mm 9009276 0031287 138 mm
: qf Designation: D 4327-84 *
AMERICAN SOCIETY FOR TESTING AND MATERIALS
1016 Race St., Philadelphia, Pa, 19103
Reprinted from the Annual Book of ASTM Standards, Copyright ASTM,
If pt listed in the current combined Index, will appear in the next edition,
Standard Test Method for
ANIONS IN WATER BY ION CHROMATOGRAPHY'*
‘This standard is istued under the fixed designation D-4327; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A nurber in parentheses indicates the year of last reapproval
‘A superscript epsilon (¢) indiates an editorial change since the las revision or repproval,
1. Scope
1.1 This test method? covers the sequential
determination of fluoride, chloride, nitrite, phos-
phate, bromide, nitrate, and sulfate ions in water
by ion chrdmatography.
1.2 This test method is applicable to drinking,
ground and surface waters, boiler and cooling
waters, domestic and industrial wastes, and at-
mospheric precipitation samples.
1.3 Concentrations as low as 0.1 mg/L can be
determined depending upon the anions to be
quantitated. Utilizing a 100-1L sample volume
and a sensitivity of 3 uS/em full scale, the follow-
ing are approximate detection limits:
Approximate
‘Detection Limit,
‘Anion mg/L
Fluoride (>) on
Chloride (Cr) ‘ 4
Nitrite (NO) 02
Phosphate (PO) o1
Bromide (BF) 02
Nitrate (NOS) 02
Sulfate (S007) 02
If lower detection levels are required, the sen-
sitivity may be improved by using a lower scale
setting (<3 S/cm) or a larger sample injection
(100 BL). The analyst must assure optimum
instrument performance to maintain a stable
baseline at more sensitive conductivity full-scale
settings,
1.4 The upper limit of this test method is
dependent upon total anion concentration and
may be determined experimentally as described
in Annex Al. These limits may be extended by
appropriate dilution.
1.5 It is the users’ responsibility to assure the
validity of this test method to other matrices.
16 It is the analysts responsibility to assure
the validity of the method for untested matrices.
1.7 This standard may involve hazardous ma-
terials, operations, and equipment. This standard
does not purport to address all of the safety prob-
lems associated with its use, Iti the responsibil.
ity of: whoever uses this standard to consult and
establish appropriate safety and health practices
and determine the applicability of regulatory limi-
tations prior to use.
2. Applicable Documents
2.1 ASTM Standards:
D 1066 Practic for Sampling Steam?
D1129 Definitions of Terms Relating to
Wate?
D 1193 Specification for Reagent Water?
D 2777 Practice for Determination of Preci-
sion and Bias of Methods of Committee D-
19 on Water
D 3370 Practices for Sampling Water?
E 200 Preparation, Standardization, and Stor-
age of Standard Solutions?
E 260 Recommended Practice for General
Gas Chromatography Procedures
E 355 Recommended Practice for GC Terms
and Relationships!
‘This test method is under the juridiction of ASTM Com-
mittee D-19 on Water and isthe direc responsfblity of Subcom-
mittee D19.05 on Inorganic Constituents in Water.
josie elton approved Fe. 28, 198, Published March
The following references may be consulted for additional
information:
Small, H, Steveas, T. S, and Bauman, W. C., “Novel Ton
Exchange Chromatographic Method Using Conductimetric
Detection.” Analytical Chemistry, Vol 47, 1975, p. 1801.
Stevens, T.S, Turkeluon, V.T. and Alve, W. Rey=Determina-
tion of Anions in Boiler Blow Down Water with lon Chro-
‘matopraphy,” Analytical Chemistry, Vol 49, 1977, p. 1176.
Sawicki, E, Mulik, J.D, and Wigensein, E., Editors, Jon
Chromatographic Analysis of Environmental Polbwants, AN
Arbor Science Publishes, Ann Arbor, Mich, 1978.
Mulik, J. D. and Sawicki, E, Editors, lon Chromatographic
Avalbie from Dionex Cor, 1228 Tan Way, Sunayvale,
Cali 94086, or an equivalent may be used.conductivity celf equipped with a itieter capable
of reading from 0 to 1000 pS/cm on a linear
scale.
7.1.5 Pump, capable of delivering a constant
flow of approximately 2 to 5 mL/min throughout
the test and tolerating a pressure of 200 to 700
psi (1.3 to 4.8 MPa). rd
7.2 Recorder, compatible with the detector
output with a full-scale response time in 2 s or
less.
