You are on page 1of 5

[Downloaded free from http://www.jicdro.org on Sunday, December 31, 2017, IP: 93.76.25.

162]

Short Communication

Is primary stability a predictable parameter for


loading implant?
Ratnadeep Patil1,2, Dimple Bharadwaj1
1
Department of Clinical Dentistry, Smile Care, Mumbai, Maharashtra, India, 2Department of Prosthodontics and Biomaterial, Adjunct Professor, Centre for Dentistry
and Oral Hygiene Section, Oral Function, Prosthodontics and Biomaterial, University Medical Centre, The University of Groningen, Groningen, The Netherlands

Access this article online


ABSTRACT
Website: www.jicdro.org
Implant stability is important for osseointregration; without it, long-term success cannot be achieved. DOI: 10.4103/2231-0754.176264
Continuous monitoring in a quantitative and objective manner is important to determine the status of implant Quick Response Code:
stability. Measurement of implant stability is a valuable tool for making decisions pertaining to treatment
protocol and it also improves dentist-patient communication. Owing to the invasive nature of histological
analysis, various others methods have been proposed such as radiographs, cutting torque resistance,
reverse torque, and resonance frequency analysis (RFA). This review focuses on the objectives and various
methods to evaluate implant stability.

Key words: Insertion torque (IT), primary stability, secondary stability

INTRODUCTION of the implant is, the surface properties of the implant, the
lubrication of the preparation (blood), and also the design of
The use of dental implants has become widespread and a
predictable treatment modality for the restoration of missing
teeth. It has become a part and parcel of routine therapy;
considering this, success of implant dentistry depends on the
stability of the implant, whether biological or mechanical.
Initial stability at placement (primary stability) and the
development of osseointegration in the following healing
process (secondary stability) are two important factors for
implant success [Figure 1].

It is important to verify the status of implant-bone interface


to decide the loading time of the implant. Generally, clinicians
evaluate primary stability using the percussion test or using
their own perception during the implant placement process.
In particular, peak insertion torque (IT) and resonance
Figure 1: primary stability comes with old bone. secondary stability comes with
frequency analysis (RFA) are the most used globally.[1] Peak IT new bone
is measured while the implant is inserted. The IT corresponds
to a combination of the cutting friction of the tip of the This is an open access article distributed under the terms of the
implant in the bone, and the friction between the implant Creative Commons Attribution-NonCommercial-ShareAlike 3.0
surface and the preparation in the bone. If the osteotome is License, which allows others to remix, tweak, and build upon the
work non-commercially, as long as the author is credited and the
narrow or the bone quality is high, the torque will be higher.
new creations are licensed under the identical terms.
The torque will also depend on how sharp the cutting tip
For reprints contact: reprints@medknow.com
Address for correspondence:
Dr. Dimple Bhardwaj, 13 Geetanjali, 234 SV Road, Bandra (West),
Mumbai - 400 050, Maharashtra, India. Cite this article as: Patil R, Bharadwaj D. Is primary stability a predictable
parameter for loading implant?. J Int Clin Dent Res Organ 2016;8:84-8.
E-mail: dimple_bh@yahoo.com

84 © 2016 Journal of the International Clinical Dental Research Organization | Published by Wolters Kluwer - Medknow
[Downloaded free from http://www.jicdro.org on Sunday, December 31, 2017, IP: 93.76.25.162]

Patil and Bharadwaj: Primary stability predictable parameter for loading implants?

