Professional Documents
Culture Documents
ServiceManual Biolyzer-300
ServiceManual Biolyzer-300
Service Manual
Analyticon
Biotechnologies AG
Am Muehlenberg 10
35104 Lichtenfels - Germany
info@analyticon-diagnostics.com
www.analyticon-diagnostics.com
WARRANTY
The term of warranty of this analyzer is one year from the date of purchase.
Our company shall not be responsible for the following failures and damages for the
warranty period.
1. Failure/damage caused by the result of misuse.
2. Failure/damage caused by repair or alternation performed by any company other
than our company.
3. Failure/damage caused by the phenomenon which is due to other than our
product.
4. Failure/damage caused by the condition beyond the normal operating condition of
this analyzer such as the power supply and the installation environment.
5. Failure/damage caused by fire, earthquake, flood damage or other natural
disaster.
6. Failure/damage caused by shift or transportation performed by other than our
company after our installation.
TRADEMARKS
Windows XP Professional, Windows Vista Business, Windows Vista Home Premium,
Windows 7 and Microsoft .NET FrameWork2.0 are registered trademarks of Microsoft
Corporation.
REVISION RECORD
The latest edition supersedes any preceding ones. If you have old editions, discard
them to avoid possible confusion.
Table of Contents
Foreword 0-1
1. Safety precautions 0-2
2. Precautions for Use 0-5
3. <Warning and caution labels> 0-13
4. Installation environment 0-14
5. Handling of barcode reader (option: Sample barcode reader) 0-24
6. Statement for EMC requirement 0-25
Chapter 1 Overview 1-1
1.1 General 1-1
1.2 Appearance 1-2
1.3 Built-in Units 1-4
1.3.1 Function for each unit 1-4
1.3.2 Unit description 1-8
Chapter 2 Unit description 2-1
2.1 Chassis unit (CHS) 2-2
2.1.1 Frame (FRM) 2-2
2.1.2 Cover (CVR) 2-4
2.1.3 Power supply unit (PSU) 2-6
2.1.4 Control unit (CNU) 2-8
2.2 Reaction table block (RTB) 2-9
2.2.1 Incubation reaction unit (IRU) 2-9
2.2.2 Detector unit (DTR) 2-11
2.2.3 Washing unit (WU) 2-13
2.3 Stirrer unit (MIX) 2-15
2.3.1 Stirrer unit 1 (MIX1) and Stirrer unit 2 (MIX2) 2-15
2.4 Pipette unit (PT) 2-16
2.4.1 Sample Pipette unit (SPT) 2-16
2.4.2 Reagent Pipette unit (RPT) 2-18
2.5 Trough unit (TR) 2-20
2.6 Sample container unit (ASP) 2-22
2.7 Reagent container unit (RCU) 2-24
2.8 Pump unit (PP) 2-27
2.8.1 Sample pump unit (SPP) 2-27
2.8.2 Reagent pump unit (RPP) 2-29
2.8.3 Wash pump unit (WPP) 2-31
2.8.4 Subtank (SUBTK) 2-33
2.9 Supply water unit (SWU) 2-34
Table of Contents
Foreword
FURUNO Electric Company thanks you for selecting and purchasing the FURUNO
Fully Automated Clinical Chemistry Analyzer.
Prior to use, please read this service manual carefully and follow the recommended
procedures for operation and maintenance.
The analyzer is designed to measure human specimen (Serum, Plasma, and Urine).
It enables to operate by connecting to PC and print out the measurement results.
Precaution
The service manual is for clinical laboratory technologists, who operate the
analyzer.
1
Foreword
1. Safety precautions
1.1 Prevention of damage and ignition
Please observe the following precautions to prevent damage and ignitio n of the
analyzer.
When changing the installation site, contact our servicing or sales division.
Do not detach the front, the rear, or the side cover during measurement.
If any leakage inside the analyzer, contact our servicing division. Careless
action may cause electric shock.
For halogen lamp replacement, turn the power off and wait for the lamp to ge t
cold for more than 30 minutes. Then perform replacement. Direct contact
with the heated lamp may cause burn injury.
For safety reasons, follow the instructions on the labels attached to the
analyzer and the service manual.
2
Foreword
Do not look directly at the lighted halogen lamp or the laser beam from the
barcode reader. It may cause eye injury.
Do not detach the front, the rear, or the side cover during measurement.
3
Foreword
Some reagents are strong acid or strong alkali. Be careful not to touch them
directly with hands or spill over your clothes. If they mistakenly spill over
your skin or clothes, immediately wash thoroughly with water and soap. If
they contact eyes, flush thoroughly with plenty of water and consult a
physician.
Install appropriate power supply wiring not to affect other major instruments
(such as surgical instruments) or the facilities from the system failure.
4
Foreword
Note that there are some cases where a significant clinical affect is achieved
depending on chemical substances which coexist with samples (such as
medication, anticoagulant, and preservative).
B. Storage of samples
5
Foreword
Fibrins contained in serum may clog the pipettes. When separating serum,
make sure that blood is clotted enough. Samples used for measurement
shall be free of fibrin.
For storing, handling, and use of reagents, calibrators, and quality control
samples, follow the instructions of the relevant manufacturers and sales
representatives.
For stability after opening the package, refer to the relevant manufacturers
and sales representatives.
6
Foreword
F. Analysis performance
Make sure that there is no bending or loops for tube connection be tween
connector and tank. Inappropriate connection may cause troubles including
tube disconnection during operation, leakage, etc.
Make sure to prevent air bubbles from entering purified water into the
analyzer.
Do not install the analyzer near the equipment that generates an abnormal
noise.
Turn off a portable radio or cellular phone near the analyzer. Abnormal noise
or electromagnetic waves generated from the portable radio or cellular phone
may cause malfunction of the analyzer.
Do not install the analyzer near the medical equipment that may generate
electromagnetic waves.
Do not handle monitor, keyboard, and mouse with wet hand (water, reagents,
etc) to prevent failure.
7
Foreword
Calibration results
QC measurement data
Measurement results
Follow the instructions on the manual for m aintenance checks and periodical
parts replacement.
The used parts after replacement may be infected. For disposal, refer to the
relevant manufacturers and sales representatives and follow your company or
national regulation.
N. Backup of data
O. Virus check
Do not use the analyzer for purposes other than those intended.
8
Foreword
Q. Cleaning of covers
When external covers such as covers, work table, and touch panel display
become dirty, wipe and clean them with a neutral detergent moistened gauze.
R. Other cautions
S. Operation problems
In the event of operation problems, do not play with the analyzer and leave
repair work to authorized experts.
When applying the ISE unit, do not turn OFF the analyzer. L Solution is
automatically discharged into the ISE electrodes every five minutes to prevent
the electrodes from drying.
Check all the electrodes are properly installed into the ISE module prior to
operation. If not properly installed, L solution may leak inside the analyzer
and cause mechanical errors.
9
Foreword
The following is general precautions for use of medical electrical equipm ents. Please
read this manual to fully understand and handle properly.
If the equipment is not used in a manner specified by the manufacturer, the protection
may be impaired.
Secure the space (minimum 300mm) at the left hand of analyzer to turn off the
analyzer at emergency situation easily.
In the case of carrying the analyzer (weight: approx. 120 kg), at least four
persons should lift the analyzer by holding tightly the grip areas at the bottom
of analyzer.
Make sure that all necessary electrical cables are correctly and completely
connected.
10
Foreword
Do not exceed the required period and dosage for the diagnosis and
treatment.
Turn off the power after every operational switch and control is restored to its
pre-use state as directed.
Do not apply extra force when disconnecting the line cord plugs from
receptacles.
Storage areas:
(1) Keep the analyzer away from any other water.
(2) Avoid areas where the analyzer may be adversely affected by atmospheric
pressure, temperature, humidity, ventilation, sunlight, dust, air containing
salt or sulfur, etc.
(3) Do not expose the analyzer to inclination, vibration, shock (including shock
during transportation), etc. and pay attention to the state of safety.
(4) Do not store the analyzer in the storage area of chemicals or a place where
any gas is likely to be generated.
Organize and store the accessory parts and cords after they have been
cleaned.
Keep the analyzer clean not to cause any inconvenience to the next use.
6. In the event of trouble, do not play with the analyzer and leave repair work to
authorized experts.
Make sure to check that the analyzer operates normally and safely when it is
reused after being kept unused for some time.
11
Foreword
9. The following precautions shall be taken when the cleaning procedure will be
performed.
10. For disposal of the analyzer, please contact our servicing division in advance.
11. When moving the analyzer, please contact our servicing division in advance.
12. When USB flash drive is used as the external storage device, don't remove the
flash drive from USB port while the flash drive's indicator is lighting. Otherwise,
the data in the flash drive may be broken and/or the trouble on the analyzer may
cause.
12
Foreword
The following labels are affixed on the major hazardous places within the analyzer and
added in their described parts of this manual.
13
Foreword
4. Installation environment
Installation operation (such as unpacking, installation, and operation check)
will be performed by our approved service representative.
In a place that is not exposed to direct wind, rain or any other water. (The
analyzer shall be installed indoors.)
In a dustless place.
On the floor that is strong enough to support the weight of the whole analyzer
system (approx. 120kg).
Not near noise sources (facility or equipment). Avoid placing the analyzer
under the fluorescent light.
At the height of 2,000 meters above sea-level or lower. (Use the analyzer only
indoors.)
Do not store the analyzer in the storage area of chemicals or a place where
any gas is likely to be generated.
B. Temperature/humidity conditions
Indoor temperature: 15 to 30ºC, Temperature variation: less than 2ºC per hour.
Install the analyzer in a place with good ventilation or with ventilation system.
Do not install the analyzer in a place subject to direct wind from air
conditioning.
14
Foreword
4.1 Space
Refer to the figure below for space required for installation including space for
maintenance and for exhaust ventilation.
A space of at least 300mm at the left side of analyzer to easily power off the
analyzer at the emergency situation.
A space of at least 150mm at the back of the analyzer for adequate ventilation.
A space of at least 500mm at the front of the analyzer for operation and
maintenance.
15
Foreword
16
Foreword
High conc. wastewater Wastewater from washing unit Silicon tube Analyzer – High conc.
Low conc. wastewater Wastewater from washing unit Silicon tube Analyzer – Low conc.
The water supply system and the low conc. wastewater drainage system need
to be placed prior to the analyzer installation.
Place the high conc. wastewater tank for high conc. waste.
17
Foreword
1 CA-270 1 Analyzer
B. Accessories list
Standard
25S3254(GB) (1)
25S3206(BS) (1)
operational PC
supply tube
1 Wash solution 1
supply tube *2
1 Wash solution 2
supply tube *2
tube
tube
(20L) *3
(20L) *3
(10L) *3
No. 3
No. 9
18
Foreword
Standard
F2
C. Options list
1 External tank External tank sensor table 25-040-S3702 1 Sensor for wash
sensor solution
supply tube
supply tube
(20L) sensor
(20L) sensor
(10L) sensor
external tank
sensor table
(20L) *3
(20L) *3
(10L) *3
19
Foreword
*2 When selecting external tank sensor (option), the tubes will be enclosed with the options. However,
when not selecting, the tubes will be enclosed with the accessories.
*3 When selecting external tank sensor (option), the external tanks (purified water tank, Low conc.
wastewater tank, and High conc. wastewater tank) will be enclosed with the options. However, when
*4 When selecting external tank sensor (option), the external tanks (wash solution 1/2 tanks (2L)) will be
enclosed with the options. However, when not selecting, the tanks (wash solution 1/2 tanks (5L) will
20
Foreword
Windows 7
Ethernet 10/100/1000Base-T
Make sure that the external storage device will not cause any system failure
from virus, etc. We do not guarantee any system failure caused by the
computer virus.
21
Foreword
22
Foreword
4.7 System
23
Foreword
Do not remove the covers. The adjustment of the barcode readers will be
performed by our approved service representative.
Laser beam from the sample barcode reader is exposed in the above arrow direction
during operation.
24
Foreword
25
Chapter 1: Overview
Chapter 1 Overview
1.1 General
The analyzer is a desktop fully automated clinical chemistry analyzer with throughput
of 270 tests per hour. Perform operation by connecting to an external operational PC
and print out the measurement results from a printer. This analyzer enables to
connect to a host computer when needed (ASTM Protocol capable).
Seventy-two cuvettes, where the samples are reacted with reagents, are located on the
Incubation reaction unit (IRU), the inside of which is kept at 37ºC constantly. After
mixing reagent and sample in the cuvette in the Incubation reaction unit, perform
absorbance measurement for the reaction liquid every 13 seconds. Wavelength(s) (to
a maximum of 2 wavelengths) used for a measurement can be chosen from 12
pre-specified wavelengths.
The sample tray is detachable and can accommodate maximum 40 sample tubes and
10 cup-holders. The reagent tray is also detachable and can accommodate maximum
50 bottles. The Reagent container has a cooling function to keep a constant
temperature (8 - 15ºC).
Barcode readers for reagent and sample are available to be mounted (as option).
Figure 1.1-1
1-1
Chapter 1: Overview
1.2 Appearance
Figure 1.2-1
Figure 1.2-2
1-2
Chapter 1: Overview
Figure 1.2-3
Figure 1.2-4
1-3
Chapter 1: Overview
Figure 1.3-1
Figure 1.3-2
1-4
Chapter 1: Overview
Figure 1.3-3
2. Detector unit
Measures the absorbance of the reaction liquid during reaction process.
This disperses the light from the halogen lamp as a source of light with a
grading method and measures 12 wavelengths at one time.
5. Stirrer unit
It is paddle-type stirrer to mix the reagent and sample within the cuvette .
For stirring speed, select from 5 levels (low to fastest).
6. Washing unit
Cleans the cuvettes after use.
7. Sample container
Accommodates sample tubes or cups.
It enables to hold 40 sample tubes and 10 sample cups (max.). A barcode
reader (option) allows sample identification from the barcode labels on the
sample tubes.
1-5
Chapter 1: Overview
8. Reagent container
Holds 50 reagent bottles (max.). It is refrigerated at 8-15ºC when the analyzer
is ON. A barcode reader (option) allows reagent identification from the
barcode labels on the reagent bottles.
1-6
Chapter 1: Overview
1-7
Chapter 1: Overview
1-8
Chapter 2: Unit description
The following are structurally similar units. Once you understand one unit, you easily
understand the other.
Note that figures illustrated in this chapter are drastically simplified for broad
understanding.
See also
Chapter 14 “Electrical connections” Electrical connections such as PC boards,
sensors, and motors
Chapter 15 “Fluidic System Diagram” Fluidic connections such as tubes, pumps,
and valves
Chapter 16 “Sensor List” Lists up functions of each sensor
2-1
Chapter 2: Unit description
B. Components
2-2
Chapter 2: Unit description
Figure 2.1-1
Figure 2.1-2
2-3
Chapter 2: Unit description
Prevents leak of the laser light from the sample BCR (option).
B. Components
2-4
Chapter 2: Unit description
Figure 2.1-3
Figure 2.1-4
2-5
Chapter 2: Unit description
AC power is supplied to the internal power supply pack via the noise filter and the
power switch. The power supply pack converts the AC power to required DC power (5V,
12V, or 24V) and supplies it to each unit. AC power input range is within 100V to 240V
to meet the voltage variation by countries.
When the power switch is ON, only the power (12V) for the RCU Peltier module (24V2)
and the ISE and the power (5V) for the CNU are supplied. When operational software
starts up by the control unit, the power (24V) for the motors and heaters and the power
(12V) for the halogen lamp are supplied.
B. Components
switch
Power supply Generates DC power for internal need from AC power provided outside. Mounts two types
PSU fan Cools off the power supply pack (multi-slot type). Has a function that detects fan stopping.
2-6
Chapter 2: Unit description
POWER-CN Controls inrush current and lighting voltage limitation of the halogen lamp.
PCB
Figure 2.1-5
2-7
Chapter 2: Unit description
The CNU houses the CNT-IBM PCB and the CNT-CN PCB and controls the analyzer
itself.
B. Components
CNT-IBM Controls the operation (measurement data/each sensor results/reading mclock, etc./each
PCB calculation method) and performs Ethernet communication with the operational PC.
CNT-CN PCB Relays each signal between CNT-IBM PCB and each of the PC boards.
Figure 2.1-6
2-8
Chapter 2: Unit description
Holds 72 semi-disposable cuvettes on the cuvette table and rotates the table to bring
the requested cuvette to the SPT (RPT) dispensing position where the SPT (RPT)
dispenses sample (reagent) into the cuvette.
To promote the incubation (reaction between sample and reagent), the cuvette table is
heated.
B. Components
Cuvette Table Holds 72 cuvettes. The table is kept at 37 ± 0.1℃ with the heater to promote the
At initialization, the cuvette table is driven to bring the cuvette #1 to the reagent dispensing
position.
Cuvette The cuvettes are made of resin that can pass light from the halogen lamp.
SENSA PCB Photo interrupter. Detects the zero position of the cuvette table (The zero -position
indicates the state when the cuvette # 1 comes to the R1 reagent dispensing position.)
Rotary Detects the rotational position of the IRU. This enables the DTR to perform photometric
Heater Three silicone rubber heaters are attached to the cuvette table for keeping the table warm.
TSP PCB Temperature sensors. This sensor is attached to the cuvette table and measures
temperature of the table. This is for controlling the incubation (reaction) liquid inside the
cuvettes kept at 37±0.1°C. A sensor is installed on each of the three divided heaters to
Thermal Prevents the IRU from overheat. When the temperature exceeds 76ºC, the fuse blowout
fuses leads to stop the power supply to the heaters. A fuse is installed on each of the three
Slip ring Relays the heater’s power of the cuvette table and temperature sensor signals.
IRU-CN2 Relays the signals from the heater’s power and the temperature sensors on the rotating
2-9
Chapter 2: Unit description
Figure 2.2-1
Figure 2.2-2
2-10
Chapter 2: Unit description
The DTR measures the absorbance change of the reaction solution that indicates
incubation (reaction) process in a cuvette. The light from the halogen lamp passes
through the heat-ray absorbing filter, is condensed by the lens. The transmissive light
through a cuvette hits the mirror to change its direction by 90 degree. After passing
through the imaging lens and the grating slot, it is dispersed by the grating mirror and
measured the absorbance in the 12 wavelengths at one time. While the IRU rotates
one cycle, the DTR measures the absorbance of 72 cuvettes at one time and outputs
the result data (72 cuvettes × 12 wavelengths) to the CNU.
B. Components
Halogen lamp This is a light source used for measuring the absorbance of the reaction mixture.
Grating Disperses the light passing through a cuvette with its different reflection angles by
mirror wavelengths.
PD-AMP2 Receives the light dispersed by the grating mirror as the photodiodes by wavelengths.
PCB
PD-AMP1 Converts and amplifies photocurrent of the PD-AMP2 PCB into voltage.
PCB
AD-MPX PCB Amplifies light-receiving signals by electronic volume value set at each wavelength, and
DET-CPU Controls the operation of the DTR. Transmits the measurement data to the CNU.
PCB
2-11
Chapter 2: Unit description
Figure 2.2-3
Figure 2.2-4
Figure 2.2-5
2-12
Chapter 2: Unit description
The WU consists of 4 pairs of drain and supply nozzles (WU1 to WU4) and 1 wipe
nozzle (WU5). After measurement, discharge the reaction liquid in the cuvette and
rinse inside the cuvette off with wash solution or purified water.
