Culture Documents
Humidifier VH-3000
Version: A01(1.09)
2.0 Specification
2.1 Mechanical
2.2 Electrical Specification
2.3 Standards & Approvals
2.4 Temperature Control
2.5 Temperature Measurement
2.6 Temperature Display
5.0 Circuit
5.1 Diagram
5.2 Control Thermistors
5.3 Operation -
5.3.1 Primary Side -
5.3.1.1 PCB-ACPWR
5.3.1.2 AC/DC Power Module
5.3.2 Secondary Side -
5.3.2.1 PCB-MB
5.3.2.2 Control Panel
5.3.2.3 PCB-CNR (Board of Connection of Heater Wire & Temp. Probe)
5.3.3 Heater Plate -
5.3.3.1 Heater Pad
5.3.3.2 Thermostat
5.3.3.3 Thermistor
5.3.4 Heater Wire Cable
5.3.5 Heater Wire Capacity
5.3.6 Temp. Probe
5.4 Emergency of Power Cutout
8.0 Service
8.1 Introduction
8.2 Part Removal-
8.2.1 PCB and Control Panel
8.2.2 Power Module
8.2.3 PCB-CNR
8.3 Heater Plate –
8.3.1 Thermostat
9.0 Calibration
9.1 Appearance Inspection
9.2 Various Functional Tests
9.2.1 Control Panel and LED Tests
9.2.2 Input Rating Detection
9.2.3 Function Tests
9.2.3.1 Heating Function
9.2.3.2 Temperature Reading
9.2.3.3 Temperature Probe Connection Test
9.2.3.4 Heater Wire Adaptor Connection Test
9.2.3.5 Silence Function
9.2.3.6 Heater Plate Temperature Control
9.3 Temperature Probe Accuracy Test
9.4 Warning Buzzer Sound
9.5 Rated Input Current
1.2 Symbol
Consult
Caution: High
operating
temperature
instruction
Power On/Off
Type BF
(Stand by)
Caution
Electrical Shock Protective earth
Hazard
Heater
Stand by
wire mode
Silence Unlock
Exclamation
Increase.
Mark
Decrease CE Mark
IEC-60601-1-2
Medical electrical equipment, Part 1: General requirements for safety. 2. Collateral
standard: Electromagnetic compatibility - Requirements and tests
ISO-8185
Humidifiers for medical use - General requirements for humidification systems
3.2.2 Unlock
Press “unlock” until the LED is lit, that users can operate the humidifier.
The lock will automatically resume if no action is taken within 10 sec.
3.2.4 Silence
The button is used to turn off audio alarm for 120 sec. When the airway temp.
exceeds 42℃, silence is limited up to 30 sec.
Thermistor
Heated Wire
Wye
Heating Pad Thermistor
Heating Plate
Thermistor
Thermostat Power Cable
Chamber
Connection PCB
Secondary
Primary
VH-3000
5.3.3.2 Thermostat
Cutout Temperature: 120±7℃
Reset: the reset button is black and located under the heater plate
5.3.3.3 Thermistor
Resistance: 10KΩ@25℃
B Value: 3435
Temperature/Resistance
℃ KΩ ℃ KΩ
30 1.814 36 1.411
31 1.739 37 1.355
32 1.667 38 1.301
33 1.598 39 1.249
34 1.533 40 1.199
35 1.471 41 1.152
6.1.2 Procedure
Install temperature
Install Heater
Prepare the Install Heater simulator
Wire Power
humidifier Wire (VH30T000-4038) onto
Cable
temperature seat
If the humidifier is failed to pass the test, please call the distributor or agent
locally for service.
6.2.2 Procedure
(1) Connect the temperature probe on VH-3000 humidifier.
(2) Place both the airway and chamber probes in a container of water (at
approx. 40℃).
(3) Put the accurate thermometer in the water.
