You are on page 1of 8

Scandinavian Journal of Clinical & Laboratory Investigation, 2014; 74: 673–680

ORIGINAL ARTICLE

Automated white blood cell counts in cerebrospinal fluid using


the body fluid mode on the platform Sysmex XE-5000

AIHONG LI1, ELISABETH GRÖNLUND2 & GÖRAN BRATTSAND1


1Departmentof Medical Biosciences, Clinical Chemistry, and 2Department of Pathology, Umeå University Hospital,
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

Umeå University, Umeå, Sweden

Abstract
Background. The Sysmex XE-5000 offers automated quantification of red blood cells and white blood cells (WBCs) in
body fluids, with differentiation of polymorphonuclear cells (PMNs) and mononuclear cells (MNCs). Methods. We
evaluated automated WBC counting in cerebrospinal fluid (CSF) using the body fluid mode on the Sysmex XE-5000,
comparing it with flow cytometry as the reference method, and also with manual counting by microscopy. Experimental
analysis for linearity and limit of detection was performed by diluting isolated WBCs in cell-free CSF. To study the ability
to discriminate between PMNs and MNCs, samples were spiked using MNCs separated from peripheral blood. Com-
parison of WBC counts between a counting chamber and the XE-5000 was performed for 198 CSF samples. Results. In
the experimental set-up, within-run (CV 19%) and between-day imprecision (CV 15.3%) in quantitating total number of
WBC on XE-5000 was acceptable for WBC counts ⱖ 25 ⫻ 106/L. Compared with expected cell counts, mean bias was
For personal use only.

⫹ 2.6% for flow cytometry, ⫹ 5.5% for XE-5000 and ⫺ 73.2% for manual counting. Differentiation between PMNs and
MNCs was in concordance with flow cytometry. In comparisons of clinical CSF samples, overall agreement between the
XE-5000 and manual counting was observed in 81% of the samples, but mean difference in WBC differentiation was higher
for PMN (51.1 ⫻ 106/L) than for MNC (7.95 ⫻ 106/L). Conclusion. Despite limited precision at low WBC counts, XE-5000
could be a favourable alternative to the labour-intensive, time-consuming and less reliable manual counting and cuts
turnaround times in routine CSF-based diagnosis.

Key Words: Leukocyte count, neutrophils, monocyte, cerebrospinal fluid, automated exam

Introduction would delay the reporting of result. Other disadvan-


tages of microscopic evaluation are the high degree
In the majority of laboratories in Sweden, cells in
cerebrospinal fluid (CSF) have traditionally been of imprecision and the requirement for skilled
analyzed in Türk’s staining solution (Gentian violet/ personnel [3–9]. Since differential counting of
acetic acid) by manual microscopic counting in a WBCs in CSF into polymorphonuclear cells (PMNs)
Bürker chamber as described by Berggren Söderlund and mononuclear (MNCs) may aid in clinical
et al. [1]. The method provides simultaneous differ- decision-making regarding antibiotic therapy, it is
ential counting of white blood cells (WBCs) without important to report CSF-WBC counts together with
cytospin preparation. In contrast, document H56-A MNC and PMN counts to provide valuable diagnos-
of the Clinical and Laboratory Standards Institute tic information in various medical conditions.
(CLSI) recommends that erythrocytes and WBCs The XE-5000 offers a protocol for the quantifica-
should be counted unstained in a counting chamber, tion of red blood cells (RBCs) and WBCs in body
while morphological examination of nucleated cells fluid mode, and also differentiation between PMNs
is performed on air-dried and stained cytospins [2]. and MNCs. Moreover, there is no need for pretreat-
Morphological examination is no doubt better using ment of samples. Thus, the XE-5000 is an attractive
the CLSI procedure, but this tedious procedure method for automated analysis of WBCs, PMNs and
would be very difficult to use as a 24-h service, and MNCs in most body fluids. Only a few studies have

Correspondence: Aihong Li, M.D., Ph.D., Department of Medical Biosciences, Clinical Chemistry, Building 6M, ground floor, Umeå University Hospital,
Umeå University, SE-90185 Umeå, Sweden. Tel: ⫹ 46 (0) 90 7852507. Fax: ⫹ 46 (0) 90 778197. E-mail: aihong.li@medbio.umu.se

(Received 6 February 2014 ; accepted 26 June 2014 )

ISSN 0036-5513 print/ISSN 1502-7686 online © 2014 Informa Healthcare


DOI: 10.3109/00365513.2014.939994
674 A. Li et al.

