Service Manual

Humidifier VH-3000

Version: A01(1.09)

VADI Medical Technology Co., Ltd


5F-2, No. 492-19, Sec.1, Wan Shou Rd., Kwei Shan Hsiang, Taoyuan Hsien, Taiwan
Tel: +886-2-8209-2833
Fax: +886-2-8209-1152
website: www.vadimedical.com.tw

VADI Medical Technology Co. Ltd 1


Table of Content
1.0 Definition of Symbols
1.1 Introduction
1.2 Symbol
1.3 Condition of Transport and Storage

2.0 Specification
2.1 Mechanical
2.2 Electrical Specification
2.3 Standards & Approvals
2.4 Temperature Control
2.5 Temperature Measurement
2.6 Temperature Display

3.0 Mode & Control


3.1 Mode
3.1.1 Heater Wire Mode
3.1.1.1 Airway Temp. Set
3.1.1.2 Chamber Temp. Set
3.1.1.3 Warm up Mode
3.1.2 Non-Heater Wire Mode
3.1.3 Stand by Mode
3.2 Control -
3.2.1 Power Switch
3.2.2 Unlock
3.2.3 Heater Wire Mode
3.2.4 Silence

4.0 Alarm & Indicator


4.1 Heater Wire Mode -
4.1.1 Temp. Probe Alarm
4.1.2 Heater Wire Alarm
4.1.3 Airway Temp. High Alarm
4.1.4 Airway Temp. Low Alarm

VADI Medical Technology Co. Ltd 2


4.1.5 Chamber Temp. High Alarm
4.1.6 Chamber Temp. Low Alarm

4.2 Non-heater Wire Mode -


4.2.1 Chamber Temp. High Alarm
4.3 Overheating Protection –
4.3.1 Temperature Upper Limit on Heater Plate
4.3.2 Alarm of Overheating Temp. of Heater Plate
4.3.3 Temperature of Supply Cutout of Heater Plate
4.4 Audio Alarm
4.4.1 ISO9703-1/-2

5.0 Circuit
5.1 Diagram
5.2 Control Thermistors
5.3 Operation -
5.3.1 Primary Side -
5.3.1.1 PCB-ACPWR
5.3.1.2 AC/DC Power Module
5.3.2 Secondary Side -
5.3.2.1 PCB-MB
5.3.2.2 Control Panel
5.3.2.3 PCB-CNR (Board of Connection of Heater Wire & Temp. Probe)
5.3.3 Heater Plate -
5.3.3.1 Heater Pad
5.3.3.2 Thermostat
5.3.3.3 Thermistor
5.3.4 Heater Wire Cable
5.3.5 Heater Wire Capacity
5.3.6 Temp. Probe
5.4 Emergency of Power Cutout

6.0 Confirmation of Temp. Display & Measurement


6.1 Temperature Display on VH-3000 Humidifier
6.1.1 Requirement

VADI Medical Technology Co. Ltd 3


6.1.2 Procedure
6.2 Measurement of Temperature Probe -
6.2.1 Requirement
6.2.2 Procedure

7.0 Performance Check


7.1 Requirement
7.2 Installation
7.3 Temperature Setting
7.4 Items to be confirmed

8.0 Service
8.1 Introduction
8.2 Part Removal-
8.2.1 PCB and Control Panel
8.2.2 Power Module
8.2.3 PCB-CNR
8.3 Heater Plate –
8.3.1 Thermostat

9.0 Calibration
9.1 Appearance Inspection
9.2 Various Functional Tests
9.2.1 Control Panel and LED Tests
9.2.2 Input Rating Detection
9.2.3 Function Tests
9.2.3.1 Heating Function
9.2.3.2 Temperature Reading
9.2.3.3 Temperature Probe Connection Test
9.2.3.4 Heater Wire Adaptor Connection Test
9.2.3.5 Silence Function
9.2.3.6 Heater Plate Temperature Control
9.3 Temperature Probe Accuracy Test
9.4 Warning Buzzer Sound
9.5 Rated Input Current

VADI Medical Technology Co. Ltd 4


9.6 Burn-in Test

10.0 Trouble Shooting


10.1 Solution -
10.1.1 Low Temperature
10.1.2 High Temperature
10.1.3 Temperature change

Appendix A. Spare Parts List


Appendix B. Wiring Scheme

VADI Medical Technology Co. Ltd 5


1.0 Definition of Symbols ________________________________________
1.1 Introduction
VH-3000 humidifier is a device that is intended to add moisture and to warm the
breathing gases for administration to a patient. Chambers, gas heated, and prefilled
humidifiers are included in this device.