Stage 1: WU1: aspirates and discharges reaction liquid (drain nozzle) -> supplies
wash solution (supply nozzle)
Stage 2: WU1: aspirates and discharges wash solution (drain nozzle) -> supplies
wash solution (supply nozzle)
Stage 3: WU2: aspirates and discharge wash solution (drain nozzle) -> supplies
purified water (supply nozzle)
Stage 4: WU3: aspirates and discharges purified water (drain nozzle) -> supplies
wash solution (supply nozzle)
WU2: aspirates and discharges wash solution (drain nozzle)
Stage 5: WU3: supplies wash solution (supply nozzle)
Stage 6: (Presoaking in wash solution)
Stage 7: WU4: aspirates and discharge wash solution (drain nozzle) -> supplies
purified water (supply nozzle)
Stage 8: WU4: aspirates and discharge purified water (drain nozzle) -> supplies
purified water (supply nozzle)
Stage 9: (Presoaking in purified water at the Stage 7 -> performs water blank
measurement)
Stage 10: (Presoaking in purified water at the Stage 7 -> performs water blank
measurement)
Stage 11: WU5: discharges purified water and wipe the cuvette (water blank
measurement -> wipe nozzle)
Stage 12: WU5: discharges purified water and wipe the cuvette (wipe nozzle)
Each of WU1 to WU4 is equipped with a liquid level sensor to prevent the overflow of
the wastewater in the cuvette by the discharging and supplying proced ure.
2-13
Chapter 2: Unit description
B. Components
Supply Installed in each of WU1 to WU4 (total 4 supply nozzles). Supplies wash solution or
Drain nozzle Installed in each of WU1 to WU4 (total 4 drain nozzles). Aspirates and discharge reaction
liquid or wash solution in cuvettes while moving downward. It also functions as electrodes
Wipe nozzle Installed in the WU5. A wipe tip is equipped at the tip of the nozzle to wipe inside the
cuvette.
SENSA PCB Photo interrupter. Detects the vertical zero position (the upper limit).
WU-SEN This sensor detects the remaining volume of the cuvette in the WU 1 to WU4 after
PCB aspiration and discharging procedure. Detects the remaining volume by monitoring
WU-CN PCB Relays the signals between the SENSA PCB and the WU-SEN PCB.
Figure 2.2-6
2-14
Chapter 2: Unit description
After sample and reagent are dispensed into a cuvette, the paddle -type stirrer comes
down into the cuvette and stirs the mixture by rotating so that reaction may progress
evenly. The MIX consists of two mixing stirrers: MIX1 for R1 reagent and MIX2 for R2
reagent. The stirrer is cleaned at the trough with purified water after every use. At
initialization, the stirrer is in the position for washing in the trough.
B. Components
Stirrer Comes down into a cuvette and stirs the reaction mixture in the cuvette evenly.
SENSA PCB Photo interrupter. The sensor detects the stirrer is positioned at the trough.
Figure 2.3-1
2-15
Chapter 2: Unit description
The SPP in this unit aspirates and dispenses samples and others in the ASP, IRU, and
ISE. The pipette is cleaned in the SPT trough after use. At initialization, the pipette is
placed in the washing position in the trough.
B. Components
Pipette Performs vertically and horizontally rotation by the SPT_U and SPT_R motors. The
(Rotation) Photo interrupter. Detects the zero position for the rotation (the position where the
(Vertical) Photo interrupter. Detects the up-and-down zero position (the upper limit).
The pipette itself functions as a sensor. Detects that the tip of the pipette comes into
the sample liquid. When the sensor detects the pipette insertion, the SPT_U motor
stops, and then begins pipette aspiration and dispensing. Insertion of the tip of the
Photo interrupter. Normally the gobo of the pipette rests in the photo interrupter and
the photo interrupter is OFF. When the pipette tip reaches the bottom of a sample tube,
the gobo of the pipette is lifted up, resulting that the photo interrupter goes ON. This
sensor is used for protection of the pipette. When the sensor output turns ON, the
MOTOR-SEN-CN Synthesizes the signals of each sensor and each motor and relays to other units.
PCB
2-16
Chapter 2: Unit description
Figure 2.4-1
2-17
Chapter 2: Unit description
The RPP in this unit aspirates reagent or others in the RCU and dispenses it into the
cuvette in the IRU. This also dispenses purified water for washing cuvette in the IRU.
The used pipette is washed at the RPT 1 or 2 trough. At initialization, the pipette is
placed at the washing position in the trough.
B. Components
Pipette The RPT makes vertical, horizontal, and rotary movement by the RPT_U and the
(Rotation) Photo interrupter. Detects the zero position for the rotation (the position where the
(Vertical) Photo interrupter. The sensor on this PCB detects the up-and-down zero position
The pipette itself functions as a sensor. Detects that the tip of the pipette comes
into the reagent liquid. When the sensor detects the pipette insertion, the RPT_U
motor stops, and then begins pipette aspirating and discharging. Detects the
Photo interrupter. Normally the gobo of the pipette rests in the photo interrupter
and the photo interrupter is OFF. When the pipette tip reaches the bottom of a
reagent bottle, the gobo of the pipette is lifted up, resulting that the photo
interrupter goes ON. This sensor is used for protection of the pipette. When the
MOTOR-SEN-CN Synthesizes the signals from each sensor and each motor and relays to other
PCB units.
2-18
Chapter 2: Unit description
Figure 2.4-2
2-19
Chapter 2: Unit description
Trough
1. To dispose of dummy that is not used for measurement in the pipette (SPT and
RPT).
2. To dispose of the washing water inside the pipette (SPT and RPT).
3. To wash the surface of the SPT and the RPT pipettes and the MIX1 and the MIX
2 stirrers with purified water, and to dispose of the waste.
Trough chamber
2. The storage of the dew condensation water generating inside the RCU. The
stored waste sends to the drain system for disposal.
B. Components
Trough Total 4 troughs are placed at the operational area of the each pipette (SPT and RPT) and
each stirrer (MIX1 and MIX2). The surface of the SPT and the RPT pipettes can be washed
Trough Stores the waste from the PT and the MIX troughs and the dew condensatio n water from
chamber the RCU, and sends it to the external tanks. Detects an overflow by the float sensor.
Figure 2.5-1
2-20
Chapter 2: Unit description
Figure 2.5-2
2-21
Chapter 2: Unit description
The Sample container unit (ASP) holds 40 sample tubes and 10 sample cups. It rotates
its sample tray to bring a requested sample to the SPT aspiration position.
The sample position # is marked on the tray and the SPT starts aspiration from the
sample mounted in the sample position #1.
The sample tray rotates one position at a time during operation. If any rerun is required,
it moves the requested sample to the dispensing position.
B. Components
Outer tray Removable from the sample tray. The tray can hold up to 40 sample tubes. (outer layer: 20
tubes, inner layer: 20 tubes) Since the barcode reader (BCR) is equipped at the outside of
the ASP, the barcode label on the sample needs to be faced outward.
Inner tray Removable from the sample tray. The tray can hold up to 10 sample cups. No reading
Hall sensor Detects the ASP lid condition, open/close. A hall sensor is equipped at each lid of the
switch
ASP motor Stepping motor. Rotates the sample tray and moves the requested sample tube (normal
SENSA PCB Photo interrupter. Detects the zero position of the sample tray. (The zero-position indicates
Barcode Identifies samples by reading barcode labels attached to the sample tubes.
reader
(CCD/laser)
(option)
2-22
Chapter 2: Unit description
Figure 2.6-1
Figure 2.6-2
2-23
Chapter 2: Unit description
The RCU holds up to 50 reagent bottles and rotates to move the requested reagent to
the RPT aspiration position.
The data read from the BCR (option) sends to the CNU to save the reagent position
and its content.
The RCU is cooled by the Peltier modules and the reagents inside are kept at the
temperature 8 to 15℃.
B. Components
Reagent Tray The reagent tray is detachable and can accommodate up to 50 bottle s:
Since the BCR is equipped at the outside of the RCU, the barcode label on the reagent
Peltier module Cools off the RCU with two Peltier modules, and keeps inside the container at a
Cooler fan for A fan used to cool off the radiation fin for Peltier module. Since each of the Peltier
Peltier module modules has one fan, total 2 fans are equipped. This fan has a function that detects a
rotation stopping.
TSP PCB Monitors the temperature inside of the RCU. Based on the information of the
temperature, the RCU drives the Peltier modules to keep 8 to 15°C inside of the RCU.
Heater Defogs the BCR reading window glass. Connecting to the heater, stops the power supply
RCU motor Stepping motor. Rotates the reagent tray and moves the requested reagent to the RPT
aspiration position.
SENSA PCB Photo interrupter. Detects the rotational zero position of the reagent tray. (The
zero-position indicates the state when reagent position # 1 i s at the reagent dispensing
position.)
Barcode Identifies reagents by reading the barcode labels on the reagent bottles.
reader (BCR)
(option)
2-24
Chapter 2: Unit description
Figure 2.7-1
2-25
Chapter 2: Unit description
Figure 2.7-2
2-26
Chapter 2: Unit description
The SPP consists of the sample syringe (SPPS) and the wash syringe (SPPW).
This unit performs aspiration and dispensing of a sample or purified water through
up-and-down movement of a plunger. Since a motor is installed for each of the SPPS
and the SPPW, each syringe operates independently.
B. Components
Components Function and Operation
Plunger Two plungers (for SPPS and SPPW) are equipped. Perform up-and-down movement
SPP motor Stepping motor. Two motors (for SPPS and SPPW) are equipped.
SENSA PCB Photo interrupter. Two sensors (for SPPS and SPPW) are equipped. Detects the upper
NC: SPPS
Clot detection Helps to detect the syringe clog level with the pressure sensor.
(option) PRE-SEN-CN PCB: processes the pressure sensor signals at the sub -CPU and sends it to
the CNU.
2-27
Chapter 2: Unit description
Figure 2.8-1
Figure 2.8-2
2-28
Chapter 2: Unit description
The RPP consists of the reagent syringe (RPPR) and the wash syringe (RPPW).
This unit performs aspiration and dispensing for reagent and purified water through
up-and-down movement of a plunger. Since a motor is installed for each of the RPPS
and the RPPW, each syringe operates independently.
B. Components
Plunger Two plungers (for RPPR and RPPW) are equipped. Perform up-and-down movement
RPP motor Stepping motor. Two motors (for RPPR and RPPW) are equipped. Gives each of the
SENSA PCB Photo interrupter. Two of up zero-position sensors (for RPPR and RPPW) are equipped.
NC: RPPR
2-29
Chapter 2: Unit description
Figure 2.8-3
Figure 2.8-4
2-30
Chapter 2: Unit description
The WPP consists of 4 syringe pumps (two for wash solution and the other two for
purified water) and dispenses it into cuvettes.
B. Components
Syringe Performs syringe aspirating and dispensing by plunger pumping and connects to
dispensing nozzles of the WU. Supplies wash solution to the WU1 and WU3 and purified
WPP motor Performs the syringe aspirating and dispensing of wash solution or purified water by
plunger pumping. Each of the plungers is connected to its motor and operates
simultaneously.
WPP-EV Four solenoid valves are connected to each syringe. Connected to each syringe, performs
Solenoid aspirating and dispensing by switching the valve to the nozzle, the wash solution, or the
COM: Syringe
NC: Nozzle
SENSA PCB Photo interrupter. Detects the upper limit position of the plunger. (Starts aspiration of wash
2-31
Chapter 2: Unit description
Figure 2.8-5
2-32
Chapter 2: Unit description
The SUBTK stores purified water to remove air bubbles, and sends it to the SPP, the
RPP, and the WPP.
B. Components
Figure 2.8-6
2-33
Chapter 2: Unit description
The SWU supplies purified water and wash solution and discharges the waste.
B. Components
pump Supplies purified water to subtank, and each trough of the SPT, the RPT, the MIX1, and
the MIX2.
Check valve Total 4 check valves are equipped as connections between each of the supply pumps
and its trough. Prevents reverse flow of purified water supplied to each trough.
Inline filter N Total 4 filters are equipped as connections between the drain nozzles (WU 1 to WU4)
and the drain syringe pumps for the WU. Removes extraneous particles from waste.
Figure 2.9-1
2-34
Chapter 2: Unit description
tank
Supply tube 1 tube for purified water tank (Mesh filter installed to prevent any
contaminations)
2 tubes for wash solution tanks (Mesh filters installed to prevent any
contaminations)
2-35
Chapter 2: Unit description
2.11 Option
2.11.1 ISE unit (ISE)
A. Functions
The ISE is installed on the right side of the analyzer and is used to determine the
concentration of electrolytes (sodium, potassium and chloride) contained in serum,
blood plasma or urine using ion electrodes.
The sample supplied at the SPT is performed measurement automatically.
B. Components
Ion electrodes Consists of four electrodes, Na+, K+, Cl- and Reference. Measures the
Pump tube Controlled by the ISE module. Sends and drains liquids (such as H
ISE module Controls the pump operation. Measures signals from the ion electrodes and
ISE liquids Six kinds of liquids for the ISE. Each liquid is installed at the designated
position.
dispensed into the ISE every 5 minutes to prevent the electrodes from
drying.
H Solution: Installed inside the ASP. Dispensed into the ISE by the SPT.
ISE Cleaner: Wash solution exclusively for the ISE. Installed inside the
Urine Diluent: Used exclusively for diluting a urine sample and installed
inside the RCU. The RPT dispenses it into an IRU cuvette and mixes the
solution. Then dispenses the solution into the ISE by the SPT.
2-36
Chapter 2: Unit description
Figure 2.11-1
2-37
Chapter 2: Unit description
A. Functions
This unit enables to remove air bubbles from purified water and sends to the SPP and
the RPP.
B. Components
Degasser Removes air and dissolved gases through nonporous, special extra-fine
Vacuum pump Diaphragm pump. Evacuates the inside of the degasser module.
gauge
Figure 2.11-2
2-38
Chapter 2: Unit description
A. Functions
This unit enables to monitor liquid level inside the external tank.
B. Components
External tank Holds wash solution tank 1 and 2 that equipped with liquid level senso r for
rack each.
Connectors to liquid lever sensors for the purified water tank and
cable.
Purified water Float sensor. The sensor is inserted inside of the tank to detect the
OFF: 4L or less
High conc. Float sensor. The sensor is inserted inside of the tank to detect the
Low conc. Float sensor. The sensor is inserted inside of the tank to detect the
Wash solution Photo interrupter. Installed at the lower part of the tank outside and
2-39
Chapter 2: Unit description
Figure 2.11-3
2-40
Chapter 3: PC boards
Chapter 3 PC boards
3.1 General
This chapter explains each PC board in the analyzer.
A. Functions of PC boards
ASP/RCU-DRV 25P4102 Drives the motors of the ASP and the RCU. RCU
CNT-CN PCB 25P4103 Relays signals between the CNT-IBM PCB and CNU
CNT-IBM PCB 25P3521 Controls operation of the analyzer under the CNU
DET-CPU PCB 25P3514 Controls operation of the DTR. Transmits the data DTR
to the CNU.
F-LED PCB 25P3528 Displays analyzer status with 2 LED lamps. CHS
IRU_CN1 PCB 25P3510 Converts sensor signals of the units (SPT, RPT, CHS
IRU_CN2 PCB 25P3218 Relays power for driving the IRU heaters. Relays IRU
3-1
Chapter 3: PC boards
LLED PCB 25P3142 Light emitting board to detect the liquid volume of External tank
LLRD PCB 25P3141 Light receiving board to detect the liquid volume External tank
(Option)
SEN_LL/DL PCB 25P3716 Detects the liquid level and low limit level of the RPT, SPT
SENSA PCB 25P3207 Detects position through photo interrupter. ASP, RCU, IRU,
MIX1/2, RPT,
WPP, WU
WU nozzle.
3-2
Chapter 3: PC boards
Figure 3.3-1
3-3
Chapter 4: Unpacking and installation
1. Dimensions:
870mm (w) × 670mm (d) × 625mm (h) (not including the projection part)
2. Mass:
120kg (Max.)
B. Environmental condition
Do not expose the analyzer to direct heat or cool air from air conditione r, air
hole of heater, etc.
C. Power requirements
D. Operational PC
Refer to each operator’s manual for power supply voltage and power cons umption.
Refer to each operator’s manual for power supply voltage and power consumption.
4-1
Chapter 4: Unpacking and installation
Capacity: 5L/h
Connection tube:
4-2
Chapter 4: Unpacking and installation
Figure 4.2-1
Figure 4.2-2
Figure 4.2-3
4-3
Chapter 4: Unpacking and installation
4. Remove the 4 M8 bolts (WAF: 13mm, 2 at each right and left side) on the sleeve
side of the cardboard box.
Figure 4.2-4
Figure 4.2-5
6. Lift the sleeve part of the cardboard box and remove it. .
Figure 4.2-6
4-4
Chapter 4: Unpacking and installation
Figure 4.2-7
8. Pull out the two holding plates on the both right and left sides.
Figure 4.2-8
Figure 4.2-9
10. Remove the caps and the tapes at the tube connections.
4-5
Chapter 4: Unpacking and installation
4.3 Installation
Four persons or more are required to carefully lift and carry the analyzer from the
palette holding the grip areas on the bottom of the analyzer and place it at the
designated place.
The installation table or workbench should be solid enough to keep the analyzer (it
weighs approx. 120kg) stable.
Figure 4.3-1
4-6
Chapter 4: Unpacking and installation
Figure 4.3-2
2. Open the top cover and remove all the fixing tapes on the lids of the Sample
container and the Reagent container.
Figure 4.3-3
3. Remove the protective material attaching on the inner side of the lids of the
Sample container and the Reagent container.
Figure 4.3-4
Figure 4.3-5
4-7
Chapter 4: Unpacking and installation
Figure 4.3-6
Follow the procedure below to dismount the protective material for transportation use.
Hold the shaft part and lift the reagent pipette. (DO NOT hold the arm part.)
Carefully move the reagent pipette to the front without hitting the tip of the
reagent pipette at the trough or the work panel.
After removing the protective material, move the reagent pipette to the trough
position.
Figure 4.3-7
4-8
Chapter 4: Unpacking and installation
Hold the shaft part and lift the sample pipette. (DO NOT hold the arm part.)
Carefully move the sample pipette to the front without hitting the tip of the
sample pipette at the trough or the work panel.
Figure 4.3-8
3. Washing unit
After removing the protective material, move the sample pipette to the trough
position.
Figure 4.3-9
When removing the protective material, pay attention not to hit at other parts
such as the sample pipette, the reagent pipette, and the stirrer unit.
For the protective materials, keep and store them together wit h the
accessories appropriately.
When transferring the analyzer to some other place, make sure to reattach
the protective material appropriately.
4-9
Chapter 4: Unpacking and installation
Figure 4.3-10
Rotate the incubation reaction unit and remove the fixing tapes for the
cuvette dustproof cover.
For removing the fixing tapes, lift up the edge of the cuvette dustproof cover
onto the work table and remove the tapes at the several points accordin gly
by rotating the incubation reaction unit. Total 8 pieces of the fixing tapes on
the cuvette dustproof cover.
Figure 4.3-11
Make sure that there are no fixing tapes remained inside of the incubation
reaction unit.