7.3 Syringe, minimum capacity of 2 mL and
equipped with a male pressue fitting.
1.4 Eluent and Regenerant Reservoirs—Suit-
able containers for storing eluents and regener-
ant.
7.5 Integrator—It may be desirable to inte-
grate the area under the chromatogram, Different
integrators can perform this task when compati-
ble with the electronics of the detector meter or
recorder. If an integrator is used, the maximum
area measurement must be within the linear
range of the integrator.
8. Reagents
8.1 Purity of Reagents—Reagent grade chem-
icals shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents shall
conform to the specification of the Committee
on Analytical Reagents of the American Chemi-
cal Society, where such specifications are avail
able. Other grades may be used, provided it is
first ascertained that the reagent is of sufficiently
high purity to permit its use without lessening
the accuracy of the determination.
8.2 Purity of Water—Unless otherwise indi-
cated, references to water shall be understood to
mean reagent water conforming to Specification
D 3193, Type I. Column life may be extended
by passing Type II water through a 0.22-m filter
prior to use.
8.3 Anion Working “Solutions—Prepare_&
blank and at least three different working solu-
tions containing the following combinations of
anions. The combination anion solutions must
be prepared in volumetric flasks. See Table 1.
8.3.1 Prepare a high-range standard solution
by diluting the volumes of each anion specified
in Table | together to 1 L with water.
8.3.2 Prepare the intermediate-range standard
solution by diluting 10.0 mL of the high-range
standard solution (see 8.3.1) to 100 mL with
water.
wa yuutiess
4327
Uuab265 TOO mi
8.3.3 Prepare the low-range standard solution
by diluting 20.0 mL of the intermediate-range
standard solution (see 8.3.2) to 100 mL with
water.
8.4 Eluent—Dissolve 1.0080 g of sodium bi-
carbonate (0.003 M NaHCOs) and 1.0176 g of
sodium carbonate (0.0024 Mf NaCOs) in water
and dilute to 4 L with water.
8.5 Stock Solutions:
8.5.1 Bromide Stock Solution (1.00 mL=1.00
mg Br”)}—Dry approximately 2 g of sodium bro-
mide (NaBr) for 6 h at 150°C and cool in a
desiccator. Dissolve 1.2877 g of the dried salt in
water and dilute to 1 L with water.
8.5.2 Chloride Stock Solution (1.00 mL = 1.00
mg Cl’)}—Dry sodium chloride (NaCl) for 1 hat
600°C and cool in a desiccator. Dissolve 1.6484
gof the dry salt in water and dilute to { L with
water.
8.5.3 Fluoride Stock Solution (1.00 mL = 1.00
mg F-)—Dissolve 2.2160 g of sodium fluoride
(NaF) in water and dilute to 1 L with water. Store
in chemical resistant glass or polyethylene.
8.5.4 Nitrate Stock Solution (1.00 mL = 1.00
mg NO3")—Dry approximately 2 g of sodium.
nitrate (NaNO;) at 105°C for 24 h. Dissolve
exactly'1.3707 g of the dried salt in water and
dilute to 1 L with water.
8.5.5. Nitrite Stock Solution (1.00 mL = 1.00
mg NOs)—Place approximately 2 g of sodium
nitrite (NaNO,) in a 125-mL beaker and dry to
constant weight (about 24 bh) in a desiccator
containing concentrated H,S0, (sp gr 1.84). Dis-
solve 1.4998 g of the dried salt in water and dilute
to I L with water. Store in a sterilized glass bottle
(Notes 1 and 2). Refrigerate and prepare
monthly.
Nore 1—Nitite is easily oxidized, especially in the
presence of moisture, and only fresh reagents are to be
used.
(ore 2—-Prepare sterile bottles for storing nitrite
solutions by heating for 1 h at 170°C in an air oven.
8.5.6 Phosphate Stock Solution (1.00 mL. =
1,00 mg PO,~")—Dissolve 1.4330 g of potassium
dihydrogen phosphate (KH2PO,) in water and
dilute to 1 L with water.
