Professional Documents
Culture Documents
EEG-1250
Electroencephalograph
TD.EEG-1250_A
This technical data may be revised or replaced by Nihon Kohden at any time without notice.
Some products may not be available in your country.
Specifications
Data Acquisition
JE-921A electrode junction box
Number of input jacks
EEG input jacks 25
Extra inputs 4
Bipolar input jacks (pairs) 7
The extra jacks (X1-X4) can be used as bipolar jacks. To select extra jack or
bipolar jack, refer to Section 5 “System Programs”.
DC input jacks 4
SpO2 connector 1
CO2 connector 1
Input impedance
EEG input/extra input 100 MΩ
DC input 1.5 MΩ
Input circuit current 5 nA or less
Internal noise level
EEG input/extra input 1.5 µVp-p or less (0.53 to 60 Hz)
DC input 10 mVp-p or less
CMRR
EEG input/extra input 105 dB or greater (60 Hz)
Bipolar input 100 dB or greater (60 Hz)
Low-cut filter 0.08 Hz (Time constant: 2 s)
High-cut filter 300 Hz (−18 dB/oct)
Offset tolerance ±600 mV
A/D conversion 16 bits (97 nV/LSB)
Sampling and hold All electrodes at the same time
Sampling frequency 100, 200, 500, 1000 Hz, selectable
Data Processing
Sensitivity
EEG INPUT OFF, 1, 2, 3 (2.5), 5, 7, 10, 15, 20, 30, 50, 75, 100, 150, 200 μV/mm
DC INPUT OFF, 10, 15, 20, 30, 50, 75, 100, 150, 200 mV/mm
Time constant (Low-cut filter) 0.001, 0.003, 0.03, 0.1, 0.3, 0.6, 1.0, 2.0, 5.0, 10.0 s
0.016, 0.03, 0.08, 0.16, 0.27, 0.53, 1.6, 5.3, 53, 159 Hz (−6 dB/oct)
High-cut filter 15, 30, 35, 60, 70, 120 (−12 dB/oct), 50 (RAPID), 300 Hz (−18 dB/oct)
AC filter 50 or 60 Hz, (rejection ratio: 1/25 or more)
Calibration waveform
Waveform shape 0.25 Hz step wave or 10 Hz sine wave
Voltage 2, 5, 10, 20, 50, 100, 200, 500, 1000 μV (×1000 for DC input signal)
Impedance check
Indication on the screen All electrodes are displayed on the screen in electrode position layout.
Impedance for each electrode is displayed and electrodes with impedance higher
than the preset impedance threshold are highlighted.
Indication on LED LEDs on the electrode junction box with impedance higher than the preset
impedance threshold light.
Impedance threshold 2, 5, 10, 20 and 50 kΩ
2/12
Pattern 36 sets of programmable montages combined with programmable individual
amplifier settings
Reference electrode selector A1 → A2, A1 ← A2, A1 ↔ A2, A1 + A2, VX, AV (with unsuitable electrode
deletion function), Aav, Org, SD, BN and OFF
Display
Display size 23.8 inches
Display type Touch panel
Display resolution Up to 1920 dots × 1080 lines
Number of display channels Up to 64 and one mark channel
Waveform display color 16 colors
Waveform position adjustment Provided
Waveform sweep speed 0.1, 0.2, 0.5, 1, 2, 5, 10, 15, 20, 30, 60 s or 2, 3, 5 min/page
Timing mark 0.1, 1 s
Time scale off, 0.2, 1 s
Event mark Displays at the bottom of the screen
EEG scale Provided
Photic Stimulator
Optional LS-120A photic stimulator control unit
Maximum flash energy 1.28 J/Single
Stimulation modes 3 automatic (30 steps, programmable), manual, and single
Mode of operation Continuous operation with intermittent loading
Duty cycle Max. 5 minutes continuous operation in 30 minutes
Automatic stimulation
Stimulus rate 0.5, 1 to 33 (1 Hz steps), 50 and 60 Hz
Stimulation period 1 to 99 seconds in 1 second steps
Pause period 1 to 30 seconds in 1 second steps
Safety
Safety standard:
EEG-1250 (Op No. 50A) electroencephalograph
IEC 60601-1: 1988
IEC 60601-1 Amendment 1: 1991
IEC 60601-1 Amendment 2: 1995
IEC 60601-1-1: 2000
IEC 60601-1-2: 2001
IEC 60601-1-2 Amendment 1: 2004
3/12
IEC 60601-2-26: 2002
IEC 60601-2-40: 1998*
* Applied only for bipolar inputs (X1 to X4) on the electrode junction box when
acquiring EMG.
Electromagnetic Compatibility
EEG-1250 (Op No. 50A) electroencephalograph
IEC 60601-1-2: 2001
IEC 60601-1-2 + Amendment 1: 2004
CISPR11 GROUP 1 CLASS A: 2003 Amendment 1: 2004, Amendment 2: 2006
Power Requirements
Line voltage AC 100 to 240 V ±10%
Line frequency 50/60 Hz
Power consumption 400 VA or less
4/12
Operation Conditions
Composition CC-125A PC unit, JE-921A electrode junction box, LS-703A flash assembly,
LS-120A photic stimulator control unit, KC-125A cart, SM-120A isolation unit
Temperature 10 to 35˚C (50 to 95˚F)
Humidity 30 to 80% (noncondensing)
Atmospheric pressure 700 to 1060 hPa
Electromagnetic Emissions
The EEG-1250’s essential performances in EMC standard satisfy the following criteria.
