You are on page 1of 9

Original Paper

Caries Res 2014;48:566–574 Received: August 5, 2013


Accepted: November 26, 2013
DOI: 10.1159/000357596
Published online: July 29, 2014

A Comparative Study of Different


Radiographic Methods for Detecting
Occlusal Caries Lesions
Elif Tarım Ertas a Ebru Küçükyılmaz b Hüseyin Ertaş c Selçuk Savaş b
       

Meral Yırcalı Atıcı a   

Departments of a Oral and Maxillofacial Radiology, b Pedodontics and c Endodontics, Faculty of Dentistry,
     

Izmir Katip Çelebi University, Izmir, Turkey

Key Words a diagnostic D1 threshold. The Az values of the CBCT system


Computed tomography · Dental caries · Diagnosis · were found to be statistically higher than those of the other
Digital · Radiography imaging modalities at a diagnostic D3 threshold (p > 0.05);
no significant differences were found among the other im-
aging modalities. All radiographic methods showed similar
Abstract sensitivities, specificities and accuracy in detecting D1
Objectives: The aim of this in vitro study was to compare the threshold. The CBCT system showed higher sensitivity and
diagnostic accuracy of different radiographic imaging mo- accuracy in detecting dentine lesions. Conclusions: Within
dalities in detecting occlusal caries lesions. Materials and the limitations of this study, CBCT exhibited better perfor-
Methods: Under standardized conditions, 125 extracted hu- mance in detecting deep occlusal caries lesions than the
man permanent molar teeth with sound or occlusal caries other radiographic systems. © 2014 S. Karger AG, Basel
lesions were radiographed using a conventional film system
(F-speed), a direct digital imaging system (complementary
metal oxide semiconductor sensor), an indirect digital imag-
ing system (photostimulable phosphor plate) and a cone In recent years, the prevalence of dental caries has de-
beam computed tomography system (CBCT). Two observers clined worldwide in a remarkable way [Rathore et al.,
scored the resultant images for the presence or absence of 2012]. However, accurately diagnosing occlusal pit and
caries. Then, the teeth were histologically prepared and a fissure caries and detecting incipient lesions are important
definite diagnosis was determined by stereomicroscopic as- issues for clinicians [Souza-Zaroni et al., 2006]. Using a
sessment. The area under the receiver operating character- visual/tactile method when diagnosing occlusal caries is
istic curve (Az), sensitivity, specificity and accuracy of each the oldest and most common method preferred by den-
imaging modality were calculated, as well as the intra- and tists in clinical practice [Rathore et al., 2012]. Since occlu-
interexaminer reproducibility. Results: For both thresholds, sal caries can progress without visible breakdown of the
interexaminer agreement were higher for CBCT. For intraex- enamel structure [Poorterman et al., 2000], visual exami-
aminer agreement, observers had different scores for both nation alone is not always sufficient for diagnosing occlu-
thresholds, but the scores were generally higher for CBCT. sal caries. In the literature, although visual inspection has
Similar Az values were achieved with all imaging methods at been reported to be highly specific [Heinrich-Weltzien et
169.230.243.252 - 11/27/2014 3:27:14 PM

© 2014 S. Karger AG, Basel Elif Tarım Ertaş, DDS, PhD


UCSF Library & CKM

0008–6568/14/0486–0566$39.50/0 Department of Oral and Maxillofacial Radiology


Faculty of Dentistry, Izmir Katip Çelebi University
Downloaded by:

