AUTOCLAVE
(Steam Sterilizer)
ALP Co.,Ltd.
Tokyo Japan
Contents
1. Preliminary notes P.3
4. Installation P.9 to 15
8. Maintenance P.44 to 51
9. Specifications P.52
2
1. Preliminary notes
• Thank you for choosing ALP products.
Keep this manual at your hand for your quick reference whenever operating this autoclave.
• Do not use this autoclave for improper purpose. Strict observance of this manual is required.
.& WARNING :Failure to this warning may cause death or serious injury of person .
(S) Prohibited
o Strictly Observed
3
• How to use autoclave
&WARNING
o This autoclave sterilizes objects with high temperature and high pressure.
Do not operate putting following materials and I or dangerous substances
(Refer to 11.3 Dangerous substances list) in chamber .
• Explosive materials •Flammable materials • Acid or Alkaline materials
•Corrosive materials •Non heat-resistance materials
o In case of sterilization of saline objects such as seawater, discharge saline water after
sterilization. Then put fresh water and operate several times without load. If you spill saline
objects, be sure to wipe off.
Failure to do them may cause explosion due to corroded chamber and pipes.
o Do not disassemble and I or modify autoclave.
Autoclave being pressure vessel. disassemble and I or modification of autoclave may cause accident and I or
breakage.
o Do not move autoclave holding lid nor push it hard.
Lid may jolt out of alignment which leads to serious accident and I or trouble.
o Do not get autoclave wet.
If electric component get wet, it may be broken. get a shock. and even cause fire.
• Installation
&WARNING
o Do not use power strip. (Refer to 4. 4 Power cable connection and protective grounding)
If power cable is bundled or tangled and lor wrongly connected. these may cause fire and I or get a shock.
o Protective grounding conductor must be connected to protective grounding terminal.
Connection to gas pipe or water pipe may cause explosion. get a shock and lor breakage.
o Do not install autoclave in the place where chemicals such as flammable materials, explosive
materials, and lor corrosive gas are kept.
Installing autoclave in above mentioned place may cause serious accident and/or trouble of autoclave.
&CAUTION
o This autoclave should be used indoors. Do not install it outdoors.
There are many possibilities of malfunction and/or breakage of autoclave if it is used outdoors.
o This autoclave must be installed on level and hard floor.
Installation without stability may cause malfunction and lor trouble.
• Cleaning
&WARNING
o Do not use cleaning products for cleaning the inside of the chamber.
It may cause explosion due to corroded metallic parts.
&CAUTION
o Do not use liquid which corrode metal and/or paint (EX. thinner, hypochlorous acid, chlorine
dioxide, acetyl hydroperoxide) for cleaning the exterior.
Metallic parts may corrode.
o Do not splash chlorine chemical and/or cleaning products (EX. hypochlorous acid, chlorine
dioxide, acetyl hydroperoxide) to the exterior during cleaning the floor.
Metallic parts may corrode.
. -
• Notes on the operation
~WARNING
o Be sure to close lid to prevent pinching dusts before operation.
If there is something between lid and chamber which prevents lid from closing, the lid may burst at high pressure
which cause fatal accident or serious injury. Then operator may get burned by steam leaked from chamber.
o Be sure to slide lid lock lever slowly to close position (left end) before operation.
If lock is not enough, operator can not start operation (Er4 is shown). But if operator can start operation by
malfunction of detection part, lid may explode at high pressure which cause fatal accident or serious injury.
o Objects (including disposable plastic bag) must be put in stainless wire basket or stainless bucket
when sterilization.
If operator loads objects into chamber directly, it may cause choking pipes which may lead to serious accident.
<9 Do not open lid forcibly during operation.
Opening lid forcibly during operation may cause fatal accident or serious injury.
<9 Do not open lid until chamber pressure goes down to "OMPa" when operation completed or
operation interrupted.
Steam spouts out of chamber by residual pressure which may cause burn injury.
o In case of sterilizing liquid objects, cool down objects enough after sterilization. Then unload
objects slowly.
liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber. Therefore extreme temperature change when unloading may cause high temperature liquid
spouting. It may cause burn injury.
o In case of sterilizing liquid objects, metallic or heat-resistance glass container must be used.
A glass container without heat-resistance may be easy to break by extreme temperature change and operator may
be injured by this breakage.
<9 In case of sterilizing liquid objects (EX. liquid media), do not close container with lid or bung which
has no breathabllltv.
It may cause rise of pressure in container, because temperature in container cool down slowly. Especially glass
container is easy to explode if it is damaged.
<9 Do not touch autoclave more than necessary.
If operator touches autoclave more than necessary, it may cause serious accident.
<9 Do not open drain valve under following conditions.
- during operation
- chamber temperature is high
- chamber pressure is not OMPa
It may cause burn injury by blow-out of boiling water and / or steam, water lack, and/or malfunction.
<9 Do not touch power switch(breaker) by wet hand.
Operator may get electric shock.
~CAUTION
o Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove, operator may get burned.
<9 Do not touch exhaust bottle and exhaust hose during air evacuation and exhaust.
It may cause burn injury. Do not touch them until cooling down enough.
<9 Do not touch stainless lid directly when opening or closing lid.
It may cause burn injury. Do not touch it until cooling down enough.
o Genuine accessories and spare parts made by ALP Co., Ltd. must be used.
Using other company's products may cause accident and lor breakage of autoclave.
5
• Notes on power failure
~WARNING
('9 Do not open lid until pressure gauge indicates OMPa and TEMPERATURE DISPLAY indicates below
80°C when operation interrupted.
Steam spouts out of chamber by residual pressure which may cause burn injury.
('9 Do not open drain valve under following conditions.
- chamber temperature is high
- chamber pressure is not OMPa
It may cause burn injury by blow-out of boiling water and / or steam. water lack. and/or malfunction .
•• Be sure to set automatic exhaust OFF when sterilizing liquid object or glass container.
(Do not press EXHAUST KEY.)
Quick reduction of chamber pressure may cause breakage of glass container or blow-out of hot liquid .
•• When sterilizing liquid, it takes longer to cool. When opening the lid, pay attention not to get burned
then unload the objects avoiding the impact to them.
liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber.
~CAUTION
•• Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove. operator may get burned .
•• Manual exhaust under 121°C is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.
6
3. Part names
Main body
Lid cover
Lid lock lever
Operation panel
:.
Power switch
(Breaker)
Drain outlet
Exhaust bottle
Power cable
7
/
1. Pressure gauge 14. STERILIZE LAMP
It indicates chamber pressure. The character lights up during sterilizing process. The
2. TEMPERATURE DISPLA Y character blinks at sterilizing setting change
It displays temperature in the chamber or setting condition.
temperature. Also. it displays the date of an operation 15. EXHAUST LAMP
reservation. The character blinks during exhaust process.
3. TIMER DISPLA Y The character lights up before exhaust process in
It displays the remaining time to the process case of automatic exhaust ON.
completion. "down count" Also. it displays the time of 16. WARM LAMP
an operation reservation and error. The character lights up during warming process. The
4. START KEY character blinks at warming setting change
Press it to start operation. condition.
