You are on page 1of 9

Journal of Medical Imaging and Radiation Oncology  (2019) –

MEDICAL IMAGING—REVIEW ARTICLE

Radiation dose and risk to the lens of the eye during CT


examinations of the brain

Journal of Medical Imaging and Radiation Oncology


Rebekah Poon1 and Mohamed K Badawy1,2
1 Department of Medical Imaging and Radiation Sciences, School of Primary and Allied Health Care, Faculty of Medicine, Nursing and Health
Sciences, Monash University, Clayton, Victoria, Australia
2 Monash Imaging, Monash Health, Clayton, Victoria, Australia

R Poon BRadMedImag; MK Badawy PhD, Summary


MAppSci, BSc.
Cataracts are the leading cause of blindness and visual disability worldwide.
Correspondence Of the known contributing factors to this condition, ionising radiation is con-
Dr Mohamed K Badawy, Monash Imaging, sidered the primary concern in a radiological context given the particular
Monash Health, 246 Clayton Road, Clayton, radiosensitivity of the lens of the eye. In light of the substantially increased
Vic. 3168, Australia. application of computed tomography in brain imaging, an investigation of the
Email: mohamed.badawy@monashhealth.org relevent literature is warranted to assess thresholds, lens radiation doses and
dose reduction techniques in respect to the cataractogenic risk of such exami-
Conflict of interest: The authors declare no nations. The value and very existence of a lens dose threshold is debatable
conflict of interest or financial funding. given different considerations of radiation dose, latency, opacity classifications
and historical sample populations, though ICRP guidelines suggest a threshold
Submitted 14 April 2019; accepted 19 August of 0.5 Gy. Documented CT-specific radiation doses to the eye following scans
2019. of the brain are highly variable between studies (2–130 mGy), primarily owing
to discrepancies in scanning technique. These findings, when coupled with the
doi:10.1111/1754-9485.12950 relative ambiguity of known threshold values, present difficulties in assessing
the overall risk of cataracts following serial CT examinations to the head. In
the absence of definitive risk evaluations, a cautionary approach is advised.
The implementation of gantry tilt along the supraorbital margin is recom-
mended as standard practice on account of its highly effective radiation dose
reduction outcomes. Organ-based tube modulation and reductions in tube
current may also be considered beneficial. Bismuth eye shielding is only
advised where gantry tilting is unachievable, and in such cases, ensure careful
adherence to appropriate shield placement and infection control measures.

Key words: computed tomography; lens dose; radiation cataracts; radiation


dose optimisation.

the exponentially increasing demand for CT scans, which


Introduction contributes to approximately one half of the overall effec-
Rapid technological advances in computed tomography tive radiation dose from medical procedures per capita in
(CT) within recent years have seen its contribution to the United States.2
diagnostic imaging services increase significantly, with While diagnostically valuable and relatively quick to per-
over 3.2 million examinations reported in Australia in form, CT scans of the head do not come without risks. In
2017.1 Of these, almost 500,000 cases involved CT scans conjunction with the consequences associated with direct
of the brain, 5.7% of which were performed on paediatrics irradiation of the brain, the formation of visually impairing
and young people ages 0–24 years old. Despite continu- opacities and radiation cataracts in the lens of the eye
ous endeavours in technological advances and awareness may be of concern.3 When paired with a lack of standard-
in regards to radiation safety, it has been purported that ised scanning techniques and the growing tendency for
the average medically induced radiation dose per capita is brain scans to be over-referred,4,5 the potential for radia-
increasing, particularly in well-developed countries. For tion cataracts to become a more widespread phenomenon
the most part, this upward trend can likely be attributed to becomes tangible. This paper aims to assess the evidence

