Professional Documents
Culture Documents
BAERD GEN-007
REV. C
A.Bourgeois D.Rivest
LES RENSEIGNEMENTS, LES DONNÉES ET DESSINS TECHNIQUES DIVULGUÉS DANS LES THE INFORMATION, TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY
PRÉSENTES SONT LA PROPRIÉTÉ EXCLUSIVE DE BOMBARDIER INC. OU SES GROUPES, OF BOMBARDIER INC. OR ITS GROUPS, DIVISIONS OR SUBSIDIARIES OR CONTAIN PROPRIETARY RIGHTS
DIVISIONS OU FILIALES CORPORATIVES OU COMPORTENT DES DROITS DE PROPRIÉTÉ
APPARTENANT À AUTRUI ET NE PEUVENT ÊTRE DIVULGUÉS À QUICONQUE NI UTILISÉS PAR QUI OF OTHERS AND ARE NOT TO BE USED OR DISCLOSED TO OTHERS WITHOUT THE WRITTEN CONSENT OF
QUE CE SOIT SANS LE CONSENTEMENT ÉCRIT DU TITULAIRE DES DROITS. LE DÉTENTEUR DES THE PROPERTY OWNER. THE RECIPIENT OF THIS DOCUMENT, BY ITS RETENTION AND USE, AGREES TO
PRÉSENTES S’ENGAGE, DANS TOUTES CIRCONSTANCES, À GARDER CONFIDENTIELS LES HOLD IN CONFIDENCE THE TECHNICAL DATA AND DESIGNS CONTAINED HEREIN. THE FOREGOING SHALL
DONNÉES ET DESSINS TECHNIQUES QU’ELLES CONTIENNENT. CE QUI PRÉCÈDE NE S’APPLIQUE NOT APPLY TO PERSONS HAVING PROPRIETARY RIGHTS TO SUCH INFORMATION, TECHNICAL DATA OR
PAS AUX PERSONNES QUI ONT DES DROITS DE PROPRIÉTÉ SUR CES RENSEIGNEMENTS,
DESIGNS TO THE EXTENT THAT SUCH RIGHTS EXIST. D.4204 1995-02
DONNÉES TECHNIQUES ET DESSINS DANS LA MESURE OÙ CES DROITS EXISTENT.
REVISION HISTORY
LTR DATE REVISED BY APPROVED BY PAGES AFFECTED
Comments: Section 1.1: Editorial change from DER comments see eBCR #7940.
Table 2: Editorial change from DER comments see eBCR #7940.
Section 2.1: Add BAEMM-001
Section 3: SAC acronym added next to proper definition and definition of use
added, definition of expendable and non-expendable T/C added
Section 4.1.1: Editorial change from DER comments see eBCR #7940.
Section 4.2: High point instead of mid point
Section 4.3: Editorial change from DER comments see eBCR #7940.
Section 4.4: Chemical tank should be Class C, Editorial change from DER
comments see eBCR #7940.
Section 4.5: Tº interval conversion to ºC corrected
Section 4.5.1: Editorial change from DER comments see eBCR #7940.
Section 4.5.2: Full reworded to cover all the requirements for to re-use of SAT,
TUS and load thermocouples
Section 5.1: Correct clerical error in T/C part number
Section 8.2: TUS to be conducted once a year at min and max operating Tº is
added
Section 8.2.1: Editorial change from DER comments see eBCR #7940.
Section 8.2.2: Editorial change from DER comments see eBCR #7940.
Section 8.2.3: Align number of T/C and location for all equipment other the
quench tank.
Table 3: The tank unit volume are changed to liter (US gallon)
Section 8.2.4, 8.3, 9.1, 9.2: Editorial change from DER comments see eBCR
#7940.