E-Mail karger@karger.com
TR–35180 Izmir (Turkey)
www.karger.com/cre
E-Mail dteliftarim @ yahoo.com
al., 2005; Reis et al., 2006], these methods are subjective, [Kamburoğlu et al., 2011], at a lower cost and with lower
which normally leads to low sensitivity and reliability re- absorbed doses than with the conventional computed to-
sults [Ketley and Holt, 1993; Costa et al., 2002]. mography used in medical radiology [Tyndall and Ra-
In order to overcome difficulties during diagnosis and thore, 2008], a new technology that uses a 2D sensor and a
enable better detection of occlusal caries, radiographic cone-shaped beam in place of the fan-shaped X-ray beam
examinations, especially conventional film and digital in- used for conventional computed tomography has been de-
traoral radiography, are the most easily accessible tech- veloped [Haiter-Neto et al., 2008]. The development of
niques for improving caries detection in routine clinical cone beam computed tomography (CBCT) has been revo-
practice [Kamburoğlu et al., 2010, 2011]. Recent develop- lutionary in that this technology [Rathore et al., 2012] of-
ments in imaging systems and the production of new sen- fers a number of potential advantages over conventional
sor types and advanced software programs offer increas- tomography, including easier image acquisition, higher
ing clinical advantages [Kayipmaz et al., 2011]. Although image accuracy, fewer artifacts, lower effective radiation
the image resolution of conventional film is superior to doses (up to 15 times lower than those of conventional
that of digital images [Parks and Williamson, 2002], this computed tomography scans), faster scan times and great-
technique requires more radiation to produce an image er cost effectiveness [Scarfe and Farman, 2008; Tyndall and
of diagnostic quality; therefore, many professionals are Rathore, 2008]. This technique could be applied in several
now replacing conventional film radiographs with digital dental diagnostic areas, such as implant treatment, cranio-
radiography, due to its many advantages [Cederberg et facial anomalies, endodontic treatment, orthodontics,
al., 1998; Paurazas et al., 2000; Williams, 2001; Kitagawa periodontology [Arai et al., 1999; Ziegler et al., 2002; Dan-
et al., 2003; Naoum et al., 2003; Kayipmaz et al., 2011]. forth, 2003] and caries diagnosis [Kayipmaz et al., 2011].
There are several digital radiographic systems currently In the literature, the results of studies using CBCT for car-
used in dental practice as an alternative to film-based radi- ies detection are not consistent with each other. In some
ography [Pontual et al., 2010]. The most common direct studies, promising results have been reported in the detec-
digital imaging systems use solid state sensors – either a tion of caries lesions [Haiter-Neto et al., 2008], while in
charge-coupled device (CCD) or a complementary metal another study, a limited CBCT system was found to be
oxide semiconductor (CMOS); indirect digital imaging superior to conventional film and storage phosphor radi-
systems use photostimulable phosphor (PSP) plates, also ography for the in vitro assessment of approximate caries
known as storage phosphor plates [Kamburoğlu et al., 2010]. lesion depth [Akdeniz et al., 2006]. Tsuchida et al. [2007]
Although digital systems have a number of advantages and Haiter-Neto et al. [2008] had raters score images of
[Wenzel, 1995; Hintze et al., 2002; Jacobsen et al., 2004], teeth made with a CBCT unit and found no benefit over
such as lower exposure dose, reduced working time from film for detecting incipient proximal surface caries [Young
image exposure to image display (no wet processing is et al., 2009]. Thus, there are no clear conclusions regarding
involved), lack of destroyed processing artifacts often ex- the value of CBCT for detecting caries in the literature. The
perienced with conventional film, and possible image purpose of the present study was to compare the caries di-
quality enhancements, such as contrast and density mod- agnostic accuracy of conventional film radiography (F-
ulation, which might increase diagnostic accuracy [Wen- speed film), a direct digital imaging system (CMOS sen-
zel, 1995, 2000; Hintze et al., 2002; Pai and Zimmerman, sor), an indirect digital imaging system (PSP plate) and a
2002], all of these systems are able to provide two-dimen- CBCT unit that provides high-resolution images with the
sional (2D) information about dental tissues and diseases smallest voxel size (0.075 mm) on the market for the in vi-
[Wenzel, 2000]. Another disadvantage of 2D radiographs tro determination of occlusal caries.
is that the apparent depth can also vary as a function of
X-ray beam angulation [van der Stelt et al., 1989; Chad-
wick et al., 1999] and radiographic density [Versteeg et Materials and Methods
al., 1997], which can lead to variations in perception. The
perceived lesion depth can lead dentists to erroneously This study was approved by the Research Ethics Committee of
believe that caries has progressed or regressed, resulting Izmir Katip Çelebi University (registration number 2013-39).
in unnecessary restorative intervention or delay in treat-
Sample Selection
ment [Akdeniz and Gröndahl, 2005]. A total of 125 extracted permanent molar teeth exhibiting com-
Due to the high demand for a technique that can pro- plete root formation were included in this study. Altered physical
vide three-dimensional (3D) data at the tooth level properties in the tooth structure, large cavitated surfaces and den-
169.230.243.252 - 11/27/2014 3:27:14 PM

