F1334 - 97 A - Weighted Sound Level of Vacuum Cleaners

You might also like

You are on page 1of 7

Designation: F 1334 – 97

Standard Test Method for


Determining A-Weighted Sound Power Level of Vacuum
Cleaners1
This standard is issued under the fixed designation F 1334; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope F 655 Specifications for Test Carpets and Pads for Vacuum
1.1 This test method covers the overall A-weighted sound Cleaner Testing4
pressure level emitted by small portable upright, canister, and 2.2 ANSI Standards:
combination vacuum cleaners intended for operation in domes- S1.4-1983 Specifications for Sound Level Meters5
tic and commercial applications. S12.31-1990 Precision Methods for the Determination of
1.2 The measurement is performed on a stationary vacuum Sound Power Levels of Broad Band Noise Sources in
cleaner in a semi-reverberant room. This test method deter- Reverberant Rooms5
mines sound power by a comparison method for small noise S12.32-1990 Precision Methods for Determination of
sources, that is, comparison to a broad band reference sound Sound Power Levels for Discrete Frequency and Narrow
source. Band Noise Sources in Reverberant Rooms5
1.3 This test method describes a procedure for determining S12.33-1990 Engineering Methods for Determination of
the approximate sound power level of small noise sources. This Sound Power Levels of Noise Sources in a Special
test method uses a non-special semi-reverberant room. Reverberant Test Room5
1.4 Results are expressed as sound power level in decibels. 2.3 ISO Standard:
1.5 The values stated in inch-pound units are to be regarded 3741, 3742, and 3743 are similar to and may be used in
as the standard. The SI units given in parentheses are for place of ANSI S12.31, S12.32 and S12.33 respectively5
information only. 2.4 IEC Standard:
1.6 This standard does not purport to address all of the 704.1 Test Code for the Determination of Airborne Acous-
safety concerns, if any, associated with its use. It is the tical Noise Emitted by Household and Similar Electrical
responsibility of the user of this standard to establish appro- Appliances5
priate safety and health practices and determine the applica- 3. Terminology
bility of regulatory limitations prior to use.
3.1 Definitions:
2. Referenced Documents 3.1.1 population—the total of all of the units of the particu-
2.1 ASTM Standards: lar model and/or type of vacuum cleaner being tested.
C 634 Definitions of Terms Relating to Environmental 3.1.2 population sample or sample—three or more units,
Acoustics2 randomly taken from the population.
E 177 Practice for Use of the Terms of Precision and Bias in 3.1.3 test unit or unit—a single vacuum cleaner of the
ASTM Test Methods3 model and/or type being tested.
E 691 Practice for Conducting an Interlaboratory Study to 3.1.4 Unless otherwise indicated, definitions are in accor-
Determine the Precision of a Test Method3 dance with Terminology C 634.
F 608 Method for Evaluation of Carpet-Embedded Dirt
4. Significance and Use
Removal Effectiveness of Household Vacuum Cleaners
(Laboratory Method)4 4.1 The test results enable the comparison of A-weighted
sound emission from vacuum cleaners when tested under the
condition of this test method.
1
This test method is under the jurisdiction of ASTM Committee F-11 on Vacuum 5. Test Room Requirements
Cleaners and is the direct responsibility of Subcommittee F11.25 on Sound 5.1 The test room shall be semi-reverberant. It shall contain
Measurement.
Current edition approved Nov. 10, 1997. Published August 1998. Originally sufficiently little sound absorption material so the requirements
published as F 1334 – 91. Last previous edition F 1334 – 91.
2
Annual Book of ASTM Standards, Vol 04.06.
3 5
Annual Book of ASTM Standards, Vol 14.02. Available from American National Standards Institute, 11 W. 42nd St., 13th
4
Annual Book of ASTM Standards, Vol 15.07. Floor, New York, NY 10036.

Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.

