You are on page 1of 5

American Journal of Infection Control 45 (2017) e69-e73

Contents lists available at ScienceDirect

American Journal of Infection Control American Journal of


Infection Control

j o u r n a l h o m e p a g e : w w w. a j i c j o u r n a l . o r g

Major Article

Effectiveness of cleaning-disinfection wipes and sprays against


multidrug-resistant outbreak strains
Nikki Kenters BSc a,*, Elisabeth G.W. Huijskens PhD, MD a,b, Sophie C.J. de Wit BSc a,
Joost van Rosmalen PhD c, Andreas Voss PhD, MD d,e
a
Department of Infection Prevention and Control, Albert Schweitzer Hospital, Dordrecht, The Netherlands
b
Department of Medical Microbiology, Albert Schweitzer Hospital, Dordrecht, The Netherlands
c Department of Biostatistics, Erasmus MC, Rotterdam, The Netherlands
d
Department of Medical Microbiology, Radboud University Medical Centre, Nijmegen, The Netherlands
e
Department of Medical Microbiology, Canisius-Wilhelmina Hospital, Nijmegen, The Netherlands

Key Words: Background: Hospital rooms play an important role in the transmission of several health care–
Cleaning-disinfecting wipes associated pathogens. During the last few years, a number of innovative cleaning-disinfecting products
Cleaning-disinfecting sprays have been brought to market. In this study, commercially available products combining cleaning and dis-
Outbreak strains
infection were compared, using 2 different application methods. The aim was to determine which product
ATP
was most effective in simultaneous cleaning and disinfection of surfaces.
CFU
Methods: Seven cleaning-disinfecting wipes and sprays based on different active ingredients were tested
for their efficacy in removal of microbial burden and proteins. Efficacy was tested with known Dutch out-
break strains: vancomycin-resistant enterococci (VRE), Klebsiella pneumoniae OXA-48, or Acinetobacter
baumannii.
Results: For all bacteria, ready-to-use cleaning-disinfecting products reduced the microbial count with
a log10 reduction >5 with a 5-minute exposure time, with the exception of a spray based on hydrogen
peroxide. Omitting the aforementioned hydrogen peroxide spray, there were no significant differences
between use of a wipe or spray in bacterial load reduction. Using adenosine triphosphate (ATP) mea-
surements, a significant difference in log10 relative light units (RLU) reduction between various bacteria
(P ≤ .001) was observed.
Conclusions: In general, a >5 log10 reduction of colony forming units (CFU) for tested wipes and sprays
was obtained for all tested bacteria strains, with exception of hydrogen peroxide spray and VRE. Al-
though ATP may show a difference between pre- and postcleaning, RLU reduction does not correlate with
actual CFU reductions.
© 2017 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier
Inc. All rights reserved.

It is becoming apparent that cleaning and disinfection of patient health care–associated pathogen have a greater risk factor of getting
environments in hospitals is extremely important. Multidrug- colonized or infected with the same pathogen.3
resistant microorganisms are emerging globally.1 Hospital rooms play Improved terminal cleaning and disinfection of rooms leads to
an import role in the transmission of several health care–associated decreased rates of pathogen transmission.4 Still, multiple studies have
pathogens, including (methicillin-resistant) Staphylococcus aureus, demonstrated that <50% of hospital rooms are adequately cleaned
vancomycin-resistant enterococci (VRE), and norovirus.2 All these and disinfected when chemical germicides are used.4-6
microorganisms persist in the environment for an extended amount Over the last few years, a number of innovative cleaning and dis-
of time.3 Research has shown that patients admitted to a hospital infecting products have come on to the market. Ultraviolet
room that was previously occupied by a patient that harbored a disinfection and hydrogen peroxide vapor devices are becoming
common in hospitals. These devices are an asset to terminal dis-
infection of patient rooms.7,8 Ready-to-use cleaning-disinfecting
wipes and sprays are becoming regular for cleaning and disinfec-
* Address correspondence to Nikki Kenters, BSc, Albert Schweitzer Hospital, Albert
Schweitzerplaats 25, 3318 AT Dordrecht, The Netherlands.
tion in hospitals. Ready-to-use products can be used for routine and
E-mail address: n.kenters@gmail.com (N. Kenters). terminal cleaning and disinfection. The ease of use of the wipes and
Conflicts of interest: None to report. sprays has the potential to save time and reduce barriers for health

