Applications
The instrument is designed for routine and
diagnostic testing like IR, PI, DD, DAR & SV of high
voltage electric equipment like Transformers,
generators, motors, cables, bushings, LA's,
switchgears, etc. Trending of data can reveal vital
information about insulation quality and enables
the user to take informed decision.
www.motwane.com
Standards & Approvals
5K PI Technical Specification Safety In accordance with EN 61010-1
Test Parameter Specifications EMC In accordance with EN 61326-1
Test Regimes (Automatic) : IR, Leakage Current, PI, Type Test In accordance with IS 11994
DAR, DD, SV & BDV Ingress Protection IP 65
Test Voltage : 250V to 5KV in 50V CE & CAT IV 600V
steps Safety of Accessories with IEC61010-31
Test Voltage Accuracy : ±5%
Short Circuit Current : 3 mA Physical Specifications
Three Programmable timers : T1, T2 &T3 Dimensions : 195mm (H) x 290mm
(max. 99 min 59 sec.) (W) x 360mm (L)
Weight : 5Kg. (excluding
Measurement Ranges accessories)
Insulation Resistance
Analog Display : 1K? to 5 T? Environmental Specifications
Digital Range (3 Digit display) : 10K? to 10T? Operating Temperature : 0°C to 55°C
Accuracy (25°C) : ±5% (up to 999 G? ) Storage Temperature : -20°C to 70°C
±20% (1T? to 10T? ) Relative Humidity : <95%RH, Non
Guard condensing
2% error guarding <300 K? leakage with 100 M? load
Indications & Annunciations
Leakage Current Audio visual indication for HV presence
Range : 0.5nA to 3mA LED indication for HV presence
Accuracy : ±5% (>5nA to 3mA) Battery status indication, Battery charging indication
±20% (0.5nA to1nA) Audible discharge annunciation
Capacitance Mains / Battery Indication
Range : 10nF to 50µF
Accuracy : ± (10% +5nf) of rdg. Display details
Real-time display of Insulation Resistance
Voltage Measurement After end of test scrolling display of test results including
Range : Upto 600V DC resistance, leakage current, time & capacitance
Upto 600V AC rms
Accuracy : ±5% of rdg
Accessories
Standard Accessories
Power Supply 3 x 3m test leads for Earth, Guard and Line connection
AC mains with large insulated clamps
Voltage : 230V ± 10%, 1.5m Main power cord
Frequency : 50Hz ± 10% Mains protected fuse
Instruction Manual
Battery RS 232 communication Cable
Internal Rechargeable Battery USB Cable
Backup : Approx. 6 hours of Carrying Bag
continuous testing at
DBS/PC/IR-05/Rev-03
5KV on a load of 10 G? Optional Accessories
Charging time : 4 hours (approx.) 3 X 10m test leads with large insulated clamps
RS 232 to USB converter
Interference Rejection Software CD
1mA rms per KV test voltage
Notes