1
F 1334
of 5.2 can be met. It should be large enough to meet the 6.4 Reference Sound Source—The reference sound source
dimension requirements of 5.2. Check the calibration of each shall meet the requirements of Section 9 of ANSI S12.31.
microphone according to the instrument manufacturer’s direc- 6.5 Instrumentation:
tions.6 6.5.1 Thermometer, accurate to within 62°C.
5.2 Identify a location on the test room floor which can 6.5.2 A means of measuring relative humidity, accurate to
accommodate six equally spaced microphone positions 60° within 62 % over the range used.
apart located at a height of 60 in. (1.5 m) above the floor on a 6.5.3 Barometer, accurate to within 62 %.
12 ft (3.6 m) diameter circle the center of which is the center
of the sound source. These positions shall result in a standard 7. Operation of Vacuum Cleaner
deviation of the six sound pressure measurements of not more 7.1 Run-In—Operate new test cleaners continuously for at
than 2.3 dB when measuring the reference sound source. This least 1 h prior to testing. Canister cleaners shall be run open
location is suitable for the vacuum cleaner and reference sound with no hose attached. Upright and power nozzles shall be run
source for this test method. so that the rotating brush does not engage the carpet or the
5.3 Environmental—Ambient test conditions within the test floor.
room shall be controlled to within 20 6 4°C and 30 to 70 % 7.2 Warm-Up—Operate the cleaners for 10 min just prior to
relative humidity. making sound pressure level measurements in the same con-
5.4 Also, any room which has qualified in accordance with figuration as described in 7.1.
ANSI S1.26, S12.31, S12.32, S12.33, ISO 3741, 3742 and 7.3 Test Configuration:
3743 may be used to measure the sound power levels of 7.3.1 The vacuum cleaner shall be configured for the carpet
vacuum cleaners.7 cleaning mode.
5.5 The measured A-weighted sound pressure levels shall be 7.3.2 The dust bag or primary filter shall be new.
corrected for the influence of background noise according to 7.3.3 All belts shall be new at the start of the run-in.
Table 2. 7.3.4 Rotating agitator type cleaner including power nozzle
shall use the same setting as specified in Test Method F 608 for
6. Instrumentation and Equipment cleaning multilevel carpet that is as follows:
6.1 Acoustical Instrumentation—The sound measurement 7.3.4.1 If various settings are provided, set the motor speed
system shall be as specified in ANSI S1.4-1983. setting, suction regulator, or nozzle height, or combination
6.2 Voltage Regulator System—The regulator shall be ca- thereof, using the manufacturer’s specified setting for cleaning
pable of maintaining 61 V, the rated voltage (61 %) and multilevel carpet. Momentary or instantaneous speeds are not
frequency (61 Hz) with sinusoidal wave form. to be used.
6.3 Test Carpet and Pad—The test carpet and pad shall be 7.3.5 For straight air canister cleaners use the same setting
54 in. wide (137 cm) and 72 in. (183 cm) long. The carpet shall specified in Test Method F 608 for cleaning multilevel carpet,
be multilevel. Carpet and pad shall be as specified in Specifi- that is as follows:
cation F 655. 7.3.5.1 If various settings are provided, the motor setting,
suction regulator, or nozzle height, or combination thereof,
NOTE 1—For this test method two standard size 27 by 72-in. (69 by
183-cm) long test carpets could be placed side by side to make it 54 by using the manufacturer’s specific setting for cleaning multi-
72-in. (138 by 183-cm) long. It is recommended that the two pieces of test level carpet. If no specific instructions are given, or if judged
carpet be taped to the floor, side by side with the pile running in the same to be inadequate or unclear, position the nozzle on the carpet so
direction. the maximum suction is provided, taking care to maintain the
tilt angle throughout the test. Momentary or instantaneous
speeds are not to be used.
6
Further information provided in ANSI S1.10-1966 (R1986) Method for the 7.3.6 Voltage—Tests are to be conducted at the nameplate
Calibration of Microphones, available from American National Standards Institute,
11 W. 42nd St., 13th Floor, New York, NY 10036.
voltage (61 %) and frequency (61 Hz), when measured at the
7
Further information provided in ANSI S1.26-1978 (R1989) Method for the appliance plug. For cleaners with dual nameplate voltage
Calibration of the Absorption of Sound by the Atmosphere, available from American ratings, conduct sound tests at the highest voltage.
National Standards Institute, 11 W. 42nd St., 13th Floor, New York, NY 10036.
8. Location of Sound Sources and Equipment
TABLE 2 Corrections for Background Noise Levels 8.1 Carpet and Pad—Lay the carpet and pad directly on the
Correction to be Subtracted floor over the location defined in 5.2 with the geometric center
Difference Between Sound Pressure from Sound Pressure Level
Level Measured with Sound Source Operating Measured with Sound Source
of the carpet coinciding with the test location.
and Background Noise Operating to Obtain Sound 8.2 Locate the vacuum cleaner or the reference sound
Level Alone, dB Pressure Level Due to Sound source at the position determined in 5.2. All modules (generally
Source Alone, dB
the motor/blower and its housing in the base of an upright unit,
Less than 6 No correction allowed or in the canister of a canister unit) shall be adequately secured
6 1
7 1 in a manner which will not affect the sound pressure readings.
8 1 8.2.1 Reference Sound Source Location—Position the cen-
9 0.5 ter of the radiating portion (fan blades, for an aerodynamic
10 0.5
Greater than 10 0 source) over the location defined above. The reference sound
source is placed directly on the carpet.