Table 4: Editorial change from DER comments see eBCR #7940
(eBCR # 8025)
DISPOSITION OF STOCK:
MATERIALS: Use
PARTS PROCESSED: Use as is
B DOC:
REV:
BAERD GEN-007
C
PAGE: 1
DATE: 2011-09-21
TABLE OF CONTENTS
1 SCOPE.................................................................................................................... 4
1.1 CLASSIFICATION ......................................................................................... 4
2 APPLICABLE DOCUMENTS ................................................................................. 6
2.1 BOMBARDIER AEROSPACE DOCUMENTS ............................................... 6
2.1.1 Bombardier Aerospace Specifications ............................................... 6
2.1.2 Bombardier Aerospace Manuals........................................................ 6
2.2 INDUSTRY SPECIFICATIONS ..................................................................... 6
3 DEFINITIONS ......................................................................................................... 7
4 FACILITIES............................................................................................................. 9
4.1 INSTRUMENTATION REQUIREMENTS FOR MATERIAL
PROCESSING EQUIPMENT ........................................................................ 9
4.1.1 General .............................................................................................. 9
4.2 REFRIGERATORS...................................................................................... 10
4.3 QUENCH BATHS ........................................................................................ 10
4.4 CHEMICAL PROCESSING EQUIPMENT ................................................... 10
4.5 TEMPERATURE SENSORS ....................................................................... 10
4.5.1 Thermocouples ................................................................................ 11
4.5.2 Re-use of Thermocouples for SAC, TUS and load .......................... 12
4.5.2.1 General ..................................................................................... 12
4.5.2.2 Recalibration ............................................................................. 12
4.5.2.3 Re-use of Thermocouples for TUS............................................ 12
4.5.2.4 Re-use of Thermocouple for SAC ............................................. 12
4.5.2.5 Load sensors ............................................................................ 13
4.6 TEST INSTRUMENTS................................................................................. 13
5 MATERIALS ......................................................................................................... 14
5.1 AIRCRAFT AND PROCESS MATERIALS .................................................. 14
5.2 MISCELLANEOUS ...................................................................................... 14
TABLE OF CONTENTS
6 HEALTH AND SAFETY........................................................................................ 15
7 REQUIREMENTS ................................................................................................. 15
8 TESTING PROCEDURE....................................................................................... 15
8.1 GENERAL ................................................................................................... 15
8.2 TEMPERATURE UNIFORMITY SURVEY................................................... 15
8.2.1 Equipment Conditioning................................................................... 16
8.2.2 Recording Data................................................................................ 16
8.2.3 Test Sensor Numbers and Locations............................................... 17
8.2.4 Qualification Temperature Uniformity Survey .................................. 18
8.3 INSTRUMENTATION SYSTEM ACCURACY CHECK ................................ 19
9 QUALITY CONTROL............................................................................................ 20
9.1 RECORDS................................................................................................... 20
9.2 QUALITY CONTROL REQUIREMENTS ..................................................... 21
LIST OF TABLES
TABLES
TABLE 1 PROCESS EQUIPMENT CLASSES................................................................ 5
1 SCOPE
This specification establishes the requirements for control of thermal
processing equipment and instruments. This specification applies but is not
limited to, heat treating and hot forming equipment, autoclaves, cavity and
platen presses, refrigerators, chemical and electrochemical processing
equipment, etc.
This specification also establishes procedures for temperature uniformity
surveys, system accuracy checks and system certification of thermal processing
equipment.
1.1 CLASSIFICATION
2 APPLICABLE DOCUMENTS
Except where a specific issue is indicated, the issue of the following documents
in effect at the time of manufacture shall form a part of this specification to the
extent indicated herein.
3 DEFINITIONS
Accuracy The maximum deviation of the instrumentation
being tested from the corrected readings of a
traceable standard.
Controller Permanently installed instrument used to maintain
the set process temperature during any thermal
processes.
Correction Factor The number of degrees, determined from the most
recent calibration which must be added or subtracted
from the sensor and/or instrument reading to obtain
NIST true temperature.
