Intended Use
2
Contents
3
Warning and Safety Information
Combics complies with the European – Note on installation: Using the Equipment in Legal
Council Directives as well as inter- The operator shall be responsible Metrology:
national regulations and standards for for any modifications to Sartorius – When the indicator is connected to
electrical equipment, electromagnetic equipment and must check and, if a weighing platform and the resulting
compatibility, and the stipulated safety necessary, correct these modifications. weighing instrument is to be verified,
requirements. On request, Sartorius will provide make sure to observe the applicable
information on the minimum operating regulations regarding verification.
§ To prevent damage to the equipment, specifications (in accordance with Make sure to read the brochure stored
read these operating instructions the Standards for defined immunity to on the enclosed CD first: “Guide to
thoroughly before using your Combics. interference). Verification of Weighing Instruments."
When connecting a Sartorius weighing
! Do not use the equipment in a $ If there is visible damage to the platform, make sure to observe the
hazardous location. equipment or power cord: unplug the permitted weighing ranges as listed in
equipment and lock it in a secure place the Declaration of Conformity.
! The indicator may be opened only by to ensure that it cannot be used for
authorized service technicians who the time being – If any of the verification seals are
have been trained by Sartorius and who damaged, make sure to observe the
follow Sartorius’ standard operating Industrial protection ratings for national regulations and standards
procedures for maintenance and repair. the housing: applicable in your country in such
– CISL models are rated to IP44 cases. In some countries, the
! Make absolutely sure to unplug the (IP65 with option L1) verification will become null and void
indicator from power before you – CIS models are rated to IP67 and the equipment must be re-verified.
connect or disconnect any electronic
peripheral devices to or from the – Connect only Sartorius accessories and
interface port. options, as these are optimally designed
for use with your Combics.
! If you use electrical equipment in
installations and under ambient – Do not expose the indicator to
conditions requiring higher safety aggressive chemical vapors or to
standards, you must comply with extreme temperatures, moisture,
the provisions as specified in the shocks, or vibration.
applicable regulations for installation
in your country. – Clean your Combics only in accordance
with the cleaning instructions (see
– Warning When Using Pre-wired RS-232 “Care and Maintenance").
Connecting Cables! RS-232 cables
purchased from other manufacturers $ If you have any problems with your
often have incorrect pin assignments Combics:
for use with Sartorius weighing systems. contact your local Sartorius office,
Be sure to check the pin assignments dealer or service center.
against the chart below before
connecting the cable, and disconnect
any lines identified differently from NEMA Protection rating on CIS
those specified by Sartorius. models:
– Protection of the indicators according
– Use only extension cords that meet the to NEMA is ensured only if the rubber
applicable standards and have a protec- gasket is installed and all connections
tive grounding conductor. are fastened securely (including the
caps on unused sockets). Weighing plat-
– Disconnecting the ground conductor is forms must be installed and tested by a
prohibited! certified technician.
– If you install an interface port or
battery connector after setting up your
Combics, keep the protective cap(s) in
a safe place to be used for protecting
the interface port or battery connector
when not in use, or prior to shipment.
This will protect the data interface or
battery connector from vapors, moisture
and dust or dirt.
4
Getting Started
The indicator is available in various ver- Equipment Supplied Conditioning the Indicator:
sions. If you have ordered special – Indicator Moisture in the air can condense on the
options, the indicator will be supplied surfaces of a cold indicator whenever
with these options premounted at the – Operating Instructions (this manual) it is brought to a substantially warmer
factory. place. If you transfer the indicator to
– Special accessories listed on the bill a warmer area, make sure to condition it
Storage and Shipping Conditions of delivery, if ordered for about 2 hours at room temperature,
Allowable storage temperature: leaving it unplugged from AC power.
-10 ...+40°C (+14°F ... + 104°F) Setting Up the Indicator Afterwards, if you keep the indicator
The indicator is designed to provide connected to AC power, the constant
– Unpackaged equipment may lose reliable weighing results under normal positive difference in temperature
its accuracy when exposed to strong ambient conditions. When choosing between the inside of the indicator and
vibration. Excessive vibration may a location to set up your indicator, the outside will practically rule out the
compromise the safety of the observe the following so that you will effects of moisture condensation.
equipment. be able to work with added speed and
– Do not expose the indicator accuracy: Seal on Indicators Verified for Use
unnecessarily to extreme temperatures, in Legal Metrology in the EU*:
moisture, shocks, or vibration. – Avoid placing the indicator in close EU legislation requires that a control
proximity to a heater or otherwise seal be affixed to the verified device.
Unpacking exposing the indicator to heat or direct The control seal consists of a sticker
§ After unpacking the equipment, sunlight. with the “Sartorius" logo. This seal will
please check it immediately for any be irreparably damaged if you attempt
visible damage. – Protect the indicator from drafts that to remove it. If the seal is broken, the
come from open windows or doors. validity of the version becomes null
$ If you detect any damage, proceed and void, and you must have your scale
as directed in the chapter entitled – Avoid exposing the indicator to extreme re-verified.
“Care and Maintenance” under “Safety vibrations during weighing.
Inspection." * including the Signatories of
– Protect the indicator from aggressive the Agreement on the European
$ If you will need to ship your indicator chemical vapors. Economic Area
later, save all parts of the packaging
because only the original packaging – Do not expose the indicator to extreme
provides the best protection for moisture over long periods.
shipment.
Turn off the power if you do not
$ Before shipping, be sure to disconnect need to use the indicator with other
all cables to prevent damage. equipment.
Note:
Warranty This equipment has been tested and
Do not miss out on the benefits of our found to comply with the limits for
full warranty. Please contact your local a Class A digital device, pursuant to
Sartorius office or dealer for further Part 15 of the FCC rules. These limits
information. If available, complete the are designed to provide reasonable
warranty registration card, indicating protection against harmful interference
the date of installation, and return the when the equipment is operated in
card to your Sartorius office or dealer. a commercial environment. This
equipment generates, uses and can
radiate radio frequency energy and, if
not installed and used in accordance
with the instruction manual, may
cause harmful interference to radio
communications. Operation of this
equipment in a residential area is likely
to cause harmful interference in which
case the user will be required to correct
the interference at his own expense.
Changes or modifications not expressly
approved by Sartorius AG could void
the user's authority to operate the
equipment.
5
General View of the Equipment
10
9
8
7 4
6 5
16
15
14
6
Installing the Weighing Platform
The connecting cable should be installed by a certified technician who has received special
training from Sartorius.
!Installation work that affects the NEMA rating must be performed with extreme care.
!Any installation work that does not conform to the instructions in this manual will result
in forfeiture of all claims under the manufacturer’s warranty
!Always make sure the equipment is disconnected from AC power before performing
any installation, maintenance or repair work!
!The cable gland (NEMA rating) is already installed on the indicator. Please use extreme
caution when performing any work on the equipment that affects this cable gland!
§ Use the connecting cable from the weighing platform to connect the indicator.
!The cable gland (NEMA rating) is already installed on the indicator. Please use extreme
caution when performing any work on the equipment that affects this cable gland.
Use a torque wrench. Torque applied for this cable gland: 5 Nm.
§ Strip off the insulation at the cable end and attach the cable as follows:
– Route the cable through the cable gland.
– Properly tighten the screw fasteners of the cable gland.
– Remove the casing from a section of the cable end (see illustration). The shield (1)
must have contact with the clamps (2).
7
§ Attach the cable to the weighing platform as follows:
– Expose approximately 5 cm (2 inches) of the wires for installation.
– Strip the casing from approximately 1 cm (1/2 inch) of the wires and attach ferrules
to the wires.
– Install the ferrite ferrule over wires 1 to 6
– Securely attach the wires to the screw terminals.
Wiring Diagram
No. Signal name Meaning
1 BR_POS Bridge supply voltage, positive
2 SENSE_POS Sense (+)
Bridge supply voltage
3 OUT_POS Measuring voltage, positive
4 OUT_NEG Measuring voltage, negative
5 SENSE_NEG Sense (-)
Bridge supply voltage
6 BR_NEG Bridge supply voltage, negative
!Note:
Please refer to the data sheet or operating instructions for the load cell or weighing platform
in question for details on the assignment of wire colors to signals. Disconnect any lines that
are not used.
$ The indicator is powered through the pre-installed power cord. The power supply is built
into the indicator, which can be operated with a supply voltage of 100V to 240V.
Make sure that the voltage rating printed on the manufacturer´s ID label is identical to that
of your local line voltage. If the voltage specified on the label or the plug design of the
AC adapter does not match the rating or standard you use, please contact your Sartorius
office or dealer. The power connection must be made in accordance with the regulations
applicable in the U.S.
§ To power a protective class 1 device, plug the power cord into an electrical outlet that is
properly installed with a protective grounding conductor.
Safety Precautions
If your local AC output does not have a protective grounding conductor have a certified
electrician install equivalent protection according to installation requirements. You may not
neutralize the protective grounding effect by using an extension cord that lacks a protective
grounding conductor.
Warmup Time
To deliver exact results, the device must warm up for at least 30 minutes after initial
connection to AC power or after a relatively long power outage. Only after this time will
the device have reached the required operating temperature.
!The indicator may be opened only by authorized service technicians who have been
trained by Sartorius and who follow Sartorius’ standard operating procedures for
maintenance and repair.
8
Installing the Verification Adapter for Use in Legal Metrology
(on verfiable models only)
! Disconnect the indicator from power (unplug the power cord from the wall socket)
§ Installation
CISL1 | CISL2 models: Connect the 25-pin D-Sub connector (connecting cable
YCC02-RB01) from the battery pack to the second port (“UNICOM")
CIS1 | CIS2 models: Please see “Pin Assignment Chart" in this manual (via connecting
cable YCC02-RB02 or using option L2)
Display symbols:
Fully charged:
Completely drained:
! Disconnect the indicator from power (unplug the power cord from the wall socket)
$ Installation
CISL2 models:
– Connect the 25-pin D-Sub connector from the battery pack to the second port
(“UNICOM")
– To connect both the bar code scanner and the external battery pack:
Use the YTC01 T-connector
CIS2 models: Please see “Pin Assignment Chart" in this manual (via connecting cable
YCC02-BR02 or using option 118)
9
Operating Design
10
Operating Elements: Combics 1 (Combics 2: see next page)
11
Operating Elements: Combics 2 (for Combics 1, see previous page)
Keys to the Right of the Display Input Through the Digital Input/ Error Codes and Messages
c r O w Output Interface If a key is inactive, this is indicated as
There are four output control lines and follows:
These keys are used for initializing one input line for use with the check- – The error code “-------" is displayed
applications, and are described in weighing and classification applications. for 2 seconds. The display then returns
detail in the following chapters. You can assign one of the following to the previous screen content.
functions to the input control line – An acoustic signal (double-beep)
Input Through the Weighing (configuration in the operating menu): is emitted.
Platform
You can store the weight on the – p key Dynamic errors are indicated by an
weighing platform as a tare weight. – p key (> 2 sec) error code (e.g. inf 09) until the error
is corrected.
Input Through a Bar Sode Scanner – ) key Temporary errors are indicated for
(Combics 2 only) – ) key (> 2 sec) 2 seconds.
You can use a bar code scanner to – k key
read in the reference weight for the – n key “Fatal" errors are displayed continuously
active application (counting, weighing (until you restart the Combics).
in percent, neutral measurement). Data – O key
input through the bar code scanner is The same applies in all operating
handled by the Combics in the same modes (weighing, application programs,
manner as keypad input. Input Through the COM Port operating menu).
In general, you can use all bar code The Combics is equipped with an SBI Error codes and messages are described
formats that are compatible with the interface for data transfer. You can in detail in the chapter entitled “Error
connected bar code scanner. define certain parameters for this Codes and Messages."
interface (generate printout, time-
dependent auto print, ID codes).
This is a simple ASCII interface. The
functions are described in detail in
the chapter entitled “Operating
the Combics," under “Data Output
Functions.” You can define parameters
for the interface (resolution of the
output, time-dependent automatic
output, ID)
12
123 4 5 6 7 8 9 10 11 12 1 2 3 4
2 2
124853 10
printer 10
13 14 15 16 15 17
1 Symbol for the routine in progress 11 R1, R2 for multiple range mode 1 Selected menu item (in this example,
2 O sign for the value displayed 12 Active weighing platform "Printer," for configuring the
3 Indicates that the weighing platform (Combics 2 only) connected printer)
has been zeroed 13 Bar graph 2 Indicates that the menu has
4 Weight or calculated value 14 Program symbols submenu(s)
(main display) (for checkweighing, classification) 3 Indicates that this is the currently
5 This digit is not valid when the 15 Application symbols (Combics 2 only) selected setting
equipment is used in legal metrology 16 Auto: Reaction of application 4 Highest menu level
6 Weight unit of the value displayed program triggered by weight value;
7 GMP printing mode active Opt: Reference sample updating was
8 Data in tare memory performed
9 Gross value shown in main display 17 Numeric display (e.g., reference value)
10 Printing mode active 18 Battery symbol
13
Data Output Functions Digital Input/Output Interface
You can choose from four forms of The digital I/O interface is supported
data output: by the Checkweighing and Classification
applications.
– Printer
– Digital I/O interface Checkweighing
– COM port Four data outputs transfer the following
– LEDs information on the weight values: “less
than," “equal to," “greater than" and
Printer “set." You can define whether the out-
You can connect one or two strip put lines are always active, or are acti-
printers or one or two label printers to vated only at stability, only within the
the Combics. If you use a YDP02, checkweighing range, only within the
YDP03, or Universal printer, you can checkweighing range at stability, or not
configure certain parameters for the at all (“off"); please see page 52.
printer (such as baud rate, parity, stop
bits, handshake mode and data bits). Classification
Four data outputs transfer the following
The printout can be formatted by the information on the weight values:
user. The printout should be formatted “less than", “equal to", “greater than"
only after the desired application and “set". You can define whether the
has been configured, as some of the output lines are always active, or are
printout positions are application- activated only at stability (or are “off");
dependent. please see page 58.
Press the p key to print the settings – “Set” if you are using the Checkweighing
of the current menu level on a strip application or “exceed a minimum load”
printer or a universal printer. All if you are using the Classification
submenus under the current menu application.
level are included on the printout.
In all other applications and for
Combics 2 only: For the Totalizing and Combics 1, the “set” output indicates
Net-total Formulation applications, you that the indicator is ready to operate.
can also configure printouts of total
(results) and individual or component COM Port
values. The Combics is equipped with an SBI
interface for data transfer. You can
With a strip printer, universal printer or define certain parameters for this
a label printer, you can define whether interface (generate printout, time-
the printout includes a GMP header and dependent auto print, ID codes).
GMP footer (field for operator signature).
With a label printer, this function This is a simple ASCII interface.
requires the menu setting for ISO/GMP The functions are described in detail
printouts: Always for one measurement in the chapter entitled “Operating
result (7.11.2). GMP stands for Good the Combics," under “Data Output
Manufacturing Practice. Functions.”
LEDs
The Combics 2 has 3 LEDs on the
right side of the display window. These
are for use with the Checkweighing and
Classification applications. This check-
weighing display shows the relationship
of the current weight value to the
tolerance limits in checkweighing;
with the classification application, they
indicate how the weight value is
classified.
14
Settings
1.18
1.18.1
CAL.
50.00
etc.
15
Navigating in the Operating Menu (Examples)
Example: Adapting the Combics to the place of installation, menu code 1. 1. 4, “Extreme vibration", for weighing platform WP1
(in this example, using the Combics 2).