evaluated the body fluid mode on the XE-5000 washed twice in phosphate-buffered saline (PBS)
[9–13]. Boer et al. were the first to demonstrate the and resuspended in PBS. The cell suspensions in
general good agreement between manual counting different experiments had WBC counts in the range
and XE-5000 [10]. However, high imprecision at low 2–6 ⫻ 109 /L as measured in the normal mode of the
WBC counts (⬍ 20 cells/mm3) was observed. There- XE-5000.
fore, a manual count of CSF samples with WBCs ⬎ 5 Normal CSF routine diagnostic samples (RBC
and ⬍ 20 per μL was recommended to increase diag- ⬍ 100 ⫻ 106/L, PMN ⬍ 1 ⫻ 106/L and MNC ⬍ 5 ⫻ 106
nostic accuracy [10]. De Jonge et al. also found a /L), with no turbidity, were selected for preparing
high imprecision with WBC counts ⬍ 10 cells per μL cell-free CSF. Supernatant after centrifugation
and poor discrimination of MNCs from PMNs [11]. (2000 g 5 min) was transferred into a new tube for
Paris et al. studied several different types of body each sample and kept at ⫺ 20°C until use. A pool
fluids and found a good correlation between auto- of cell-free CSF was prepared by mixing about
mated and manual counting methods for all types of 30 supernatants for further use of dilution.
fluids except CSF [12]. These studies suggested that Between May of 2009 and April of 2010, 198
XE-5000 can be suitable for automated quantifica- CSF samples were examined for routine diagnostic
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

tion of WBCs in a defined diagnostic setting. purposes by WBC counting using the XE-5000 and
However, a recent study by the German Society the manual counting method. At least 1 mL CSF was
for Clinical Chemistry and Laboratory Medicine collected in plain sterile tubes and the samples were
(DGKL) evaluated cell analysis on Sysmex XT-4000i requested to arrive in the laboratory within 30 min
and XE-5000 with 20 DGKL CSF controls [14]. after collection. After arrival in the laboratory, all
Conformity of WBC counts was obtained with automated and manual analyses were completed as
Fuchs-Rosenthal chamber as the reference. How- soon as possible preferably within 1 h. First, cell
ever, significantly increased PMN counts (2.2-fold counts were performed by the manual method as
on average) were demonstrated with Sysmex described below. Then, if enough material was left
XT-4000i and XE-5000 compared to flow cytome- over (130 μL), automated cell counts were performed
try, indicating that WBC differentiation into MNC in the open body fluid mode of a Sysmex XE-5000
For personal use only.

and PMN is unreliable [14]. Thus, the XE-5000 haematology analyzer.


body fluid system has not been recommended as a Results of WBC counts from patient CSF sam-
suitable alternative to manual differential cytologic ples were divided into three subgroups including a
workup for CSF [14]. Results of examination with range of normal clinical reference (0–5 ⫻ 106
the XE-5000 in 65 CSF samples with tumor cells WBCs/L), low WBC count (6–20 ⫻ 106 WBCs/L)
and 126 CSF samples without tumour cells sug- and high WBC count (⬎ 20 ⫻ 106 WBCs/L). Overall
gested that measurement of high-fluorescence cells agreement between manual counting and XE-5000
is not an appropriate diagnostic test for intrathecal was calculated with consideration of above reference
tumor cells [15]. The disparity between these previ- cut off level (WBC ⬎ 5 ⫻ 106 WBCs/L).
ous reports indicates that more requires to be known
regarding interpretation of results for routine usage. Microscopic analysis
In the present study, with the combined use of
manual counting and flow cytometry, which was Microscopic cell counting was performed by an
used as the reference method, we evaluated the experienced technician using a Bürker counting
performance of the body fluid mode on the Sysmex chamber (Paul Marienfeld GmbH & Co. KG Lauda-
XE-5000. Use of an experimental set-up and also königshofen, Germany) in the case of the perfor-
clinical samples showed that the XE-5000 could be mance experiments.
a favourable alternative for routine quantification of Cell counts in clinical samples were performed in
WBCs, PMNs and MNCs in CSF. clinical routine by technicians who have received
special training but had different degrees of experi-
ence. For each sample, 10 microscopic fields
Materials and methods (A-square) were examined with a ⫻ 40 objective lens.
50 μL CSF and 50 μL Türk’s solution (Merck,
Sample preparation
Darmstadt Germany) were mixed and 10 μL of the
WBCs were obtained using RBC lysis buffer from mixture was analyzed in a counting chamber. Ten
whole blood samples newly taken from healthy fields were evaluated for PMNs and MNCs. It often
donors. We randomly selected WBCs in the range of takes at least 10 min for one sample.
5–9 ⫻ 109 cells/L which consisted of approximately
60% PMNs and 40% lymphocytes/monocytes. One
Automated measuring on the XE-5000
millilitre whole blood was mixed gently for 5–10 min
with 50 mL lysis buffer (8.26 g NH4Cl, 1 g KHCO3, The Sysmex XE-5000 haematology analyser (Sysmex
and 33.6 mg EDTA in 1 L deionized water), followed Corporation, Kobe, Japan) incorporates a body fluid
by centrifugation at 400 g for 5 min. Cell pellets were mode that allows instant quantification of WBCs and
Automated WBC counts in CSF on XE-5000 675

