Professional Documents
Culture Documents
▪ Summary
In order to provide convenience to our customers, the maintenance manual of the electrolyzer
has been recently revised based on the results of repeated experiments and tests over the
years. The revised content is summarized and delivered to our valued customers, so please
refer to the following explanation.
1/3
Doc No : HGS-BWTS-SL-20-0001
The Notice of Revised Maintenance Manual for an Electrolyzer
Before
Acid Cleaning
After
Acid Cleaning
2/3
Doc No : HGS-BWTS-SL-20-0001
The Notice of Revised Maintenance Manual for an Electrolyzer
※ Since HiBallast will have been continuously improved, if you have any suggestions or other
useful information with operation and maintenance, please don’t hesitate to contact Green
Solution Team.
[End]
3/3
Doc No : HGS-BWTS-SL-20-0001
BALLAST WATER TREATMENT SYSTEM
MAINTENANCE MANUAL
OF HIBALLASTTM SYSTEM
Page
DOCUMENT NO. HiB-M01002
127
RESPONSIBLE PARTY Hyundai Heavy Industries Co., Ltd.
Revision Sheet
TABLE OF CONTENTS
1. General ...................................................................................................................................................................................................... 10
3.2.2. EM (Electrolyzer module) cleaning procedure by using cleaning kit (optional) ...................................... 25
3.2.4. EM (Electrolyzer module) cleaning procedure by manual for each plate ...................................................... 40
3.2.5. EM (Electrolyzer module) cleaning procedure by manual for internal module assembly ..................... 43
11.3. Calibration............................................................................................................................................................................................... 92
Appendix A .......................................................................................................................................................................................................1
Appendix B .......................................................................................................................................................................................................1
Appendix C .......................................................................................................................................................................................................1
LIST OF FIGURES
LIST OF TABLES
REFERENCE MANUAL
ABBREVIATION
Abbreviations Descriptions
AODD Air Operated Double Diaphragm
AL Alarm
AMS Alarm Monitoring System of a vessel
APT After Peak Tank
BS Existing Ballast System of the vessel
BWP Ballast water Pump
BWTS Ballast Water Treatment System
CC Control Cabinet
CCR Cargo Control Room
DP Deferential Pressure
ECR Engine Control Room
EM Electrolysis Module
EMC Electromagnetic Compatibility
ESD Emergency Shut-down
EU Electrolysis Unit
FIFRA The Federal Insecticide, Fungicide, and Rodenticide Act
FU Filter Unit
GPS Global Positioning System
GS Gas Separator
GSP General Service Pump
HAZOP Hazard and Operability study
HMI Human-Machine Interface
IAS Integrated Automation System
IMO International Maritime Organization
IP International Protection
J/B Junction Box
LOP Local Operating Panel
MCC Motor Control Cabinet
Abbreviations Descriptions
MMO Mixed Metal Oxide
MTBF Mean Time Between Failure
NU Neutralizing Unit
PC Personal Computer
PLC Programmable Logic Controller
PMS Power Management System of a vessel
PSU Practical Salinity Unit
RF Rectifier
RH Relative Humidity
SD Shut-Down
SKID The place electrolyzer unit and gas separator installed
SCU System Control Unit
ST Strainer
SWFU Seawater Feeding Unit
TBD To be determined
TRO Total Residual Oxidant
TRC Total Rated Capacity(Total treatment Rated Capacity)
TSU TRO Sensor Unit
USCG United State Coast Guard
WBP, BWP Water Ballast Pump
BO Before Operation
AO After Operation
AR As Required
1. General
To ensure the best performance of equipment, the following inspection and maintenance works are
recommended. The following table summarizes inspection schedule and maintenance period. The more details
about installation, inspection, maintenance and spare part can be found at each unit’s document.
⚫ The personnel for installation, commissioning, operation and maintenance of the HiBallast system and
its equipment must read and understand this manual including the health and safety section.
Depending on the qualification of the personnel, additional training might be necessary.
⚫ The installation of the HiBallast system has to be carried out in accordance with local regulations and
accepted code of good practice.
⚫ The installation, commissioning, operation and maintenance of the HiBallast system might only be
carried out by authorized and trained personnel.
⚫ Strictly comply with all relevant rules for accident prevention and local health and safety regulations.
⚫ Check all relevant safety procedures before you switch on the HiBallast system.
⚫ HHI accepts no responsibility for any damage arising from incorrect installation, incorrect use,
incorrect maintenance or own modifications of the HiBallast system.
⚫ Always disconnect the HiBallast system from the power supply before disassembling and
disconnecting.
⚫ All maintenance of the HiBallast system must be logged in the ship’s maintenance log.
To carry out any maintenance or calibration work described in this manual and reference manual, BWTS
controller to be changed to maintenance or calibration mode in compliance with USCG 46 CFR § 162.060-20
(e) (2). Please refer to operation manual [7.6 Maintenance status indication function] for more information.
⚫ Hazard warning
⚫ Hazard warning
2. Maintenance schedule
Follow maintenance schedule is to secure performance, safety and durability of the HiBallast system and its components.
Maintenance work can be classified as follows
⚫ Inspection
⚫ Periodic functional test
⚫ Maintenance: replacement or repair
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
L/B : 1hr
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
insulation guides for power
feeders.