2
F 1334
8.2.2 Vacuum Cleaner Location: 10.6 A minimum of two additional test units of the same
8.2.2.1 Center the principal module over the source location model must be selected in accordance with the sampling
as determined in 5.2. statement of Section 9. Repeat 10.1-10.4 for each additional
8.2.2.2 Take care to ensure the rotating brush model does test unit. See Annex A1 for a procedural example and whether
not damage the carpet seam (that is, center upright on carpet on additional units need to be tested.
center of carpet with the rotating brush perpendicular to seam). 10.7 The best estimate of sound power level for the popu-
8.2.2.3 Second Module—If there is a second module not lation of the vacuum cleaner model being tested is the
integral with the basic unit (for example, the power nozzle), it arithmetic mean of the sound power level of the sample
shall be positioned on the side of the principle module away population meeting the requirements of the sampling statement
from the center of the room. Position the second module in in Section 9.
such a way that, to the maximum extent possible, its sound
emissions are not shielded by the principal module, and vice 11. Calculation of Sound Power Levels for the
versa. For example, a position shall be such that air exhaust Comparison Method
noise will radiate into the room in a direction other than toward 11.1 Calculate the sound power level produced by the
the second module. source in each octave band within the frequency range of
8.2.3 Handle—Any operating handle shall be separately interest as follows. Subtract the band source pressure level
supported in a position such that the handle grip is 31 in. (79 produced by the reference sound source (corrected for back-
cm) above the carpet surface. Take care that the support ground noise according to 5.5) from the band sound pressure
structure does not introduce additional rattles, etc. There shall level of the source under test (corrected for the background
be a rigid clamping of the handle to the support structure. noise according to 5.5). Add the difference to the known sound
power level produced by the reference sound source:
9. Sampling
LW 5 LWr 1 Lp 2 Lpr (2)
9.1 Test a sufficient number of samples of each vacuum
cleaner model until a 90 % confidence level is established where:
within6 2.0 dBA of the mean value. Test a minimum of three LW 5 the octave-band or A-weighted sound power level,
samples. in decibels, produced by the source under test,
Lp 5 the average octave-band or A-weighted sound pres-
10. Procedure sure level, in decibels, produced by the source under
10.1 With the vacuum cleaner operating according to Sec- test, as determined in accordance with 10.1,
tion 7 of this test method, take six sound pressure level LWr 5 the known octave-band or A-weighted sound power
(A-weighted) readings from the positions in accordance with levels, in decibels, produced by the reference sound
5.2 (six is defined as the minimum number of microphone source, and
positions) with the microphone directed towards the sound Lpr 5 the average octave-band or A-weighted sound pres-
source. The use of more microphone positions is permissible. sure level, in decibels, produced by the reference
Calculate the average sound pressure level: sound source, as determined in accordance with