)
Display for selecting the application APPL is shown
16
( Return to next higher menu level
)
Display for selecting the application APPL is shown
17
) Show user password menu item
18
Operating Menu Overview (Parameters)
= Setting|function on Combics 2 only
* Factory setting
W User-defined setting
Menu
Z* Counting
3. 5. Minimum load for automatic taring and automatic printout
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
19
Z* Counting
3. 9. Resolution for calculation of average piece weight
3. 9. 1 * Display accuracy
3. 9. 2 Display accuracy + 1 decimal place
3. 9. 3 Display accuracy + 2 decimal places
3. 9. 4 Internal resolution
Z nm Neutral Measurement
3. 5. Minimum load for automatic taring and automatic printout
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
20
Z nm Neutral Measurement
3.10. Decimal places for display of results
3.10. 1 * None
3.10. 2 1 decimal place
3.10. 3 2 decimal places
3.10. 4 3 decimal places
21
V Averaging (Animal Weighing)
3.19. Animal activity
3.19. 1 0.1 % of the animal/object
3.19. 2 * 0.2% of the animal/object
3.19. 3 0.5 % of the animal/object
3.19. 4 1% of the animal/object
3.19. 5 2% of the animal/object
3.19. 6 5% of the animal/object
3.19. 7 10% of the animal/object
3.19. 8 20% of the animal/object
3.19. 9 50% of the animal/object
3.19.10 100 % of the animal/object
O Checkweighing
3. 5. Minimum load for automatic taring
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
4. 2. Checkweighing range
4. 2. 1 * 30 to 170%
4. 2. 2 10% to unlimited
22
O Checkweighing
4. 5. Parameter input
4. 5. 1 * Min, max, target
4. 5. 2 Only target with percentage limits
O CL Classification
3. 5. Minimum load for automatic taring and automatic printout
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
4. 8. Number of classes:
4. 8. 1 * 3 classes
4. 8. 2 5 classes
4. 9. Parameter input
4. 9. 1 * Weight values
4. 9. 2 Percentage
23
O CL Classification
4.10. Automatic printout of results
4.10. 1 * Off
4.10. 2 On
L Weighing in Percent
3. 5. Minimum load for automatic taring and automatic printout
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
24
L Weighing in Percent
9. 1. Restore factory settings of all application programs
9. 1. 1 Yes
9. 1. 2 * No
3.17. Printout of individual component data when value stored in totalizing memory
3.17. 1 Automatic printout of results off
3.17. 2 * Generate printout with complete standard configuration each time O is pressed
3.17. 3 Generate printout with complete standard configuration only once when O is pressed
∑ Totalizing
3. 5. Minimum load for automatic taring and automatic printout
3. 5. 1 1 digit
3. 5. 2 2 digits
3. 5. 3 5 digits
3. 5. 4 * 10 digits
3. 5. 5 20 digits
3. 5. 6 50 digits
3. 5. 7 100 digits
3. 5. 8 200 digits
3. 5. 9 500 digits
3. 5.10 1000 digits
25
∑ Totalizing
3. 7. Automatic tare: 1st weight tared
3. 7. 1* Off
3. 7. 2 On
3.17. Printout of individual component data when value stored in totalizing memory
3.17. 1 Automatic printout of results off
3.17. 2 * Print individual transaction by pressing O
3.17. 3 Print components of transaction by pressing O
26
setup Device Parameters
code Password prompt
(only when password is active: 2 sec. on)
_ (cursor flashes) Enter the password
1. 2. Application filter
1. 2. 1 * Final readout
1. 2. 2 Filling mode
1. 2. 3 Low filtering
1. 2. 4 Without filtering
1. 3. Stability range
1. 3. 1 4 digit
1. 3. 2 * 1 digit
1. 3. 3 1 digit 1)
1. 3. 4 * 2 digits 1)
1. 3. 5 4 digits 1)
1. 3. 6 8 digits 1)
1. 4. Stability delay
1. 4. 1 No delay
1. 4. 2 * Short delay
1. 4. 3 Average delay
1. 4. 4 Long delay
1. 5. Taring 1)
1. 5. 1 Without stability
1. 5. 2 * After stability
1. 6. Auto zero
1. 6. 1 * On
1. 6. 2 Off
1. 7. Weight unit 1 2)
1. 7. 1 Long ton Io
1. 7. 2 Grams /g
1. 7. 3 Kilograms /kg
1. 7. 4 Carats /ct 1)
1. 7. 5 Pounds /lb
1. 7. 6 Ounces /oz
1. 7. 7 Troy ounces /ozt 1)
1. 7. 8 Hong Kong taels /tlh 1)
1. 7. 9 Singapore taels /tls 1)
1. 7.10 Taiwanese taels /tlt 1)
1. 7.11 Grains /GN 1)
1. 7.12 Pennyweights /dwt 1)
1. 7.13 Milligrams /mg 1)
1. 7.14 Parts per pound //lb 1)
1. 7.15 Chinese taels /tlc 1)
1. 7.16 Mommes /mom 1)
1. 7.17 Austrian carats /K 1)
1. 7.18 Tola /tol 1)
1. 7.19 Baht /bat 1)
1. 7.20 Mesghal /MS 1)
1. 7.21 Tons /t
1
) = not available on scales verified for use in legal metrology
2
) = depends on type of weighing platform
27
wp-1 1
Weighing Platform 1
1. 8. Display accuracy 1
1. 8. 1 * All digits
1. 8. 2 Reduced by 1 decimal place for load change
1. 8. 3 Reduced by 1 decimal place 1)
1. 8. 4 Reduced by 2 decimal places 1)
1. 8. 5 Reduced by 3 decimal places 1)
1. 9. Calibration, adjustment
1. 9. 1 * Ext. calibration/adjustment; default weights
1. 9. 3 Ext. calibration/adjustment;
weight can be selected under menu item 1.18.1
1. 9.10 Key blocked
1
) = not available on scales verified for use in legal metrology
2
) = only on scales verified for use in legal metrology
28
wp-1 1
Weighing Platform 1
3. 1. Weight unit 2 1)
3. 1. 1
3. 1. 2 Grams /g
3. 1. 3 Kilograms /kg
3. 1. 4 Carats /ct 2)
3. 1. 5 Pounds /lb
3. 1. 6 Ounces /oz
3. 1. 7 Troy ounces /ozt 2)
3. 1. 8 Hong Kong taels /tlh 2)
3. 1. 9 Singapore taels /tls 2)
3. 1.10 Taiwanese taels /tlt 2)
3. 1.11 Grains /GN 2)
3. 1.12 Pennyweights /dwt 2)
3. 1.13 Milligrams /mg 2)
3. 1.14 Parts per pound //lb 2)
3. 1.15 Chinese taels /tlc 2)
3. 1.16 Mommes /mom 2)
3. 1.17 Austrian carats /K 2)
3. 1.18 Tola /tol 2)
3. 1.19 Baht /bat 2)
3. 1.20 Mesghal /ms 2)
3. 1.21 Tons /t
3. 2. Display accuracy 2
3. 2. 1 * All digits
3. 2. 2 Reduced by 1 decimal place for load change
3. 2. 3 Reduced by 1 decimal place
3. 2. 4 Reduced by 2 decimal places
3. 2. 5 Reduced by 3 decimal places
3. 3. Weight unit 3
Counting menu as for weight unit 2
3. 4. Display accuracy 3
Counting menu as for display accuracy 2
COM1 2 Interface 1
Display designation of this menu level: 2
off * Off
wp2 5) Weighing Platform 2
rs-232 * RS-232
sbi-std SBI standard version
SBI-app SBI for legal metrology
bpi-232 * xBPI-232 1)
Numeric menu as for WP-1 ...
1. 1. ...
1. 9. 4 Internal calibration/adjustment
adc-232 ADC-232 1)
Numeric menu as for WP-1 ...
1. 1. ...
1. 9. 4 Internal calibration/adjustment
5. 2. Parity
5. 2. 2 Space 3)
5. 2. 3 * Odd
5. 2. 4 Even
5. 2. 5 None 4)
1 3 5
) = depends on type of weighing platform ) = not with setting 5. 6. 2 (8 bits) ) = not permitted in legal for trade applications
2 4
) = not available on scales verified for use in legal metrology ) = not with setting 5. 6. 1 (7 bits)
29
datprot Data protocol
sbi * SBI standard version
5. 3. Number of stop bits
5. 3. 1 * 1 stop bit
5. 3. 2 2 stop bits
5. 4. Handshake mode
5. 4. 1 Software handshake
5. 4. 3 * Hardware handshake, 1character after CTS
bpi-232 xBPI-232
ind. 2 Index 2
1 2.1 Ind. 2.1
1 2.2 Ind. 2.2
1 2.3 Ind. 2.3
1 2.4 Ind. 2.4
30
datprot Data protocol
mp8 “MP8 interface" emulation
pari Parity
5. 2. 1 Mark
5. 2. 2 Space
5. 2. 3 * Odd
5. 2. 4 Even
5. 2. Parity
5. 2. 2 Space 1)
5. 2. 3 Odd
5. 2. 4 Even
5. 2. 5 * None 2)
5. 4. Handshake mode
5. 4. 1 Software handshake
5. 4. 3 * Hardware handshake, 1 character after CTS
5. 2. Parity
5. 2. 2 Space
5. 2. 3 * Odd
5. 2. 4 Even
5. 2. 5 None
5. 4. Handshake mode
5. 4. 1 Software handshake
5. 4. 3 * Hardware handshake, 1 character after CTS
1) not with setting 5. 6. 2. (8 bits) 2) not with setting 5. 6. 1 (7bits)
31
printer Printer configuration
YDP03 YDP03-0CE
5. 1. Baud rate
5. 1. 4 * 1200 baud
5. 1. 5 2400 baud
5. 1. 6 4800 baud
5. 1. 7 9600 baud
5. 1. 8 19,200 baud
5. 2. Parity
5. 2. 2 Space
5. 2. 3 * Odd
5. 2. 4 Even
5. 4. Handshake mode
5. 4. 1 Software handshake
5. 4. 3 * Hardware handshake, 1 character after CTS
YDP02IS YDP02IS
line * Strip printer
label Label printer
5. 2. Parity
5. 2. 2 Space 1)
5. 2. 3 Odd
5. 2. 4 Even
5. 2. 5 * None 2)
5. 4. Handshake mode
5. 4. 1 * Software handshake
5. 4. 3 Hardware handshake, 1 character after CTS
YDP04IS * YDP04IS
line* Strip printer
label Label printer
LabFF Label print with manual feed
32
unicom 3
Interface 2 (Optional)
Display designation of this menu level: 3
off * Off
wp-2 Weighing Platform 2
rs-232 RS-232
sbi-std SBI standard version
SBI-app SBI for legal metrology
bpi-232 * xBPI-232 1)
Numeric menu as for WP-1
1. 1. ...
1. 9. 4 Internal calibration/adjustment
adc-232 ADC-232 1)
Numeric menu as for WP-1
1. 1. ...
rs-485 * RS-485
is-485 * For connecting Sartorius IS weighing platforms 1)
Numeric menu as for WP-1
1. 1. ...
1. 9. 4 Internal calibration/adjustment
adc-485 ADC-485 1)
Numeric menu as for WP-1
1. 1. ...
1. 9. 4 Internal calibration/adjustment
datprot Data protocol
sbi * SBI-standard version
Numeric menu as for COM1
5. 1. ...
bpi-232 xBPI-232
bpi-485 xBPI-485
0 to 31 Network address: From 0 to 31 inclusive
1
) = menu depends on connected scale or weighing platform
33
unICoM 3 Interface 2 (optional)
analog Analog Data Output Port for PLC Operation
8.12. Output value
8.12. 1 * Net value
8.12. 2 Gross value
5
barcode Bar Code
Display designation of this menu level: 5
ref * Store reference value
tare Store value directly as tare value
6
prtprot Printouts
Display designation of this menu level: 6
7. 4. Enter header line
7. 4. 1 (blank) Header line 1 (max. 20 characters),
input example: “MEYER"
7. 4. 2 (blank) Header line 2 (max. 20 characters),
input example: “STEEL"
7. 5. COM1 interface
7. 5. 1 * 1 printout
7. 5. 2 2 printouts
7. 7. COM1 interface
Configure standard printout (press p, O)
7. 7. 1 * 1) Header lines 1, 2
7. 7. 2 * 1) Date and time
7. 7. 3 * 1) Initialization data for the application program
7. 7. 4 * 1) Scale/platform designation
7. 7. 5 * 1) Result from the application program
34
prtprot 6
Printouts
7. 9. COM1 interface:
Print results when c pressed in
Totalizing and Net-total applications
7. 9. 1 * 1) Header lines 1, 2
7. 9. 2 * 1) Date and time
7. 9. 4 * 1) Scale/platform designation
7. 9. 5 * 1) Result from the application program
7.14. FlexPrint
7.14. 1 * Off
7.14. 2 On
8. 3. Keys
8. 3. 1 * Accessible
8. 3. 2 Blocked
8. 8. Display lighting
8. 8. 1 * On
8. 8. 2 Off
8. 8. 3 Switch off according to the setting under
menu code 8. 9., “Timer"
8. 9. Timer
8. 9. 1 * After 1 + 1 minute not in use
(after 1 min.: warning* displayed for 1 minute)
8. 9. 2 After 2 + 2 minutes not in use
(after 2 min.: warning* displayed for 2 minutes)
8. 9. 3 After 5 + 5 minutes not in use
(after 5 min.: warning* displayed for 5 minutes)
1
) More than one can be selected
35
utilit. 7 Operating the Combics
8.11. First platform displayed
8.11. 1 * Weighing platform WP-1
8.11. 2 Weighing platform WP-2
time Time
Input example: 10.07.41 (hh.mm.ss)
date Date
Input: 05.01.02 (mm.dd.yy); non U.S. mode: mm.dd.yy Input: 01.05.02 (dd.mm.yy); U.S. mode: mm.dd.yy
term Indicator
TN02 1 Model
10405355 2 Serial number
1.0101.10.01 3 Version number of the indicator
01.24.01 4 Software version
flexinf FlexPrint
APPLSET 1 File name 2)
id 123 2 ID 2)
v 123 3 Version 2)
36
Operating the Combics
Automatic Printing
The first weight value that exceeds
the preset minimum load is printed
automatically.
37
Example
Switch on the Combics 2; zero the platform; tare the container weight; fill container; toggle display to gross weight, second weight unit,
or 10-fold increased resolution
8888888 188
01 188
00 188
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
500 188
00 188
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
38
8 Display for tared platform with filled container
1202 188
k Toggle the display; depending on the settings, the display now shows either
1702 188
(in this example, 50 g container weight + 120.2 g substrate), or
8 Net weight in the second weight unit (in this example, kg), or
01 188
12023 188
8 Previous display
1202 188
G# + 170.2 g
T + 50.0 g
N + 120.2 g
--------------------
39
Calibration and Adjustment Features Preparation
Purpose Which of the following features are § Press the e key to switch on the
The accuracy of weighing results must available depends on the connected Combics
be carefully controlled. This is achieved weighing platform:
through calibration and adjustment. § While the self-test of the display
Calibration technically means to deter- – External calibration/adjustment with segments is running, press the ) key
mine the difference between the weigh- the default weight value or standard
ing instrument readout and the actual weight (1.9.1) (not available on verified § Open the Setup menu: Press the k
weight on the platform to determine weighing platforms) key twice
the accuracy. This does not involve
making any changes in the weighing – External calibration/adjustment with § Select the Setup menu: Press the
platform. a user-defined weight (1.9.3) ) key
Adjustment means to bring a weighing (not available on verified weighing
instrument to the level of accuracy platforms) § Select weighing platform 1
required for its use. (by pressing ))
– Internal adjustment for an IS platform
Configuring Functions for Use of the configured as WP2 (1.9.4) With the Combics 2, you have the option
Weighing Platform in Legal Metrology of selecting weighing platform 2:
(for service personnel only) – Block the function of the ) key when – Select Com-1 or Unicom (depending
Proceed as follows to use the Combics pressed > 2 seconds to prevent use of on which interface the second platform
in legal metrology: the functions described above (1.9.10) is connected to)
– Select WP-2
– Block external adjustment – Calibration followed by automatic – Press ) twice
adjustment (1.10.1) (not available on
Preparation verified weighing platforms) 1 Weighing platform parameters
§ Remove the covering panel on the 1.1 ... 1.8
left-hand side of the back of the – Calibration, optionally followed by
indicator. (manually triggered) adjustment (1.10.2) 1.9 Calibration function
(not available on verified weighing
platforms) 1.9.1 External calibration/
§ Move the switch to the right.
adjustment with default weight
> Switch on the left: – Calibration prompt: flashing W symbol 1.9.3 External calibration/
For legal metrology (external (1.15.2). If more than one platform is adjustment with user-defined
adjustment blocked) connected, the platform number is also weight
(Factory setting on verified models) displayed.
1.9.4 Internal adjustment
> Switch on the right: External adjust- – Block external calibration/adjustment (for WP2 only)
ment accessible (1.16.2) 1.9.10 Block the function of
) key when pressed > 2 sec.
– No external adjustment on verified
weighing platforms 1.10 Calibration/adjustment sequence
1.10.1 Calibration/adjustment
– Altitude and degree of latitude or
in one operation
gravitational acceleration displayed
after Cal is shown when the calibration 1.10.2 Calibration; adjustment
procedure is started, if these values have triggered manually
been entered in the Combics by service
personnel and are supported by the 1.11 ... 1.14
connected platform. 1.15 Calibration prompt
For each of these parameters, the term
is displayed first (Altitud, Latitud, 1.15.1 Off
or Gravity) for approx. 1.15.2 On
1 second, and then the corresponding
value is displayed continuously until 1.16 External calibration/adjustment
you press ).
1.16.1 Accessible
1.16.2 Blocked
1.18 Enter calibration weight
1.18.1 User input
40
Example
External calibration and manual adjustment with default weights
Settings in the Setup menu: Please see the "Preparation" section on the previous page
1.9.1; 1.10.2; 1.16.1
Press and hold ) Start calibration (e.g., when calibration prompt flashes: W)
cal000
If the setting Setup: utilit.: 8.12.2 is selected, the display now shows ALTITUD
for 1 second, followed by the corresponding value. Press ) to confirm; then
LATITUD is displayed for 1 second followed by the value. If the gravitational
acceleration has been entered as a valid parameter, then this value is displayed rather
than the altitude and latitude.