other highly fluorescent cells in CSF, pleural fluid, on FACSCanto, XE-5000 and manual counting.
ascites, and synovial fluid without any treatment or Because of the short half- life of the leukocytes we
pretreatment.When switching from ‘full blood mode’, also tested a later time point at 3 h in the experimen-
the system performs background checks and a sepa- tal set-up to find whether there are differences of
rate washing step to avoid cross-contamination. It WBC counts at different levels between these two
takes approximately 195 s for one CSF sample on time points.
the XE-5000. At least 130 μL CSF was required for
analysis. All samples were assessed in the open mode WBC differentiation. Peripheral blood (PB) from a
in the system’s DIFF channel. healthy donor was used for the experiments.
WBCs were prepared as described above and diluted
at four expected WBC counts (10, 20, 100 and
Flow cytometric analysis 1000 ⫻ 106/L). Cell analysis was performed on the
Flow cytometric analysis was performed on a XE-5000, flow cytometry and by manual counting.
FACSCanto II Flow Cytometer (BD Biosciences, To further study differentiation between PMNs
San José, CA, USA). In this analysis, we used a and MNCs, mononuclear cells were separated by
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

special test tube: Trucount (BD Biosciences). These centrifugation on a density gradient (Lymphoprep,
tubes contain a known number of particles, which Pharmacia, Uppsala, Sweden) and washed twice
acts as a reference to determine the absolute number with PBS, and resuspended in PBS before dilution.
of residual leukocytes in blood components. The These cells were further diluted into a pool of
number of leukocytes was determined from a calcu- cell-free CSF. To diminish the effect of pH on cell
lation based on the number of particles and the stability, a sterile-filtered cell-free CSF pool (pH 7.2)
sample volume. One hundred microliters of CSF was was prepared by incubation at 37°C overnight in 5%
required for analysis. Differentiation between PMNs CO2. Four expected WBC concentrations (5, 10, 50
and MNCs on the FACSCalibur (BD Biosciences) and 500 ⫻ 106/L) were prepared for cell analysis
was analyzed using CD45/SSC gating according to using the XE-5000, flow cytometry and manual
counting.
For personal use only.

Kleine et al. [16].

Performance studies Statistical analysis

Limit of blank (LoB). The LoB was established in the Bias was calculated as [(mean obtained values ⫺
background analysis by measuring 20 different expected values)/expected values] ⫻ 100. For calcu-
samples of cell-free CSF, twice for each sample. lation of imprecision, the Excel 2003 was used for
data analysis. For linearity, Passing-Bablok regresson
Functional sensitivity. To evaluate within-run impreci- anlysis was used and data were plotted using the
sion, we diluted WBCs into cell-free CSF at the Method Validator (version 1.1). In comparison
expected counts of 5, 25 and 50 ⫻ 106 WBCs/L. between two methods in clinical CSF samples, mean
WBC counting was determined on the XE-5000 by difference was obtained using Method Validator
repeatedly measuring samples 20 times and the CVs (version 1.1). The Spearman correlation test was
at the expected counts were calculated. Manual used for correlation between two methods by SPSS
counting was also performed but samples could be (version 18).
only evaluated 3–7 times at each of the three levels
due to repeatedly continuous microscopying, the
examiner was unable to properly evaluate the subse- Results
quent samples. Limit of blank (LoB)
To test between-day accuracy, we performed
analysis by diluting the Sysmex control (e-CHECK) In the background analysis, 20 cell-free CSF samples
in saline solution at an expected count of 30 ⫻ 106 were evaluated on the XE-5000. In three of 20 sam-
WBCs/L and measuring WBC counts in the open ples, 1 ⫻ 106 PMNs/L were detected, whereas the
body fluid mode on the XE-5000 over 19 consecu- remaining samples had no WBCs. This corresponds
tive days. to LoB of ⬍ 2 ⫻ 106/L for WBCs.

