About Instrument
Profile Projector with 300mm Dia screen is suitable for all type of
a 300mm dia rotary screen with a cross hair line, which is rotatable through
How to Operate
Unpack the instrument and place it on the top of the table. Plug the wire to
220v AC mains and switch on the instrument. Place the specimen on the ceter
of the glass plate of X-Y bed and bring it to the optical path in between contour
light & objective lens with the help of micrometers. Now viewing on the screen
move the focusing knob upward until the shadow of the specimen is focused on
the screen. Adjust the surface lights to illuminate the object properly. Now
measure the relative position with the help of micrometer scale with the
reference of cross hair line appeared on the screen. Rotate the screen and note
the reading of vernier for to get the measurement of the angle of an object.
Dinesh Scientific
2 Dynamics of Machine
Laboratory
Procedural Steps
Observe the work piece trough eye piece and focus the object.
Then measure the relative position by measuring the travel of the work
2. Measurement of Angle
The angle are measured by successively setting fiducial line situated in the
plane of the eye piece along with armof the image, or through points indicating
the angle and noting form a protector scale (goniometer) the angle through
second for lateral measurments. We can determine the least count of the
Dinesh Scientific
3 Dynamics of Machine
Laboratory
For Example,
100
Revolving Screen: The function of the 360 degree rotable screen is to measure
the angle of the object, we can determine its count in the following way: -
60
Things to Remember
Use the instruments in complete dust proof room for better results & sharp
image. Never forget to cover the instrument with dust cover after use. Always
check the fuse provided at the back of the intruments in case of the non
Dinesh Scientific
4 Dynamics of Machine
Laboratory
working of illumination. The image will be hazy if the optical surface of the
objective is unclean. In such case clean the front TIP of the lens with help of
cotton clothes.
PROFILE PROJECTOR
300MM
Dinesh Scientific