1.2 Symbol

Consult
Caution: High
operating
temperature
instruction

Power On/Off
Type BF
(Stand by)

Caution
Electrical Shock Protective earth
Hazard

Heater
Stand by
wire mode

Silence Unlock

Exclamation
Increase.
Mark

Decrease CE Mark

1.3 Condition of Transport and Storage


Condition of transport and storage: -10~50C, 5~95% RH
Recommended environmental pressure:700 hPa to 1060 hPa

VADI Medical Technology Co. Ltd 6


2.0 Specification ________________________________________________
2.1 Mechanical
Dimension : 135mm(W) *160mm(D)* 145mm(H)
Weight : 1.48Kg

2.2 Electrical Specification


Input Voltage: 100-120V or 200-240V / 350 VA
Frequency: 50/60 Hz Sine Wave
Heater Wire Supply: 24 ±2.4V
Heater Plate Output: 150 W at 110 V or 150W at 220V

2.3 Standards & Approvals


IEC-60601-1-1
Medical Electrical Equipment - Part 1: General requirements for basic safety
and essential performance.

IEC-60601-1-2
Medical electrical equipment, Part 1: General requirements for safety. 2. Collateral
standard: Electromagnetic compatibility - Requirements and tests

ISO-8185
Humidifiers for medical use - General requirements for humidification systems

2.4 Temperature Control


PID (Proportional Integral Derivative) control of Heater Wire (in Heater Wire mode
operation only) and Heater Plate.

2.5 Temperature Measurement


The temperature measurements of airway& chamber are performed by the thermistors
of temperature probe.

VADI Medical Technology Co. Ltd 7


2.6 Temperature Display
The temperature display is indicated by 6 pcs of 7-segment LED. The upper 3 pcs of
7-segment LED are used to display the temperature of airway. The rest of 7-segment
LED are used to display the chamber temperature.
Temperature Measurement Range: 10~70℃
(While <10℃, it shows “Lo ”)
(While >70℃, it shows “Hi”)
(While <28℃ and no temperature increase in 30 seconds, it shows “no Flo ”)
Measured Temp. Accuracy:25~45℃ ±0.3℃

VADI Medical Technology Co. Ltd 8


3.0 Mode & Control ________________________ _____________________
3.1 Mode
3.1.1 Heater Wire Mode
3.1.1.1 Airway Temp. Set
Choose the airway temp. set by pushing “+”and “-“ buttons. The range is
from 31 to 40℃ with a scale of 1℃. The airway temp. set is lit when the
temperature is settled

3.1.1.2 Chamber Temp. Set


Choose the Chamber temp. set by pushing “+”and “-“ buttons. The
range is from +2 to -5℃ with a scale of 1℃. The Chamber temp. set is lit
when the temperature is settled.
The chamber temp. is a sum of the airway temp. set and the chamber
temp. set.

3.1.1.3 Warm up Mode


After initial turn on, warm up mode is entered. In this mode, in the initial
10 min, Temp. Low will be disabled. The Temp. Low alarm line will be
gradually decreased at the rate of 0.3℃/min. About 75 min, the warm up
mode will be exited. Any change in setting will result in warm up mode.

3.1.2 Non-Heater Wire Mode


Press “Unlock” until the LED is lit. And then turn off “Heater Wire Mode”.
Press “+” or “-” on Airway temp. set to select the power level of heater
plate as shown as below:
Level 1 2 3 4 5 6 7 8 9
Temp. 45 48 52 56 60 64 68 71 75

3.1.3 Stand by Mode


While the function is activated, the temperature of heater plate will keep 40 ℃ at
most and the power of heater wire will held on 20% of power duty at most for 15
min. After 15 min, Stand by mode will be automatically exited.
To press Stand by mode again, resume original function.