Linearity. WBCs isolated from blood were diluted in


Functional sensitivity
a pool of cell-free CSF at eight different expected cell
counts between 5 ⫻ 106/L and 5000 ⫻ 106/L (5, 10, Figure 1 shows the within-run imprecision on
25, 50, 250, 500, 1000 and 5000 ⫻ 106/L). WBC XE-5000 in quantitating total number of WBCs at
counts were evaluated on the XE-5000, the FACS- three expected cell counts, with a CV of 32% at
Canto and manual counting. The expected WBC 5 ⫻ 106 WBCs/L, 19% at 25 ⫻ 106 WBCs/L, and
counts were plotted against the measured cell counts 15.8% at 50 ⫻ 106 WBCs/L. Thus, the functional
676 A. Li et al.

sensitivity on XE-5000, defined as the lowest level 0 h (slope ⫽ 0.37). The mean bias was ⫺ 52.9% for
with total CV of less than 20%, was ⱖ 25 ⫻ 106 manual counting.
WBCs/L.
The total WBC counts by manual counting
showed lower numbers of WBCs compared with Differentiation between MNCs and PMNs
expected cell counts and the results from the
XE-5000. High imprecision was shown by manual Table I summarizes the results of WBC differentia-
counting, with a CV of 35% for 5 ⫻ 106 WBCs/L, tion on the FACSCanto, XE-5000 and manual
38% for 25 ⫻ 106 WBCs/L and 40% for 50 ⫻ 106 counting. FACSCanto and XE-5000 showed lower
WBCs/L. bias for MNC count in both PB dilutions and
Between-day accuracy was tested by diluting the isolated MNC dilutions than manual counting when
Sysmex control (e-CHECK) at an expected count of compared with the expected MNC counts. A high
30 ⫻ 106 WBCs/L. Median WBC count on the bias, especially at low concentrations of MNC counts,
XE-5000 was 30 ⫻ 106/L (range 24–39 ⫻ 106/L) was found when microscopic analysis was applied.
and the mean count was 31 ⫻ 106/L. Between-day Manual counting improved at expected MNC count
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

imprecision was 15.3%. of 46 ⫻ 106/L.


In PB dilutions, PMN analysis on the FACS-
Canto showed preferable bias (ⱕ 11%) at all four
Linearity expected PMN counts tested. On the XE-5000, high
bias (33%) was shown only at low expected PMN
Linearity was tested at expected concentrations of count (12.8 ⫻ 106/L). Bias varies from 25–56% in an
WBCs from 5 ⫻ 106/L to 5000 ⫻ 106/L. There expected range of PMN counts (6.4–64 ⫻ 106/L)
were good agreements between expected WBCs when manual counting was used. In isolated MNC
and measured cells on FACSCanto (y ⫽ 0.86x ⫹ 3.9, samples, PMN analysis showed high bias in the
R2 ⫽ 0.998) and XE-5000 (y ⫽ 0.92x ⫹ 0.2, majority of PMN counts despite of these methods
R2 ⫽ 0.996) regarding total number of WBCs used in the experimental set-up. However, the main
For personal use only.

(Figure 2A, 2B). The mean bias was ⫹ 2.6% for purpose using isolated MNC from blood was to test
FACSCanto and ⫹ 5.5% for XE-5000 compared to accuracy of MNC analysis. Figure 3 shows a scatter
expected cell counts. However, manual counting plot from the XE-5000 (panels A and B) and from
generally gave lower counts than expected and the the FACSCanto (panels C and D) of the MNC and
cell counts from the XE-5000 and FACSCanto. A PMN cell populations before and after purification
low slope value by Passing-Bablok regression analysis of MNCs.
was shown in Figure 2C (y ⫽ 0.37x – 2.0, R2 ⫽ 0.988). The results raised the question of whether it
At low level of WBC counts ⬍ 50 ⫻ 106/L, manual might be easier to miscount PMNs at low quantities
counting failed to detect any cells (Figure 2C). The of WBCs by microscopic analysis.
mean bias was ⫺ 73.2% for manual counting.
In the experimental set-up, we found that the
total number of WBCs was lower at 3 h than at 0 h
on the XE-5000 and FACSCanto (Figure 2D and Comparison of patient samples
2E). The mean bias was ⫺ 7.4% for FACSCanto Comparison of results from chamber counting and
and ⫺ 34.4% for XE-5000. Interestingly, manual from the XE-5000 was performed in 198 CSF
counting showed slightly higher WBC counts samples. WBC counts was significantly correlated
(slope ⫽ 0.39, Figure 2F) at this time point than at between manual counting and XE-5000 (r ⫽ 0.833,
p ⬍ 0.001). Mean difference was 40.8 ⫻ 106/L
(range 14.5–67.1). WBC count by the XE-5000 was
classified to the same subgroup as manual counting
in 161 out of 198 samples (81%) as shown in
Table II. We identified 37 samples with discordant
results. Seven samples showed lower WBC counts on
the XE-5000 than by manual counting, whereas
30 samples had elevated WBC counts on the
XE-5000.
Cell analysis in WBC differentiation between
manual counting and XE-5000 in clinical CSF
samples showed that mean difference was higher
for PMNs (51.1 ⫻ 106/L, range from ⫺ 11.6 to 114)
Figure 1. Within-run imprecision for WBC counts on the XE-5000
in the experimental set-up. WBCs isolated from peripheral blood
than for MNC (7.95 ⫻ 106/L, range from ⫺ 14.6 to
were diluted at three expected counts of 5, 25, and 50 ⫻ 106 30.5, data not shown).
WBCs/L.
Automated WBC counts in CSF on XE-5000 677
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