You are prompted to place the required weight on the platform (e.g., 10 kg)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
The difference between the weight value and the true mass is displayed, with O sign.
External calibration
Targ. + 50.00 lb If you press ( to cancel calibration without performing adjustment, a printout is generated
Diff. - 0.01 lb
----------------------
41
-------------------- A GMP-compliant printout is generated
14.01.2002 13:00
Model CW2P1-30ED-LCE
Ser.no. 12345678
Vers. 1.1007.12.1
BVers. 01-25-01
--------------------
External calibration
Targ. + 50.00 lb
Diff. - 0.01 lb
External adjustment
Diff. + 0.00 lb
--------------------
14.01.2002 13:02
Name:
--------------------
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
42
Operation
43
Operating the Combics 2
– Automatic taring (first weight on The value for the reference sample
platform tared): Off (3.7.1) quantity remains active in the reference
memory until you overwrite it or until
– Application started automatically with you select a different application. It also
most recent initialization data when remains stored after you switch off the
Combics switched on: Off (manual start) Combics.
(3.8.2)
44
Operating the Combics 2
45
Preparation
§ Press the e key to switch on the
Combics
§ While all segments are lit, press the
) key
§ Open the numeric menu: Press the
) key twice
3 Application parameters
3.5 Minimum load for automatic taring
and automatic printing
3.5.1 1 digit
3.5.2 2 digits
3.5.3 5 digits
3.5.4 10 digits
3.5.5 20 digits
3.5.6 50 digits
3.5.7 100 digits
3.5.8 200 digits
3.5.9 500 digits
3.5.10 1000 digits
3.6 Minimum load for automatic
initialization
3.6.1 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.7 Automatic taring (first weight tared)
3.7.1 Off
3.7.2 On
3.8 Application started automatically with
most recent initialization data when
Combics switched on
3.8.1 On (automatic start)
3.8.2 Off (manual start)
3.9 Resolution for calculation of
reference value
3.9.1 Display resolution
3.9.2 Display resolution + 1
3.9.3 Display resolution + 2
3.9.4 Internal resolution
3.11 Storage parameter
3.11.1 At stability
3.11.2 At higher stability
3.12 Reference sample updating
3.12.1 None
3.12.2 Automatic
3.13 Reference platform
3.13.1 Off
3.13.2 WP1
3.13.3 WP2
46
Operating the Combics 2
Example
Determine piece count; menu setting: weigh in the defined reference sample quantity with the minimum load set to 100 digits
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Counting: Z
Select parameter 3.6.7 100 digits: Parameter 3.6.7
Operation
00 10
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
00 10
Press r (repeatedly, if necessary) to set the desired reference sample quantity (in this example, 20)
47
8
00 20
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
If the weight is too light, an error code Reduce the minimum load setting (3.6) or
appears in the main display inf 29 increase the reference sample quantity and the number of parts in the container
20 20
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
38 20
0pt 38 20
G# + 610.0 g
T + 200.0 g
N + 410.0 g
Qnt 38 pcs
--------------------
48
Operating the Combics 2
– Minimum load for automatic initializa- – By entering the reference weight (i.e.,
tion: 1 digit (3.6.1) the weight of 1 unit) using a bar code
scanner. In this case, the value is stored
– Automatic taring (first weight on automatically without pressing O.
platform tared): Off (3.7.1)
The value for the reference weight
– Application started automatically with remains active in the reference memory
most recent initialization data when until you overwrite it or until you select
Combics switched on: Off (manual start) a different application. It also remains
(3.8.2) stored after you switch off the Combics.
– Reference value: 1
49
Minimum Load Neutral Measurement with Two Preparation
The minimum load at initialization is Weighing Platforms § Press the e key to switch on the
configured in the Application settings You can use two weighing platforms Combics
(3.5: minimum load for automatic taring simultaneously with the Neutral § While all segments are lit, press the
and automatic printing; 3.6: minimum Measurement application. When using ) key
load for automatic initialization). two platforms, you can choose from § Open the numeric menu: Press the
You can choose from the following ten the following operating modes: ) key twice
levels for this setting: – Neutral Measurement with two
platforms of the same type 3 Application parameters
1 digit (no minimum load) – Neutral Measurement with one 3.5 Minimum load for automatic taring
2 digits reference platform and one weighing and automatic printout
5 digits platform
10 digits 3.5.1 1 digit
20 digits Neutral Measurement with Two 3.5.2 2 digits
50 digits Platforms of the Same Type 3.5.3 5 digits
100 digits Use this mode to measure different 3.5.4 10 digits
200 digits types of sample material with different 3.5.5 20 digits
500 digits weights. For example, measure the 3.5.6 50 digits
1000 digits lighter-weight samples on one platform 3.5.7 100 digits
and the heavier samples on the other. 3.5.8 200 digits
The “digits" here refer to the weighing After starting a counting routine on 3.5.9 500 digits
platform intervals in the connected one platform, press n to toggle to 3.5.10 1000 digits
weighing platform. If the interval the other platform, re-initialize the 3.6 Minimum load for automatic
of the connected platform is 1 g, for application and begin a new measuring initialization
example, and 1000 digits are required, operation. 3.6.1 1 digit
you must place at least 1000 g 3.6.2 2 digits
(= 1000 intervals = 1000 digits) on the You can define which of the two 3.6.3 5 digits
weighing platform for initialization. platforms is active when the Combics 3.6.4 10 digits
is switched on (8.11). This setting is 3.6.5 20 digits
If the interval of the connected platform independent of automatic initializiation 3.6.6 50 digits
is 5 g, you have to place at least 5000 g of the Neutral Measurement application. 3.6.7 100 digits
on the platform for initialization to 3.6.8 200 digits
achieve the number of digits required 3.6.9 500 digits
in the above example. Neutral Measurement with 3.6.10 1000 digits
One Reference Platform and 3.7 Automatic taring (first weight tared)
You can increase the resolution for One Weighing Platform
determining the reference weight In this mode, the reference platform is a 3.7.1 Off
by changing the Application settings high-resolution weighing platform with 3.7.2 On
for resolution (3.9) to use display a relatively low maximum capacity. The 3.8 Application started automatically with
resolution + 1 decimal place, display other platform is used for measuring most recent initialization data when
resolution + 2 decimal places, or large amounts, and has a high capacity Combics switched on
internal resolution. With the setting with a relatively low resolution. 3.8.1 On (automatic start)
for “Display resolution + 1 (2) decimal 3.8.2 Off (manual start)
place(s)", the resolution of the net value This allows you to both determine
3.9 Resolution for calculation of
is increased by one or two decimal the reference sample quantity with
reference value
places; “Internal resolution" uses the high resolution; i.e., very precisely,
maximum resolution available in the and to measure large amounts of parts, 3.9.1 Display resolution
weighing platform. without requiring an expensive high- 3.9.2 Display resolution + 1
resolution, high-capacity weighing 3.9.3 Display resolution + 2
Once the limit is exceeded by the load, platform. 3.9.4 Internal resolution
initialization can begin. If the load on 3.10 Decimal places displayed
platform is too light for initialization, The system toggles automatically to 3.10.1 None
the following will occur when you try the reference platform for initializiation 3.10.2 1 decimal place
to initialize the application: (REF is displayed). Following initializa- 3.10.3 2 decimal places
tion, the system toggles to the measur- 3.10.4 3 decimal places
– The error inf 29 is displayed. ing platform.
3.11 Storage parameter
– A warning signal (double-beep) 3.11.1 At stability
is emitted 3.11.2 At higher stability
3.13 Reference platform
– The weighing platform is not initialized
3.13.1 Off
3.13.2 WP1
– The preset reference value is stored
3.13.3 WP2
50
Operating the Combics 2
Example
Determine the length of an amount of electrical cable after weighing in the defined reference unit value
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Neutral Measurement: Z nm
Select Neutral Measurement
Operation
00 1
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
00 1
Press r (repeatedly, if necessary) to set the desired reference value (in this example, 2)
51
8
00 2
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
If the weight is too light, an error code Reduce the minimum load setting (3.5) or
appears in the main display inf 29 increase the reference sample value setting and the number of units in the container
2 2
Remove the reference material and place the sample to be measured on the weighing
platform (in this example, 8 units)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 2
G# + 734.1 g
T + 200.0 g
N + 534.1 g
Qnt 8 o
--------------------
52
Operating the Combics 2
53
Display Modes Digital I/O Interface Checking Range
The result of a measurement is The digital I/O interface is supported 30 % Target weight 170%
displayed in the weight display mode by the Checkweighing application.
or tolerance limit display mode. You Four data output ports relay the Lower Upper
limit limit
can toggle between the two display following information (see diagram):
modes by pressing w. – Lighter
– Equal
– Weight Display Mode – Heavier
The weighed values are always shown – Set Lighter
on the main display, even if the The “Set” output normally changes
corresponding results are over or under its voltage level when the load is near Equal
the tolerance limits. the target weight. Alternatively, the
“Ready-for-Use” function (4.3.2) can
In the bar graph, the symbols for lower be assigned to this output.
limit, target value and upper limit are
Heavier
displayed. In the application program, you can
define whether these control ports
The corresponding LED lights up: – are deactivated (4.4.1)
Yellow: Weighing result > upper – always activated (4.4.2)
tolerance limit – activated at stability (4.4.3)
Green: Weighing result lies within – activated within the check range (4.4.4)
the tolerance limit – activated at stability within the check Set
Red: Weighing result < lower range (4.4.5)
tolerance limit
Ready to operate
For example, this enables you to show
If no LED lights up: the weighed or measuring result on a
– the application is not yet completely simple external indicator, similar to the
initialized, or 3 LEDs on the Combics 2.
– weighing result lies outside of the Digital I/O Interface
check range, or All data output ports have a high Control line “SET”: set and control lines:
– weighing platform has not reached voltage level when: always/always at stability
stability – The application has not been initialized
– The weighing platform is not at stability
– Tolerance Limit Display Mode and the menu setting “at stability ...” Checking Range
(4.4.3 and 4.4.5) was selected 30 % Target weight 170%
The same as the weight display mode, – The load is not within the check-
Lower Upper
but: weighing range (4.4.4) limit limit
– LL appears on the main indicator
when the weighing result is less than Lighter
the lower tolerance limit
– HH appears on the main indicator
for weighing result is greater than
the upper tolerance limit
Equal
– Bar Graph
The bar shows a logarithmic display
of the current load if the weight is any-
where from 0 to the minimum load, and Heavier
a linear display for weights beyond that
range.
Set
Ready to operate
54
Operating the Combics 2
Setup
§ Turn on the Combics by pressing e
§ While all segments are lit, press the
) key
§ Select the Numeric menu: press the
) key 2 x
3 Application Parameters
3.5 Minimum capacity for automatic
taring and for automatic printout
3.5.1 1 digit
3.5.2 2 digits
3.5.3 5 digits
3.5.4 10 digits
3.5.5 20 digits
3.5.6 50 digits
3.5.7 100 digits
3.5.8 200 digits
3.5.9 500 digits
3.5.10 1000 digits
3.7 Automatic taring (first weight tared)
3.7.1 Off
3.7.2 On
3.8 Automatic start mode:
When switched on, the application
starts automatically with the previous
initialization data:
3.8.1 Automatic mode (On)
3.8.2 Manual mode (off)
4 Application Parameters 2
4.2 Check range
4.2.1 30% to 170%
4.2.2 10% to infinity
4.3 Switch control line “SET” to
4.3.1 Output “SET”
4.3.2 Ready
4.4 Control Lines
4.4.1 Off
4.4.2 Always activated
4.4.3 Activated at stability
4.4.4 Activated within check range
4.4.5 Activated at stability within
the check range
4.5 Parameter Input
4.5.1 Min, max, target value
4.5.2 Target value only with limits
in percent
4.6 Automatic Printout
4.6.1 Off
4.6.2 On
4.6.3 Only if within the
tolerance limits
4.6.4 Only if outside the
tolerance limits
55
Example
Initialize checkweighing by storing a weighed value as the target value, tolerance limit display mode in percent
e Switch Combics on
8888888 188
)
The display for selecting the application appears
appl000
Select application appl Symbol for Checkweighing: O
Select Checkweighing
If desired, change the configurations in see the paragraph “Setup" on the previous page
the Setup menu:
Weighing Mode
Place a sample on the weighing platform for target value (here: 100.0g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The weight of the sample is indicated (percentage for limit value, here: 10%)
1000 10
1000 10
Press r (repeatedly, if necessary) to set the desired percentage for the tolerance limits (minimum, maximum)
56
Operating the Combics 2
8 The middle bar graph segments flash, percentage for limits (in this example: 5%)
1000 15
Press and hold the w key to view the stored results if desired (Info mode)
1000 inf
5 Inf
950 inf
Press and hold down w to exit the menu target value, minimum and maximum (exit Info mode)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
The weighed value appears in the main indicator, the bar graph segments are not lit up.
00
57
Example: Checkweigh a sample with an unknown weight in the tolerance limit display mode.
If desired, w if the weight display mode is activated, toggle to the tolerance limit display mode
The weighing platform is unloaded and LL (weight too light) appears on the main display
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
If the weight is greater than the maximum, HH (weight too heavy) is indicated
If the weight is within the tolerance range (here: 103.2 g), the weighed value is indicated
58
Operating the Combics 2
59
Display Digital I/O Interface Min. load Max. load
The result of a given measurement The Classification application supports Class
is shown as either a weight value or the digital input/output interface. There 1 2 3
the class number. You can toggle are 4 control lines, or outputs, which
between these two display modes by are activated as follows (see also the Class 1
pressing w. diagram below):
60
Operating the Combics 2
61
Example
Initializing the Classification application by storing weight values for 5 classes
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Classification: O cl
Select Classification
Application settings: Please see the “Preparation" section on the previous page
4.8.2, 4.9.1
Operation
Place weight for upper limit of Class 1 on platform (in this example, 10 g)
The minimum load has already been configured in the Application menu
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
100 1
8 The readout shows the current weight and the first four segments of the bar graph flash
100 1
62
Operating the Combics 2
Place the weight for Class 2 on the platform (in this example, 20 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The readout shows the current weight and the “minus" indicator for the bar graph flashes
200 2
Place the weight for Class 3 on the platform (in this example, 30 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The readout shows the current weight and the four segments around the “=" sign
300 3
of the bar graph flash
Place the weight for Class 4 on the platform (in this example, 40 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The readout shows the current weight and the “plus" indicator for the bar graph flashes
400 4
class 4 4
63
Press and hold w View class delimiters (Info mode)
100 1
200 1
Upper limit of Class 2 is displayed
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The main display shows the weight values and the numeric display shows (Class) “0"
00 0
64
Operating the Combics 2
Example:
Determine the class of a given weight
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 The display shows which class the weight falls into; in this example, Class 2 (10 g – 20 g)
class 2 2
173 2
G# + 17.3 g
T + 0.0 g
N + 17.3 g
Class 2
--------------------
65
Weighing in Percent L To determine the weight of a sample Storage Parameter
When the Weighing in Percent relative to a reference weight, you The reference weight is stored when
application is selected, you can use must enter the reference weight in the the platform has stabilized. “Stability" is
your weighing platform to obtain weighing platform. There are 2 ways defined as the point at which
weight readouts in percent which are to do this with the Combics weighing fluctuation of a measured value lies
in proportion to a reference weight. platform: within a defined tolerance range. The
L is displayed as the weight unit. narrower the tolerance range, the more
– By placing the amount of reference stable the platform is at “stability".
Features: material on the weighing platform This setting is also applied when you
– Store reference weight from the value specified by the reference percentage zero the weighing platform.
on the weighing platform and pressing O to calculate and
store it. You can set the parameter for “Stability
– Enter the reference weight using a bar at initialization" to “At stability"
code scanner The reference percentage is shown in (normal tolerance range) or “At higher
the numeric display. stability" (limited tolerance range).
– Enter tare weight using a bar code
scanner How the reference weight is calculated If you select “higher stability", the
depends on the resolution specified in reference weight stored will be
– Toggle the display between percentage the Application settings (3.9). It is either more accurate and the results more
and weight by pressing w rounded off in accordance with the dis- reproducible, but the response time of
play resolution, or stored with 10-fold the weighing platform might be longer.
– Show percentage as loss or residue or 100-fold resolution, or with the max-
imum internal resolution of the weigh-
– Values displayed with up to 3 decimal ing platform. Press the r key to
places change the reference weight; then con-
tinue as described above.
– Info mode
– By entering the reference weight (in this
Factory settings: example, the weight of 1 unit) using a
– Minimum load for automatic taring and bar code scanner. In this case, the value
automatic printing: 10 digits (3.5.4) is stored automatically without pressing
O.
