WI/06 Procedure For Calculation of Uncertainty of Sensor With Probe
WI/06 Procedure For Calculation of Uncertainty of Sensor With Probe
1.1 SCOPE
The following method of calculating the uncertainty is based on a
measurement Uncertainty in compliance to " Expression of Uncertainty &
confidence in measurement.", UKAS M3003.
1.2 PARAMETERS
The parameter which are used for calculating uncertainty, are Uexp is
Expanded Uncertainty.
Uexp = k x UC
And
Where,
Ua – Uncertainty due to repeatability
Umas – Uncertainty due to the master sensor
Ustb – Uncertainty due to Stability of liquid bath / dry block
Uun – Uncertainty due to non uniformity of liquid bath / dry block
Ures – Uncertainty due to resolution of UUC
Udrf – Uncertainty due to drift of the master instrument
Uin.m - Uncertainty due to inhomogeneity of thermocouple ( if used ).
UC – Combined uncertainty
Uexp – Expanded Uncertainty
1.3 CALCULATIONS
Combined Uncertainty
UNCERTAINTY BUDGET
Combined Uncertainty
∞
Degrees of freedom
Normal
Expanded Uncertainty k=2