To check out the crucible opening time it is required to use the “service mode 3” . To switch on the service mode keep the
“3”-key pressed while you switch on the device until the display shows the software version.
Now press the “ casting key “ .
Right after pressing the casting key the display will show you the crucible opening time.
After pressing the “cast “ key again you release the cylinder and the “cast” key can be pressed
again to lift the cylinder for checking the opening time again .
The value should be between 15 and 17 .
If the values are different to the original adjustment correct them to the values showed below.
If you should not be able to adjust the value for the crucible opening time correctly (Y11) the valve is defective and need
to be replaced.
Page 1/3
WARNING!
Page 2/3
Removing of cover of crucible chamber
Remove the top cover as shown on page 2. Loose the counter nut (A) and unplug the pressure
tubes (B) to remove the valve. Mount the new valve and follow the adjustment instructions on page
1. Spare part number: 14153