– Minimum load for automatic initializa-
tion: 1 digit (3.6.1) The value for the reference weight
remains active in the reference memory
– Automatic taring (first weight on until you overwrite it or until you select
platform tared): Off (3.7.1) a different application. It also remains
stored after you switch off the Combics.
– Application started automatically with
most recent initialization data when
Combics switched on: Off (manual start)
(3.8.2)
66
Operating the Combics 2
67
Example
Weighing in percent after weighing in the defined reference amount
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Weighing in Percent: L
Select Weighing in Percent
Operation
00 1000
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
00 1000
Press r (repeatedly, if necessary) to set the desired reference percentage (in this example, 50)
68
Operating the Combics 2
00 500
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
If the weight is too light, an error code Reduce the minimum load setting (3.5) or
appears in the main display inf 29 increase the reference percentage and the amount of sample material in the container
50 50
Remove the reference material and place the sample to be measured on the weighing
platform (in this example, 174 percent)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
79 50
G# + 0.641 kg residue, or
T + 0.200 kg
N + 0.441 kg
Prc 21.00 %
--------------------
or
G# + 0.641 kg loss.
T + 0.200 kg
N + 0.441 kg
D 79.00 %
--------------------
69
Averaging (Animal Weighing) V In order to calculate an average Minimum Load
When the Averaging application is weight, you need to perform several The minimum load at initialization is
selected, you can use your weighing subweighing operations. These are configured in the Application settings
platform for calculating weights as performed consecutively as soon as the (3.5: minimum load for automatic taring
the average of a number of individual weight of one sample differs from that and automatic printing; 3.6: minimum
weighing operations. These individual of the previous sample by less than the load for automatic start). You can
operations are also known as preset deviation limit. choose from the following ten levels for
“subweighing operations." The value for this deviation is entered this setting:
in the Application settings (3.19) as
This application is used to determine a percentage and is another parameter, 1 digit (no minimum load)
weights under unstable ambient in addition to the minimum load, for 2 digits
conditions or for weighing unstable starting the averaging routine. 5 digits
samples (such as live animals). You can also define a minimum weight 10 digits
in the Application settings (3.6). This 20 digits
Features: is especially helpful when using the 50 digits
– Manual start of averaging routine “automatic start" function. 100 digits
Whether averaging starts automatically 200 digits
– Automatic start of averaging routine or manually is also defined in the 500 digits
Application settings (3.18). 1000 digits
– Press the r key to select the desired
number of subweighing operations There are four modes for calculating The “digits" here refer to the weighing
average weights: platform intervals in the connected
– Toggle the display between result of weighing platform. If the interval of the
the last measurement and weight by – Manual start with preset number of connected platform is 1 g, for example,
pressing w. subweighing operations and 1000 digits are required, you must
place at least 1000 g
Factory settings: To use this mode, place the sample or (= 1000 intervals = 1000 digits) on the
– Minimum load for automatic taring and animal on the weighing platform and weighing platform for initialization.
automatic printing: 10 digits (3.5.4) press O.
If the interval of the connected platform
– Minimum load for automatic start: – Manual start with the number of is 5 g, you have to place at least 5000 g
1 digit (3.6.1) subweighing operations entered on the platform for initialization to
manually achieve the number of digits required in
– Automatic taring (first weight on the above example.
platform tared): Off (3.7.1) To use this mode, place the sample/
animal on the weighing platform, Once the limit is exceeded by the load,
– Application started automatically with press r to select the number of initialization can begin. If the load on
most recent initialization data when subweighing operations (5, 10, 20, 30, platform is too light for initialization,
Combics switched on: Off (manual start) 40 or 50) and then press O. the following will occur when you try to
(3.8.2) initialize the application, the averaging
– Automatic start with preset number of routine does not start.
– Start of averaging routine: Manual subweighing operations
(3.18.1)
The subweighing operations begin when
– Deviation of successive subweighing the weight of the sample/animal on the
operations: 0.2% (3.19.2) weighing platform exceeds the minimum
load preset in the Application settings.
– Automatic printout of results: Off
(3.20.1) – Automatic start with the number
of subweighing operations entered
– Display of results static when platform manually
unloaded: Display fixed until unload
threshold reached (3.21.1), or press c To use this mode, press r and enter
and then O to restart. the number of subweighing operations
before placing the sample or animal on
– Number of subweighing operations: 10 the weighing platform.
If the automatic taring function is
active, the weight of the first load is
stored as the tare value, and averaging
begins only when the second load is
placed on the weighing platform
(provided this weight exceeds the preset
minimum load).
The number of subweighs remains
active in the reference memory, and is
automatically loaded when you select
Averaging again after having used
another application. It also remains
stored after you switch off the Combics.
70
Operating the Combics 2
71
Example
Determining the average weight of mice, with manual start and 20 subweighing operations
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Averaging: V
Select Averaging
Operation
00 10
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
00 10
Press r (repeatedly, if necessary) to set the desired number of measurements (in this example, 20)
72
Operating the Combics 2
00 20
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
O Begin averaging
240 14
8 At the conclusion of the series, the calculated mean is displayed continuously together
246 0
with the ! symbol, until the platform is unloaded or c is pressed.
T + 200.000 g
xNet + 24.163 g
--------------------
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
8 Once the weighing platform is unloaded (i.e., load = less than 1/2 the minimum load),
00 20
the current weight value is displayed.
73
Net-Total Formulation R Minimum Load Net-Total Formulation with
When the Net-Total Formulation The minimum load at initialization is 2 Weighing Platforms
application is selected, you can weigh in configured in the Application settings This mode is used for weighing large
different components up to a defined (3.5: minimum load for automatic and small components at the same
total. You can also print out the total taring; 3.6: minimum load for storing time.
weight and the individual weights of value). You can choose from the
the components. following ten levels for this setting: This makes it possible to toggle from
the small-component platform to
If the components are filled in separate 1 digit (no minimum load) the large-component platform during
containers, please use the Totalizing 2 digits measurement. Once you toggle to
application (see page 71). 5 digits the large-component platform, the
10 digits ( and ) keys are available until
Features: 20 digits a component is stored. For example,
– Automatic or manual taring of first 50 digits you can tare a partially-filled container
load on the platform 100 digits taken from the small-component
200 digits platform on the large-component
– Minimum load for automatic taring 500 digits platform.
1000 digits
– Maximum 199 components The content of the component memory
The “digits" here refer to the weighing in the small-component platform is
– Current transaction number shown platform intervals in the connected transferred to the large-component
on the numeric display weighing platform. If the interval of the platform, and the weight value is
connected platform is 1 g, for example, converted, if necessary, to the other
– Press w to toggle the display: and 1000 digits are required, you must weight unit. The Component and
– In the component mode: Show the place at least 1000 g Additive display modes are both
weight of the current component (= 1000 intervals = 1000 digits) on available in the large-component
(shown approx. 1 second after it the weighing platform for the value platform.
is stored, after which the platform to be stored.
is tared) The readout from the active platform is
– Additive mode: Show the previous Once the limit is exceeded by the load, stored in the component memory. The
weight plus the current component initialization can begin. If the load on result is shown in the active weight
weight (shown approx. 1 second after platform is too light for initialization, unit; for example:
it is stored, after which the platform the following will occur when you try to 1243 g (from a platform with 1 interval)
is tared) initialize the application: + 1.40 kg (from a platform with 5 intervals)
= 2.643 kg (calculated result)
– Press and hold the w key to show – The error inf 29 is displayed.
precise content of component memory – A warning signal (double-beep) When you press c to stop a
in the info mode is emitted measurement series, the tare memories
– The value is not stored for both platforms are cleared, unless
– Automatic or manual printout of the large-component platform is an SBI
individual component weights or set of instrument. In this case, the platform
component weights when value stored is only tared.
Factory settings:
– Minimum load for automatic taring and
automatic printing: 10 digits (3.5.4)
74
Operating the Combics 2
Preparation
§ Press the e key to switch on the
Combics
§ While all segments are lit, press the
) key
§ Open the numeric menu: Press the
) key twice
3 Application parameters
3.5 Minimum load for automatic
taring and automatic printing
3.5.1 1 digit
3.5.2 2 digits
3.5.3 5 digits
3.5.4 10 digits
3.5.5 20 digits
3.5.6 50 digits
3.5.7 100 digits
3.5.8 200 digits
3.5.9 500 digits
3.5.10 1000 digits
3.6 Minimum load for storing value
3.6.1 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.7 Automatic taring (first weight tared)
3.7.1 Off
3.7.2 On
3.17 Individual/component printout
when value stored
3.17.1 Printing function off
3.17.2 Print standard configuration
every time by pressing O
3.17.3 Print standard configuration
once by pressing O
75
Example
Weigh in 3 components, then view total weight of components (Additive mode)
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Net-Total Formulation: R
Select Net-Total Formulation
Application settings: Please see the “Preparation" section on the previous page
3.5.1; 3.7.1; 3.17.3
Operation
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
500 11
00 11
76
Operating the Combics 2
Add the first component to the container (in this example, 68.3 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
683 11
8 The platform is tared automatically and the transaction counter is increased by 1 (to “2")
00 12
Add the second component to the container (in this example, 82.7 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
827 12
8 The platform is tared automatically and the transaction counter is increased by 1 (to “3")
00 13
77
Add the third component to the container (in this example, 77.1 g)
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
771 13
3052 13
8 The platform is tared and the readout shows the stored weight plus the current weight
4000 14
w Toggle the display back to the Component mode to view the current weight
00 14
78
Operating the Combics 2
79
Display Printout Preparation
Toggle the display between current In the Application settings you can § Press the e key to switch on the
weight and total weight by pressing configure whether a printout is Combics
w. The numeric display shows the generated manually, by pressing p
transaction counter. (3.17.1), or automatically when a § While all segments are lit, press the
weight value is stored in the totalizing ) key
When you have toggled to the total memory (3.17.2 or3.17.3).
weight display, you can add the current § Open the numeric menu: Press the
weight on the weighing platform When you generate a printout manually ) key twice
to the totalizing memory (manually by pressing p, the transaction counter
by pressing O or automatically, value is not printed. 3 Application parameters
depending on the Application setting). 3.5 Minimum load for automatic
For automatic printouts, you can define taring and automatic printing
Press and hold the key w (> 2 sec) to whether a printout is generated after
3.5.1 1 digit
toggle to the Info mode. The total of all each individual transaction (3.17.2)
3.5.2 2 digits
gross values is displayed. Press w or includes all totalized components
3.5.3 5 digits
again to view the total of all net values. (3.17.3).
3.5.4 10 digits
Press the w key again to show the
3.5.5 20 digits
transaction counter in the main display Each printout starts with the header
3.5.6 50 digits
(in this example: n 15 ). lines and the date and time, followed by
3.5.7 100 digits
Press the w key again to return to a dotted line, before printing the gross,
3.5.8 200 digits
the original status. tare, net and transaction counter values.
3.5.9 500 digits
Press and hold the key w again to
3.5.10 1000 digits
exit the Info Mode before the above The printout of components includes
sequence is completed. the header lines and date and time 3.6 Minimum load for automatic storage
only on the first printout. Subsequent 3.6.1 1 digit
printouts include only the gross, tare, 3.6.2 2 digits
net and transaction counter values (as 3.6.3 5 digits
well as the serial number, if configured). 3.6.4 10 digits
The individual and component printouts 3.6.5 20 digits
are configured under 7.7 and 7.8. 3.6.6 50 digits
3.6.7 100 digits
The total data record is printed when 3.6.8 200 digits
you clear the totalizing memory. The 3.6.9 500 digits
total data record includes the data from 3.6.10 1000 digits
the gross totalizing memory (*G, the 3.7 Automatic taring (first weight tared)
net totalizing memory (* N, the total
transaction count (n) and a dotted line. 3.7.1 Off
This printout always includes the lines 3.7.2 On
configured for Application settings 3.8 Application started automatically with
7.9.x and 7.10.x, regardless of whether most recent initialization data when
individual or component printing is Combics switched on
configured. 3.8.1 On (automatic start)
3.8.2 Off (manual start)
3.16 Weight stored automatically
3.16.1 Off
3.16.2 On
3.17 Automatic individual or component
printout when value stored
3.17.1 Off
3.17.2 Print individual transaction
by pressing O
3.17.3 Print all components of
transaction by pressing O
80
Operating the Combics 2
Example
Totalizing weight values with printout of components
Configure Settings
8888888 188
)
Display for selecting the application is shown
appl000
Open Application selection appl Symbol for Totalizing: s
Select Totalizing
Application settings: Please see the “Preparation" section on the previous page
3.17.3
Operation
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
1029 0
1029 1
In the numeric display: Transaction counter (in this example: 1)
81
Sartorius Header line 1 (user-definable)
Goettingen Header line 2 (user-definable)
06.02.2002 12:15:03
--------------------
G# + 102.9 g
T + 0.0 g
N + 102.9 g
n 1
2058 1
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
2152 1
8 The numeric display shows the transaction counter (in this example: 2).
2152 2
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
82
Operating the Combics 2
4001 2
8 The numeric display shows the transaction counter (in this example: 3).
4001 3
G# + 400.1 g Component printout for the third transaction is generated, as configured under 7.7.x or 7.8.x
T + 0.0 g
N + 400.109 g
n 3
Comb
ics 2
n -
CF
0
T RE
F
Fn
OK
7182 3
83
Data Output Functions
These two interfaces can be configured in the Setup program for different output
functions, such as connection to a printer or a second weighing platform (WP),
or for data communication with a PC.
With the optional UniCOM multifunction interface installed, you can activate one of the
following functions in addition to the COM1 interface:
– RS-232 interface
– RS-485/RS-422 interface
– Analog output (voltage/current interface)
The indicator is divided into 8 sections. Information about the scale or weighing platform,
the application being used and the sample weighed is output in the following sections:
b – The busy symbol, as long as internal operations are being carried out at the press of a key
U – The zero-setting symbol: identifies “zero” as a weight value (after zeroing the scale or the
active weighing platform)
5.234 – The current weight (on verified scales or platforms with e = d, the last digit is bordered for
identification as a legal value)
17.23 – Calculated values (only on Combics 2, e.g., for the Weighing in Percent application)
Unit
The following is displayed in this section of the indicator:
pcs – ID for additional units (only on Combics 2, e.g. for the Counting application)
B/G NET – Identification of the gross value or data in the tare memory (net value)
PT – Identification of manual tare input (via bar code scanner) for tare information
W1 – Display of the active weighing platform when 2 platforms are connected (only for
the Combics 2 model)
The display flashes when the weighing platform prompts isoCAL calibration and adjustment
W12 – When the timer is active (menu code 8.9), this symbol flashes to indicate that one-half of
the preset time period has elapsed.
84
Symbols for Printing, GMP Printout and Battery Charging Display
The following is displayed in this section of the indicator:
Application Symbols
The following is displayed in this section of the indicator:
85
Interface Port
Purpose Specifications
There are two interface ports available
with the Combics indicator: Type of interface port: Serial interface port
– Standard COM1 data interface (built in) Operating mode: Full duplex
– UniCOM universal interface (optional)
Standard: COM1: RS-232,
These two interfaces can be configured UniCOM 1): RS-232, RS-422/485,
in the Setup program (see “Settings”) Interface connector: CISL1 and CISL2 indicators (IP44 protection):
for different input/output functions, 25-contract D-Sub female connector
such as connection to a printer, Alibi CIS1 and CIS2 indicators (IP65/67 protection):
memory, PC, checkweigher, second The free ends of the cable are connected to terminal screws
weighing platform, or for control (terminal screws) inside the housing; the cable is routed
command input (for example, using into the housing via a cable gland.
a foot switch). The optional UniCOM Transmission rates: 150, 300, 600, 1200, 2400, 4800, 9600 and 19,200 baud
universal interface can be used for (depending on the operating mode)
RS-232, RS-485 or RS-422 communica- Number of data bits: 7, 8
tion, or as a voltage/current interface
for analog output. A bar code scanner Parity: Space, odd, even, none (depending on the operating mode)
(Combics 2 only) or an external battery Number of stop bits: 1 or 2 stop bits
pack can be conneced to the female Handshake mode: Software (XON/XOFF), hardware (1 character after CTS)
UniCOM port (on CIS1 and CIS2 models,
use the corresponding terminal screws). Communication mode: SBI, xBPI-232 2), xBPI-485 1)2),
MP8 binary 3), SMA
Available printers: – YDP01IS
Features – YDP01IS-Label
Built-in standard COM1 port and – YDP02
optional UniCOM universal port: – YDP03
– CISL1 and CISL2 indicators – YDP02IS
(IP44 protection): – YDP02IS-Label
Connect via a 25-contact D-Sub female – Universal
connector. – YDP04IS
If you wish to connect a second – YDP04IS-Label
peripheral device to the same interface – Alibi memory, YAM01IS
port, you will need to use a T-connector Network address 4): 0, 1, 2, ..., 31
(see “Accessories”). SBI: Manual printing: Without stability, after stability, configurable printout
– CIS1 and CIS2 indicators SBI: Automatic printing: Without stability, at stability, time-dependent at user-defined
(IP65/67 protection): intervals
Route the connecting cable of the
peripheral device via a cable gland SBI: Output format: 16 characters, 22 characters
into the indicator. Then connect the Printout of application Output of a configurable printout
free ends of the cable using the data:
terminal screws .