© 2019 The Royal Australian and New Zealand College of Radiologists 1


R Poon and MK Badawy

regarding radiation dose thresholds to the lens of the eye, varying results. These reports have predominantly
the estimated radiation dose to the lens of the eye follow- looked at cases of atomic bomb exposure, radiotherapy
ing CT scans of the brain, and the most effective strategies patients and occupationally exposed medical profession-
in reducing these radiation doses. By assessing the most als, as well as Chernobyl survivors, building contamina-
relevant and up to date literature, the authors seek to tions, diagnostic imaging examinations and even
develop an understanding as to the current risk status of astronauts.7,11 Overall, the consensus suggests that the
radiation-induced cataracts, and how any risks may be formation of radiation-induced cataracts is deterministic
minimised in future practice. (a tissue effect), with a threshold response which esca-
lates in severity with increased dose and reduced
latency.6,8,11 In response to the data presented in the
Radiation cataracts studies above, the International Commission on Radio-
The lens of the eye is considered one of the most logical Protection (ICRP) introduced threshold values
radiosensitive tissues in the body, with the primary con- under which it was assumed that cataracts are not a
cern being the development of cataracts and lens opaci- common risk factor. This threshold was initially intro-
ties following radiation exposure.6 The term ‘cataract’ duced as a lifetime equivalent dose of 15 Sievert (Sv) in
refers to the clouding of the lens of the eye, and it is the 1977, however, was lowered to 5 Sv for acute exposure
leading cause of blindness and visual disability world- (and 8 Sv for a protracted or fractionated exposure) in
wide.7 It is proposed that the condition arises from errors 1984, and eventually to an absorbed dose of 0.5 Gray
in cell division following damage to the epithelial layer of (Gy) in 2012.12 This reduction was made following the
the lens. Cataracts are categorised as either cortical, development of newer research studies which were sug-
nuclear or posterior subcapsular (PSC) opacities depend- gestive of lower threshold limits. These studies noted
ing on their anatomical position.8 Contributing risk fac- that evidence supporting the previous threshold had a
tors are age, radiation exposure, genetics, the use of scarcity of data for low-dose investigations, significant
systemic steroids and physiological conditions such as variations in the type of radiation exposure, short follow-
diabetes.6,7 Radiation-induced cataracts are most com- up periods which were inconsiderate of latent effects,
monly associated with PSC opacities. However, it is and potentially inept methods of detecting early lens
important to bear in mind that once a cataract has developments.6,8
formed, it is difficult to definitively identify the opacity as It is worth noting, however, that the value of this
radiation-induced. It should also be noted that associa- threshold is still a controversial and dynamic affair.
tions have been made between radiation exposure and Suggestions have been made by multiple parties that the
cortical lens changes, albeit to a markedly lesser threshold is non-existent and that cataractogenic effects
extent.9 may be purely stochastic. Research conducted by Neri-
As with several other radiation-induced effects, the ishi et al.13 is a fundamental component in such an argu-
latency of cataractogenic effects has an inverse relation- ment, where the incidence of cataract surgeries was
ship with dose, as well as an affiliated age modulating investigated among atomic bomb survivors approxi-
disposition.8 Once the progression of cataracts reaches a mately 50 years after the event. The best estimate
stage of ‘visual impairment’, the only means of treatment threshold of 0.1 Gy was reported, with a zero value
is via surgical removal of the damaged lens and replace- threshold included within 95% confidence intervals (CI).
ment with an artificial substitute. While this surgery is This confirmed and extended conclusions made by Naka-
considered non-invasive and has a relatively high suc- shima et al.14 the year prior, who also speculated a lack
cess rate, the economic and societal burden of this pro- of threshold within 95% CI. More recently, it has even
cedure is predicted to worsen accordingly in the context been proposed that these thresholds may have been
of booming ageing populations.7,8 Furthermore, the conservative given that some participants were relatively
operation carries risks of infection, suprachoroidal haem- young at the time of re-evaluation and therefore may
orrhaging, retinal detachment and unsuccessful overall not yet have developed operable cataracts. Some sub-
visual restoration.10 As a consequence of the anatomical jects were also reported to have forgone surgery in light
location of radiation-induced PSC opacities, it has been of age or medical complications, which may have further
suggested that the potential exists for significant visual distorted the actual incidence of cataract development.8
impairment to occur despite minor cellular changes,6 and In a later study involving low-dose exposures to radio-
thus, an investigation into the origins of radiation catar- logical technologists in the United States, Chodick
acts is warranted. et al.15 also reported a correlation between dose and cat-
aract formation with no discernible threshold, which fur-
ther supports the notion that cataractogenic effects may
Radiation threshold for cataractogenic be more stochastic than first assumed.
effects The 2007 study performed by Neriishi et al.13 also
Extensive bodies of work have been collated over the addressed one of the most significant issues associated
years analysing the effects of radiation on the eye with with radiation cataract studies; the point at which lens

