You are on page 1of 9

UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE

prototype pulse oximeter design


Luz Janeth Guio Figueroa (51785)
María Angélica Martínez Rojas (48580)
María Camila Pérez (61246)
Alejandro Fernández Rosas (62889)
Daniel Alejandro Morales

each beat, which means that the presence of


SUMMARY: A pulse oximeter is a device arterial pulse is necessary for that the device
that emits light with two wavelengths of 660 recognizes some signal. By comparing the
nm (red) and 940 nm (infrared) which are light absorbed by the pulsatile wave with
respectively characteristics of oxyhemoglobin respect to basal absorption, the percentage of
and reduced hemoglobin. Most of the light is oxyhemoglobin is calculated. Absorption is
absorbed by the connective tissue, skin, bone only measured on a pulse wave, which
and venous blood in a constant amount, minimizes the influence of tissues, veins and
producing a small increase of this absorption capillaries on the result. The pulse oximeter
in the arterial blood with each beat, which measures the saturation of oxygen in tissues,
means that the presence of an arterial pulse is has a transducer with two parts, a light emitter
necessary so that the device recognizes some and a photodetector, usually in the form of a
signal. By comparing the light absorbed clamp and which is usually placed on the
during the pulsatile wave with respect to basal finger, then it is expected to receive the
absorption, the percentage of oxyhemoglobin information on the screen: Oxygen saturation,
is calculated. Only net absorption is measured heart rate and pulse curve.
during a pulse wave, which minimizes the
influence of tissues, veins and capillaries on KEY WORDS: Pulse, light, saturation,
the result. The pulse oximeter measures the oxygen.
saturation of oxygen in tissues, has a
transducer with two parts, a light emitter and 1 NOMENCLATURE
a photodetector, usually in the shape of a clip
and that is usually placed on the finger, then it Traditional nomenclature:
is expected to receive the information on the
screen: Oxygen saturation, heart rate and SaO2 (saturation of hemoglobin in arterial
pulse curve. blood)
HbO2 (Oxihemoglobin)
PALABRAS CLAVE: Pulso, luz, Hb (Reduced hemoglobin)
saturación, oxigeno. PaO2 (partial pressure of oxygen in the arterial
blood)
ABSTRACT: A pulse oximeter is a O2 (Oxygen)
device that emits light with two wavelengths of
660 nm (red) and 940 nm (infrared) that are 2 INTRODUCTION
characteristic of oxyhemoglobin and reduced
hemoglobin. Most of the light is absorbed by This article presents a detailed analysis of
the connective tissue, skin, bone and blood in a pulse oximetry or pulse oximetry, being a
constant amount, producing a small increase device that consists of a thimble and a
of this absorption in the arterial blood with computer that graphs the saturation of oxygen

1
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

in the blood. These teams monitor non- was not used clinically. In 1964 Shaw
invasive oxygen saturation. Oxygen saturation assembled the first absolute measurement
(SaO2) is the% ratio between the concentration oximeter using eight wavelengths of light,
of oxygenated hemoglobin (HbO2) and marketed by Hewlett Packard, its use was
reduced hemoglobin (HbR). This parameter limited to lung functions and to sleep
denotes the amount of O2 that passes from the laboratories, due to its cost and size. The pulse
alveoli to the blood and dissolves is the tissues oximetry was developed in 1974, by Takuo
and body fluids. The oximeters measure the Aoyagi and Michio Kishi, bioengineering of
O2 assimilation of the patient non-invasively. Nihon Kohden using the ratio of red to infrared
light absorption of pulsating components at the
3 METHODOLOGY measurement site. A surgeon, Susumu
Nakajima and his associates first tested the
3.1 PULSIOXIMETRO – device on patients, which was reported in
PULSE OXIMETRE 1975. This device was marketed by Biox in
1981 and Nellcor in 1983. Biox was founded
STORY in 1979 and introduced the first pulse oximeter
in 1981. Biox initially focused on respiratory
In 1930 it begins to investigate on the care, but when it discovered that its devices
saturation of oxygen through light. In the were being used in operating rooms to monitor
Second World War the interest is renewed due oxygen levels, Biox expanded its marketing
to the hypoxia that the pilots presented. In resources to focus on operating rooms at the
1960, the first oximeter was designed and end of 1982. A competitor, Nellcor (now part
marketed in 1970. In 1935 Matthes developed of Covidien, Ltd.), began to rival Biox by the
the first wavelength device to measure oxygen market of operating rooms in 1983. Before the
saturation with red and green filters, which introduction of this device, the oxygenation of
were then changed to red and infrared filters. a patient could only be measured by means of
This was the first device for the measurement gases in the arterial blood, a single point of
of oxygen saturation. measurement, which takes a few minutes of
processing in a laboratory (In the absence of
oxygenation, brain damage begins in 5
minutes, with brain death after terrier within
another 10 to 15 minutes). In 1987, the
standard of care for the administration of
general anesthesia in the United States
included pulse oximetry. The use of the pulse
oximeter spread rapidly in the hospital, first in
the operating room and later in the recovery
rooms and intensive care units. Pulse oximetry
is of great value in the unit of neonates where
Fig. 1 Pulsioxímetro de Mattes-1940 patients require the continuous determination
of the level of oxygenation, as they can have
In 1949 Wood added a pressure capsule to side effects due to excess oxygen.
squeeze the blood out of the ear and thus get
zeroed in an effort to obtain absolute oxygen
saturation when the blood was readmitted. The
concept was similar to the current pulse
oximeter but was difficult to implement due to
unstable photocells and light sources. Method