Keep and store the cuvette dustproof cover together with the accessories
appropriately.
4-10
Chapter 4: Unpacking and installation
4.3.5 Level
The analyzer must be installed on the leveled table or workbench. Turn the height
adjustment bolt and perform level adjustment of the analyzer.
1. Open the lid of the sample container and pull out the sample tray.
Figure 4.3-12
3. Adjust the four height adjustment bolts on the bottom of the analyzer to enter
the air bubble of the level into the center of the level. The height adjustment
bolts is adjusted to 28mm before shipping from factory.
Figure 4.3-13
4. Adjust any of the appropriate height adjustment bolts to adjust the level within
the range 28 to 43mm (from initial height: 28mm).
Figure 4.3-14
4-11
Chapter 4: Unpacking and installation
When the height of the height adjustment bolt becomes lower than 28mm,
the Peltier element of the reagent container may cause damages.
Figure 4.3-15
G. Cable connection
1. Power connection
For power cable connection, apply power supply code (length: 2m) attached to
the analyzer and install a 3-pronged power cable with a ground terminal.
Figure 4.3-16
4-12
Chapter 4: Unpacking and installation
Plug power code plug directly in the outlet. Using Improper wiring such as
extension code or power strip may cause the fire or malfunction. It also adversely
affects the measurement results.
Keep the analyzer away from any other water. Prevent exposure to water drops
or rain to cause fire, electric shock, malfunction, or corrosion.
Avoid to damage on the power code such as scratches or under the heavy
objects. It may cause short-circuits, fire, electric shock, or malfunction.
Power supply is required AC100V to 120V, 50/60Hz, 6.5A (Max.) for the
analyzer itself.
Refer to each operator’s manual for power supply voltage and power
consumption.
4-13
Chapter 4: Unpacking and installation
H. Tubes
Connect the analyzer to each external tank using each of the accompanying
supply/drain tubes.
Figure 4.3-17
4-14
Chapter 4: Unpacking and installation
Two tube types: one for No. 1, 2 and 3 and the other for No. 4, 5, and 6.
Once each of the tubes No. 1, 2, and 3 is connected, do not change the
connection. If changing the connection, it causes contamination.
4-15
Chapter 4: Unpacking and installation
Figure 4.3-18
Insert a silencer into each tip of the tubes (# 4 and 5 ) of the high conc. waste
and the low conc. waste as shown below.
Figure 4.3-19
Push each of silencers accompanying in the accessory box into the tube No.
2 for the high conc. waste and No. 3 for the low conc. waste and insert them
into each of the wastewater tanks.
4-16
Chapter 4: Unpacking and installation
The tip of the supply tube is positioned at the bottom of the purified water
tank.
Enough purified water in the purified water tank
The tip of the supply tube is positioned at the bottom of each wash solution
tank.
Enough wash solution in each of the wash solution tank
The tip of the drain tube is positioned at the upper part of the wastewater
tank and with enough space from the liquid level.
Enough capacity in the wastewater tank
Figure 4.3-20
4-17
Chapter 4: Unpacking and installation
Figure 4.3-21
4-18
Chapter 4: Unpacking and installation
Figure 4.3-22
Figure 4.3-23
The analyzer in standard specification does not have connectors for the
external tank sensors (option).
4-19
Chapter 4: Unpacking and installation
Validity date for each reagent and sample is 1 month after each preparation.
3. Add purified water into the measuring flask and make 100mL solution.
2. Add purified water into the measuring flask and make 500mL solution.
2. Add the (B) solution into the measuring flask and make 100mL solution.
2. Add the (B) solution into the measuring flask and make 100mL solution.
4-20
Chapter 4: Unpacking and installation
2. Empty the low conc. wastewater tank (20L) and the high conc. wastewater tank
(10L).
3. Prepare a wash solution 1 tank (5L) and a wash solution 2 tank (5L) and fill each
of wash solution 1 and 2 (both diluted to 200 times) for each tank. (Fill 25mL of
each wash solution and 5L of purified water into the tank. Shake the tank well to
dissolve it.)
4. Insert the supply tube for wash solution 1 into the wash solution1 tank and the
supply tube for wash solution 2 into the wash solution2 tank respectively.
5. Insert drain tubes (No. 3 and 4) for the low conc. waste into the low conc.
wastewater tank and a drain tube (No.2) for the high conc. waste into the high
conc. wastewater tank.
6. Insert a supply tube for purified water into the purified water tank.
4-21
Chapter 4: Unpacking and installation
1. Turn the switch ON at the left side of the analyzer and reboot the operational
PC.
3. Enter the log-in name and the password. Press [->] key.
Figure 4.5-1
4. After successfully log-in under the EL level, the [Monitor] screen in the [Run
(F5)] appears.
4-22
Chapter 4: Unpacking and installation
2. Click on the [Sequence] screen name (white area) at upper -left part of the
screen.
Figure 4.5-2
3. Enter the open channel code for RCU. Press the [ENTER] key.
The open channel code for RCU is automatically changed each year.
Figure 4.5-3
5. Specify “System setting” area at the right part of the screen as follows:
4-23
Chapter 4: Unpacking and installation
7. Click the “Shutdown” icon. Click the [Login] button in the pop-up message.
Figure 4.5-4
4.5.4 Prime
1. Select [Maintenance (F10)] > [Sequence] screen.
2. Tick on the “Prime” and enter “10” at the right side of the “Execute” icon. Click
the “Execute” icon.
Figure 4.5-5
Prior to measurement, fill the analyzer with the purified water and remove the
air bubbles within the tubes completely.
4-24
Chapter 4: Unpacking and installation
1. Click the [Wash cuvette] button in the [Maintenance (F10)] > [Sequence] screen.
Perform cuvette wash.
Figure 4.5-6
B. Automatic gain
Perform automatic gain in 20 minutes or more after the analyzer powered ON.
1. Click the [Automatic Gain] button in the [Maintenance (F10)] > [Performance]
screen. Perform automatic gain.
2. Check that each wavelength at the “Gain” is kept within 8.5 to 9.5 V.
Figure 4.5-7
1. Click the [Check water blank] button in the [Maintenance (F10)] > [Sequence]
screen. Perform water blank measurement.
Figure 4.5-8
4-25
Chapter 4: Unpacking and installation
3. Select “340” in “Wave Length" from drop-down menu at the upper-left part of the
screen.
4. Check that the absorbance value of each cuvette is under 5,000mAbs/10 and
not displayed in red.
Figure 4.5-9
4-26
Chapter 4: Unpacking and installation
Figure 4.5-10
3. Register the reagent bottle position as follows. Then click the “Save” icon.
Double-click the reagent name in the Pos. “01” and select “SPTS” from
drop-down menu. Select “R1” as the Reagent Type.
Double-click the reagent name in the Pos. “02” and select “RPTS” from
drop-down menu. Select “R1” as the Reagent Type.
Double-click the reagent name in the Pos. “03” and select “RPT2S” from
drop-down menu. Select “R1” as the Reagent Type.
4-27
Chapter 4: Unpacking and installation
Figure 4.5-11
2. Specify the chemistry parameter setting for the dye test as follows. For every
chemistry parameter setting is done, click the “Save” icon.
4-28
Chapter 4: Unpacking and installation
Figure 4.5-12
2. Specify the calibration setting for the dye test as follows. For each calibration
setting is done, click the “Save” icon.
4-29
Chapter 4: Unpacking and installation
Figure 4.5-13
Select “01” at the Start (Pos.) and tick on the “SPTS” on the method list on
the right side of the screen. Select “20” at the “Number” from drop -down
menu and click the “Save” icon.
Register the following orders as indicated below in the same procedure.
4-30
Chapter 4: Unpacking and installation
Place a reagent bottle (70mL) containing 30mL or more of the (B) solution
prepared in “4.4.1 Preparation of sample and reagent” on the position # 1.
Place a reagent bottle (70mL) containing 30mL or more of the (C) solution
prepared in “4.4.1 Preparation of sample and reagent” on the position # 2.
Place a reagent bottle (70mL) containing 15mL or more of the (D) solution
prepared in “4.4.1 Preparation of sample and reagent” on the position # 3.
Open the reagent bottle prior to measurement and close the bottle
immediately after measurement.
Place sample cups containing 0.5mL or more of the (A) solution prepared in
“4.4.1 Preparation of sample and reagent” on the position # 1 and # 2.
Place sample cups containing 2.0mL or more of the (B) s olution prepared in
“4.4.1 Preparation of sample and reagent” on the position # 3, # 4, # 5, and #
6.
4-31
Chapter 4: Unpacking and installation
Figure 4.6-1
4-32
Chapter 4: Unpacking and installation
Figure 4.7-1
2. Select “Normal” in the Sample Category from drop-down menu at the “Search
for” area of the screen.
3. Specify the search dates “From” and “To” from drop-down menu and enter the
“Round No”. When not specifying the search date, remove tick mark at “Enable”.
When not specifying the round no. enter “*”.
4. Select “Monitor” at the Output Result from drop-down menu and click the
“Search” icon.
5. Check that the dye test results are displayed in the screen.
7. Click the […] button to display the My File Open Dialog box and specify the
folder to be saved.
8. Click the “Search” icon and transfer the data into the specified folder. The data
is saved in csv. format.
9. Sum up the measurement results (the absorbance values of total 20 tests Sum
up the measurement results (the absorbance values of total 120 tests (each 2
running for the SPTS, the RPTS, and the RPT2S tests where N=20) and
calculate the average (AV), the standard deviation (SD), the max. and the min.
values, and the coefficient of variance (CV). When the coefficient of variance
(CV) is within the following acceptable range, the results meet the criteria.
4-33
Chapter 4: Unpacking and installation
Criteria:
When applying the Degasser module, SPTS<2.0%, RPTS<1.0%,
RPT2S<1.0%
When not applying the Degasser module, SPTS<2.0%, RPTS<1.0%,
RPT2S<3.0%
10. If the results do not meet the criteria, some malfunction may occur. Check that
the all procedures, setting details, and the analyzer status and perform dye test
again.
4-34
Chapter 4: Unpacking and installation
Figure 4.8-1
Level: L3
New password: CA-270
4. Click the “Shutdown” icon. Click the [Logout] button in the pop-up message.
5. Check that the newly created account is effective to log in from the log -in
screen.
4-35
Chapter 4: Unpacking and installation
4-42
Chapter 4: Unpacking and installation
4-43
Chapter 4: Unpacking and installation
2. Click the “Shutdown” icon. Click the [Power Off] button in the pop-up message.
Figure 4.9-1
3. Turn OFF the power supply switch at the left side of the analyzer.
4-36
Chapter 4: Unpacking and installation
1. Operation environment
2. Installation site
4. Data
4-37
Chapter 4: Unpacking and installation
4-38
Chapter 4: Unpacking and installation
4-39
Chapter 4: Unpacking and installation
4-40
Chapter 4: Unpacking and installation
4-41
Chapter 4: Unpacking and installation
4-42
Chapter 4: Unpacking and installation
4-43
Chapter 5: Cleaning
Chapter 5 Cleaning
This chapter describes cleaning instructions for each unit.
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
In manual operating, hold the neck of shaft for the unit and raise or rotate the
arm.
Prevent the cuvettes on the Incubation reaction unit from contamination of any
dust or liquids.
In the case of using alcohol, should not be close to fire or high temperat ure
parts/units, because the alcohol is flammable liquid.
2. Wipe the pipette from top to bottom with the alcohol moistened gauze.
When wiping the pipette, do not apply extra force to it. It is very important that
the pipette is vertically set so that the analyzer can function correctly. Improper
alignment may cause the damages of the pipette and/or inaccurate analysis
results.
5-1
Chapter 5: Cleaning
Figure 5.1-1
3. Insert the cleaning pointer of the pipette cleaning jig into the poi nt of the pipette
to clean the inside of the pipette.
Two needles of the pipette cleaning jig on both sides: the thin (for the sample
pipette) and the thick type (for the reagent pipette). Distinguish the needle
type from the intended usage part.
Figure 5.1-2
4. Wipe the pipette cover entirely with the neutral detergent moistened gauze.
Figure 5.1-3
5-2
Chapter 5: Cleaning
5. After cleaning inside the Sample pipette, perform the pipette wash with purified
water by pressing the [Sample Pipette Wash] button in the [Run (F5)] > [Daily
Maintenance] screen.
Prior to the sample pipette wash, place wash solution C-1 (diluted to 100 times
with purified water) in the Sample container and register the order. For the
procedure, refer to operator’s manual “6.3.7 Sample pipette wash.”
5-3
Chapter 5: Cleaning
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
Prevent the cuvettes on the Incubation reaction unit from contamination of any
dust or liquids.
In the case of using alcohol, should not be close to fir e or high temperature
parts/units, because the alcohol is flammable liquid.
2. Wipe the supply/drain nozzles with the alcohol moistened gauze gently. Avoid
bending the nozzles while cleaning.
When wiping the nozzle, do not apply extra force to it. It is very important that
the nozzle is vertically set so that the analyzer can function correctly. Improper
alignment may cause the damages of the nozzle and/or inaccurate analysis
results.
3. Wipe the Washing unit base with the neutral detergent moistened gauze.
Figure 5.2-1
5-4
Chapter 5: Cleaning
5.3 Stirrer
A. Check the following points.
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
Prevent the cuvettes on the Incubation reaction unit from contamination of any
dust or liquids.
2. Wipe the stirrer with the neutral detergent moistened gauze gently. Avoid
bending the stirrer while cleaning.
3. Wipe the Stirrer unit cover with the neutral detergent moistened gauze.
Figure 5.3-1
5-5
Chapter 5: Cleaning
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
1. Check that the reagent pipette is not positioned on the Sample container. If the
pipette is on the Sample container, move it away slowly from the Sample
container.
2. Remove the sample trays (both Inner and Outer) and the lid.
3. Wipe the sample trays (both Inner and Outer) and the lid with the neutral
detergent moistened gauze.
4. Wipe the inner surface of the Sample container with the neutral detergent
moistened gauze.
5. Attach the sample trays (both Inner and outer) and the lid in the reverse
procedure of the step 2.
5-6
Chapter 5: Cleaning
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
1. Confirm that the reagent pipette is not positioned on the Reagent container. If
the pipette is on the Reagent container, move it away slowly from the Reagent
Container.
3. Wipe the reagent tray and the lid with the neutral detergent moistened gauze.
4. Wipe the inner surface of the Reagent container with the neutral detergent
moistened gauze.
Figure 5.5-1
5. Attach the reagent tray and the lid in the reverse procedure of the step 2.
5-7
Chapter 5: Cleaning
B. Cleaning
Before the cleaning procedure, make sure to turn off the analyzer.
1. Wipe the surface of the work table and the outer surface with the neutral
detergent moistened gauze.
Figure 5.6-1
5-8
Chapter 5: Cleaning
B. Cleaning
1. Discharge purified water, wastewater, or wash solution from each of the external
tanks.
2. Pour the Wash solution C-1 diluent (diluted to 100 times with purified water) into
the external tank.
The diluent volume for each tank varies depending on the tank size.
Wash thoroughly inside the external tank with purified water to completely
remove wash solution component.
5-9
Chapter 5: Cleaning
When applying the Degassing unit (option), perform tube wash every 2 weeks.
A. Preparation
Wash solution C-1 diluent (diluted to 100 times with purified water) (5 L)
Prepare wash solution C-1 diluent followed by the instructions on the label.
For preparing diluent, use purified water (ion-exchanged water). If using tap
water, brown deposits (corroded iron) or white deposits (calcium chloride) may
be generated after several days passed.
Do not make the product contact with zinc plating or metals including iron,
ferric oxide, copper, copper alloy, etc. to prevent corrosion of them.
When re-sampling the product into another container, use the container that
completely clean by rinsing it with pure water and dry it.
5-10
Chapter 5: Cleaning
D. Cleaning
1. Click the [Tube wash] button in the [Run(F5)] > [Dail y Maintenance] screen.
Figure 5.8-1
Figure 5.8-2
3. Click the “Print” icon to print out the procedure on the screen.
4. Perform cleaning procedure. For details, refer to the following “E. Cleaning
procedure”.
5-11
Chapter 5: Cleaning
E. Cleaning procedure
Figure 5.8-3
(1) Connect the tubes from the wash solution 1 and 2 supply to the system
water tank. Click the “Execute” icon.
2. Preparatory operation
Figure 5.8-4
5-12
Chapter 5: Cleaning
(1) Discharge the purified water from the system water tank. Fill the tank
with 5 L of the wash solution C-1 diluent.
(2) Connect the supply tubes for the system water and the wash s olution 1
and 2 to the system water tank filled with the wash solution C-1 diluent.
(3) Click the “Shutdown” icon and power OFF the analyzer.
(4) Dismount the Side panel R. Discharge the remaining liquid in the
subtank.
Dismounting procedure for subtank
Loosen the 3 screw lids at the upper part of the subtank and disconnect the
subtank float sensor assembly and its tubes from the subtank.
Figure 5.8-5
(5) Install the empty subtank and mount the Side panel R to original position.
Mounting procedure for subtank
Connect the subtank float sensor assembly and its tubes to the subtank and
tighten the 3 screw lids.
Figure 5.8-6
5-13
Chapter 5: Cleaning
Figure 5.8-7
(1) Confirm that the supply tubes for the system water and the wash solution
1 and 2 are connected to the system water tank filled with the wash
solution C-1 diluent. Click the “Execute” icon.
Figure 5.8-8
(1) Disconnect the supply tubes for the system water and the wash solution
1 and 2 from the system water tank filled with the wash solution C-1
diluent.
5-14
Chapter 5: Cleaning
(2) Discharge the wash solution C-1 diluent from the system water tank and
rinse inside the tank with purified water. Then fill the tank with purified
water up to the full level.
(3) Connect the supply tubes for the system water and the wash solution 1
and 2 to the system water tank filled with purified water. Click the
“Execute” icon.
Figure 5.8-9
(1) Click the “Shutdown” icon and power OFF the analyzer.
(2) Dismount the Side panel R and discharge the remaining liquid at the
subtank.
(3) Install the empty subtank and mount the Side panel R to original
position.
(4) Power ON the operational PC and the analyzer.
5-15
Chapter 5: Cleaning
6. Ending procedure
Figure 5.8-10
(1) Check that the system water supply tube is properly connected to the
system water tank.
(2) Connect the supply tubes for the wash solution 1 and 2 are connected to
the relevant wash solution tanks. Click the “Execute” icon.
5-16
Chapter 5: Cleaning
5.9 Cuvettes
Perform water blank measurement to check any dirt on the cuvettes.
1. Select the [Run (F5)] > [Daily Maintenance] screen. Click the [Check water
blank] button to perform water blank measurement.
Figure 5.9-1
2. Check each of the cuvette status in the [Maintenance (F10)] > [Water Blank]
screen.
The absorbance results of all wavelengths for all cuvettes are displayed.
The values exceeding the threshold value among the selected wavelengths
indicates in red. The values exceeding the threshold value among other
wavelengths indicates in yellow.
Figure 5.9-2
5-17
Chapter 5: Cleaning
3. Confirm that the values of all cuvette are within the threshold value.
B. Cuvette cleaning
1. Select the [Run (F5)] > [Daily Maintenance] screen. Click the [Wash cuvette]
button to perform the cuvette cleaning.