2 © 2019 The Royal Australian and New Zealand College of Radiologists


CT Brains and the risk of cataracts

opacification results in clinically relevant visual disability. Nonetheless, across these three studies, a repeated
While affiliations had been made between radiation verdict was reached that the cataractogenic risk of CT
exposure and cataract formation in the past, the point at scans was likely increased with multiple examinations.
which these opacifications become visually impairing to This concept was supported by a retrospective analysis
the point of requiring actionable outcomes had yet to be conducted by Yuan et al.21 via the use of a database
quantifiably studied. Given the aforementioned depen- published by the National Health Research Institutes in
dency of dose versus latency, early stages of opacifica- Miaoli Taiwan. It is worth noting that this study was also
tion may not initially result in visual impairment at low inclusive of scans to the neck as well as the head.
doses and therefore could be considered inconsequential. Despite having a more accurate record of the number of
The point at which surgery is warranted, however, is cir- scans performed, and therefore eliminating the unrelia-
cumstantial and subjective depending on the extent to bility associated with self-reporting, the radiation doses
which cataracts impact one’s quality of life, and this in administered to the patients remained unknown. The use
itself may potentially skew the analyses. of a national database did, however, allow for compar-
Conversely, arguments have also been made that the ison of the sample population with age and sex-matched
ICRP’s lowering of the dose threshold was not suffi- equivalents who had never been irradiated, and from
ciently evidenced in the context of chronic expo- these comparisons, it was suggested that the threshold
sures.7,16 These views have been put forward based on for cataractogenic effects might be as low as 0.25 Gy.
multiple considerations, such as the variability of self- For a more realistic indication as to the risk of cataracts
reporting studies and differences in the opacity classifi- following CT scans of the head, a thorough examination
cation system if one was even used. Age at exposure of the actual radiation dose to the lens of the eye during
also needs to be recognised as an essential component these scans needs to be conducted.
given that some studies were built on the follow-up of
individuals irradiated as children, a population which is
known to be more sensitive to cataractogenic effects
Radiation Dose to the Lens of the Eye
than adults.17 Furthermore, despite the dose/latency
During CT Scans of the Head
relationship revealing a higher incidence of cataracts Current literature regarding the specific dose to the lens
where follow-up studies extended their follow-up peri- of the eyes following CT scans of the brain is variable
ods, consideration must also be given to the naturally and somewhat complicated to navigate. This variability is
high occurring frequency of cataracts in the older gener- primarily due to inconsistencies in scanning technique,
ations. It is argued by Thome et al.7 that the conse- CT scanners, means of dose measurement and study
quences of a possibly unsubstantiated reduction in methodology. As per Table 1, recorded dose values range
threshold dose may result in economic afflictions. These from 2 to 130 mGy. Of the trials which utilised a degree
may be in the form of potential requirements for addi- of gantry tilt as a component of their reference scanning
tional shielding, dosimetry devices and other radiation technique, doses were significantly lower overall, mainly
safety equipment, as well as radiation safety training where alignment could be achieved along the supraor-
for occupationally exposed workers. bital margin such that the lenses of the eyes were no
Very little data specific to CT and diagnostic imaging is longer within the primary irradiation beam. However, of
known to contribute to such threshold discussions these studies, the majority of those who were able to
and that which does exist relies heavily on self-reporting achieve adequate alignment were phantom trials.3,22
and radiation doses which had not been directly deter- While the use of anthropomorphic phantoms in such
mined. As reported in the Beaver Dam Eye Study Cohort studies allows precise documentation of absorbed dose,
in 1993, an odds ratio (OR) of 1.45 for PSC opacities and variations in patient size and position are not able to be
1.28 for nuclear opacities was calculated in relation to taken into account. Furthermore, errors in patient posi-
head CT scans; however, past medical imaging history tioning, inappropriate gantry tilt and movement before
was self-reported, and thus, specific dose values were or during the scan are all reflective of clinical realities
unable to be resolved.18 Subsequent re-evaluation of this and may influence the degree to which the lens of the
same sample population with at least a five-year latency eye is included in the primary scanning range.23 In vivo
demonstrated similar figures; however, when CT scans dosimetry and clinical trials are therefore the only way of
were compared to all other forms of head X-ray imaging, determining the actual dose to the patient. However,
the results proved statistically insignificant.19 In a study dose comparisons between these studies are multivari-
conducted by Hourihan et al.20 in Australia, no clear cor- ate and require considerations of individual scanning
relation could be made between a history of single head parameters, different patient positioning and resulting
CT scans and the incidence of cataracts. However, this image quality.
study had a relatively broad spectrum of latency between Studies which indicated no gantry tilt or unsuccessfully
individuals at the time of the survey (1–18 years), and attempted gantry tilt23–25 were generally nondescript in
again this data was reliant on self-reporting, and dose defining the exact extent to which the lens of the eye
values were not made known. was included in the primary irradiation beam. Even in

© 2019 The Royal Australian and New Zealand College of Radiologists 3


R Poon and MK Badawy

Table 1. Reported radiation dose to the lens of the eye from CT scans to the head

References Scanner Method Gantry tilt Lens dose (mGy)

McLaughlin Toshiba GX Xpress Clinical trial (1 thermoluminescent Supraorbital line 6.0


and Mooney26 dosimeter [TLD] on the right eye only)
Heaney and Norvill22 Toshiba Medical Systems Phantom trial (2 TLDs on the right eye Hard palate† 120.0
Asteion 4 slice only) Reid’s baseline‡ 130.0
Supraorbital line§ 15.0
Abdeen et al.24 16 slice (Lightspeed 16; General Clinical trial (metal–oxide– Angulation to the Sequential: 61.5
Electric) and 64 slice (VCT; semiconductor field-effect transistor skull base was Helical: 43.5
General Electric) [MOSFET] system taped to the attempted
patients’ eyelid)
Wang et al.25 Dual-source Definition Flash; Phantom trial (1 optically stimulated Not specified 32.2
Siemens Health care luminescence detector per eye)
Lai et al.3 Discovery CT750 HD with Phantom trial (5 TLDs per eye) 0-degree tilt 30.7
Adaptive Statistical Iterative 30-degree cranial tilt 2.4
Reconstruction (ASiR)
Nikupaavo et al.29 128 slice Somatom Definition Phantom trial (1 MOSFET per eye) Angled from the skull base to 18.3
AS+; Siemens Health care the radix nasi
Ploussi et al.23 16 slice (Brightspeed; GE Health Clinical trial (2 TLDs per eye) Angulation to the supraorbital 0.8-1 yo: 10.5
care) line was attempted (but 2.0-4.9 yo: 29.9
unsuccessful in some cases) 5.5-15.5 yo: 34.2

†Hard palate: a line running parallel to the hard palate.