VADI Medical Technology Co. Ltd 9


3.2 Control
3.2.1 Power Switch
Pushing the button to turn on/off the humidifier.
Note: The default is as shown as follows-
Mode: Heater Wire Mode
Airway Temp. Set: 39℃
Chamber Temp. Set: -2℃

3.2.2 Unlock
Press “unlock” until the LED is lit, that users can operate the humidifier.
The lock will automatically resume if no action is taken within 10 sec.

3.2.3 Heater Wire Mode


The button is used to switch heater wire mode or non-heater wire mode.

3.2.4 Silence
The button is used to turn off audio alarm for 120 sec. When the airway temp.
exceeds 42℃, silence is limited up to 30 sec.

VADI Medical Technology Co. Ltd 10


4.0 Alarm & Indicator_____________________________________________
※Note: All the alarms listed as follows pass the warm up mode.
4.1 Heater Wire Mode
4.1.1 Temp. Probe Alarm
Under Heater wire mode, the humidifier will alarm and the probe disconnected
indicators flashes. if temp. probe is disconnected or broken. In the meantime,
both airway and chamber temp. show “---“.
A temperature probe alarm immediately shuts down both heater wire and heater
plate.
Once the alarm condition has been removed, the humidifier will enter warm up
mode.

4.1.2 Heater Wire Alarm


Under heater wire mode, the humidifier will alarm and the cable disconnected
indicator flashes if heater wire or cable is disconnected or broken.
A heater wire alarm immediately shuts down both heater wire and heater plate.
Once the alarm condition has been removed, the humidifier will enter warm up
mode.

4.1.3 Airway Temp. High Alarm


The green LED display will flash and an audible alarm will sound if the gas
temperature being delivered to the patient is 2℃ higher than the airway temp.
set. A high temperature alarm instantly disables both heater circuits and heater
plate.
When airway temperature is greater than 42℃, the power to heater plate and
heater wire is cutoff simultaneously. It is also applied to warm up mode.

4.1.4 Airway Temp. Low Alarm


The green LED display will flash and an audible alarm will sound if the gas
temperature being delivered to the patient is 4℃ lower than the airway temp.
set. A low temperature alarm instantly disables both heater circuits and heater
plate.

VADI Medical Technology Co. Ltd 11


4.1.5 Chamber Temp. High Alarm
The reading on the Chamber Temp. will flash and an audible alarm will sound if
the gas temperature at chamber is 4℃ higher than the chamber temperature
set. A high temperature alarm instantly disables both heater circuits and heater
plate.

4.1.6 Chamber Temp. Low Alarm


The reading on the Chamber Temp. will flash and an audible alarm will sound if
the gas temperature at chamber is 4℃ lower than the chamber temperature set.
A low temperature alarm instantly disables both heater circuits and heater plate.

4.2 Non-heater Wire Mode


4.2.1 Chamber Temp. High Alarm
When temperature on the heater plate is over 110℃, the power to heater plate is
cutoff and the audio alarm sounds.

4.3 Overheating Protection


4.3.1 Temperature upper Limit on Heater Plate: 100℃
4.3.2 Alarm of Overheating Temp. of Heater Plate: 110℃. The power to heater plate is
disabled.
4.3.3 If the Temperature of Heater Plate greater than 120±7℃, the main active supply
is cutoff.

4.4 Audio Alarm


The audio alarms follow ISO 9703-1/-2.

VADI Medical Technology Co. Ltd 12


5.0 Circuit______________________________________________________
5.1 Diagram
Temp. Probe

Thermistor
Heated Wire
Wye
Heating Pad Thermistor

Heating Plate
Thermistor
Thermostat Power Cable
Chamber

Temp. Probe & Power Cable

Connection PCB

AC/DC Power Control


ACPWR Module PCB
Interface

Secondary
Primary

VH-3000

5.2 Control Thermistors


Three standard unique thermistors control all temperature functions:
 One thermistor is mounted in airway temp. probe at the end of the delivery hose to
measure temperature. This temperature is indicated by the LED display on the
humidifier control panel.
 Another thermistor mounted in chamber probe measures the temperature at the top
of the humidifying chamber. In heater wire mode this thermistor is used to monitor
and control the chamber temperature.
 The other thermistor is permanently mounted in intimate contact with the heater
plate, which prevents overheating.

VADI Medical Technology Co. Ltd 13


5.3 Operation
5.3.1 Primary Side
5.3.1.1 PCB-ACPWR
(1) Filter out the ripple in the mains.
(2) Supply power to heater plate for heating.