2) Check cable connection
Electrolyzer Internal Inspection See 3.2.3 L/B : 4hr x
●
2M
Electrolyzer module cleaning See3.2.2, 3.2.4, 3.2.5 Every 60hrs
ballasting
operation
time or 3
● ● PIC month
whichever is
earlier
L/B : See
3.2.3 ~ 3.2.5
Replace insulation guide and Periodic replacement after L/B : 2hr
electrode end plate nut o-ring inspection ●
(if any)
Gas separator vent system See 3.4 L/B : 4hr x
● PIC
inspection 2M
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
Check H2 vent pipe Check if vent pipe and head Refer to
is blocked or not troubleshooti
ng manual
●
3.1 (3)
L/B : 1hr
Seawater Feed Cleaning the filter inside of Remove particles on the L/B : 2hr
●
Unit strainer surface of filter
Check the seawater feed pump Check any noise during L/B : 0.5hr
●
operation
Check seal, bearing, sleeve, shaft, Check any leakage during L/B : 0.5hr
●
and impeller operation
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
Check piping and all valves for Check any leakage during L/B : 0.5hr
●
leakage operation
Neutralization Fill up tank, if less than half L/B : 3hr x
Check the tank level ●
Unit of full 2M
See the output value of L/B : 0.5hr
Check the injection pump ●
pump(s)
Check the injection pump Check any leakage ● L/B : 0.5hr
Check the lubricant oil First 1year, then every 2years ● PIC L/B : 0.5hr
Check piping and all valves Check any leakage during L/B : 1hr
●
operation
Check injection nozzle Check and clean it if L/B : 1hr
● ●
necessary
Check back-pressure valves and Check and clean it if L/B : 1hr
check valves especially for necessary ●
blockage
Filter Unit Back-flushing filter element L/B : 8hr x
Check any visible damage(s) ● PIC
2M
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
Check any noise during L/B : 0.5hr
Check any noise of gear unit 2W ●
operation
Check any leakage during L/B : 0.5hr
Check any leakage of gear unit ●
operation
Check the vent valve Cleaning vent valve outlet ● L/B : 1hr
L/B : 2hr x
Check fixing bolts for tightness Annually ●
2M
Seals Leakage, Visual check ● L/B : 0.5hr
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
filling recommended if sailing
period is long to protect filter
element from corrosion
Electric Cabinet Check all cabinet doors are closed ● L/B : 1hr
FREQUENCY BY WHOM
COMPONENTS ACTION INSTRUCTION NOTES
D W M 6M Y BO AO AR (LEVEL)
clean the dirty one for future
use.
Replacement of pump tubes and L/B : 1hr
Replace if necessary ● PIC
check valves
See reference
Check any clogging and
Check any tubes and check valves ● PIC manual
leakage
L/B : 0.5hr
Check compressed air supply Only for
Compressed air & purge
and purge controller normal ● explosion
controller check
operation proof type
Flow Meter L/B : 6hr x
Check flow meter electrodes Check and clean if necessary ●
2M
Cables Check all cable connections L/B : 2hr
Check the cable connection status ●
in visual
Note
Above all maintenance works should be performed and managed by crew. Furthermore, it is recommended that some items marked by PIC are controlled by qualified
person such as chief officer or chief engineer. If any problem or defective part is found or replacement is required, please refer to HiBallast troubleshooting manual and
reference manual for more information. If further information or support is required, please contact manufacturer.
AUTION! The warranty does not apply if the product (a) has been altered, except by the maker or its authorized representative, (b) has not been installed,
operated, repaired, or maintained in accordance with instructions supplied by the maker, (c) has been subjected to abnormal physical or
electrical stress, abnormal environmental conditions, misuse, negligence, or accident or (d) has provided for training purposes.
Especially, the following works must be conducted in accordance with the above PMS for stable operation, equipment protection and safety.
- Electrolyzer visual inspection and cleaning
- H2 gas detector calibration
- Replacement of the TRO reagent
- Normal operation status, leakage and connections shall be checked during operation after any maintenance work.
A periodic calibration should be carried out by user to secure performance, safety and durability of the
HiBallast system.
2.2. Troubleshooting
To assess system functionality, specific parameters should be monitored periodically and biological indicator
compliance monitoring sampling of ballast water must be conducted. The required parameters to be monitored
and guidance for sampling are contained in Appendix B
3. Electrolysis Unit
If the BWTS was stopped by emergency shutdown, the remaining water inside of
electrolysis unit is solution containing hypochlorite ions. So if drain is required for
maintenance after ESD, manual flushing should be performed by crew before drain.
Othewise this drain water should be neutralized properly.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid inside EM. Protect your skin and wear
safety goggles.
Please make sure that the proper safety precautions are taken, when using the cleaning
agents according to the material safety data sheet. Also ensure that only trained and
instructed personnel is allowed to perform this cleaning-especially pay attention to the
mixing procedure and potential chemical reactions.
Procedure 1 Make flushing line-up from sea water feed unit to ballast tank by valves open
in manual.
3.2. Electrolyzer
Electrolyzer is a core equipment of EU, which produces Sodium Hypochlorite through the electrolyzation
of seawater.
3.2.1. Configuration
For instance, the HiBallast system is composed of several electrolyzer modules connected in a row as
follows.
Figure. 1 Electrolyzer
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your personal protective equipment.
Do not expose your eyes and skin to the liquid inside EM. Protect your skin and wear
safety goggles.
Please make sure that the proper safety precautions are taken, when using the cleaning
agents according to the material safety data sheet. Also ensure that only trained and
instructed personnel is allowed to perform this cleaning-especially pay attention to the
mixing procedure and potential chemical reactions.