If you wish to connect a second
peripheral device to the same interface Factory settings: Depends on device configured; for example,
port, use a separate cable gland to route “Data record”, “SBI”
the connecting cable of this device into Transmission rate: 1200 baud
the indicator and connect the free ends
Number of data bits: 7 bits
of the cable using terminal screws.
Parity: Odd
Stop bits: 1 stop bit
! Warning When Using Pre-wired Handshake: Hardware handshake, 1 character after CTS
RS-232 Connecting Cables:
The pin assignments in the RS-232 Activation of data output: Print on request (manual) after stability
cables purchased from other manu- Time-dependent 1 display interval
facturers are often incompatible automatic printing:
with Sartorius scales. Check the pin Output format: 22 characters
assignments against the chart before
connecting the cable, and disconnect
any lines that have a different assign- ! If necessary, use an external power source to power peripheral devices.
ment. Failure to do so may damage
or even completely ruin your scale
and/or peripheral device. 1)
optional universal data interface
2)
xBPI communication mode always set to 9600 baud, 8-bit, odd parity, 1 stop bit
3)
only the standard COM1 port
4)
Network address applies only to the xBPI communication mode
86
Analog UniCOM interface (optional) Preparation
Standard: 4 to 20 mA, 0 to 20 mA, 0 to 5V § For pin assignments, see the section
entitled “Pin Assignment Chart.”
Power supply: Internal or external
Factory settings: 4 to 20 mA, internal power supply
Interface connector: CISL1 and CISL2 indicators (IP44 protection): Options for Connecting Peripherals
25-contract D-Sub female connector You can connect the following printers
CIS1 and CIS2 indicators (IP65/67 protection): to the standard COM1 port or the
The free ends of the cable are connected to terminal screws optional UniCOM universal port:
inside the housing; the cable is routed into the housing
via a cable gland. – YDP02 (user-definable settings of
the interface parameters)
– YDP03 (user-definable settings of
COM1 Off * the interface parameters)
WP2 1) RS-232 SBI standard version – YDP01IS (strip or label printer)
SBI verified version – YDP02IS (strip or label printer)
XBPI-232 * – YDP04IS (strip or label printer)
ADC-232 – Universal printer (user-definable
Data protocol SBI * settings of the transmission parameters)
XBPI-232 – Alibi memory, YAM01IS
MP8 binary
SMA
Printer YDP01IS Line * ! If necessary, use an external power
Label source to power peripheral devices.
Label, man. form feed
YDP02
YDP03 In addition, the following devices can
YDP02IS Line * be connected to the standard COM1
Label port:
Universal
YDP04IS Line * – Foot switch / hand switch
Label – PC (RS-232 interface)
Label, man. form feed – 2nd weighing platform (only for
YAM01IS Combics 2, RS-232 interface port)
(Alibi memory) – External checkweighing display
External YAM01IS (“traffic-light” display) via digital I/O
Alibi memory (Sartorius standard)
UniCom Off *
WP2 1) RS-232 SBI standard version The connector of the UniCOM universal
SBI verified version port allows the following standard
XBPI-232 * devices to be connected:
ADC-232 – A rechargeable battery pack
RS-485 IS-485 * – A bar code scanner (only for Combics 2)
ADC-485
Data protocol SBI * The following devices can be connected
XBPI-232 to the optional UniCOM universal port:
XBPI-485
SMA – Rechargeable battery pack
Printer YDP01IS Line * – PC (RS-232 interface)
Label – 2nd weighing platform (only for
Label, man. form feed Combics 2, can be switched from the
YDP02 RS-232 to RS-485 mode)
YDP03 – 2nd printer (external power source
YDP02IS Line * required)
Label – Remote display
Universal – Bar code scanner / keyboard interface
YDP04IS Line * – Digital I/O
Label – Current interface (0/4-20 mA)
Label, man. form feed
YAM01IS
(Alibi memory)
Analog output
External YAM01IS
Alibi memory
* = Factory setting
1)
only for Combics 2
87
Combics 2 enables a second weighing Preparation Generating SBI Data Output
platform to be connected, either to the Configuring the Interface Port:
COM1 port or to the UniCOM universal (The previous page shows a general Under items 6.1 and 6.3 in the
port. view) operating menu, you can define
which data is output when an output
The standard COM1 port is operated in § Turn on the indicator: press e command is received:
the RS-232 mode. The second weighing – The displayed value, with or without
platform can be operated in any of the > A self-test is run. stability check
following modes: – Automatic printout of the displayed
– SBI § While the self-test is running, press ) value with or without stability or
– xBPI-232 (factory setting) time-dependent automatic data
– ADC-232 > Access the menu output.
– Output of a configurable printout
The optional UniCOM port can be oper- § Select Setup (SETUP): press k several defined using menu items 7.7.x to 7.10.x
ated either in the RS-232 mode or in times (see next page). You can configure your
the RS-485 mode. You can operate a printout by putting together several
second weighing platform in any of the > SETUP is displayed blocks of information. Please refer to
following modes: the section entitled “Configuring Print-
– SBI (RS-232 mode) § Open the Setup menu: outs” in this chapter for detailed
– XBPI-232 (RS-232 mode) Press ) instructions and sample printouts.
– ADC-232 (RS-232 mode)
– IS485 (RS-485 mode, xBPI § Select the desired interface port Unless you have selected a configurable
communication mode, factory setting) (“COM1” or “UniCOM”): printout (menu setting 6.1.7), the
– ADC-485 (RS-485 mode) press k several times current value displayed on the indicator
(weight with unit, calculated value,
> The selected interface port COM-1 numeric or alphabetic ID) is output.
The standard COM1 port or the or UNICOM is displayed
universal UniCOM port or both can To configure the print mode, select
be used as printer ports. § Access the interface menu: the “Manual/automatic data output"
Press ) setting in the interface configuration
menu (menu item 6.1);
The data record can be set to any of > For the COM1 and UniCOM ports, options include “Print on request"/
the following communication modes: the following settings are possible “Automatic printout", either with or
– SBI (factory setting) (see overview on the previous page): without the stability, and “Configurable
– xBPI-232 printout” (6.1.7). If you select the time-
– xBPI-485 (only optional UniCOM port) § Select the desired setting: dependent automatic output mode,
– MP8-binary (only COM1 port) you also need to define the number of
– SMA – Open the menu: press ) display intervals under menu item 6.3.
– Select setting: press k (several times,
You can operate the COM1 port and if necessary) As a rule, data is output only when the
the UniCOM port independently of one – Confirm and save setting: press ) scale has stabilized (factory setting). If
another (i.e., for transferring data and The selected menu item is identified you prefer to have data output without
controlling equipment via a PC while by a circle (o). stability, you need to configure the
simultaneously outputting data to the – To exit the menu item level and return print mode of the interface accordingly
COM1 printer port). to the next higher level, press ( (menu item 6.1.1 or 6.1.4).
– To exit the menu completely, press )
In the SBI communication mode, for more than 2 sec. Line format:
you can control a display unit and A printed line can have up to 20
a connected weighing platform by characters. The first 6 characters, called
sending ESC commands from a PC the header, identify the subsequent
to the communications port (COM1 value. You can disable header printout,
or UniCOM). For further details, see in which case a printed line contains up
the section entitled “Data Input to 14 characters (menu item 7.2; factory
Format” in this chapter. setting: With header).
88
g
el tin
lev u pag
e
lin
e
word y set ite
m
nu n nu nu tor nu
Me M e Me Me Fac M e
6 7 7 4 Enter header 741 Header 1 Printing
742 Header 2
7 5 COM1 port: 751 * 1 printout There are several actions that generate
Number of printouts 752 2 printouts the command for outputting data to
7 6 UniCOM port: 761 * 1 printout the printer port:
Number of printouts 762 2 printouts – Pressing the p key. If you are working
7 7 COM1 port: 2) 771 * Header(s) 1,2 with the operating menu, the menu
Configure standard or 772 * Date / time 1) settings are printed (see chapter entitled
component printout 773 * Initialization data of the “Settings”).
(p and O keys 1)) application – Sending the SBI command “Esc k P _”.
774 * Scale ID Please see the section entitled
775 * Result of the application “Data Input Format” in this chapter
7 8 UniCOM port: 2) 781 * Header(s) 1,2 for details.
Configure standard or 782 * Date / time 1) – In some applications, pressing a given
component printout 783 * Initialization data of the key (e.g., to store a value or start a rou-
(p and O keys 1)) application tine) also generates a print command.
784 * Scale ID In this case, a configurable printout is
785 * Result of the application generated with the application-specific
7 9 COM1 port: 1) 2) 791 * Header(s) 1,2 data. For more details, refer to the
Configure printout of 792 * Date / time 1) descriptions of the applications and to
components/single component 794 * Scale ID the section entitled “Configuring Print-
(c key) for Totalizing and 795 * Result of the application outs”, which contains sample printouts.
Net-total applications
7 10 UniCOM port: 1) 2) 7 10 1 * Header(s) 1,2 The P and b symbols are displayed
Configure printout of 7 10 2 * Date / time 1) when data is being output to the
components/single component 7 10 4 * Scale ID printer port.
(c key) for Totalizing and 7 10 5 * Result of the application
Net-total applications Line format:
7 11 GMP printout 7 11 1 * Off One printed line consists of a maximum
7 11 2 On, for 1 measured result of 20 characters. The first 6 characters,
7 11 3 One, for several measured called the ‘header', identify the
results subsequent value.
7 12 Date/time stamp 7 12 1 * Off
without time 1) 7 12 2 On
7 13 Auto-print one value 7 13 1 * Off
at stability 3) 7 13 2 On Configuring Printouts
7 14 Flexprint 7 14 1 * Off
7 14 2 On To configure a printout, access the
9 1 Reset menu parameters for 911 Reset numeric menu (level 6; see chart on the
printouts to factory settings 912 * Do not reset left). See the chapter entitled “Settings”
for details on accessing the numeric
menu.
* = factory setting
1)
= only for Combics 2 You can configure a different printout
2)
= several selections are possible for each port. Each printout is made
3)
= when the minimum load is exceeded (menu item 3.5.x) up of various blocks of information,
selected by activating or de-activating
them in the menu:
– Headers: header 1, header 2
– Date, time (not for Combics 1)
– Dotted line and blank line (for the
Weighing application).
You cannot de-activate this block. It is
printed to separate the first block from
the ones that follow.
– Initialization data (e.g., reference
sample quantity, reference piece weight)
followed by a blank line.
This block is not included on the
printout of results from the Totalizing
and Net-Total Formulation applications
(menu items 7.9.x and 7.10.x).
– Serial number of the load cell
– Results: Gross, tare, net values; blank
line and application-specific result (e.g.,
piece count), followed by a dotted line.
89
Use the menu to select individual blocks Block 3: Initialization Data GMP Printout
of information (multiple selections Which data is included in this block Select the menu code for GMP-
possible; factory setting: all blocks). depends on the application. In the compliant printout (7-11) to add a
A block that you have not selected is Weighing application, for example, GMP header and a footer bracketing
omitted on the printout. this block is empty; in the Counting the measured result on the printout. If
application, the reference sample the GMP printout is activated (factory
quantity and the reference piece weight setting: GMP printout de-activated), the
Information blocks are printed. The block is terminated A symbol is displayed on the indicator
The individual blocks of information are with a blank line. If the initialization until the printout is generated. You
shown below with detailed explanations. data block is activated, the following have a choice of the following settings:
Samples of complete printouts are is printed for the Counting application – GMP printout off (factory setting)
provided at the end of this section. (example): – GMP printout for 1 measured result
nRef 10 pcs – GMP format for several measured
Block 1: Headers wRef + 0.035 kg results
You can define 2 headers, each with
20 characters per line (e.g., for printing Block 4: Serial Number of the If you have activated the GMP printout,
your company’s name). Load Cell the GMP header is printed as soon as
The following characters are available: If this is activated, the following the request is received to print out the
“0” to “9”, “A” to “Z”, “-” and “ ” information is printed (example): first measured value. The GMP footer
(space). To input headers, select menu Ser.no. 1234567890 is printed either after each measured
codes 7-4-1 (header 1) and 7-4-2 result (GMP footer for 1 measured
(header 2). Block 5: Results result; menu code setting 7.11.2) or
You need to activate printout of the Which data is included in this after the last measured result of a
headers to have these lines included on block depends on the application. If series, when you press the p key for
the printout. The following is an exam- provided in the application, the gross, more than 2 seconds (GMP for several
ple of two header lines: net and tare weights are usually measured results; menu code setting
ACE HARDWARE printed, followed by a blank line. In 7.11.3). In this case, the A symbol
GOETTINGEN the Counting application, the piece remains displayed until the GMP footer
In this example, the company name count is printed as the result. This is printed.
is centered on the printout. This was block is terminated by a dotted line.
achieved by entering blank spaces at If the results block is activated, the If you toggle to a different platform
the beginning and end of the text. following is printed (example for the (only for Combics 2) while a GMP print-
Counting application): out of several measured results is being
Block 2: Date/Time generated, the GMP footer for the plat-
When this block is activated for B# + 1.402 kg form used up to that point is generated
printing, the following is included T + 0.200 kg when you press n. The GMP header
on your printout (example): N + 1.202 kg for the other platform is included on
21.01.2001 16:02 the next printout generated.
Qnt 34 pcs
If you are using the Combics 1 -------------------- A GMP-compliant printout is generated
indicator, this block is not available. automatically at the conclusion of
calibration/adjustment routines, as well
To obtain a standardized time stamp as when you set or clear a preload.
(e.g., for documentation within a
completely automated system), you can If you use a label printer for GMP-
suppress the printout of the time stamp compliant printouts, a single label might
in the “Date / time” information block. not be long enough for the data printed.
This function is controlled under menu Select menu item 7.11.3 to activate
code 7 –12, “Time not printed". The automatic “label feed” following each
factory setting is “Off” (i.e., the time is printout of a GMP header plus results.
included on the printout). If you select
“On" for this menu item, the time The following page shows 3 samples of
stamp can be inserted by a higher-level GMP headers and footers. On Combics
controller or central computer to 1 models, the date and time line is not
maintain consistent time stamping. included. Please refer to the following
This setting is especially important for section, “Sample Printouts", for samples
communication with a PC. of complete printouts.
90
Weighing platform WP1: Sample Printouts
-------------------- Dotted line
14.01.2002 09:43 Date/time 1) For details on the individual information
Model CW2P1-30ED-LCE Combics model blocks, see “Configuring Printouts" on the
Ser.no. 12345678 Combics serial no. previous page. For details on configuring
Vers. 1.1007.12.1 Software release for applications the header lines, refer to the section
BVers. 01-25-01 Software release for basic version describing the application in question.