‡Reid’s baseline: a line joining the superior border of the external auditory meatus of the infra-orbital margin.
§Supraorbital line: a line running from the upper border of the orbits to the midpoint of the external auditory meatus, parallel to the base of the
skull.

studies where the value of the tilt was specified,3 the present arguments regarding the value and existence of
region of the lens included in the primary beam was a threshold value are still ongoing.
not explicitly determined. Given that the natural posi- Worth noting is that the radiation risk of opacity for-
tion of the head/phantom can vary substantially with mations is difficult to differentiate from the risk of visual
or without the addition of a head holder, particularly in impairment. The majority of the suggested dose values
clinical settings in regards to patient size, presentation for cataracts formation are in regards to the formation of
and anatomy, the inclusion of the lens changes accord- opacities, which in and of themselves may not present as
ingly, and thus, the comparability of these findings in a concerning factor. As mentioned above, studies have
an attempt to converge on an ‘average’ lens dose is been performed to address such matters based on the
complicated. incidence of cataract surgeries.13,14 The concept of visual
Another consideration to take into account is the impairment or deterioration, however, is highly subjec-
method of dose measurement. The number and place- tive and likely dependent on an individual’s sight toler-
ment of dosimeters differ from each trial, which could ance. Consequently, self-reporting and access to surgical
affect the consistency and reproducibility of the results intervention may affect dose-effect conclusions. The risk
with only one or two devices. As well as this, the inclu- of global visual disability may also be confounded by
sion of the dosimeter in the scout images, as opposed to existing co-contributing factors such as diabetes or sys-
being attached after the scout images have been taken, temic steroid use, making the influence of radiation
could have implications on the end dose;3,26 however, effects alone challenging to appreciate.
admittedly its contribution would likely be relatively low. In light of these discussions, it is therefore challenging
to pinpoint the exact cataractogenic risk posed by CT
Cataractogenic Risk of Brain CT examinations of the brain. In the absence of such a
definitive argument, it is therefore advisable that all
Despite the majority of the aforementioned CT brain practical measures be taken to reduce the radiation dose
studies involving lens radiation doses much lower than to the lens of the eye to as low as reasonably achievable.
the speculated 0.5 Gy threshold dose, Suzuki et al.27
and Wang et al.25 raise the concern that patients may
often be subjected to multiple examinations, as well as
Radiation reduction strategies
repeat imaging and/or multi-phase contrast studies. As a
Gantry tilt
result, these cumulative radiation doses may still have
the potential to provoke a cataractogenic response. Fur- The inclusion of the orbits is deemed mostly unnecessary
thermore, this must be considered in the context that in the majority of CT brain examinations, and so it stands