8.5.7 Sulfate Stock Solution (1.00 mL = 1.00
‘Reagent Chemical, American Chemical Society Spei
cations” Am. Chemica! Soe, Washinglon, De, For sues
tions on the tesiog of agents not ied bythe Amenexs
(Chemical Soccry, se “Reagent Chemicals and Stands,”
bby oseph Rosin, D. Van Nostrand Co, lnc, New York, NY,
nd the “United States Pharmacopeci)
mg SO.)—Dry sodium sulfate’(Na;SO.) for 1
‘h at 105°C and cool in a desiccator. Dissolve
1.4790 g of the dried salt in water and dilute to
1 L with water.
8.6 Suppressor Regenerant Solution—Cau-
tiously add 111 mL of concentrated H:SOx (sp
gx 1.84) to approximately 600 mL of water. Cool
the solution and dilute to 4 L with water.
9. Sampling
9.1 Collect the sample in accordance with ap-
plicable Practice D 1066 and Practices D 3370.
9.2 Analyze the samples as soon as possible
after collection, Preservation by refrigeration at
4°C is required.
9.3 Filter the samples containing particulates
through a prewashed 0.22-um filter prior to anal-
ysis to avoid fouling or clogging the resin col-
umns.
10. Calibration
10.1 Determination of Retention Times:
10.1.1 The retention time for each anion is
determined by injecting a standard solution con-
taining only the anion of interest and noting the
time required for a peak to appear on the chro-
matogram. Retention times vary with operating
conditions and are influenced by the concentra-
tion of ion(s) present. Prepare separate standard
solutions in accordance with Table 2 by transfer-
ring by pipet the designated aliquots of stock
solutions prepared in Section 8 (8.5.1 through
8.5.7) into separate I-L volumetric flasks. Ana-
lyze each standard of interest as defined in Sec-
tion 11. Note the time in minutes for each peak
to appear on the chromatogram.
Note 3—Some operators have reported unvsually
large shifis in retention time for nitrate with changes in
concentration. If this occurs, care must be taken to
ensure integration of the correct peak when integration
is used for calculation.
10.1.2 Concentrations other than those listed
in Table 2 may be used if they better approximate
concentrations expected in the samples. Those
concentrations listed will give about midscale
response with a 1-V recorder input and a con-
ductivity meter full-scale setting of 10 uS/cm.
10.1.3 Although retention times may vary
from one column to another and are influenced
bby such factors as temperature and eluent flow
rates, the elution order of the anions is always as
follows when (8.4) using the standard CO3"/
mE SUUSE /b UUSbeU eee
Dasa
HCO}" eluent: F, CI", NOy
soe, :
10.2 Analyze the blank and each of the pre-
pared calibration solutions described in 8.3 in
accordance with the defined procedure (see Sec-
tion 11).
POL, Br, NO,
Nore 4—Ifthe concentrations of the sample ions of
interest are known or estimated, the concentration of
standard solutions prepared for instrument calibration
may be varied to better approximate or bracket the
concentration range of interest. Anions of no interest
‘may be omitted
Nore $—The intermediate-range combination
anion standard (see 8.3.2) may be used to verify reso-
lution of all seven anions.
Nore 6—Each analytical curve should be established
using only one scale setting. Changing the scale setting
‘may result in a slight change in the slope of the analyt-
ical curve, :
10.3 Prepare analytical curves for each anion
of interest by plotting on linear graph paper peak
height or peak area versus the nominal concen-
tration of the anion standard solution.
Nore 7—Some operators have reported a shift ia
slope of the phosphate calibration curve at approxi-
rately 30 mg/L PO,,Ifsucha shiftin slope is observed,
auditional standard solutions covering the entite range
of corcentration should be prepared and analyzed in
corder-to accurately define the slope of the curve. If an
integrators being used, it may be necessary to manually
calculate phosphate concentrations above 30 mg/L POs
in order to oblain maximum accuracy.
11. Procedure
11.1 Set up the ion chromatograph in accord-
anee with the manufacturer's instructions. See
Annex Al for information concerning possible
configurations for instrument set-up.
Nore 8—The detector ranges are variable. Normal
operating ranges are from 3 10 100 wS/cm full scale.
‘The range setting required for analysis will depend on
the concentration of anions in the sample and should
be chosen accordingly
11.2 Equilibrate the system by pumping
eluent through all three columns until a stable
baseline is obtained (approximately 15 to 20
min). This equilibration can normally be accom-
plished while the samples and standards are being
prepared.
11.3 Set up the instrument as directed by the
manufacturer, Load 2 to 3 mL of sample into
the sample entry port using a syringe (see Note
7). Inject the sample into the eluent stream and
record the ion chromatogram (refer to Fig. 2).