the implant itself. The formula of higher IT torque translating the IT values, between conical and cylindrical implants can
into higher primary stability may not always be true because be explained by the different contact surface areas among
the quantity and quality of bone varies significantly among the thread geometry of these implants.[5]
patients.[1]
It is clear that higher ITs fulfill the desire to achieve a high
The torque values that we get during the placement of degree of mechanical stability as interpreted through manual
the implant is the first contact of the implant surface with perception. It is typical for manufacturers to provide some
the bone; these values will not be the same all along the guidance on optimal IT, with some implant designs being
osteotome and so we require a smart physiodispenser, which specifically tailored to deliver higher ITs. This yields a sense
will give us a graphical representation of torque value of of comfort for the clinician that the implant is initially
each implant thread, which gets embedded inside the bone. “stable.”[8] However, such a high torque has not been shown
The current trends and the changes required to measure to be propitious to the surrounding bone. Numerous studies
the biological and mechanical stability will be explored in have been published, which clearly demonstrate that the
this article. critical pressure at this high torques leads to microfracture
of the bone, with a net resorption in the cortical zone and
Primary stability an unfavorable, indeed delayed healing process with reduced
Primary stability is defined as the absence of mobility in the BIC. Such a response might well shift the onset for secondary
bone bed after the implant has been placed.[2] It depends on stability and thereby delay or extend the period of potential
the mechanical engagement of an implant within the fresh vulnerability. This is clearly counter to the goal we are trying
bone socket. During the early stages of healing, mechanical to achieve with immediate or even early loading protocols
stability decreases and biological stability increases. In where we want to transfer from simple mechanical fixation
an osseointegrated implant, the stability depends on the to full osseointegration in the shortest possible time.[8]
biological component.[3] Primary stability is important for
good secondary stability as it prevents the formation of Bone density and quality
connective tissue layer between implant and bone, ensuring Bone quality is often referred to as the amount (and their
bone healing. topographic relationship) of cortical and cancellous bone
in which the recipient site is drilled. A poor bone quantity
A key factor for the implant primary stability is the
and quality have been indicated as the main risk factors for
bone-to-implant contact (BIC)[4] and thus, factors such as
implant failure as it may be associated with excessive bone
implant shape, length, and diameter that cause an increase
resorption and impairment in the healing process, compared
in the contact area between the implant and bone may
with higher density bone.[2] Clinical studies have reported
increase the implant primary stability. Furthermore, the
dental implants in the mandible to have higher survival rates
quality of bone bed plays an important role in shaping
compared to those in the maxilla, especially for the posterior
the BIC area.[5]
maxilla. It has been shown that achieving optimum primary
Factors influencing primary stability:[6] stability in soft bones is difficult and is also related to a higher
Implant geometry. implant failure rate for the implants placed in such bones.
Bone density and quality. Thus, the density of the surrounding bone seems to play an
Surgical protocol (osteotomy preparation) including the skill essential role in high occlusal forces and therefore, the high
of the surgeon. BIC percentages of a thin, “carpet”-like bone in contact with
the implant surface seems to be not clinically significant
Implant geometry compared to the lower rate of BIC in a thick bone.
Mechanical stability occurs where friction occurs between
the implant and the surrounding bone, giving rise to Surgical protocol
resistant torque at time of insertion. This resisting torque The clinical perception of primary implant stability is frequently
is proportional to the effort required to seat the implant; it based on the cutting resistance of the implant during its
depends largely on the characteristics of the implant, the insertion. Different surgeons have different preparation
density of the bone, and size of the osteotomy, as it pertains protocols, depending on the patient bone densities. Among
to the diameter of the implant. The implants with treated the surgical factors that influence osseointegration, implant
surfaces show greater roughness, a higher friction coefficient, bed preparation is of critical importance. Drilling the implant
and demand a larger IT than machined implants.[7] The results bed not only causes mechanical damage to the bone but also
of the surface roughness and friction coefficients are in increases the temperature of the bone directly, adjacent to
accordance with the results of the IT. The difference, across the implant surface.[7] Mechanical and thermal damage to

Journal of the International Clinical Dental Research Organization | January-June 2016 | Vol 8 | Issue 1 85
[Downloaded free from http://www.jicdro.org on Sunday, December 31, 2017, IP: 93.76.25.162]

Patil and Bharadwaj: Primary stability predictable parameter for loading implants?