0196-6553/© 2017 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
http://dx.doi.org/10.1016/j.ajic.2017.04.290
e70 N. Kenters et al. / American Journal of Infection Control 45 (2017) e69-e73

Table 1
Disinfecting-cleaning wipe and spray ingredients

Wipe-spray Composition* Product Manufacturer and manufacturer location


Wipe and spray A Glucoprotamin 26 g/100 g (1.5%) Incidin plus wipes Ecolab, Nieuwegein, The Netherlands
Wipe and spray B Hydrogen peroxide (Hi-speed H2O2): 15 mg/g (CAS 77-22-841) Aseptix Sterimax Aseptix, Loenen a/d Vecht, The Netherlands
Sporicide wipes
Wipe and spray C Ethanol 140 mg/g, propane-2-ol 100 mg/g; propane-1-ol 60 mg/g, Bacillol 30 tissues Hartmann, Nijmegen, The Netherlands
N-alkyl amino propyl glycine (CAS 1397 34-65-9) 5 mg/g
Spray D Didecyldimonium chloride, benzalkonium chloride, Formula 429 spray Formula 429, Amsterdam, The Netherlands
polyaminopropyl biguanide, dimethicone

*Active ingredient(s).

Table 2
care workers to apply these ready-to-use products. Because anti-
Mean log10 bacterial load reduction and mean log10 RLU reduction of cleaning-
microbial resistance is becoming a big threat for effective treating disinfection products with 95% CIs
of infections, cleaning and disinfection products are of great im-
Product Log10 CFU reduction Log10 RLU reduction
portance to reduce transmission between patients. Currently, there
is a lack of evidence that these products are truly effective in clean- Wipe A 5.77 (95% CI, 5.61-5.94) 1.98 (95% CI, 1.85-2.11)
Spray A 5.74 (95% CI, 5.58-5.90) 1.95 (95% CI, 1.82-2.08)
ing and disinfecting at the same time. The aim of this study was
Wipe B 5.58 (95% CI, 5.41-5.74) 2.27 (95% CI, 2.14-2.39)
to compare the effectiveness of commercially available products in Spray B 5.33 (95% CI, 5.16-5.49) 1.82 (95% CI, 1.69-1.94)
simultaneous cleaning and disinfection with 2 different applica- Wipe C 5.56 (95% CI, 5.40-5.73) 1.84 (95% CI, 1.71-1.97)
tion methods. Spray C 5.69 (95% CI, 5.53-5.85) 1.78 (95% CI, 1.65-1.91)
Spray D 5.72 (95% CI, 5.56-5.89) 1.60 (95% CI, 1.46-1.76)

MATERIALS AND METHODS CFU, colony forming units; CI, confidence interval; RLU, relative light unit.