2
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

wavelength, photo detectors and


microprocessors. The oximeter sensor is
usually placed on the fingers or toes. There are
oximeters that you can have at home, which
are small, easy to use and work with batteries
or are sometimes rechargeable. If you have any
illness that requires the use of these
instruments you can buy an oximeter to be
Fig. 2 Pulsioxímetro OF Nellcor-1993 aware of oxygen saturation levels, according to
the indications given by the doctor. When you
1. ¿ what is the pulse pulse oximeter? use it at home you should be calm or ask the
same person to put the oximeter, then place the
It is a non-invasive method that allows to oximeter, preferably on the index finger and
determine the percentage of oxygen saturation wait for the device to throw the results, on the
of the hemoglobin in a patient's blood with the screen, a heart it will indicate the pulsations of
help of photoelectric methods. the person and the abbreviations Spo2 will
To perform this technique, the pulse oximeter indicate the level of oxygen saturation.
is placed on a part of the body that is relatively
translucent and has a good blood flow, for 2. ¿ what is its characteristics?
example the fingers of the hand or the foot or
the earlobe. The pulse oximeter emits lights A pulse oximeter is a particularly convenient
with wavelengths, red and infrared that pass and non-invasive measuring instrument.
sequentially from an emitter to a detector Normally, it has a pair of small light-emitting
photo through the patient. The absorbance of diodes (LEDs) of a photodiode (A photo-diode
each wavelength caused by arterial blood is a semiconductor constructed with a PN
(pulsatile component) is measured, excluding junction.) The fundamental structure of the
venous blood, skin, bones, muscle, fat. With commonly referred to as semiconductor
these data it will be possible to calculate the electronic components, mainly diodes, is called
oxygen saturation in blood. PN junction. and transistors), sensitive to the
incidence of visible or infrared light) through a
translucent portion of the patient's body,
usually a finger or the lobe of an ear. One of
the LEDs is red, with a wavelength of 660 nm,
and the other is in the infrared, 905, 910, or
940 nm.

Fig. 3 Hemoglobin in the blood

Pulse oximeters perform a spectrophotometric


evaluation of the oxygenation of hemoglobin,
by measuring the light transmitted through a
capillary bed, in other words, the oximeter
measures the changes in light absorption that
result from the pulsations of the blood. The
oximeters have a detection system consisting
of light-emitting diodes(LED) a single Fig. 4 parts of the pulse oximeter