-------------------- Dotted line
HEADER1
HEADER2
1)
14.01.2002
--------------------
G# + 1.402 kg
T + 0.200 kg
N + 1.202 kg
--------------------
91
Counting Application Percentage as loss Result outside “OK" range; “weight"
The block for initialization data printout
contains the reference sample quantity HEADER1
and reference piece weight. The results HEADER2 HEADER1
block shows gross, net and tare weights, 14.01.2002 09:43 HEADER2
as well as the calculated piece count. -------------------- 14.01.2002 09:43
pRef 100 % --------------------
HEADER1 Wxx% + 2.100 kg Setp + 1.300 kg
HEADER2 Min + 1.235 kg
14.01.2002 09:43 G# + 0.641 kg Max + 1.365 kg
-------------------- T + 0.200 kg
nRef 10 pcs N + 0.441 kg G# + 1.200 kg
wRef + 0.035 kg T + 0.000 kg
D 21 % N + 1.200 kg
G# + 1.402 kg --------------------
T + 0.212 kg Lim - 7.69 %
N + 1.190 kg Diff.W- 0.100 kg
Checkweighing Application: --------------------
Qnt 34 pcs The block for initialization data
-------------------- contains the nominal, minimum and Result outside “OK" range (too light);
maximum weights. The results block “threshold" printout
Neutral Measurement Application: always contains the gross, net and
The block for initialization data tare weights. The other results can HEADER1
contains the reference number and be displayed in two ways: HEADER2
reference weight. The results block – Result as weight: 14.01.2002 09:43
shows gross, net and tare weights, The deviation from the nominal weight --------------------
as well as the calculated piece count. is given both as a percentage and as an Setp + 1.300 kg
absolute (weight) value, whether the Min + 1.235 kg
HEADER1 result lies within the “OK" range or not. Max + 1.365 kg
HEADER2 – Result in relation to configured
14.01.2002 09:43 thresholds: G# + 1.200 kg
-------------------- The deviation from the nominal weight T + 0.000 kg
Ref 2 o is given both as a percentage and as N + 1.200 kg
wRef + 1.200 kg an absolute (weight) value if the result
lies within the “OK" range. Stat LL
G# + 14.700 kg If the result is outside the defined --------------------
T + 0.300 kg tolerance limits, the last line indicates
N + 14.400 kg the status is indicated as follows: Result outside “OK" range (too heavy);
“HH" = too heavy; “LL" = too light. “threshold" printout
Qnt 12 o
-------------------- Result within “OK" range; “weight" or HEADER1
“threshold" printout HEADER2
Weighing in Percent Application: 14.01.2002 09:43
The block for initialization data HEADER1 --------------------
contains the reference percentage and HEADER2 Setp + 1.300 kg
reference weight. The results block 14.01.2002 09:43 Min + 1.235 kg
shows gross, net and tare weights, -------------------- Max + 1.365 kg
as well as the percentage, shown as Setp + 1.300 kg
either the loss or the residual amount. Min + 1.235 kg G# + 1.400 kg
Max + 1.365 kg T + 0.000 kg
Percentage as residue N + 1.400 kg
G# + 1.312 kg
HEADER1 T + 0.000 kg Stat HH
HEADER2 N + 1.312 kg --------------------
14.01.2002 09:43
-------------------- Lim + 0.92 %
pRef 100 % Diff.W+ 0.012 kg
Wxx% + 2.100 kg --------------------
G# + 1.859 kg
T + 0.200 kg
N + 1.659 kg
Prc 79 %
--------------------
92
Sorting Application: Total data record Individual printout when storing
The block for initialization data a component in tare memory
contains the die upper limits of Classes HEADER1 (by pressing O).
1 through 4. The results block contains HEADER2 Example: Print 2nd component
gross, net and tare weights, as well as 14.01.2002 09:43
the class that the sample belongs to -------------------- HEADER1
(1 through 5, where Class 5 means that n 3 HEADER2
the upper limit of Class 4 was exceeded). S-Comp+ 3.400 kg 14.01.2002 09:43
Cont.T+ 0.200 kg --------------------
HEADER1 -------------------- Cmp002+ 1.000 kg
HEADER2
14.01.2002 09:43 Individual / component printout Individual printout of components
-------------------- by pressing p
Lim1 + 10.000 kg If you press O, the header is printed Example: Print 2nd component:
Lim2 + 11.000 kg only once. Each component is printed
Lim3 + 12.000 kg automatically when you press O to HEADER1
Lim4 + 13.000 kg store it. HEADER2
If you are using a label printer, make 14.01.2002 09:43
G# + 9.700 kg sure a single label is large enough for --------------------
T + 0.000 kg the list of all components. For printer G# + 2.400 kg
N + 9.700 kg models YDP01IS and YDP04IS, you can T + 0.200 kg
configure manual form feed in the T2 + 2.200 kg
Class 11) operating menu. With the YDP02IS N + 0.000 kg
-------------------- printer, form feed is automatic after
each print command (fixed setting).
1)
Sample are sorted into 5 classes. Totalizing Application:
Any weight that exceeds the upper If an automatic printout is generated There is no data for the “initialization
limit defined for Class 4 is designated when you store a component, the data" block. If this block is activated
as Class 5, if the application is component weight is equal to the for the printout, a blank line is output.
configured for 5 classes rather than 3. current net weight. This is why Which values are shown in the results
components rather than net weights block depends on the program
are printed. operating status at the time of printing.
Animal Weighing Application: The following options are available:
The block for initialization data Example with menu setting 3.17.3
contains the number of measurements active, and 3 components weighed – Results printout (press c):
used for averaging. The results block (the total data record is shown above): Prints data in gross totalizing memory
contains the tare weight and the mean (“G"), net memory (“N") and the
value. HEADER1 number of transactions (“n").
HEADER2
HEADER1 14.01.2002 09:43 – Automatic standard/components
HEADER2 -------------------- printout (when O is pressed)
14.01.2002 09:43 Cmp001+ 1.200 kg
-------------------- – Manual standard/components printout
mDef 8 Cmp002+ 2.000 kg (when p is pressed)
93
When “manual" printing is configured GMP-compliant Printout ‘Clear preload' record
(print by pressing p), the transaction The GMP printout is made up of --------------------
counter value is not included on the 3 areas (see also the section entitled 14.01.2002 13:50
printout. “GMP Printout", above): Model CW2P1-30ED-LCE
– GMP header Ser.no. 12345678
Printout of components – Printout of data record Vers. 1.1007.12.1
Example with 3 transactions: (e.g., from the Weighing application) BVers. 01-25-01
– GMP footer --------------------
HEADER1
HEADER2 Linearization record Clear preload
14.01.2002 09:43 -------------------- completed
-------------------- 14.01.2002 13:00 --------------------
G# + 1.400 kg Model CW2P1-30ED-LCE 14.01.2002 13:52
T + 0.200 kg Ser.no. 12345678 Name:
N + 1.200 kg Vers. 1.1007.12.1
n 1 BVers. 01-25-01 --------------------
--------------------
G# + 3.400 kg
T + 0.200 kg Linearization Printout from Weighing application
N + 3.200 kg Wt.1 + 20.00 lb with multiple results
n 2 Wt.2 + 40.00 lb Example: 2 results:
Wt.3 + 60.00 lb --------------------
G# + 4.400 kg Wt.4 + 80.00 lb 14.01.2002 09:43
T + 0.200 kg completed Model CW2P1-30ED-LCE
N + 4.200 kg -------------------- Ser.no. 12345678
n 3 14.01.2002 13:02 Vers. 1.1007.12.1
Name: BVers. 01-25-01
Total data record (when c pressed) --------------------
with data as in printout above: --------------------
HEADER1
HEADER1 HEADER2
HEADER2 Calibration/adjustment record 14.01.2002 09:43
14.01.2002 09:43 -------------------- --------------------
-------------------- 14.01.2002 13:50 G# + 2.400 kg
Model CW2P1-30ED-LCE T + 0.200 kg
*G 9.200 kg Ser.no. 12345678 N + 2.200 kg
*N + 8.600 kg Vers. 1.1007.12.1 --------------------
n 3 BVers. 01-25-01 HEADER1
-------------------- -------------------- HEADER2
14.01.2002 09:44
Individual printout when storing External calibration --------------------
a component in tare memory Targ. + 50.00 lb G# + 3.400 kg
(by pressing O). Diff. - 0.01 lb T + 0.300 kg
Example: Print 2nd transaction External adjustment N + 3.100 kg
Diff. + 0.00 lb --------------------
HEADER1 -------------------- 14.01.2002 09:45
HEADER2 14.01.2002 13:52 Name:
14.01.2002 09:43 Name:
-------------------- --------------------
G# + 2.400 kg --------------------
T + 0.200 kg
N + 2.200 kg
n 2 ‘Set preload' record
--------------------
14.01.2002 13:50
Individual printout by pressing p Model CW2P1-30ED-LCE
Example: Print 2nd transaction: Ser.no. 12345678
Vers. 1.1007.12.1
HEADER1 BVers. 01-25-01
HEADER2 --------------------
14.01.2002 09:43
-------------------- Set preload
G# + 2.400 kg completed
T + 0.200 kg --------------------
N + 2.200 kg 14.01.2002 13:52
Name:
--------------------
94
Data Output Format (Line Format) Error Codes
You can output the values displayed in the line for measured Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
values and the weight unit with or without a data ID code. Select * * * E r r * * # # * * * * CR LF
menu item 7.2 to define whether the data ID code is included. or * * * E r r * # # # * * * * CR LF
Examples: *: Space
+ 235 pcs Without data ID code #: Error code number (2 or 3 digits)
Qnt + 235 pcs With data ID code
95
Error Codes
Automatic Data Output (SBI)
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 You can have results of measurement printed automatically1. You
S t a t * * * * * E r r * * # # * * * * CRLF can configure the autoprint function to print at certain intervals
(measured in display updates2) and define whether printing is
or S t a t * * * * * E r r * # # # * * * * CRLF dependent on stability of the weighing instrument3. How often
the display is updated depends on the operating status and model
*: Space of the equipment.
#: Error code number (2 or 3 digits)
Examples:
N + 153.00 g Net weight
ID code characters I 1 Meaning Stat Display blank
G# Gross value Stat L Display underload
N Net weight Stat H Display overload
T Application tare memory 1 Setting:
1) 3)
T2 Application tare memory 2 Automatic output without stability (menu item 6.1.4)
Diff Difference between maximum and minimum Automatic output with stability (menu item 6.1.5).
Factory setting: Manual with stability;
Targ. Exact calibration weight (value) i.e., automatic output function off.
Nom. Exact calibration weight with SBI output 2)
Time-dependent output:
nRef Reference sample quantity Intervals: 1, 2, 10 or 100 display updates
pRef Reference percentage (menu item 6.3-x; x=1, 2, 4, or 7).
Factory setting: 6.3.1; 1 display update
wRef Reference piece weight
Qnt Result from Counting (piece count) and
Neutral Measurement applications
mDef Target value for Animal weighing
x-Net Result from Animal weighing
Setp Target value for Checkweighing
Diff.W. Absolute difference (e.g., in kg)
in checkweighing
Lim Deviation in % in Checkweighing
Max Upper limit for Checkweighing
Min Lower limit for Checkweighing
Stat Status
Classx Sorting in classes
Limx Class limits
D Percentage (loss)
Prc Percentage (residue)
Wxx% Reference percentage weight
Cmpxxx Component no. xxx
Cont.T Contents of tare memory
in Net-total Formulation
S-Comp Total weight
in Net-total Formulation
PT2 Preset tare
n Transaction counter
*G Sum of gross weights in Totalizing
*N Sum of net weights in Totalizing
Ser.no Serial number of platform or indicator
1)
depends on scale type; e.g., not all units and characters are
available on scales verified for use in legal metrology.
96
Data Input Format Character Command Handshake
You can connect a computer to your kT_ Trigger T key function (tare)
indicator to send commands controlling The scale interface (Sartorius Balance
kNW_ Trigger n key function Interface = SBI) has transmit and receive
scale functions and applications via the (toggle weighing platforms)
interface port. All commands use the same buffers. You can define the handshake
format, starting with the ESC character kZE_ Trigger ( key function parameter in the Setup menu:
(ASCII 27) and ending with a carriage (zero)
return (CR; ASCII 13) and a line feed x1_ Output model designation – Hardware handshake (CTS/DTR)
(LF; ASCII 10). The total length of a of active weighing platform – Software handshake (XON, XOFF)
command is anywhere from 4 characters Example: “LP6200S-0C "
(1 command character between the start x2_ Output serial no. Hardware Handshake
and end described above) to 7 characters of active weighing platform
(4 command characters). Example: “0012345678 ” With a 4-wire interface, 1 more
character can be transmitted after CTS
x3_ Output software version (clear to send).
The table below shows the available
of active weighing platform
command characters; each command must
Example: “00-20-04 ” Software Handshake
be flanked by the start and end characters
as described above. x4_ Output software version
Example: the command character for of indicator The software handshake is controlled
output is “P" (“Output readout to port"). Example: “01-25-02 ” via XON and XOFF. When a device is
To trigger this command, send the string: x9_ Output serial no. of switched on, XON must be transmitted
“ESC P CR LF". indicator to enable any connected device to
Example: “0012345678 ” communicate.
Character Command x10_ Output model of indicator When the software handshake is
K Weighing mode 1 Example: configured in the Setup menu, the
L Weighing mode 2 “CW2P4-1500RR-LCE ” hardware handshake becomes active
M Weighing mode 3 z1_ Activate input for printout after the software handshake.
header 1
N Weighing mode 4
z2_ Activate input for printout The data transmission sequence is
O Block keys as follows:
header 2
P Output readout to data
interface The “underline" character is ASCII 95. Scale ––– byte –––> Computer
Q Output acoustic signal (trans- ––– byte –––> (receiving
Format for entering printout header lines: mitting ––– byte –––> device)
R Unblock keys
ESC z x a…a_CR LF where x=[header line] device) ––– byte –––>
T Tare and zero 1 or 2; a…a p to 20 characters text, <–– XOFF –––
(combination tare function) followed by the carriage return and line ––– byte –––>
f3_ Zero (see also the feed characters. ––– byte –––>
“kZE_" command) ...
f4_ Tare without zeroing (Pause)
(see also the “kT_" command) Synchronization ...
<–– XON –––
i_ Information about the ––– byte –––>
Data is communicated between the
indicator ––– byte –––>
indicator and a computer in the form of
Example of output: ––– byte –––>
messages (telegrams) consisting of ASCII
“CI2/012502/1" ––– byte –––>
characters, transmitted via the interface.
Meaning:
For error-free data communication, the
Indicator: Combics 2,
settings for baud rate, parity, handshake
Software version: 012502, Transmitting Device
mode and character format must be the
active weighing platform: 1
same at both ends.
kF1_ Soft key F1: trigger k key Once XOFF has been received, it pre-
function You can configure the interface settings vents further transmission of characters.
kF2_ Soft key F2: trigger c key in the Setup menu so that they match When XON is received, it re-enables the
function (Combics 2 only) those of the computer. You can also define transmitting device to send data.
kF3_ Soft key F3: trigger r key parameters in the indicator to make data
function (Combics 2 only) output dependent on various conditions.
Details on conditional data output are Receiving Device
kF4_ Soft key F4: trigger O key provided in the sections describing the
function (Combics 2 only) application programs. To prevent too many control commands
kF5_ Soft key F5: trigger w key from being received at one time, XON
function (Combics 2 only) If you do not connect a peripheral device is not transmitted until the buffer
kP_ Trigger p key function to the indicator interface port, this will not is almost empty.
Output to printer port generate an error message.
97
Pin Assignment Charts Models CIS1 and CIS2:
Terminals on the PCB
Models CISL1 and CISL2 COM1: 11-20 1-10 UniCOM: 1-4 1-10
(IP44 protected):
Front view
98
Second connection: For connecting an external battery pack and a bar code scanner 1)
(optional UNICOM interface not installed)
1 2 3 4 5 6 7 8 9 10
Top view
1)
Combics 2 models only
* depends on the UniCOM used
99
§ Connect the cable as follows:
– Route the cable through the cable gland.
– Tighten down the cable gland in accordance with the applicable regulations.
– Strip the casing from a section of the cable end (see illustration). The shield (1)
must have contact with the clamps (2).
Cabling diagram
Connection assignments for the cable from the indicator to an RS-232 PC interface
100
Cabling Diagram (Adapter Cable for PC)
Indicator PC
V24
PC
25 pin or 9 pin
— 3 2 3
+
4
+
— 5
20 4
2
3 2
20
5 8
6 6
14
7
7 5
101
Service
) Select “Setup"
If a password is required at this point (“CODE" appears in the indicator for 2 sec,
then the password prompt); proceed as below next to the described “password prompt."
) and confirm.
) and confirm.
S The Service mode is activated: An “S" is displayed in the upper right corner of the indicator.
102
Service Menu – 1-20 Place of adjustment – 11-7 Lock units:
When the Service mode is activated, the (geographical latitude and 11-7-1: lo-g to.
service technician can access additional altitude, alternatively the 11-7-2: Grams
menu items that are not visible when acceleration of gravity at the 11-7-3: Kilograms
the Service mode is deactivated. place of installation): ...
1-20-1: Latitude in degrees 11-7-21:Tons
The following advanced functions are 1-20-2: Height in meters above
available: sea level – 11-8 Calibration/adjustment unit:
1-20-3: Acceleration of gravity 11-8-1: User-definable unit
In the text menu (Setup): 1-20-4: Save values for 1-20 11-8-2: Grams
The following menu items are behind 11-8-3: Kilograms
the menu items Date (“DATE") and Code ...
(“CODE"): – 9-1 Factory setting / menu reset 11-8-21:Tons
– Service date “S-DATE" 9-1-3: Load standard configuration
– Memory number “MEM-NO" 9-1-4: Load verification configuration – 11-10 Save A/D converter
– Serial number of the Combics “SER-NO" configuration data:
11-10-1:Yes
In the numeric menu: – 11 Configuring the Analog/Digital 11-10-2:No*
The WP1 setup menu (“WP-1") and Converter:
the interfaces COM1 (“COM1") and – 12-1 Store platform serial number
UniCOM (“WUNICOM") is extended by – 11-1 Classes: (verified platform connected
the following additional setting options 11-1-4: Class l at WP2):
for configuring weighing platforms 12-1-1: Store serial number
WP-1 (“WP-1") and WP2 (“WP-2"): – 11-2 1st weight unit 12-1-2: Inactive (standard WP)*
(Copy from menu page 1-7):
– 1-9 Calibration/adjustment 11-2-1: lo-g to.
functions: 11-2-2: Grams
1-9-5: Internal linearization 11-2-3: Kilograms
(only for WP-2 connected to ...