5. STOP KEY 17. COMPLETE LAMP
Press it to stop operation on the way. Press it to reset The character lights up with the buzzer beeping at
previous operation before starting operation. the completion of all the processes. The character
6. PATTERN/ENTER KEY blinks and reports whether lid can be opened when
Press it to fix an operation pattern and each setting. pressing STOP KEY during operation.
7. TEMP/TIME KEY
Press it to check setting temperature. setting time PROCESS INDICATION BAR
and the operation pattern. Press it to change setting They light up, blink. light off according to the status of
display's blink. process.
8. CHANGE ..•• KEY 18. HEA TING-1 PROCESS LAMP
Press it to increase the figure of the setting display. It blinks from start temperature of the sterilization to
9. CHANGET KEY the end of the air evacuation.
Press it to decrease the figure of the setting display. It blinks from start temperature of the dissolving to
10. EXHAUST KEY the dissolving temperature.
Press it to switch automatic exhaust ON and OFF or 19. HEA TlNG-2 PROCESS LAMP
set an automatic exhaust starting temperature. Press It blinks from the end of the air evacuation to the
it to exhaust manually. sterilization temperature.
11. START TIMER KEY 20. STERILIZATION PROCESS LAMP
Press it to make a reservation of operation. Set an It blinks during sterilization.
operation starting time of month. day. hour. and 21. COOLlNG-1 PROCESS LAMP
minute. It blinks to 100"C after sterilization.
12. READY LAMP 22. COOLlNG-2 PROCESS LAMP
The character lights up at the reservation condition. It blinks from 99"C after sterilization to the
The character blinks at setting change condition. completion.
13. HEAT LAMP 23. WARMING PROCESS LAMP
The character lights up during heating or dissolving It blinks during warming.
process. The character blinks at dissolving setting 24. COOLlNG-3 PROCESS LAMP
change condition. It blinks from 99"C after warming to the completion.
8
4. Installation
4. 1 Transportation
~WARNING
CS> Do not pinch hands I feet when moving autoclave.
There is a risk of a fracture or bruise.
~CAUTION
The autoclave must be moved by two or more persons.
~ Do not hit and lor get shock to autoclave nor hold lid cover. (Fig. 1)
It may cause breakage or damage of autoclave.
o In case of lifting the autoclave, be sure to use a heavy machinery. Do not lift by humans.
It may cause a falling accident.
o In case of lifting the autoclave by a forklift, be sure to insert the fork under the bottom plate
avoiding the casters. (Fig.2)
o In case of lifting the autoclave by a crane, be sure to hang the bands to four corners of the
bottom plate. Put cushioning between bands and the autoclave. (Fig.S)
9
4. 2 Installation location
Autoclave must be installed as follows.
Refer to 4.3-4.5.
Up to ceiling or duct
Wall
~ surface
-L00-L-L-L-L-L-L-L..,L~~-L-L-L-L-L.L....L....L..- Wall Pay attention to electric
surface components and alarms
150mm or more of the ceiling. 800mm or more
I
100mm or more
Wall surface
or other O:=iiIHIiiiHIiIHiiiIHiIiiiiiiIii=:O
0:::::::::::::::::::::::::::-
~
instrument 150mm or more
r----, Exhaust
Pour water to exhaust bottle
up to LOW level. and insert
r:
I
cooling
tank
exhaust hose into the bottle. I
I
Then. insert the bottle into I
main body. Do not bend the I
I
hose when insertin the bottle. I
I
I
I
I
10
~WARNING
(9 Do not install autoclave in the place where chemicals such as flammable materials, explosive
materials, and lor corrosive gas are kept.
Installing autoclave in above mentioned place may cause serious accident and/or trouble of autoclave.
o This autoclave must be installed to a place where operator can read pressure gauge, or it may
cause serious trouble.
(9 Do not bend exhaust hose when setting exhaust bottle, or it may cause serious accident and / or
breakage.
~CAUTION
o This autoclave must be installed on level and hard place.
o Keep 100 to 150 mm space between autoclave and wall or other instruments.
o Install autoclave avoiding humid and dusty place and keep away from direct sunlight.
o Lock stoppers of front casters.
(9 Do not install autoclave under electric components and alarms of the ceiling, or it may cause
malfunction of them by steam.
o Be sure to use exhaust bottle, or steam from chamber is splattered to outside which may cause
burning.
o Pour water to exhaust bottle up to LOW level.
If water level is under LOW level, steam spouts out directly which may cause burning.
(9 Autoclave must not be installed outdoors.
Breakage of autoclave may happen if it is installed outdoors.
(9 Do not use exhaust bottle for drain.
It may cause burn injury by blow-off of boiling water.
11
4. 4 Power cable connection and protective grounding
Lid is not opened until putting the power on. Slide lid lock lever to open position (right end)
after power on. Then open lid and unload all contents inside chamber.
(Refer to 4.5 Standard accessories)
Do not press START KEY during this operation. Otherwise contents inside chamber may
burn out due to the water lack.
&WARNING
o Be sure to connect the power cable to the breaker which meets specifications of the autoclave.
Grounding cable must be connected to grounding terminal.
~ Do not bend and / or bundle power cable.
Improper wiring may cause fire or power leakage due to bad electrical failure or malfunction.
o Do not put heavy things on power cable.
Improper wiring may cause fire or power leakage due to bad electrical failure or malfunction.
o Do not connect grounding cable to gas pipe, water pipe, telephone line or lighting rod.
Improper connection may cause fire or power leakage due to bad electrical failure or malfunction.
If connector with grounding terminal is not provided. make the connector with grounding terminal.
o Do not touch power connector by wet hand.
Operator may get electric shock.
~CAUTION
o Power cable must be singly-connected to the commercial power with proper voltage and
amperage.
Improper connection may cause malfunction or trouble.
12
4. 5 Standard accessories
CDRead the cautions of sticker which covers power switch(breaker), then remove it.
® Turn on power switch, then slide lid lock lever to open position (right end) and open lid.
@ Unload all standard accessories as below.
@ Set every accessory and pour water into exhaust water cooling tank following below
instructions.
o 0 0 0 0
,
Base plate, 1pc.
Exhaust bottle(3L), 1pc.
13
.Base plate hook (for CL-40L only)- •• Keep it with this instruction manual.
Purpose of use: Tool for unloading base plate from bottom of chamber at the time of
maintenance.
II \V
E/I\
E
0
~
Funnel 0
.•....•
0
C'?
Plug
Figure 1 Figure 3
14
Keep the silicon
hose straight.
Figure 2
~WARNING
o Be sure to set base plate during operation.
Operation without base plate may cause burning of object which leads to fire.
o Do not bend exhaust hose when setting exhaust bottle, or it may cause serious accident and / or
breakage.
~CAUTION
e Do not pour too much water into exhaust cooling tank.
It makes overflow which may cause malfunction and/or breakage of autoclave.
o Water
Be sure to stopple the funnel after pouring water into exhaust cooling tank.
in exhaust cooling tank reduces faster therefore frequent water supply is required.
o Beburning.
sure to use exhaust bottle, or steam from chamber is splattered to outside which may cause
15
5. Operation method
&WARNING ----------------------,
o Be sure to read this chapter carefully and understand well before operation and use autoclave
properly, or it may cause fatal accident at worst as well as mechanical trouble .