3
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

3. types of pulse oximeters Fig. 6 portable pulse oximeter

There are two different methods by which the wrist pulse oximeter: It also provides an
lights can be transmitted and read through the accurate reading and the sensor is placed
oximeter. Despite the differences that may around the wrist, as if it were a watch; It is
exist between these two methods, there is still quite useful for people who are not
no evidence that one is more effective than the comfortable with a sensor on their finger all
other. the time.
The different types of reflective technology
presented in pulse oximeter models allow
oxygen saturation readings to be taken from
other parts of the body, which becomes useful
in some situations. In addition to the
differences in the way light is detected, they
may be different in size, shape and certain
uses according to the objective of the oximeter. Fig. 7 Wrist pulse oximeter
finger pulse oximeter: It is the type of pulse
oximeter most used in homes since they
provide reading of oxygen saturation and Table pulse oximeter: Se usa frecuentemente
pulse, with only placing them on the tip of any en hospitales y laboratorios para estudiar a los
finger. It usually works with batteries and is pacientes en estudios de conducta del sueño.
very practical, as it can be carried around De igual forma utilizan un sensor que se coloca
easily. en el dedo, pero el monitor es mucho más
grande que el convencional o portátil ya que
ofrece información adicional a los médicos.

Fig. 5 Finger pulse oximeter

Portable pulse oximeter: It is usually a more


sophisticated equipment and its use is more
common in hospitals. Since the device is
formed by a sensor that is placed on the finger Fig. 8 Table pulse oximeter
and that is connected to a portable device that
is the one that throws the results of the search.
4. ¿how does it work?
The pulse oximeter is based on the fact that the
color of the blood varies depending on how
saturated oxygen it is, due to the optical
properties of the hemoglobin group of the
hemoglobin molecule. When the hemoglobin
molecule releases oxygen.

4
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

pink, acquiring a more bluish tone and letting


less red light through. Thus, the pulse oximeter
determines the oxygen saturation by
spectrophotometrically measuring the "degree"
of blue of the arterial blood and expresses this
"blue" in terms of saturation. Since the amount
of oxyhemoglobin is related to the red
coloration of the blood, it is stronger how
much more oxyhemoglobin contains blood,
and more tenuous how much less
oxyhemoglobin is present. Because the
absorption of light from tissues and from
venous blood is constant, any change in the
absorption of light, between a given time and a
subsequent one, is due exclusively to arterial
blood. The pulse oximeters measure the Tabla 1 saturation, pressure of oxygen in the
relationship, in a time interval, between the blood
absorption differences of the red and infrared
lights. This relationship is directly linked to the Clinical interpretation: On the contrary,
saturation of oxyhemoglobin. above 95%, large increases in PaO2 do not
imply significant increases in oxygen
5. ¿What is it for? saturation. PaO2 (mmHg) 100% 677 100 95%
80 90% 60 80% Below these values, small
Pulse oximetry can allow us to assess whether decreases in PaO2 cause significant
oxygen levels (or oxygen saturation) in the saturations. O2 Saturation 98.4% A saturation
blood are adequate in various circumstances of 90%, corresponds to PaO2 60 mm Hg. 48
such as in surgery, other procedures involving 73% 40 60% 30 50%.
sedation (for example, the bronchoscope), Saturation porcentaje:
supplemental oxygen adjustment as necessary, The critical point that should give the alarm
the effectiveness of medications for the lungs signal is that of saturations lower than 95%
and the patient's tolerance to higher levels of (lower than 90 or 92% when there is previous
activity. Other reasons may include, but are not chronic lung disease) these patients should
limited to, the following: Mechanical receive immediate treatment. PaO2 (mmHg)%
ventilation: use of a respirator to sustain Saturation> 95% No immediate action. 95-
breathing sleep apnea: Periods of interruption 90% Immediate treatment and monitoring.
of breathing during sleep, medical problems, According to the answer, assess referral to the
such as myocardial infarction, congestive heart hospital. Patients with chronic respiratory
failure, anemia, lung cancer, asthma or disease tolerate saturations well around these
pneumonia. values. <90% Severe ill. Severe hypoxia
Oxygen therapy + treatment and transfer to the
hospital. <80% Assess intubation and
mechanical ventilation. In children with <92%:
Refer to the hospital even if they improve with
initial maneuvers, since their response to
treatments is more uncertain.