Different Radiographic Methods for Caries Res 2014;48:566–574 567


UCSF Library & CKM

Detecting Caries DOI: 10.1159/000357596


Downloaded by:
tal restorations were not included in the study sample. The occlu- at a fixed 110 kVp setting, automated adjusted milliamperes and a
sal surfaces ranged from sound to varying degrees of fissure, dis- scan time of 36 s. The volumetric data from the CBCT system were
coloration and possible microscopic breakdown of the surface reconstructed and sectioned into 0.075-mm pieces in the mesio-
structure; however, none of the teeth showed macroscopic signs of distal tooth plane.
cavity formation with exposure into the dentine. The teeth were
stored in 0.1% thymol solution for less than 3 months from the Evaluation of Radiographic Methods
time of extraction. For the study, the teeth were radiographed us- The images were organized into groups of images from a par-
ing four different radiographic methods. ticular exposure setting and system. Two observers (one pedodon-
tist and one oral and maxillofacial radiologist) independently
Conventional Film Radiography viewed the image groups in random order. The oral and maxillofa-
A bitewing technique using size 2 (3 × 4 mm) F-speed film (CF- cial radiologist with 10 years of experience in image interpretation
SPEEDX, Medex Medical Imaging, Nice, France) and standard- provided a training session to familiarize the other observer with
ized bitewing projection geometry was used. The intraoral X-ray the presentation of all imaging methods and was given brief infor-
unit (eXTtend; MyRay, Imola, Italy) was operated at 65 kV, 7 mA mation about the use of the NNT viewing software (version 3.10,
with 2.5 mm aluminium equivalent filtration. Focus-to-film dis- QR srl, Verona, Italy) on how to observe the CBCT images. Then
tance was 30 cm. Each tooth was mounted in a block of silicone the observers discussed and scored a number of teeth with carious
paste to ensure a reproducible geometry. While the vinyl polysi- lesions with different radiographic methods for calibration. The
loxane putty was still soft, the film holder was pressed into it, and digital and CBCT images were displayed using the dedicated soft-
once hardened, the putty allowed quick realignment of the speci- ware of each imaging system incorporated into the same computer.
men as well as CMOS sensor, PSP plate and F-speed conventional Observation conditions were optimized by using the same com-
films. A 20-mm-thick soft tissue equivalent Plexiglas block was puter monitor to display the images; the display ratio was 1:1. View-
placed close to the tooth and facing the X-ray tube to simulate scat- ing distance was kept constant, at about 50 cm, for all observers, and
ter radiation and beam attenuation from soft tissues. The F-speed the lights were dimmed during the observations. The observers
films were exposed for 0.25 s to generate an optimal density sub- were not given the option of performing any image enhancements
jectively for caries detection [Sogur et al., 2011]. The exposed films to avoid the production of a variety of different digital images.
were automatically processed (Velopex Extra-X; Medivance In- Conventional film radiographs were examined on a light box
struments Ltd., London, UK) using fresh Kodak developer and and at 2× magnification. A period of at least 1 day separated each
fixer solutions. viewing session. All tooth surfaces were examined for the presence
of carious lesions on the occlusal surfaces, using a five-point con-
Direct Digital Radiography – CMOS Sensor fidence rating scale: 0 = no caries; 1 = radiolucency extending to
Standardized images of the teeth were obtained using the same the outer half of the enamel; 2 = radiolucency extending to the in-
intraoral X-ray unit, but with a size 1 (37 × 24 mm) CMOS sensor ner half of the enamel; 3 = radiolucency extending to the outer half
(DIGORA Toto; SOREDEX, Milwaukee, Wisc., USA) with the of the dentine; 4 = radiolucency extending to the inner half of the
same standardized projection geometry. The CMOS sensors were dentine. Intraobserver agreement was assessed by having each ob-
exposed for 0.12 s to generate an optimal density subjectively for server view all images twice, with a 2-week interval between view-
caries detection [Sogur et al., 2011]. The direct digital imaging sys- ing to eliminate memory bias.
tem provides three types of diagnostic modes (dentoenamel, peri-
odontal and endodontic) at two levels (high and low) of spatial Histological Validation
resolution. In this study, the exposures with the CMOS sensor were The teeth were individually embedded in acrylic (Vipcril; Vipi,
performed in the dentoenamel high-resolution mode, which was São Paulo, Brazil) and serially sectioned into 700-μm-thick sections
recommended by the manufacturer to provide high-contrast im- in the mesiodistal direction, using a water-cooled 200-μm diamond
ages for caries detection. band. The histological examination was performed by one of the
study authors (E.K.), with experience from several previous in vitro
Indirect Digital Radiography – PSP Plate System studies, using a stereomicroscope with a magnification of 10–20×
Standardized images of the teeth were obtained using the same under reflected light (Olympus SZ61, Tokyo, Japan).
intraoral X-ray unit, but with size 2 (31 × 41 mm) VistaScan blue Caries was defined as being present when demineralization was
storage phosphor plates (Dürr Dental, Bietigheim-Bissingen, observed, seen as a white or discolored (yellow/brown) area. The
Germany) with the same standardized projection geometry. The histological criteria for caries lesion depth were: 0 = no caries; 1 =
PSP plates were exposed for 0.12 s to generate an optimal density demineralization extending to the outer half of the enamel; 2  =
subjectively for caries detection [Sogur et al., 2011]. The plates demineralization extending to the inner half of the enamel; 3  =
were later scanned in a VistaScan scanner. The PSP plates were demineralization extending to the outer half of the dentine; 4 =
stored in lightproof envelopes during the exposure and scanned demineralization extending to the inner half of the dentine. As a
immediately after exposure using the VistaScan scanner. The high- measure of histological assessment reliability, all of the sections
resolution scan mode was selected from the scanner setup menu, were re-examined after 10 days. Almost perfect agreement was
as recommended by the manufacturer for most diagnostic tasks. reached between the first and second histological examinations.

CBCT Statistical Analysis


Each tooth block was also radiographed using a NewTom 5G The analyses of the caries detected on the radiographs were
CBCT system (Verona, Italy) with a 6 × 6 cm field of view in the performed at two different thresholds: enamel and dentine caries
high-resolution denture scan mode. The voxel size was 0.075 mm3 lesions (D1 threshold; sound vs. decayed) and dentine caries le-
169.230.243.252 - 11/27/2014 3:27:14 PM

568 Caries Res 2014;48:566–574 Tarım Ertas/Küçükyılmaz/Ertaş/Savaş/


UCSF Library & CKM

DOI: 10.1159/000357596 Yırcalı Atıcı


Downloaded by:
Color version available online
among each other, receiver operating characteristic (ROC) analy-
ses were performed, and the area under the ROC curve (Az) at the
1.0 D1 and D3 thresholds was calculated. Comparison of ROC curves
was calculated with the statistics program MedCalc 9.3.0.0 (Mar-
iakerke, Belgium). The analyses were performed using the SPSS
0.8 statistics program for Windows, version 11.5 (SPSS Inc., Chicago,
Ill., USA). For all statistical analyses, the level of significance was
p < 0.05.
0.6
Sensitivity