16
Before operation
~WARNING
o In case of sterilization of objects such as electrolyte-containing saline (EX. sea water), be sure to
discharge water in chamber and clean it after sterilization complete. Then put fresh water and
operate several times.
Failure to maintain after sterilization may cause corrosion of bottom of chamber by electrolytic corrosion which
leads to rot.
o Be sure to check if pressure gauge indicates "OMPa".
If pressure gauge does not work properly, please contact local distributor.
If pressure gauge is wrong. it may cause serious accident and/or breakage of autoclave.
o Do not bend exhaust hose when setting exhaust bottle, or it may cause serious accident and / or
breakage.
o Do not touch power switch(breaker) by wet hand.
Operator may get electric shock.
OPEN CLOSE
Drain valve
Inside of the ]
[ housing hole
~CAUTION
o Be sure to close drain valve.
If water leaks from chamber. it may cause water lack. malfunction and/or breakage of autoclave.
17
5. 3 Water supply
5. 3. 1 Water supply
Proper level
Heater
&WARNING
\S) Do not open lid until chamber pressure goes down to "OMPa".
Steam spouts out of chamber by residual pressure which may cause burn injury.
&CAUTION
o Water level in camber is getting lower during operation. Check water level in every operation and
pour water to the same level if it's under proper level.
o Water put in chamber must be distilled one or purified one. Frequent cleaning of chamber is
requested when using city water which generates scale in it.
\S) Do not spill water on the autoclave when pouring it. If you spill water, wipe up it.
o Change water for sterilizing frequently. Water must be changed every day in case of daily use.
o Do not use well water and industrial water which may cause corrosion and/or smudge.
Failure to use proper water may decrease life of the heating element and/or the chamber.
18
&CAUTION
o Water level in exhaust bottle rises every operation. Keep the level between LOW and HIGH
at any time.
Operating over HIGH level is a cause of burn injury due to boiling over from exhaust bottle.
Operating under LOW level is also a cause of burn injury due to steam blowing.
OAt the time of drainage from exhaust bottle, check the bottle if it is cold enough, or it may cause
burn injury.
Exhaust bottle keeps high temperature soon after air evacuation and exhaust.
5. 4 Load object
CD Objects must be put in the stainless wire basket or the stainless bucket (optional
accessory), then set in the chamber slowly.
* In case of CL-40L, use a step as below figure when loading the stainless wire
basket(optional accessory) or the stainless bucket (optional accessory) into the
chamber.
fJJ
~------n!.\.-1)
._-------'
19
*Do not put too much objects in the chamber.
Too much objects in the chamber may cause sterilization incomplete.
(Quantity of objects to be put inside the chamber shall be less than 60% of the capacity
of chamber)
~WARNING
o Be sure to use the stainless wire basket, the stainless wire basket with pan or the stainless
bucket.
In case of sterilization without above accessories. the chamber pressure become out of control
due to the objects choke up the pipe which leads to serious accident such as explosion.
o Be sure that objects do not make contact with the inside of the lid.
lor serious accident.
If the lid is not closed properly. it may cause serious injury by blowing steam
~CAUTION
o In case of sterilization of flagons such as vial container, screw cap bottle, conical flask, take out
cap or screw down it.
(Do not close mouth of flagons with impervious material such as silicon or rubber.)
o In case of sterilization of liquid objects, quantity of object to be put inside container shall be less
than 80% of capacity of container.
Too much object in container may cause bumping when objects are taken out of the chamber after operation
complete.
20
.About a disposable plastic bag
When using a disposable plastic bag or other kind of bags, please be strictly observe following
warnings/ cautions.
~WARNING
o Be sure to put the bag into Stainless bucket or Stainless wire basket. (Fig. 1)
If the bag is put into chamber directly, autoclave may cause serious explosion. When exhaust holes located in the
wall of chamber are chocked by the bag, safety devices may not activate.
Following safety devices may not work due to choking of exhaust holes in chamber.
- Over pressure detector (Error message is shown on the display and buzzer beeping)
- Safety Valve (Steam evacuates out of chamber)
& Do not put too much bags into chamber. (Fig. 2,Fig.3)
Only single pack of bag must be put in chamber. (Fig.1)
Putting too much bags into chamber may cause serious explosion. When exhaust holes located in the wall of
chamber are chocked by the bag, safety devices may not activate.
o Be sure to fold back the mouth of the bag inside of Stainless bucket or Stainless wire basket.
(Fig.4, Fig.5)
If the mouth of bag is folded back outside of Stainless bucket or Stainless wire basket, autoclave may cause
serious explosion. When exhaust holes located in the wall of chamber are chocked by the bag, safety devices may
not activate.
& The bag must not be caught between lid and mouth of chamber.
Operator may burn by spouting hot water and/or steam which may leads to explosion, because the bag bursts in
chamber.
x x
21
o x
Fig.4 Fig.5
&CAUTION
o Do not put too much object in the bag.
(Quantity of object shall be less than 60-70% of bag capacity.)
Putting too much objects into the bag may cause improper steam penetration which leads to activation of
safety devices in addition to the sterilization incomplete.
o The bag must have open mouth.
When closing the mouth of the bag, residual air may cause sterilization incomplete.
If the bag need to be tied with band for reason of protective from hazardous substance, be sure to secure space
for the steam going through the bag. Or safety devices may activate in addition to the sterilization incomplete.
o For sure sterilization, put 100mQ to 200mQ water into bag.
Steam is generated in the bag which leads to getting better sterilization.
o Be sure to put the bag into genuine stainless bucket or stainless wire basket with pan made by
ALP Co., Ltd. when you put objects with sharp edge such as broken petri dishes, test tubes,
needles into the bag directly.
Using stainless wire basket may cause spillage of waste water in the chamber and breakage of the autoclave.
o Be sure to check the water level of chamber before operation.
Insufficient water in chamber may cause melting bag and stick on the base plate by directly heated from
heating element located in bottom of chamber.
o Sterilization time / temperature shall be longer/higher as the Quantity of objects is increased.
Sterilization conditions may vary depending on the objects. Use the chemical indicator and so on to get suitable
sterilization condition for each object.
Water
(approx 200m!)
Heating element
Fig.6 Fig.7
Recommended method If you need to close the mouth of the bag to protect
from hazardous substance etc., instruction by Fig.7 is
one of the solution.
22
.How to use stainless bucket (optional parts)
~WARNING
o Load stainless bucket in the center of chamber.
If the space between the bucket and the chamber's wall is not enough. exhaust holes located in the wall of
:. chamber may choke. It leads to serious accident.
Following safety devices may not work due to choking of exhaust holes in chamber.
- Over pressure detector
(Error message is shown on the display with buzzer beeping and power to heater is cut off.)
- Safety Valve (Steam evacuates out of chamber even if over pressure detector doesn't work oroperly.)
~CAUTION
(9 Do not stack stainless buckets up in the chamber..
If the space between buckets is not enough. the sterilization may be incomplete as air convection in the chamber
is insufficient.
(9 Do not overload objects which are difficult to penetrate steam (linen, gauze, etc+-), or it may
cause sterilization incomplete.
o Put water (200m!) into stainless bucket, or it may cause sterilization incomplete.