H J
10.1.
1 Nm
Lm 5 10 log N ( 10Li/10 (1)
m i51 12. Information
where: 12.1 General—Record the name and location of the test
LM 5 octave-band or A-weighted sound pressure level laboratory, including the date and time of the measurements.
averaged over all microphone positions, for a single 12.2 Test Room—Record the description of the room con-
source location, dB, struction, dimensions, configurations, and deployment of ab-
Li 5 octave-band or A-weighted sound pressure level for sorptive materials, etc.
the ith microphone position, dB, and 12.3 Equipment—Maintain recorded diagram of the acous-
Nm 5 number of microphone positions. tical data acquisition system. This shall include the model
10.2 Replace the vacuum cleaner with the reference sound number and serial number of all microphones, preamplifiers,
source. Operate the sound source according to the manufactur- filters, meters, etc. Describe microphone cables specifically.
er’s recommendations. Using the same microphone positions Record the calibrator model number and serial number, output
as in 10.1, take sound pressure level readings from each of the frequency and calibrated level. Record any other pertinent
microphone positions. Calculate the average sound pressure equipment information.
level by using the Eq 1 in 10.1. 12.4 Geometry—Record the source location point and the
10.3 Using the same test vacuum cleaner repeat 10.1 and microphone positions.
10.2 two additional times for a total of three test runs. 12.5 Vacuum Cleaner—Record the manufacturer, model
10.4 Calculate the value of A-weighted sound power level name and number, and unit serial number.
for each sound pressure level recorded according to the 12.6 Environment—Record the temperature, relative hu-
“comparison method” in Section 11. midity, and barometric pressure.
10.5 The sound power value for each individual test unit is 12.7 Calibration Check—Record the actual readout level
the average of three test runs meeting the repeatability state- with the calibrator on the microphone, both at the beginning
ments in Section 14. and end of the measurement period, to the nearest 0.1 dB, or as