Limitation This cleaning manual is only available when the electrolysis unit is equipped with
1. Cleaning kit connection nozzle
2. Electrolyzer isolation valve on the outlet of electolyzer train
3. Isolation valve between gas separator and electrolyzer
Parts and Electrolyzer modules, circulation pump, hose, flange connector, neutralization power
equipment (Sodium Bicarbonate), Fresh water, Drum
required
As an
alternative,
10% SAF
ACID can be
used.
5 PPE
How to make 10% SAF ACID as the acid solution for cleaning.
Required
neutralization
HiBallast Acid solution total volume Required neutralization
chemical
Model (L) chemical(kg)
≤ HiB-2000 320 13
≤ HiB-4000 480 20
≤ HiB-6000 660 27
≤ HiB-8000 860 35
≤ HiB-10000 1000 40
Procedure 1 Isolate EU unit and drain solution containing hypochlorite ions inside EU and other
parts.
⚫ Drain solution in EU
1) Open the drain valve (DV-1111)
2) After draining the solution, close the drain valve (DV-1111)
3 Connect the hose from the tank to the inlet/ outlet of the electrolyzer piping.
1) Remove the blind flange from the nozzles
2) Connect the valve with hoses
5 Make line-up.
In case that the EU has multiple electrolyzer module trains such as HiB-4000 or
Hib-6000, circulation cleaning should be done one by one by closing and opening
electrolyzer outlet manual valve. Below figure shows an example of electrolyzer
circulation line-up.
Connect hose to
7 After leakage check and before use acid chemical, all flange and hose connection to
be covered with anti-splash tape.
8
Wear the full PPE (Personal Protection Equipment) before use of chemical!
Start the circulation pump and pour the acid solution into the tank using barrel
pump or hand pump. Do not pour all acid into the tank at once. 10L per 1 min. is
Use of more than the recommended concentration may affect the life of
electrodes.
10 Take a small amount of the sample chemical through tank drain valve and bottle.
Do not pour the neutralization powder at once. Otherwise, too much bubble
will be generated by chemical reaction. Add 500g at a time.
14 Circulate the neutralized chemical by running circulation pump for 30 minutes per
electrolyzer train.
This chemical circulation should be also carried out one by one for all trains
15 Take small amount of the sample chemical through the tank drain valve.
18 Close the manual valve on the cleaning kit connection nozzle of EU.
20 Drain the remaining neutralized water to bilge tank by opening tank drain valve.
21 Disconnect cleaning kit from EU and place the blind flange on the cleaning kit
connection nozzle of EU
23 Check the final position of all manual valves in EU for normal operation.
24 Write the date on the record book after completing cleaning work. Please see
appendix C for the record book
1) After cleaning work, all manual valves for electrolyzer module outlet should be set as marked position to
ensure even flow of the modules. The required manual valve position is set by maker before delivery and correct
position can be checked by marking on the indicator of each manual valves.
2) All manual valves should be re-installed, if it is disassembled, to its original position just as it was.
Generally scales are produced by sea water electrolysis as a by-product and these
scales could be accumulated in the each electrolyzer modules inside as time goes by. If
the scales are not removed properly and accumulated too much, internal electrode
module can be damaged by over voltage and heating. So, periodic inspection &
cleaning should be carried out by crews in accordance with the maintenance manual.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations.
Wear your Personal Protective Equipment.
The liquid remaining in the EM may splash. Wear safety gloves for handling.
The liquid may cause eye and skin irritation. Keep channel covers closed at all times.
The HG may remain. Wear clean safety gloves.
4 Check inside of EM
If accumulated scale is founded, rake out it from inside of EM
5 Assemble the gasket and flange and tighten the bolts and nuts
8 Write the date on the record book after completing cleaning work. Please see
appendix C for the record book
Take care when handling Ti-electrode to prevent contamination with scales and
sediments stuck on electrode. This can lower the efficiency of EU performance.
So cleaning shall be required with special caution and care because of using acid
solution for cleaning.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid inside EM. Protect your skin and wear
safety goggles.
Please make sure that the proper safety precautions are taken, when using the cleaning
agents according to the material safety data sheet. Also ensure that only trained and
instructed personnel is allowed to perform this cleaning-especially pay attention to the
mixing procedure and potential chemical reactions.
Parts and * Required acid solution for cleaning is 3 %(vol.) of hydrochloric acid (HCl) or 10% SAF
equipment ACID
required * A bath for one (1) electrolyzer module: 0.5 m (W) x 1.5m (L) x 0.6 m (H) acid cleaning
bath.
Initial filling level should not be over 50% of bath volume considering overflow of acid
solution during cleaning.
Electrode plate
3 Pick out each electrode plate carefully from the electrolyzer module(s).
4 In order to clean the electrode plate(s), put them in a bath with the chosen
cleaning acid and leave them for 12hr (Max.) respectively until you see a
cleaning effect desired.
* Recommended immersion time: max. 12 hr
* Required acid solution for cleaning: ab. 3 % (vol.) hydrochloric acid (HCl)
or 10% SAF ACID (alternative)
Caution! Use of more than the recommended concentration may affect the
life of electrodes.
5 After cleaning, the elements are taken out from the bath, and you have to
ensure to clean off any remaining acid from the elements surface by suitable
method.
6 After cleaning, please assemble carefully the electrode plates in the correct
way. And then install flanges accordingly.
When the electrodes are reassembled in the electrolyzer after cleaning, the anode and cathode of the electrode
must be clearly identified and installed in their original position. If the anode and cathode are placed incorrectly,
the life of the electrode is seriously affected.
3.2.5. EM (Electrolyzer module) cleaning procedure by manual for internal module assembly
Take care when handling Ti-electrode to prevent contamination with scales and
sediments stuck on electrode. This can lower the efficiency of EU performance.