The sterilization time shall be longer if you sterilize without water.
(9 Do not put too much objects in the chamber for sterilization.
Quantity of objects to be put in the chamber shall be less than 60% of the chamber capacity.
o Sterilization time / temperature shall be longer/higher compared with stainless wire basket, as
evacuating air's efficiency of sterilization bucket is not good.
Use the chemical indicators and so on to get suitable sterilization condition.
spaces
bucketStainless
Water
(about 200m!)
Chamber
Objects
Base plate
Heater
23
5. 5 Close and lock the lid
~WARNING
o Be sure lid gasket has no dusts and crack.
Wipe off dusts on gasket with a soft cloth.
If you find crack on gasket. please contact local distributor.
If you leave crack one unchanged. it may cause burst at worst as well as leakage of steam.
o Be sure to slide lid lock lever slowly to close position (left end) before operation.
If lock is not enough. operator can not start operation (Er4 is shown). But if operator can start operation by
malfunction of detection part. lid may explode at high pressure which cause fatal accident or serious injury.
(9 Do not slide the lid lock lever by force.
Forcing the lever to move may damage to mechanical locking device. it may cause unexpected accident.
~CAUTION
o Soon after previous operation (in case that TEMPERATURE DISPLAY indicates 50-79°C after
opening lid). wait for a minute with closing lid. then slide the lid lock lever.
Reactive force arises when closing lid because the air which was caught by lid is swollen by heat of chamber.
Therefore. it's necessary to keep the lid closing for a minute to discharge swollen air in chamber.
CD Press STOP KEY to reset completion process of previous operation, in case you start
new operation continuously.
~CAUTION
o Be sure to press STOP KEY before setting and starting operation.
If completion process of previous operation is not reset. you can not set and start new operation.
24
5. 7 Set operation pattern, value of temperature and time
5. 7. 1 Operation pattern setting
Initial condition
Steri I ization Press PATTERN/ENTER KEY.
pattern
.:
r-------
,-----
I
I
I -,
_--;-;::' ",.--C~' I
MON
8 I' I
DAY °c
EXHAUST
STERILIZE
-
~~
TIME ~~
I
I START II HEAT \c::=:J \\
TIMER I, D
'READY
-- .•.
I -, I I 8
EXHAUST _STERILIZE
_ ---::::'
",.--C~'
lION DAY
°c I'
1-' • -, 1-' START
TIMER
II
I, HEAT \\-,
D WARM
U·Cu 'READY
-- .•.
L l
I\
\ i10N
nn~8
, L'
§
DAy
c
c::=:J
~
I
-----1/
, START
I
TIMER HEAT \\-, WARM
READ{
D
-- .•.
• Sterilization pattern
This operation pattern is for sterilizing objects .
25
5. 7. 2 Temperature / time setting
Setting range
Sterilization pattern
CL-32S/32L : Sterilization ternp.: 100°C to 140°C
Sterilization time: Ominute to 48hours
CL-40S/ 40M/ 40L : Sterilization temp.: 100°C to 137°C
Sterilization time: Ominute to 48hours
Sterilization/Warming pattern
CL-32S/32L : Sterilization temp.: 100°C to 140°C
Sterilization time: Ominute to 48hours
Warming ternp.: 40°C to 60°C
Warming time: Ominute to 48hours
CL-40S/ 40M/ 40L : Sterilization ternp.: 100°C to 13rC
Sterilization time: Ominute to 48hours
Warming ternp.: 40°C to 60°C
Warming time: Ominute to 48hours
If you want to operate continuous run (long run over 48hours), contact local distributor.
~CAUTION
o In case object is a large amount of liquid or object is put in container which is difficult to
evacuate air, temperature of object rises slowly. In this case, set sterilization temperature
higher or set sterilization time longer.
Use the chemical indicators and so on to check sterilization condition for object periodically.
26
5. 7. 2 Temperature / time setting
Setting range
Sterilization pattern
CL-32S/32L : Sterilization temp.: 100°C to 140°C
Sterilization time: Ominute to 48hours
CL-40S/ 40M/ 40L : Sterilization ternp.: 100°C to 137°C
Sterilization time: Ominute to 48hours
Sterilization/Warming pattern
CL-32S/32L : Sterilization ternp.: 100°C to 140°C
Sterilization time: Ominute to 48hours
Warming ternp.: 40°C to 60°C
Warming time: Ominute to 48hours
CL-40S/ 40M/ 40L : Sterilization temp.: 100°C to 137°C
Sterilization time: Ominute to 48hours
Warming ternp.: 40°C to 60°C
Warming time: Ominute to 48hours
If you want to operate continuous run (long run over 48hours), contact local distributor.
~CAUTION
o In case object is a large amount of liquid or object is put in container which is difficult to
evacuate air, temperature of object rises slowly. In this case, set sterilization temperature
higher or set sterilization time longer.
Use the chemical indicators and so on to check sterilization condition for object periodically.
26
When press ~ sh ift tocJ
Initial condition
(In case of Steri Iization / Warming pattern)
In case of other patterns, operate in the same way.
/ ¥r
When press( / ~ sh ift
,
,/
, //
/
to: ~
r
----I
1\
r-----~\
~
\
".,,,
LI· CLI
PATTERN
~
I
HR MIN
---1'
To the mode of
D temperature / timer setting.
I
I
I
I
I
I
I
I
I
I
L_,
I
----,
J
I
/
"
I ,
Ie
MON
LI·CLI
. ",
I
I
DAY'-
°c
8XHAUST
I
, \
"
~E
HEAT
/
~WARM \ PATTERN
~
HR MIN
"
!EAO~
/
.
(//"
,
."
/
/
/
' .....
/
II
-,
e I 18
0c EXHAUST
ST~RllfzE
,iiIiiiiiiiii\
~~
MON DAY
I
L
1 ,
JI
In ,
LI· _,LI
HR
• _,
- ,- ,
MIN
START
TIMER
-,
HEAT
JEAO:(
/
~-\ WARM
.
"
,
..•
~
-:.
/
.iNTER
/
[SZJ~
Change steri Iization time
D by CHANGEA ••• KEY.
-
STERILI ZE
'-
L
J
I
,
I,
MON
IC·
<ru»
LILI
DAY
HR MIN
----I
MON
t:
o I, I 8
LI
DAY
-
STERILI ZE
HEAT
______ JI
HR MIN
I "JEAO:(/
• Automatic exhaust
Automatic exhaust enables to exhaust steam to outside automatically after sterilization to cool
down Quickly. It's necessary to set in each operation depending on objects as follows.
For appliance(except glassware) ••. on
For glassware, liquid or agar medium, other high viscosity substance .•• of F
Even in case of on, automatic exhaust starting temperature can be set according to boiling
point of objects to prevent bumping. Setting range of automatic exhaust starting temperature
is 102 to 121°C.
~CAUTION
o Set automatic exhaust or not in each operation depending on object to be sterilized.
The object which is insufferable under rapid pressure and temperature reduction may burst in chamber in case
of automatic exhaust.
.Manual exhaust
Either automatic exhaust oFF, or automatic exhaust on after sterilization complete, you can
exhaust manually by pressing EXHAUST KEY before the temperature in the chamber reaches to
automatic exhaust starting temperature.