Figure 2. Linearity of WBC counts on the FACSCanto, the XE-5000 and manual counting. Agreement between expected WBC count
and measured cell count is shown by scatter plot (Passing-Bablok regression analysis). The identity line (y ⫽ x) is drawn. The x-axis
represents expected WBC counts. The y-axis represents measured cell counts. Cells were immediately analysed after dilution on the
FACSCanto (A), the XE-5000 (B), and by manual counting (C). Cell counts were performed 3 h after dilution on the FACSCanto (D),
the XE-5000 (E), and by manual counting (F). Diluted cells were kept at room temperature.
For personal use only.

Discussion differentiation between MNCs and PMNs, com-


In the present study, we found that the open body pared to manual counting. In experiments we found:
fluid mode of the Sysmex XE-5000 was a favourable (1) that within-run and between-day accuracy was
method for determination of WBC counts and for acceptable at WBC counts of ⱖ 25 ⫻ 106/L, (2) that

Figure 3. Scatter plot of cell populations before and after separation of mononuclear cells. Differential counts of PMNs and MNCs on
the XE-5000 before isolation of MNCs from peripheral blood (panel (A) normal mode on DIFF channel) and after isolation of MNCs
from peripheral blood (panel (B) body fluid mode on DIFF channel). Differential counts on the FACSCanto before (C) and after (D)
isolation of MNCs from peripheral blood. R2 represents the population of lymphocytes, R3, represents the population of monocytes and
R4 represents the population of granulocytes. This Figure is reproduced in color in the online version of The Scandinavian Journal of
Clinical & Laboratory Investigation.
678 A. Li et al.

Table I. WBC differentiation by FACSCanto, XE-5000 and manual counting in the experimental set-up.

MNCs ⫻ 106/L (bias %) PMNs ⫻ 106/L (bias %)


Expected Expected
MNCs ⫻ 106/L FACSCanto XE-5000 Manual PMNs ⫻ 106 /L FACSCanto XE-5000 Manual
PB*
3.6 4.2 (17) 3.5 (⫺ 3) 0 (⫺ 100) 6.4 6.8 (6) 8 (25) 8 (25)
7.2 7.3 (1) 6.5 (⫺ 9) 0 (⫺ 100) 12.8 12.9 (1) 17 (33) 20 (56)
36 37.9 (5) 32 (⫺ 11) 24 (⫺ 33) 64 67.3 (5) 55 (⫺ 13) 90 (41)
360 366.5 (2) 337 (⫺ 6) 340 (⫺ 6) 640 711.5 (11) 646 (1) 594 (⫺ 7)

MNCs#
4.6 4.8 (4) 5.5 (20) 0 (⫺ 100) 0.4 0.7 (75) 5 (1150) 0 (⫺ 100)
9.2 7.9 (⫺ 14) 8.5 (⫺ 8) 0 (⫺ 100) 0.8 0.7 (⫺ 13) 1.5 (88) 0 (⫺ 100)
46 46.4 (1) 42 (⫺ 9) 50 (9) 4 0.5 (⫺ 88) 4 (0) 10 (150)
460 450 (⫺ 2) 395 (⫺ 14) 456 (⫺ 1) 40 4.5 (⫺ 89) 24 (⫺ 40) 34 (⫺ 15)

MNCs, mononuclear cells; PMNs, polymorphonuclear cells. *PB, peripheral blood before dilution had 6.1 ⫻ 109 WBCs/L: 64% PMNs
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

and 36% MNCs. The expected MNCs were obtained by diluting PB in cell-free CSF. #MNCs, mononuclear cells including monocytes
and lymphocytes were separated from the PB above using Lymphoprep and had 895 ⫻ 106 WBCs/L: 8% PMNs and 92% MNCs after
separation. The isolated MNCs were diluted in cell-free CSF.

linearity showed a good agreement between expected sensitivity for the manual counting method was not
number of cells and measured cell counts, and (3) investigated systematically, preliminary results
that differentiation between MNCs and PMNs was showed high imprecisions (CV varied from 35–40%)
in good concordance with the reference method at all three expected WBC counts tested in this study.
(flow cytometry) except at low PMN counts. In com- In a recent report, a slightly lower CV of 12.5% for
parisons of clinical CSF samples, there was good the XE-5000 compared with 15.2% for the manual
For personal use only.