Figure 5.9-3
2. Click the [Check water blank] button again. Confirm that the values of all cuvette
are within the threshold value.
After cuvette cleaning, if the value still indicates out of the threshold values,
needs to replace it with the new one.
If more than 24 cuvettes with out of the threshold values, measurement cannot
start.
5-18
Chapter 5: Cleaning
Place the ISE cleaning (ISE Cleaner) in the Sample container and register the
order prior to ISE cleaning. For the registration procedure, refer to “6.3.9. ISE
cleaning (option).”
ISE activation is automatically performed during ISE cleaning. Place the ISE
activator in the Sample container and register the order. For the registration
procedure, refer to “6.3.11 ISE activation (option).”
A. Cleaning
1. Check that the ISE cleaning and the ISE activator are placed on the designated
positions that registered in the [Test selection] screen, and then click the [ISE
Cleaning] button in the [Run (F5)] > [Daily Maintenance] screen.
5-19
Chapter 6: Periodical checks
6-1
Chapter 6: Periodical checks
6-2
Chapter 6: Periodical checks
6-3
Chapter 6: Periodical checks
6-4
Chapter 6: Periodical checks
6-5
Chapter 6: Periodical checks
6-6
Chapter 6: Periodical checks
6-7
Chapter 7: Maintenance screens
To login as the engineer level, it is required to enter the provided password from the
manufacturer in the [Setup Password] screen.
7.2 Sequence
Figure 7.2-1
7-1
Chapter 7: Maintenance screens
Figure 7.2-2
1. Initialization:
Place back each unit to its default position.
3. Wash cuvette:
Perform the cuvette cleaning.
6. Ascend/Descend Stirrer:
Move the Stirrer unit up or down.
7. Ascend/Descend Washer:
Move the Washing unit up or down.
8. ISE Prime:
Fill the tubes with the L solution to remove the air bubbles in the tubing. Perform
priming at the specified number of times.
7-2
Chapter 7: Maintenance screens
9. ISE Calibration:
Perform 2-point calibration after registering the position of the L solution in the
[Run (F5)] > [Test Selection] screen.
7-3
Chapter 7: Maintenance screens
Figure 7.2-3
Perform pipette wash for the sample pipette or the reagent pipette. Select purified
water (system water) or wash solution.
1. Sample:
Perform washing for the sample pipette. Specify the wash by ticking “System
water” or “Wash solution” (registered in the Sample Container).
2. Reagent:
Perform washing for the reagent pipette. Specify the wash by ticking “System
water” or “Wash solution” (registered in the Reagent Container).
7-4
Chapter 7: Maintenance screens
Figure 7.2-4
Perform pump test for each pump with specifying the operation time. This funct ion is
for checking whether each pump works properly.
The following pumps can operate at the respective area:
MIX1: a pump for supplying purified water to the Stirrer unit 1 trough
MIX2: a pump for supplying purified water to the Stirrer unit 2 trough
SPT-T: a pump for supplying purified water to the Sample pipette trough
RPT-T: a pump for supplying purified water to the Reagent pipette trough
Figure 7.2-5
Specify the mode and the position for each motor and perform motor operation. This
function is for checking whether the motor works properly and the position adjustment
after the unit replacement.
1. Motor:
Select the motor from drop-down menu.
7-5
Chapter 7: Maintenance screens
2. Mode:
Select the operation mode (4 kinds) from drop-down menu.
Move: rotates the motor to move the unit to the position specified at the
“Position”.
Check: move the motor outside of the original position and returns it to the
original position.
3. Position:
When selecting the “Move” at the “Mode”, select the motor designated position
from drop-down menu.
7-6
Chapter 7: Maintenance screens
7-8
Chapter 7: Maintenance screens
E. Prime area
Figure 7.2-6
Fill the tubes with purified water or wash solution to remove the air bubbles in the tubes.
After priming, it is available to select the execution of the sleep mode or shut down.
1. Prime:
Perform priming at the specified number of times.
Figure 7.2-7
Click the [Reagent Barcode Adjustment] button. The following pop-up message
appears.
Figure 7.2-8
7-9
Chapter 7: Maintenance screens
Place the reagent bottle at the position specified by the pop -up message. Click
the [OK] button. Then start scanning the barcode label of the reagent bottle.
Perform total 20 scans.
When 20 scans successfully completed and all of the scan results are
recognized equal, the following pop-up message appears indicating the
scanned results.
Figure 7.2-9
When more than one out of 20 scans failed or any of the scan results are not
recognized equal, the following pop-up message appears.
To scan barcode label successfully, adjust the barcode reader position.
To rescan the barcode label, click the [Re-Scan] button in the pop-up message.
Figure 7.2-10
Place the reagent bottles labeled barcodes at all of the specified positions. Click
the [OK] button and start barcode scan. Perform scanning for all positions twice.
The following message appears when scans are completed.
7-10
Chapter 7: Maintenance screens
Figure 7.2-11
When the barcode scans twice are successfully completed, it is recognized the
barcode scan functions properly.
To rescan the barcode label, click the [Re-Scan] button in the pop-up message.
Figure 7.2-12
Place the sample tube at the position specified by the pop-up message. Click
the [OK] button. Then start scanning the barcode label on the sample tube.
When 20 scans successfully completed and all of the scan results are
recognized equal, the following pop-up message appears indicating the
scanned results.
7-11
Chapter 7: Maintenance screens
Figure 7.2-13
When more than one out of 20 scans failed or any of the scan results are not
recognized equal, all 20 scan results are displayed in the following pop -up
message.
To scan barcode label successfully, adjust the barcode reader position.
To rescan the barcode label, click the [Re-Scan] button in the pop-up message.
Figure 7.2-14
Place the sample tubes labeled barcodes at all of the specified positions # 1 to #
40. Click the [OK] button and start barcode scan. Barcode scan will be
performed for the position # 1 to 40 twice.
The following message appears when scans are completed.
7-12
Chapter 7: Maintenance screens
Figure 7.2-15
When the barcode scans twice are successfully completed, it is recognized the
barcode scan functions properly. To rescan the barcode label, click the
[Re-Scan] button in the pop-up message.
7-13
Chapter 7: Maintenance screens
Figure 7.2-16
This area is available to perform each of the unit operation s that preprogrammed.
One button can be registered for one operation. (Four buttons available)
CA-270MotorMoveCommandFile
(motor unit number), (mode number), (position), (remaining seconds)
(motor unit number), (mode number), (position), (remaining seconds)
・
(max. 50 lines)
・
9999
7-14
Chapter 7: Maintenance screens
Figure 7.3-1
This screen is available for checking the cuvette water blank results.
Figure 7.3-2
7-15
Chapter 7: Maintenance screens
1. Wave length:
Select the wavelength for viewing the results from drop-down menu.
2. Date time:
Select the date and time for viewing the results from drop-down menu. Latest 20
results are available to view.
Figure 7.3-3
The water blank results for each cuvette are displayed on this screen. When water
blank measurement is completed, this area is updated the values.
When the cuvette is beyond the threshold value in the specified wave length s the
cuvette # is displayed in red. When it is beyond the threshold value in other wave
lengths, it is displayed in yellow. The cuvette beyond the threshold value is not used for
measurement. At least 24 usable cuvettes are required for measurement start.
Figure 7.3-4
1. Cuvette No.:
Enter the cuvette # to display its cumulative data.
7-16
Chapter 7: Maintenance screens
2. Graph:
The water blank results for each cuvette are displayed in graph format. The
results up to latest 20 results are available to be displayed.
3. Max./Min.:
Specify the graph display area.
Figure 7.3-5
2. Save:
Saves the water blank results. This button is available only after the water blank
measurement.
3. Print:
Print out the displayed water blank results.
4. Output history:
Saves the water blank results as a file. The up to last 3-month water blank
results can be saved. When saving the results, the mean values are saved after
calculating the results in a day.
5. Cuvette replacement:
When water blank measurement needs to be performed again after cuvette
replacement, click this button prior to replacement. The following pop -up
message appears.
7-17
Chapter 7: Maintenance screens
Figure 7.3-6
Perform the cuvette replacement and click the [OK] button. The following
pop-up message appears for confirmation.
Figure 7.3-7
7-18
Chapter 7: Maintenance screens
7.4 Performance
Figure 7.4-1
Figure 7.4-2
7-19
Chapter 7: Maintenance screens
1. Graph
2. Temperature
Incubator: Displayed the values from the three temperature sensors in the
Incubator.
Reagent Container: Displayed the values from the three temperature sensors
in the Incubator.
3. Offset
Incubator: Enter the offset value for each temperature sensor in the Incubato r.
Enter the value in units of 0.01 ºC.
Reagent Container: Enter the offset value for each temperature sensor in the
Reagent container. Enter the value in units of 0.01 ºC.
Figure 7.4-3
1. Wave Length
Displays the wavelengths in the Detector unit.
2. Gain
Displays the gain value for each wave length.
3. Voltage
Displays the voltage of the halogen lamp by the indicated gain value.
4. Automatic Gain
7-20
Chapter 7: Maintenance screens
Automatically perform the gain adjustment for the halogen lamp. After halogen
lamp replacement, gain adjustment needs to be performed.
5. Check Lamp
Perform halogen lamp power check. The results will be displayed at the
“Voltage”. If it is under the specified value, it is displayed in r ed.
7-21
Chapter 7: Maintenance screens
7.5 Sensor
Figure 7.5-1
This screen is available to check ON/OFF state for each sensor in the analyzer. It is
effective when performing mechanical check in the analyzer.
A. Sensor/pulse
SPT to WU4: Perform liquid level sensor check for the SPT, the RPT and the
WU1 to 4.
2. Read Start:
Starts to obtain the current sensor state.
7-22
Chapter 7: Maintenance screens
B. Sensor state
Displays the ON/OFF state for each sensor. When the sensor is ON, it is displayed in
red. When it is OFF, it is displayed in yellow.
Following are the sensors displayed on the screen.
<Sensor list>
Unit Sensor Description
CHS RIGHT MIDDLE LEAK Leak sensor at the middle layer of the right side.
RIGHT LOWER LEAK Leak sensor at the lower layer of the right side.
LEFT MIDDLE LEAK Leak sensor at the middle layer of the left side.
CHASSIS FAN Chassis fan
PSU FAN POWER Power supply cooling fan
IRU ORIGIN Zero-Point Sensor
DTR FAN Cooling fan for halogen lamp
WU WU ORIGIN Up Zero-Position Sensor
ASP ORIGIN Zero-Point Sensor
FRONT COVER Emergency Sample Lid Sensor
ALL COVER ASP Lid Sensor
RCU ORIGIN Zero-Point Sensor
COVER RCU Lid Sensor
PELTIER1 FAN RCU Peltier1 fan
PELTIER2 FAN RCU Peltier2 fan
MIX 1 ORIGIN Zero-Position Sensor
MIX 2 ORIGIN Zero-Position Sensor
SPT UP ORIGIN Up Zero-Position Sensor
ROTATION ORIGIN Rotational Zero-Position Sensor
DOWN LIMIT Down Limit Sensor
RPT UP ORIGIN Up Zero-Position Sensor
ROTATION ORIGIN Rotational Zero-Position Sensor
DOWN LIMIT Down Limit Sensor
SPP SPPS ORIGIN SPPS Zero-Position Sensor
SPPW ORIGIN SPPW Zero-Position Sensor
RPP RPPR ORIGIN RPPR Zero-Position Sensor
RPPW ORIGIN RPPW Zero-Position Sensor
WPP ORIGIN WPP Zero-Position Sensor
SUB TANK FULL Subtank Full Sensor
POSITION Tank Detection Sensor
CHAMBER OVERFLOW Trough Chamber Liquid Level Sensor
DC POWER 24V1A MAIN 24V Power Supply
24V1B MAIN 24V Power Supply
24V2A ASP/RCU 24V Power Supply (ASP/RCU)
24V2B ASP/RCU 24V Power Supply (ASP/RCU)
TANK WATER Liquid level sensor for purified water tank
(OPTION)
WASH-1 Liquid level sensor for wash solution tank 1
WASH-2 Liquid level sensor for wash solution tank 2
7-23
Chapter 7: Maintenance screens
LOW CONC. WASTE Liquid level sensor for low conc. wastewater
tank
HIGH CONC. WASTE Liquid level sensor for high conc. wastewater
tank
DEAERATOR PRESSURE Pressure detection sensor inside the Degassing
(OPTION) unit.
7-24
Chapter 7: Maintenance screens
Figure 7.6-1
By dispensing the specified volume with each pipette and entering the actual volume
on this screen, perform volume adjustment between the indicated dispensing volume
and the actual volume.
Since the entered volumes have been adjusted at the factory shipment, do not
change the volume basically except if necessary.
After saving the values, reboot the analyzer and the operational PC.
A. Volume Adjustment
1. Sample Pipette:
Enter actual volumes for measurement when dispensing 1.5 µL and 20 µL with
the sample pipette.
2. Reagent Pipette:
Enter actual volumes for measurement when dispensing 20 µL and 100 µL with
the reagent pipette.
1. Sample:
Enter the actual volume for measurement when dispensing an ISE sample (53
µL).
7-25
Chapter 7: Maintenance screens
2. Diluted Sample 1:
Enter the actual volume for measurement when dispensing an ISE diluted
sample (53 µL).
7-26
Chapter 7: Maintenance screens
Figure 7.7-1
This screen is available to adjust the photometry position for each cuvette in the
Incubator.
The offset position for each cuvette are displayed.
2. Click the “Save” icon to save the offset values displayed on the screen in file
format. Saved as “csv” format.
Click the […] button to specify the destination folder to be saved.
7-27
Chapter 7: Maintenance screens
Figure 7.8-1
1. User Name:
Select the user name from drop-down menu.
2. User Level:
Displays the user level that selected from the “User Name”.
3. Date:
Enter the period to search for the operation log. Specify the date by clicking the
calendar.
[Enable]: When not ticking the checkbox, it is not available to specify the search
start and end dates.
4. Output Result:
Specify the results output from drop-down menu.
5. “Search” icon:
Searches for the results and saves.
7-28
Chapter 8: Parts replacement
For dismounting/mounting explained in this chapter, the following items are required:
Note: Hold the neck of the shaft when vertically rotating SPT/RPT/MIX1/MIX2 unit by
hand.
8-1
Chapter 8: Parts replacement
Required Tools
Item Description
Screwdriver Required types:
+ Screwdrivers No.1 and No.2
+ Long Screwdriver No.2
+ Short Screwdriver No.2
- Screwdriver 6 mm
Torque Screwdriver No.2
Hex wrench Required sizes:
1.5 mm, 2 mm, 2.5 mm, 3 mm
Small ratchet -
set
Ratchet Required sizes for packing:
wrench 13 mm
Box Required sizes:
screwdriver 5.5 mm, 7 mm, 8 mm
Spanner Required sizes:
5.5mm, 10mm, 13mm, 14mm, 14.2mm, 17mm, 19mm, 22mm
Long-nose -
pliers
Nipper -
Pliers -
Monkey -
wrench
Tube cutter Required for tubing
Retractable -
knife
Air blower Required for cleaning lens and filters.
Gap gauge Required for clearance adjustment in the pulley.
Tweezers -
8-2
Chapter 8: Parts replacement
2. The Mosaic covers are divided in the following figure. For replacement of parts
or units, dismount the Mosaic covers as required.
Figure 8.1-1
8-3
Chapter 8: Parts replacement
Figure 8.1-2
1. Loosen the two screws (M3×8) and dismount the side panel L (or side panel R)
by sliding it backward.
Figure 8.1-3
8-4
Chapter 8: Parts replacement
2. Loosen the four screws (M3×8) and dismount the front panel by sliding it
forward.
Figure 8.1-4
1. Remove the five screws (M3×8) and dismount the rear panel.
Figure 8.1-5
2. Remove the two screws (M3×8) and dismount the upper panel.
Figure 8.1-6
8-5
Chapter 8: Parts replacement
2. Remove the four screws (M3×8) for the auto hinge and the four screws (M3×8)
for the free stop hinge and dismount the top cover from the upper panel.
Figure 8.1-7
3. Remove the two screws (M3×8) and the four screws (M3×6) and dismount he
auto hinge.
Figure 8.1-8
2. Remove the air filters that are detachable using the hook and the loop fasteners.
Figure 8.1-9
8-6
Chapter 8: Parts replacement
3. Remove the nine screws (M3×8) and dismount the MAIN-DRV PCB.
Figure 8.1-10
3. Remove the four screws (M3×8) and dismount the IRU_CN1 PCB.
Figure 8.1-11
1. Dismount the side panel L, the side panel R, and the front panel.
8-7
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the F-LED PCB.
Figure 8.1-12
3. Remove the two screws (M3×30) and dismount the CHS fan assembly.
Figure 8.1-13
8-8
Chapter 8: Parts replacement
2. Remove the screw (M3×8) and the connecting lines from the leak sensor.
Dismount the leak sensor assembly.
Figure 8.1-14
3. Remove the two screws (M3×8) and the connecting lines. Dismount the sponge
from the leak sensor assembly.
Figure 8.1-15
8-9
Chapter 8: Parts replacement
2. Remove the screw (M3×8) and the connecting lines from the leak sensor.
Dismount the leak sensor assembly.
Figure 8.1-16
3. Remove the two screws (M3×8) and the connecting lines. Dismount the sponge
from the leak sensor assembly.
Figure 8.1-17
8-10
Chapter 8: Parts replacement
3. Remove the three screws (M3×8) and dismount the PSU assembly.
Figure 8.1-18
2. Loosen the three screws (M3×8) and dismount the PSU cover.
Figure 8.1-19
8-11
Chapter 8: Parts replacement
Figure 8.1-20
4. Remove the four screws (M4×10) at the rear of the PSU assembly and dismount
the Multi-slot power supply.
Figure 8.1-21
2. Remove the two screws (M3) on the side of the Multi-slot power supply and
disconnect the connector from the PSU fan. Dismount the PSU fan.
Figure 8.1-22
8-12
Chapter 8: Parts replacement
4. Remove the four screws (M3×8) and dismount the single supply board.
Figure 8.1-23
4. Remove the four screws (M3×8) and dismount the POWER-CN PCB.
Figure 8.1-24
8-13
Chapter 8: Parts replacement
F. Dismounting Fuses
1. Turn the fuse holder as shown picture below and pull the fuse out.
Figure 8.1-25
2. Replace the fuses with the new ones that enclosed in the accessory kit.
Figure 8.1-26
8-14
Chapter 8: Parts replacement
2. Remove the upper two screws (M3×8) and loosen the lower two screws (M3×8).
Dismount the CNU cover.
Figure 8.1-27
Figure 8.1-28
8-15
Chapter 8: Parts replacement
1. Remove the four screws (M3×8) and dismount the CNT-CN PCB.
Figure 8.1-29
2. Remove the three screws (M3×8) and the four hex spacers and dismount the
CNT-IBM PCB.
Figure 8.1-30
8-16
Chapter 8: Parts replacement
When mounting the Mosaic-IRU, handle carefully not to tuck cables. For the
cables, tie them with a clip.
1. Dismount the side panel L, the side panel R, the front panel, and all Mosaic
covers.
2. Dismount the ASP assembly, the SPT assembly, the MIX1 assembly, and the
ISE assembly (option).