So cleaning shall be required with special caution and care because of using acid
solution for cleaning.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid inside EM. Protect your skin and wear
safety goggles.
Please make sure that the proper safety precautions are taken, when using the cleaning
agents according to the material safety data sheet. Also ensure that only trained and
instructed personnel is allowed to perform this cleaning-especially pay attention to the
mixing procedure and potential chemical reactions.
Parts and * Required acid solution for cleaning is ab. 3 % (vol.) hydrochloric acid (HCl) or 10% SAF
equipment ACID (alternative)
required * A bath for one (1) electrolyzer module: 0.5 m (W) x 1.5m (L) x 0.6 m (H) acid cleaning
bath.
Initial filling level should not be over 50% of bath volume considering overflow of acid
solution during cleaning.
Internal module
→ →
→ →
Clamp by using a
4 In order to clean the electrode plate(s), put them in a bath with the chosen cleaning
acid and leave them for 12hr (Max.) respectively until you see a cleaning effect
desired.
* Recommended immersion time: max. 12 hr
* Required acid solution for cleaning: ab. 3 % (vol.) hydrochloric acid (HCl) or
10% SAF ACID (alternative)
Use of more than the recommended concentration may affect the life of
electrodes.
5 After cleaning, take it out from the bath and clean inside of the module(s) by
flushing with fresh water. Then check inside of the module before assembly.
6 When assembly, please put the internal module in the correct way.
And then install flanges, power feeder and cables accordingly.
When the electrodes are reassembled in the electrolyzer after cleaning, the anode and cathode of the electrode
must be clearly identified and installed in their original position. If the anode and cathode are placed incorrectly,
the life of the electrode is seriously affected.
Take care when cleaning inside gas separator to prevent scales piling up inside GS.
This can cause downgrade of the efficiency of EU performance.
So cleaning shall be required with special caution and care because of acid liquid
remaining inside GS.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid inside GS. Protect your skin and wear
safety goggles. Ordinary glass or plastic is sufficient.
Procedure 1 Make sure that the electrolysis unit is stand-by (not in operation).
4 Fill the separator with sea water using sea water feed pump
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid inside GS. Protect your skin and wear
safety goggles. Ordinary glass or plastic is sufficient.
Frequency If required
3.5. Blower
Take care when checking the blower to able to cause the electric shock to the human.
Electric shock may happen by the malfunction of the motor.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the mixed HG around blower. Protect your skin
and wear safety goggles. Ordinary glass or plastic is sufficient.
Frequency Check if the blower needs checking when hearing abnormal noise.
Take care when checking the seawater feed pump, electric shock may happen by
malfunction.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid in the pipe line.
Purpose To maintain the seawater feed pump to keep the functionality of EU continuously
and safely.
Frequency Check if the seawater feed pump needs maintenance when hearing abnormal noise.
Procedure 1 Isolate the seawater feed pump using the manual valves.
2 Drain the seawater in the pipe line around seawater feed pump.
Take care when disassembling for cleaning the plate heat exchanger, the noise in
unlocking the bolts from the plates is no problem. Be careful hanging the plate
hanger to guide bar. (It can be solved by drive.)
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid in plate heat exchanger.
Procedure 1 Isolate plate heat exchanger using the manual valves in in/outlet of the plate
heat exchanger and on/off pneumatic valve in inlet of plate heat exchanger are
closed.
3 Unlock the middle part with spanner and then up and down
4 For example, if someone wants unlock the 8 bolted machine, he should unlock
upper 4 bolts firstly by 10mm individually, then unlock the 4 lower bolts by
10mm. Push the bolts to the fixed frame and pull out them from it. Move the
moving frame.
6 Pull the plate by the hanger and pick out the plate with lowing upper port
7 When the plates are firmly jammed, take apart it using drive or something like
that
Chemical cleaning
- Beside regular disassemble cleaning; it can be cleaned in assembled
situation as Fig.2.
- In cleaning brine, the brine should be expelled completely and then run
the system with pure water at normal speed: till it is clean enough
(Fig.3).
- In cleaning brine with hot water, follow the sequence above and loose
the bolts for gasket’s expansion 100plates/5min.
The flow rate of cleaning water should higher than driving situation.
- In chemical cleaning, the temperature and density of soda are not that
critical but in nitric case they are important. Nitric acid should not be
touched with anything, except stainless steel.
- If the PH of fluid is below than 5, it needs it needs cleaning and should
be changed to over 7.
4.3. Strainer
Take care when replacing the filter basket in strainer to prevent particles piling up on
the mesh screen of the filter basket and increase pressure on that. This can cause to
prevent seawater as the electrolysis source from inflowing to the EU.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Purpose To clean sediments and particles piling up on the mesh screen of the filter basket.
Procedure 1 Isolate strainer using the manual valves in in/outlet of strainer are closed.
6 Eliminate the particles and sediments using the fresh water in vessels.
5. Neutralization Unit
Take care when checking the NP to able to cause the electric shock to the human.
Electric shock may happen by the malfunction of the motor.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not expose your eyes and skin to the liquid remaining in NT. Protect your skin
and wear safety goggles. Ordinary glass or plastic is sufficient.
Disassembly
2 Drain the acid liquid in the pipe line around injection pump.
5 Loosen the joint flange of the suction and discharge parts, and take out the
valve (ball seat, ball guide, check ball).
6 If there is any abnormal scar or wear-out in the check ball or ball seat,
exchange it.
Assembly
2 Assemble the joint flange of the suction and discharge parts, and tighten the
valve.