5/12
Electromagnetic Immunity
The EEG-1250’s essential performances in EMC standard satisfy the following criteria.
This Model EEG-1250 is intended for use in the electromagnetic environment specified below.
The customer or the user of the EEG-1250 should assure that it is used in such an environment.
Electromagnetic environment -
Immunity test IEC 60601 test level Compliance level
guidance
Electrostatic discharge ±6 kV contact ±6 kV contact Floors should be wood, concrete or
(ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/ ±2 kV for power supply ±2 kV for power supply Mains power quality should be that
burst lines lines of a typical commercial or hospital
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output environment.
lines lines
Surge ±1 kV differential mode ±1 kV differential mode Mains power quality should be that
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode of a typical commercial or hospital
environment.
Voltage dips, short <5% UT (>95% dip in UT) <5% UT (>95% dip in UT) Mains power quality should be that
interruptions and for 0.5 cycles for 0.5 cycles of a typical commercial or hospital
voltage variations on environment.
power supply input lines 40% UT (60% dip in UT) 40% UT (60% dip in UT) If the user of the EEG-1250 requires
IEC 61000-4-11 for 5 cycles for 1 cycle continued operation during power
mains interruptions, it is recommended
70% UT (30% dip in UT) 70% UT (30% dip in UT) that the EEG-1250 be powered from an
for 25 cycles for 25 cycles uninterruptible power supply.*
6/12
IEC 60601 test
Immunity test Compliance level Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the EEG-1250,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
*1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EEG-1250 is used exceeds the applicable RF
compliance level above, the EEG-1250 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the EEG-1250.
*2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than for all other functions.
7/12
When the JE-921A electrode junction box is connected to the EEG-1250 electroencephalograph
NOTE 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE 2: It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to
assure that they meet the minimum specification.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength outside the shielded location in which the JE-921A that is connected to
the EEG-1250 is used exceeds 3 V/m, observe the JE-921A that is connected to the EEG-1250 to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as relocating the JE-921A
that is connected to the EEG-1250 or using the shielded location with a higher RF shielding effectiveness and filter
attenuation.
NOTE
The JE-921A that is connected to the EEG-1250 should be used in a shielded location with an RF filter
attenuation of 31 dB over the frequency range 80 MHz to 2.5 GHz. If the JE-921A that is connected to the
EEG‑1250 is used outside shielded location, use it in a location where the measured field strength is less than
0.3 V/m. To measure shielding effectiveness, consult a service engineer.
8/12
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
The EEG-1250 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the EEG-1250 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the EEG-1250 as recommended below,
according to the maximum output power of the communications.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
9/12
Standard Accessories
10/12
Options and Consumables
Options
When ordering the following options, specify the model. When ordering
consumables, specify the supply code. When the supply code is not provided
with the option or consumable, specify the model number or code number.
Photic Stimulation
• Photic stimulator control unit, LS-120A
• Flash lamp assembly, LS-703A, LS-706A
• Visual Stimulator Adapter, YL-103A, YL-104A
Software
• Digital video software, QP-110AK
• Review program, QP-112AK
• Trend program, QP-160AK
Digital Video
• Camera capture unit, QI-120A
Cart/Hardware/Electrical
• Stand, KC-001A (for the electrode junction box and flash lamp assembly)
• Holder, DI-120A
• Cart, KC-125A
• Holder, DI-121A
• Isolation unit, SM-120A
11/12
Consumables
Electrodes and Paste
Model Supply Code Description
Z-401CE F510 Elefix EEG paste 400 g, 3 pcs/set
NE-113A H503A EEG disk electrode, 26 pcs/set, 1.5 m
NE-114A H503B EEG disk electrode, 26 pcs/set, 0.8 m
NE-134A H526 Collodion EEG disk electrode,
12 pcs/set
NE-311A H540A EEG ear clip electrode, 2 pcs/set
BE-911A – EEG disk electrode (shielded type), 24 electrode
leads
BE-912A – EEG disk electrode (shielded type), 12 electrode
leads
– H068 ECG clip-on electrode
BC-112B K512A ECG electrode lead, 2 pcs/set
SpO2 Measurement
Model Supply Code Description
JL-550T1 – SpO2 adapter*
JL-550T2 Y095A SpO2 adapter*
TL-201T P225F Finger probe BluPRO
TL-220T P225G Multi-site probe BluPRO
TL-260T P205A Multi-site Y probe
TL-271T P203A SpO2 probe (for adult, disposable)
TL-272T P203B SpO2 probe (for child, disposable)
TL-273T P203C SpO2 probe (for neonate/adult, disposable)
TL-274T P204D SpO2 probe (for adult/infant)
TL-051S P228A SpO2 probe (for adult and neonate, 0.8 m)
TL-052S P228B SpO2 probe (for adult and neonate, 1.6 m)
TL-061S P229A SpO2 probe (for adult, child and infant, 0.8 m)
TL-062S P229B SpO2 probe (for adult, child and infant, 1.6 m)
* For differences between JL-550T1 and JL-550T2, contact your Nihon Kohden
representative
CO2 Measurement
Model Supply Code Description
TG-921T3 P908 CO2 sensor kit
YG-120T V921 Nasal Adapter (for nasal breathing)
YG-121T V922 Nasal Adapter (for naso-oral breathing)
12/12