Figure 8.2-1
4. Disconnect the connectors and tubes from the other units and pull out the RTB
assembly as indicated in the arrow direction.
Figure 8.2-2
8-17
Chapter 8: Parts replacement
There are two cuvette types (resin or glass) available as option. Replacement
procedure is same for both cuvette types. However, the shape and the replacement
cycle vary depending on each type.
Figure 8.2-3
Figure 8.2-4
Figure 8.2-5
8-18
Chapter 8: Parts replacement
Figure 8.2-6
When dismounting the cuvette holder, handle carefully in the screws not to
drop them.
Figure 8.2-7
8-19
Chapter 8: Parts replacement
When replacing,
Resin type: hold the upper handles of the cuvette.
Glass type: hold the upper part of the ground glass sides.
To avoid any scratches on the surface of the cuvette, remove the cuvette
holder first and then perform cuvette replacement.
Wear disposable gloves, mask, and eye protector to avoid any adherence of
liquids directly to the skin.
When placing the cuvette, make sure to vertically insert it on the cuvette table.
Figure 8.2-8
7. Fill the cuvette with 500µL of the Wash solution C-1 (diluted to 100 times) by
micropipette, etc.
The new cuvette needs to be rinsed with the diluted wash solution C -1 before
use.
If the cuvette not rinsed before use, it may effect on the measurement results.
8-20
Chapter 8: Parts replacement
10. Click the [OK] button in the pop-up message of the the step 3.
Figure 8.2-9
11. The pop-up message appears for the cuvette placement confirmation.
Click the [OK] button when confirming that all the cuvettes are placed on the
specified positions.
Figure 8.2-10
12. Leave the analyzer in the ready mode for more than 1 hour.
When rinsing the new cuvette with the diluted wash solution C -1. It needs
more than 1 hour presoak.
8-21
Chapter 8: Parts replacement
Figure 8.2-11
2. Move the cuvette to be replaced to the front position by rotating the Incubation
Reaction unit to right and left manually.
Before rotating the Incubation reaction unit, confirm that the wash nozzle or
the stirrer paddle is not inserted into the cuvette. If any of them is in its proper
position, raise each of the unit.
When dismounting the cuvette holder, handle carefully in the screws not to
drop them.
Figure 8.2-12
8-22
Chapter 8: Parts replacement
When replacing,
Resin type: hold the upper handles of the cuvette.
Glass type: hold the upper part of the ground glass sides.
To avoid any scratches on the surface of the cuvette, remove the cuvette
holder first and then perform cuvette replacement.
Wear disposable gloves, mask, and eye protector to avoid any adherence of
liquids directly to the skin.
When placing the cuvette, make sure to vertically insert it on the cuvette table.
Figure 8.2-13
8-23
Chapter 8: Parts replacement
9. Select “Wash bottle” from drop-down menu and click the [Pour water into
cuvettes] button.
Place the reagent bottle filled with the wash solution C -1 (diluted to 100 times)
in the RCU.
Register the diluted wash solution C-1 as “Wash Solution” at the Reagent code
for Cuvette Water Placement in the [System Setup 1] screen.
Figure 8.2-14
10. Leave the analyzer in the ready mode for more than 1 hour.
When rinsing the new cuvette with the diluted wash solution C-1, it needs
more than 1 hour presoak.
8-24
Chapter 8: Parts replacement
3. Select “System Water” from drop-down menu and click the [Pour water into
cuvettes] button.
Figure 8.2-15
Figure 8.2-16
8-25
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the Mosaic front, and the
Mosaic-IRU.
Figure 8.2-17
Figure 8.2-18
Figure 8.2-19
8-26
Chapter 8: Parts replacement
5. Remove the two screws (M3×8) and the screw (M3×40) to dismount the spacer.
Dismount the slip ring.
Figure 8.2-20
1. Dismount the side panel L, the side panel R, the front panel, and all Mosaic
covers.
3. Remove the two screws (M3×8) and dismount the Slip ring base assembly.
Figure 8.2-21
5. Remove the four screws (M3×10) and pull out the Cuvette table assembly by
lifting up the handle.
Figure 8.2-22
8-27
Chapter 8: Parts replacement
1. Temporarily fix the Cuvette table assembly by tightening the four screws
(M3×10).
Figure 8.2-23
4. Insert the IRU table adjuster into the hole for position adjustment.
Figure 8.2-24
8-28
Chapter 8: Parts replacement
Insert the IRU table adjuster into the hole for position adjustment to the bottom
and fix the screw.
Figure 8.2-25
6. Fix the Cuvette table assembly by tightening the four screws (M3×10).
Figure 8.2-26
7. Remove the screw (M3×12) and dismount the IRU table adjuster.
8-29
Chapter 8: Parts replacement
2. Remove the two hex socket head bolts (M5×12) and dismount the IRU motor
plate.
Figure 8.2-27
3. Remove the four screws (M4×12) and dismount the IRU/ASP motor assembly.
Figure 8.2-28
4. Remove the two screws (M3×8) on the Encoder base 2 and dismount the timing
belt.
Figure 8.2-29
8-30
Chapter 8: Parts replacement
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
Figure 8.2-30
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
Figure 8.2-31
4. Adjust the tension by moving the IRU motor plate in the arrow direction.
6. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
8-31
Chapter 8: Parts replacement
Figure 8.2-32
Figure 8.2-33
4. Pull down and dismount the Rotary encoder assembly with plate.
Figure 8.2-34
8-32
Chapter 8: Parts replacement
5. Remove the two screws (M3×8) and dismount the Rotary encoder assembly
from the plate.
Figure 8.2-35
Prior to mounting the Rotary encoder assembly, check that the rotary encoder
is tightly placed against the IRU shaft. Then perform the mounting procedure.
1. Dismount the side panel L, the side panel R, and the Mosaic-IRU.
2. Remove the two screws (M3×6) and dismount the sensor springs fixing on the
temperature fuses.
Figure 8.2-36
Figure 8.2-37
8-33
Chapter 8: Parts replacement
2. Remove the two screws (M3×6) and dismount the sensor springs fixing on the
TSP PCB to be replaced.
Figure 8.2-38
Figure 8.2-39
8-34
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.2-40
Figure 8.2-41
8-35
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the Mosaic -front, and the
Mosaic-IRU.
3. Remove the three screws (M3×6) and dismount the IRU_CN2 PCB.
Figure 8.2-42
8-36
Chapter 8: Parts replacement
3. Loosen the screw (M3×35) on the lamp case. (The screw on the lamp case
cannot be removed.)
Figure 8.2-43
Figure 8.2-44
5. Loosen the two screws (M3×6) and dismount the halogen lamp.
Figure 8.2-45
8-37
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the DTR fan holder.
Figure 8.2-46
4. Remove the two screws (M3×30) and dismount the DTR fan assembly.
Figure 8.2-47
8-38
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the front panel, and all Mosaic
covers.
3. Disconnect the connectors (DTR-FAN, DTR, and LAMP) from the DTR fan
assembly, the DET_CPU PCB, and the halogen lamp respectively.
6. Remove the three screws (M3×10) and dismount the IRU case.
Figure 8.2-48
7. Remove the three screws (M4×10) and dismount the DTR assembly.
Figure 8.2-49
8-39
Chapter 8: Parts replacement
2. Hold the wipe nozzle by fingers and pull out the wipe tip downward slowly.
1. Apply uniformly the liquid gasket (1212) on the surface of the 2 to 3 mm width
area in the 5 to 10 mm away from the tip of the wipe nozzle.
Figure 8.2-50
It is required 8 hours for the liquid gasket (1212) hardening after application.
2. Insert the new wipe tip into the bottom of the nozzle .
4. Select the [Run(F5)] > [Daily Maintenance] screen. Click the [Initialization]
button to perform initialization.
8-40
Chapter 8: Parts replacement
6. Insert the wipe nozzle into the cuvette and adjust the wipe tip position (equal
width: A and A”, B and B”) from the inner surface of the cuvette by rotating it.
Figure 8.2-51
7. When attaching the wipe tip correctly, check that the length between the lowest
part of the wipe base and the tip of the wipe tip is 67 mm.
Figure 8.2-52
8-41
Chapter 8: Parts replacement
Figure 8.2-53
3. Loosen the two screws (M3×8) and dismount the WU nozzle base assembly.
The retaining washer is attached behind each screw.
Figure 8.2-54
When mounting,
Place the edge of the WU nozzle base assembly and the boss against the
edge of the WU drive plate assembly and fix them.
Figure 8.2-55
8-42
Chapter 8: Parts replacement
2. Loosen the hex socket head screw (M3×5) and dismount the supply nozzle and
the drain nozzle assembly.
Figure 8.2-56
The supply nozzle and the drain nozzle are tightened by one hex socket head
screw. By loosening the screw, both nozzles are dismounted at one time.
3. For the drain nozzle replacement, disconnect each connector from the
WU_SEN PCB.
WU1: connector (CN612)
WU2: connector (CN613)
WU3: connector (CN614)
WU4: connector (CN616)
Figure 8.2-57
8-43
Chapter 8: Parts replacement
1. Install the supply nozzle in the position (the WU nozzle base is positioned at
51-mm height from the bottom of the supply nozzle).
Figure 8.2-58
2. Install the drain nozzle in the position (the WU nozzle base is positioned at
55-mm height from the bottom of the drain nozzle).
Figure 8.2-59
8-44
Chapter 8: Parts replacement
Figure 8.2-60
1. Install the wipe nozzle in the position (the WU nozzle base is positioned at
64.5-mm height from the bottom of the wipe nozzle).
Figure 8.2-61
8-45
Chapter 8: Parts replacement
E. Dismounting WU assembly
Figure 8.2-62
4. Disconnect each connector from the motor and the WU_CN PCB of the WU
assembly.
Figure 8.2-63
8-46
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the WU_CN PCB.
Figure 8.2-64
3. Remove the screw (M3×8) and either one of the two screws fixing the belt catch
10.
Figure 8.2-65
4. Remove the four screws (M3×6) and dismount the WU_U/PT_R motor
assembly.
Figure 8.2-66
8-47
Chapter 8: Parts replacement
Figure 8.2-67
Figure 8.2-68
2. Measure the belt tension by gently hitting the midst of t he belt span with a
screwdriver, etc.
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
4. Adjust the tension by moving the WU_U/PT_R motor assembly in the arrow
direction.
Figure 8.2-69
8-48
Chapter 8: Parts replacement
3. Remove the four screws (M3×8) and dismount the WU_SEN PCB.
Figure 8.2-70
3. Remove the two screws (M3×8) and dismount the WU_CN PCB.
Figure 8.2-71
8-49
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.2-72
Figure 8.2-73
8-50
Chapter 8: Parts replacement
2. Loosen the hex socket head screw (M3×3) and dismount the stirrer paddle.
Figure 8.3-1
Figure 8.3-2
8-51
Chapter 8: Parts replacement
2. Attach the upper part of the MIX paddle jig on the MIX arm base.
Attach the upper part of the MIX paddle jig on the MIX arm base precisely. If
incorrect fixing, it may cause malfunction in the Stirrer unit.
Figure 8.3-3
3. Tighten the hex socket head screw (M3×3) and fix the stirrer paddle.
8-52
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, and all Mosaic covers.
2. Disconnect the connectors (MIX1U from MIX1, and MIX2U from MIX2) from
MIX_U motor assembly, the connectors (MIX1K from MIX1 and MIX2K from
MIX2) from MIX_K motor assembly, and the connector (UP) from the SENSA
PCB.
Figure 8.3-4
Figure 8.3-5
8-53
Chapter 8: Parts replacement
4. Remove the three screws (M4×10) and dismount the MIX unit assembly.
Figure 8.3-6
Figure 8.3-7
Figure 8.3-8
8-54
Chapter 8: Parts replacement
4. Remove the two screws (M3×8) and dismount the MIX_K motor assembly.
Figure 8.3-9
2. Remove the four screws (M3×10) and dismount the MIX motor plate.
Figure 8.3-10
3. Remove the four screws (M4×10) and dismount the MIX_U motor assembly.
Figure 8.3-11
8-55
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.3-12
Figure 8.3-13
8-56
Chapter 8: Parts replacement
1. Remove the PT arm cover. The PT arm cover is fixed in its arm base by the
inside tabs.
Figure 8.4-1
3. Fix the pipette by spanner (subtense 5.5mm) and rotate the PT joint (tube side)
by hand.
Figure 8.4-2
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
8-57
Chapter 8: Parts replacement
4. Loosen the hex socket head screw (M3×5) and pull out the SPT pipette
assembly.
Figure 8.4-3
1. Follow the reverse procedure of the step 2 to 4 and install the new SPT pipette
assembly.
When tightening the hex socket head screw, pay attention to the tigh tening
torque.
Tighten the screw at 90-degree rightward when placing the screw against the
pipette.
8-58
Chapter 8: Parts replacement
The height from the tip of the pipette to the PT arm base is 128.5mm.
Make sure to move smoothly up and down (in the arrow direction) with holding
the pipette.
Figure 8.4-4
8-59
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R and all Mosaic covers.
2. Turn the M6 joint counterclockwise and remove the tube from the SPPS
assembly.
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Figure 8.4-5
8-60
Chapter 8: Parts replacement
3. Disconnect the two connectors (J1, J6) from the MOTOR-SEN_CN PCB.
Figure 8.4-6
4. Remove the three screws (M4×10) and dismount the SPT assembly.
Figure 8.4-7
8-61
Chapter 8: Parts replacement
2. Loosen the two screws (M3×8) and dismount the SPT stopper.
Figure 8.4-8
3. Remove the three screws (M3×10) and dismount the WU_U/PT_R motor
assembly.
Figure 8.4-9
8-62
Chapter 8: Parts replacement
Figure 8.4-10
Figure 8.4-11
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned range, take the following
steps:
4. Adjust the tension by moving the WU_U/PT_R motor assembly in the arrow
direction.
Figure 8.4-12
8-63
Chapter 8: Parts replacement
2. Remove the four screws (M4×15) and dismount the PT_U motor assembly.
Figure 8.4-13
3. Remove the screw (M3×8) and either one of the two screws fixing the belt catch
10.
Figure 8.4-14
Figure 8.4-15
8-64
Chapter 8: Parts replacement
Figure 8.4-16
6. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
8. Adjust the tension by moving the PT_U motor assembly up and down in the
arrow directions.
Figure 8.4-17
8-65
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the MOTOR_SEN_CN PCB.
Figure 8.4-18
3. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.4-19
8-66
Chapter 8: Parts replacement
Figure 8.4-20
3. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.4-21
8-67
Chapter 8: Parts replacement
Figure 8.4-22
8-68
Chapter 8: Parts replacement
Figure 8.4-23
Figure 8.4-24
Figure 8.4-25
8-69
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, and all Mosaic covers.
2. Turn the M6 joint counterclockwise and remove the tube connecting to the
RPPR assembly.
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Figure 8.4-26
8-70
Chapter 8: Parts replacement
3. Disconnect the two connectors (J1, J6) from the MOTOR-SEN_CN PCB.
Figure 8.4-27
4. Remove the three screws (M4×10) and dismount the RPT assembly.
Figure 8.4-28
8-71
Chapter 8: Parts replacement
8-72
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, and the Mosaic -front.
Figure 8.5-1
3. Disconnect the two tubes (one for the supply tube, the other for the drain tube )
from the SPT trough assembly.
Figure 8.5-2
8-73
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, and the Mosaic -front.
Figure 8.5-3
3. Disconnect the two tubes (one for the supply tube, the other for the drain tube)
from the RPT trough assembly.
Figure 8.5-4
8-74
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the MIX trough.
Figure 8.5-5
3. Disconnect the two tubes (one for the supply tube, the other for the drain tube)
from the MIX trough assembly.
Figure 8.5-6
8-75
Chapter 8: Parts replacement
1. Dismount the side panel R, the rear panel, the Mosaic-front, and the
Mosaic-right.
Figure 8.5-7
5. Remove the two screws (M3×8) and dismount the Trough chamber assembly.
Figure 8.5-8
8-76
Chapter 8: Parts replacement
Figure 8.5-9
3. Remove the cap at the upper part of the Trough chamber assembly and
dismount the Trough chamber float sensor.
Figure 8.5-10
8-77
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, all Mosaic covers, and the front
panel.
2. Disconnect the connectors (LED and CN113) from the ASP/RCU-DRV PCB and
the F-LED PCB.
Figure 8.6-1
3. Disconnect the connectors (CN111, CN118, CN119, CN120, and CN122 for
BCR: option) from the ASP/RCU-DRV PCB of the RCU assembly.
Figure 8.6-2
8-78
Chapter 8: Parts replacement
5. Remove the four screws (M4×10) and dismount the front frame.
Figure 8.6-3
6. Remove the four screws (M4×10) and dismount the ASP assembly.
Figure 8.6-4
8-79
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, all Mosaic covers, and the front
panel.
2. Disconnect the connectors (CN119, CN120) from the ASP/RCU-DRV PCB of the
RCU assembly.
Figure 8.6-5
4. Dismount the ASP cover assembly, the ASP inner tray, and the ASP outer tray.
5. Remove the three hex socket head screws (M4×15) and dismount the S tray up
base.
Figure 8.6-6
8-80
Chapter 8: Parts replacement
6. Remove the two hex socket head screws (M4×6) and dismount the Tray lower
base.
Figure 8.6-7
7. Remove the four screws (M3×8) and dismount the ASP case assembly.
Figure 8.6-8
8-81
Chapter 8: Parts replacement
2. Remove the four screws (M4×15) and dismount the IRU/ASP motor assembly.
Figure 8.6-9
Figure 8.6-10
8-82
Chapter 8: Parts replacement
Figure 8.6-11
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
4. Adjust the tension by moving the IRU/ASP motor assembly in the arrow
direction.
Figure 8.6-12
8-83
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the front panel.
5. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.6-13
Figure 8.6-14
8-84
Chapter 8: Parts replacement
3. Remove the two screws (M3×15) and dismount the Interlock switch.
Figure 8.6-15
1. Dismount the side panel L, the side panel R, the front panel, and the
Mosaic-front.
3. Disconnect the connector (CN119) from the ASP/RCU-DRV PCB of the RCU
assembly.
Figure 8.6-16
8-85
Chapter 8: Parts replacement
4. Remove the screw (M2×5) and dismount the ASP hall sensor assembly.
The ASP hall sensors in the ASP cover and the STAT cover are common parts
in the ASP hall sensor assembly.
Figure 8.6-17
1. Dismount the side panel L, the side panel R, the front panel, a nd the
Mosaic-front.
3. Disconnect the connector (CN120) from the ASP/RCU-DRV PCB of the RCU
assembly.
Figure 8.6-18
8-86
Chapter 8: Parts replacement
4. Remove the screw (M2×5) and dismount the ASP hall sensor assembly.
The ASP hall sensors in the ASP cover and the STAT cover are common parts
in the ASP hall sensor assembly.
Figure 8.6-19
1. Dismount the side panel L, the side panel R, the front panel, and the
Mosaic-front.
Figure 8.6-20
8-87
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the front panel, and the
Mosaic-front.
Figure 8.6-21
8-88
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the Mosaic -front, and the front
panel.
2. Disconnect the connectors (CN111, CN118, CN119, CN120, and CN122) from
the ASP/RCU-DRV PCB.