overall agreement between the XE-5000 and manual Fuchs-Rosenthal chamber was demonstrated [13].
counting in the majority of cases. In discordant sam- However, high imprecision with WBC counts of
ples, WBC counts were often higher on the XE-5000 ⬍ 10/μL on the XE-5000 was reported [15]. Boer
than by manual counting. et al. reported total precision on the XE-5000 (a CV
Requirements for WBC counts are based on the of 14.9% for a mean WBC count of 37/mm3) [10].
lowest result with clinical significance. According to In this study, we found a similar degree of between-
CLSI H56-A, this was in the range of 6–10 ⫻ 106/L in day imprecision of 15.3% for a mean WBC count of
adult lumbar CSF samples [2], which has been widely 31 ⫻ 106/L.
used by laboratories and accepted in clinical practice. With consideration on stability of WBCs in CSF
In this study, we found that the LoB (defined as ⫹ 2 we tested WBCs diluted in CSF at room temperature
SD of blank) for WBC (MNCs/PMNs) on the from 0–6 h and found a tendency of a decrease in
XE-5000 was ⬍ 2 ⫻ 106/L. Linearity for WBC count- total number of WBCs mainly due to loss of PMNs
ing in the experimental system covered the expected (data not shown). In the experimental set-up, we
cell counts in the majority of clinical CSF samples. An showed decreases of WBC counts after 3 h. However,
acceptable agreement was found in most experiments our experimental set-up of the stability test is of lim-
between the expected number of WBCs and the ited value because patient samples would be prefer-
results from the XE-5000 and FACSCanto, while ably used. One previous study found that the decay
manual counting in general gave fewer WBCs. We in WBCs in CSF reached 40% after 2 h at room
believe that our method of manual counting may suf- temperature [17]. In another study, after 90 min,
fer from an unpredictable amount of under-recovery. lymphocyte counts were reduced by 65% while decay
Data on the analysis of within-run accuracy of monocytes and neutrophils was more rapid and
revealed an acceptable CV (19%) at a WBC count reached 90% [18]. The data collected indicate that
of 25 ⫻ 106/L, but poor precision (CV ⫽ 32%) at the immediate analysis of CSF is important to pro-
a WBC count of 5 ⫻ 106/L. Although functional vide accurate cell counts and differentiation of WBCs,
MNCs, and PMNs.
Table II. Agreement between manual and automated (XE-5000) Accuracy in discrimination between MNCs and
WBC counting in 198 clinical cerebrospinal fluid samples.
PMNs by XE-5000 has not been critically evaluated
0–5 ⫻ 106 6–20 ⫻ 106 ⬎ 20 ⫻ 106 previously. Boer et al. described that WBC counting
Manual/XE5000 n WBCs/L WBCs/L WBCs/L
on the XE-5000 provided the highest rate of ‘correct
0–5 ⫻ 106 WBCs/L 121 96 25 0 negatives’ (or the lowest rate of ‘false positives’) of
6–20 ⫻ 106 WBCs/L 25 6 14 5 all analyzers reported so far [10]. However, they rec-
⬎ 20 ⫻ 106 WBCs/L 52 0 1 51
ommended manually counting of all CSF samples
Total 198 102 40 56
with WBC counts of ⬎ 5 and ⬍ 20 ⫻ 106/L due to the
Automated WBC counts in CSF on XE-5000 679

high degree of imprecision at low WBC counts tended to be higher than by manual counting.
(⬍ 20 ⫻ 106 WBCs/L). Our findings in the experi- This is in agreement with the observation of under-
mental set-up showed that our manual method of recovery with manual counting seen in the experi-
differentiation between MNCs and PMNs in this mental set-up.
range of WBC count could not improve the accuracy. The reason for our manual counting method
Furthermore, differential counting by microscopic apparently showing under-recovery is unclear. We
analysis in clinical CSF samples agree our experi- believe that cells may stick to the walls of tips or
mental finding. Paris et al. also found that there was tubes in the process of sample dilution in Türk’s
a poor correlation in WBC differential count between solution and transfer to the counting chamber. Why
the automated XE-5000 method and the manual recovery for the manual method in the experimental
method [12]. de Jonge et al. reported that for PMNs set-up appears to have been better in samples
below 20 ⫻ 106 /L, agreement was poor with a large after 3 h of storage is not easy to explain but we
positive bias using the XE-5000 compared with man- speculate that the deteriorating cells in these samples,
ual counting [11]. However, good agreement was particularly PMNs, may be less prone to activation,
shown with higher counts of WBCs, PMNs and aggregation and stickiness. One limitation of our
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