■ Replacement of Diaphragm
Disassembly
2 Drain the acid liquid in the pipe line around injection pump.
Operate the motor at a low speed. When the stroke length is 100%, grab the
outer diameter of the diaphragm and turn it counter-clockwise to remove the
6
diaphragm from the slider. Since it is a 2-head type, the other diaphragm can
be disassembled in the same way.
Assembly
1 Operate the motor at a low speed. When the stroke length is 100%, grab the
outer diameter of the diaphragm and turn it clockwise to assemble the
diaphragm to the slider. Turn it until it is tightly fixed.
2 Since it is a 2-head type, the other diaphragm can be assembled in the same
way.
3 Tighten the pump head-fixing bolt evenly (30kgf.cm), diagonally so that there
is no gap between the pump head and the gear box flange.
It requires checking and filling the NT with NA before the system run especially de-
ballasting. It is available to be filled additional supplement of NA during operation if
necessary.
(If an uncertified NA solution is used, it may cause performance degradation in the
system)
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not inhale liquid. Protect your skin and wear safety gears. Ordinary glass or
plastic is sufficient. Keep away from incompatibles such as oxidizing agents, acid,
and alkalis. Keep container tightly closed. Keep container in a cool area.
*Refer the Material Safety Data Sheet (MSDS) for more information.
Purpose To fill the Neutralization Tank.
Parts and Neutralization agent solution and a hand pump (Spare part for NA).
equipment required One (1) drum: approx. 200 L (256kg)
5 Open the NA drum and prepare the hand pump in right position adequately.
8 Clean up around the neutralization unit, and close the manual valve of NT.
Validity date of 1 The validity date of product is NOT specified, but it is recommended to check
Purchase of product If additional purchase is required, please contact to HHI Global Network or
1
(NA) bwts@hhi.co.kr before leaving the current port.
6. Control Cabinet
1) Take care when handling stainless steel to prevent contamination with mild steel
materials, dust, shavings and the like as these small particles rust quickly and
discolour the surface of the stainless steel. This can destroy the protective oxide
film and render it liable to pitting corrosion.
2) Ensure that damaged devices must be cut off and disconnected from the power
source.
3) Before effecting any work or repairing on the device, there generally must be
executed a complete disconnection from power supply.
4) High operating voltage, firing impulse and high operating current!
5) Danger of life or severe personal injury through electric shock at improper
application!
6) Be sure the earth is ALWAYS connected. In accordance to the EN50178-
discharge current via the protective earth conductor-one of the measures must be
fulfilled related the EN50178.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
6.2. Description
The Control Cabinet (CC) is an enclosure that contains LOP (Local Operation Panel) and all PLC modules such
as CPU & I/O module. The PLC system is to control all actuators and to communicate with the rectifier. The
following figure shows CC.
1) Take care when handling stainless steel to prevent contamination with mild steel
materials, dust, shavings and the like as these small particles rust quickly and
discolour the surface of the stainless steel. This can destroy the protective oxide
film and render it liable to pitting corrosion.
2) Ensure that damaged devices must be cut off and disconnected from the power
source.
3) Before effecting any work or repairing on the device, there generally must be
executed a complete disconnection from power supply.
4) High operating voltage, firing impulse and high operating current!
5) Danger of life or severe personal injury through electric shock at improper
application!
6) Be sure the earth is ALWAYS connected. In accordance to the EN50178-
discharge current via the protective earth conductor-one of the measures must be
fulfilled related the EN50178.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
7.2. Description
The Motor Control Cabinet (MCC) is an enclosure that contains power supply module, several circuit breakers
and magnet contactors for motor on/off control. This panel communicates with the CC (Control Cabinet). The
following figure shows MCC.
Set value detecting over current for motor protection is listed in the table below
Rate(kW)
See the electric wiring diagram of “MCC-100”.
Set(kW)
Above set value can be adjusted by using setting dial of EOCR (Electrical Over-Current Relay) as figure below.
8. Rectifier (RF)
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
8.2. Description
The rectifier (RF) is an enclosure that contains transform, power switching devices, control board and roof
type cooling unit, which is used to supply power to the electrolyzer module.
The rectifier allows several operation types, local, remote and remote/local mode. The HiBallast system uses
normally remote operation mode of the rectifier by using Ethernet communication.
The rectifier has a protection function such as high voltage, high current, short circuit of output power, high
temperature of device, internal high temperature and so on.
8.3. HiPower
8.3.1. Troubleshooting
The warning lamp on the front panel of the rectifier blinks when an error occurs. The error has different
meaning according to lamp flickering interval.
“0.5sec interval”: Indicates over current alarm when input/output power of the rectifier goes over higher than
rate power.
“Blinking”: Indicates several alarms including over voltage, internal high temperature alarm, SCR high
temperature alarm and unbalancing alarm of output current.
8.4. GiPower
8.4.1. Troubleshooting
The display on the front panel of the rectifier will shows error codes when an error occurs as below.
9. Filter Unit
For more detail information related on maintenance of the filter unit, refer to the chapter 8 in the attached
document, “Operating Instruction”, copyright by HYDAC Process Technology GmbH. Hereafter.