3
F 1334
closely as the instrumentation permits. 14.1.4.1 The expected standard deviation of repeatability of
12.8 Ambient Sound Pressure Level—Record the ambient the calculated sound power level results within a laboratory, Sr,
overall (A-weighted) sound pressure levels at each of the has been found to be 0.7 dBA.
microphone locations to the nearest 0.1 dB. 14.1.4.2 The 95 % repeatability limit within a laboratory, r,
12.9 Reference Sound Source—Record the overall has been found to be 1.9 dBA, where r 5 1.96 2 2 .
A-weighted sound pressure levels at each of the microphone =2 (Sr).
positions to the nearest 0.1 dB. Include a copy of the sound 14.1.4.3 With 95 % confidence, it can be stated that within
power data calibration sheet as supplied from the source a laboratory a set of calculated sound power level results
manufacturer. derived from testing a unit should be considered suspect if the
12.10 Vacuum Cleaner—Record the overall A-weighted difference between any two of the three values is greater than
sound pressure levels at each of the microphone locations to the respective value of the repeatability limit, r.
the nearest 0.1 dB.
14.1.4.4 If the absolute value of the difference of any pair of
12.11 Record any other pertinent data or comments.
calculated sound power level results from three test runs
13. Test Report performed within a single laboratory is not equal to or less than
13.1 Report the following information: the respective repeatability limit, r, that set of test results shall
13.1.1 A description of the test samples used and the means be discarded.
used to distinguish them from other similar specimens (make, 14.1.5 Reproducibility (Multiday Testing and Single Opera-
model, serial number, manufacturing date), tor within Multilaboratories)—The ability to repeat the test
13.1.2 Approximate size and weight of the models tested within multiple laboratories.
and whether an operator was present during the sound level 14.1.5.1 The expected standard deviation of reproducibility
measurements, and of the average of a set of calculated sound power level results
13.1.3 Average A-weighed sound power level (calculated) between multiple laboratories, SR, has been found to be 1.2
shall be reported to the nearest decibel. dBA.
14.1.5.2 The 95 % reproducibility limit within a laboratory,
14. Precision and Bias 8
R, has been found to be 3.4 dBA, where R 5 1.96 =2 (SR).
14.1 Precision—The following precision statements are
14.1.5.3 With 95 % confidence, it can be stated that the
based on interlaboratory tests involving ten laboratories and
average of the calculated sound power level results from a set
three test units:
of three test runs performed in one laboratory, as compared to
14.1.1 The statistics have been calculated as recommended
a second laboratory, should be considered suspect if the
in Practice E 691.
difference between those two values is greater than the respec-
14.1.2 The following statements regarding repeatability
tive value of the reproducibility limit, R.
limit and reproducibility limit are used as directed in Practice
E 177. 14.1.5.4 If the absolute value of the difference between the
14.1.3 The standard deviations of repeatability and repro- average of the calculated sound power level results from the
ducibility of the calculated sound power level results have been two laboratories is not equal to or less than the respective
derived from ten sets of data, where each set of three test runs reproducibility limit, R, the set of results from both laboratories
has been performed by a single analyst within each of the ten shall be discarded.
laboratories on multidays using the test unit which produced 14.2 Bias—No justifiable statement can be made on the bias
the largest value of average sound power level of 89.4 dBA. of the method to measure and calculate the A-weighted sound
14.1.4 Repeatability (Single Operator and Laboratory; power of household vacuum cleaners since the true value of the
Multiday Testing)—The ability of a single analyst to repeat the characteristics cannot be established by an acceptable refereed
test within a single laboratory. method.

8
15. Keywords
Supporting data are available from ASTM Headquarters, Request RR: F11-
1011. 15.1 A-weighted; sound power level; vacuum cleaners

4
F 1334

ANNEX

(Mandatory Information)

A1. DETERMINATION OF THE POPULATION MEAN HAVING A90 % CONFIDENCE INTERVAL9

A1.1 Theory: TABLE A1.1 Percentiles of the t Distribution1010

A1.1.1 The most common and ordinarily the best estimate df t0.95

of the population mean, µ, is simply the arithmetic mean, X̄, of 1 6.314


2 2.92
the individual scores (measurements) of the units comprising a 3 2.353
sample taken from the population. The average score of these 4 2.132
units will seldom be exactly the same as the population mean; 5 2.015
6 1.943
however, it is expected to be fairly close so that in using the 7 1.895
following procedure it can be stated with 90 % confidence that 8 1.86
the true mean of the population, µ, lies within 62 dBA of the 9 1.833
10 1.812
calculated mean, X̄, of the sample taken from the population as 11 1.796
stated in Section 9. 12 1.782
A1.1.2 The following procedure provides a confidence in- 13 1.771
14 1.761
terval about the sample mean which is expected to bracket µ, 15 1.753
the true population mean, 100(1 − a) % of the time where a is
the chance of being wrong. Therefore, 1 − a is the probability
or level of confidence of being correct.
A1.1.3 The desired level of confidence is 1 − a 5 0.90 or A1.1.6 The following equations establish the upper and
90 % as stated in Section 11.1. Therefore a 5 0.10 or 10 %. lower limits of an interval centered about X̄ that will provide
A1.1.4 Compute the mean, X̄, and the standard deviation, s, the level of confidence required to assert that the true popula-
of the individual scores of the sample taken from the popula- tion mean lies within this interval:
tion: CIu 5 X̄ 1 ts/=n (A1.5)
1 n CIL 5 X̄ 2 ts/=n
X̄ 5 n ( Xi (A1.1)
i51

