of Laboratory Medicine, Algemeen Ziekenhuis
alyzers and serve the same population as the one included in our Alma, Eeklo, Belgium
12
study. Department of Laboratory Medicine, Algemeen Ziekenhuis
Sint-Dimpna, Geel, Belgium
13
AC K N OW L E D G E M E N T Department of Diagnostic Sciences, Ghent University, Ghent,
The authors wish to thank Liesbeth Seaux for the analysis of samples Belgium
and the practical help during this study. The authors Matthijs Oyaert,
Veronique Stove, Marleen Vandevenne, Jan Van den Bossche and Correspondence
Lisa Florin provided samples from healthy volunteers from 3 different Veronique Stove, Department of Laboratory Medicine,
hospitals and were responsible for data analysis. The authors Marleen Ghent University Hospital, Corneel Heymanslaan 10, 9000
Vanden Driessche, Rowan Claeys, Ludo Marcelis, Johan Robbrecht, Gent, Belgium.
Nadia Jacobs, An Nijs, Karen Janssen, Annemie Van Ruymbeke and Email: veronique.stove@ugent.be
Guy de Mûelenaere contributed to this study by providing additional
datasets of samples from healthy volunteers from 8 other hospitals. ORCID
Lisa Florin  https://orcid.org/0000-0002-1374-4936
C O N FL I C T O F I N T E R E S T Matthijs Oyaert  https://orcid.org/0000-0002-0240-0679
The authors have no competing interests. Veronique Stove  https://orcid.org/0000-0002-2447-971X

Lisa Florin1 REFERENCES
1
Matthijs Oyaert 1. Park SH, Park CJ, Lee BR, et al. Establishment of age-and gen-
Marleen Vandevenne2 der-specific reference ranges for 36 routine and 57 cell population
data items in a new automated blood cell analyzer, Sysmex XN-
Jan Van den Bossche3
2000. Ann Lab Med. 2016;36(3):244-249.
Marleen Vanden Driessche 4 2. Arbiol-Roca A, Imperiali CE, Montserrat MM, et al. Reference inter-
Rowan Claeys5 vals for a complete blood count on an automated haematology anal-
Ludo Marcelis6 yser Sysmex XN in healthy adults from the southern metropolitan
area of Barcelona. Ejifcc. 2018;29(1):48-54.
Johan Robbrecht7
3. Clinical Laboratory Standards institute. CLSI EP28-A3c Defining,
Nadia Jacobs8 establishing, and verifying reference intervals in the clinical labora-
An Nijs9 tory. Approved Guideline – Third Edition. In: 2010.
Karen Janssen10 4. Wakeman L, Al-Ismail S, Benton A, et al. Robust, routine hae-
matology reference ranges for healthy adults. Int J Lab Hematol.
Annemie Van Ruymbeke11
2007;29(4):279-283.
Guy de Mûelenaere12
5. Nordin G, Martensson A, Swolin B, et al. A multicentre study of ref-
Veronique Stove1,13 erence intervals for haemoglobin, basic blood cell counts and eryth-
rocyte indices in the adult population of the Nordic countries. Scand
1
Department of Laboratory Medicine, Ghent University J Clin Lab Invest. 2004;64(4):385-398.
6. Ichihara K. Statistical considerations for harmonization of the
Hospital, Ghent, Belgium
global multicenter study on reference values. Clin Chim Acta.
2
Department of Laboratory Medicine, Algemeen Ziekenhuis Sint- 2014;432:108-118.
Maarten, Mechelen, Belgium 7. Corporation S. Reference ranges analysis document for XN series.
3
Department of Laboratory Medicine, Antwerp University 2014; https​://www.sysmex.de/filea​dmin/media/​f101/Produ​k tfly​
er/Wisse​nszen​trum/XN-Series_Refer​ence_Ranges.pdf
Hospital, Antwerp, Belgium
|
6       LETTER TO THE EDITOR

8. Tarallo P, Humbert JC, Mahassen P, Fournier B, Henny J. 11. Sigma metric Analysis of a Sysmex XN. https​://www.westg​ard.com/
Reticulocytes: biological variations and reference limits. Eur J sigma-metric-sysmex-xn.htm. Accessed July 23, 2018.
Haematol. 1994;53(1):11-15.
9. Wu S, Wang R, Zhao Y, et al. The relationship between self-rated
health and objective health status: a population-based study. BMC S U P P O R T I N G I N FO R M AT I O N
Public Health. 2013;13:320. Additional supporting information may be found online in the
10. Schoorl M, Chevallier M, van der Ploeg T, van Pelt J. Multicenter verifi- Supporting Information section. 
cation of the Sysmex XN-Series. Int J Lab Hematol. 2017;39(5):489-496.

You might also like