9.1. Maintenance
PARTS MEASURES OR ACTION FREQUENCY
Check gear unit noise for
Geared motor During operation
changes
Check fixing bolts are securely
Geared motor Annually
tightened
At the regular inspection of the
Filter inspection
installation
Back-flushing filter At the regular inspection of the
Visual inspection
Filter elements installation
Cleaning with high-pressure
Possibly after visual inspection
cleaner(water)
At the regular inspection of the
Function check
Shut-off valves installation
Replace As required
At the regular inspection of the
Function check
Position switch installation
Replace As required
Visual inspection of the internal
Back-flushing filter with At the regular inspection of the
corrosion protection
corrosion protection installation
layer
If corrosion protection layer is
Repair corrosion protection layer As required, as soon as possible
damaged
If corrosion protection layer is
Contact HYDAC Process
irreparably As required, as soon as possible
Technology GmbH
damaged
Dismount the filter elements (see section 10.4 “Removing the filter elements”)
Secure the filter elements from outside to inside (see Figure 5)
Warning!
Clean the filter elements from outside to inside (see Figure 5)
✓ Start the cleaning of the elements at a distance of the water jet of approximately 1m.
✓ Then slowly reduce the distance to app. 50 cm.
✓ The minimum distance between element and water jet pistol may not be less than 50 cm.
✓ The maximum water temperature should not exceed 60°C.
✓ No cleaning agents that could damage the filter material may be used.
✓ The minimum pressure of the water jet should be 20 bar
✓ The maximum pressure may not exceed 100 bar.
Typical cleaning agents are e.g. citric acid or hydrochloric acid. The kind of cleaning
agent to be used can be defined by the kind of deposit. A sour cleaning is better
suitable for mineral deposits (scaling). Alcaline cleaning agents are more suitable for
organic impurities.
Please make sure that the proper safety precautions are taken, when using
the cleaning agents according to the material safety data sheet. Also ensure that
only trained and instructed personnel is allowed to perform this cleaning -
especially pay attention to the mixing procedure and potential chemical
reactions.
In order to clean the elements, put them in a bath of the chosen cleaning acid and leave
them in for a maximum time of 1h respectively until you see a desired cleaning effect.
After the elements are taken out from the bath, you have to ensure to clean off any
remaining acid from the elements surface by suitable methods. A second high pressure
cleaning could be finally necessary to remove the dispensed dirt from the mesh
material.
Disposal of the cleaning acid is not covered by this manual, but has to be done in
accordance with safety regulations onboard and according to the applicable
regulations.
Summary: after application of above mentioned alkaline cleaning incl. a dwell time in
the bath of 15min. please thoroughly pressure wash the elements before use.
5 Please hang the chain block and sling belt on the upper plate
9.5. Troubleshooting
Only EPT:
-Set up compressed air supply
Shut-off valve does Compressed air supply of the
-Compressed air supply min. 4 bar
not open pneumatic actuator is not
max. 6 bar
connected
It requires drain sea water and filling fresh water. By filling fresh water, the filter
elements can be protected from sea water corrosion. It’s to be carried out after
ballasting.
(If an uncertified NA solution is used, it may cause performance degradation in the
system)
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Do not inhale liquid. Protect your skin and wear safety gears. Ordinary glass or
plastic is sufficient. Keep away from incompatibles such as oxidizing agents, acid,
and alkalis. Keep container tightly closed. Keep container in a cool area.
Procedure 1 Drain sea water inside of filter by manual drain valve open.
3 Close fresh water supply valve when the pressure transmitter indicates abt.
1bar.
10. Flowmeter
● Error messages which occur during measurement are always indicated via the current and, depending
on the preset relay function, also via the relay output.
● In addition, a system or process error is always indicated when the red LED flashes.
Switch off the power supply before unscrewing the cover to the electronics compartment.
10.1.4. Procedure:
Catalytic sensor of DA-500 is built as one-body molded structure of sensor cap and sensor filter for the
protection of the sensor. Therefore, the sensor exchanging work should be done by the method of exchanging
sensor, filter and the one-body structure of sensor cap. Sensor exchange should be done according to following
procedure by referring to AMP BOARD.
11.2. Caution
● For sensor exchanging work or separating board, power should be cut firstly.
● After checking the possibility of remaining combustible gas or flammable materials near, and cutting power,
then you can do sensor exchanging.
● If the direction of sensor cable and Sensor Board connector socket is different, they cannot be inserted.
● Sensor cap and detector cover should be tightened in order to maintain the pressure-resistant explosion-
proof structure.
11.3. Calibration
Gas sensor’s feature is able to be changed little by little in accordance with the environment such as temperature,
humidity and wind, or the lapse of used time by sensing response degree. This feature is common to all sensors,
and the optimum condition of sensor should be maintained though a periodic calibration.
For the optimum condition of gas detector, regular inspections which includes calibration test are needed
according to the recommended schedule. The recommended calibration schedules are as follows:
Standard gas with density range from 10 %LEL to 100 %LEL must be used for span cali
bration. With the high density gas beyond the sensing range, sensor calibration is impossib
le, and with the excessiveness, some damages may happen to the sensor or life span of the senso
r may be extremely shortened.(basic setting = gas density: 50%LEL, measuring range 0 ~ 100%L
EL)
After exchange of a sensor at the site, the SPAN Calibration should be carried out after 1
0 minutes of power approval.
Take care when checking the TRO sensor to be able to cause the electric shock to the
human. Electric shock may happen by the malfunction of the device.
Safety Strictly comply with all relevant rules for accident prevention and local health and
safety regulations. Wear your Personal Protective Equipment.
Lamp
Key cover
By-pass key
Warning: Do not apply power to the Purge Control X-Purge controller unless the area has been
properly tested and is known not to contain explosive materials.
By-Pass Key
Warning: Use of the by-pass key disables the safety features and safe operation of this instruction.
It is imperative that this operation be performed only after the area has been tested and known to
be safe.