Results
0.4
Distribution of Lesions
Source of the curve
Conventional film A total of 125 occlusal surfaces were examined in the
0.2 PSP study. According to the histological examination, the sta-
CMOS
CBCT tus of the 125 occlusal surfaces was: 31 (24.8%) sound, 11
Reference line (8.8%) with caries lesions extending into the outer half of
0 the enamel, 47 (37.6%) with caries lesions extending into
0 0.2 0.4 0.6 0.8 1.0
1 – Specificity
the inner half of the enamel, 29 (23.2%) with caries lesions
extending into the outer half of the dentine, and 7 (5.6%)
with caries lesions extending into the inner half of the
dentine.
Fig. 1. Area under the ROC curve values of all radiographic detec-
tion methods at the D1 threshold. Diagonal segments are pro-
duced by ties. Occlusal Lesions at the D1 Threshold
The sensitivity scores of the CBCT images of occlusal
lesions at the D1 threshold were higher than those of the
Table 1. Sensitivity, specificity, accuracy and Az scores of all radio-
other imaging methods, while the specificity scores were
graphic detection methods at the D1 threshold lowest. However, the diagnostic accuracy of all radio-
graphic systems was assessed using the Az, and similar
Test method Az Sensitivity Specificity Accuracy accuracy scores were obtained with all imaging methods
at the D1 threshold. A comparison of the Az values showed
Conventional film 0.851 0.798 0.903 0.824 that the differences between the radiographic methods
PSP 0.824 0.713 0.935 0.768
CMOS 0.834 0.830 0.839 0.832 were not significant (p = 0.401 between conventional film
CBCT 0.801 0.926 0.677 0.864 and PSP plate, p = 0.725 between conventional film and
CMOS sensor, p = 0.338 between conventional film and
CBCT, p = 0.840 between PSP plate and CMOS sensor,
p  = 0.663 between PSP plate and CBCT, p  = 0.447 be-
tween CMOS sensor and CBCT) (fig. 1). Table 1 shows
sions (D3 threshold), based on the histological evaluation as the
gold standard. The appropriate cutoff point for the D1 threshold the sensitivity, specificity, accuracy and Az values of all
was a score of 1 or above, considering the gold standard scores 1, radiographic methods at the D1 threshold.
2 and 3 as evidence of disease for all radiographic methods. The
appropriate cutoff point for the D3 threshold was a score of 3 or Occlusal Lesions at the D3 Threshold
above, considering the gold standard scores 3 and above as evi- The sensitivity scores of the CBCT images of occlusal
dence of disease for all radiographic methods. Data were presented
and analyzed separately for each examiner. lesions at the D3 threshold were higher than those of the
Inter- and intraexaminer reliabilities were calculated using Co- other imaging methods. At the D3 threshold, similar
hen’s kappa test after collapsing the results into two categories: D1 specificity and accuracy scores were obtained with all im-
and D3 thresholds. For each observer and each radiographic mo- aging methods. A comparison of the Az values showed
dality, the sensitivity (cumulative percentage of carious enamel le- significant differences between CBCT and the other ra-
sions identified among those that had carious lesions), specificity
(cumulative percentage of sound surfaces identified among those diographic methods (p  = 0.029 between conventional
who had sound surfaces) and accuracy (percentage of correct film and CBCT, p = 0.023 between PSP plate and CBCT,
scores) were computed. To compare the performance of methods p = 0.014 between CMOS sensor and CBCT). There were
169.230.243.252 - 11/27/2014 3:27:14 PM

Different Radiographic Methods for Caries Res 2014;48:566–574 569


UCSF Library & CKM

Detecting Caries DOI: 10.1159/000357596


Downloaded by:
Color version available online
Table 2. Sensitivity, specificity, accuracy and Az scores of all radio-
1.0 graphic detection methods at the D3 threshold

Test method Az Sensitivity Specificity Accuracy


0.8
Conventional film 0.739a 0.500 0.978 0.840
PSP 0.730a 0.472 0.989 0.840
0.6
CMOS 0.717a 0.444 0.989 0.832
0.875b
Sensitivity

CBCT 0.806 0.944 0.904

Different superscript letters show a significant difference be-


0.4
tween methods.
Source of the curve
Conventional film
0.2 PSP
CMOS
CBCT Table 3. Interexaminer kappa scores (with standard errors) of all
Reference line radiographic detection methods at the D1 and D3 thresholds
0
0 0.2 0.4 0.6 0.8 1.0
Test method D1 D3
1 – Specificity

Conventional film 0.473 (0.071) 0.822 (0.070)


PSP 0.654 (0.068) 0.804 (0.071)
Fig. 2. Area under the ROC curve values of all radiographic detec- CMOS 0.518 (0.083) 0.763 (0.079)
tion methods at the D3 threshold. Diagonal segments are pro- CBCT 0.833 (0.061) 0.980 (0.020)
duced by ties.

no differences among the other methods at the D3 thresh- that intraoral analogue X-ray film, direct and indirect
old (p = 0.899 between conventional film and PSP plate, digital X-ray modalities (CMOS sensor and PSP plate)
p = 0.732 between conventional film and CMOS sensor, and CBCT performed similarly in detecting occlusal car-
p = 0.560 between PSP plate and CMOS sensor) (fig. 2). ies at the D1 threshold, and that CBCT performed better
Table 2 shows the sensitivity, specificity, accuracy and Az at the D3 threshold.
values of all radiographic methods at the D3 threshold. In the present study, the occlusal surfaces ranged
In the kappa analysis of interexaminer agreement be- from sound to varying degrees of fissure, discoloration
tween the observers, conventional film radiography had and possible microscopic breakdown of the surface
the lowest scores at the D1 threshold, while the CBCT im- structure; however, none of the teeth showed macro-
ages had higher scores at both the D1 and D3 thresholds scopic signs of cavity formation with exposure into den-
compared with the other methods. At the D3 threshold, tine, as we believe that if diagnostic differences be-
repeatability between the observers was similar for all im- tween radiographic systems are to be found, their accu-
aging methods and higher than the scores obtained at the racy in detecting subtle pathological changes should be
D1 threshold (table 3). tested.
There was a high level of intraexaminer agreement be- An in vitro model was preferred in this radiological
tween the two assessments of the observers for the CBCT study, as ideal patient positioning is not always possible
images at the D3 threshold. The overall intraexaminer in an in vivo study and absolute reproducibility is limited.
scores were compatible between the observers at both In addition, image quality may vary from one patient to
thresholds. Table 4 shows the intraexaminer agreement another. Another advantage of this in vitro model is that
scores of both observers at both thresholds. the occlusal surfaces can be exposed to X-rays repeatedly
and ideal positioning of the specimen with exact repro-
ducibility is possible. However, the results of the present
Discussion study might not correlate with clinical situations. This is
because in clinical settings, movement of the patient de-
The present research focused mainly on the diagnostic creases image resolution, restorations in the occlusal
accuracy and reliability of observers regarding images ob- plane can cause metallic streaking artifacts in the occlusal
tained using different X-ray systems. This study found plane, and the other head and neck structures can result
169.230.243.252 - 11/27/2014 3:27:14 PM