MNCs (⬎ 20 ⫻ 106 WBCs/L). Positive bias was only study is that we analyzed artificial samples. Cells in
seen for the lower WBC counts (⬍ 20 ⫻ 106/L) [11]. patient samples are neutrophils, lymphocytes and
This may be explained by some debris or fragments monocytes from blood. Manipulation of leukocytes
counted as PMNs. Thus, it was not recommended in RBC lysis buffer and transfer to cell-free CSF
that differential analysis (PMNs, MNCs) should be may, however, affect cell integrity, cell adherence,
reported to the clinicians if the WBC count was and recovery. The buffering capacity of cell-free CSF
below 10 ⫻ 106 WBCs/L [11]. Regarding cell counts is low, and the pH will increase as CO2 in solution
obtained with flow cytometry, some low MNC counts evaporates [20], something that may affect recovery.
and some high PMN counts were observed on the On the other hand, if this kind of handling really
Sysmex XT-4000i and the XE-5000 in 20 DGKL does affect recovery, the robustness of the manual
CSF controls [14]. Furthermore, poor likelihood counting method must be called into question.
For personal use only.

ratios, both positive (3.0) and negative (0.29), were Automated analysis of a single CSF sample on
recently demonstrated in detection of tumour cells the XE-5000 only takes approximately 195 s,
in CSF by the XE-5000, suggesting that evaluation including switching mode and washing steps. A
by microscope seems mandatory [15]. Our experi- recent study found that CSF cell analysis using the
ments support that microscopic analysis is necessary XE-5000 gives a 7.5-fold reduction in turnaround
in distinguishing between malignant and non-malig- time compared to counting with the Fuchs-
nant cells, because a high bias between expected and Rosenthal counting chamber [13]. This may be an
automated MNC and PMN counts on the XE-5000 important additional advantage for clinicians who
was found at low WBC counts below 10 ⫻ 106/L. often have to make therapeutic decisions as quickly
Manual microscope examination, preferably using as possible.
stained cytospin preparation, is no doubt required In summary, the results of our experiments and
for malignant CSF samples. comparisons in clinical CSF samples indicate that
When comparing CSF samples from patients, a the XE-5000 may offer a relatively robust and accu-
recent study found that agreement between the man- rate method for automated routine quantification of
ual method and the XE-5000 method reached 95% WBCs, MNCs and PMNs in CSF in a diagnostic
for the lowest level of WBCs (ranging from 0–10 setting. Despite its limitations, the accuracy of
cells/μL) and it was 100% for highest level (⬎ 200 XE-5000 is superior to that of the manual counting
WBCs/μL) [19]. However, MNC counts in the low- method.
est category was overestimated by the automated
method [19]. We observed an overall agreement
between manual counting and counts from the Acknowledgements
XE-5000 in 81% of CSF clinical samples. We also
observed a high frequency of elevated WBC counts We thank Anita Hurula and all the technicians at the
by automated counting compared to manual count- Department of Clinical Chemistry, Umeå University
ing. The discordant WBC counts may have been due Hospital, Sweden, for performing the analyses.
to the explanations given above, and to the perfor-
mance of manual counting by several technicians,
some of whom were very experienced and some of Declaration of interest: The authors report no
whom were less experienced. The lack of agreement conflict of interest. The authors alone are responsible
may also have been a result of the time delay between for the content and writing of the paper.
manual counting and analysis on the XE-5000, which This study was supported by grants from the
possibly affected cell sedimentation or cell survival. County Council of Västerbotten, Umeå, Sweden
WBC counts (MNCs, PMNs) on the XE-5000 often (ALF) Grant number (ALF no 7000468).
680 A. Li et al.