If you press EXHAUST KEY again, you can stop exhaust.
~CAUTION
•• Manual exhaust under 121°C is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.
28
WON
-' c I
L _'°c DAY
-
I \
STERILIZE
G ~ TIWE
~~
tHEAT
II ·:111
LI· eLl
Ifl WIN
\=~ PATTERN
~ [2J~
~ To the mode of exhaust setting
~[iJ
[2J~
~ Press PATTERN/ENTER KEY to fix the setting.
II
LI
DAY
°c
eXHAUST
~[iJ
[2J~
START
TIMER
on
-:==--..",IflO------..,.W=IN
I
I C
______ C
__ -_'_°c
I eXHAUST
I S~E
~[iJ \EXHAUST
~::~ I \=\\
110O DAY \
II
LI·CLI
II .:J I HEAT
\'l ~
[\7](
v 1 PATTERN S<!>TAR
T )
Ifl WIN
29
5. 8 Press START KEY
Press START KEY to start operation.
CD The heater is on with "the output lamp of heater" on TEMPERATURE DISPLAY lighting up, and
HEATING-1 PROCESS LAMP blinking.
(2)Mixture of steam and air in chamber is evacuated to exhaust bottle with rise of temperature. This is
the process of filling chamber with saturated steam, not abnormal condition. The mixture steam
stops evacuating at 100°C to 120°C by closing exhaust valve which is controlled by
microprocessor.
When the air evacuation finishes, HEATING-2 PROCESS LAMP blinks as below chart.
Afterwards, temperature and pressure in chamber rise quickly to setting sterilization temperature.
@ When the temperature in chamber reaches to setting temperature, sterilization timer starts to work,
and remaining time of sterilization is counted down on TIMER DISPLAY.
Afterwards, the autoclave keeps the temperature and pressure constantly during sterilization time.
STERILIZA TION PROCESS LAMP and the output lamp of heater blink during sterilization as below
chart.
Start operation.
Press START KEY.
D
~
~ MIN
D
G ~
I II
I LI
I I
I·c
eXHAUST
1: ~ =
~~~rI\l ~
;, .J;' I ~I/
LI.LLI ~
~AT "=,,
\) \)~
PATTERN (v] ~
~
HR MIN
@ D
emalnlng time 0
ster i I i zat i on
~WARNING
o Do not touch autoclave more than necessary.
If operator touches autoclave more than necessary. it may cause serious accident.
30
5. 9 Operation complete
I " "
I LI I
I
°c
eXHAUST
I ,X:A~TE)
~E -,
~ TIME
~[i]
lION
II · II II
LI• LI LI
DAY
~i~~~ I HEAT
I
\\ r===J \\
\\ \\
PATTERN
~ [7J~
HR WIN
E)
'-----M-:-oN---oAY--'
o " "
LI °c
eXHAUST
I ~E
\EXHAUST
~ TIME
II · II II ~i~~~
II HEAT \r===J \\ PATTERN
LI. LI LI I \\ ~
HR WIN COMPLETE
@ After sterilization time has elapsed, complete alarm sounds then power to heater is cut off. In case
of setting automatic exhaust ON as above chart, automatic exhaust starts after temperature in
the chamber reaches to automatic exhaust starting temperature to decrease pressure forcibly in
chamber.
(In case of automatic exhaust OFF, EXHAUST LAMP does not blink and autoclave does not exhaust
steam automatically. When you press EXHAUST KEY, autoclave exhausts steam.)
COOLlNG-1 PROCESS LAMP blinks from sterilization temperature to 100oe.
After that COOLlNG-2 PROCESS LAMP brinks from ggoe to completion.
@ When TEMPERATURE DISPLAY indicates so-e. complete alarm sounds and COMPLETE LAMP
lights up.
® Press STOP KEY to release the lid lock. Then the lid can be opened and COMPLETE
LAMP turns off.
*In case of interruption of operation, the lid can not be opened till TEMPERATURE
DISPLAY indicates below 80°C too.
31
&WARNING
o Be sure to set automaticexhaust OFF when sterilizing liquid object or glass container.
(Do not press EXHAUST KEY.)
Quick reduction of chamber pressure may cause breakage of glass container or blow-out of hot liquid.
(9 Do not slide the lid lock lever by force nor open lid till TEMPERATURE DISPLAY indicates below
80°C.
Lid lock can not be released. Because it is mechanically-locked. Forcing the lever to move may damage to
mechanical locking device, it may cause unexpected accident.
&CAUTION
o Manual exhaust under 1210C is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.
o Be sure to press STOP KEY to release lid lock after operation complete.
If you don't press STOP KEY, lid lock is not released.
(9 Do not slide the lid lock lever to the right side before you release the lid lock.
The lid lock may not be released which leads to Er4.
32
.Checking operation pattern, value of temperature and time during operation
I' MON
2 51~8
§=~ DAY
G
~[IJ
" . '-LI
L'·
HR
'" MIN
START
TIMER
/'
tHEAT
\
~~~
PATTERN
~ [2J[iJ
Steri Iization
temperature D
" /
I
I ,
c- I
I
°c
8 S?
RIl
W G
~[IJ
[2J[iJ
Warming
temperature D
t: LI
J "
°c
8 I-\ STERILIZE
G
~[IJ
~M\ [2J[iJ
EXHAUST LAMP blinks when setting automatic exhaust ON.
The display returns to previous indication after 5 seconds elapsed.
33
5. 10 Unload object (below 80°C in chamber)
~WARNING -------------------,
o Do not open lid until pressure gauge indicates OMPa and TEMPERA TURE DISPLA Y indicates below
80"C.
Steam spouts out of chamber by residual pressure which may cause burn injury.
•• When sterilizing liquid, it takes longer to cool. When opening the lid, pay attention not to get burned
then unload the objects avoiding the impact to them.
liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber.
•• In case of breakage or spill of objects, be sure to clean inside of chamber to use autoclave safely,
or it may decrease life of heating element and/or chamber, then causes corrosion of chamber and
serious accident.
~CAUTION
•• Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove, operator may get burned.
o If you spill liquid when unloading objects, be sure to wipe off it.
It may cause malfunction or trouble of autoclave.
34
5. 11 Close and lock the lid
CD Drain water in chamber by laying down the lever of drain valve in case autoclave is not
used for a long time.
® Turn off power switch(breaker).
• Stop operation
~WARNING
IS> Do not open lid until pressure gauge indicates OMPa and TEMPERATURE DISPLAY indicates below
so-e.
Steam spouts out of chamber by residual pressure which may cause burn injury.
o When sterilizing liquid, it takes longer to cool. When opening the lid, pay attention not to get burned
then unload the objects avoiding the impact to them.
liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber.
o In case of breakage or spill of objects, be sure to clean inside of chamber to use autoclave safely,
or it may decrease life of heating element and/or chamber, then causes corrosion of chamber and
serious accident.
~CAUTION
o Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove. operator may get burned.
o Manual exhaust under 121°C is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.
35
5. 13 Other functions
Autoclave has calendaring function. It's necessary to set date and time as follows according
to the location.
