CEP Services Pte Ltd Certificate No.

: LA-2017-0641-C
30B Hillview Terrace
Singapore 669248 Issue No. : 5

Date : 30 March 2020

Page : 1 of 3

Field of Testing : Calibration and Measurement

MEASURED QUANTITIES/ CALIBRATION AND


INSTRUMENTS CALIBRATED/ METHOD MEASUREMENT CAPABILITY
RANGE OF MEASUREMENT (CMC *)

A. Dimensional
1A Angle Measurement / CEP In-House developed method,
Calibration of Inclinometer CM-001 (Ver 5, dated 02/03/20),
Probe and Portable Automated Vertical Rotary Table.
Tiltmeter
0 ° to 30 ° 25 "

1B Angle Measurement / CEP In-House developed method,


Calibration of Angular CM-001A (Ver 1, dated 02/03/20),
Measuring Device, Digital Automated Vertical Rotary Table.
Level and Tilt sensor
-20 ° to 20 ° 0.006 °

2 DC Volt Measurement / CEP In-House developed method,


Calibration of Inclinometer CM-002 (Ver 2, dated 01/11/17),
and Tiltmeter readout unit DC Calibrator.
-5 V to 5 V 0.4 mV

B. Force Measurement
Calibration of Force CEP CM-003 (Ver 7, dated
Transducers and Load Cells 20/08/19) with reference to ISO
(Compression only) 376: 2011 E.

200 kN to 2000 kN 2 MN Force Calibration Machine. (1.10 % to 1.00 %) of reading


500 kN to 10000 kN 10 MN Force Calibration Machine. (0.74 % to 0.36 %) of reading

The SAC Programme is managed by Enterprise Singapore


230 Victoria Street, #09-00 Bugis Junction Office Tower, Singapore 188024 Tel: +65 6278 6666 Fax: +65 6659 0640
www.sac-accreditation.gov.sg
Certificate No. : LA-2017-0641-C Issue No. : 5

Date : 30 March 2020 Page : 2 of 3

MEASURED QUANTITIES/ CALIBRATION AND


INSTRUMENTS CALIBRATED/ METHOD MEASUREMENT CAPABILITY
(CMC *)
RANGE OF MEASUREMENT

C. Vibration Measurement
Calibration of Vibration CEP CM-004 (Ver 3, dated
Monitoring Device 20/08/19) with reference to ISO
16063-21:2003 (E), Vibration
a) Radial (Longitudinal) Calibration System.
b) Transverse
c) Vertical

25.4 mm/s @ 2 Hz 2.5 %


25.4 mm/s @ 3 Hz 2.5 %
25.4 mm/s @ 4 Hz 2.0 %
25.4 mm/s @ 10 Hz 2.0 %
25.4 mm/s @ 30 Hz 2.0 %
25.4 mm/s @ 60 Hz 2.0 %
25.4 mm/s @ 125 Hz 2.5 %
25.4 mm/s @ 250 Hz 2.5 %

D. Frequency Measurement
Calibration of Vibrating Wire CEP CM-005 (Ver 2, dated
Readout Unit 15/08/19), Function Generator.

1000 Hz to 6000 Hz @ 1 Vpp 0.1 Hz


to 5 Vpp

The SAC Programme is managed by Enterprise Singapore


230 Victoria Street, #09-00 Bugis Junction Office Tower, Singapore 188024 Tel: +65 6278 6666 Fax: +65 6659 0640
www.sac-accreditation.gov.sg
Certificate No. : LA-2017-0641-C Issue No. : 5

Date : 30 March 2020 Page : 3 of 3

MEASURED QUANTITIES/ CALIBRATION AND


INSTRUMENTS CALIBRATED/ METHOD MEASUREMENT
RANGE OF MEASUREMENT CAPABILITY (CMC *)

E. Pressure Measurement
Calibration of Pressure CEP CM-006 (Ver 3, dated
Measuring Device 04/10/19), Pressure Controller

1. Analogue Gauge Pressure


Indicator
0 psi to 100 psi (Gauge)
0.08 % F.S.
2. Digital Gauge Pressure
Indicator

0 kPa to 700 kPa (Gauge) 0.05 % F.S.


0 psi to 100 psi (Gauge) 0.05 % F.S.
100 psi to 145 psi (Gauge) 0.07 % F.S.

3. Vibrating Wire Piezometer


0 kPa to 700 kPa (Gauge) 0.10 % F.S.

* CMC is expressed as an expanded uncertainty estimated at a level of confidence of approximately 95%.

Approved signatories:
Mr Darrell Teo )
Mr Terence Teo ) All items in the schedule
Mr Lim Wee Kiat )

Note :

This laboratory is accredited in accordance with the recognised International Standard ISO/IEC 17025.
A laboratory's fulfilment of the requirements of ISO/IEC 17025 means the laboratory meets both the
technical competence requirements and management system requirements that are necessary for it
to consistently deliver technically valid calibrations. The management system requirements in ISO/IEC
17025 are written in language relevant to laboratory operations and operate generally in accordance with
the principles of ISO 9001.

The SAC Programme is managed by Enterprise Singapore


230 Victoria Street, #09-00 Bugis Junction Office Tower, Singapore 188024 Tel: +65 6278 6666 Fax: +65 6659 0640
www.sac-accreditation.gov.sg

You might also like