Expendable
Thermocouples Thermocouples made of wire covered by plastic or
fabric. Insulation is usually glass braid or ceramic
fiber cloth.
Heat Sensitive Marker Marker used to witness the maximum temperature
achieved during a thermal operation. Usually used
during hot forming or preheating operations.
High Limit Instrument Temperature instrument that causes shutdown of the
heat source when a safe temperature is exceeded.
Instrumentation System
Accuracy Check (SAC) An on-site comparison of the process
instrumentation readings (i.e. controller, recorder,
indicator, high limit or process sensors) with
readings of test instrumentation to determine if the
measured deviations are within the requirements of
Table 4 of this specification.
Load Sensor Sensor that is attached to the production parts or a
representation of the production parts, that supplies
temperature data to a working instrument.
Non-Expendable
Thermocouples Thermocouples that are not covered with fabric or plastic
insulations but are typically bare thermocouple wire
inserted in ceramic insulators and sometimes inserted in a
tube for protection.
4 FACILITIES
4.1.1 General
4.2 REFRIGERATORS
4.5.1 Thermocouples
For process and test thermocouples, use gage and alloy that is consistent with
the accuracy, sensitivity and thermocouple life required for the applicable
process conditions. Thermocouples shall be protected from deterioration
caused by furnace atmosphere, bath liquids and physical abuse.
Thermocouple calibration intervals specified in Table 4, are the maximums
permitted. However, compliance with these intervals does not relieve the user
of the responsibility for ensuring that excessive drift above Table 4
requirements has not occurred under the particular conditions (environment,
time, and temperature) of exposure. Shorter intervals shall be used until an
operating history is developed to justify the maximum limits or establish limits
for the particular normal, and extremes of exposure conditions.
Thermocouples covered by this specification shall be of the special limit error
type of ASTM E230.
It is permissible to use thermocouples made from calibrated wire rolls in lieu of
individually calibrated thermocouples.
Each stock length of insulated thermocouple wire shall be calibrated as a roll in
accordance with Section 4.5 and as follows:
For rolls of wire up to 3000 ft., 1 thermocouple from each end of rolls
of wire shall be calibrated.
For rolls of wire over 3000 ft. in length, calibrate one thermocouple
from each end and every 3000 ft. between.
Results shall meet the requirements of this section and in addition, all
thermocouples from the same roll must agree within 1°F (0.6ºC) for each
calibration temperature. If results do not agree, calibrate thermocouples from
intermediate locations. Only segments that meet the 1°F (0.6ºC) requirement
shall be used.
Each thermocouple shall be identified with a tag that refers to the related
calibration laboratory report number.
The corrections shown in the calibration laboratory report of a given number
shall be used for all temperature readings taken on thermocouples tagged with
the given number.
4.5.2.1 General
- For re-use of Types K and E thermocouples above 500ºF (260ºC), the depth of
insertion shall be equal to, or greater than the depth of insertion of any previous
use.
- The re-use of any thermocouple is prohibited unless the insulation remains intact
and the wires, including the hot junction, are not damaged.
4.5.2.2 Recalibration
- Recalibration of any expendable base metal test thermocouples is prohibited.
- Recalibration of Types K and E thermocouple is prohibited.
- Recalibration of base metal thermocouple J and N shall be initially and every 3
months maximum.
- Recalibration of base metal thermocouples used for load sensors is prohibited.
- Recalibration of noble thermocouples (B, R, S) shall be initially and every 6
months maximum
- Expendable base metal thermocouples are limited to a single use above 1800ºF
(982ºC).
- Non-expendable thermocouples shall be recalibrated per 4.5.2.2. Type K and E
shall be discarded after 3 months.
5 MATERIALS
All aircraft materials and process materials listed in this section shall be
controlled in accordance with BAEMM-001.