Figure 8.7-1
3. Disconnect the connector (CN113) to the F-LED PCB from the ASP/RCU-DRV
PCB.
4. Disconnect the connector (CN103) to the PSU from the ASP/RCU -DRV PCB.
5. Disconnect the connector (CN101) to the CNU from the ASP/RCU -DRV PCB.
6. Disconnect the connector (CN102) to the ISE (option) from the ASP/RCU-DRV
PCB.
8-89
Chapter 8: Parts replacement
7. Remove the four screws (M4×10) and dismount the front frame.
Figure 8.7-2
Figure 8.7-3
Figure 8.7-4
8-90
Chapter 8: Parts replacement
3. Remove the three screws (M4×10) and dismount the RCUBCR assembly.
Figure 8.7-5
1. Dismount the side panel L, the side panel R, and the fron t panel.
3. Remove the six screws (M3×8) and dismount the ASP/RCU-DRV PCB.
Figure 8.7-6
8-91
Chapter 8: Parts replacement
2. Remove the four screws (M4×15) and dismount the RCU motor assembly.
Figure 8.7-7
Figure 8.7-8
8-92
Chapter 8: Parts replacement
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
Figure 8.7-9
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
4. Adjust the tension by moving the RCU motor assembly in the arrow direction s.
Figure 8.7-10
8-93
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the RCU drv plt on the
ASP/RCU-DRV PCB.
Figure 8.7-11
5. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.7-12
8-94
Chapter 8: Parts replacement
Figure 8.7-13
8-95
Chapter 8: Parts replacement
3. Remove the four screws (M3×8) and dismount the mounting parts including the
RCU fan assembly.
Figure 8.7-14
4. Remove the two screws (M3×35) and dismount the RCU fan assembly out of the
mounting parts.
Figure 8.7-15
When mounting, mount the RCU fan to the downward arrow direction.
Figure 8.7-16
8-96
Chapter 8: Parts replacement
1. Dismount the RCU cover assembly and the RCU tray assembly.
The screw area of this assembly is covered with the silicon rubber for
waterproof. Remove the silicon rubber with tweezers first. Then remove the
screws.
Figure 8.7-17
5. Dismount the RCU peltier assembly (PE1, PE2) at the bottom of the RCU.
Figure 8.7-18
8-97
Chapter 8: Parts replacement
6. Wipe the heat sink with alcohol-moistened gauze and apply a thin layer of the
heat sink (HSC-50) on the PE cold plate evenly.
Figure 8.7-19
7. Wipe the facing surfaces of both the RCU peltier assembly and the RCU with
alcohol-moistened gauze and apply a thin layer of the heat sink (HSC-50) on the
surface evenly.
8. Coat the silicon rubber over the surface of the screws for waterproof.
Tighten the two screws (M3×10) and mount the RCU peltier assembly. Cover
the tightened screws with the silicon rubber.
Figure 8.7-20
9. Fill the clearance between the RCU case and the RCU peltier assembly with the
silicon rubber.
Figure 8.7-21
8-98
Chapter 8: Parts replacement
1. Dismount the side panel L, the side panel R, the front panel, and the
Mosaic-front.
3. Disconnect the connector (CN116) from the ASP/RCU-DRV PCB of the RCU
assembly.
Figure 8.7-22
5. Remove the two screws (M2×5) and dismount the RCU hall sensor.
Figure 8.7-23
8-99
Chapter 8: Parts replacement
4. Turn the three M6 joints counterclockwise and remove the three tubes
connecting to the syringe S1.4 and the SPP/RPP solenoid valve (COM, NO).
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Figure 8.8-1
8-101
Chapter 8: Parts replacement
5. Remove the four screws (M3×8) and dismount the SPPS assembly.
Figure 8.8-2
4. Turn the three M6 joints counterclockwise and remove the three tubes
connecting to the SPP/RPP solenoid valve (COM, NO) and the clot detection
sensor.
Figure 8.8-3
8-102
Chapter 8: Parts replacement
5. Remove the four screws (M3×8) and dismount the SPPS assembly.
Figure 8.8-4
2. Turn the M6 joint counterclockwise and remove the tube connecting to the
syringe S1.4.
Figure 8.8-5
8-103
Chapter 8: Parts replacement
3. Remove the hex socket head screw (M4×22) and dismount the syringe block.
Figure 8.8-6
4. Remove the two screws (M3×6) and dismount the fixing plate for syringe head.
Figure 8.8-7
Figure 8.8-8
8-104
Chapter 8: Parts replacement
2. Turn the M6 joint counterclockwise and remove the tube connecting to the
SPP/RPP solenoid valve.
Figure 8.8-9
3. Remove the two screws (M3×8) and dismount the SPP/RPP solenoid valve.
Figure 8.8-10
8-105
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.8-11
Figure 8.8-12
8-106
Chapter 8: Parts replacement
The SPPW and the RPPW is a common unit as the SPPW/RPPW assembly.
3. Turn the M6 joint counterclockwise and remove the tube connecting to the
syringe S7.2.
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Figure 8.8-13
8-107
Chapter 8: Parts replacement
4. Remove the three screws (M3×8) and dismount the SPPW/RPPW assembly.
Figure 8.8-14
2. Remove the hex socket head screw (M4×22) and dismount the syringe block.
Figure 8.8-15
8-108
Chapter 8: Parts replacement
3. Remove the two screws (M3×6) and dismount the fixing plate for syringe head.
Figure 8.8-16
Figure 8.8-17
8-109
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.8-18
Figure 8.8-19
8-110
Chapter 8: Parts replacement
2. Remove the two screws (M3×6) and dismount the belt catch 10.
Figure 8.8-20
Figure 8.8-21
Figure 8.8-22
8-111
Chapter 8: Parts replacement
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
Figure 8.8-23
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
4. Loosen the screw (M3×15) and adjust the tension by moving the shaft pulley
vertically as shown in the figure below.
Figure 8.8-24
5. After adjusting the belt tension within the acceptable range, tighten the four
screws (M4×12) and the hex nut (M3).
8-112
Chapter 8: Parts replacement
4. Turn the three M6 joints counterclockwise and remove the tubes connecting to
the syringe R3.2 and the SPP/RPP solenoid valve (COM, NO).
Figure 8.8-25
5. Remove the three screws (M3×8) and dismount the RPPR assembly.
Figure 8.8-26
8-113
Chapter 8: Parts replacement
2. Turn the M6 joint counterclockwise and remove the tube connecting to the
syringe R3.2.
Figure 8.8-27
3. Remove the hex socket head screw (M4×22) and dismount the syringe block.
Figure 8.8-28
4. Remove the two screws (M3×6) and dismount the valve holder R-1.
8-114
Chapter 8: Parts replacement
Figure 8.8-29
Figure 8.8-30
Figure 8.8-31
8-115
Chapter 8: Parts replacement
2. Turn the M6 joint counterclockwise and remove the tube connecting to the
SPP/RPP solenoid valve.
Figure 8.8-32
3. Remove the two screws (M3×6) and dismount the SPP/RPP solenoid valve.
Figure 8.8-33
8-116
Chapter 8: Parts replacement
2. Remove the two screws (M3×8) and dismount the SENSA PCB.
Figure 8.8-34
Figure 8.8-35
8-117
Chapter 8: Parts replacement
The SPPW and the RPPW is a common unit as the SPPW/RPPW assembly.
3. Turn the M6 joint counterclockwise and remove the tube connecting to the
syringe S7.2.
During this procedure, any liquid dripping may be recognized from the tube
connections or the pipette.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Figure 8.8-36
4. Remove the three screws (M3×8) and dismount the SPPW/RPPW assembly.
Figure 8.8-37
8-118
Chapter 8: Parts replacement
8-119
Chapter 8: Parts replacement
3. Turn the eight M6 joints counterclockwise and remove the tubes connecting to
the WPP solenoid valve.
During this procedure, any liquid dripping may be recognized from the tube
connections or the nozzles.
If the M6 joint is too tight to loosen, use long-nose pliers. When using
long-nose pliers, cover the M6 joint with a cloth, etc. to prevent damage.
Prevent deformation or damage of the tube tip.
Figure 8.8-38
8-120
Chapter 8: Parts replacement
5. Remove the three screws (M3×8) and dismount the WPP assembly.
Figure 8.8-39
The WPP assembly consists of the four syringe pumps (WPP1 to WPP4).
The dismounting procedure for the syringe W7.2 (WPP1 to WPP4) is same as
that for the WPP solenoid valve.
2. Remove the hex socket head screw (M4×22) and dismount the syringe block.
Figure 8.8-40
8-121
Chapter 8: Parts replacement
Figure 8.8-41
4. Remove the two hex spacers and dismount the fixing plate for syringe head.
Figure 8.8-42
6. Turn the WPP solenoid valve counterclockwise and dismount it from the syringe
R3.2. Dismount the valve packing.
Figure 8.8-43
8-122
Chapter 8: Parts replacement
2. Remove the two screws (M3×6) and dismount the SENSA PCB.
Figure 8.8-44
Figure 8.8-45
8-123
Chapter 8: Parts replacement
2. Remove the two screws (M3×6) and dismount the belt catch 10.
Figure 8.8-46
Figure 8.8-47
8-124
Chapter 8: Parts replacement
6. Move up the shaft pulley and dismount the WPP timing belt 160.
Figure 8.8-48
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
Figure 8.8-49
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
8-125
Chapter 8: Parts replacement
4. Loosen the screw (M3×15) and adjust the tension by moving the shaft pulley up
and down in the arrow directions.
Figure 8.8-50
5. After adjusting the belt tension within the acceptable range, tighten the four
screws (M4×12) and the hex nut (M3).
8-126
Chapter 8: Parts replacement
Figure 8.8-51
Figure 8.8-52
8-127
Chapter 8: Parts replacement
5. Move up the shaft pulley and dismount the WPP timing belt 192.
Figure 8.8-53
2. Measure the belt tension by gently hitting the midst of the belt span with a
screwdriver, etc.
Figure 8.8-54
The tension of the timing belt may vary depending on the measuring spots.
Perform measurement several times consecutively.
If the belt tension is out of the above-mentioned acceptable range, take the
following steps:
8-128
Chapter 8: Parts replacement
4. Loosen the screw (M3×15) and adjust the tension by moving the motor holder
up and down in the arrow directions.
Figure 8.8-55
5. After adjusting the belt tension within the acceptable range, tighten the four
screws (M4×12) and the hex nut (M3).
8-129
Chapter 8: Parts replacement
2. Remove the three screw caps at the upper part of the subtank and dismount the
subtank float sensor assembly and its tubes.
Figure 8.8-56
8-130
Chapter 8: Parts replacement
Figure 8.9-1
3. Pull out the SWU assembly halfway out of the analyzer and disconnect the three
connectors (SWU-LEAK, SWU1, and SWU2).
Figure 8.9-2
8-131
Chapter 8: Parts replacement
4. Remove the 10 tubes connecting to the diaphragm pumps, the two tubes from
the relay connectors, and the tube from the SWU chamber (option as the ISE
drain tube).
Figure 8.9-3
8-132
Chapter 8: Parts replacement
2. Remove the screw (M3×8) and the connecting lines from the leak sensor
assembly. Dismount the leak sensor assembly.
Figure 8.9-4
3. Remove the two screws (M3×8) and dismount the sponge from the leak sensor
assembly.
Figure 8.9-5
8-133
Chapter 8: Parts replacement
The SWU assembly consists of the diaphragm pump of each of the WU (1 -5),
the SPT, the RPT, the MIX (1-2), and the SUBTK.
Figure 8.9-6
8-134
Chapter 8: Parts replacement
3. Remove the two screws (M3×8) and dismount the diaphragm pump.
Figure 8.9-7
4. Remove the two screws (M3×6) and dismount the pump holder from the
diaphragm pump.
Figure 8.9-8
8-135
Chapter 8: Parts replacement
The check valves are installed at the four diaphragm pumps (SPT, RPT, MIX1
and MIX2).
Figure 8.9-9
2. Disconnect the tubes connecting at the upper and the lower parts of the check
valve. Replace the check valve with the new one.
The liquid flows in one direction in the check valve. Mount the check valve in the
direction of the arrow on the side of the check valve pointing to the analyzer.
Figure 8.9-10
8-136
Chapter 8: Parts replacement
The inline filter N is installed at each of the four diaphragm pumps (WU1 to 4).
Figure 8.9-11
2. Disconnect the tubes from both of the upper and the lower parts of the inline
filter N. Replace the inline filter N with the new one.
The liquid flows in one direction within the inline filter N. Mount the inline filter N
in the direction of the arrow on the side of the inline filter N pointing to the
diaphragm pump.
Figure 8.9-12
8-137
Chapter 8: Parts replacement
8.10 Option
Prior to dismounting the ISE unit, click the [ISE Drain] button from the [Daily
Maintenance] screen to perform drainage of the remaining liquid in the ISE
unit.
1. Dismount the side panel L, the side panel R, and the Mosaic-front.
Figure 8.10-1
3. Disconnect the connector (ISE) from the rear of the ISE unit.
Figure 8.10-2
8-138
Chapter 8: Parts replacement
4. Remove the two screws (M3×8) and pull out the SWU.
Figure 8.10-3
Figure 8.10-4
6. Remove the three screws (M3×8) and dismount the ISE unit.
Figure 8.10-5
8-139
Chapter 8: Parts replacement
8-140
Chapter 8: Parts replacement
A. Dismounting DGU
Two tubes from the S-R divider assembly (one between the S-R divider
assembly and the SPPS, the other between the S-R divider assembly and the
RPPR)
Figure 8.10-6
One tube from the degasser module (M6 joint between the degasser module
and the subtank)
Figure 8.10-7
3. Disconnect the connector (OP-EV) from the degassing solenoid valve assembly.
8-141
Chapter 8: Parts replacement
4. Remove the three screws (M3×8) and pull out the DGU.
Figure 8.10-8
The silicon tube from the Degassing pump (between the degassing pump and
the trough chamber)
Figure 8.10-9
6. Disconnect the connector from the rear of the pressure SW and the connector
(OF-PUMP) from the Degassing pump.
8-142
Chapter 8: Parts replacement
One connecting line from the upper part of the degasser module to the M6 joint
Figure 8.10-10
One connecting line (silicon tube) from the rear of the degasser module
Figure 8.10-11
Figure 8.10-12
8-143
Chapter 8: Parts replacement
Figure 8.10-13
8-144
Chapter 8: Parts replacement
2. Disconnect the connecting line from the rear of the pressure SW (DP-101Z).
Figure 8.10-14
3. Remove the two screws (M3×6) and dismount the pressure SW (DP-101Z).
Figure 8.10-15
8-145
Chapter 8: Parts replacement
2. Disconnect the two connecting lines from the degassing solenoid valve.
Figure 8.10-16
4. Remove the two screws (M2.5×5) from the bottom of the degassing solenoid
valve and dismount the Degassing solenoid valve assembly.
Figure 8.10-17
8-146
Chapter 8: Parts replacement
Figure 8.10-18
3. Remove the two screws (M3×8) and dismount the Degassing pump assembly
with the Base APN30.
Figure 8.10-19
4. Remove the two screws (M3×8) from the bottom of the Base APN30 and
dismount the Degassing pump assembly.
Figure 8.10-20
8-147
Chapter 8: Parts replacement
4. Remove the two screws (M3×6) and dismount the CDS PCB.
Figure 8.10-21
8-148
Chapter 9: Position adjustment of unit
9-1
Chapter 9: Position adjustment of unit
A. Jigs
9-2
Chapter 9: Position adjustment of unit
B. Required Tools
Item Type
Screwdriver + Long Screwdriver No.2
Screwdriver No. 2
Hex Wrench Required size: 1.5 mm
9-3
Chapter 9: Position adjustment of unit
Position adjustment
SPT O ∆ ∆ ∆ × × ×
ASP × O × × × × ×
SPT trough × × O × × × ×
Unit name
ISE × × × O × × ×
RPT × × × × O ∆ ∆
RCU × × × × × O ×
RPT trough × × × × × × O
Position adjustment
MIX1 O ∆ × × × ×
MIX1 trough × O × × × ×
Unit name
MIX2 × × O ∆ × ×
MIX2 trough × × × O × ×
WU × × × × O
IRU × × × × × O
O: Adjustment required, △: Check required. Only if misalignment, adjustment required.
X: No adjustment or check required.
9-4
Chapter 9: Position adjustment of unit
9.3 Preparation
Perform the following preparation prior to the unit position adjustment.
1. Power the analyzer OFF and shut down the operational PC.
2. Dismount all the mosaic covers. For the procedure, refer to “9.1.1. Cover (CVR)
replacement.”
3. Perform level adjustment of the analyzer. For the procedure, refer to “4.3.4
Level adjustment.”
6. Hold and move the SPT, the RPT, the MIX1, or the MIX2 to the trough area by
hand. Make sure to position at the enough space (not to hit other parts.)
9-5
Chapter 9: Position adjustment of unit
2. Remove the curettes (# 22, 24 and 25). Mount the RPT positioning jig at the #
22, MIX positioning jig at the #24, and the SPT positioning jig at the #25.
Figure 9.4-1
5. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
7. Put the ASP cover dummy jig into the shaft hole of the ASP interlock to release
the ASP interlock.
Figure 9.4-2
9-6
Chapter 9: Position adjustment of unit
Select [07 SPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [06 SPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [07 SPTU] at the Motor, [Pulse] at the Mode, and enter “220” at the
Pulse Number. Click the “Execute” icon.
10. Place the height adjustment jig against the MIX positioning jig and the RPT
positioning jig.
Figure 9.4-3
11. Place the height difference A (lower part) of the height adjustment jig under the
tip of the SPT pipette assembly.
Check that the tip of the SPT pipette assembly is positioned at the range of the
height difference A. When it is not positioned, perform vertical position
adjustment of the SPT to position the tip of the pipette at the range of the
height difference A. When adjusting vertical position of the unit, the entered
pulse number becomes the pulse offset value. Record the moving pulse
number.
9-7
Chapter 9: Position adjustment of unit
Figure 9.4-4
Select [07 SPTU] at the Motor, [Pulse] at the Mode, and enter the designated
value at the Pulse Number. Click the “Execute” icon.
9-8
Chapter 9: Position adjustment of unit
12. Refer to “10.3 Analyzer software update” and connect to analyzer software
using the [FFFTP].
Figure 9.4-5
14. Select the file: “mapoffset.txt” and click the “download” icon to save it on the
operational PC holder. After saving the file, double-click the “mapoffset.txt” file
from the operational PC holder.
Figure 9.4-6
15. Modify the value at the “+XX” of “SPTU_OFFSET +XX↑” as specified when
position adjusting. When modifying the value, carefully enter the value not to
change the numeric order in the text.
The “+XX” part in the text has been entered the value that adjusted at the
factory shipment.
9-9
Chapter 9: Position adjustment of unit
Figure 9.4-7
17. Select the modified “mapoffset.txt” file from the drop-down menu at the [FFFTP]
file and click the “Upload” icon to upload to “/home/ca270/sysboot/system”.
18. Select “Disconnect” from the “File” menu at the [FFFTP] screen and disconnect
the connection with the analyzer software.
19. Power the analyzer OFF and shut down the operational PC.
9-10
Chapter 9: Position adjustment of unit
Select [07: SPTU] at the Motor and [Search] at the Mode and click the
“Execution” icon.