5
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

¿How is it used?
Pulse oximetry can be done on an outpatient
basis or as part of your stay in a hospital. The
procedures may vary depending on your
condition and your doctor's practices. First,
information should be obtained on the correct
use of each model, and if it is necessary to
know how to adapt the needs we have to the
correct model, since in the market there are Fig. 9 Translucent body parts in adults
many different models with a wide range of
work possibilities through different programs. Children: (less than 20 Kg. Of weight)
Remove nail paints in the case of using Average portion of the foot on the big toe or on
thimble sensors. A clamp-like device called a the thumb. If the patient has poor peripheral
probe (works as a clothes pin but does not circulation, try on the lobe of the ear or even
pinch) will be placed on the finger or lobe of on the nose.
the ear. Alternatively, an adhesive probe may
be placed on your forehead or finger. The
patient will be explained what the
measurement consists of, insisting on the need
to move the finger minimally and not move the
sensor. The probe can be left in place for
continuous monitoring or can be used to obtain
a single reading. Unless continuous monitoring
is done, the probe will be removed after the
test. Carry out the measurement far from an
important light source, spotlights, etc. Clip
sensors should not compress excessively, as Fig. 10 Translucent body parts in children
they could alter the measurement. You have to
choose the device according to the size of the 1. Measurement Range
patient. There are sensors for children and • Measurement range of Oxygen Artificial
adults. Use a very deep place that completely Saturation (%): Between 0% and 100%
covers the sensor detector. • Resolution of the O2 Sat Arterial
measurement: 1%
Adults: Heart or ring fingers of the non- • Accuracy of O2 Sat Artery measurement: 2%
dominant hand. If you have a canalized radial in levels between 70% and 100% Saturation
artery, put the sensor there whenever the • Pulse measurement range: Between 30 bpm
reading is correct (thus we only immobilize and 250 bpm (beats per minute)
one hand). It can be placed on the foot (finger • Resolution of the Pulse Measurement: 1 beat
next to the fat one of the foot), but it is per minute
necessary to verify that it does not have bad • Accuracy of the pulse measurement: 2 lpm or
circulation in the lower extremities. 2% of the measurement

Benefits or Advantages of the pulse oximeter


and Disadvantages
Also the use of a pulse oximeter or oximeter is
not only for people who suffer from some kind
of disease but for those who also want to
improve their physical condition, for example

6
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

for people who practice sports, using a finger essential For the device to work, there must be
oximeter while training, can help them to pulsatile flow. It can be improved with heat,
create an exercise routine, according to the massage, local vasodilator therapy, removing
strength of your body. tight clothing, not placing the tension cuff on
Advantages: The pulse oximeter is a very the same side as the transducer.
useful method for monitoring, but it does not 6. The venous pulse: right heart failure or
replace the information provided by arterial tricuspid regurgitation.
blood gas, which is much more sensitive and 7. The increase of the venous pulse can alter
complete. the reading, the device must be placed above
Advantages with respect to arterial blood the heart.
gases: 8. Fetal hemoglobin does not interfere.
• Provides instant and continuous monitoring. 9. Obstacles to the absorption of light: nail
• Non-invasive or painful. polish (remove with acetone), pigmentation of
• Easy to use. the skin (use the 4th finger or the ear lobe).
• Economically affordable. Equipment Maintenance
• Portable and manageable. • Do not immerse it in water or in cleaning
Disadvantages: Disadvantages with respect to solutions or sterilize it.
arterial blood gases: • Monitor the battery level
• Does not provide data on patient ventilation. • If the sensor is for multiple use of patients, it
• Does not provide data on pH. must be cleaned between applications
• Critically ill patients usually have poor according to local regulations and
peripheral perfusion. manufacturer's recommendations.
• In case of excessive ambient light
1. Limitations of Pulsioximetry (phototherapy, xenon, infrared, fluorescent,
Alterations of hemoglobin: operating theater or fiber optic), protect the
• Dyes and pigments in the reading area sensor with an opaque material to avoid falsely
(painted nails). • Sources of external light. high values.
• Peripheral hypoperfusion. • If there is poor quality of the signal or
• Anemia. artifacts due to patient movements, relocate the
• Increased venous pulse. sensor to an area with less mobility.
• Does not detect hyperoxia. • The optical interference produces an
• It does not detect hyperventilation. inaccurate measurement in the absence of a
Current devices are very reliable, when the correct opposition between the two diodes and
patient has saturations greater than 80%. The part of the light does not pass through the
situations that can give rise to erroneous sensor, the curve is correct but the value
readings are: obtained.
1. Severe anemia: Hemoglobin must be less
than 5 mg / dl to cause false readings. Fix it:
2. Interference with other electrical devices. • Selecting the appropriate sensor for the
3. Intravenous contrasts can interfere if they patient.
absorb light of a wavelength similar to that of • Relocating the sensor properly.
hemoglobin.
4. Intense ambient light: xenon, infrared, Preventive Maintenance: After the First World
fluorescent. War, it was stated that maintenance not only
5. Poor peripheral perfusion due to had to correct the breakdowns, but had to
environmental cold, decreased body anticipate them, guaranteeing the proper
temperature, hypotension, vasoconstriction. It functioning of the machine.
is the most frequent cause of error since it is