the tissue surrounding the implant during drilling can have Resonanace frequency analysis
a destructive effect on the initial state of the cavity housing RFA measures the stability by apply a bending load,
the implant.[6] A correlation between cutting resistance at which mimics the clinical load and direction and provides
implant placement and primary resonance frequency (RF) information about the stiffness of the implant-bone junction.
values was reported for maxillary implants.[9] It evaluates the micromobility or displacement of the implant
in bone under a lateral load, applying microscopic lateral
Subjective evaluation forces to the implant with a vibrating transducer.[11,8] The
In clinical work, primary stability can be evaluated by first commercial version of the RFA technique (Osstellt,
percussion test, Periotest, RFA and placement torque. Integration Diagnostic AB, Goteborg, Västergötland and
Bohuslän, Sweden).
Percussion test
The percussion test may involve the tapping of a mirror The results are given as implant stability quotients (ISQs),[10]
handle against the implant carrier and is designed to elicit which are affected by three main factors:
a ringing sound from the implant as an indication of good • The stiffness of the implant fixture
stability or osseointregration. [10] However, this method • The interface with surrounding tissue
may be subjective according to the examiner and may give • The design of the transducer and the total effective
inaccurate measurements for implants because of the high implant length above bone level.
rigidity of implants and the lack of periodontal ligaments.[3]
The stiffer the interface between the bone and implant, the
The result is displayed digitally and audibly as Periotest values
higher the frequency and higher the frequency, higher is the
(PTVs) on a scale of –8 (low mobility) to 50 (high mobility).
ISQ level. The ISQ unit is based on the underlying RF and
Peak insertion torque ranges from 1 (lowest stability) to 100 (highest stability). We
The IT used during placement of implants was measured already know from the literature that an implant can tolerate
through a surgical handpiece; it can be used to predict implant a degree of micromotion, thought to be 100-150 μm, and
survival and to estimate healing time.[11] However, this method this is what ISQ measures.
is nonsubjective, noninvasive, and extensively used in clinical
The RFA values are still not definitive as no actual threshold
practice during implant placement to assess primary stability.
value has been established to differentiate a stable, integrated
Peak IT gives us a static measurement, which is taken only
implant from a failing/failed implant; however, it has been
once while force that is required for each and every thread
suggested that an ISQ value above 57 at 1 year after loading
of the implant to go through the bone will not be at the
represents a successful implant outcome[7] while a value
same torque. Thus, it allows only a single measurement at
below 50 indicates a risk of implant failure.[12]
implant insertion and cannot be used for evaluating secondary
stability. IT only assesses condition at the time of implant There is a lack of correlation between IT and the ISQ as
placement.[9] measured by RFA in an implant that is driven in at 30 ncm
and has the same ISQ as the one that required 100 ncm of
Periotest torque. Since ISQ is measuring axial stiffness, could it be that
Periotest (Gulden-Medizintechnik, Bensheiman der axial stiffness is far more relevant than rotational friction in
Bergstrae, Hesse, Germany) is an electronic instrument ensuring implant integration? 
originally designed to perform quantitative measurements
of the damping characteristics of the periodontal ligament It has been postulated that implants with low ISQ values yield
surrounding a tooth, thereby establishing a value for its more marked increases in ISQ values with time than implants
mobility.[9] As the outcome of Periotest measurements with high ISQ values, indicating that differences in RFA values
is influenced by the distance from the striking point to between implants may diminish with time.
the first bone contact, it is evident that placement of the
It may be speculated that similar bone densities will result
implant in the vertical dimension, abutment height, the
with time as a consequence of remodeling and adaptation
level of marginal bone loss, and the striking position on
to function.
the abutment/implant are critical factors for accuracy and/or
reproducibility.[9] Single readings of Periotest determinations Although extensively used in clinical research as one
are of limited clinical value and have not been demonstrated parameter to monitor implant stability, it has to be
to reflect on the nature of the bone/implant interface. By realized that RFA is affected by factors such as bone tissue
performing repeated measurements of the same implant characteristics and implant sink depth, diameter, and surface
over time, implant stability may be confirmed.[9] characteristics. Research indicates that implants yielding high

86 Journal of the International Clinical Dental Research Organization | January-June 2016 | Vol 8 | Issue 1
[Downloaded free from http://www.jicdro.org on Sunday, December 31, 2017, IP: 93.76.25.162]

Patil and Bharadwaj: Primary stability predictable parameter for loading implants?