570 Caries Res 2014;48:566–574 Tarım Ertas/Küçükyılmaz/Ertaş/Savaş/


UCSF Library & CKM

DOI: 10.1159/000357596 Yırcalı Atıcı


Downloaded by:
Table 4. Intraexaminer kappa scores (with standard errors) of all radiographic detection methods at the D1 and
D3 thresholds

Test method D1 D3
examiner 1 examiner 2 examiner 1 examiner 2

Conventional film 0.872 (0.044) 0.610 (0.073) 0.546 (0.111) 0.647 (0.100)
PSP 0.824 (0.051) 0.727 (0.059) 0.548 (0.106) 0.834 (0.072)
CMOS 0.513 (0.077) 0.714 (0.064) 0.666 (0.093) 0.965 (0.035)
CBCT 0.776 (0.067) 0.935 (0.037) 0.918 (0.040) 0.938 (0.035)

in scattering into the field of interest, thereby reducing a perfect test, and anything near 0.5 is a poor test result
contrast [Ricketts et al., 1995]. [Hintze et al., 2003]. In the present study, the AUC values
The accuracy of 2D systems is well established in the of the four radiographic methods were around 0.80–0.85
literature [Rathore et al., 2012]; however, the accuracy of at the D1 threshold, indicating that none of the methods
both intraoral analogue X-ray film and digital X-ray sen- was superior to the others and that they performed well
sor measurements is limited by the 2D nature of the tech- at the D1 threshold. In addition, the CBCT AUC values
nology. The lower radiation doses, superimposition of were higher than those of the other methods at the D3
anatomical structures and patient-related factors that af- threshold, and the scores were close to those of the other
fect caries diagnosis are inevitable factors with 2D sys- methods.
tems [Kamburoğlu et al., 2011]. In addition, the differ- In recent years, several studies have been carried out
ences in mass between small, incipient lesions and the to evaluate the accuracy of CBCT in detecting caries le-
surrounding tissues are so small that they do not reflect sions on proximal and occlusal surfaces, with varying re-
density differences with 2D images [Akdeniz and Grön- sults [Haiter-Neto et al., 2008; Young et al., 2009;
dahl, 2005]. In the past decade, the use of CBCT in den- Kamburoğlu et al., 2010; Qu et al., 2011]. In 2007, Kalath-
tistry has become more and more widespread, as the ingal et al. published a study comparing a CBCT device
CBCT technology overcomes the irradiation geometry (SIDEXIS sensor; Sirona Dental Systems, Bensheim,
problems that can cause errors in caries diagnosis with 2D Germany) and conventional film radiography in the de-
imaging [Akdeniz and Gröndahl, 2005]. tection of proximal caries, and their results showed no
In this study, the observers were calibrated for each differences between the two methods. In the same year,
radiographic interpretation, but some differences be- Tsuchida et al. reported that the accuracy of the 3D Ac-
tween the observers still occurred, which may be due to cuitomo in evaluating incipient proximal caries was not
their levels of experience and training with these radio- superior to that of intraoral films. Similar to their previ-
graphic methods. ous finding, Haiter-Neto et al. [2008] reported no differ-
In the present study, ROC analysis was used to evalu- ences in specificity or overall true scores among the meth-
ate the diagnostic performance of four radiographic ods when comparing the diagnostic accuracy of two
methods. In this analysis, significant differences among CBCT systems – NewTom 3G (Verona, Italy) and 3DX
the areas under the ROC curves of the competing tech- Accuitomo – with one digital PSP (DIGORA fmx) and
niques were compared [Kantor et al., 1989], and it was one conventional film system (Kodak Insight) in detect-
found that the area under the curve (AUC) reflected di- ing occlusal caries.
agnostic performance more comprehensively than sensi- Contrary to previous findings, Young et al. [2009] re-
tivity and specificity did, which are considered only one ported that Accuitomo 3DX images were superior to
cutoff point [Kositbowornchai et al., 2004]. The ROC CCD projection images in detecting lesions extending
curve distinguishes between the inherent capacities of the into the dentine on occlusal and proximal surfaces. How-
observers to under- and overread when interpreting im- ever, the authors concluded that the 3DX images did not
aging; therefore, this analysis provides the most mean- offer significantly superior information for detecting
ingful approach to comparing the diagnostic perfor- proximal surface caries limited to the enamel compared
mance of two or more different imaging modalities [Ra- with CCD projection imaging. The possible explanation
thore et al., 2012]. In this analysis, an area of 1 represents for these differences in study results is that both Tsuchida
169.230.243.252 - 11/27/2014 3:27:14 PM