852-4875175 22 AWG, Chromel-Alumel Thermocouple
852-4875183 22 AWG, Copper-Constantin Thermocouple
852-4875191 20 AWG, Chromel-Alumel Thermocouple
MIL-W-5846 Chromel-Alumel Thermocouple
MIL-W-5908 Copper-Constantin Thermocouple
The use of other thermocouples is allowed as long as they meet the
requirements of this specification in its entirety.
5.2 MISCELLANEOUS
Materials listed in this section shall meet only those requirements stated
specifically against them in this section. BAEMM-001 is not applicable.
Heat sensitive markers must have an accuracy of ±1% of their stated
temperature.
Heat sensitive markers must be supplied as a lot, and shall be certified in
accordance with ISO 10012-1.
At Bombardier, each lot of markers along with the calibration report shall be
sent to the metrology authority for identification and approval prior to being
released for production.
7 REQUIREMENTS
Bombardier sites, and subcontractors shall process in strict accordance with
this specification and all sub-tier specifications, modified if necessary only by
change formally approved by BAMPE in response to request for deviation
using form BT0213-01.
8 TESTING PROCEDURE
8.1 GENERAL
Readings of the test sensors shall be started immediately after insertion of the
sensors in the equipment. Readings shall be continued with sufficient
frequency (maximum interval of five minutes between readings), to determine
when the hottest test sensor approaches the minimum of the equipment
temperature tolerance range as specified in Table 1 and Table 4.
From this stage until stable thermal conditions are reached in the equipment,
readings of the test sensors shall be continued at intervals of two minutes, in
order to detect any overshooting.
After stable thermal conditions are reached in the equipment at the control
temperature, readings of the test sensors shall be continued at intervals of five
minutes for sufficient time to determine the recurrent temperature pattern and
in any case for not less than thirty minutes.
BT0217-04 MW97 1999-11
B DOC:
REV:
BAERD GEN-007
C
PAGE: 17
DATE: 2011-09-21
9 0.5 working zone volume in ft 3 225
ft 3 OR
The qualification TUS shall be conducted in accordance with Section 8.2 with
the following additions.
The qualification TUS shall be conducted at the minimum and maximum
operation temperatures of the equipment.
Where the equipment is used for different thermal processes (e.g. solution heat
treatment and aging), the qualification TUS shall be conducted at the minimum
and maximum operating temperatures of each process.
For each item of equipment, the position and quantity of test sensors to be used
shall be determined by the Metrology authority. A sketch showing the pattern
and location of test sensors together with a description of the equipment, the
dimensions of the work zone and the thermal process involved shall be kept in
file and made available to Bombardier upon request.
The qualification TUS shall include an Instrumentation System Accuracy
Check in accordance with section 8.3 of this specification. This system
accuracy check shall be performed at each TUS temperature with the thermal
processing equipment empty (free of any load).
Unless other otherwise specified, the initial qualification temperature
uniformity survey is not applicable to the following types of equipment:
Drying ovens
Pre-heating equipment
9 QUALITY CONTROL
9.1 RECORDS
Process records shall be maintained on file for the time specified in the
applicable process specification. If no minimum time is specified in the
applicable process specification, process records shall be retained for a
minimum of 5 years.
