Professional Documents
Culture Documents
January, 2021
1 Introduction
2 Sieve Size analysis
Theoretical Background
Purpose of the test
Apparatus
Sample preparation and test specimen
Test Procedure
Test computation
3 Hydrometer Analysis
Theoretical Background
Purpose of test
Apparatus and Reagent
Sample preparation and test specimen
Test Procedure
Test computation
4 Combined analysis
the results of particle size analysis are presented in the form of the
cumulative distribution of the percentage soil passing.
Mass of soil retained (on each sieve) = Mass of soil + sieve - Mass of
sieve [column 5]
Cumulative mass retained (on a given sieve) = total mass of soil
coarser than the sieve opening = the sum of all mass retained on all
coarser sieves[column 6]
Cumulative retained percentage = ratio of cumulative mass retained
column6
to the total mass multiplied by 100 [column7 = Totalmass x100]
percentage passing = 100 - cumulative percentage retained [column 8
= 100 - column 7]
The results of presented graphically in a semi-log plot where the
particle size is in log scale
Theoretical Background
based on stocks law; which relates the terminal velocity of a spherical
particle falling in a viscous fluid to the particles diameter.
X At the end of 1 min set the cylinder in a convenient location and take
hydrometer readings
X The hydrometer readings should be done for following time intervals
measured from the beginning of sedimentation: T = 2, 4, 8, 15, 30,
60, 120, 1440 min, RH
Along with the test cylinder a calibration cylinder is prepared to
account for errors in
Dispersing Agent (Sodium Hex-Meta Phosphate),
Temperature and
Meniscus
The reading is recorded as the same time and interval as the test
cylinder. RC