You are on page 1of 12

Technical Data

EEG-1250
Electroencephalograph

TD.EEG-1250_A

This technical data may be revised or replaced by Nihon Kohden at any time without notice.
Some products may not be available in your country.
Specifications

Data Acquisition
JE-921A electrode junction box
Number of input jacks
EEG input jacks 25
Extra inputs 4
Bipolar input jacks (pairs) 7
The extra jacks (X1-X4) can be used as bipolar jacks. To select extra jack or
bipolar jack, refer to Section 5 “System Programs”.
DC input jacks 4
SpO2 connector 1
CO2 connector 1
Input impedance
EEG input/extra input 100 MΩ
DC input 1.5 MΩ
Input circuit current 5 nA or less
Internal noise level
EEG input/extra input 1.5 µVp-p or less (0.53 to 60 Hz)
DC input 10 mVp-p or less
CMRR
EEG input/extra input 105 dB or greater (60 Hz)
Bipolar input 100 dB or greater (60 Hz)
Low-cut filter 0.08 Hz (Time constant: 2 s)
High-cut filter 300 Hz (−18 dB/oct)
Offset tolerance ±600 mV
A/D conversion 16 bits (97 nV/LSB)
Sampling and hold All electrodes at the same time
Sampling frequency 100, 200, 500, 1000 Hz, selectable

Data Processing
Sensitivity
EEG INPUT OFF, 1, 2, 3 (2.5), 5, 7, 10, 15, 20, 30, 50, 75, 100, 150, 200 μV/mm
DC INPUT OFF, 10, 15, 20, 30, 50, 75, 100, 150, 200 mV/mm
Time constant (Low-cut filter) 0.001, 0.003, 0.03, 0.1, 0.3, 0.6, 1.0, 2.0, 5.0, 10.0 s
0.016, 0.03, 0.08, 0.16, 0.27, 0.53, 1.6, 5.3, 53, 159 Hz (−6 dB/oct)
High-cut filter 15, 30, 35, 60, 70, 120 (−12 dB/oct), 50 (RAPID), 300 Hz (−18 dB/oct)
AC filter 50 or 60 Hz, (rejection ratio: 1/25 or more)
Calibration waveform
Waveform shape 0.25 Hz step wave or 10 Hz sine wave
Voltage 2, 5, 10, 20, 50, 100, 200, 500, 1000 μV (×1000 for DC input signal)
Impedance check
Indication on the screen All electrodes are displayed on the screen in electrode position layout.
Impedance for each electrode is displayed and electrodes with impedance higher
than the preset impedance threshold are highlighted.
Indication on LED LEDs on the electrode junction box with impedance higher than the preset
impedance threshold light.
Impedance threshold 2, 5, 10, 20 and 50 kΩ

2/12
Pattern 36 sets of programmable montages combined with programmable individual
amplifier settings
Reference electrode selector A1 → A2, A1 ← A2, A1 ↔ A2, A1 + A2, VX, AV (with unsuitable electrode
deletion function), Aav, Org, SD, BN and OFF

Display
Display size 23.8 inches
Display type Touch panel
Display resolution Up to 1920 dots × 1080 lines
Number of display channels Up to 64 and one mark channel
Waveform display color 16 colors
Waveform position adjustment Provided
Waveform sweep speed 0.1, 0.2, 0.5, 1, 2, 5, 10, 15, 20, 30, 60 s or 2, 3, 5 min/page
Timing mark 0.1, 1 s
Time scale off, 0.2, 1 s
Event mark Displays at the bottom of the screen
EEG scale Provided

Photic Stimulator
Optional LS-120A photic stimulator control unit
Maximum flash energy 1.28 J/Single
Stimulation modes 3 automatic (30 steps, programmable), manual, and single
Mode of operation Continuous operation with intermittent loading
Duty cycle Max. 5 minutes continuous operation in 30 minutes

Automatic stimulation
Stimulus rate 0.5, 1 to 33 (1 Hz steps), 50 and 60 Hz
Stimulation period 1 to 99 seconds in 1 second steps
Pause period 1 to 30 seconds in 1 second steps

Manual stimulation Manually set frequency and stimulation period


Photic frequency 0.5 Hz, 1 to 33 Hz in 1 Hz steps, 50 and 60 Hz
Stimulation time 1 to 99 s in 1 second steps and continuous stimulation (FREE: Max. 5 min)

Pulse mode Normal, random, and double


Random stimulation 1 to 33 Hz in 1 Hz steps within ± 50%
Single stimulation Manual key operation single stimulation or automatic single stimulation by
external trigger signal
Trigger input TRIGGER IN connector (1 to 5 V)
Trigger output TRIGGER OUT connector (3 V or more)

Safety
Safety standard:
EEG-1250 (Op No. 50A) electroencephalograph
IEC 60601-1: 1988
IEC 60601-1 Amendment 1: 1991
IEC 60601-1 Amendment 2: 1995
IEC 60601-1-1: 2000
IEC 60601-1-2: 2001
IEC 60601-1-2 Amendment 1: 2004

3/12
IEC 60601-2-26: 2002
IEC 60601-2-40: 1998*
* Applied only for bipolar inputs (X1 to X4) on the electrode junction box when
acquiring EMG.