Œ
n n
where:
n ( X2i 2 ~i ( Xi!2 CI 5 Confidence Interval (U − upper limit; L − lower
i51 51 limit),
s5 (A1.2)
n~n 2 1! X̄ 5 mean score of the sample taken from population,
t 5 t statistic from Table A1.1 at 95 % confidence level,
where:
s 5 standard deviation of the sample taken from the
n 5 number of units tested, and
Xi 5 the value of the individual test unit score of the i8th population, and
n 5 number of units tested.
test unit. As will be seen in the procedural example to
A1.1.7 It is desired to assert with 90 % confidence that the
follow, this is the average value of the results from
true population mean, µ, lies within the interval, CIu to CIL,
three test runs performed on an individual test unit
centered about the sample mean, X̄. Therefore, the quantity ts/
with the resulting set of data meeting the repeatability
requirements of Section 14. =n shall be less than some value, A, as stated in 9.1, as
established as 2 dBA for all cases.
A1.1.5 Determine the value of the t statistic for n − 1
degrees of freedom, df, from Table A1.1 at a 95 % confidence NOTE A1.2—Generally, the value of A is stated as a percentage of the
level. 10 estimated population mean. As agreed to by ASTM Committee F-11 on
Vacuum Cleaners, in cooperation with Committee E-33 on Environmental
NOTE A1.1—The value of t is defined as t1−a/2 and is read as “t at 95 % Acoustics, the value of 2 dBA has been established.
confidence”.
A1.1.8 As n → `, ts/ =n → 0. As this relationship
t statistic 5 t12a/2 5 t0.95 (A1.3) indicates, a numerically smaller confidence interval may be
where: obtained by using a larger number of test units, n, for the
1 2 a/2 5 1 2 0.10/2 5 1 2 0.05 5 0.95, or 95 % (A1.4) sample. Therefore, when the standard deviation, s, of the
sample is large and the level of confidence is not reached after
9
Natrella, Mary Gibbons, Experimental Statistics, National Bureau of Standards
testing three units, a larger sample size, n, shall be used.
Handbook 91, U.S. Government Printing Office, Washington DC, 1963, pp. 2-1 to
2-3. A1.2 Procedure—A graphical flow chart for the following
10
Adapted by permission from Introduction to Statistical Analysis (2d ed.) by W. procedure is shown in Fig. A1.1.
J. Dixon and F. J. Massey, Jr., Copywright, 1957. McGraw-Hill Book Co., Inc.
Entries originally from Table III of Statistical Tables by R. A. Fisher and F. Yates, A1.2.1 Select three units from the population for testing as
1938, Oliver and Boyd, Ltd., London. the minimum sample size.

5
F 1334

FIG. A1.1 Testing Procedure Flowchart

A1.2.2 Obtain individual test unit scores by averaging the A1.3 Example—The following data is chosen to illustrate
results of three test runs performed on each of the three how the mean value of sound power, X̄, for the population of
individual test units. The data set resulting from the three test a vacuum cleaner model is derived. The calculated sound
runs performed on each individual test unit shall meet the power level test results from three test runs on each unit are
respective repeatability requirement found in Section 14. required to have a repeatability limit not exceeding 1.94 dBA
A1.2.3 Compute X̄ and s of the sample. as indicated in Section 14.
A1.2.4 A 5 2.0 dBA (for all cases).
A1.3.1 Select three test units from the vacuum cleaner
A1.2.5 Determine the statistic t for n − 1 degrees of freedom
model population. A minimum of three test runs shall be
from Table A1.1, where n 5 the number of test units.
performed using each test unit.
A1.2.6 Compute ts/ =n for the sample and compare it to
the value of A. A1.3.2 Test run scores for test unit No. 1:
A1.2.7 If the value of ts/ =n > A, an additional unit from Test Run No. 1 5 85.5 dBA
Test Run No. 2 5 83.4 dBA
the population shall be selected and tested, and the computa- Test Run No. 3 5 85.1 dBA
tions of A1.2.2-A1.2.6 repeated.
A1.2.8 If the value of ts/ =n < A, the desired 90 % A1.3.3 Maximum spread 5 85.5 − 83.4 5 2.1 dBA.
confidence level has been obtained. The value of the final X̄ A1.3.3.1 This value is greater than the repeatability limit
may be used as the best estimate of the sound power rating for required in Section 14. The results shall be discarded and three
the population. additional test runs performed.