COM1 and UniCOM) 11-2-21:Tons
1-9-6: External linearization with
default weights – 11-3 Range selection:
1-9-7: External linearization with 11-3-1: Single range scale
user-defined weights 11-3-2: Multi-interval scale1)
1-9-8: Set preload 11-3-3: Multiple range scale1)
1-9-9: Clear preload
– 11-4 Metrological data for
– 1-18 Enter the calibration and single range scale:
linearization weights: 11-4-1: Scale interval d
1-18-2: Enter Lin. weight 1 11-4-2: Verification scale interval e
1-18-3: Enter Lin. weight 2 11-4-3: Minimum capacity
1-18-4: Enter Lin. weight 3 11-4-4: Maximum capacity
1-18-5: Enter Lin. weight 4
– 11-5 Metrological data for
– 1-19 Adjustment without weight Multi-interval scale:1)
(Enter the ID codes for the 11-5-1: Scale interval d
load cell(s): 11-5-2: Verification scale interval e
1-19-1: Nominal capacity 11-5-3: Minimum capacity
1-19-2: Resolution 11-5-4: Range 1
1-19-3: Sensitivity in mV/V for cell 1 11-5-5: Range 2
(or mean averaged over all cells) 11-5-6: Range 3
1-19-4: Sensitivity in mV/V for cell 2 11-5-7: Maximum capacity
1-19-5: Sensitivity in mV/V for cell 3
1-19-6: Sensitivity in mV/V for cell 4 – 11-6 Metrological data for
In the case of several cells, Multiple range scale:1)
either the individual values are 11-6-1: Scale interval d
entered in 1-19-3 ... 1-19-6 or 11-6-2: Verification scale interval e
the mean of all cells in 1-19-3. 11-6-3: Minimum capacity
1-19-7: Store the values for 1-19 11-6-4: Range 1
11-6-5: Range 2
11-6-6: Range 3
11-6-7: Maximum capacity
1
) Not permitted in legal for trade
applications * = Factory setting
103
Configuring the Analog/ Configure the maximum capacity for Menu item 11-1 (Class)
each weighing range in a suitable
Digital Converter weight unit, without in decimal places. When using the “Standard configura-
Select the desired weight unit under the tion", this menu item is turned off.
11-7 “Lock units" menu item. For pur-
Purpose poses of display in the metrology line, When using the “Trade configuration,"
Configure the analog/digital converter the maximum capacity for all ranges (for weighing platforms scale verified
by selecting the required parameters must be configured so that they can be for use in trade/legal metrology)
in the Setup menu to adapt Combics displayed with a maximum of four dig- only menu item 11-1-4 is displayed.
for use with all commercially available its other than “0" for all accessible
strain-gauge load cells and analog units. Accuracy class l is already selected;
Sartorius CAPP, CAPS, IWA, IU, IF when you press ) to select menu
platforms. Access to settings for A/D The same applies to releasing and item 11-4-4, a circle (“o”) is shown
converter configuration is password- locking weight units 1 and 2. The indicated that this setting is active.
protected (Service-code). weight unit selected for configuration
cannot be excluded.
Features Range Selection (menu item 11-3)
With the menu access switch open, you Testing and Configuration for Use
can configure most of the following in Legal Metrology The weighing range of the weighing
parameters: A metrology label is included in the platform can be divided into one or
equipment supplied with the Combics. multiple ranges as desired. This is
– Toggling between weighing mode After the A/D converter has been suc- established by selecting the menu
“standard" or “trade" (scale verified for cessfully configured, the metrological item 11-3:
use in trade/legal metrology) data for all of the ranges on the label
can be keyed in. Attach the label below – Single range (11-3-1)
– Verification scale interval e the display window and seal with the The entire weighing range is divided
water-tight acetate overlay. into scale intervals which are dependent
– Scale interval d on the lowest scale interval d and the
maximum capacity. The readability for
– Minimum capacity the entire weighing range is always the
lowest scale interval d.
– Maximum capacity – Multiple range scale (11-3-2)
Scales with two or three weighing
– Maximum capacity for each weighing ranges. If the lower weighing range
range is exceeded, the scale switches to
the higher weighing range (lower
– Verification scale interval e for each resolution) and remains there. You
weighing range can only switch back to the lower
weighing range (higher resolution),
– Accessibility weight units when the scale is completely unloaded
You additionally need to attach the
label “The counting and neutral meas- after pressing the ( key.
These parameters are not reset when – Multi-interval scale (11-3-3)
you restore the factory settings. urement feature is not legal for trade”
below the display window on Combics 2 The “multi-interval scale" function
indicators.. divides the weighing range into as many
Depending on the weighing mode as 4 ranges with different readability.
selected, the configuration options for For the Trade configuration, only in
parameters not listed above are the same Under menu item 1-7, you can check to
make sure that only permissible weight the accuracy classes l permitted.
as they are in a corresponding Sartorius The readability automatically adjusts
weighing instrument that cannot be units are accessible.
each time the weight on the load
switched between “standard" or “trade" receptor exceeds a defined range limit.
modes. When a weight causes the weighing
Description of the Individual
Menu Items instrument to change from a higher to
Note on Settings a lower range, the higher resolution
Configure the A/D converter in the Menu item 9-1-3 / 9-1-4:
Select and load the configuration data is automatically selected. After taring,
numeric menu (weighing platform 1: the highest possible resolution of the
Level 1, weighing platform 2: Level 2 by first range is restored, even when the
selecting menu item 11-x). Please refer Before starting the actual A/D converter
configuration, you must define whether weighing platform is loaded.
to the illustration on the next page. For
accessing the numeric menu, also refer the weighing platform is to be used in
to the chapter entitled “Configuring the “standard" mode or in the “trade"
the Combics." mode (scale verified for use in trade/
legal metrology):
The A/D converter can be configured – Standard configuration
only in the Service mode. (Menu item 9-1-3)
Make sure you close the menu access or
switch after configuring the A/D con- – Trade configuration (Menu item 9-1-4)
verter; otherwise, “H” (too heavy) and
“L” (too light) cannot be displayed.
104
Setup WP-1 9.1.3 Standard 11.3 Ranges 11.3.1 Single-range scale
Access only configuration 11.3.2 Multiple range scale
with Service 11.3.3 Multi-interval scale
password
11.4 Single-range 11.4.1 Scale interval d
scale 11.4.4 Maximum capacity
1
) not permitted for legal for trade applications
105
The ranges are divided as desired by Maximum capacity Max. cap. Adjustment without Weights
selecting the assigned menu item with (menu item 1-19)
key k and confirm with the ) key. The maximum capacity is the maximum
A circle (o) appears behind the selected load that may be placed on the weigh- The parameters for “Nominal capacity"
menu item as a marker. ing instrument. If the weight exceeds in kg (menu item 1-19-1) (specification
this capacity, the weighing instrument of the strain-gauge beam), “Resolution"
will display H. in kg (menu item 1-19-2) and “Sensitiv-
Scale Interval d The scale intervals are calculated from ity" in mV/V (menu item 1-19-3) are
the maximum load and the lowest scale converted to internal quantities. Once
The lowest scale interval d indicates the interval d. In legal for trade applica- the A/D converter configuration data
weighing instrument's resolution. You tions, the number of verification scale has been stored, these parameters can
can only enter it in the scale intervals divisions may not be greater than no longer be read out. For platforms
1, 2, 5, 10, 20 etc. If you have selected 10,000 e. In the “standard" weighing with multiple load cells, the sensitivity
the “Trade" setting (accuracy class l) mode, as opposed to “trade," you can of the additional cells is entered by
the lowest scale interval d is not define a “SuperRange" to increase the selecting menu item 1-19-4 (cell 2) to
prompted because it is equal to the permitted intervals: menu item 1-19-6 (cell 4).
verification scale interval e. For example, a maximum capacity of Save the entered data by selecting the
50,00 lb combined with a smallest menu item 1-19-7.
interval d of 0.01 lb, yields 50,00 scale
Verification Scale Interval e intervals. These parameters, however,
may be influenced by physical ! Important Note:
When “Standard configuration” is limitations. Please do not forget to set the menu
active, this menu item is not displayed. access switch (15) to “locked" when you
have finished configuring the
The verification scale interval e indicates Lock units (menu item 11-7) A/D converter and performing the
the resolution of the weighing instru- adjustment (calibration/adjustment
ment in legal metrology. With accuracy With this function, you can make and linearization).
class l, the verification e is particular weighing units inaccessible
the same as scale interval d. Thus, with during weighing. Accessible units are Weighing results can no longer be
classes l, the lowest scale interval d is marked by a circle (o) on the indicator influenced by the Combics after the A/D
not prompted. (multiple range selection is possible). converter has been configured.
The range of available weighing instru-
ment functions is defined in the A/D
Minimum capacity Min. cap. Calibration/Adjustment Unit converter. The following functions, for
(Menu item 11-8) example, can be activated:
When “Standard configuration” is – Read weighing results
active, this menu item is not displayed. The calibration unit defines the weight – Tare
unit with which calibration must be – Adjustment
The minimum weight for weighing performed. This unit will be used even – Read tare value
instruments of accuracy class l is if a different unit has been selected for – Save/delete tare input
20 e and for accuracy class, 10 e. the weighing mode.
Caution: The minimum capacity limits
have the function of warning the
equipment operator that loads below Store settings
these limits lead to cumulative addition (menu item 11-10)
of the tolerances, which can result in
significant errors of measurement. In Save the A/D converter configuration
Germany, for example, weighing below by selecting menu item 11-10-1.
the minimum capacity is not permitted.
Once these parameters have been
configured, the A/D converter in
Range 1, Range 2, Range 3 conjunction with the weighing cell is
defined as a weighing instrument. The
Enter the limits for the individual A/D converter, in conjunction with the
ranges. When a given limit is exceeded, weighing platform, can now be used
the accuracy changes. like any standard weighing platform.
The following applies for all limits:
Range 1 < Range 2 < Range 3 In addition, the weight unit must be
< Max. cap. defined and the weighing platform
Thus, you can divide the weighing adjusted (calibration/adjustment
range into 4 ranges. The display resolu- and linearization). For more detailed
tion changes in intervals of 1, 2, 5, 10, instructions, see the chapter entitled
20 etc. The lowest resolution is the “Calibrating in the Service Mode."
lowest scale interval d. Ranges you do
not require should be set to 0.
106
A/D Converter Configuration (Examples)
Setup
(see also the chapter on “Operating the Combics", section entitled “Calibration, Adjustment")
S § Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (toward the interface ports) (“accessible" position).
§ Set to Service mode (see the corresponding section at the beginning of this chapter).
Configuration for using the scale in legal metrology (menu item 9-1-4).
107
) Confirm selection.
If the “Trade" configuration was activated, then at this point the menu line 11-1
(Select Class) is displayed.
Otherwise, i.e. in “Standard" configuration, this menu line is skipped and the menu line
11-2 (1st weight unit) is displayed. In this case, the user must use the k key to select
the menu line 11-3 and proceed as of described behind the figure in menu line 11-3.
) If the Combics is to be configured for use in legal metrology (i.e., the “Trade" configuration
was activated), the accuracy class is already configured (class l)
The setting for menu item 11-1-4 for class l is marked by a circle (o) on the indicator.
) Confirm selection.
In the present example, the menu item 11-3-1 (“single range") is pre-set.
k)
if necessary, Select “Multiple range scale", confirm selection.
For more detailed instructions on how to select the weighing range, see the section on
“Description of the Individual Menu Items" at the beginning of this chapter.
Enter the scale interval, range limits (only required for multiple range scale or
k multi-interval scale) and the maximum load:
Depending on the weighing range selected, the relevant menu line is displayed, i.e.
for “single range" (menu item 11-3-1): Input to menu page 11-4,
for “Multiple range scale" (menu item 11-3-2): Input to menu page 11-5,
for “Multi-interval scale" (menu item 11-3-3): Input to menu page 11-6.
The other two menu lines are not accessible.
In the present example, 11-3-1 (single range) is active. Therefore, the program opened
the menu line 11-4 for input. The menu lines 11-5 and 11-6 are not accessible.
If the “Trade configuration” is active (class l), then the menu item for setting the scale
interval d is skipped in the submenus 11-4, 11-5 and 11-6 and the second menu item
(for setting the verification scale interval e) is opened, since e = d for these scales.
This menu item is described on the next page.
108
Menu item 11-4-1 is displayed.
If the “Trade” configuration is active (class l), menu item 11-4-2 is shown
(see the last paragraph on this page).
Exit an activated menu item without storing entry (For example in the case of an inadvertent
selection): Press the ( key.
Move the cursor 1 position to the right (in the present example from the 1st zero to the
decimal point): Press the ) key. This confirms the character (number, minus sign or
decimal point) at the previous cursor position. If the cursor is at the far right, pressing )
confirms the input.
Move the cursor 1 position to the left: Press the ( key. If the cursor is in the 1st position
(all the way to the left), then you exit the menu item.
Delete the displayed contents and overwrite with your own input: press the k key.
A zero is written in the first position.
Every time the k key is pressed, you scroll one page forward (1, 2 ... 9).
In the 1st position (all the way to the left), there is a minus sign “-" after the number “9"
if you want to enter a negative number (... 8, 9, -, 1, 2 ...). In all other cursor positions,
there is a decimal point “." after the number “9" which can be selected (... 8, 9, ., 1, 2 ...).
If the default value displayed (here: “0.001 kg") corresponds with the desired scale interval,
then store the value by pressing the ) key (for longer than 2 sec.)
Otherwise use the k and ) keys (if necessary, also the ( and p) keys) to overwrite
the default value and confirm by pressing the ) key.
If the “Trade configuration” is active (class l), this menu item is opened automatically
(if “Multiple-range scale” is set: 11-5-2; with “Multi-interval scale”: 11-6-2). If the
“Standard configuration” is active, this item is not shown. See also the remarks on the
previous page.
109
) Show the configured verification scale interval e.
In accuracy classes l and m the verification scale interval e = d (see also the notes
at the beginning of the section). That is why the program shows this as default value.
k) When the “Trade configuration" is active, call up the menu item 11-4-3 (minimum capacity)
(if “Multiple-range scale” is set: 11-5-3; with “Multi-interval scale”: 11-6-3).
This menu item is not displayed when the “Standard configuration" is selected.
) When the “Trade configuration" is selected, activate the menu item by pressing ), set the
minimum capacity and confirm the entered value by pressing and holding the ) key (> 2sec).
k) Open the next menu item. If “Single-range scale” is set, this is 11-4-4, the last menu item
(maximum capacity).
If “Multiple-range scale” or “Multi-interval scale” is set rather than the A/D converter setting
“Single-range scale”, then the range delimiters must be entered under 11-5 (for the former)
or 11-6 (for the latter). Each of these menu lines has 3 items to choose from, to set limits 1,
2 and 3. The last menu item (11-5-7 or 11-5-8) sets the maximum capacity.
)
Menu item 11-10-2 is currently active (“Do not save configuration").
110
The settings for the A/D converter configuration are stored. For a short time, the “Busy"
symbol appears on the indicator, then the Combics is restarted.
After the self-test when all segments are displayed, the Combics is in the weighing mode.
Adjust the scale: see the chapter entitled “Operating the Combics", under “Calibration,
Adjustment" or the Service Manual.
Following A/D converter configuration, change the menu access switch from
“open" to “locked":
§ if necessary, remove the cap from the back on the left of the Combics housing.
§ Move the switch to the left (to the “locked" position).
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment"
or the Service Manual.
Setup
(see also the chapter entitled “Operating the Combics", under “Calibration, Adjustment")
§ Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (“open" position).
§ Service mode setting up (see corresponding section at the beginning of this chapter).
) and confirm.
In the picture shown, the tare at power on is set: the menu item 1-13-1 is selected.
111
k) Switch off tare at power on: Call up menu item 1-13-2.
)(( Save setting for menu item 1-13-2 and return to menu “1."
Call up “Standard menu". If the “Trade configuration” is active, the scale cannot be adjusted
without weights by entering the load cell data.
) confirm selection.
Configuration: range selection (single range, multiple range scale, multi-interval scale),
scale interval d and maximum capacity Max. cap. (if necessary, also enter range limits).
In the present example, the menu item 11-3-1 (“single range") is pre-configured.
k)
If necessary, press to select “Multi-interval scale", confirm selection.
From more detailed explanations on selecting the range, please refer to the section
entitled “Description of the Individual Menu Items" at the beginning of this chapter.
Enter the scale interval, range limits (only required in the multiple range or multi-interval
k mode) and maximum capacity:
112
Depending on the weighing range selected, the relevant menu line is displayed, i.e.
for “single range" (menu item 11-3-1): Input to menu page 11-4,
for “Multiple range scale" (menu item 11-3-2): Input to menu page 11-5,
for “Multi-interval scale" (menu item 11-3-3): Input to menu page 11-6.