4 © 2019 The Royal Australian and New Zealand College of Radiologists


CT Brains and the risk of cataracts

to reason that emphasis should be made on excluding


Bismuth shielding
the lens of the eyes from the primary irradiation beam
where possible.28 In cases where this is adequately Bismuth shielding is effective in reducing radiation dose
achieved, any dose to the lens of the eye should there- to superficial tissues, including the lens of the eye. Dose
fore solely be the result of scattered radiation.26 Cur- reductions of approximately 30% are possible assuming
rently, the overarching consensus suggests that tilting a 0-degree gantry tilt technique.3,22,25,26,31 However, the
the gantry along the supraorbital line, based on topo- commercial viability of single-use shields is a compelling
gram planning, is the most effective method of dose concern. While the concept of multi-use shields may pre-
reduction, with reductions reported as high as sent itself as a more economically feasible alternative,
92%.3,22,28,29 For this sort of outcome to be achievable, the maintenance of suitable infection control measures
additional forward tilting of the head with the aid of a must then become a consideration.25,26 Erroneous place-
dedicated head holder or sponge may also be necessary ment of the shields on the face before scout imaging
for scanners where the degree of gantry tilt is limited.24 may also result in a compensatory increase in tube cur-
No literature was found at the time of writing regard- rent, and therefore increase in lens dose, where auto-
ing the dose reduction potential while solely using for- matic exposure control (AEC) is utilised.29
ward head tilt, which involves aligning the supraorbital The use of bismuth shields not only reduces the dose
line perpendicularly to the scanning table with zero gan- to anterior structures such as the eye, but it also attenu-
try tilt. While this practice has been observed clinically ates what would be potentially diagnostic photons from
by the authors and should theoretically achieve the same reaching the detector,32 giving rise to streaking, scatter
geometric advantages as gantry tilting, we hypothesize artefacts and altered CT numbers. In addition to dimin-
that forward head tilting alone may be less effective in ishing the overall image quality, these undesirable
lens radiation dose reduction than appropriately utilised changes in CT number values also have the potential to
gantry tilting. This is because the positioning of the influence the outcomes of certain quantitative diagnostic
patient may have more of an operator bias and is based examinations, such as brain perfusion analyses.25 As a
on an external visual assessment of the head as opposed means to overcome this, it has been suggested that
to being planned on a topogram image. With the lens of increasing the separation between the surface of the eye
the eye being such a small structure, even a few degrees and the shield may significantly decrease the associated
of deviation could potentially affect the overall radiation image artefacts.25 Increasing the distance of the shield
lens dose. Moreover, patients with large necks, neck stiff- from the skin surface, however, also slightly reduces its
ness or trauma history may struggle to achieve such dose reduction efficacy. In any case, the presence of
positioning. artefacts which originate from appropriately placed
Some studies have proposed that the application of shields are generally thought to be contained to the orbi-
gantry tilt may cause difficulties in being able to appreci- tal region and therefore are not believed to have any sig-
ate pathology in the base of the skull as a result of beam nificant impact on the regions of the brain typically
hardening and partial volume averaging.3,22 Nikupaavo required for diagnostic imaging.29,31 Dose reduction ver-
et al.29 even found an absolute difference of 3 Hounsfield sus image quality considerations should be made in this
Units (HU) in the posterior fossa where this technique situation.
was used, which, given that contrast between grey and The benefit of bismuth shielding in examinations
white matter can vary as little as 4-5 HU, could be con- where adequate gantry tilt is obtainable has at times
sidered substantial. However, helical acquisitions coupled been considered unjustified given that some studies
with a thin slice CT technique have been shown to reduce demonstrate no discernible differences with or without
such effects.22,24,30 Interestingly, earlier research con- the addition of shields.22 However, McLaughlin and
ducted by Yeoman et al.28 noted that the presence of Mooney26 did find an 18% decrease in dose with bis-
posterior fossa artefacts in their study was not consid- muth shielding relative to a gantry tilted reference
ered statistically significant and that it was also possible scan. Nikupaavo et al.29 also found the addition of bis-
for these artefacts to occur where scans were inclusive muth shielding partially consequential in its combined
of the lenses. use with gantry tilt, with overall results demonstrating
It is best to note that the gantry tilting technique may a further 8% decrease compared to gantry tilt alone.
not be as effective in isolation where full exclusion of the Nonetheless, given that the dose associated with
lens of the eyes is clinically unachievable, such as with appropriate gantry tilt could be considered relatively
patients wearing an immobilisation collar. Additionally, modest in the first place, some as low as 2 mGy, the
many scanners are incapable of gantry tilt, and certain rationale for its use in these circumstances may be dis-
protocols may necessitate a non-angled gantry. In such putable. Furthermore, the use of shielding may not be
cases, it is recommended that alternative dose reduction advisable for emergency patients, paediatrics or dis-
techniques, as discussed below, are considered where abled individuals who are unable to tolerate the use of
practicable. the shield.

© 2019 The Royal Australian and New Zealand College of Radiologists 5


R Poon and MK Badawy

of modulation was arranged such that the tube current


Organ-based tube current modulation (TCM)
was decreased by 75% for the anterior 120 degrees of
Organ-based TCM involves modulation of the tube cur- the tube rotation (where it is assumed the eyes would be
rent based on longitudinal or angular position around the in a correctly positioned patient). For the remaining 240
patient. In a study conducted by Wang et al.25, this form degrees, the current was increased by 25% to give the

Table 2. The effectiveness of varying strategies for the reduction of radiation dose to the lens of the eye

References Base Scan Dose Reduction Strategy Lens Dose Image Quality
Reduction (%)

Yeoman et al.28 Non-shielded scan with no gantry Angulation to the supraorbital line 87% No significant differences in posterior
tilt fossa artefacts compared to a non-
tilted reference scan
McLaughlin and Non-shielded scan with angulation Bismuth shielding 18% Not assessed
Mooney26 from the superior orbital margin
to the base of the skull
Heaney and Dose reductions associated with Angulation to the supraorbital line 87% Not assessed
Norvill22 angulation to the supraorbital Bismuth shielding and angulation along: 43%
margin were calculated against -Hard palate† 48%
the doses recorded when angling -Reid’s baseline‡ 0%
to the hard palate or Reid’s -Supraorbital line§
baseline.
Dose reductions associated with
bismuth shielding were calculated
against the doses recorded along
its corresponding angulation
plane when unshielded.
Abdeen et al.24 Sequential scan with attempted Helical scan with attempted angulation 29% Image quality is comparable
angulation to the base of the skull to the base of the skull
Wang et al.25 Non-shielded scan with no organ- Bismuth shielding 26% Image noise increased for all reduction
based TCM or decreased tube (1 layer) 42% methods.
current utilised. No angulation of Bismuth shielding 30% CT numbers increased with the use of
the scanning plane was specified (2 layers) 47% the bismuth shield
Organ-based TCM 30%
Bismuth shielding and organ-based
TCM
Decreased tube current
Ryska et al.2 Non-shielded scan with no gantry Iterative reconstruction protocol with 33% Image noise slightly increased relative
tilt using a filtered back projection associated reduction in tube current to standard filtered back projection,
protocol (reference mAs value set (reference mAs value lowered to 200 but differences in diagnostic quality
to 300) mAs) not deemed significantly significant
Lai et al.3 Non-shielded scan with no gantry Bismuth shielding 24% Significant drift in mean CT number
tilt Bismuth shielding with face shield (air 22% observed with bismuth shielding
gap) 32% alone, the effects of which gradually
Bismuth shield with bolus 35% decreased as air gaps and bolus were
Bismuth shield with bolus and air gap 92% introduced
30-degree cranial tilt
Nikupaavo Non- shielded scan with no gantry Bismuth shield after scout image 17% Organ-based TCM demonstrated an
et al.29 tilt or organ-based TCM. Bismuth shield before scout image 19% increase in noise in the posterior and
Bismuth shield with gantry tilt 55% central regions of the brain relative to
Gantry tilt 47% the reference scan (absolute change
Gantry tilt with half the recommended 14% of 1.5 HU).
angle 28% An absolute difference of 3HU
Organ-based TCM 54% observed in the posterior fossa where
Organ-based TCM with gantry tilt gantry tilt was used.