Different Radiographic Methods for Caries Res 2014;48:566–574 571


UCSF Library & CKM

Detecting Caries DOI: 10.1159/000357596


Downloaded by:
et al. [2007] and Haiter-Neto et al. [2008] used a popula- [2008], who reported that the sensitivity of the two CBCT
tion of teeth in which the vast majority of the proximal systems (Accuitomo and NewTom) was high, but that
surface lesions were limited to the enamel, whereas Young their specificity was low, resulting in an increase in false-
et al. [2009] evaluated both enamel and dentinal lesions positive results. Similarly, in their study, Young et al.
equally. The different distribution of lesion depths would [2009] found more false-positive results in intact teeth,
make detecting lesions more challenging, especially when thus resulting in significantly lower average specificity
they are located in enamel. In a recent study by Rathore scores for the 3DX images than for the CCD images. A
et al. [2012], no statistically significant differences were possible explanation for this might be the beam-harden-
reported between the Sirona Galileos (Sirona Dental Sys- ing artifacts that appear in the pericoronal area and may
tems) CBCT and conventional film radiography in de- mislead observers in distinguishing pathologies with low
tecting occlusal caries. density, such as caries lesions [Tsuchida et al., 2007]. In
The results of the present study are consistent with the addition, sound dentine under some cusps may appear
data in the literature in that there were no differences be- artifactually more radiolucent than the surrounding den-
tween the four imaging methods in detecting occlusal tine on the CBCT images, and this difference between
caries at the D1 threshold (sound/decayed), whereas enamel and dentine density might lead observers to mis-
CBCT was superior in detecting dentine lesions. interpret these areas as dentinal occlusal caries, thereby
CBCT systems offer a feature whereby an area can be elevating the sensitivity scores while depressing the spec-
evaluated in three planes at the same time (axial, sagittal ificity scores [Young et al., 2009; Kayipmaz et al., 2011].
and coronal), and sections can be created with different It is possible that better training of observers regarding
slice thicknesses and intervals. In the present study, we this misinterpretation might improve the observers’ spec-
chose the mesiodistal plane to validate occlusal surfaces, ificity scores without significantly depressing their sensi-
as in this section, occlusal surface demineralization can tivity scores [Young et al., 2009].
be evaluated clearly [Haiter-Neto et al., 2008]. However, In clinical practice, when referring a patient for CBCT
it is critical that if the operator fails to choose the slice examination, one of the most important issues to keep in
that shows the deepest extension of the lesion, he/she mind is the effective radiation dose. Although the effec-
may misdiagnose the lesions [Akdeniz and Gröndahl, tive dose is somewhat lower than that of medical com-
2005]. It has been suggested that the small voxel sizes puted tomography, the dose is relatively higher than in
and thin slices are the potential advantages of a CBCT conventional imaging alternatives and intraoral exami-
system in caries diagnosis [Kayipmaz et al., 2011]. The nations. It is critical that the potential patient benefits
CBCT device used in the present study (NewTom 5G) from a radiographic examination be balanced against the
offers the smallest voxel size (0.075 mm3) available on risk of exposure to ionizing radiation [Ludlow et al.,
the market, and this is the first study in the literature to 2003], and it is fundamental that diagnostic radiology and
evaluate accuracy at the smallest voxel size. However, CBCT procedures should be reserved for selected cases
no statistically significant differences were found among [Farman, 2005]. It should also be noted that CBCT is not
the four radiographic methods in the diagnosis of oc- considered suitable for routine caries diagnosis, because
clusal caries lesions at the D1 threshold, even with the the large majority of patients have metallic restorations
smallest voxel size. Consistent with this finding, in a re- that cause artifacts due to beam hardening and scatter
cent study, Kamburoğlu et al. [2010] compared intra- that simulates recurrent caries, and because the quality of
oral digital CCD sensor images and CBCT images ob- CBCT images can be seriously affected by patient motion
tained at different voxel resolutions in the in vitro detec- [Clifton et al., 1998].
tion of occlusal caries, and the authors concluded that In conclusion, the NewTom 5G CBCT exhibited high-
there were no statistically significant differences be- er sensitivity for detecting occlusal lesions than the intra-
tween ultra-resolution images and high- and low-reso- oral systems did, but the overall accuracy scores were sim-
lution images. ilar at the D1 threshold. CBCT was more accurate in de-
In the present study, although the CBCT images had tecting lesions at the D3 threshold than the intraoral
the highest sensitivity scores, the CBCT specificity scores systems. However, it can be concluded that because of the
were lowest at the D1 threshold, which means that CBCT possible disadvantages of CBCT systems in clinical prac-
performs better when detecting true carious lesions, but tice, including relatively higher effective radiation doses
weakly when detecting real sound surfaces. The present as well as possible misdiagnosis due to the artifacts caused
finding is consistent with the result of Haiter-Neto et al. by movement and metallic artifacts, intraoral radiogra-
169.230.243.252 - 11/27/2014 3:27:14 PM

572 Caries Res 2014;48:566–574 Tarım Ertas/Küçükyılmaz/Ertaş/Savaş/


UCSF Library & CKM

DOI: 10.1159/000357596 Yırcalı Atıcı


Downloaded by:
phy methods are found to be sufficient and have practical, the manuscript: E. Tarım Ertas. Critical revision of the manu-
economical and low-dose features that are valuable for script for important intellectual content: H. Ertaş, E. Tarım
Ertas. Approval of the version of the manuscript to be pub-
the routine assessment of caries. lished: E. Tarım Ertas, H. Ertaş, E. Küçükyılmaz, S. Savaş, M.
Yırcalı Atıcı.
Author Contributions

Conception and design of study: E. Tarım Ertas. Acquisition Disclosure Statement


of data: S. Savaş, M. Yırcalı Atıcı. Analysis and/or interpretation
of data: E. Tarım Ertas, E. Küçükyılmaz, H. Ertaş. Drafting of The authors declare that they have no conflicts of interest.

