ABSTRACT
Objective: To evaluate the effect of cortical bone thickness on the maximum insertion and re-
moval torque of different types of self-drilling mini-screws and to determine if torque depends on
Table 1. Principal Properties and Dimensions of the Mini-screws Used in This Study
Type A Type B Type C
Screw design Pure cylindrical Combined cylindrical part ⫹ taper part Combined cylindrical part ⫹ taper part
Corporation Biomaterials Korea Jeil Medical Ortholution
Body length, mm 7.0 8.0 7.0
Thread length, mm 6.0 7.0 6.0
Thread diameter, mm 1.5 1.6a 1.6 (2.0)
Taper length, mm — 1.5 2.5
Chemical composition Ti-6Al-4V Ti-6Al-4V Ti-6Al-4V
a
X@XType B screw has a constant outer thread diameter as 1.6 mm with an increasing inner core diameter in a taper part.
In all thickness samples, MIT was in order of type C or implant, and bone density of the insertion area are
⬎ type B ⬎ type A (Table 5). mentioned in many reports as factors that affect the
Figure 4. Time-removal torque graph and the definitions of mea- Figure 5. Superimposition of time-insertion torque graphs by cortical
surement values regarding the removal torque test. bone thickness in type A.
Table 4. Cortical Bone Insertion Time (s) According to Cortical Bone Thickness
1 mm 1.5 mm 2 mm Tukey’sa
Type A 94 ⫾ 0.55 144 ⫾ 5.37 268 ⫾ 14.21 1 ⬍ 1.5 ⬍ 2 mm
Type B 68 ⫾ 2.05 103 ⫾ 5.79 130 ⫾ 6.04 1 ⬍ 1.5 ⬍ 2 mm
Type C 122 ⫾ 1.48 132 ⫾ 38.10 159 ⫾ 39.97 1 ⬍ 1.5 ⬍ 2 mm
Tukey’s Type B ⬍ A ⬍ C Type B ⬍ C ⫽ A Type B ⬍ C ⬍ A
a
XStatistical significance is determined by Tukey’s studentized range test (P ⬍ .05).
cortical bone reaches the maximum diameter of the than 1.5 mm, the increase of torque is reduced (Fig-
screw, the increase in torque is mainly affected by the ure 6).
insertion depth on medullary bone. Hence, with the in- Type C also showed an increase in insertion torque
crease of cortical bone, the torque value remains con- with the increase in cortical bone thickness, which was
stant, (Figure 5). With the increase of cortical bone more evident than with type B. According to macro-
thickness, only the time taken to perforate the cortical scopic measurements, the length of the tapered part
bone increased, and the increased rate of torque value was about 2.5 mm, which led to a significant increase
against time was similar. However, as the diameter of in torque values within the 1.5-mm to 2-mm range
the type A mini-screw used in this study was 1.45 mm, (Figure 7).
which is smaller than that of types B and C, there are Although many biomechanical studies have report-
limitations to comparisons of absolute torque values. ed that insertion torque affects the stability of screws,
Type B showed an increase in insertion torque value it is difficult to assert that the insertion torque is pro-
with the increase in cortical bone thickness. However, portional to the stability of mini-screws. Lawes et al27
the increase in MIT between 1.5 mm and 2 mm was suggested that a high insertion torque of a tapered
less than the increase between 1.0 mm and 1.5 mm. bone screw decreases loosening at the interface with
The diameter of the upper part of the screw increases bone. In contrast, Frost28 mentioned that integrity
slightly and transitions into the soft tissue contact area, could not be maintained if too much bone change oc-
which was measured to be approximately 1.5 mm. It curred. Hansson and Werke29 also mentioned that
is because of this part that cortical bone thickness and bone resorption occurs when forces are excessive and
MIT increases. However, with cortical bone thicker exceed the physiologic limit. In other words, when
the nasal floor and tooth root using the skeletal anchorage 19. Heidemann W, Gerlach KL, Gröbel KH, Köllner HG. Influ-
system in dogs. Angle Orthod. 2003;73:158–166. ence of different pilot hole sizes on torque measurements
5. Kanomi R. Mini-implant for orthodontic anchorage. J Clin and pullout analysis of osteosynthesis screws. J Cranio-
Orthod. 1997;31:763–767. maxillofac Surg. 1998;26:50–55.
6. Park HS, Kwon TG, Kwon OW. Treatment of open bite with 20. Homolka P, Beer A, Birkfellner W, Nowotny R, Gahleitner
microscrew implant anchorage. Am J Orthod Dentofacial Tschabitscher M, Bergmann H. Bone mineral density mea-
Orthop. 2004;126:627–636. surement with dental quantitative CT prior to dental implant
7. Park HS, Kwon OW, Sung JH. Microscrew implant anchor- placement in cadaver mandibles: pilot study. Radiology.
age sliding mechanics. World J Orthod. 2005;6:265–274. 2002;224:247–252.
8. Ohmae M, Saito S, Morohashi T, et al. A clinical and his- 21. Baker D, London RM, O’Neal R. Rate of pull-out strength
torical evaluation of titanium mini-implants as anchors for gain of dual-etched titanium implants: a comparative study
orthodontic intrusion in the beagle dog. Am J Orthod Den- in rabbits. Int J Oral Maxillofac Implants. 1999;14:722–728.
tofacial Orthop. 2001;119:489–497. 22. Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen P. Pull-
9. Costa A, Raffaini M, Melsen B. Miniscrews as orthodontic out strength of monocortical screws placed in the maxillae
anchorage: a preliminary report. Int J Adult Orthod Ortho- and mandibles of dogs. Am J Orthod Dentofacial Orthop.