EEG-1250 (Op No. 50K) electroencephalograph


IEC 60601-1: 2005 + Amendment 1: 2012
IEC 60601-2-26: 2012
IEC 60601-1-6: 2010 + Amendment 1: 2013
IEC 662366: 2007 + Amendment 1: 2013
Type of protection against electric shock: Class I
Degree of protection against electric shock:
When using the JE-921A electrode junction box (and JE-922A mini junction
box)
EEG input: Type BF applied part
Bipolar input: Type BF applied part
SpO2 input: Type BF applied part
(only when using the SpO2 adapter and probe)
CO2 input: Type BF applied part
(only when using the CO2 sensor kit)
Degree of protection against harmful ingress of water:
Non-protected (IPX0)
Degree of safety of application in the presence of a flammable anaesthetic mixture with air or oxygen or nitrous
oxide: Not suitable for use in the presence of a flammable anaesthetic mixture with air
or oxygen or nitrous oxide
Mode of operation: Intermittent: LS-703A/706A (5 minutes of operation, more than 20 minutes of
interval)
Continuous: Other instruments
Mechanical strength: Indoor stationary equipment

Electromagnetic Compatibility
EEG-1250 (Op No. 50A) electroencephalograph
IEC 60601-1-2: 2001
IEC 60601-1-2 + Amendment 1: 2004
CISPR11 GROUP 1 CLASS A: 2003 Amendment 1: 2004, Amendment 2: 2006

EEG-1250 (Op No. 50K) electroencephalograph


IEC 60601-1-2 : 2007
IEC 60601-2-26: 2012
IEC 61000-3-2: 2014
CISPR11: 2009 + A1: 2010 GROUP 1 CLASS A

Dimensions and Weight


CC-125A PC unit 575 (W) × 393 (H) × 63 (D) mm, 11 kg
JE-921A electrode junction box 185 (W) × 72 (D) × 167 (H) mm, 1 kg
SM-120A isolation unit 150 (W) × 140 (H) × 320 (D) mm, 12 kg

Power Requirements
Line voltage AC 100 to 240 V ±10%
Line frequency 50/60 Hz
Power consumption 400 VA or less

4/12
Operation Conditions
Composition CC-125A PC unit, JE-921A electrode junction box, LS-703A flash assembly,
LS-120A photic stimulator control unit, KC-125A cart, SM-120A isolation unit
Temperature 10 to 35˚C (50 to 95˚F)
Humidity 30 to 80% (noncondensing)
Atmospheric pressure 700 to 1060 hPa

Transport and Storage Conditions


Composition CC-125A PC unit, JE-921A electrode junction box, LS-703A flash assembly,
LS-120A photic stimulator control unit, KC-125A cart, SM-120A isolation unit
Temperature −20 to +60˚C (−4 to +140˚F)
(For LS-703A flash lamp assembly, JE-921A electrode junction box, SM-120A
isolation unit: −20 to +65˚C (−4 to +149˚F))
Humidity 10 to 80%
(For LS-703A flash lamp assembly, JE-921A electrode junction box, SM-120A
isolation unit: 10 to 95%)
Atmospheric pressure 700 to 1060 hPa

Electromagnetic Emissions
The EEG-1250’s essential performances in EMC standard satisfy the following criteria.

EEG-1250 is intended for use in the electromagnetic environment specified below.


The customer or the user of the EEG-1250 should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment guidance


RF emissions Group 1 The EEG-1250 uses RF energy only for its internal function. Therefore, its
CISPR 11 RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions Class A The EEG-1250 is suitable for use in all establishments, excluding domestic
CISPR 11 establishments and those directly connected to the public low-voltage
Harmonic emissions Class A power supply network that supplies buildings used for domestic purposes.
IEC 61000-3-2
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3

5/12
Electromagnetic Immunity
The EEG-1250’s essential performances in EMC standard satisfy the following criteria.

This Model EEG-1250 is intended for use in the electromagnetic environment specified below.
The customer or the user of the EEG-1250 should assure that it is used in such an environment.