5.3.1.2 AC/DC Power Module


This is to provide power to secondary side.

5.3.2 Secondary Side


5.3.2.1 PCB-MB
This is to manage the setting of control interface, the alarm, the power
output and the signal from the thermistors.

5.3.2.2 Control Panel


This is an interface to let users operate the humidifier.

5.3.2.3 PCB-CNR (Board of Connection of Heater Wire & Temp. Probe)


There are 2 sockets for Temp. Probe and heater wire power output.
One of them is blue (6 Pins) for Temp. Probe, the other is yellow (4
Pins) for heater wire power output.

5.3.3 Heater Plate


5.3.3.1 Heater Pad
(1) 100~120V (50/60Hz): Resistance: 80Ω
(2) 200~240V (50/60Hz): Resistance: 320Ω

5.3.3.2 Thermostat
Cutout Temperature: 120±7℃
Reset: the reset button is black and located under the heater plate

5.3.3.3 Thermistor
Resistance: 10KΩ@25℃
B Value: 3435

VADI Medical Technology Co. Ltd 14


5.3.4 Heater Wire Cable
Specification -
(1) 1 to 2: Dual Heater Wire
(2) 1 to 1: Single Heater wire

5.3.5 Heater Wire Capacity


Specification –
Maximum load for Single Heater Wire: >8Ω
Maximum load for Dual Heater Wire: >8Ω

5.3.6 Temp. Probe


Resistance: 2.25KΩ@25℃
B Value: 3977
Accuracy: +/-0.3℃ (25℃~45℃)

Temperature/Resistance
℃ KΩ ℃ KΩ
30 1.814 36 1.411
31 1.739 37 1.355
32 1.667 38 1.301
33 1.598 39 1.249
34 1.533 40 1.199
35 1.471 41 1.152

5.4 Emergency of Power Cutout


Unplug the power cable to cutout the power if you can’t switch off the humidifier.

VADI Medical Technology Co. Ltd 15


6.0 Confirmation of Temp. Display & Measurement___________________
In order to avoid any deviation of temperature measurement and temperature display, please
perform the following tests to ensure the accuracy of the measurement.

6.1 Temperature Display on VH-3000 Humidifier


6.1.1 Requirement -
(1) Temperature Simulator (item no: VH30T000-4038)
(2) Heater Wire Power Cable
(3) Heater Wire

6.1.2 Procedure

Install temperature
Install Heater
Prepare the Install Heater simulator
Wire Power
humidifier Wire (VH30T000-4038) onto
Cable
temperature seat

Switch on the Turn the mode to See the display


humidifier Heater Wire Mode temperature

Chamber Temp. Airway Temp.


(38℃) (40℃)

Pass If not, Fail Pass If not, Fail


(37.9~38.1℃) (39.9~40.1℃)

If the humidifier is failed to pass the test, please call the distributor or agent
locally for service.

VADI Medical Technology Co. Ltd 16


6.2 Measurement of Temperature Probe
6.2.1 Requirement -
(1) A VH3000 Humidifier
(2) A Temperature Probe, item no VH-30T150
(3) A Accurate Thermometer (Accuracy ± 0.5 degrees)
(4) A container of water at approximately 40℃, or a stirred water bath at 40℃

6.2.2 Procedure
(1) Connect the temperature probe on VH-3000 humidifier.
(2) Place both the airway and chamber probes in a container of water (at
approx. 40℃).
(3) Put the accurate thermometer in the water.
(4) Make sure the water is constantly stirred, and wait approximately 30
seconds for the temperature to stabilize.
NOTE: If a probe fault condition exists, the humidifier will alarm. Then
replace a new one.
(5) The airway and chamber temp. will be displayed the measured reading.
(6) Compare the displayed temperature with the accurate thermometer. The
temperature difference(between the measured temperature and the
thermometer) should less than 1.5℃. If the difference is larger than 1.5℃,
the probe should be replaced.