Warning: If the by-pass key can be used to test parts of instrument while the door open. Use
caution as power is live inside the enclosure.
Upon turning the key, the indicator will display a FLASHING RED lamp. This is an indication of the
maintenance mode. Power is retained to the instrument. The key cannot be removed in this mode of operation.
After maintenance has been performed the door should be closed and latched. Air pressure will once again
pressurize the enclosure. The maintenance key can now be turned and removed and the key cover replaced. The
purge controller will display a SOLID GREEN lamp indicating a safe operation and power is kept applied to the
instrument.
Check Valve
Cap Assembly "A"
Flow
Tube"B"
Each indicator reagent will last a maximum of 3 months based on reagent aging when kept in the
operating cooling chamber.
Remove the cap and add the contents of the DPD powder bottle (small brown bottle).
30 seconds
this will empty the cuvette and stop any flow of water.
3) Remove the cap on both bottles replace with the blue cap supplied with the CLX-XT.
Be sure to replace the reagents in the correct location as labeled on the inside of the CLX-XT. The
buffer is installed on the left and the indicator is installed on the right side. The suction tube for both
reagents will reach the bottom of the bottles.
The system will automatically return to normal operation after it has primed.
Use caution when changing the reagents as they are caustic. These reagents will stain
clothing and anything they contact. After changing the reagents,
operators should wash their hands.
Tubing Replacement
1) Press SERVICE to stop the flow of sample water and drain the cuvette.
2) Remove and retain the thumb screw on top of the pump; pull the pump hammer and spring up and
out of the way.
5) Replace the hammer and spring back into place and secure with the
thumb screw.
Check the condition of cuvette and change if it appears badly soiled or discolored. Follow the steps
below:
4. Install the new cuvette by pushing it firmly in place and turning the knurled top clockwise until the
cuvette is held securely.
The flow rate on the CLX-XT was factory adjusted and should not need adjustment. Installation
variances may affect the flow. The optimal flow rate through the CLX-XT may be adjusted if needed.
The flow is adjusted by removing regulator vinyl cap and turning the adjustment screw on the pressure
regulator. To assist in this adjustment follow the procedure shown below:
2. Wait for the display to read HOLd, then press Mode/ Exit.
4. CLX-XT will drain, and then pulse in water while a count is displayed on the screen.
The flow test determines if the flow rate is suitable for proper operation. Loosen the locking nut then
adjust the pressure regulator using a coin or a large flat blade screwdriver. Press either the or
button while in the FLOW routine to display a new flow rate. Please note that only ¼ turn incremental
adjustments should be made to the regulator on each attempt.
If the message is LO, turn the regulator control clockwise. If the message is HI, turn the regulator
counterclockwise. If the message is Good, no adjustment is required. Tighten the locking nut after
adjustment and replace the regulator vinyl cap. To return to normal operation, press the button.
T-Strainer Cleaning
A T- strainer is integral to the instrument and must be checked occasionally. When necessary it must be
removed and cleaned.
Steps:
1. Press SERVICE to stop the flow of sample water and drain the cuvette.
2. Ensure the source water is turned off.
3. The T-strainer is clamped to the intake regulator. The removal requires the use of a flat blade
screwdriver as shown in the photo. Once the clamp is opened the T-strainer can be removed.
4. Disconnect the T-strainer and regulator from the tubing and clear of the case.
5. Clean the T-strainer screen and replace.
6. Be certain the clamp is tightened fully and the pumping connections are complete.
7. Turn source water back on.
8. Return to normal operation.
12.4. Troubleshooting
The purpose if calibration is to ensure that a measuring device is accurate, reliable and consistent. If any
fault occurs, the message is displayed in the bottom row of the LCD as following message.
13. Calibration
The purpose of calibration is to ensure that a measuring device is accurate, reliable and consistent. So
Calibration shall be performed by user regularly in order to maintain performance and normal operation
condition of HiBallast system. Instruments intended for calibration are figured out in BWTS CALIBRATION
SCHEUDLE.
Please refer to the documents, 200. BWTS CALIBRATION SCHEDULE, are provided individually.
BWTS technical support can be provided by maker (HHI). Please contact us via below e-mail address.