References mode on the Sysmex XE-5000 for counting leukocytes and


erythrocytes in cerebrospinal fluid and other body fluids.
[1] Berggren Söderlund M, Blom A, Carlson J, Dahlbäck B, Clin Chem Lab Med 2010;48:665–75.
Hansson L, Harper P, Hillarp A, Hultdin J, Jeppsson J, [12] Paris A, Nhan T, Cornet E, Perol JP, Malet M, Troussard X.
Ljungdahl L, Malm J, Sterner G, Tryding N, Vinge E, Performance evaluation of the body fluid mode on the
Eriksson S, Christensson A, van Hage M. Akuta platform Sysmex XE-5000 series automated hematology
likvoranalyser. Laurells Klinisk Kemi i praktisk medicin analyzer. Int J Lab Hematol 2010;32:539–47.
2003;8:e:364–5. [13] Zimmermann M, Ruprecht K, Kainzinger F, Heppner FL,
[2] Szamosi DI, Bautista JM, Cornbleet J, Glasser L, Rothe G, Weimann A. Automated vs. manual cerebrospinal fluid cell
Sandhaus L, Key M, Meloni-Ehrig A, Culp NB, counts: a work and cost analysis comparing the Sysmex
Dougherty W. Body fluid analysis for cellular composition; XE-5000 and the Fuchs-Rosenthal manual counting cham-
Approved Guideline. H56-A. 2006;26. ber. Int J Lab Hematol 2011;33:629–37.
[3] Aune MW, Sandberg S. Automated counting of white [14] Kleine TO, Nebe CT, Lower C, Geilenkeuser WJ,
and red blood cells in the cerebrospinal fluid. Clin Lab Dorn-Beineke A. Cell analysis in cerebrospinal fluid (CSF)
Haematol 2000;22:203–10. using Sysmex® hematology analyzers XT-4000i and
[4] Mahieu S, Vertessen F, Van der Planken M. Evaluation of XE-5000: evaluation with CSF controls of the Joint German
ADVIA 120 CSF assay (Bayer) vs. chamber counting of Society for Clinical Chemistry and Laboratory Medicine
cerebrospinal fluid specimens. Clin Lab Haematol 2004; (DGKL). Cytometry A 2012;81:255–64.
Scand J Clin Lab Invest Downloaded from informahealthcare.com by Nyu Medical Center on 06/18/15

26:195–9. [15] Zimmermann M, Otto C, Gonzalez JB, Prokop S,


[5] Salinas M, Rosas J, Iborra J, Manero H, Pascual E. Ruprecht K. Cellular origin and diagnostic significance of
Comparison of manual and automated cell counts in EDTA high-fluorescent cells in cerebrospinal fluid detected by the
preserved synovial fluids. Storage has little influence on the XE-5000 hematology analyzer. Int J Lab Hematol 2013;
results. Ann Rheum Dis 1997;56:622–6. 35:580–8.
[6] Andrews J, Setran E, McDonnel L, Kussick S, Wood BL, [16] Kleine TO, Nebe CT, Lower C, Lehmitz R, Kruse R,
Sabath DE. An evaluation of the cell-dyn 3200 for counting Geilenkeuser WJ, Dorn-Beineke A. Modifications of haema-
cells in cerebrospinal and other body fluids. Lab Hematol tology analyzers to improve cell counting and leukocyte
2005;11:98–106. differentiating in cerebrospinal fluid controls of the Joint
[7] Aulesa C, Mainar I, Prieto M, Cobos N, Galimany R. Use German Society for Clinical Chemistry and Laboratory
of the Advia 120 hematology analyzer in the differential Medicine. Cytometry A 2009;75:688–91.
cytologic analysis of biological fluids (cerebrospinal, perito- [17] Chow G, Schmidley JW. Lysis of erythrocytes and leuko-
neal, pleural, pericardial, synovial, and others). Lab Hematol cytes in traumatic lumbar punctures. Arch Neurol 1984;
For personal use only.

2003;9:214–24. 41:1084–5.
[8] Barnes PW, Eby CS, Shimer G. An evaluation of the utility [18] Dux R, Kindler-Rohrborn A, Annas M, Faustmann P,
of performing body fluid counts on the coulter LH 750. Lab Lennartz K, Zimmermann CW. A standardized protocol for
Hematol 2004;10:127–31. flow cytometric analysis of cells isolated from cerebrospinal
[9] Sandhaus LM, Ciarlini P, Kidric D, Dillman C, fluid. J Neurol Sci 1994;121:74–8.
O’Riordan M. Automated cerebrospinal fluid cell counts [19] Perne A, Hainfellner JA, Womastek I, Haushofer A,
using the Sysmex XE-5000: is it time for new reference Szekeres T, Schwarzinger I. Performance evaluation of the
ranges? Am J Clin Pathol 2010;134:734–8. Sysmex XE-5000 hematology analyzer for white blood cell
[10] Boer K, Deufel T, Reinhoefer M. Evaluation of the XE-5000 analysis in cerebrospinal fluid. Arch Pathol Lab Med
for the automated analysis of blood cells in cerebrospinal 2012;136:194–8.
fluid. Clin Biochem 2009;42:684–91. [20] Kristensen SR, Salling AM, Kristensen ST, Hansen AB.
[11] de Jonge R, Brouwer R, de Graaf MT, Luitwieler RL, Unrecognized preanalytical problem with the spectrophoto-
Fleming C, de Frankrijker-Merkestijn M, Sillevis Smitt PA, metric analysis of cerebrospinal fluid for xanthochromia.
Boonstra JG, Lindemans J. Evaluation of the new body fluid Clin Chem. 2008;54:1924–5.

You might also like