References
Akdeniz BG, Gröndahl HG: Proximal caries rescence and visual examination: an in vivo tor and charge-coupled device intraoral X-ray
depth: agreement between limited cone-beam comparison. Eur J Oral Sci 2005;113:494–498. detectors using subjective image quality. Den-
CT, storage phosphore and film radiography. Hintze H, Frydenberg M, Wenzel A: Influence of tomaxillofac Radiol 2003;32:408–411.
Hacettepe Dishekimligi Fakultesi Dergisi number of surfaces and observers on statisti- Kositbowornchai S, Basiw M, Promwang Y, Mor-
2005;29:7–12. cal power in a multiobserver ROC radio- agorn H, Sooksuntisakoonchai N: Accuracy
Akdeniz BG, Gröndahl HG, Magnusson B: Accu- graphic caries detection study. Caries Res of diagnosing occlusal caries using enhanced
racy of proximal caries depth measurements: 2003;37:200–205. digital images. Dentomaxillofac Radiol 2004;
comparison between limited cone beam com- Hintze H, Wenzel A, Frydenberg M: Accuracy of 33:236–240.
puted tomography, storage phosphor and caries detection with four storage phosphor Ludlow JB, Davies-Ludlow LE, Brooks SL: Do-
film radiography. Caries Res 2006; 40: 202– systems and E-speed radiographs. Dento- simetry of two extraoral direct digital imaging
207. maxillofac Radiol 2002;31:170–175. devices: NewTom cone beam CT and Ortho-
Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shi- Jacobsen JH, Hansen B, Wenzel A, Hintze H: Re- phos Plus DS panoramic unit. Dentomaxillo-
noda K: Development of a compact computed lationship between histological and radio- fac Radiol 2003;32:229–234.
tomographic apparatus for dental use. Dento- graphic caries lesion depth measured in im- Naoum HJ, Chandler NP, Love RM: Convention-
maxillofac Radiol 1999;28:245–248. ages from four digital radiography systems. al versus storage phosphor-plate digital im-
Cederberg RA, Tidwell E, Frederiksen NL, Ben- Caries Res 2004;38:34–38. ages to visualize the root canal system con-
son BW: Endodontic working length assess- Kalathingal SM, Mol A, Tyndall DA, Caplan DJ: trasted with a radiopaque medium. J Endod
ment. Comparison of storage phosphor digi- In vitro assessment of cone beam local com- 2003;29:349–352.
tal imaging and radiographic film. Oral Surg puted tomography for proximal caries detec- Pai SS, Zimmerman JL: Digital radiographic im-
Oral Med Oral Pathol Oral Radiol Endod tion. Oral Surg Oral Med Oral Pathol Oral aging in dental practice. Dent Today 2002;21:
1998;85:325–328. Radiol Endod 2007;104:699–704. 56–61.
Chadwick BL, Dummer PM, van der Stelt PF: Kamburoğlu K, Kurt H, Kolsuz E, Öztaş B, Tatar Parks ET, Williamson GF: Digital radiography: an
The effect of alterations in horizontal X-ray I, Çelik HH: Occlusal caries depth measure- overview. J Contemp Dent Pract 2002; 3: 23–
beam angulation and bucco-lingual cavity ments obtained by five different imaging 39.
width on the radiographic depth of approxi- modalities. J Digit Imaging 2011; 24: 804– Paurazas SB, Geist JR, Pink FE, Hoen MM, Stei-
mal cavities. J Oral Rehabil 1999; 26: 292– 813. man HR: Comparison of diagnostic accuracy
301. Kamburoğlu K, Murat S, Yüksel SP, Cebeci AR, of digital imaging by using CCD and CMOS-
Clifton TL, Tyndall DA, Ludlow JB: Extraoral ra- Paksoy CS: Occlusal caries detection by using APS sensors with E-speed film in the detec-
diographic imaging of primary caries. Dento- a cone-beam CT with different voxel resolu- tion of periapical bony lesions. Oral Surg Oral
maxillofac Radiol 1998;27:193–198. tions and a digital intraoral sensor. Oral Surg Med Oral Pathol Oral Radiol Endod 2000;89:
Costa AM, Yamaguti PM, De Paula LM, Bezerra Oral Med Oral Pathol Oral Radiol Endod 356–362.
AC: In vitro study of laser diode 655 nm diag- 2010;109:e63–e69. Pontual AA, de Melo DP, de Almeida SM, Bosco-
nosis of occlusal caries. ASDC J Dent Child Kantor ML, Zeichner SJ, Valachovic RW, Reiskin lo FN, Haiter Neto F: Comparison of digital
2002;69:249–253, 233. AB: Efficacy of dental radiographic practices: systems and conventional dental film for the
Danforth RA: Cone beam volume tomography: a options for image receptors, examination se- detection of approximal enamel caries. Den-
new digital imaging option for dentistry. J Ca- lection, and patient selection. J Am Dent As- tomaxillofac Radiol 2010;39:431–436.
lif Dent Assoc 2003;31:814–815. soc 1989;119:259–268. Poorterman JH, Weerheijm KL, Groen HJ, Kals-
Farman AG: ALARA still applies. Oral Surg Oral Kayipmaz S, Sezgin OS, Saricaoglu ST, Can G: An beek H: Clinical and radiographic judgement
Med Oral Pathol Oral Radiol Endod 2005; in vitro comparison of diagnostic abilities of of occlusal caries in adolescents. Eur J Oral Sci
100:395–397. conventional radiography, storage phosphor, 2000;108:93–98.
Haiter-Neto F, Wenzel A, Gotfredsen E: Diagnos- and cone beam computed tomography to de- Qu X, Li G, Zhang Z, Ma X: Detection accuracy of
tic accuracy of cone beam computed tomog- termine occlusal and approximal caries. Eur J in vitro approximal caries by cone beam com-
raphy scans compared with intraoral image Radiol 2011;80:478–482. puted tomography images. Eur J Radiol 2011;
modalities for detection of caries lesions. Ketley CE, Holt RD: Visual and radiographic di- 79:e24–e27.
Dentomaxillofac Radiol 2008;37:18–22. agnosis of occlusal caries in first permanent Rathore S, Tyndall D, Wright J, Everett E: Ex vivo
Heinrich-Weltzien R, Kuhnisch J, Ifland S, molars and in second primary molars. Br comparison of Galileos cone beam CT and in-
Tranaeus S, Angmar-Mansson B, Stosser L: Dent J 1993;174:364–370. traoral radiographs in detecting occlusal car-
Detection of initial caries lesions on smooth Kitagawa H, Scheetz JP, Farman AG: Comparison ies. Dentomaxillofac Radiol 2012; 41: 489–
surfaces by quantitative light-induced fluo- of complementary metal oxide semiconduc- 493.
169.230.243.252 - 11/27/2014 3:27:14 PM