bwts@hhi.co.kr
sales@hyundai-gs.com
service@hyundai-gs.com
bwts-tech@hyundaiwelding.com
bwts-cs@hyundaiwelcing.com
【 END 】
Appendix A
Equipment List
Electrolysis unit
1 Blower B-117A
2 Blower B-117B
3 Expension-Joint
4 Venting pipe
5 Drain Assy
6 Drain Tank
7 Drain pump (Magnetic Pump) P-115
8 2way Solenoid Valve(drain pump) XV-1141
9 Filter & Regulator
10 NON-Return Valve
11 Union
12 Male connector
13 Male connector
14 Elbow connector
15 Elbow connector
16 Male connector
17 Elbow connector
18 Male connector
19 Socket
20 Hex Nipple
21 Connector
22 Ball valve MV-1141
23 Cable Gland
24 Bolt Assy (HIB-2000)
25 Gasket Assy (HIB-2000)
30 Strainer S-172A
31 Strainer S-172B
Neutralizing Agent unit
32 Neutralizing Agent Tank V-131
33 Neutralizing Agent Injection Pump P-132
34 Neutralizing Agent Injection Pump P-232
35 Y-Strainer S-1321
36 Expension-Joint
37 Expension-Joint
38 Expension-Joint
39 Reduce Union
40 Hex Nipple
41 Male Connector
42 Male Connector
43 Tube
Miscellaneous parts
44 Male Connector
45 Globe valve
46 Y-strainer
47 Back Pressure Valve
Spare parts
48 Sea Water Feed Pump seal set
49 Neutralizing Agent Injection Pump seal set
50 Strainer O-Ring
51 Neutralization Agent Solution
52 Hand Pump
PPE
Valve List
Electrolysis unit
1 Manual Butterfly valve MV-1181
2 Manual Ball valve DV-1121
3 Manual Ball valve DV-1111
4 Control valve (Intermediate butterfly) FCV-1131
5 Control valve (Intermediate butterfly) FCV-2131
6 Control valve (Intermediate butterfly) FCV-3131
6 Manual Butterfly valve MV-1111
7 Manual Butterfly valve MV-1133
8 Manual Butterfly valve MV-1134
8 Swing Check NRV-1111
9 Auto vent Valve AV-1121
10 Auto vent Valve AV-1121
11 On-off valve XV-1171A
12 On-off valve XV-1171B
SPACER
Seawater feed unit
13 Manual Butterfly valve MV-1711A
14 Manual Butterfly valve MV-1711B
15 Manual Butterfly valve MV-1721A
16 Manual Butterfly valve MV-1721B
17 Manual Butterfly valve MV-1722A
18 Manual Butterfly valve MV-1722B
19 Manual Butterfly valve MV-1731
20 Manual Butterfly valve MV-1732
21 Manual Butterfly valve MV-1733
22 On-off valve XV-1732
23 On-off valve XV-1731
24 Swing Check NRV-1711
25 Swing Check NRV-1712
Neutralizing Agent unit
26 Manual Ball valve MV-1313
27 Manual Ball valve MV-1321
28 Manual Ball valve MV-2321
Sensor List
Electrolysis Unit
1 Conductivity Transmitter CIT-1111
2 Conductivity Sensor Indicator CIT-1111
3 Level Switch LS-1111
4 Level Switch LS-1141
5 Pressure Indication Transmitter PIT-1121
6 Flow meter FIT-1131
7 Flow meter FIT-2131
8 Flow meter FIT-3131
8 Hydrogen gas detector AIT-1121
9 Hydrogen gas detector AIT-1122
10 Hydrogen gas detector AIT-1123
The above list could be different according to vessel’s type and application of HiBallast.
Neutralization
Neutralization Agent (N.A.) - As required
Unit
1) Seal set 1/unit
Neutralization
2) Check ball set 1/unit
Injection Pump
3) Diaphragm 1/unit
TRO Sensor
Reagent kit, Tubing kit 1/unit
Unit
1) Gasket set 1/unit
Electrolyzer
2) Power feeder O-ring set 8/unit
Unit
3) Power feeder ass’y set 8/unit
Seawater Feeding 1) Mechanical seal set 1/unit
Pump 2) Bearing set 1/unit
Only for damaged or
Elements -
Filter Unit corroded part
Seal set 1/unit
Pneumatic valves Sol. valve & regulator 1/unit
Pneumatic diaphragm
Diaphragm 1/unit
pumps
Appendix B
Filter
TRO for Salinity of Temperature Main ballast TRO for
Rectifier current Rectifier voltage differential
ballasting feed water of feed water water flow deballasting
pressure (DP)
Date
5.5 ~ 10 ppm ≥15 PSU 1) ≥10 ℃ 2) < Max. current < Max. voltage < TRC (m3/h, < 0.5 bar 3) < 0.1 ppm
(DD.MM.YYYY)
(A/B/C/D) (A/B/C/D) Treatment Rated
(㎍/l) (㎍/l)
Capacity, Project
specific)
(Sample) 05.04.2019 Avg. 8 20 15 250/250 120/120 4000 Avg. 0.2 Avg. 0.03
Filter
TRO for Salinity of Temperature Main ballast TRO for
Rectifier current Rectifier voltage differential
ballasting feed water of feed water water flow deballasting
pressure (DP)
Date
5.5 ~ 10 ppm ≥15 PSU 1) ≥10 ℃ 2) < Max. current < Max. voltage < TRC (m3/h, < 0.5 bar 3) < 0.1 ppm
(DD.MM.YYYY)
(A/B/C/D) (A/B/C/D) Treatment Rated
(㎍/l) (㎍/l)
Capacity, Project
specific)
1) Dedicated sea water tank or salt tank should be used in low salinity area for ballasting.
2) Min. 4℃ with reduced treatment capacity.
3) Filter will start back-flushing when DP is above 0.5 bar
H2 gas detector for vent pipe H2 gas detector for room H2 gas detector for cofferdam
Date
Remark
(DD.MM.YYYY)
AIT-1121 AIT-1122 AIT-1123
Note
Ballast water discharge from the vessel will be subject to limitations in following table.
Microorganism category Performance standards
Organism, size ≥50 μm <10 living organisms/m3
Organism, 10≤ size <50 μm <10 living organisms/mL
Toxicogenic Vibrio cholera <1 cfu/100 mL
Escherichia coli <250 cfu/100 mL
Intestinal Enterococci <100 cfu/100 mL
To ascertain compliance with above limitations, EPA is establishing the following biological indicator compliance monitoring.
Instrument or Standard
Measurement EPA Method ASTM ISO Other
Analysis Method
Appendix C
Cumulative Train No. 1 Train No.2 Train No.3 Train No.4 Train No.5
Date operation Visual Visual Visual Visual Visual
Cleaning Cleaning Cleaning Cleaning Cleaning
time (hrs) inspection inspection inspection inspection inspection
(Sample) 05.04.2019 70hrs Done Done Done Done NA NA NA NA NA NA