†Hard palate: a line running parallel to the hard palate.


‡Reid’s baseline: a line joining the superior border of the external auditory meatus of the infra-orbital margin.
§Supraorbital line: a line running from the upper border of the orbits to the midpoint of the external auditory meatus, parallel to the base of the
skull.

6 © 2019 The Royal Australian and New Zealand College of Radiologists


CT Brains and the risk of cataracts

same compensatory tube current–time product as with


Helical versus sequential scanning
the reference scan. This specific allocation of increased
current to the posterior and lateral positions was to allow Investigation as to the implications of helical versus
for adequate penetration to these regions. sequential scanning techniques has also been attempted
The advantage of using angular TCM is that it avoids in the past24 with statistically significant results implying
compensational increases in tube current based on dense that helical scanning is associated with lower radiation
materials such as shielding, and so the use of this tech- doses to the lens of the eye (about two thirds of that
nique in conjunction with bismuth shields has been sug- seen in sequential scans). An increased kVp to the poste-
gested as beneficial.25 However, Nikupaavo et al.29 rior fossa during sequential scans and differences in
recommend that the combination of organ-based TCM scanning geometry are thought to be the main contribu-
and bismuth shielding should be used with caution tors to these outcomes. In terms of the use of helical
granted that the increased tube output at the posterior multislice CT scanning in conjunction with a dose reduc-
aspect of the patient could cause significant backscatter tion technique such as gantry tilting, Geleijns et al.36
into the patient with the presence of the shield. This makes an interesting note commenting that this combi-
approach also requires the optimal positioning of the nation of practices may in effect be counterintuitive
patient’s head with their eyes pointed directly upwards. depending on the extent of over-ranging in the z-direc-
It may therefore not be appropriate for use in emergen- tion. While this is a theoretically plausible argument, it
cies or with paediatric or disabled individuals where posi- has yet to be specifically evidenced. Therefore, it is still
tioning circumstances may not be ideal. suggested that helical brain CT acquisitions remain the
more advantageous scanning technique with regards to
irradiation of the eyes when compared to their sequential
Global reduction of tube current
scanning equivalent.
A phantom study performed by Wang et al.25 utilising a While all of the above techniques have been reported
global reduction of tube current produced similar reduc- to significantly reduce the dose to the lens of the eye
tions in lens dose to those found with tube current mod- (see Table 2), it is important to note that this may not
ulation or bismuth shielding (~30%). This approach necessarily correlate to an absolute reduction in the for-
involved a 30% reduction in the tube current over the mation of radiation cataracts. No explicitly direct links
full 360-degree rotation such that the same dose would have been made at the time of writing between the use
be given to the lens of the eye if a single layer of bismuth of these strategies and a decreased risk of cataracto-
shielding had been used. In the results that followed, genic effects; the reduction in risk is instead inferred
previously mentioned CT number errors and image arte- owing to the discussion above as to whether cataracts
facts associated with bismuth shielding were eliminated, are considered as a deterministic or stochastic effect.
although the detected levels of noise were increased by The dose reduction percentages as outlined in Table 2
approximately 1 HU when compared to the reference should be observed with caution given that they may
scan. An additional reported benefit of the reduced tube vary depending on whether the reference scan had a
current technique was a ~29% reduction in the overall 0-degree gantry tilt or was aligned to the supraorbital
CTDIvol relative to the reference scan, which should be margin.
expected given the linear proportionality between the
CTDIvol and tube current. Given these advantages, if the
observed slight increases in noise can be diagnostically
Conclusion and Recommendations
tolerated, this method could potentially be considered In the absence of a wholly agreed upon lens dose thresh-
more beneficial than the use of bismuth shielding, partic- old evidenced by all relevant and up to date studies, it is,
ularly in scanners not equipped with Organ-based TCM therefore, prudent to act with caution and reinforce the
or iterative reconstruction. utilisation of dose reduction techniques. Important to
More recently, reductions in global tube current, with- note, however, is that the use of the strategies outlined
out significant compromises in image quality, have above is only recommended on the condition that image
become more feasible following the introduction of more quality remains unaffected or at least remains diagnosti-
effecient iterative reconstruction methods. In a clinical cally acceptable since being able to address the clinical
trial, Ryska et al.2 observed an overall 33% decrease in concern remains the primary focus.
average effective dose with such techniques, with an Overall, the implementation of gantry tilt along the
ensuing 33% decrease in dose detected at the eyes. supraorbital margin is recommended as standard prac-
Other studies have also demonstrated potential reduc- tice owing to the current evidence which suggests it to
tions in effective dose with the use of iterative recon- be the most effective dose reduction technique for the
struction;33–35 however, direct inferences regarding any lens of the eye. This strategy also has the additional ben-
reductions in dose to the lenses of the eyes were not efit of not requiring any further financial considerations
specified in these cases. which may arise with methods such as single-use