ISQ values during follow-up appear to maintain stability. Low to implant stability. The value which we get during the
or decreasing ISQ values may be indicative of developing placement of the implant is the first contact of the implant
instability. However, no established normative range of ISQ surface with the bone; this torque value will not be the
values is available as yet. same all along the osteotome and thus, we require a smart
physiodispenser, which would record the torque value at
Consequently, a single determination of the ISQ value each thread level and a graph can be attained helping us to
does not define bone/interface characteristics or provide a decide whether we can load an implant immediately or if
quantitative evaluation of bone tissue integration. Similarly, we need to wait. Although we have fixed IT values, implant
no prognostic value for developing implant instability can with low IT will not be removed and it has to be left to be
be attributed to RFA. Hence, at the present time, the validity healed. Due to biology of bone, dynamics process takes
and relevance of RFA for clinical use have to be questioned.[9] place which will lead to ossteointegration and there will be
The destructive methodologies, such as removal torque micromechanical fixation.
assessment and pullout and pushout techniques are generally Osseo integration is the basis of a successful endosseous
used only in preclinical applications. While these methods implant. The process itself is quite complex and there are
may be of value as research techniques, they are of limited many factors that influence the formation and maintenance
value in clinical use owing to ethical concerns associated of bone at the implant surface.[7] The stiffness of the implant
with the invasive nature of such methodology. is a function of its geometry and material composition
(length, diameter, and overall shape). Second, the stiffness of
Biological stability
the implant tissue interface depends on the bond between
The gradual shift from primary stability to biological stability is
the surface of the implant and the surrounding bone. The
poised at around 3 weeks; this is seen as to be the least stable
stiffness of the surrounding tissue is determined by the
time point where viscoelastic stress relaxation of the bone,
ratio of cancellous to cortical bone and the density of the
along with remodeling, results in the loss of primary stability.[13]
bone with which an implant engages. Stiffness found at the
While secondary stability is the progressive increase in stability
bone-to-implant interface changes over time; thus, primary
related to biologic events at the bone-to-implant interface
stability decreases with time and mechanical stability takes
such as new bone formation and remodeling,[14] it is absent
over. During this period of transition between primary and
at the time of implant placement and increases with time.[15]
secondary stability, the implant faces the greatest risk of
Secondary stability is a biological phenomenon, the result
micromotion and consequent failure. It is estimated that this
of the healing that takes place around the implant, i.e., the
period in humans occurs roughly 2-3 weeks after implant
osseointregration. This process is dependent on many factors,
placement when osteoclastic activity decreases the initial
e.g., the implant surface properties, loading conditions, and
mechanical stability of the implant but not enough new
the individual host response
bone has been produced to provide an equivalent or greater
Bone-to-implant contact amount of compensatory biological stability.[8] This is related
The concept of initially placing more bone within the to the biologic reaction of the bone to surgical trauma
during the initial bone remodeling phase; bone and necrotic
immediate vicinity of the implant surface has been termed
materials resorbed by osteoclastic activity is reflected by
initial bone-to-implant contact (IBIC). Maximizing IBIC has
a reduction in implant stability quotient (ISQ) value. This
two major benefits:
process is followed by new bone apposition initiated by
1. The greater the IBIC, the greater the mechanical stability,
osteoblastic activity, therefore leading to adaptive bone
thus enhancing the implant’s ability to withstand
remodeling around the implant.[4]
micromovement while secondary stability develops and
2. Reducing the osteogenic migration distance decreases Hypothetically, if the level of primary stability can be
the time for osteoconduction to occur. increased and the rate of osseointegration at the same time
DISCUSSION can be accelerated, the dip in total stability can be reduced
and the implant is made less susceptible to micromovement
The seating torque is the final torque value that is achieved and potential failure. Our goal must be the rapid onset
when the implant is inserted. Depending on the implant of secondary stability, with minimal critical pressure to
design and bone properties, the value can be higher or the poorly vascularized cortical bone so that unfavorable
lower. Torque measures the rotational friction between the resorptive responses and delayed healing are avoided. At
implant and the bone, together with the force required to the same time, we need to employ an objective measure of
cut the bone. However, torque does not necessarily correlate constraint that reliably ensures that the implant can tolerate