Local time has already been set. But this time data will be lost about one month of no
electrical power supply to autoclave. Then calendar is reset to Jan. 1st, 2000 A.D. 0:00. (Er6
indicated)
Set up current date and time as following procedure.
, C EXHAUST STERILIZE
- IQO[1 ~ § ~~
D ~
t:....:::..:..J
C_'·c I
---""'::::::------:o=,y
e
1\ W
HR IIIN
. J II II II G [~][1\ 1 ~
L LI LI LI c
eXHAUST
I HR 111M
I @~ / ~AD\
r~~v [¢ 1
36
5. 13 Other functions
Autoclave has calendaring function. It's necessary to set date and time as follows according
to the location.
Local time has already been set. But this time data will be lost about one month of no
electrical power supply to autoclave. Then calendar is reset to Jan. 1st, 2000 A.D. 0:00. (Er6
indicated)
Set up current date and time as following procedure.
2 51'c8,s=\E G§~~
---MQN:::::-----:D="
n ':J',
LI·CLI
I
1 \=~--r.;;,; ~ ~<l>
TIMER
START HEAT PATTERN \7
V
START
HR MIN
o
LI L
J 0
_"c 81/ T
=
~
G~H/\l~
~~®
1/5MQN:)5~1 @I/ ~=~ ~ V ST~T
HR WIN ~EAD{
36
5. 13. 2 Reservation operation
Autoclave has function of reservation operation which can be set starting time of operation
with calendar timer. If you set reservation operation for next day, you will be able to unload
objects from chamber next morning. This function save time especially busy morning. You
can use it as following procedure accordingly.
__
-,I-I 8I
_
L
_
' °c
EXHAUST _
STERILIZE
\ IQ"O:Tl ~~
~
TIME
~ 1\ ~ STOf'
L::,@
,~ • :J ,;Y
u.cu
fYT:: II
,
HEAT \YWARM
~
\ PATTERN
~
(V 1 ~~
~ IIIIN
&CAUTION
•• Be sure to check drain valve is closed and water level in chamber before reservation operation.
There is a possibility that autoclave breaks down due to water lack.
37
5. 14 For the reliable sterilization
In case of sterilizing following containers which hinder the circulation and penetration of
steam, the temperature rises slower than temperature in chamber. Because residual air in
following containers is difficult to be evacuated.
Also it may cause Er1 (over pressure error) .
Especially, in case of using disposable plastic bag and stainless bucket, refer to
5.4 Load object to prevent sterilization incomplete and/or danger of over pressure.
To operate assured sterilization, be sure to check result of the sterilization by the chemical
indicators and so on.
38
6. Safety device
&WARNING
(S) Do not disassemble or adjust safety valve since it has been adjusted before shipment.
It may cause fatal accident because of abnormal performance.
(S) Do not touch power switch(breaker) by wet hand.
It may cause electric shock.
&CAUTION
(S) Do not pour water into chamber until heater cools down enough after water lack, or it may cause
heater trouble.
6. 1 Safety valve
CDKeep away from autoclave to avoid steam blowing from bottom plate.
® After losing steam, turn off power switch(breaker). Then let chamber cool naturally.
@Contact local distributor. When contacting the distributor, be sure to inform model and serial number.
39
7. Trouble shooting
~WARNING ----------------,
o If the situation of autoclave correspond to following item(CDor<2) when you ask for repair
autoclave, make autoclave and accessories to be non-polluting condition.
CD Part of autoclave is exposed to infectious substance or radioactive substance.
(2) Blood or other hazardous components for human body adhere to part of autoclave.
(9 Do not touch power switch(breaker) by wet hand.
It may cause electric shock.
~CAUTION
o When water lack alarm sounds, turn off power switch(breaker) and leave autoclave as it is at
least one hour because the chamber is high temperature. Then pour water into chamber, or it
may cause heater trouble.
In case of trouble happening despite you operate correctly following this instruction manual,
check the cause and suitable treatment stated on below table. If you cannot find out proper
article on below table, or if it is difficult to treat, turn off power switch(breaker). Then
contact local distributor and notify serial number of the autoclave.
7. 1 Error list
*Refer to 7.2 Trouble shooting about causes and countermeasures of the errors.
40
7. 2 Trouble shooting
TROUBLE CONDITION CAUSES COUNTERMEASURES
A. Power is not on. 1. Power cable is disconnected. 1. Be sure to connect cable.
2. Misconnection of power cable. 2. Be sure to connect cable properly.
3. Burning of connecting part in power 3. Contact local distributor.
outlet or breaker.
B. Power is on, but the breaker has tripped 1. Electrical leakage. 1. Contact local distributor.
immediately. 1-1. Insulation deterioration of heating
element.
1-2. Water leakage at attaching portion of
hea ting element.
1-3. Water seeping into breaker.
2. Short circuit. 2. Contact local distributor.
2-1. Short circuit of heating element.
2-2. Short circuit of others.
C. After pressing START KEY, chamber 1. Breakage of heating element. 1. Contact local distributor.
temperature does not rise. 2. Burning of wire connected to heating 2. Contact local distributor.
element
D. When pressing START KEY, error sound 1. Completion process of previous operation 1. Press STOP KEY to reset completion
beeping and operation does not start. is not reset. process of previous operation.
Then press START KEY again.
E. The temperature in chamber rises, but it 1. Wrong electricity supply. 1. Check the required voltage of autoclave,
does not rise to setting temperature. then supply proper electricity to the
autoclave.
F. The temperature in chamber rises to the 1. Too much capacity of objects in chamber. 1. Quantity of objects to be put inside of
set temperature, but sterilization has chamber shall be less than 60% of capacity
not been completed due to inside the of the chamber. And each objects must
chamber does not become filled with keep enough space between them.
saturated steam. 2. Shortage of sterilization time. 2. Increase sterilization time appropriately
in case of sterilization of liQuid objects
and objects being large Quantity.
3. Residual air inside container prevents 3-1. In case of objects being put in a
penetration of steam. disposable plastic bag, Quantity of
objects to be put inside disposable
plastic bag shall be less than 60% of
capacity of the chamber. Disposable
plastic bag must have open mouth.
3-2. Containers shall not be superimposed
each other as possible as you can.
Then set its open mouth downward.
4. Air evacuation is insufficient due to bent 4. Do not bend exhaust hose.
exhaust hose.
G. Evacuating residual air does not stop 1. Breakage of exhaust valve. 1. Contact local distributor.
around 100"C. And the temperature 2. This autoclave is installed over 1OOOm 2. This autoclave can not be used over
does not rise. above sea level. 1000m above sea level.
H. Abnormal noise. 1. Dusts are attached to lid gasket. 1. Stop operation and take away dusts.
2. Lid is out of alignment forcibly. 2. Contact local distributor.
I. Steam leakage from lid. 1. Deterioration of lid gasket. 1. Contact local distributor.
(cracks in the gasket or hardening)
2. Lid and chamber are out of alignment 2. Contact local distributor.
.- caused by lid opened too much or moving
main body holding only lid.
J. No drainage. 1. Drain pipe is choked. 1. Contact local distributor.
K. Drainage does not stop. 1. Breakage of drain valve. 1. Contact local distributor.
(Piece of glass or other objects came into
valve and damaged "Teflon ball" of valve.)