Cleaning-disinfecting wipes and sprays


and allowed to react with the reagents in the cuvette for 10 seconds.
Seven cleaning-disinfecting wipes were obtained from differ- The swabs were placed into the Clean-Trace NG Luminometer (3M;
ent manufacturers. The wipes and sprays are currently used in health Neuss, North Rhine-Westphalia, Germany), and relative light units
care facilities around Europe, except spray D is not yet, to our knowl- (RLU) were recorded.
edge, available. Specifications of wipes and sprays are summarized To measure CFUs, the contaminated tile was wiped with a cloth
in Table 1. or sprayed and then wiped with a paper towel with the products
described in Table 2. A single technician performed all tests. A stan-
Bacteria isolates dardized sweeping technique was used, starting in the left upper
corner performing a meander-like pattern, with 4 turns, ending in
Vancomycin-resistant Enterococcus faecium (isolate from 2013 the right lower corner.
outbreak at Canisius-Wilhelminia Hospital, The Netherlands), Kleb- The tile was then placed into the neutralizer. The neutralizer con-
siella pneumoniae OXA-48 (isolate from 2011 outbreak at Maasstad sisted of lecithin 3 g/L, L-histidine 1 g/L, and saponin 30 g/L in diluent
Hospital, The Netherlands), and Acinetobacter baumannii (ATCC, (tryptone, pancreatic digest of casein 1.0 g/L, and sodium chloride
Rockville, MD) were used as test organisms. Strains were grown over- 8.5 g/L). After 2 minutes of rest in the neutralizer (10 mL), and 3
night at 37°C on blood agar. minutes of horizontal shaking (150 rpm) with glass beads (15 g,
5 mm), an aliquot of the suspension was plated on tryptic soy agar.
Efficacy of cleaning-disinfectant products in removal of After the incubation time of 24 hours at 37°C on tryptic soy agar,
microorganisms CFU were counted.
To measure ATP after the treatment of cleaning-disinfection, prod-
Bacterial isolates were suspended in physiologic saline and ad- ucts on the tile were swabbed again as previously described and
justed to a McFarland standard of 0.5. Test organisms were then then recorded.
added to 2 different test soils. All tests were performed in tripli-
cate, including a positive control per test organism. The colony Statistical analyses
forming units (CFU) found in the positive control were used for
analysis. Analysis of variance was used with the log10 reduction of CFU
The first test soil contained 3% bovine serum albumin with 0.3% and RLU as the dependent variable and bacteria, wipes A-C and
sheep erythrocytes, and the second test soil contained 12% bovine sprays A-D, and level of soil (3% and 12%) as the independent vari-
serum albumin with 10% sheep erythrocytes.9 The test solution used ables. Two-way interaction effects of bacteria, wipe-spray, and level
consisted of 1 mL of bacterial suspension, 0.2 mL of soil solution, of pollution were included in the model when statistically signif-
and 0.8 mL diluent. Standardized ceramic tiles (3709/PA00; Villeroy icant. Tukey method for multiple comparisons of means was used
& Boch, Mettlach, Saarland, Germany) measuring 5 × 5 cm were used to evaluate the differences between the different products and bac-
as the test surface. Tiles were sterilized before use. Tiles were con- teria. The results of the analysis of variance models are presented
taminated with 0.1 mL of test suspension. Test suspension was evenly using estimated marginal means, which are the predicted values
spread over the whole area of the tile and was left to dry for 1 hour of the dependent variable adjusted for the effects of the indepen-
at room temperature under laminar airflow. dent variables. Association between CFU and RLU was assessed using
To measure adenosine triphosphate (ATP), the tiles were swabbed Spearman rank correlation. All statistical analyses were carried out