© 2019 The Royal Australian and New Zealand College of Radiologists 7


R Poon and MK Badawy

bismuth shields. The supplementary global reduction of 8. Kleiman NJ. Radiation cataract. Ann ICRP 2012; 41: 80–
tube current, the use of iterative reconstruction, and the 97.
application of organ-based TCM could also be beneficial 9. Hamada N, Fujimichi Y. Role of carcinogenesis related
in reducing the radiation dose to the lens of the eye. If mechanisms in cataractogenesis and its implications for
sufficient gantry tilting is not clinically feasible, the use ionizing radiation cataractogenesis. Cancer Lett 2015;
of bismuth eye shields is then advised. However, care 368: 262–74.
should be taken in regards to appropriate placement and 10. Stein JD. Serious adverse events after cataract surgery.
Curr Opin Ophthalmol 2012; 23: 219–25.
cleanliness of the shields to minimise artefacts and infec-
11. Shore RE, Neriishi K, Nakashima E. Epidemiological
tion control implications. A possible avenue for future
studies of cataract risk at low to moderate radiation
research on cataractogenic risks would be for longitudi-
doses: (not) seeing is believing. Radiat Res 2010; 174:
nal studies to be performed solely in regards to medical
889–94.
radiation exposures, with the use of comprehensive
12. Stewart F, Akleyev A, Hauer-Jensen M, et al. ICRP
health care and imaging databases to account for radia-
publication 118: ICRP statement on tissue reactions and
tion doses, numbers of scans, comorbidity factors and early and late effects of radiation in normal tissues and
the incidence of reported opacities or cataract surgeries. organs–threshold doses for tissue reactions in a
The compilation of such data would, however, be a con- radiation protection context. Ann ICRP 2012; 41: 1–
siderably extensive task and decades of follow-up infor- 322.
mation would be required in order to make an informed 13. Neriishi K, Nakashima E, Minamoto A, et al.
assessment. Postoperative cataract cases among atomic bomb
survivors: radiation dose response and threshold.
Radiat Res 2007; 168: 404–8.
References 14. Nakashima E, Neriishi K, Minamoto A. A reanalysis of
atomic-bomb cataract data, 2000–2002: a threshold
1. Medicare Australia. Medicare Australia statistics: analysis. Health Phys 2006; 90: 154–60.
January 2017 to December 2017 [Internet]. 2018 15. Chodick G, Bekiroglu N, Hauptmann M, et al. Risk of
[updated Oct 2018; cited 18 Oct 2018]. Available from: cataract after exposure to low doses of ionizing
http://medicarestatistics.humanservices.gov.au/statistic radiation: a 20-year prospective cohort study among US
s/do.jsp?_PROGRAM=%2Fstatistics%2Fmbs_group_sta radiologic technologists. Am J Epidemiol 2008; 168:
ndard_report&DRILL=on&GROUP=5&VAR=service 620–31.
s&STAT=count&RPT_FMT=by+state&PTYPE=calyea 16. Broughton J, Cantone M, Ginjaume M, Shah B,
r&START_DT=201701&END_DT=201712 Czarwinski R. Implications in dosimetry of the
2. Ryska P, Kvasnicka T, Jandura J, Klzo L, Grepl J, Zizka J. implementation of the revised dose limit to the lens of
Reduction of effective dose and organ dose to the eye the eye. Radiat Prot Dosimetry 2015; 164: 70–4.
lens in head MDCT using iterative image reconstruction 17. Chen W, Hwang J, Chen M, Chang W. Lenticular
and automatic tube current modulation. Biomed Pap opacities in populations exposed to chronic low-dose-
Med Fac Univ Palacky Olomouc Czech Repub 2014; rate gamma radiation from radiocontaminated buildings
158: 265–72. in Taiwan. Radiat Res 2001; 156: 71–7.
3. Lai CWK, Chan HY, Cheung TP, Wong TH. Reducing the 18. Klein B, Klein R, Linton K, Franke T. Diagnostic x-ray
radiation dose to the eye lens region during CT brain exposure and lens opacities: the Beaver Dam Eye
examination: the potential beneficial effect of the Study. Am J Public Health 1993; 83: 588–90.
combined use of bolus and a bismuth shield. 19. Klein BE, Klein RE, Moss SE. Exposure to diagnostic x-
Radioprotection 2015; 50: 195–201. rays and incident age-related eye disease. Ophthalmic
4. Melnick ER, Shafer K, Rodulfo N, et al. Understanding Epidemiol 2000; 7: 61–5.
overuse of computed tomography for minor head injury 20. Hourihan F, Mitchell P, Cumming R. Possible associations
in the emergency department: a triangulated qualitative between computed tomography scan and cataract: the
study. Acad Emerg Med 2015; 22: 1474–83. Blue Mountains Eye Study. Am J Public Health 1999;
5. Klang E, Beytelman A, Greenberg D, et al. Overuse of 89: 1864–6.
head CT examinations for the investigation of minor 21. Yuan M, Tsai DC, Chang SC, et al. The risk of cataract
head trauma: analysis of contributing factors. J Am Coll associated with repeated head and neck CT studies: a
Radiol 2017; 14: 171–6. nationwide population-based study. Am J Roentgenol
6. Rehani MM, Vano E, Ciraj-Bjelac O, Kleiman NJ. 2013; 201: 626–30.
Radiation and cataract. Radiat Prot Dosimetry 2011; 22. Heaney D, Norvill C. A comparison of reduction in CT
147: 300–4. dose through the use of gantry angulations or bismuth
7. Thome BC, Chambers MD, Hooker WA, Thompson RJ, shields. Australas Phys Eng Sci Med 2006; 29: 172–8.
Boreham RD. Deterministic effects to the lens of the eye 23. Ploussi A, Stathopoulos I, Syrgiamiotis V, et al. Direct
following ionizing radiation exposure: is there evidence measurements of skin, eye lens and thyroid dose during
to support a reduction in threshold dose? Health Phys pediatric brain CT examinations. Radiat Prot Dosimetry
2018; 114: 328–43. 2018; 179: 199–205.