41
TROUBLE CONDITION CAUSES COUNTERMEASURES
L. Safety valve activates. 1. Residual air is not evacuated enough and 1-1. Quantity of objects to be put inside of
(Steam blowing from bottom plate.) pressure in chamber rises abnormally due chamber shall be less than 60% of
to additional pressure by residual air. capacity of the chamber. And each
objects must keep enough space
between them.
1-2. In case of objects being put in a
disposable plastic bag, quantity of
objects to be put inside disposable
plastic bag shall be less than 60% of
capacity of the chamber. Disposable
plastic bag must have open mouth.
1-3. Containers shall not be superimposed
each other as possible as you can.
Then set its open mouth downward.
1-4. Do not bend exhaust hose.
2. Chamber pressure surpassed set 2. Contact local distributor.
pressure of safety valve due to the
malfunction of temperature sensor.
M. Pressure gauge does not indicate 1. Deterioration of pressure gauge. 1. Contact local distributor.
correct pressure value.
(Pressure gauge does not indicate OMPa
when the lid is opened.)
N. The lid lock lever can not be released 1. Power switch(breaker) is OFF. 1. Turn on power switch(breaker).
2. In operation or chamber temperature is 2. Press STOP KEY on the condition that the
over 80CC. TEMPERA TURE DISPLA Y indicates below
80CC after operation complete.
3. COMPLETE LAMP lights up after 3. Press STOP KEY to release the lid lock.
operation.
O. Movement of the lid lock lever is hard. 1. The lid lock lever is slid by force when lid is 1. Slide the lid lock lever slowly to close
not closed completely. position (left end) with pressing front part
of the lid cover.
(Especially TEMPERATURE DISPLA Y
indicates over 60CC.)
2. Insufficient space between lid and clamps. 2. Contact local distributor.
P. ErO is shown on the display. 1. Memory error of control unit. 1-1. Power off, then power on again.
1-2. If problem still exists after taking above
countermeasure, contact local
distributor.
Q. Er1 is shown on the display with alarm 1. Autoclave works without water in 1. Turn off power switch(breaker) and leave
and operation stops. chamber. autoclave as it is at least one hour.
(Er1 is shown when water lack preventive Because the chamber is high temperature.
device or over pressure detector works.) Then pour water into chamber.
2. Drain valve is opened. 2. Close drain valve.
3. Breakage of water lack preventive 3. Contact local distributor.
device or over pressure detector.
4. Over pressure detector activated.
4-1. Residual air in the capacious container 4-1. Containers shall not be superimposed
or disposable plastic bag swelled. each other as possible as you can.
Then set its open mouth downward.
In case of objects being put in a
disposable plastic bag, quantity of
objects to be put inside disposable
plastic bag shall be less than 60% of
capacity of the chamber. Disposable
plastic bag must have open mouth.
4-2. Air evacuation is insufficient due to bent 4-2. Do not bend exhaust hose.
exhaust hose.
42
TROUBLE CONDITION CAUSES COUNTERMEASURES
R. Er210r Er22 is shown on the display 1. Breakage of control sensor or object 1. Contact local distributor.
with alarm. temperature sensor.
S. Er3 is shown on the display with 1. In a sterilization process, temperature 1-1. The distance between chamber wall and
alarm. in chamber rises 20C higher than objects shall be kept more than 10mm.
setting temperature in certain period 1-2. Oleaginous objects must not be
of time. sterilized. (Oil causes unexpected rise in
This function is reset after dropping in ternperature.)
temperature within range of ±2OC from 1-3. Contact local distributor.
setting temperature.
T. Er4 is shown on the display with alarm. 1. Lid lock lever is not slid to close position 1. Slide the lid lock lever slowly to close
(left end) completely. position (left end) with pressing front part
of the lid cover.
(Especially TEMPERATURE DISPLA Y
indicates over 60OC.)
2. Lid lock lever is slid by force when the lid 2. Slide the lid lock lever slowly to close
is not closed completely. position (left end) with pressing front part
of the lid cover.
(Especially TEMPERATURE DISPLA Y
indica tes over 60OC.)
3. Lock position jolts out of alignment by 3. If problem still exists after taking above
doing same operation as above several countermeasures, contact local
times. distributor.
4. The lid lock lever is slid to the right side 4. Slide the lid lock lever to the close position
before you release the lid lock. (left end).
U. Er6 is shown on the display. 1. Battery for clock shutoff. 1. Turn off power switch(breaker) and turn
on it again. Set the right time.
(Refer to 5.13.1 Date and time setting)
2. IC in the control board is broken. 2. If problem still exists after taking above
countermeasure, contact local distributor.
V. Er1 is shown on the display. 1. Abnormal water supply - Sensor does not 1. Open faucet.
(only for automatic water supply detect water level for 3 minutes
function) due to the trouble of water supply.
(The city water faucet is closed)
W. Er8 is shown on the display. 1. The printer breaks. 1. Contact local distributor.
(only for printer function)
X. Er8 and Erg are shown on the display. 1. Paper-out condition on the printer. 1. Set the new paper.
(only for printer function) 2. Printer cover is opened. 2. Close printer cover.
43
8. Maintenance
.Regular Maintenance
8. 1 Daily maintenance
CD Replacement of water
Be sure to replace water in the chamber every day as follows.
Failure to this replacement reduce life span of chamber
due to choking pipes and/or oxidizing water.
~WARNING -----------------,
<9 Do not open lid until chamber pressure goes down to "OMPa".
Steam spouts out of chamber by residual pressure which may cause burn injury.
44
I -- ~ ~~ _===~====~==~==~==========~_=====~====
_
L!}. CAUTION
o When replacing water in chamber, check the chamber temperature
cause burn injury.
if it is cold enough, or it may
o over
Empty chamber, drain whole water in chamber, then close lid when autoclave is not used for
a week.
o Beplate.sureIf tothecheck water level in the chamber. In case of lack, pour water up to level of the base
water is impure, replace it.
o Water put in the chamber must be distilled one or purified one. Frequent cleaning of the chamber
is requested when using city water which generates scale in it.
o DoDo not spill water when pouring it. If you spill water, wipe off it.
o not use well water and industrial water which may cause corrosion and/or smudge.
Failure to use proper water may decrease life of the heating element and/or the chamber.
o Drainage must be carried out several times.
Overflow from chamber drain tray may cause burning.
o Use base plate hook when unloading base plate.(CL-40L only)
® Check of lid clamping unit
• How to check
1. Open the lid.
2. When sliding the lid lock lever slowly to the close position (left end), check all clamps
turn around to the locking positions. If the clamps are at locking positions, you can
see them from the slit of the silicon rubber pan. (Fig.1, Fig.2)
3. Slide the lid lock lever to the open position (right end), then close the lid.
4. Slide the lid lock lever to the close position with pressing the front part of the lid
cover, check if the lid lock lever is slid smoothly.
L!}.WARNING
o Do not use autoclave if clamps do not turn around to locking positions.
If you leave clamps unchanged, it may cause fatal accident. Contact local distributor.
o Do not use autoclave if the lid lock lever is not slid smoothly.
If you leave the lid lock lever unchanged, it may cause fatal accident. Contact local distributor.
o Do not open lid until chamber pressure goes down to "OMPa".