Note 9—Most plastic disposable syringes, if used,
-cU
come in packages that ae labeled “sterile,” but sterile
does not necessarily imply clean of anions; and there-
fore, all syringes should be flushed with’ water and
sample or standard before use to mizimize contami-
nation.
11.4 When the suppressor column becomes
partially exhausted, regenerate with suppressor
Tegenerant solution (8.6) in accordance with
manufacturer's instructions.
12. Calculations
12.1 Refer the peak height or area noted for
the anion(s) of interest to the appropriate analyt-
ical curve(s) (see 10.3) to determine the anion
concentration in milligrams per litre.
Anion concentration, mg/L = Ax F
where:
A= milligrams per litre read from appropriate
TABLE 1 Preparation of Standard Solutions for Instrument Calibratios
4327
calibration curve and
F= dilution factor if sample was diluted prior to
analysis,
13. Precision and Bias?
13.1 The collaborative test of this test method
was performed in reagent water by 15 laborato- {
ries, one operator each. Fourteen operators made
uplicate determinations at each level for a total
of 29 determinations at each level for each of
seven anions,
13.2 The precision and bias of this test
method for each anion in reagent water is ex-
pressed in Tables 3 through 9.
13.3 These data may not apply to other water
matrices
* Supporting data has ben ied and are on loan from ASTM.
Request RR:DI9-1105,
Tigh Range Standard Gee 83.1)
Milt of Each "termediate Range Low-Range Staid-
Aion GaSe Sotasos, Anion Concenza. Sunda, m/L (see ar alle
4 fe 832) 533
1a) Diluted 12 = me _
1000 mi. E
Fuodde (F i 0 10 2
Ghieride (Cr) 10 10 10 02
Nitaite (NOP) 20 20 ° oa
Phosphate (P00) 50 50 10
Bromide (BF) 0 0 a
Nitrate (NOS) 30 20 06
Sule $0." 100 100 100 20}
‘TABLE 2 Preparation of Standard Soltions for
‘Determination of Retention Times
Volume of —_
Stock Solution (1 mL = Stock Solution Anion Concea-
1.00 me) erie of “‘raton, me/t
Wate mi
Fioeide 3 3
Goride 4 4
Nivite 10 10
Phosphate 50 30
Bromide 10 wo
Nite 30 30
Sulfate 50 50mm 5005275 9031252 STS mw
qh 4327
TABLE 3. Foride
‘Amount Adéed, mg Amount Found, mg Ss s oe
Bias, Signifcazt
Le fo 95% Level
019 om ‘O08 ‘O02 +158 7
049 048 0.05 03 = 20 0
43 454 052 046 ra 0
212 aa 21 06 $09 10
130 142 9 2 =53 ys
322 30 86 33 =25 Bo
TABLE 4 Chloride
‘Amount Added, mg Amount Found, mg S a i Statiscally
i Bias, % Significant
oe on 93% Lvl
O17 026 010 008 4529 s
076 086 038 out 4132 0
110 172 oz 043 + 20
549 565 82 24 +29 20°
455 an 46 B +35 0
10% 1099 9 4a $24 20
TABLE 5 Nivite
‘Siatsically
‘Amount Added, mg Amount Found, mg si s
i Bias, % Significant
poe oe : 95% Level
04s 0.08 008 200 s
0.99 085 on ~s46 s
218 194 1 Hi10 ye
667 652 4s =23 sno
434 42 2 +18 20
937 952 4“ +16 0
TABLE 6 Bromide
- Satatally
‘Amount Added, mg Amount Found, mg Ss s 5
ins, % Sigiscant
ae BA 95% Level
025 025 008 002 00 20
046 045 0.04 0.04 22 20
137 29 10 06 =s8 ye
437 aa 39 25 35 20
63 63 3 " m7 20
750 139 3 uN +12 80
TABLE 7. Phosphate
‘Statistically
‘Amount Added, mg Amount Found, me Ss s. Fas, % ay
POUL Poy 95% Level
018 010 ‘006 003 wa vs
049 om 0.15 017 =306 ye
m2 23 Lt 08. =245 ye.