8 © 2019 The Royal Australian and New Zealand College of Radiologists


CT Brains and the risk of cataracts

24. Abdeen N, Chakraborty S, Nguyen T, et al. Comparison image quality and posterior fossa artifacts in routine
of image quality and lens dose in helical and sequentially brain imaging with standard protocols. Eur Radiol 2008;
acquired head CT. Clin Radiol 2010; 65: 868–73. 18: 1720–6.
25. Wang J, Duan X, Christner JA, Leng S, Grant KL, 31. Hopper K, Neuman JD, King S, Kunselman A.
McCollough CH. Bismuth shielding, organ-based tube Radioprotection to the eye during CT scanning. Am J
current modulation, and global reduction of tube current Neuroradiol 2001; 22: 1194–8.
for dose reduction to the eye at head CT. Radiology 32. Geleijns J, Wang J, McCollough C. The use of breast
2012; 262: 191. shielding for dose reduction in pediatric CT: arguments
26. McLaughlin DJ, Mooney RB. Dose reduction to against the proposition. Pediatr Radiol 2010; 40: 1744–
radiosensitive tissues in CT. Do commercially available 7.
shields meet the users' needs? Clin Radiol 2004; 59: 33. Bodelle B, Wichmann JL, Scholtz JE, et al. Iterative
446–50. reconstruction leads to increased subjective and
27. Suzuki S, Furui S, Ishitake T, et al. Lens exposure objective image quality in cranial CT in patients with
during brain scans using multidetector row CT scanners: stroke. Am J Roentgenol 2015; 205: 618–22.
methods for estimation of lens dose. Am J Neuroradiol 34. Cheng B, Xing H, Lei D, et al. Impact of iterative model
2010; 31: 822–6. reconstruction combined with dose reduction on the
28. Yeoman LJ, Howarth L, Britten A, Cotterill A, Adam EJ. image quality of head and neck CTA in children. Sci Rep
Gantry angulation in brain CT: dosage implications, 2018; 8: 12613.
effect on posterior fossa artifacts, and current 35. Moorin RE, Gibson DAJ, Forsyth RK, Fox R. The impact
international practice. Radiology 1992; 184: 113–6. of iterative reconstruction on computed tomography
29. Nikupaavo U, Kaasalainen T, Reijonen V, Ahonen SM, radiation dosimetry: evaluation in a routine clinical
Kortesniemi M. Lens dose in routine head CT: setting. PLoS ONE 2015; 10: e0138329.
comparison of different optimization methods with 36. Geleijns J, Salvado Artells M, Veldkamp W, Lo  pez
anthropomorphic phantoms. Am J Roentgenol 2015; Tortosa M, Calzado CanteraA. Quantitative assessment
204: 117–23. of selective in-plane shielding of tissues in computed
30. Ertl-Wagner B, Eftimov L, Blume J, et al. Cranial CT with tomography through evaluation of absorbed dose and
64-, 16-, 4- and single-slice CT systems-comparison of image quality. Eur Radiol 2006; 16: 2334–40.

© 2019 The Royal Australian and New Zealand College of Radiologists 9

You might also like