VADI Medical Technology Co. Ltd 17


7.0 Performance Check__________________________________________
*Note: This check is to make sure VH-3000 humidifier work properly after service or half a
year.
7.1 Requirement -
(1) VH-3000
(2) Chamber
(3) Breathing Circuit
(4) Gas supply
(5) Temp. Probe
(6) Ambient Temp. controlled around 25℃

7.2 Installation -
(1) Use heater wire breathing circuit to link the chamber installed on the humidifier and
the gas supply.
(Please fill water into the chamber to the max. water level)
(2) Plug the temperature Probe.
(3) Place the probe on the chamber Outlet and airway probe seats on the breathing
circuit.
(4) Activate the gas supply and the suggested flow is 10±1LPM.
(5) Activate the humidifier and turn it into Heater Wire Mode.

7.3 Temperature Setting :


Airway Temp. Set: 39℃
Chamber Temp. Set: -2℃
Waiting time: 75 minutes after installation completed

7.4 Items to be confirmed


(1) Airway Temp.: 39±2℃
(2) Chamber Temp.: 37±2℃
(3) Heater Wire Alarm: Disconnect Heater Wire and humidifier will alarm and the cable
disconnected indicator flashes.
(4) Temp. Probe Alarm: Disconnect the probe and humidifier will alarm and the probe
disconnected indicators flashes. In the mean time, both airway and chamber temp.
will show “---“.
*Note: If one of the items fails, please contact local dealer or qualified service
engineers.
VADI Medical Technology Co. Ltd 18
8.0 Service_____________________________________________________
8.1 Introduction
This is to provide the methods of repairing the humidifier. The following parts are main
components: Please refer to appendix A.
(1) PCB-MB
(2) Power Module
PCB-ACPWR
PCB-VRM
AC/DC Power Supply
(3) PCB-CNR (Board of Connection of Heater Wire & Temp. Probe)
(4) Control Interface
* Caution -
A. Please make sure the power cable is removed from power outlet before service.
B. Electrostatic protect device must be equipped before service.
C. Please refer to section 6.0 and 7.0 to check after service.
D. Please beware of the temperature of heater plate before service.

8.2 Part Removal


8.2.1 PCB and Control panel
(1) The control panel and heater plate are separated by removing the 4 fixing
screws from the rear of the lower case

VADI Medical Technology Co. Ltd 19


(2) Remove all the wires that link to PCB.

(3) Separate the upper case and PCB(MB) by removing the 4 fixed screws

(4) Locate the flat cable linked PCB(MB) to control panel, and then remove the
flat cable by pressing the latch.

Caution: Do not damage the flat cable.

VADI Medical Technology Co. Ltd 20


8.2.2 Power Module
(1) Remove all the connectors from the module.
(2) Take the module and the lower case apart by removing the 4 fixed screws.

8.2.3 PCB-CNR (Board of Connection of Heater Wire & Temp. Probe)


Remove the cables on the board. And loose the blue and yellow plastic nuts to
take the board from the case.

8.3 Heater Plate


8.3.1 Thermostat
The thermostat will cut off if the temperature of the plate is over 120±6℃. Please
push the button on the thermostat to reset.

Manual Reset thermostat

VADI Medical Technology Co. Ltd 21


9.0 Calibration __________________________________________________
After Repair Table
Test Item Content After Repair
9.1 Appearance Inspection ●
9.2 Various Functional Tests
9.2.1 Control Panel and LED Tests ●
9.2.2 Input Rating Detection ●
9.2.3 Function Tests
9.2.3.1 Heating Function ●
9.2.3.2 Temperature Reading ●
9.2.3.3 Temperature Probe Connection Test ●
9.2.3.4 Heater Wire Adaptor Connection Test ●
9.2.3.5 Silence Function ●
9.2.3.6 Heater Plate Temperature Control ●
9.3 Temperature Probe Accuracy Test ●
9.4 Warning Buzzer Sound
9.5 Rated Input Current ●
9.6 Burn-in Test ●

Periodic Calibration Table


Periodic Calibration
Test Item Content
Six Months One Year
9.1 Appearance Inspection ●
9.2 Various Functional Tests
9.2.1 Control Panel and LED Tests
9.2.2 Input Rating Detection ●
9.2.3 Function Tests
9.2.3.1 Heating Function
9.2.3.2 Temperature Reading ●
9.2.3.3 Temperature Probe Connection Test
9.2.3.4 Heater Wire Adaptor Connection Test
9.2.3.5 Silence Function
9.2.3.6 Heater Plate Temperature Control ●
9.3 Temperature Probe Accuracy Test ●
9.4 Warning Buzzer Sound
9.5 Rated Input Current
9.6 Burn-in Test