Whenever temperature uniformity survey, instrumentation system accuracy
check, calibration of working or testing instruments or sensors has been
performed successfully, a calibration sticker shall be attached to the newly
calibrated equipment. This calibration sticker shall show the following
information, as applicable:
Date of calibration
Type of calibration
Quality stamps
EQUIPMENT
CLASS AND ACCEPTANCE
TEST PROCEDURE FREQUENCY
TYPE CRITERIA [1]
CLASS TYPE
5 A Monthly ± 5ºF (2.8ºC) inside
B the working zone
10 A Monthly ± 10ºF (5.6ºC) inside
B the working zone
Temperature 15 A Monthly ± 15ºF (8.3ºC) inside
Uniformity Survey (TUS) B, C the working zone
per section 8.2 25 A Semi-Annually ± 25ºF (13.9ºC) inside
[2] [3] [4] [5] [6] [14] the working zone
B, C Quarterly ± 25ºF (13.9ºC) inside
D, E
50 B, C Semi-Annually ± 50ºF (27.8ºC) inside
D, E the working zone
Instrumentation System Accuracy 5 A Weekly ± 2ºF (1.1ºC) with a
Check (SAC) per Section 8.3 B maximum instrument
[4] [6] [7] [8] [9] [14] adjustment of ± 3ºF
(1.7ºC)
10 A Weekly ± 3ºF (1.7ºC) with a
B maximum instrument
adjustment of ± 5ºF
(2.8ºC)
15 A Weekly ± 4ºF (2.2ºC) with a
B maximum instrument
C adjustment of ± 5ºF
(2.8ºC)
25 A Monthly ± 5ºF (2.8ºC) with a
[10] maximum instrument
adjustment of ± 5ºF
(2.8ºC)
B, C Weekly ± 5ºF (2.8ºC) with a
D, E maximum instrument
adjustment of ± 5ºF
(2.8ºC)
EQUIPMENT
CLASS AND ACCEPTANCE
TEST PROCEDURE FREQUENCY
TYPE CRITERIA [1]
CLASS TYPE
Instrumentation System Accuracy 50 B, C Monthly ± 10ºF (5.6ºC) with a
Check (SAC) per Section 8.3 [11] D, E maximum instrument
[4] [6] [7] [8] [9] [14] adjustment of 1% of the
scale
Calibration of Working 5 All Monthly ± 2ºF (1.1ºC) or ± 0.3 %
Instruments [12] [15] 10 All Monthly of the instrument
range whichever is
15 All Quarterly greater.
25 All Quarterly ± 2ºF (1.1ºC) for
50 All Semi-Annually digital instruments.
Calibration of Test Instruments N/A N/A Quarterly ± 1ºF (0.6ºC) up to
1000ºF (537.8ºC)
± 0.2% above 1000ºF
(537.8ºC)
Calibration of Secondary N/A N/A Yearly ±0.3ºF (0.2ºC) or
Standards ±0.05% of reading,
whichever is greater
Calibration of Primary Standards N/A N/A 3 years ±0.1ºF (0.06ºC) or
±0.015% of reading,
whichever is greater
Calibration of Working N/A N/A Before ± 2ºF (1.1ºC) or 0.4% of
Permanent Sensors installation reading
Load and Test Sheathed N/A N/A 6 Months ± 2ºF (1.1ºC) or 0.4%
per Section 4.5 Thermocouples of reading
[13]
Noble Non- N/A N/A 3 Months ± 2ºF (1.1ºC) or 0.4%
Sheathed of reading
Thermocouples
NOTES:
[1] Acceptance criteria are inclusive of sensor or recording instrumentation correction factors.
[2] The frequency of temperature surveys for equipment classes 5, 10, 15 and 25 may be reduced to a
maximum of semi-annually. The frequency reduction may be implemented, provided that the
equipment meets at least one of the following requirements:
a) There must be at least 2 recording sensors in each work zone, as defined in Table 2 for
instrumentation type A, or as defined in Table 2 with, in addition, duel working
BT0217-04 MW97 1999-11
B DOC:
REV:
BAERD GEN-007
C
PAGE: 24
DATE: 2011-09-21
EQUIPMENT
CLASS AND ACCEPTANCE
TEST PROCEDURE FREQUENCY
TYPE CRITERIA [1]
CLASS TYPE
instruments, which means that a check sensor is permanently installed in the thermal
equipment in such a way that it measuring device is within 3 inches of the equipment
working sensor for instrumentation type B. In both cases, there must be a history of
continuously satisfactory performance, including a record of consecutive satisfactory
thermal surveys over a period of 6 months immediately prior.