6
F 1334
A1.3.4 Test run scores for test unit No. 1: s 5 1.656 dBA (A1.7)
Test Run No. 4 5 84.9 dBA A1.3.10 A 5 2.0 dBA
Test Run No. 5 5 85.1 dBA
Test Run No. 6 5 85.8 dBA A1.3.11 Degrees of freedom, n − 1 5 3 − 1 5 2
t0.95 statistic 5 2.920 (A1.8)
A1.3.5 Maximum spread 5 85.8 − 84.9 5 0.9 dBA.
A1.3.5.1 This value is less than the repeatability limit A1.3.12 tsl =n 5 2.920 (1.656)/ =3 5 2.792 dBA
requirement of Section 14. A1.3.13 2.792 > 2.0
A1.3.6 Unit No. 1 score 5 (84.9 + 85.1 + 85.8)/3 5 85.3 A1.3.13.1 The requirement that ts/ =n < A has not been
dBA. met because s is large. Therefore, an additional test unit from
the population shall be tested.
NOTE A1.3—If it is necessary to continue repeated test run sets A1.3.14 Score of test unit No. 4 5 86.3
(7,8,9 − 10,11,12—etc.) because the spread of data within a data set is not 85.27 1 88.53 1 87.41 1 86.34
less than the repeatability limit requirement stated in Section 14, there A1.3.15 X̄ 5 4 5
may be a problem with the test equipment, the execution of the test
procedure, or any of the other factors involved in the test procedure. 86.9 dBA
Consideration should be given to re-evaluating all aspects of the test A1.3.16
procedure for the cause(s).
=4@~85.27!2 1 ~88.53!2 1 ~87.41!2 1 ~86.342# (A1.9)
A1.3.7 A minimum of two additional test units must be 2
2 @85.27 1 88.53 1 87.41 1 86.34#
tested, each meeting the repeatability limit requirement. For s5
4~4 2 1!
this procedural example, assume those units met the repeat-
ability requirements and the individual unit scores are: s 5 1.401 dBA (A1.10)
Score of Test Unit No. 1 5 85.27 dBA A1.3.17 A 5 2.0 dBA
Score of Test Unit No. 2 5 88.53 dBA
Score of Test Unit No. 3 5 87.41 dBA
A1.3.18 Degrees of freedom, n − 1 5 4 − 1 5 3
t0.95 statistic 5 2.353 (A1.11)
85.27 1 88.53 1 87.41
A1.3.8 X̄ 5 3 5 87.1 dBA A1.3.19 tsl =n 5 2.353 (1.401)/ =4 5 1.772 dBA
A1.3.9 A1.3.20 1.772 < 2.0 (meets requirements)
A1.3.21 Thus, the value of X̄, 86.9 dBA, represents the
=3@~85.27!2 1 ~88.53!2 1 ~87.41!2 (A1.6) sound power level score for the vacuum cleaner model tested
2
2 @85.27 1 88.53 1 87.41# and may be used as the best estimate of the sound power level
s5
3~3 2 1! rating for the population mean.

The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.

This standard is copyrighted by ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States. Individual
reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at 610-832-9585
(phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website (http://www.astm.org).

You might also like