Electromagnetic environment -
Immunity test IEC 60601 test level Compliance level
guidance
Electrostatic discharge ±6 kV contact ±6 kV contact Floors should be wood, concrete or
(ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/ ±2 kV for power supply ±2 kV for power supply Mains power quality should be that
burst lines lines of a typical commercial or hospital
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output environment.
lines lines
Surge ±1 kV differential mode ±1 kV differential mode Mains power quality should be that
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode of a typical commercial or hospital
environment.
Voltage dips, short <5% UT (>95% dip in UT) <5% UT (>95% dip in UT) Mains power quality should be that
interruptions and for 0.5 cycles for 0.5 cycles of a typical commercial or hospital
voltage variations on environment.
power supply input lines 40% UT (60% dip in UT) 40% UT (60% dip in UT) If the user of the EEG-1250 requires
IEC 61000-4-11 for 5 cycles for 1 cycle continued operation during power
mains interruptions, it is recommended
70% UT (30% dip in UT) 70% UT (30% dip in UT) that the EEG-1250 be powered from an
for 25 cycles for 25 cycles uninterruptible power supply.*

<5% UT (>95% dip in UT) <5% UT (>95% dip in UT)


for 5 s for 5 s
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical commercial
IEC 61000-4-8 or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
* It is recommended that the EEG-1250 be powered from an uninterruptible power supply if the power quality is not
stable.

6/12
IEC 60601 test
Immunity test Compliance level Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the EEG-1250,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.

Recommended separation distance

Conducted RF 3 Vrms 3 Vrms d = 1.2 P


IEC 61000-4-6 150 kHz to 80 MHz

Radiated RF 3 V/m 3 V/m d = 1.2 P 80 MHz to 800 MHz


IEC 61000-4-3 80 MHz to 2.5 GHz d = 2.3 P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the


transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).

Field strengths from fixed RF transmitters, as


determined by an electromagnetic site survey*1, should
be less than the compliance level in each frequency
range*2.

Interference may occur in the vicinity of equipment


marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
*1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EEG-1250 is used exceeds the applicable RF
compliance level above, the EEG-1250 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the EEG-1250.
*2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than for all other functions.

Avoiding Electromagnetic Interference (Impedance Respiration)


Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological
limitations do not allow immunity levels higher than 1 Vrms for conducted disturbances induced by RF fields.
Electromagnetic fields with conducted disturbances above 1 Vrms may cause measurement error. Do not use electrically
radiating equipment near the impedance respiration measurements.

7/12
When the JE-921A electrode junction box is connected to the EEG-1250 electroencephalograph

IEC 60601 test


Immunity test Compliance level Electromagnetic environment – guidance
level
The JE-921A that is connected to the EEG-1250 must
be used only in a shielded location with a minimum RF
shielding effectiveness and, for each cable that enters
the shielded location, a minimum RF filter attenuation
Radiated RF 3 V/m 0.3 V/m of 31 dB over the frequency range 80 MHz to 2.5 GHz.
IEC 61000-4-3 80 MHz to 2.5 GHz Refer to the note below this table.

Field strengths outside the shielded location from fixed


RF transmitters, as determined by an electro-magnetic
site survey*, should be less than 3 V/m. Refer to the
note below this table.

Interference may occur in the vicinity of equipment


marked with the following symbol:

NOTE 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE 2: It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to
assure that they meet the minimum specification.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength outside the shielded location in which the JE-921A that is connected to
the EEG-1250 is used exceeds 3 V/m, observe the JE-921A that is connected to the EEG-1250 to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as relocating the JE-921A
that is connected to the EEG-1250 or using the shielded location with a higher RF shielding effectiveness and filter
attenuation.

NOTE
The JE-921A that is connected to the EEG-1250 should be used in a shielded location with an RF filter
attenuation of 31 dB over the frequency range 80 MHz to 2.5 GHz. If the JE-921A that is connected to the
EEG‑1250 is used outside shielded location, use it in a location where the measured field strength is less than
0.3 V/m. To measure shielding effectiveness, consult a service engineer.

8/12
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
The EEG-1250 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the EEG-1250 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the EEG-1250 as recommended below,
according to the maximum output power of the communications.

Rated maximum output Separation distance according to frequency of transmitter (m)


power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(W) d = 1.2 P d = 1.2 P d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

System Composition for EMC Test


The EEG-1250 electroencephalograph is tested to comply with IEC 60601-1-2: 2001 and Amendment 1: 2004, and
IEC 60601-1-2: 2007 with the following composition. If any part which is not specified by Nihon Kohden is used, the EMC
specifications may not comply.