7
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

avoiding the delay caused by breakdowns and


their consequences, giving rise to what was • The pulse oximeter provides a non-invasive,
called preventive maintenance, which is what easy to use, very simple technique to evaluate
is done, before the equipment comes into the gas exchange function. Within its many
operation, in eviction of subsequent limitations, an estimate of the degree of blood
breakdowns, guaranteeing a period of reliable oxygenation can be obtained quickly with an
use. acceptable degree of accuracy for clinical
practice. It is essential that medical personnel
3.2 mounting on the breadboard. understand the use and limitations of the pulse
oximeter and that all medical personnel have
been given the appropriate training for a
correct application of the pulse oximeter in
practice.

• Los actuales resultados de las simulaciones y


de las medidas clínicas indican que para que
este método no invasor mida las
concentraciones de la oxihemoglobina (O2
Hb), de la hemoglobina reducida (HHb), de la
metahemoglobina (MetHb), y del
carboxihemoglobina (COHb) hace falta utilizar
FIG. 11 MOUNTING ON THE BREADBOARD
al menos nueve longitudes de onda en el rango
a partir del 600 nm a 940 nm. Así se consigue
una exactitud cercana 1%.

3.3 BlOCKS DIAGRAM 5 REFERENCIAS

[1] John W. Severinghaus, MD,and Yoshiyuki


Honda, MD: HISTORY OF BLOOD
GAS ANALYSIS. VII. PULSE
OXlMETRY"History of Pulse Oximetry"
[2] Malloy, Daniel (9 de enero de 2008).
«Medicare may allow sleep apnea diagnoses
from home». Pittsburgh Post-Gazette..
[3] Kirk y Bistner. Urgencias en veterinaria.
Fig. 3. BLOCKS DIAGRAM Procedimientos y terapéutica. Elsevier. 8ª
edición. 2007. Disponible en:
1.5
1.1
4 CONCLUSIONS 1.2 Libros:
1.6
1.3 Luis
1.7 Jiménez -Murillo,FJ Montero. Medicina
de urgencias y emergencias: guía diagnóstica y
• The pulse oximeter is a device that measures
the saturation of blood, so it is very important
1.4 protocolos
1.8 ACONDICIONAMIENTO
de actuación. Elsevier. 2ª edición.
to perform this test because it provides 1999. Disponible en:
information about the health of patients, it is http://books.google.es/books?
generally used in hospitals and at home, it can id=_0XjCi_q5EgC&pg=PA58&dq=pulsioxime
help a patient who I would have given him an tr
asthma attack %C3%ADa&hl=es&sa=X&ei=S69pT_bjHNG

8
UNIVERSIDAD ECCI TECNOLOGÍA BIOMEDICA I III SEMESTRE
.

AhQey9fywCg&ved=0CDUQ6AEwAA#v=on
epage&q=pulsioximetr%C3%ADa&f=false
Kirk y Bistner. Urgencias en veterinaria.
Procedimientos y terapéutica. Elsevier. 8ª
edición. 2007. Disponible en:
http://books.google.es/books?id=r-
ZyhTIaHXMC&pg=PA25&dq=pulsioximetr
%C3%ADa&hl=es&sa=X&ei=S69pT_bjHNG
AhQey9fywCg&ved=0CDwQ6AEwAQ#v=on
epage&q=pulsioximetr%C3%ADa&f=false

You might also like