The other two menu lines are not accessible.
In the present example , 11-3-1 (single range) is active. This is why the program opened
menu line 11-4 for input. Menu lines “11-5" and “11-6" are not accessible.
The configuration for the scale interval d is displayed. The flashing number marks the
current cursor position. This marked number can be changed.
The number in the input field can be changed to the desired value by repeatedly moving
the cursor and changing the respective number
If the default value displayed (here: “0.005 lb") corresponds with the desired scale interval,
store the value by pressing the ) key (press for longer than 2 sec.)
Otherwise use the k and ) keys (if necessary, also the ( and p) keys to overwrite
the default value and confirm by pressing the ) key.
In the present example, the scale interval d was set to 0.01 lb.
113
If, instead of the A/D converter configuration, the configuration “single range”, “multiple
range” or “multi-interval mode" was selected, then the respective weighing range limit
must be additionally entered into the assigned menu lines (11-5 for “multiple range scale"
and 11-6 for “multi-interval scale"). These menu lines thus contain 3 additional menu
items (the limits for the ranges 1, 2 and 3)
)
Menu item 11-10-2 (“configuration not saved") is pre-configured.
The settings for the A/D converter configuration are stored. For a short time, the “Busy"
symbol appears on the indicator, then the program returns to indicator display of menu item,
then Combics is restarted.
S After the self-test (i.e. when all segments are displayed), the Combics is in the weighing mode.
Activate the Service mode again (see the corresponding section at the beginning of
this chapter ).
Enter the nominal capacity of the load cell(s), for example 50.00 lb.
On a weighing platform that consists of several load cells, the nominal capacity must
be multiplied appropriately.
Example: The weighing platform consists of 4 load cells with 50 lb each. In this case,
the nominal capacity (= maximum capacity) 4 + 50 kg = 200 lb.
114
k) Activate menu item 1-19-2.
Enter the resolution (lowest scale interval d) (For example 0.01 lb). This value must
corresponed with the value entered under menu item 11-4-1 (and/or 11-5-1 or 11-6-1).
) Confirms setting.
) Confirms setting.
k) On a weighing platform that consists of several load cells, enter the sensitivity of
the other load cells (max. 4 load cells) under the menu items 1-19-4 (for cell 2) to
k) 1-19-6 (for cell 4). If there are less than 4 load cells for the corresponding menu items,
enter the value “0.000".
k)
k) Call up menu item 1-19-7 (save values entered on menu page 1-19).
)(( Save ID codes of the load cell(s), return to menu page “1."
Set/delete preload: see corresponding section on the subsequent pages.
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment"
or the Service Manual.
Restart: Switch the terminal off and on again.
After completing the A/D converter configuration, move the menu access switch from the
“open” to the “locked” position, as follows:
§ Remove the cover plate from the back of the indicator housing
§ Move the switch to the left (“locked”).
See also the “Calibration/Adjustment” section in the chapter entitled
“Operating the Combics”, or refer to the service manual.
115
Entering Geographical The following exception applies in The geographical data cannot be edited
Germany for scales with 3000 e: unless the menu access switch is open.
Latitude, Local Elevation If the latitude is set to 51,00° and the If the indicator is part of a verified weigh-
and Acceleration of Gravity altitude 513 m above sea level, which cor- ing system, the verification seal must be
responds to an acceleration of gravity of broken to change this data. Afterwards, the
Purpose 9.810 m/s-2, then the scale can be used scale must be re-verified.
To perform an external adjustment of anywhere in Germany (“Zone D”).
the weighing instrument at a place of If the acceleration of gravity has been
adjustment that is not the same as the These values are calculated for Germany entered, then this value takes precedence
place of installation. based on a mean value for the Earth’s over the geographical latitude and the
acceleration. The greater the precision of elevation. When the menu item 1-20-1
When a weighing instrument changes the geographical data entered, the greater is called up, the value “99999.99" is
location and is used somewhere else, its the precision achieved with the weighing displayed; when the menu item 1-20-2
sensitivity changes as well. The sensitiv- instrument; the tolerance range, however, is called up, the value “9999999" is dis-
ity is dependent on the acceleration of is also restricted accordingly (see above). played. In the converse case (geographical
the earth's gravity, which in turn is latitude and local elevation were entered),
affected by the geographical latitude Procedure the value “0.000000" is displayed for the
and the elevation of the place of instal- The geographical data stored in the menu item 1-20-3.
lation. In general, the acceleration of indicator at the factory applies to Germany
gravity increases in the direction of the (“Zone D”) (see above). To display geographical data during the
poles (i.e. with increasing degree of adjustment procedure, select menu item
latitude) and decreases the greater the Before adjusting the complete weighing 8-12-2 (factory setting: 8-12-1; “Do not
distance is from the center of the earth system, make sure the geographical show geographical data”). With this setting
(i.e. with increasing elevation). reference data stored in the indicator active, the indicator shows which set of
matches the values for the place of adjust- data is used for adjustment; either the lati-
If the precise location of the place of ment (whether at the factory or at the tude and elevation or the acceleration of
installation is known; i.e. its geographi- place of installation). If the values do not gravity. Press ) to confirm the data
cal latitude in degrees (northern or match, enter the latitude and elevation shown; or, if the data is invalid, press (
southern latitude) and the elevation under menu items 1-20-1 and 1-20-2, to cancel the adjustment routine.
in meters above sea level, then the or the acceleration of gravity under item
weighing instrument can be adjusted 1-20-3. Then adjust the scale. If the place If the display of geographical data (menu
at the factory in preparation for its of adjustment is not the same as the place item 18-12-2) is active, the adjustment
future place of installation, provided of use, enter the data describing the place procedure is as follows:
that the geographical latitude and of use. Following adjustment, close the After starting the adjustment procedure
elevation at the place of adjustment menu access switch. The scale can now be “CAL" and when the data “local elevation"
are also known. Rather than the used at the place of installation, or any- and “geographical latitude" is used, the
geographical latitudes and local where within a tolerance zone (see above) word “ALTITUD" appears for 2 seconds,
elevations of the places of adjustment around the place of installation. If the scale followed by the settings for the elevation
and installation, it is sufficient to is installed in Germany, you can enter the (in meters above sea level). Press ) to
know the acceleration of gravity at data for “Zone D” (51,00 degrees, confirm the data, or ( to cancel the
both the places. 513 m above sea level). In this case, adjustment routine. Then, the word
the scale can be used anywhere within “LONGITUD" appears for 2 seconds, fol-
The following data, describing the place Germany. This setting is recommended for lowed by the settings for the geographical
of manufacture (Sartorius in Goettingen, weighing equipment dealers who do not latitude in degrees. Press ) to confirm
Weender Landstrasse 94-108) is used as need to know the exact geographical data the data, or ( to cancel the adjustment
reference data: when delivering to customers in Germany. routine. After this, the prompt to place
– Geographical latitude: the adjustment weight appears. If, instead
51° 32’ = 51.53 degrees If a service technician performs a span of elevation and geographical latitude,
– Elevation: 151 m adjustment (for example after one or more the acceleration of gravity was entered,
– Acceleration of gravity: 9.811590 m/s-2 load cells have been replaced), it is con- the word “GRAVITY” appears for 2 seconds
If this reference data is stored in the ducted without changing the values stored after “CAL," followed by settings for the
indicator, the adjustment factor does in the menu items 1-20-1 and 1-20-2, value for the acceleration of gravity.
not need to be corrected for the place or 1-20-3. Press ) to confirm the data, or ( to
of installation. cancel the adjustment routine.
For this adjustment, either the geographi-
After closing the menu access switch, cal latitude and elevation (menu items
the geographical data is stored in the 1-20-1 and 1-20-2) or the acceleration
A/D converter. of gravity (menu item 1-20-3) for the place
of installation must be entered. This data
The tolerance zones, e.g. for a scale can be obtained from the relevant Cadaster
with 3000 e, are ± 100 km for the Agency or the State Surveying Authorities.
latitude and ± 200 m for the elevation
above sea level.
116
Enter of the place of adjustment with subsequent external adjustment (Example)
Setup
(See also the chapter entitled “Operating the Combics", under “Calibration, Adjustment")
S § Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (“accessible" position).
§ Service mode setting up (see corresponding section at the beginning of this chapter.
The picture shows that tare at power on is “on": The menu item 1-13-1 is selected.
If the geographical latitude and elevation of the place of installation are known, enter this
data by selecting menu item 1-20-1 (degree of latitude) and 1-20-2 (elevation).
To enter the acceleration of gravity at the place of installation select menu item 1-20-3.
Any value entered for the acceleration of gravity under 1-20-3 takes priority over the pair
of values “geographical latitude" and “elevation."
In the example shown here, the parameter “acceleration of gravity" (menu item 1-20-3)
has been entered. Therefore, the input fields for the menu items 1-20-1 and 1-20-2
are empty.
In the example shown here, the last geographical latitude entered was the setting
for Germany (“Zone D”). The scale can be used anywhere in Germany with this setting,
if the elevation entered under menu item 1-20-2 is 513 m above sea level.
Enter the required latitude as an absolute number (convert the angular minutes to
decimal places). Press ) to confirm the data entered.
117
)k) Call up menu item 1-20-2 (elevation of the place of installation).
The input field is empty because the parameter “acceleration of gravity" (menu item
1-20-3) has been entered.
In the example shown here, the last elevation entered was the setting for Germany
(“Zone D”). The scale can be used anywhere in Germany with this setting, if the latitude
entered under menu item 1-20-1 is 51.00.
Enter the elevation of the place of installation (in meters above sea level, and if necessary,
the reference value “151"). The elevation may also be negative under certain circumstances,
(for example, when the weighing instrument is being set up in a mine).
Confirm the entry by pressing the ) key.
In the example shown here, the parameters “geographical latitude" and “elevation"
(menu items 1-20-1 and 1-20-2) were last entered. Therefore, the input field for this
menu item is empty.
In the example shown here, the last value entered for acceleration of gravity was the setting
for Germany (“Zone D”). The scale can be used anywhere in Germany with this setting, which
is a mean value calculated for the whole of Germany.
Enter the acceleration of gravity at the place of installation (in m/s-2, if necessary, reference
value “9.811590"). Confirm the entry by pressing the ) key.
The “Busy" symbol appears in the indicator as proof that the parameters entered
have been stored.
After the self-test when all segments are displayed, the Combics is in the weighing mode.
Scale adjustment:
118
Adjusting the scale:
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment."
Example: The parameters “elevation of the place of installation" and “geographical latitude"
are entered (menu items 1-20-1 and 1-20-2).
The elevation of the place of installation entered in meters above sea level is displayed.
In this example, the display shows the elevation valid for Germany (“Zone D”).
The prompt to place the calibration/adjustment weight appears (for example, 50 lb).
Please refer to the chapter entitled “Operating the Combics", under “Calibration/Adjustment"
for more details on the calibration/adjustment procedure.
The prompt to place the calibration/adjustment weight appears (for example 50 lbg).
For more details on the Calibration/adjustment procedure, please refer to the chapter
entitled “Operating the Combics", under “Calibration, Adjustment."
119
After completing the A/D converter configuration, move the menu access switch from
“open" to “locked":
§ If necessary, remove the cap from the back on the left of the Combics housing.
§ Move the switch to the left (“locked" position).
See also the chapter entitled “Operating the Combics", under “Calibration, Adjustment"
or the Service Manual.
120
Error Codes and Messages
Error codes are shown in the main display. Err codes and messages are displayed continuously. Inf messages are displayed for 2 seconds,
after which the program then returns automatically to the weighing mode.
121
Care and Maintenance Recycling
Service $ If used in the food industry, use a Sartorius products are packaged
Regular servicing by a Sartorius cleaning agent suitable for the working to ensure safe shipment using
technician will extend the service life of environment. environmentally friendly materials. If
your indicator and ensure its continued you do not need to save the packaging
weighing accuracy. Sartorius can offer § After cleaning, wipe down the indicator after successful installation of the
you service contracts, with your choice with a soft, dry cloth. equipment, you should return it for
of regular maintenance intervals recycling.
ranging from 1 month to 2 years. ! Do not use jets of water or compressed
air to clean the indicator. For information on recycling options,
The length of the service interval including recycling of old weighing
depends on the operating conditions Replacing the Dust Cover equipment, contact your municipal
at the place of installation and on > Replace damaged dust covers. waste disposal center or local recycling
your requirements. depot.
§ Place the new dust cover on the
indicator and press down on the front If the equipment contains batteries,
Repairs and back along the edges until the remove them before disposal. Batteries
! Disconnect defective equipment from cover is firmly seated . are treated as hazardous waste and
power immediately (unplug the power must be disposed of separately. Please
cord from the wall outlet). Repair work contact your municipal waste disposal
must be performed by trained service Safety Inspection center for details.
technicians authorized by Sartorius Safe operation of the device is no
and using original Sartorius parts. longer ensured when:
Any attempt by untrained persons to
perform repairs may result in hazards – there is visible damage to the device
for the user. or the power cord
Note: In case of defects during the war-
ranty period, please return the entire – the built-in power supply no longer
indicator to the factory. functions properly
! If a cable or cable gland is defective or – the device has been stored for a
damaged, replace the entire assembly. relatively long period under unfavorable
conditions (e.g., extreme moisture)
! Do not open the indicator housing
while the indicator is electrically charged – the equipment has been subjected to
(live). Wait at least 10 seconds after rough handling during shipment
disconnecting the equipment from power
before opening the housing. Because If there is any indication that safe
the fitting surfaces of the housing parts operation of the device is no longer
directly affect the IP protection rating, warranted:
make sure to prevent any damage that
would affect the seal when opening and § Disconnect from AC power: unplug
closing the indicator housing. the equipment from the wall outlet
(mains supply)
Cleaning
The Sartorius Combics indicators are > Lock the device in a secure place to
designed in compliance with directives prevent it from further usage.
of the EHEDG (European Hygienic
Equipment Design Group) and in § In this case, notify your nearest
accordance with HACCP (Hazard Analysis Sartorius Service Center.
and Critical Control Points) standards
for prevention of contamination, which Maintenance and repair work may be
means they are particularly easy to performed only by authorized Sartorius
clean and disinfect. Service technicians who:
! Disconnect from AC power: unplug the – have access to the required service and
equipment from the wall outlet (mains maintenance manuals
supply) and disconnect the data cable
(if connected). – have attended the relevant service
training courses
! Make sure that no liquid enters the
indicator housing. ! The seals affixed to the equipment
housing indicate that the equipment
! Do not use any aggressive cleaning may be opened only by authorized
agents (solvents or similar agents). service technicians, to ensure safe and
trouble-free operation of the equipment
§ Clean the indicator using a piece of and to maintain the conditions for
cloth dampened with a detergent. warranty coverage.
122
Overview
Specifications
When used in standard applications
(as opposed to legal metrology):
– Number of scale intervals <50,000 d
In legal metrology:
Accuracy class l,
Number of verification scale divisions
when used as
– Single-range scale <10,000 e acc. to NIST
Load cell connection:
– Supply voltage 9 V (± 4.5 V)
– Bridge impedance 83 O to 2000 O
– Available sensor technology 4- or 6-conductor technology
123
Dimensions (Drawings)
90
187
49
302
124
Accessories
Product Order No.
Legal for trade printer with functions for date, time and YDP03-0CE
statistical evaluations; liquid-crystal display
Legal for trade strip and label printer with thermal print head,
paper width 101 mm, with adapter cable (12-contact round
male connector) and external power supply
Adapter cable YCC01-01CISLM3 required for
Combics model CISL indicator.
Can only be used with customizable printout
configuration tool option.
Adapter cable YCC02-D09F6 required for
Combics model CIS indicator.
– With 120 V power supply(USA) YDP12IS-0CEKV
Legal for trade strip and label printer with thermal print head,
Paper width 60 mm, with adapter cable (12-contact round
male connector) and external power supply.
Adapter cable YCC01-01CISLM3 required for
Combics model CISL indicator.
Adapter cable YCC02-D09F6 required for
Combics model CIS indicator.
– With 120V power supply (USA) YDP04IS-0CEUV
125
Product Order No.
Flexible formatting options for printouts (e.g., for bar codes Information available
with variable font size, graphics, etc.) on request
126
Product Order No.
Connecting cable with cable gland for car battery für (2m)
(for Combics models CIS only) YCC02-CB02
Characteristics:
– IS weighing platforms process weighing data independently
of the indicator
– IS platforms can be internally adjusted
– …-0CE models have a separate approval number, on a tag
affixed to the cable
– Please observe the conditions described in the instruction
manual for the platform connected
127
General Password
k repeatedly, );
k repeatedly, );
k repeatedly, );
k repeatedly, ); The password is shown on the display.
Press and hold ) ( > 2 sec.) Store input and exit the Setup menu
General password:
40414243
Service password:
202122