VADI Medical Technology Co. Ltd 22


9.1 Appearance Inspection
If a deformation, damaged part occurred, the humidifier should be replaced by the new
cases.

9.2 Various Functional Tests


9.2.1 Control Panel and LED Tests
(1) After the humidifier turned on, inspect the entire panel LEDs for normal
indication.
(2) Wait for 10 seconds after the humidifier has turned on, and then push the
power switch to check to the current software version.
(3) Push the Unlock for one second, and it will enable the function.
(4) Under the heater wire mode, the light will turn on in normal circumstance.
(5) Check airway temp. set from 31 ~ 40℃, and adjust the temperature set
buttons up △ and down ▽.
(6) Check chamber temp. set from -5 ~ +2℃. Adjust the temperature set up △
and down ▽.
(7) “Stand by” LED will be on when turning on the STBY.
(8) The “Silence” LED will turn on when the function is on.
(9) Power on/off button will turn on or off the humidifier.
(10) When switching to non-heater wire mode, the light will be off.
(11) Under the non-heater wire mode, the LED will display a number in the
range from 1 to 9.
(12) Unlock will be off after 10 seconds, all buttons will not function except
“Silence” button.
(13) Plug in the power after waiting at least 10 seconds when unplugs the
power. When turning on, the status of the machine will remain on the last
setting.

Testing Purpose:
(1) Make sure all the wires are connect correctly with push of buttons.
(2) Make sure all the panel functions can be trigger.
(3) Make sure the memory function is operational.
VH-SK501 Functional Tester
1. Simulate temperature output
2. Simulate heater wire output
3. Simulate heater plate output

VADI Medical Technology Co. Ltd 23


9.2.2 Input Rating Detection
(1) Connect the AC power supply
( APS-9501 or APS-1102 or similar item ).
(2) Set the AC power supply output power setting at 110V/50Hz or 220V/
50Hz.
(3) Connect VH3000 with VH-SK501.
(4) Set the VH3000 at heater wire mode with +39 / -2 level.
(5) Simultaneous push the VH-SK501’s green and yellow buttons for test.
The testing result should be as follow:
AC Power Supply Reading
P PF
200 ~ 240W ≧ 0.8

Testing Purpose:
Make sure heater plate power setting and power module output are correct.

VADI Medical Technology Co. Ltd 24


9.2.3 Function Tests
9.2.3.1 Heating Function
Connect the AC power supply
( APS-9501 or APS-1102 or similar item )
Connect VH3000 with VH-SK501.
Set the VH3000 at heater wire mode with +39 / -2 level.
Turn VH-SK501’s temperature to 35℃.
VH-SK501’s green and yellow lights will flash.
Press VH-SK501’s green button means maximum power of heater
wire.
Press VH-SK501’s yellow button means maximum power of heater
plate.
AC Power Supply Reading
Heater Wire Maximum Output Heater plate Maximum Output
65 ~ 85W 140 ~ 160W

Heater Wire
Maximum Output
P = 77W

Heater plate
Maximum Output
P = 154W

VADI Medical Technology Co. Ltd 25


Testing Purpose:
Make sure heater wire I/O is normal
Make sure heater plate I/O is normal.

9.2.3.2 Temperature Reading


(1) Connect VH3000 with VH-SK501.
(2) Set at each temperature setting to check the temperature reading
is within specification ( ± 0.1℃ ).
Setting (℃) 31.0 34.0 37.0 41.0
VH-3000 Reading 30.9 ~ 31.1 33.9 ~ 34.1 36.9 ~ 37.1 40.9 ~ 41.1

Temperature Testing
Testing Purpose:
Use a VH-3000 humidifier to test temperature probe accuracy.
9.2.3.3 Temperature Probe Connection Test
By disconnect VH-SK501 temperature probe, it will trigger the alarm
and the VH3000’s temperature probe LED will flash.

9.2.3.4 Heater Wire Adaptor Connection Test


By disconnect VH-SK501 heater wire cable; it will trigger the alarm
and the VH3000’s heater wire adaptor LED will flash.

9.2.3.5 Silence Function


The alarm will be off after press the silence button.