Select [06: SPTR] at the Motor and [Search] at the Mode and click the
“Execution” icon.
3. Move the SPT downward to move the tip of the SPT pipette assembly towards
the SPT positioning jig.
Select [07: SPTU] at the Motor, [Pulse] at the Mode, and enter “300” at the
Pulse Number. Click the “Execute” icon.
4. Adjust the SPT position to fit the tip of the SPT pipette assembly in the convex
area of the center of the SPT positioning jig.
Figure 9.4-8
5. After adjusting the position, tighten the three screws (M4×10) that tentatively
fastened the SPT assembly.
Select [07 SPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9-11
Chapter 9: Position adjustment of unit
Select [12: ASP] at the Motor and [Search] at the Mode and click the “Execute”
icon.
Select [07: SPTU] at the Motor and [Search] at the Mode and click th e
“Execute” icon.
Select [06: SPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
4. Rotate the SPT to the aspiration position at the ASP outer layer.
Select [06: SPTR] at the Motor, [Move] at the Mode and [SPTRASPOUT] at the
Position. Click the “Execute” icon.
5. Adjust the ASP position to fit the tip of the SPT pipette assembly in the tip of the
ASP positioning jig (OU-CEN).
Figure 9.5-1
6. Rotate the ASP to the aspiration position at the ASP inner layer.
Select [12: ASP] at the Motor and [Move] at the Mode, and [ASPIN] at the
Position. Click the “Execute” icon.
7. Rotate the SPT to the aspiration position at the ASP inner layer.
Select [06: SPTR] at the Motor and [Move] at the Mode, and [SPTRASPIN] at
the Position. Click the “Execute” icon.
8. Adjust the ASP position to fit the tip of the SPT pipette assembly in the tip of the
ASP positioning jig (IN).
9. After position adjustment of the ASP trays (outer, midst, and inner layers),
tighten the four fixing screws (M4×10) that tentatively fasten the ASP.
9-12
Chapter 9: Position adjustment of unit
Select [07: SPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [06: SPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [06: SPTR] at the Motor and [Move] at the Mode, and [SPTRTRD] at the
Position. Click the “Execute” icon.
3. Loosen the fixing screws and adjust the trough position to fit the tip of the SPT
pipette assembly at the center of the trough wash position.
Figure 9.6-1
Figure 9.6-2
Select [06: SPTR] at the Motor and [Move] at the Mode, and [SPTRDX] at the
Position. Click the “Execute” icon.
9-13
Chapter 9: Position adjustment of unit
5. Move the tip of the SPT pipette assembly to the drain outlet of the trough.
Select [07 SPTU] at the Motor, [Pulse] at the Mode, and enter “1000” at the
Pulse Number. Click the “Execute” icon.
6. Adjust the trough position to fit the tip of the SPT pipette assembly at the center
of the trough drain outlet.
7. After position adjustment of the SPT pipette assembly at the trough, tighten the
fixing screws for the trough.
9-14
Chapter 9: Position adjustment of unit
Select [07: SPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [06: SPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [06: SPTR] at the Motor and [Move] at the Mode, and [SPTRISE] at the
Position. Click the “Execute” icon.
3. Move the tip of the SPT pipette assembly 7mm towards the ISE sample port.
Select [07: SPTU] at the Motor, [Pulse] at the Mode, and enter “280” at the
Pulse Number. Click the “Execute” icon.
4. Adjust the ISE aspiration positions as indicated each position of the tip of the
SPT pipette assembly and the ISE sample port below. When adjusting, loosen
the two fixing screws (M3×8) for the ISE sample port.
Figure 9.7-1
5. Tighten the two fixing screws (M3×8) for the ISE sample port.
Figure 9.7-2
9-15
Chapter 9: Position adjustment of unit
6. Power the analyzer OFF and shut down the operational PC.
7. Dismount the cuvette holder for the cuvette # 19 to 27 on the IRU and replace
the position adjustment jig with cuvettes.
9-16
Chapter 9: Position adjustment of unit
2. Dismount the cuvettes at the positions # 22, 24, and 25. Install the RPT
positioning jig at the position # 22 and the RPT positioning jig at the position #
24, and the SPT positioning jig at the position # 25.
Figure 9.8-1
5. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
Select [09: IRU] at the Motor and [Pulse] at the Mode and enter “-5100” at the
Pulse Number. Click the “Execute” icon.
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9-17
Chapter 9: Position adjustment of unit
Select [04: RPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter “220” at the
Pulse Number. Click the “Execute” icon.
10. Place the height adjustment jig against the RPT positioning jig and the SPT
positioning jig.
Figure 9.8-2
11. Place the height difference A (lower part) of the height adjustment jig under the
tip of the RPT pipette assembly.
Check that the tip of the pipette is positioned at the range of the height
difference A. When it is not positioned, perform vertical position adjustment of
the RPT to position the tip of the pipette at the range of the height difference A.
When adjusting vertical position of the unit, the entered pulse number
becomes the pulse offset value. Record the moving pulse number.
Figure 9.8-3
9-18
Chapter 9: Position adjustment of unit
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter the designat ed
value at the Pulse Number. Click the “Execute” icon.
9-19
Chapter 9: Position adjustment of unit
12. Refer to “10.3 Analyzer software update” and connect to analyzer software
using the [FFFTP].
Figure 9.8-4
14. Select the file: “mapoffset.txt” and click the “download” icon to save it on the
operational PC holder. After saving the file, double-click the “mapoffset.txt” file
from the operational PC holder.
Figure 9.8-5
15. Modify the value at the “+XX” of “RPTU_OFFSET +XX” as specified when
position adjusting. When modifying the value, carefully enter the value not to
change the numeric order in the text.
The “+XX” part in the text has been entered the value that adjusted at the
factory shipment.
9-20
Chapter 9: Position adjustment of unit
Figure 9.8-6
17. Select the modified “mapoffset.txt” file from the drop-down menu at the left side
of the [FFFTP] and click the “Upload” icon to upload to
“/home/ca270/sysboot/system”.
18. Select “Disconnect” from the “File” menu at the [FFFTP] screen and disconnect
the connection with the analyzer software.
19. Power the analyzer OFF and shut down the operational PC.
9-21
Chapter 9: Position adjustment of unit
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [04: RPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [09: IRU] at the Motor and [Pulse] at the Mode and enter “-5100” at the
Pulse Number. Click the “Execute” icon.
4. Move the RPT downward to move the tip of the RPT pipette assembly towards
the RPT positioning jig.
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter “300” at the
Pulse Number. Click the “Execute” icon.
5. Adjust the RPT position to fit the tip of the RPT pipette assembly at the convex
area of the center of the RPT positioning jig.
Figure 9.8-7
6. After adjusting the position, tighten the three screws (M4×10) that tentatively
fastened the RPT assembly.
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9-22
Chapter 9: Position adjustment of unit
Select [13: RCU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [04: RPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [04: RPTR] at the Motor and [Move] at the Mode, and [RPTRRCUIN] at
the Position. Click the “Execute” icon.
5. Move the RPT downward to move the tip of the RPT towards the opening of the
70ml reagent bottle.
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter “500” at the
Pulse Number. Click the “Execute” icon.
6. Adjust the RCU position to position the tip of the RPT pipette assembly at the
center of the opening of the 70ml reagent bottle of the position # 1.
Figure 9.9-1
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9-23
Chapter 9: Position adjustment of unit
8. Rotate the RCU to the aspiration position at the RCU outer layer.
Select [13: RCU] at the Motor and [Move] at the Mode, and [RCURGOUTRPT]
at the Position. Click the “Execute” icon.
Select [04: RPTR] at the Motor and [Move] at the Mode, and [RPTRRCUOUT]
at the Position. Click the “Execute” icon.
10. Move the RPT downward to move the tip of the RPT pipette assembly towards
the RPT positioning jig.
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter “500” at the
Pulse Number. Click the “Execute” icon.
11. Adjust the RCU position to position the tip of the RPT pipette assembly at the
center of the opening of the 20ml reagent bottle of the position # 26.
Figure 9.9-2
12. After position adjustment of the RCU trays (outer and inner), tighten the four
fixing screws (M4×10) that tentatively fasten the RCU.
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9-24
Chapter 9: Position adjustment of unit
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [04: RPTR] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [04: RPTR] at the Motor and [Move] at the Mode, and [RPTRTRD] at the
Position. Click the “Execute” icon.
3. Loosen the fixing screws and adjust the trough position to position the tip of the
RPT pipette assembly at the center of the trough wash position.
Figure 9.10-1
Select [04: RPTR] at the Motor and [Move] at the Mode, and [RPTRDX] at the
Position. Click the “Execute” icon.
5. Move the tip of the RPT pipette assembly to the drain outlet of the trough.
Select [05: RPTU] at the Motor, [Pulse] at the Mode, and enter “1000” at the
Pulse Number. Click the “Execute” icon.
6. Loosen the fixing screws and adjust the trough position to position the tip of the
9-25
Chapter 9: Position adjustment of unit
7. After position adjustment of the RPT pipette assembly at the trough, tighten the
fixing screws for the trough.
Select [05: RPTU] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9. Power OFF the analyzer and shut down the operational PC.
10. Dismount the cuvette holder for the cuvette # 19 to 27 on the IRU and replace
the position adjustment jigs with cuvettes.
9-26
Chapter 9: Position adjustment of unit
Prior to performing the MIX1 height adjustment, make sure to dismount the
stirrer paddle.
3. Remove the cuvettes (# 23, 25 and 26). Mount the RPT positioning jig at the
position # 23, SPT positioning jig at the position #25, and the MIX positioning jig
at the position #26.
Figure 9.11-1
6. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
Select [00: MIX1U] at the Motor and [Pulse] at the Mode, and enter “1188” at
the Pulse Number. Click the “Execute” icon.
8. Place the height adjustment jig against the SPT positioning jig and the RPT
positioning jig.
9. Place the height difference B (higher part) of the height adjustment jig under the
convex part at the bottom of the mounting position of the MIX_K motor of the
MIX1 arm base.
9-27
Chapter 9: Position adjustment of unit
Move the MIX1 vertically and adjust to position the convex part at the bottom
of the mounting position of the MIX_K motor of the MIX1 arm base at the range
of the 0.3mm height difference B (higher part) of the height adjustment jig.
When adjusting vertical position of the unit, the actual moving height becomes
the pulse offset value. Record the moving height.
Figure 9.11-2
Select [00: MIX1U] at the Motor, [Pulse] at the Mode, and enter the designated
value at the Pulse Number. Click the “Execute” icon.
10. Refer to “10.3 Analyzer software update” and connect to analyzer software
using the [FFFTP].
9-28
Chapter 9: Position adjustment of unit
Figure 9.11-3
12. Select the file: “mapoffset.txt” and click the “download” icon to save it on the
operational PC. After saving the file, double-click the “mapoffset.txt” file from the
operational PC holder.
Figure 9.11-4
13. Modify the value at the “+XX” of “MIX1U_OFFSET +XX” as specified when
position adjusting. When modifying the value, carefully enter the value not to
change the numeric order in the text.
The “+XX” part in the text has been entered the value that adjusted at the
factory shipment.
Figure 9.11-5
9-29
Chapter 9: Position adjustment of unit
15. Select the modified “mapoffset.txt” file from the drop-down menu at the left side
of the [FFFTP] and click the “Upload” icon to upload to
“/home/ca270/sysboot/system”.
16. Select “Disconnect” from the “File” menu at the [FFFTP] screen and disconnect
the connection with the analyzer software.
17. Power OFF the analyzer and shut down the operational PC.
Select [00: MIX1U] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [00: MIX1U] at the Motor, [Pulse] at the Mode, and enter “880” at the
Pulse Number. Click the “Execute” icon.
4. Adjust the MIX1 position to position the MIX1 stirrer paddle at the center of the
MIX positioning jig.
Figure 9.11-6
5. After adjusting the position, tighten the three screws (M4×10) that tentatively
fastened the MIX1.
9-30
Chapter 9: Position adjustment of unit
Select [00: MIX1U] at the Motor and [Search] at the Mode and click the
“Execute” icon.
7. Dismount the cuvette holder for the cuvette # 19 to 27 on the IRU and replace
the position adjustment jig with cuvettes.
9-31
Chapter 9: Position adjustment of unit
Select [00: MIX1U] at the Motor and [Search] at the Mode and click the
“Execute” icon.
2. Loosen the two fixing screws (M3×8) for the trough and adjust the MIX1 trough
position to position the stirrer paddle at the center of the trough wash position.
Figure 9.12-1
3. After position adjusting, tighten the two fixing screws (M3×8) for the MIX trough.
9-32
Chapter 9: Position adjustment of unit
Prior to performing the MIX2 height adjustment, make sure to dismount the
stirrer paddle.
3. Remove the cuvettes (# 23, 25 and 26). Mount the RPT positioning jig at the
position # 23, SPT positioning jig at the position #25, and the MIX positioning jig
at the position #26.
Figure 9.13-1
6. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
Select [09: IRU] at the Motor, [Pulse] at the Mode, and enter “-5200” at the
Pulse Number. Click the “Execute” icon.
Select [00: MIX1U] at the Motor and [Pulse] at the Mode, and enter “1188” at
the Pulse Number. Click the “Execute” icon.
9. Place the height adjustment jig against the SPT positioning jig and the RPT
positioning jig.
9-33
Chapter 9: Position adjustment of unit
10. Place the height difference B (higher part) of the height adjustment jig under the
convex part at the bottom of the mounting position of the MIX_K motor of the
MIX2 arm base.
Move the MIX2 vertically and adjust to position the convex part at the bottom
of mounting position of the MIX_K motor of the MIX2 arm base at the range of
the 0.3mm height difference B (higher part) of the height adjustment jig. When
adjusting vertical position of the unit, the actual moving height becomes the
pulse offset value. Record the numerical value of the moving height.
Figure 9.13-2
Select [01: MIX2U] at the Motor, [Pulse] at the Mode, and enter the designated
value at the Pulse Number. Click the “Execute” icon.
9-34
Chapter 9: Position adjustment of unit
11. Refer to “10.3 Analyzer software update” and connect to analyzer software
using the [FFFTP].
Figure 9.13-3
13. Select the file: “mapoffset.txt” and click the “download” icon to save it on the
operational PC holder. After saving the file, double-click the “mapoffset.txt” file
from the operational PC holder.
Figure 9.13-4
14. Modify the value at the “+XX” of “MIX2U_OFFSET +XX” as specified when
position adjusting. When modifying the value, carefully enter the value not to
change the numeric order in the text.
The “+XX” part in the text has been entered the value that adjusted at the
factory shipment.
9-35
Chapter 9: Position adjustment of unit
Figure 9.13-5
16. Select the modified “mapoffset.txt” file from the drop-down menu at the left side
of the [FFFTP] and click the “Upload” icon to upload to
“/home/ca270/sysboot/system”.
17. Select “Disconnect” from the “File” menu at the [FFFTP] screen and disconnect
the connection with the analyzer software.
Select [01: MIX2U] at the Motor and [Search] at the Mode and click the
“Execute” icon.
Select [09: IRU] at the Motor and [Search] at the Mode and click the “Execute”
icon.
Select [09: IRU] at the Motor, [Pulse] at the Mode, and enter “-4800” at the
Pulse Number. Click the “Execute” icon
Select [01: MIX2U] at the Motor, [Pulse] at the Mode, and enter “880” at the
Pulse Number. Click the “Execute” icon.
6. Adjust the MIX2 position to position the MIX2 stirrer paddle at the center of the
MIX positioning jig.
9-36
Chapter 9: Position adjustment of unit
Figure 9.13-6
7. After adjusting the position, tighten the three screws (M4×10) that tentativ ely
fastened the MIX2 assembly.
Select [01: MIX2U] at the Motor and [Search] at the Mode and click the
“Execute” icon.
9. Dismount the cuvette holder for the cuvette # 19 to 27 on the IRU and replace
the position adjustment jig with cuvettes.
9-37
Chapter 9: Position adjustment of unit
2. Loosen the two fixing screws (M3×8) for the trough and adjust the MIX2 trough
position to position the stirrer paddle at the center of the trough wash position.
Figure 9.14-1
3. After position adjusting, tighten the two fixing screws (M3×8) for the MIX2
trough.
9-38
Chapter 9: Position adjustment of unit
Select [09: IRU] at the Motor and [Pulse] at the Mode and enter “2500” at the
Pulse Number. Click the “Execute” icon.
2. Dismount the cuvette holder for cuvette # 37 to 45 on the IRU. Remove the
cuvettes (# 37, 43 and 45).
3. Mount the WU positioning jigs at the cuvette # 37 and # 43. Mount each of the
jigs placing the WU side up. Position the hole of each jig at the center of the
IRU.
4. Mount the MIX positioning jig at the cuvette # 45. Mount the cuvette holder for
the cuvette # 37 to # 45.
Figure 9.15-1
6. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
Select [09: IRU] at the Motor, [Pulse] at the Mode, and enter “-200” at the
Pulse Number. Click the “Execute” icon.
Select [08: WU] at the Motor, [Pulse] at the Mode, and enter “50” at the Pulse
Number. Click the “Execute” icon.
9-39
Chapter 9: Position adjustment of unit
10. Place the height difference A (lower part) of the height adjustment jig under the
tip of the WU nozzle.
Move the WU vertically and adjust to position the tip of the WU1 nozzle at the
range of the 0.2mm height difference (lower part) of the height adjustment jig
as shown below. When adjusting vertical position of the unit, the actual moving
height becomes the pulse offset value. Record the numerical value of the
moving height.
Figure 9.15-2
Select [08: WU] at the Motor, [Pulse] at the Mode, and enter the designated
value at the Pulse Number. Click the “Execute” icon.
9-40
Chapter 9: Position adjustment of unit
11. Refer to “10.3 Analyzer software update” and connect to analyzer software
using the [FFFTP].
Figure 9.15-3
13. Select the file: “mapoffset.txt” and click the “download” icon to save it on the
operational PC holder. After saving the file, double-click the “mapoffset.txt” file
from the operational PC holder.
Figure 9.15-4
14. Modify the value at the “+XX” of “WU_OFFSET +XX” as specified when
position adjusting. When modifying the value, carefully enter the value not to
change the numeric order in the text.
The “+XX” part in the text has been entered the value that adjusted at the
factory shipment.
9-41
Chapter 9: Position adjustment of unit
Figure 9.15-5
16. Select the modified “mapoffset.txt” file from the drop-down menu at the left side
of the [FFFTP] and click the “Upload” icon to upload to
“/home/ca270/sysboot/system”.
17. Select “Disconnect” from the “File” menu at the [FFFTP] screen and disconnect
the connection with the analyzer software.
2. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
Select [09: IRU] at the Motor, [Pulse] at the Mode, and enter “-100” at the
Pulse Number. Click the “Execute” icon.
Select [08: WU] at the Motor, [Pulse] at the Mode, and enter “150” at the Pulse
Number. Click the “Execute” icon.
5. Adjust to position each tip of the WU1 supply nozzle and the WU4 supply nozzle
at the hole of the WU positioning jig. Tighten the two fixing screws (M4×10) to fix
the WU.
9-42
Chapter 9: Position adjustment of unit
Figure 9.15-6
Select [08: WU] at the Motor, [Search] at the Mode, and click the “Execute”
icon.