Units Cable length


PC unit CC-125A —
Electrode junction box JE-921A 5.0 m
Photic stimulator control unit LS-120AJ/AK 2.0 m
Flash lamp assembly LS-703A 5.0 m
Mini electrode junction box JE-922A 5.0 m
SpO2 adapter JL-550T 2.5 m
SpO2 probe TL-201T 1.6 m
CO2 sensor kit TG-921T3 3.5 m
Camera capture unit QI-120A 0.8 m

9/12
Standard Accessories

For the EEG-1250 (Op No. 50A)


Description, Model No. Q’ty Supply Code
Power cord UL 1 –
Earth lead 1 L912
Maker cord 1 –
Collodion EEG disk electrode, NE-133A (24 pcs/set)* 1 –
System program 1 –
Mouse 1 –
Keyboard 1 –
Electrode junction box holder 1 –
Cable clamp 1 –
One-wrap strap 10 –
Cable binder 2 –
Ferrite core 1 –
Operator’s manual (this manual) 1 –
Service manual 1 –
* The NE-133A Collodion EEG disk electrode is only available as an accessory of the
EEG-1250 (Op. No. 50A). To order these electrodes separately, order the NE-134A EEG
disk electrode (12 pcs/set, supply code: H526). The NE-134A is the same electrode as the
NE-133A; only the quantity is different.

For the EEG-1250 (Op No. 50K)


Description Q’ty Supply Code
Power cord UL* *1 2
1 –
Power cord N*2 1 –
Maker cord 1 –
System program 1 –
Mouse 1 –
Keyboard 1 –
Electrode junction box holder 1 –
Cable clamp 1 –
One-wrap strap 10 –
Cable binder 2 –
Ferrite core 1 –
Earth lead 1 L912
Operator’s manual (this manual) 1 –
Service manual 1 –
*1 Not provided in Europe.
*2 Use one of the power cords for your country or region.

10/12
Options and Consumables

Before using the options or consumables, read the accompanying document of


the product.

Options
When ordering the following options, specify the model. When ordering
consumables, specify the supply code. When the supply code is not provided
with the option or consumable, specify the model number or code number.

Mini Electrode Junction Box


• Mini electrode junction box, JE-922A

Photic Stimulation
• Photic stimulator control unit, LS-120A
• Flash lamp assembly, LS-703A, LS-706A
• Visual Stimulator Adapter, YL-103A, YL-104A

Software
• Digital video software, QP-110AK
• Review program, QP-112AK
• Trend program, QP-160AK

Digital Video
• Camera capture unit, QI-120A

Cart/Hardware/Electrical
• Stand, KC-001A (for the electrode junction box and flash lamp assembly)
• Holder, DI-120A
• Cart, KC-125A
• Holder, DI-121A
• Isolation unit, SM-120A

11/12
Consumables
Electrodes and Paste
Model Supply Code Description
Z-401CE F510 Elefix EEG paste 400 g, 3 pcs/set
NE-113A H503A EEG disk electrode, 26 pcs/set, 1.5 m
NE-114A H503B EEG disk electrode, 26 pcs/set, 0.8 m
NE-134A H526 Collodion EEG disk electrode,
12 pcs/set
NE-311A H540A EEG ear clip electrode, 2 pcs/set
BE-911A – EEG disk electrode (shielded type), 24 electrode
leads
BE-912A – EEG disk electrode (shielded type), 12 electrode
leads
– H068 ECG clip-on electrode
BC-112B K512A ECG electrode lead, 2 pcs/set

SpO2 Measurement
Model Supply Code Description
JL-550T1 – SpO2 adapter*
JL-550T2 Y095A SpO2 adapter*
TL-201T P225F Finger probe BluPRO
TL-220T P225G Multi-site probe BluPRO
TL-260T P205A Multi-site Y probe
TL-271T P203A SpO2 probe (for adult, disposable)
TL-272T P203B SpO2 probe (for child, disposable)
TL-273T P203C SpO2 probe (for neonate/adult, disposable)
TL-274T P204D SpO2 probe (for adult/infant)
TL-051S P228A SpO2 probe (for adult and neonate, 0.8 m)
TL-052S P228B SpO2 probe (for adult and neonate, 1.6 m)
TL-061S P229A SpO2 probe (for adult, child and infant, 0.8 m)
TL-062S P229B SpO2 probe (for adult, child and infant, 1.6 m)

* For differences between JL-550T1 and JL-550T2, contact your Nihon Kohden
representative

CO2 Measurement
Model Supply Code Description
TG-921T3 P908 CO2 sensor kit
YG-120T V921 Nasal Adapter (for nasal breathing)
YG-121T V922 Nasal Adapter (for naso-oral breathing)

12/12

You might also like