with a consistent pattern (up and down, left to right while the swab in R version 3.1.1 (R Core Team, Vienna, Austria), and a 2-sided sig-
was rotated). The swabs were then reinserted into their container nificance level of .05 was used.
N. Kenters et al. / American Journal of Infection Control 45 (2017) e69-e73 e71

RESULTS 95% CI, 1.97-2.14) and A baumannii (log10, 1.90; 95% CI, 1.81-1.98).
The estimated marginal mean of RLU log10 reduction per bacteria
In this study, different cleaning-disinfection products were used is shown in Figure 2 for the 12% soil level (data for 3% soil level is
to determine their effect on CFU reduction of VRE, A baumannii, and not shown). Differences in efficacy in RLU log10 reduction are found
K pneumoniae OXA-48 outbreak isolates and to evaluate their effect between the different types of products (P < .001) (Table 2). Wipe
on RLU reduction. B was found to be most effective in removing soil, with a RLU log10
reduction of 2.27 (95% CI, 2.14-2.39), whereas spray D with a re-
CFU reduction duction of 1.60 (95% CI, 1.47-1.73) was least effective. In general, RLU
reduction showed no significant difference between the identical
The estimated marginal mean log10 CFU reduction for all prod- wipes and sprays (product A, P > .99; product B, P > .99). Only for
ucts, with both the bacteria and 12% soil level, is shown in Figure 1 product B was a significant difference found (P < .001), where spray
(data for 3% soil level is not shown). The CFU reduction for K B had a lower RLU reduction.
pneumoniae OXA-48 was larger (log10, 5.77; 95% confidence inter-
val [CI], 5.66-5.87) than for A baumannii (log10, 5.64; 95% CI, 5.54- Correlation between CFU and RLU
5.75) and VRE (log10, 5.46; 95% CI, 5.36-5.57). The different products
reached a 5 log10 CFU reduction against all tested microorganisms, Overall, no significant correlation between CFU and RLU was
with the exception of spray B (CFU log10 reduction, 4.43) with VRE. found. The Spearman rank correlation coefficient was 0.024 for the
The marginal estimated mean log10 reduction of the tested prod- 3% test soil (P = .85) and 0.04 for the 12% test soil (P = .73).
ucts is shown in Table 2. Wipe A had the highest log10 CFU reduction
of 5.77 (95% CI, 5.61-5.94) and spray B the lowest log10 CFU reduc- DISCUSSION
tion of 5.33 (95% CI, 5.16-5.49). In general, no significant difference
was found between the efficacy of the wipes and sprays with the Different factors affect the efficacy of ready-to-use cleaning-
same active ingredient (product A, P < .99; product B, P = .32; product disinfecting sprays or wipes. In this study, cleaning practices were
C, P = .93), with the exception of the in vivo efficacy of product B mimicked in vitro. Secretion, excretion, and blood are often found
against VRE, for which the spray had a significantly lower efficacy within the patient environment of health care facilities. These fluids
(P < .001). often contain microorganisms that can be harmful to health care
workers or patients. We executed a comparable experiment with
ATP reduction 2 different test soils, low and high soil content, to mimic all pos-
sible clinical situations. The product instructions were followed as
Different concentrations (3% and 12%) of test soil had no signif- described in the approved manual provided by the Board for the
icant impact on RLU reduction (P = .08). Authorisation of Plant Protection Products and Biocides, The Neth-
Using ATP measurements, a significant difference in log10 RLU erlands. The instructions provided for the wipes were to moisturize
reduction between the various bacteria (P < .001) was observed. On the surface properly with the wipe to keep the surface moistened
average, the RLU reduction of K pneumoniae OXA-48 (log10, 1.70; 95% for 5 minutes. The spray had to be used 30 cm from the surface and
CI, 1.62-1.79) was lower than the RLU reduction of VRE (log10, 2.06; then wiped with a paper towel or a cloth.