Different Radiographic Methods for Caries Res 2014;48:566–574 573


UCSF Library & CKM

Detecting Caries DOI: 10.1159/000357596


Downloaded by:
Reis A, Mendes FM, Angnes V, Angnes G, Grande Tsuchida R, Araki K, Okano T: Evaluation of a Versteeg KH, Sanderink GC, Velders XL, van
RH, Loguercio AD: Performance of methods limited cone-beam volumetric imaging sys- Ginkel FC, van der Stelt PF: In vivo study of
of occlusal caries detection in permanent tem: comparison with film radiography in de- approximal caries depth on storage phosphor
teeth under clinical and laboratory condi- tecting incipient proximal caries. Oral Surg plate images compared with dental X-ray
tions. J Dent 2006;34:89–96. Oral Med Oral Pathol Oral Radiol Endod film. Oral Surg Oral Med Oral Pathol Oral Ra-
Ricketts DN, Kidd EA, Smith BG, Wilson RF: 2007;104:412–416. diol Endod 1997;84:210–213.
Clinical and radiographic diagnosis of occlu- Tyndall DA, Rathore S: Cone-beam CT diagnos- Wenzel A: Current trends in radiographic caries
sal caries: a study in vitro. J Oral Rehabil 1995; tic applications: caries, periodontal bone as- imaging. Oral Surg Oral Med Oral Pathol
22:15–20. sessment, and endodontic applications. Dent Oral Radiol Endod 1995;80:527–539.
Scarfe WC, Farman AG: What is cone-beam CT Clin North Am 2008;52:825–841. Wenzel A: Digital imaging for dental caries. Dent
and how does it work? Dent Clin North Am van der Stelt PF, Ruttiman UE, Webber RL, Clin North Am 2000;44:319–338.
2008;52:707–730. Heemstra P: In vitro study into the influence Williams CP: Digital radiography sensors: CCD,
Sogur E, Baksı BG, Orhan K, Paksoy SC, Dogan S, of X-ray beam angulation on the detection of CMOS, and PSP. Pract Proced Aesthet Dent
Erdal YS, Mert A: Effect of tube potential and artificial caries defects on bitewing radio- 2001;13:395–396.
image receptor on the detection of natural graphs. Caries Res 1989;23:334–341. Young SM, Lee JT, Hodges RJ, Chang TL, Elashoff
proximal caries in primary teeth. Clin Oral DA, White SC: A comparative study of high-
Investig 2011;15:901–907. resolution cone beam computed tomography
Souza-Zaroni WC, Ciccone JC, Souza-Gabriel and charge-coupled device sensors for detect-
AE, Ramos RP, Corona SA, Palma-Dibb RG: ing caries. Dentomaxillofac Radiol 2009; 38:
Validity and reproducibility of different com- 445–451.
binations of methods for occlusal caries de- Ziegler CM, Woertche R, Brief J, Hassfeld S: Clin-
tection: an in vitro comparison. Caries Res ical indications for digital volume tomogra-
2006;40:194–201. phy in oral and maxillofacial surgery. Dento-
maxillofac Radiol 2002;31:126–130.

169.230.243.252 - 11/27/2014 3:27:14 PM

574 Caries Res 2014;48:566–574 Tarım Ertas/Küçükyılmaz/Ertaş/Savaş/


UCSF Library & CKM

DOI: 10.1159/000357596 Yırcalı Atıcı


Downloaded by:

You might also like