b) The thermal equipment is not used for metal heat treatment (solution H.T., annealing,
age, temper, stress relief, etc.) or adhesive bonding.
c) The thermal equipment is used for the fabrication of composite components and each part
is monitored by load sensors as per the requirements of the applicable process
specification.
d) For equipment using load sensors, after completion of 2 consecutive successful periodic
TUS in accordance with this Table, provided that a written preventive maintenance plan
is established which includes a schedule for replacing working sensors and recording of
all maintenance operations
[3] Chemical process tanks, shall be surveyed on a yearly basis only. Quench baths and refrigerators do
not require temperature uniformity survey to be performed unless otherwise required in the
governing process specification.
[4] When load sensors are inserted at the glue line of the parts (minimum of 4 per load) or when sensors
are permanently attached to the bonding fixture for equipment used for adhesive bonding, it is
permissible to waive the temperature uniformity surveys and instrumentation system accuracy
checks
[5] Cavity and platen presses used for adhesive bonding that meet the equipment requirements of class
10 Type A, do not require temperature uniformity surveys.
[6] When heat sensitive markers are used as indicators of temperature on each part to be formed, it is
permissible to waive the temperature uniformity survey and instrumentation system accuracy check
for hot forming presses used for forming metal parts.
[7] For chemical processing equipment, instrumentation system accuracy check shall be performed
quarterly on a minimum period of 2 years. After 2 years of conformant system accuracy check for
chemical processing equipment, it is permissible to reduce the frequency to a maximum of yearly. If
a SAC fails, the quarterly test shall be reinstated.
[8] For drying ovens, refrigerators and quench tanks, instrumentation system accuracy check shall be
performed on a yearly basis.
[9] These requirements are applicable to regular working instrument system only. Dual working
instrument, which means that a check sensor is permanently installed in the thermal equipment in
such a way that its measuring device is within 3 inches of the equipment working sensor, shall be
tested as follows:
A comparison shall be performed daily by Quality between the reading from the working
BT0217-04 MW97 1999-11
B DOC:
REV:
BAERD GEN-007
C
PAGE: 25
DATE: 2011-09-21
EQUIPMENT
CLASS AND ACCEPTANCE
TEST PROCEDURE FREQUENCY
TYPE CRITERIA [1]
CLASS TYPE
sensor(s) and the corresponding check sensor(s).
The instrumentation system accuracy shall be as per Table 4 for the applicable equipment
class.
When load sensors are inserted at the glue line of the parts (minimum of 4 per load) or when
sensors are permanently attached to the bonding fixture for equipment used for adhesive
bonding, it is permissible to waive the daily comparison.
For thermal equipment used for fabrication of composite components, daily comparison may
be waived providing that load sensors are used as per the governing process specification.
In addition a weekly check of non-controlling working instrument shall be performed by
Quality. The requirements of this Table for the applicable temperature tolerance shall be met.
Finally, instrumentation system accuracy check shall be performed on a quarterly basis for all
classes of equipment and semi-annually when permanent check sensors are sheathed
thermocouples.
[10] For Class 25, equipment shall have instrumentation system accuracy of 0.5% maximum of reading
above 1000ºF (537.8º C)
[11] For Class 50, equipment shall have instrumentation system accuracy of 1% maximum of reading
above 1000ºF (537.8º C)
[12] For thermal equipment used for adhesive bonding, the frequency of calibration of working
instruments is 6 months.
[13] For sheathed thermocouples used as load or test sensors that have a history of continuous
satisfactory performance, including a record of 2 consecutive satisfactory calibrations the previous year,
it is permissible to reduce the frequency of inspection to a maximum of annually.
[14] For equipment used for heat treatment of metals (BAPS 168 series), it is permissible to reduce
frequency in accordance with AMS 2750 Issue D
[15] It is permissible to extend the calibration of digital instruments to quarterly for Class 10 equipment