Fig 1. Colony forming unit reduction with high environmental contamination. Mean log10 bacterial removal from tiles examining efficacy of disinfecting cleaning wipes
and spray with a 12% test soil against vancomycin-resistant enterococci, Klebsiella pneumoniae OXA-48, and Acinetobacter baumannii. Data are the estimated mean of 3 trip-
licates, and bars represent 95% prediction intervals.
e72 N. Kenters et al. / American Journal of Infection Control 45 (2017) e69-e73

Fig 2. Adenosine triphosphate reduction (ATP) with high environmental contamination. Mean log10 ATP reduction from tiles examining efficacy of disinfecting-cleaning
wipes and spray with a 12% test soil against Acinetobacter baumannii, vancomycin-resistant enterococci, and Klebsiella pneumoniae OXA-48. Data are the estimated mean of
3 triplicates, and bars represent 95% prediction intervals.

For all bacteria, the tested ready-to-use cleaning-disinfection Ready-to-use cleaning-disinfection products have the poten-
products reduced the microbial count by >5 log10 with 5-minute ex- tial to increase cleaning compliance. Especially when compared
posure time, with the exception of a spray based on hydrogen with the traditional bucket method, which involves preparing sus-
peroxide. This hydrogen peroxide spray had an average 4.43 log10 pensions and carrying around a bucket with disinfecting solution.
CFU reduction on VRE. The test products are composed differently The ease of use of ready-to-use cleaning-disinfection products
and therefore are also charged differently, and the organic surfac- removes this barrier. In a brief study by Wiemken et al, a signifi-
tants may alter the physical properties of hydrogen peroxide in terms cant increase in cleaning compliance was shown with ready-to-
of spreadability. use cleaning-disinfection wipes.11 Regarding antimicrobial resistance,
Overall, our results show low variability of the different wipes basic cleaning and disinfection is increasingly important in clini-
and sprays in reducing bacterial counts. The Board for the Authori- cal areas. Standardized testing will provide users and infection
sation of Plant Protection Products and Biocides demands an efficacy control and prevention teams with more confidence that the prod-
of a 5 log10 reduction in microbial count to get approval for the use ucts work in accordance with the claims on the labels. Easier
of a cleaning-disinfecting product. methods for better compliance of cleaning and disinfection will
According to testing standards, we used paper towels with the contribute to infection control and prevention in clinical settings.
sprays. Because of the mechanical properties of the ready-to-use A study by Alfa et al showed a reduction of health care–acquired
wipes in comparison with paper towels, some differences could have infections when compliance of 80% was reached with cleaning-
been expected. Still, in a study by Diab-Elschahawi et al, no signif- disinfection wipes.12
icant difference was found in the decontamination efficacy of The results in this study show that the 2 outcome measures, bac-
microfibers, cotton cloths, and paper towels.9 tericidal effect as CFU reduction and protein residue as RLUs, did
Wipes-based products are available in ready-to-use packaging not correlate. The CFU reduction was highest for K pneumoniae for
or as a solution that needs to be prepared in a reusable container all tested products, whereas the RLU reduction was lowest for K
to impregnate the wipes. Ready-to-use wipes should be favored over pneumoniae. K pneumoniae produces a layer of polysaccharide, which
wipes in reusable tissue containers that need to be prepared. A study forms a biofilm on surfaces.13-15 The layer of polysaccharide helps
by Kampf et al found that neglecting adequate processing of the con- the bacteria to adhere to surfaces, which presumably makes it harder
tainers contributed to frequent and heavy contamination.10 In to clean a surface properly. E faecium had on average the lowest CFU
addition, there is a chance of preparing incorrect disinfection so- reduction, but the highest RLU reduction. E faecium is a gram-
lutions that are unnecessarily toxic or less effective. The sprays are positive bacterium, which possibly has a higher tolerating desiccation
all ready-to-use, but have different factors that may change the ef- against disinfectants. Although the exact reasons are unknown, the
ficacy (eg, application method [directly on the wipe or on the different bacterial properties obviously influence the results re-
surface]) and amount sprayed on a square centimeter. The usage garding CFU and RLU reduction.
of sprays could lead to increased improper usage of the product. In other studies, a correlation was shown between aerobic colony
No literature of standardized tests for cleaning-disinfection spray count and ATP when surfaces were contaminated with
is available to date. In addition, there is still a lack of research about microorganisms.16-18 Within the clinical setting, microbial testing
the safety implications for staff working with the spray. of surfaces is expensive and time consuming and therefore only used
N. Kenters et al. / American Journal of Infection Control 45 (2017) e69-e73 e73