VADI Medical Technology Co. Ltd 26


9.2.3.6 Heater Plate Temperature Control
(1) Set VH3000 to non-heater wire mode.
(2) Using the surface temperature measuring instrument
( SE3101 + K-TYPE temperature measuring thermocouple +
CHINO software or similar software ) to record heater plate’s
temperature at the level 1, 5, and 9.
(3) To get the accurate reading for every level, wait 10 minutes after
the heater plate has reach set temperature.

VH3000
1 2 3 4 5 6 7 8 9
Level
Temperature 45 ± 48 ± 52 ± 56 ± 60 ± 64 ± 68 ± 71 ± 75 ±
Range 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃

Heaterplate
temperature
testing.

Testing Purpose:
Make sure that heater plate temperature control is function and
stable.

VADI Medical Technology Co. Ltd 27


9.3 Temperature Probe Accuracy Test
(1) Prepare a thermometer that has the accuracy of within 0.5℃.
(2) Prepare a small container with 40℃ of distilled water.
(3) Prepare a normal VH3000.
(4) Prepare a temperature probe.
(5) Place the thermometer in the water and wait the water temperature to stable.
(6) Before connecting the temperature probe on the VH3000, set the VH3000 in
heater wire mode and Stand by mode.
(7) Put temperature probe inside the water. VH3000’s temperature reading will be
stable after 30 seconds.
(8) Compare the temperature difference between thermometer and VH3000’s
temperature reading.
(9) The difference must be under ±1.5℃, otherwise the temperature probe must be
replaced.

Comparison between
thermometer and
temperature probe.

Testing Purpose: Make sure the temperature probe accuracy is validated.

VADI Medical Technology Co. Ltd 28


9.4 Warning Buzzer Sound
Note: The user can test Buzzer function at any time, if it is necessary.
A “Bee” sound should be noticed when the Unlock is turned on in normal
circumstances.

Testing Purpose:
Make sure the warning buzzer is able to be audible under abnormal circumstances.

VADI Medical Technology Co. Ltd 29


9.5 Rated Input Current

Ventilator

VH30T000-4038

Clamp meter

Ventilator
Multi meter

VH30T000-4038

a. Use an instrument (such as clamp meter, multi meter, etc.) to measure current.
Current resolution must be 0.01A and tolerance <±2%.
b. Install the instrument mentioned above properly to measure input current of
VH-3000.
c. As the picture above, connect ventilator, breathing circuit (G-214000), and
temperature simulator (VH30T000-4038) to VH-3000.
d. Set VH-3000 at heater wire mode with 39/-2 level.
e. The measurement of maximum input current must be under 0.10A.
f. Set VH-3000 at heater wire mode with 39/+2 level.
g. As described below, to input the mains and find out maximum input current
accordingly:

VADI Medical Technology Co. Ltd 30


Input voltage(mains) 50/60 Hz Input current
90V < 1.25A
100V < 1.39A
VH3000-110 110V < 1.53A
120V < 1.67A
130V < 1.81A
180V < 0.63A
200V < 0.69A
VH3000-220 220V < 0.76A
240V < 0.83A
260V < 0.90A

Testing Purpose:
Make sure heater plate power output is correct.

h. Set VH-3000 at heater wire mode with 39/-2 level.


i. Enable Stand by mode.
j. Unplug the power cable ( VH30200 ) which is connected at the end of inspiratory
circuit. And connect the expiratory circuit. According to input voltage and find out
maximum input current shown as figure 5-1.
k. Unplug the power cable ( VH30200 ) which is connected at the end of expiratory
circuit. And connect the inspiratory circuit. According to input voltage and find out
maximum input current shown as figure 5-1.
l. Plug power cables ( VH30200 ) those are connected at the end of inspiratory and
expiratory circuit. According to input voltage and find out maximum input current
shown as figure 5-1.
( Figure 5-1 )
Input voltage (mains) 50/60 Hz Input current
VH3000-110 90 – 130V < 0.60A
VH3000-220 180 – 260V < 0.30A
Testing Purpose:
Make sure power module output is correct.