Steam spouts out of chamber by residual pressure which may cause burn injury.
8. 2 Monthly maintenance
LhWARNING -------------------,
(9 Do not touch power switch(breaker) by wet hand.
It may cause electric shock.
(9 Do not open lid until chamber pressure goes down to "OMPa".
Steam spouts out of chamber by residual pressure which may cause burn injury.
o If you find a crack or hardening on the lid gasket, please contact local distributor.
If you leave crack one unchanged. it may cause burst at worst as well as leakage of steam.
46
~CAUTION
~ Do not scratch heater in bottom of chamber when cleaning.
The water which infiltrates into heater may cause breakdown.
o When cleaning, check the chamber temperature if it is cold enough, or it may cause burn injury.
~ Do not pull out the lid gasket forcibly.
It may cause steam leakage.
o Use base plate hook when unloading base plate.(CL-40L only)
• Power connector
1. Make sure if the power cable does not have any damage and there are not
heavy things on the cable.
2. Make sure if the power connector of the autoclave is well connected to the
power receptacle.
3. If the dust is attached with power connector, clean up them .
• Breaker
1. Check if the breaker(power switch) is shut off
by pushing the test button by a sharp object (pen etc.),
~WARNING
~ Do not use autoclave if power cable and/or power connector are damaged.
Contact local distributor, or it may cause electric shock.
o If the dust is attached with power connector, clean up them.
It may cause fire by heat.
~ Do not touch the power connector and breaker by wet hand.
It may cause electric shock.
47
8. 3 Yearly maintenance
• How to check
1. Detach the right side panel using phillips-head screw driver.
2. Check if the water level of exhaust cooling tank has a range of below figure.
In case of water lack, pour water into the tank referring to
4. 5 Standard accessories.
t:
3. Attach the right side panel using phillips-head screw driver.
II \V
E/I\
E
0
"<t
•....
0
0
C"?
~CAUTION
(9 Do not pour too much water into exhaust cooling tank.
It makes overflow which may cause malfunction and/or breakage of autoclave.
o Be sure to stopple the funnel after pouring water into exhaust cooling tank.
Water in exhaust cooling tank reduces faster therefore frequent water supply is required.
® Periodic inspection
Make sure you inspect the autoclave at least once in a year referring to
"Periodic inspection manual" owned by a distributor in your region.
Failure to this periodic inspection, it may cause not only shortening life of autoclave
itself, but also a fatal accident by a defect of main parts.
*The lid cover and the operation panel may turn yellow due to aged deterioration by
ultra violet.
48
8. 4 Spare parts and replacement
Refer to the following list and replacement manual of main parts.
If you will order these spare parts, please contact local distributor.
SSR AQA411YL
Heater relay
2 13
Solidstate relay for steam heater DC24Y 25A
Main circuit relay
Pressure gauge
8 GK25-AI8
0.7MPa CL- DP series
49
8.4.2 Replacement of the lid gasket
Assembly
Nut
Out~
• Check that all parts are free from dirt and clean them if
necessary.
Inner lid
Lid gasket
\1/
/
./
<;
• This dimension has to be around equality.
o A TTEN7l0N: Apply ethyl alcohol that the lid gasket is reassembled in correct position.
50
8.4.3 Replacement of pipe heater
Wiring
1.Dismantlement
1) Pull off the 3pcs. of metal fittings"E" with pliers.
2) Open the both sides of the body and fall down sideways for easy work.
3) Disconnect wiring of the heater.
4) Open assembly nuts of the heater with spanners.(Opposite side 21)
5) Dismount the heater from bottom of the chamber.
6) Take off the sensor"C" from the heater.
2.Assembling
1) Install the new heater on bottom of the chamber.
Meantime, be sure to mount 2pcs. of the gaskets as above drawing.
2) Connect the wiring.
3.Checking
1) Fill water up to the base plate in the chamber and perform the allowable
maximum temperature.
Hereby, check current of the heater.
2) When the chamber temperature is reached, check steam leakage of the
mounted positions.
3) Check dimension and resistance of the Insulator.
51
(C
•
en
"'C
(t)
'a "
10. Disposal and dismantlement
Make the autoclave and accessories to be non-polluting condition before disposal and
dismantlement. If the power cable is connected to distribution board directly, beware of electric
shock.
~WARNING
o If the situation of autoclave correspond to following item(CDor®) when you dismantle and/or
dispose of the autoclave. make the autoclave and accessories to be non-polluting condition.
CD Part of autoclave is exposed to infectious substance or radioactive substance.
® Blood or other hazardous components for human body adhere to part of autoclave.
~ Do not disconnect the power cable when the electric power source is turned on.
It may cause electric shock.
53
....•.
....•. ....•.
....•. •
....•. "C
»
"C
CJ) CD
CD ::l
_.
r+
r+ _.
Q.
::l ><
(Q
::::!!
0
~
0
::l"
Q)
:4.
AUTOMATIC
WATER
, SUPPLY ,
L l
U1
~
(,
'1
DB
~~
~~
~
•...
'!l1!..l!.J!J.!.J
~ ~
--- ~ ~
~~
sa.
--
~., .-<- ---
-"., :~~
:r-
PS
_ 0
;;
-r:r---- "
~i
MB
~~
oi ,_
~.
~i
_ 0
rfij ffi:
-
_.
.
i
0
~~
SSRI po r:
~ (J)
-- ,~ ~ -~-
t-- ".
lS> cs - r:
--- d
ririri-ril • -if....
~
H.
·m
- PE
•. .,":..I,I~., ,'"
?
el8:-~-L
.
l
.,~
. \/1/
V V
;r-
OC>.v
~
§+
ACIOOVACMN
~.
" +
SPI Z;r-
p'
~~ •
.
PE
55
11. 3 Dangerous substances list
EXPLOSIVE
Nitro glycol, Nitroglycerine, Nitrocellulose, and other nitrate esters
FLAMMABLE
Metallic lithium, Metallic potassium, Metallic sodium, Yellow phosphorus
Normal hexane, Ethylene oxide, acetone, benzene, Methyl ethyl ketone, and other
flammable substances having a flash point of -30 deg. C. or higher but lower than
INFLAMMABLE o deg. C.
LIQUID:
Methanol, Ethanol, Xylene, Amyl acetate, and other flammable
substances having a flash point of 0 deg, C. or higher but lower than 30 deg, C.
Kerosene, Gas oil, Oil of turpentine, lsopentyl alcohol, Acetic acid, and other
flammable
substances having a flash point of 30 deg. C. or higher but lower than 65 deg. C.
FLAMMABLE Hydrogen, Acetylene, Ethylene, Methane, Propane, Butane, and other flammable
GAS substances which are assumed a gaseous state at 15 deg, C. and 1 atm
56
Memo
57
Responsibility
Please follow the instructions in this manual when using this equipment. .:
We have no responsibility for the accidents or breakdown of device if the operator doesn't follow this
manual when operating this autoclave.
Note
58
1-
QC-200-S-005-7001E-01
(S) When sodium hypochlorite is used for the cleaning of floor to clean germs,
pay attention not to adhere it to the exterior.
The exterior corrode and become rusty if it adhere to the exterior.