568 32 29 23 37 ye
sit 6 m 3 =17 yes
3 63 n 9 06 ye4d
04327
(2) Bhoent reservoir
@ Pump
(@) Precolumn
() Separator column
(8) Suppressor column,
(6) Detecor
(2) Recorder or integrator, or both
FIG. 1 Schematic of lon Chromatograph
(Mandatory Information)
Al. DETERMINATION OF RANGE OF METHOD
AL The upper range of the method foran injected
aliquot varies for each sample type and is limited by
‘the capacity of the separator column. If the ionic
strength of the sample to be analyzed exceeds column
capacity, the ions of interest will not be separated as
shown in Fig. 2. Retention times will also decrease 2s
the ionic strength of the sample increases. The upper
range of the method for the ion(s) of interest in the
sample matrix may be determined by the followis
procedure. The procedure is defined for currently av
able separator coturans supplied by the appropsiate
manufactarer.
ALLL Bypass the separator column (only the sup-
pressor column is on line) and inject 100 ul of a 5.8-
B/L (0.01 meq) solution of sodium chloride (NaCI).
‘This represents 10 % of the separator column capacity,
that is 0.1 meq. Note the conductivity response.
AJ.1.2 Inject 100 pL of the sample of interest (sep-
arator column bypassed). Note the conductivity re-
sponse.
AL.L3 Ifthe conductivity response noted in AI1.2
is less than or equal to the conductivity response noted
in the sodium chloride solution in Al.1.1, column
capacity has not been exceeded.mm 9009276 003i-94 378 Ml
q) 04327
ALLA If the conductivity response noted in Al1.2 must be diluted prior to injection into the ion chro-
is greater than the conductivity response noted for the matograph. ‘
sodium chloride solution in ALLt-1, the test sample
A2. CONFIGURATIONS FOR SETTING UP AN ION CHROMATOGRAPH
A2.L Table A2.1 lists various configurations for set-; Configurations other than those listed may be used if
ting up an ion chromatograph, Choice of configuratién’, they successfully resolve all anions of interest, as de-
depends primarily upon which ions are of interest. scribed in 7.1.2 of this test method.
TABLE A21 Configurations for Setting Up 2 fon Chromatograph
low Rat,
Separtor Column SuppressorColumn! Flow Rat Comment
4xR0mme 4x 100mm? 23 extrolation of all anions
9x 100 mm
4x smn 4x 250mme OOM? AB wl ale 6 anions in about ne
9x 100 mat of 1) but wll ot eave NO and
ae :
()4xS0mm* 4x SO-mme 4x 100mm? 23 wilreselve NO} and SO¥ ony
9 100 mm?
@)sxsommt 4x S0mm 4x 100mm? 46 will eolve SOF only
9% 100 mmé & :
“In all cass, either suppresior column i suitable The 4 x 100 man will give a smaller waterdip, but the 9% 100 rom wil last
longer between regenerations. a
Dionex Part No, 30825 or equivalent.
Dionex Part No, 30827 or equivalent.
Dionex Par No. 30829 or equivalent
* Dionex Part No. 30828 or equivalent
APPENDIX
(Nonmandatory Information)
‘XL. TECHNIQUES.FOR ELIMINATING WATER DIP
X11 A method of eliminating the water dip is to eluent concentrate (solution that is 100 times more
introduce concentrations of carbonate and bicarbonate _concentraved than the eluent used for analysis) per 100
into the sample that closely approximate that of the ml of sample.
eluent used for analysis. Adjustment of sample back- X12 Standard solutions must be prepared as di-
ground may be accomplished in two different ways: rected in X1.1.2. It is important to prepare a blank
XI.L.1 Dilute the sample with eluent ifsample di- using reagent water and eluent (100:1) to compensate
lution is required prior to analysis. for any anionic impurities present.
‘XI.L2 Aéd an equivalent of 1.0 mL of a prepared
‘The American Society for Teuting and Materials toked no postion respecting the validity of any patent rights asserted in connection
with one item mentioned inthis standard. Users ofthis standard are expresly advised that determination of the validity of any sich
_nutont rights, andthe risk of infringement of such rights, are entirely their own responsibilty.
This standard is subject to revision at any time bythe responsible technica committee and must be reviewed every five years and
If not revised. eter reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional
standards and should be address 19 ASTM Headquarters. Your comments will receive cargul consideration at a meeting of the
responsible technical commitee, which you may attend. Ifyou feel that your comments have not received a fair hearing you should
‘make your views known to the ASTM Committe on Standards, 1916 Race S., Philadelphia, Po. 19103.