VADI Medical Technology Co. Ltd 31


9.6 Burn-in Test

Test Lung
Ventilator

a. As shown above, please connect VH3000 and breathing circuits to the ventilator
for simulation.
b. Set the ventilator flow for 10LPM.
c. Set the VH3000 at heater wire mode with +39 / -2 level.
d. As disconnect the temperature probe ( blue one ) which is connected on the side
of VH3000, temperature probe alarm will be sounded immediately.
e. As disconnect the heater wire cable ( yellow one ) which is connected on the side
of VH3000, heater wire cable alarm will be sounded immediately.
f. Please connect VH30T000-4038 to VH3000, temperature reading will show wye
at 40.0±0.1℃, chamber at 38.0±0.1℃.
VH30T000-4038

g. Connect back temperature probe and heater wire cable.


h. After one-hour burn-in test at least, temperature can be controlled stably at 39 ±
2℃ for wye, 37±2℃ for chamber.

Testing Purpose:
Use a VH-3000 humidifier to test temperature probe accuracy.
Make sure heater wire I/O is normal
Make sure heater plate I/O is normal.

VADI Medical Technology Co. Ltd 32


10.0 Trouble Shooting____________________________________________
10.1 Solution
10.1.1 Low Temperature
No. Problem Possible Cause Solution
During warm up mode Wait for 75 mins
Turn Stand by off and
Stand by mode on wait for 75 mins(warm
Temperature less than up mode)
1
setting, no alarm. Temperature Probe is not Install temperature
installed in place. probe in place.
Place humidifier at
Low ambient temperature.
18-26℃ environment.
Check whether the flow
The flow is too low or high.
is from 2-60 LPM
Temperature Probe is not Install temperature
Airway Temperature installed in place. probe in place.
2
alarm activated The distance between
The airway temperature
temperature probe and
probe is far away from
heater wire should be
heated wire.
from 25mm to 100mm.
The contact of the surface Check whether heater
of humidification chamber plate surface is clean
doesn’t contact heating and even.
plate well.
the temperature probe is Install probe in place.
Chamber Temperature
3 not installed in place.
alarm activated
Please contact your
The heating pad in heater
local distributor for
plate is malfunction.
service.
The flow is higher or lower Check whether the
for chamber. chamber is correct.

VADI Medical Technology Co. Ltd 33


10.1.2 High Temperature
No. Problem Possible cause Solution
Push the silence button
Sudden flow changed and wait for temp. fall
below limit.
Install longer or shorter
heater wire in
The airway temperature
inspiratory part. The
Airway temperature sensor of the temperature
1 distance between
alarm activated probe is too close to heater
temperature probe and
wire.
heater wire should be
from 25mm to 100mm.

Temp. probe broke Replace a new one

Push the silence button


Flow is not stable and
and wait for
changes suddenly.
temperature decreases.

Humidification Add the water in the

2 chamber temperature No water in the chamber chamber and wait for

high the warm up time.

Replace a new one.


Temperature probe broke .

VADI Medical Technology Co. Ltd 34


10.1.3 Temperature change
No. Problem Possible cause Solution
Consider to re-adjust
The setting of breathe rate on
the rate and mode on
ventilator is not appropriate.
ventilator.
Please adjust the
Influence from ambient
position of humidifier to
temperature.( Such as under
avoid the direct flow
the vent of air-con.)
Airway Temperature from vent.
1
alarm Check probe on
The temperature probe is not
breathing circuit in
well seated.
correct position.

Defective temperature probe. Replace a new one.

VADI Medical Technology Co. Ltd 35


Appendix A.
Spare Parts List of VH-3000 Humidifier
Item No. Description
700-VH3000UPCASE VH3000-110 Upper Case
1
700-VH3000UPCASE.2 VH3000-220 Upper Case
2 200-VH3000-04 Connection Cable (Heated Wire)
3 200-VH3000-03 Connection Cable (Temperature Probe)
4 700-43002 PCB-CNR
5 200-VH3000-06 Connection Cable (ACPWR)
200-41010-1 Power Cable (AUS)
6 200-41010-2 Power Cable (EUR)
200-41010-4 Power Cable (US)
7 300-41007 Screw M4*12
8 700-VH3000PWR.SNPG129 Power Module
9 100-41006 Power Cable Fixer
10 100-41005-2 Lower Case

VADI Medical Technology Co. Ltd 36


Appendix B.
Wiring scheme

VADI Medical Technology Co. Ltd 37