in limited settings.17,19 However, ATP testing is quick and easy to prac- References
tice on hospital surfaces. Unfortunately, ATP meters from different
suppliers have different sensitivities and can also react with the de- 1. Walsh TR, Toleman MA. The emergence of pan-resistant Gram-negative
pathogens merits a rapid global political response. J Antimicrob Chemother
tergent, which makes it hard to set a threshold for this measurement. 2012;67:1-3.
Biocides often quench the ATP signal (indicating a low RLU), whereas 2. Weber DJ, Rutala WA. The role of the environment in transmission of Clostridium
the presence of an organic load may increase the RLU value sig- difficile infection in healthcare facilities. Infect Control Hosp Epidemiol
2011;32:207-9.
nificantly. Several quality standards have been used as RLU 3. Dancer SJ. Hospital cleaning in the 21st century. Eur J Clin Microbiol Infect Dis
thresholds, ranging from 100-500.16,17,20 A standardized quality stan- 2011;30:1473-81.
dard needs to be developed. The fact that the 2 methods did not 4. Carling PC, Parry MF, Bruno-Murtha LA, Dick B. Improving environmental hygiene
in 27 intensive care units to decrease multidrug-resistant bacterial transmission.
correlate in our study does not implicate that they are not both Crit Care Med 2010;38:1054-9.
helpful for the purpose of auditing cleaning within the health care 5. Carling PC, Parry MM, Rupp ME, Po JL, Dick B, Von Beheren S, et al. Improving
setting. cleaning of the environment surrounding patients in 36 acute care hospitals.
Infect Control Hosp Epidemiol 2008;29:1035-41.
The variety of different cleaning-disinfection products brought 6. Carling PC, Briggs JL, Perkins J, Highlander D. Improved cleaning of patient rooms
to market makes testing the effectiveness a very elaborate process. using a new targeting method. Clin Infect Dis 2006;42:385-8.
At the time the research was carried out, standardized methods for 7. Anderson DJ, Chen LF, Weber DJ, Moehring RW, Lewis SS, Triplett PF, et al.
Enhanced terminal room disinfection and acquisition and infection caused by
surface disinfectants using cloths were not available. Now, 2 stan-
multidrug-resistant organisms and Clostridium difficile (the Benefits of Enhanced
dards, the European test norm EN 16615 (4-field test) and the Terminal Room Disinfection study): a cluster-randomised, multicentre, crossover
international test norm ASTM E2967-15 (wiperator), are available. study. Lancet 2017;389:805-14.
Both methods demonstrate the compatibility of the cloth and the 8. Passaretti CL, Otter JA, Reich NG, Myers J, Shepard J, Ross T, et al. An evaluation
of environmental decontamination with hydrogen peroxide vapor for reducing
active substance of the cloth. In addition, the 4-field test can give the risk of patient acquisition of multidrug-resistant organisms. Clin Infect Dis
statements on the area’s performance and distribution into the en- 2013;56:27-35.
vironment. However, Pseudomonas aeruginosa used as a test organism 9. Diab-Elschahawi M, Assadian O, Blacky A, Stadler M, Pernicka E, Berger J, et al.
Evaluation of the decontamination efficacy of new and reprocessed microfiber
in the standard has a loss of CFU caused by drying that can influ- cleaning cloth compared with other commonly used cleaning cloths in the
ence the test results. hospital. Am J Infect Control 2010;38:289-92.
The pressure of the wiperator is currently not yet comparable 10. Kampf G, Degenhardt S, Lackner S, Jesse K, von Baum H, Ostermeyer C. Poorly
processed reusable surface disinfection tissue dispensers may be a source of
with real-life situations. In October 2015, a study by Sattar et al infection. BMC Infect Dis 2014;14:37.
evaluated the new ASTM standard test protocol to demonstrate 11. Wiemken TL, Curran DR, Pacholski EB, Kelley RR, Abdelfattah RR, Carrico RM,
efficacy of disinfectant wipes.21 Sattar et al states that the et al. The value of ready-to-use disinfectant wipes: compliance, employee time,
and costs. Am J Infect Control 2014;42:329-30.
developed standard is appropriate to assess the efficiency of 12. Alfa MJ, Lo E, Olson N, MacRae M, Buelow-Smith L. Use of a daily disinfectant
disinfection wipes. The wiperator ensures greater precision and cleaner instead of a daily cleaner reduced hospital-acquired infection rates. Am
reproducibility. J Infect Control 2015;43:141-6.
13. Podschun R, Ullmann U. Klebsiella spp. as nosocomial pathogens: epidemiology,
A missed opportunity of the standards is not including a stan-
taxonomy, typing methods, and pathogenicity factors. Clin Microbiol Rev
dardized testing method for sprays because they are increasingly 1998;11:589-603.
used in health care facilities. 14. Barreto S, Zambrano M, Araque M. [Phenotypic variations of susceptibility in
The testing methods used in ASTME2967-15 and EN 16615 Klebsiella pneumoniae strains of nosocomial origin and their association with
biofilm formation] [Spanish]. Invest Clin 2009;50:221-9.
mainly differ in, manual versus machine wiping, horizontal 15. Schroll C, Barken KB, Krogfelt KA, Struve C. Role of type 1 and type 3 fimbriae
wiping versus circling wiping, and as well in contact pressure and in Klebsiella pneumoniae biofilm formation. BMC Microbiol 2010;10:179.
the duration of the application on surfaces. Both test methods 16. Boyce JM, Havill NL, Havill HL, Mangione E, Dumigan DG, Moore BA. Comparison
of fluorescent marker systems with 2 quantitative methods of assessing terminal
should be used in the future by manufacturers to assess the cleaning practices. Infect Control Hosp Epidemiol 2011;32:1187-93.
disinfectant wipes to make more reliable claims. Clinical studies 17. Mulvey D, Redding P, Robertson C, Woodall C, Kingsmore P, Bedwell D, et al.
are needed to assess if using ready-to-use wipes leads to a Finding a benchmark for monitoring hospital cleanliness. J Hosp Infect
2011;77:25-30.
reduction in the transmission of several health care–associated 18. Boyce JM. Modern technologies for improving cleaning and disinfection of
pathogens. environmental surfaces in hospitals. Antimicrob Resist Infect Control 2016;5:10.
In general, the log reduction of CFU for all tested wipes and 19. Malik RE, Cooper RA, Griffith CJ. Use of audit tools to evaluate the efficacy of
cleaning systems in hospitals. Am J Infect Control 2003;31:181-7.
sprays was (with the mentioned exception) >5 log10 and therefore 20. Whiteley R, Jacobsen P, Prior S, Skazalski C, Otten R, Johnson A. Correlation of
appropriate for clinical use. Although ATP may show a difference isokinetic and novel hand-held dynamometry measures of knee flexion and
between pre- and postcleaning, RLU reduction does not correlate extension strength testing. J Sci Med Sport 2012;15:444-50.
21. Sattar SA, Bradley C, Kibbee R, Wesgate R, Wilkinson MA, Sharpe T, et al.
with actual CFU reductions. A clinical study in a health care setting
Disinfectant wipes are appropriate to control microbial bioburden from surfaces:
would be an asset to prove efficiency of cleaning-disinfection wipes use of a new ASTM standard test protocol to demonstrate efficacy. J Hosp Infect
and sprays. 2015;91:319-25.

You might also like