(4) Make sure the water is constantly stirred, and wait approximately 30
seconds for the temperature to stabilize.
NOTE: If a probe fault condition exists, the humidifier will alarm. Then
replace a new one.
(5) The airway and chamber temp. will be displayed the measured reading.
(6) Compare the displayed temperature with the accurate thermometer. The
temperature difference(between the measured temperature and the
thermometer) should less than 1.5℃. If the difference is larger than 1.5℃,
the probe should be replaced.
7.2 Installation -
(1) Use heater wire breathing circuit to link the chamber installed on the humidifier and
the gas supply.
(Please fill water into the chamber to the max. water level)
(2) Plug the temperature Probe.
(3) Place the probe on the chamber Outlet and airway probe seats on the breathing
circuit.
(4) Activate the gas supply and the suggested flow is 10±1LPM.
(5) Activate the humidifier and turn it into Heater Wire Mode.
(3) Separate the upper case and PCB(MB) by removing the 4 fixed screws
(4) Locate the flat cable linked PCB(MB) to control panel, and then remove the
flat cable by pressing the latch.
Testing Purpose:
(1) Make sure all the wires are connect correctly with push of buttons.
(2) Make sure all the panel functions can be trigger.
(3) Make sure the memory function is operational.
VH-SK501 Functional Tester
1. Simulate temperature output
2. Simulate heater wire output
3. Simulate heater plate output
Testing Purpose:
Make sure heater plate power setting and power module output are correct.
Heater Wire
Maximum Output
P = 77W
Heater plate
Maximum Output
P = 154W
Temperature Testing
Testing Purpose:
Use a VH-3000 humidifier to test temperature probe accuracy.
9.2.3.3 Temperature Probe Connection Test
By disconnect VH-SK501 temperature probe, it will trigger the alarm
and the VH3000’s temperature probe LED will flash.
VH3000
1 2 3 4 5 6 7 8 9
Level
Temperature 45 ± 48 ± 52 ± 56 ± 60 ± 64 ± 68 ± 71 ± 75 ±
Range 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃ 3℃
Heaterplate
temperature
testing.
Testing Purpose:
Make sure that heater plate temperature control is function and
stable.
Comparison between
thermometer and
temperature probe.
Testing Purpose:
Make sure the warning buzzer is able to be audible under abnormal circumstances.
Ventilator
VH30T000-4038
Clamp meter
Ventilator
Multi meter
VH30T000-4038
a. Use an instrument (such as clamp meter, multi meter, etc.) to measure current.
Current resolution must be 0.01A and tolerance <±2%.
b. Install the instrument mentioned above properly to measure input current of
VH-3000.
c. As the picture above, connect ventilator, breathing circuit (G-214000), and
temperature simulator (VH30T000-4038) to VH-3000.
d. Set VH-3000 at heater wire mode with 39/-2 level.
e. The measurement of maximum input current must be under 0.10A.
f. Set VH-3000 at heater wire mode with 39/+2 level.
g. As described below, to input the mains and find out maximum input current
accordingly:
Testing Purpose:
Make sure heater plate power output is correct.
Test Lung
Ventilator
a. As shown above, please connect VH3000 and breathing circuits to the ventilator
for simulation.
b. Set the ventilator flow for 10LPM.
c. Set the VH3000 at heater wire mode with +39 / -2 level.
d. As disconnect the temperature probe ( blue one ) which is connected on the side
of VH3000, temperature probe alarm will be sounded immediately.
e. As disconnect the heater wire cable ( yellow one ) which is connected on the side
of VH3000, heater wire cable alarm will be sounded immediately.
f. Please connect VH30T000-4038 to VH3000, temperature reading will show wye
at 40.0±0.1℃, chamber at 38.0±0.1℃.
VH30T000-4038
Testing Purpose:
Use a VH-3000 humidifier to test temperature probe accuracy.
Make sure heater wire I/O is normal
Make sure heater plate I/O is normal.