Journal of the International Clinical Dental Research Organization | January-June 2016 | Vol 8 | Issue 1 87
[Downloaded free from http://www.jicdro.org on Sunday, December 31, 2017, IP: 93.76.25.162]

Patil and Bharadwaj: Primary stability predictable parameter for loading implants?

early or immediate loading; the accuracy of primary stability 5. Dos Santos MV, Elias CN, Cavalcanti Lima JH. The  effects  of
superficial roughness and design on the primary stability of dental
prediction is not good enough to prevent mistakes when
implants. Clin Implant Dent Relat Res 2011;13:215-23.
using an immediate loading technique. Therefore, a more 6. Atsumi M, Park SH, Wang HL. Methods used to assess implant
systematic use of objective measurements is encouraged.[12] stability: Current status. Int J Oral Maxillofac Implants 2007;22:743-54.
A simple, predictable, noninvasive test to quantify implant 7. Parithimarkalaignan S, Padmanabhan TV. Osseointegration:
An update. J Indian Prosthodont Soc 2013;13:2-6.
stability and osseointegration is highly desirable.
8. Norton M. Primary stability versus viable constraint a need to
redefine. Int J Oral Maxillofac Implants 2013;28:19-21.
Financial support and sponsorship 9. Aparicio C, Lang N P, Rangert B. Validity and clinical significance
Nil. of biomechanical testing of implant/bone interface. Clin Oral
Implants Res 2006;17(Suppl 2):2-7.
Conflicts of interest 10. Sennerby L, Meredith N. Implant stability measurements using
There are no conflicts of interest. resonance frequency analysis: Biological and biomechanical
aspects and clinical implications. Periodontol 2000 2008;47:51-66.
REFERENCES 11. Elias CN, Rocha FA, Nascimento AL, Coelho PG. Influence of
implant shape, surface morphology, surgical technique and bone
1. Bayarchimeg D, Namgoong H, Kim BK, Kim MD, Kim S, Kim TI, quality on the primary stability of dental implants. J Mech Behav
et al. Evaluation of the correlation between insertion torque and Biomed Mater 2012;16:169-80.
primary stability of dental implants using a block bone test. 12. Degidi M, Daprile G, Piattelli A. Determination of primary
J Periodontal Implant Sci 2013;43:30-6. stability: A comparison of the surgeon’s perception and objective
2. Javed F, Romanos GE. The role of primary stability for successful measurements. Int J Oral Maxillofac Implants 2010;25:558-61.
immediate loading of dental implants. A literature review. J Dent 13. Merdith N. Assessment of Implant stability as a progonostic
2010;38:612-20. determinant. Int J Prosthodont 1998;11:491-501.
3. Simunek A, Kopecka D, Brazda T, Strnad I, Capek L, Slezak R. 14. Bischof M, Nedir R, Szmukler-Moncler S, Bernard JP, Samson J.
Development of Implant stability during early healing of Implant stability measurement of delayed and immediately
immediately loaded Implant. Int J Oral Maxillofac Implants loaded implants during healing. Clin Oral Implants Res
2012;27:619-27. 2004;15:529-39.
4. Barikani H, Rashtak S, Akbari S, Fard MK, Rokn A. The effect of 15. Shadid RM, Sadaqah NR, Othman SA. Does the implant surgical
shape, length and diameter of implants on primary stability based technique affect the primary and/or secondary stability of dental
on resonance frequency analysis. Dent Res J (Isfahan) 2014;11:87-91. implants? A systematic review. Int J Dent 2014; 2014:204838.

88 Journal of the International Clinical Dental Research Organization | January-June 2016 | Vol 8 | Issue 1

You might also like