Select [09: IRU] at the Motor, [Pulse] at the Mode, and enter “-4500” at the
Pulse Number. Click the “Execute” icon.
8. Dismount the cuvette holder for the cuvette # 37 to 45 on the IRU and replace
the position adjustment jig with cuvettes.
9-43
Chapter 9: Position adjustment of unit
Select [09: IRU] at the Motor and [Search] at the Mode and click the “Execute”
icon.
2. Loosen the two fixing screws (M3×8) fixing the Wipe base.
Figure 9.15-7
Select [08: WU] at the Motor and [Search] at the Mode and click the “Execute”
icon.
4. Move the WU downward and check the position between the wipe tip and
cuvette.
Select [08: WU] at the Motor, [Pulse] at the Mode, and enter “400” at the Pulse
Number. Click the “Execute” icon.
5. Adjust the wipe tip position including its angle by moving the wipe base to
position the wipe tip at the center from each side of the cuvette.
9-44
Chapter 9: Position adjustment of unit
Adjust the wipe tip position at the even clearances (A and A", and B and B”)
between cuvette and the wipe tip by rotating it.
Figure 9.15-8
7. Select [08: WU] at the Motor and [Search] at the Mode and click the “Execute”
icon.
Select [08: WU] at the Motor and [Move] at the Mode, and [WUUDL] at the
Position. Click the “Execute” icon.
9. Check that there is approx. 1mm clearance between the wipe holder and the
wipe base. When the clearance is not appropriate (wider or smaller), loosen the
fixing screw (M3×5) and adjust the height of the wipe holder to make the 1mm
(approx.) clearance.
Figure 9.15-9
9-45
Chapter 9: Position adjustment of unit
Figure 9.16-1
3. Click the [Initialization] button in the [Maintenance (F10)] > [Sequence] screen.
Perform initialization.
4. Insert the IRU table adjuster into the hole for position adjustment.
Figure 9.16-2
9-46
Chapter 9: Position adjustment of unit
Insert the IRU table adjuster into the cuvette table to the bottom and tighten
the screw.
Figure 9.16-3
6. Fix the cuvette table assembly by tightening the four screws (M3×10).
Figure 9.16-4
7. Remove the screw (M3×12) and dismount the IRU table adjuster.
9-47
Chapter 9: Position adjustment of unit
1. Power the analyzer OFF and shut down the operational PC.
3. Mount all the mosaic covers to the original positions. For the procedure, refer to
“8.1.1. Replacement of covers (CVR).”
9-48
Chapter 10: Software setup and update
1. Click [Start] > [Settings] > [Control Panel] and display the “Control Panel”
window.
2. Click the “Network and Internet Connections” icon to display the “Network and
Internet Connections” window.
Figure 10.1-1
10-1
Chapter 10: Software setup and update
Figure 10.1-2
Figure 10.1-3
10-2
Chapter 10: Software setup and update
5. Select “Internet Protocol (TCP/IP)” in the [General] tab. Click the [Properties].
Figure 10.1-4
6. Tick "Use the following IP address:”. Specify the following IP address and click
the [OK] button.
Figure 10.1-5
7. Click the [X] button to close the window. The setup is completed.
10-3
Chapter 10: Software setup and update
1. Click [Start] > [Settings] > [Control Panel] and display the “Control Panel”
window.
2. Click the [Security] > [Security Center] > [Windows Firewall] and display the
“Windows Firewall” window.
Figure 10.1-6
3. Select [General] tab and tick “Off (not recommended).”. Click the [OK] button.
Figure 10.1-7
10-4
Chapter 10: Software setup and update
1. Click [Start] > [Settings] > [Control Panel] and display the “Control Panel”
window.
2. Click the [Security Center] > [Automatic updates] and display the “Automatic
updates” window.
Figure 10.1-8
3. Select [Automatic updates] tab and tick “Turn off Automatic Updates”. Click the
[OK] button.
Figure 10.1-9
10-5
Chapter 10: Software setup and update
1. Click [Start] > [Settings] > [Control Panel] and display the “Control Panel”
window.
2. Click the [Appearance and Themes] > [Display] and display the “Display
Properties” window.
Figure 10.1-10
Figure 10.1-11
4. Click the [Power…] button and display “Power Options Properties” window.
Figure 10.1-12
10-6
Chapter 10: Software setup and update
Figure 10.1-13
Power buttons
□ when I press the power button on my computer: Do nothing
Figure 10.1-14
10-7
Chapter 10: Software setup and update
Figure 10.1-15
10-8
Chapter 10: Software setup and update
1. Click [Start] >[Settings] > [Control Panel] and the “Date, Time, Language, and
Regional Options” icon. Click “Regional Language Options” icon and display the
window.
Figure 10.1-16
2. Select the [Regional Options] tab and specify "Standards and formats”.
(a) Select the item to use the language (country) from drop-down menu.
10-9
Chapter 10: Software setup and update
Figure 10.1-17
3. Click the [Customize…] button and display the “Customize Regional Options”
window.
Figure 10.1-18
10-10
Chapter 10: Software setup and update
Figure 10.1-19
Figure 10.1-20
5. Select the [Languages] tab and specify "Text services and input languages”.
(a) Click the [Details…] button and display “Text services and input
languages” window.
(b) Select the country name in “Default input language” from drop-down
menu.
(c) When the country name not displayed in the drop-down menu, click the
[Add] button to add the name.
(d) Click the [OK] button.
10-11
Chapter 10: Software setup and update
Figure 10.1-21
1. Click [Start] > [Control Panel] > [Printers and Other Hardware] and display the
“Printers and faxes” window.
Figure 10.1-22
2. Select and right-click the printer icon and display the “Printer Properties”
window.
Figure 10.1-23
10-13
Chapter 10: Software setup and update
1. Click [Start] > [Programs] > [Accessories] > [Communications] > [Hyper
Terminal].
Figure 10.1-24
Figure 10.1-25
10-14
Chapter 10: Software setup and update
3. The “location information” window appears. Enter the following information and
click the [OK] button.
Figure 10.1-26
4. The “Phone and Modem Options” window appears. Confirm that [My Location] is
ticked. Click the [OK] button.
Figure 10.1-27
10-15
Chapter 10: Software setup and update
Figure 10.1-28
Figure 10.1-29
7. Enter IP address (172.17.31.202) at the Host address:. Click the [OK] button.
Figure 10.1-30
10-16
Chapter 10: Software setup and update
8. After the setting completed, the communication between the operational PC and
the analyzer will be established. Activate the [Hyper Terminal] and log in the
analyzer. All programs are ready for execution.
Figure 10.1-31
9. To terminate the [Hyper Terminal], click the “Disconnect” icon from menu bar.
Figure 10.1-32
10-17
Chapter 10: Software setup and update
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
Figure 10.1-33
Figure 10.1-34
Figure 10.1-35
10-18
Chapter 10: Software setup and update
Figure 10.1-36
Figure 10.1-37
(d) Select the [Kanji Code] tab and specify the following.
Japanese Kanji Code: No conversion [S-JIS]
Filename Kanji Code: SHIFT-JIS
10-19
Chapter 10: Software setup and update
Figure 10.1-38
Figure 10.1-39
Figure 10.1-40
10-20
Chapter 10: Software setup and update
1. Click [Start] > [Control Panel] > [Network and Sharing Center].
2. Click [Manage network connections] > “Local Area Connection” icon. Right-click
the "Local Area Connection” icon and select “Properties”.
Figure 10.2-1
3. Tick “Internet Protocol Version 4 (TCP/Ipv4) and click the [Properties] button in
the [Networking] tab.
4. Tick “Use the following IP address" in the [General] tab and enter the following
information.
IP address: 172.17.31.201
Subnet mask: 255.255.0.0.
Default gateway: (blank)
Figure 10.2-2
Figure 10.2-3
2. Click the [Windows Firewall] and display the “Windows Firewall” window.
3. Click the [Change Settings] and display the “Windows Firewall Settings”
window.
Figure 10.2-4
4. Select [General] tab and tick “Off (not recommended)”. Click the [OK] button.
Figure 10.2-5
10-22
Chapter 10: Software setup and update
2. Click “Windows update” and display the “Windows Security Center” window.
3. Click “Automatic updating” and “Check for updates” and display the “Change
settings” window.
Figure 10.2-6
4. Tick “Never check for update (not recommended)”. Click the [OK] button.
Figure 10.2-7
10-23
Chapter 10: Software setup and update
2. Click “Windows update” and display the “Windows Security Center” window.
3. Click [Malware protection] > [Show my available protections] and display the
“Windows Security Center” window.
Figure 10.2-8
4. Tick “Don’t monitor my antivirus software state (not recommended)”. Click the
[OK] button.
Figure 10.2-9
10-24
Chapter 10: Software setup and update
2. Click the [Screen saver] and display the “Screen Saver Settings” window.
Figure 10.2-10
Figure 10.2-11
10-25
Chapter 10: Software setup and update
4. Click the [Change power setting…] and display “Power Options” window.
Figure 10.2-12
5. Click the [Change plan settings] at the Preferred plans and display the “Change
settings for the plan: Balanced” window.
Figure 10.2-13
6. Specify the setting in the “Change settings for the plan: Balanced” window.
(a) Turn off the display : Never
(b) Put the computer to sleep: Never
10-26
Chapter 10: Software setup and update
Figure 10.2-14
7. Click the [Change advanced power settings] and display "Power Options”
window.
Figure 10.2-15
Figure 10.2-16
10-27
Chapter 10: Software setup and update
Figure 10.2-17
Figure 10.2-18
10-28
Chapter 10: Software setup and update
1. Click [Start] > [Control Panel] > [Regional and Language Options].
2. Select the [Formats] tab in the “Regional and Language Options” window.
(a) Select the country name from drop-down menu.
Figure 10.2-19
10-29
Chapter 10: Software setup and update
3. Click the [Customize this format…] button and display the “Customize Regional
Options” window.
Figure 10.2-20
(a) Specify in the [Numbers] tab and click the [Apply] button.
Decimal symbol: “.”
Figure 10.2-21
10-30
Chapter 10: Software setup and update
Figure 10.2-22
(c) Specify in the [Date] tab and click the [Apply] button.
Short date: yy/MM/dd
Long date: yy/MM/dd
Figure 10.2-23
4. Select the [Location] tab in the “Regional and Language Options” window.
(a) Select the country name from drop-down menu.
Figure 10.2-24
10-31
Chapter 10: Software setup and update
5. Select the [Keyboards and Languages] tab in the “Regional and Language
Options” window.
(a) Click the [Change keyboards…] button and display the “Text Services and
Input Languages" window.
Figure 10.2-25
Figure 10.2-26
6. Select the [Administrative] tab in the “Regional and Language Options” window.
(a) Click the [Change system locale…] button and display the "Regional and
Language Options" window.
10-32
Chapter 10: Software setup and update
Figure 10.2-27
Figure 10.2-28
10-33
Chapter 10: Software setup and update
Figure 10.2-29
Figure 10.2-30
10-34
Chapter 10: Software setup and update
Daylight saving time is utilized depending on the countries and regions. Tick
the “Automatically adjust clock for Daylight Saving time” for adopting when it
appears in the setting window.
Figure 10.2-31
2. Click the [Change time zone…] button in the [Date and Time] tab and display the
“Time zone settings” window.
Figure 10.2-32
Figure 10.2-33
2. Select and right-click the “Printers” icon and display the “Printer Properties”
window.
Figure 10.2-34
10-36
Chapter 10: Software setup and update
1. Click [Start] > [Control Panel] > [Programs and features] and display “Program
and functions" window.
2. Click the [Turn Windows features on or off] and display the “Windows features”
window.
Figure 10.2-35
Figure 10.2-36
10-37
Chapter 10: Software setup and update
Figure 10.2-37
6. To terminate this session, click the [X] button at the upper-right corner.
10-38
Chapter 10: Software setup and update
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
Figure 10.2-38
Figure 10.2-39
Figure 10.2-40
10-39
Chapter 10: Software setup and update
Figure 10.2-41
(d) Select the [Kanji Code] tab and specify the following.
Japanese Kanji Code: No conversion [S-JIS]
Filename Kanji Code: SHIFT-JIS
Figure 10.2-42
10-40
Chapter 10: Software setup and update
Figure 10.2-43
Figure 10.2-44
10-41
Chapter 10: Software setup and update
Figure 10.3-1
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
Figure 10.3-2
5. Double-click the [update] folder at the right side of the screen and display the
[/home/CA270/update] contents.
6. Double click the folder that contains the analyzer update software at the left side
and display them.
7. Select the analyzer update software (CA270.tar.F) and click the "Upload" icon
on the tool bar. The data (CA270.tar) is transferred to the [update] folder in t he
analyzer software.
Figure 10.3-3
Figure 10.3-4
13. The data in the analyzer software is automatically updated. Wait until the
processing completed.
14. When “ca270>_" appears after the update completed, enter “logout” to
terminate [Hyper Terminal].
10-43
Chapter 10: Software setup and update
1. Select [System (F9] > [Backup Procedures]. Click the […] button on the
“Directory” for system parameters backup.
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
1. Click [Start] > [Control Panel] in Windows XP to display the “Control Panel”
window.
2. Click [Add or Remove Programs] and select [Analyzer]. Click the [Remove]
button.
1. Save the update program [ca_setup.exe] on desktop from the external memory
medium (such as USB memory).
4. After installation, the following window appears to restart the operational PC.
Click the [Yes] button. Click [Start] > [Turn Off Computer…] > [Restart].
5. Restart the operational PC. Double-click the shortcut "CAMAIN" and start the
operational software.
1. Select [System (F9] > [Backup Procedures]. Click the […] button on the
“Directory” for system parameters backup.
10-44
Chapter 10: Software setup and update
3. Press and hold the [Ctrl] and [.] keys. Click the [Power Off] button and terminate
the operational software.
4. Double click the shortcut "CAMAIN" and restart the operational software.
F. Final Confirmation
1. Select [System (F9)] > [Versions]. Check the version number of the "User
interface” is updated.
2. Select [System (F9)] > [Sequence]. Click the [Initialization] button and check the
initialization is performed properly.
10-45
Chapter 10: Software setup and update
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
Figure 10.5-1
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
5. Click the “download” icon from the toolbar and save the “sytem.txt” file into c:
drive in the operational PC.
Figure 10.5-2
6. Use text editor software (such as Notepad) to open the “system.txt” file. Modify
and save the file.
7. Select the modified “system. txt” at the “FFFTP” window (left side of screen).
8. Click the “upload” icon from the toolbar and save the modified “sytem.txt” file
into the [/home/ca270/sysboot/system] folder in the operational software.
Figure 10.5-3
9. Double-click the modified “system. txt” uploaded in the “FFFTP” window (left
side of screen) and check that the file is the modified version. After confirmatio n,
10-46
Chapter 10: Software setup and update
click the [X] button on the “FFFTP” window and exit the [FFFTP].
After replacing or repairing the SPT, the RPT, the MIX1, the MIX2, or the WU,
the offset file needs to be updated only when adjustment for the height
direction is required.
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
Figure 10.5-4
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
5. Click the “download” icon from the toolbar and save the file into c: drive in the
operational PC.
Figure 10.5-5
6. Use text editor software (such as Notepad) to open the file. Modify and save the
file.
7. Select the modified file at the “FFFTP” window (left side of screen).
8. Click the “upload” icon from the toolbar and save the modified file into the
[/home/ca270/sysboot/pmap] folder in the operational software.
9. Double-click the uploaded file in the “FFFTP” window (left side of screen) and
check that the file is the modified version. After confirmation, click the [X] button
on the “FFFTP” window and exit the [FFFTP].
10-47
Chapter 10: Software setup and update
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
Figure 10.5-6
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
5. Click the “download” icon from the toolbar and save the file into c: drive in the
operational PC.
Figure 10.5-7
6. Use text editor software (such as Notepad) to open the file. Modify and save the
file.
7. Select the modified file at the “FFFTP” window (left side of screen).
10-48
Chapter 10: Software setup and update
8. Click the “upload” icon from the toolbar and save the modified file into the
[/home/ca270/sysboot/kokyaku] folder in the operational software.
Figure 10.5-8
9. Double-click the uploaded file in the “FFFTP” window (left side of screen) and
check that the file is the modified version. After confirmation, click the [X] button
on the “FFFTP” window and exit the [FFFTP].
10-49
Chapter 11: Specification
Chapter 11 Specification
4 Type of calibration Factor, Linear, Point to point, Log-logit, Exponential, Spline, and Spline 2
8 Number of simultaneous 50 methods (Max.) + 3 electrolyte (Na, K and Cl) methods (option)
measurement
Detector Unit
Stirrer Unit
Trough Unit
Washing Unit
Subtank
Control Unit
Wastewater Tank
Degassing Unit
11-1
Chapter 11: Specification
9-2 Cuvette
Material: Resin
Quantity: 72
Number of sample tube: Outer tray: 40 Sample tubes or Sample cups (Max.)
Number of pipettes: 1
change of capacitance
Number of pipettes; 1
change of capacitance
11-2
Chapter 11: Specification
simultaneous measurement)
speeds)
Washing Unit waste pump: 5 (High concentrated wastewater: WU1, WU2, Low
Pipette Trough: 2
Stirrer Trough: 2
measurement
11-3
Chapter 11: Specification
th
10 step: Presoak in Purified water + Water blank
measurement
th
11 step: Nozzle 5 drainage (Water blank measurement
+ Wipe)
th
12 step: Nozzle 5 drainage (Wipe)
50/60Hz
- Profile Order
11 Environmental Requirement:
12 Dimensions
11-4
Chapter 11: Specification
13 Connectors to analyzer
- Power Connector
rack)
75dB (at the location 1m or longer distant from the analyzer with its cover closed)
15 Transient overvoltage
Overvoltage category II
Pollution degree 2
11-5
Chapter 11: Specification
PCB:25P4102
Glass tube fuse 5x20mm 5A/250V Slo-Blo
(ASP/RCU-DRV) F1
PCB:25P4102
Glass tube fuse 5x20mm 10A/250V Slo-Blo
(ASP/RCU-DRV) F2
PCB:25P4101
Glass tube fuse 5x20mm 10A/250V Slo-Blo
(MAIN-DRV) F1, F2
Fuse Holder
Ceramic tube fuse 3AB 10A/250V Slo-Blo
(AC input) F1, F2
11-6
Chapter 11: Specification
types
11-7
Chapter 12: Special tools and materials
For dismounting/mounting explained in this chapter, the following items are required:
12-1
Chapter 12: Special tools and materials
Required tools
Item Description
Screwdriver Required types:
+ Screwdrivers No.1 and No.2
+ Long Screwdriver No.2
+ Short Screwdriver No.2
- Screwdriver 6 mm
Torque Screwdriver No.2
Hex wrench Required sizes:
1.5 mm, 2 mm, 2.5 mm, 3 mm
Small ratchet -
set
Ratchet Required sizes for packing:
wrench 13mm
Box Required sizes:
screwdriver 5.5 mm, 7 mm, 8 mm
Spanner Required sizes:
5.5mm, 10mm, 13mm, 14mm, 14.2mm, 17mm, 19mm, 22mm
Long-nose -
pliers
Nipper -
Pliers -
Monkey -
wrench
Tube cutter Required for tubing
Retractable -
knife
Air blower Required for cleaning lens and filters.
Gap gauge Required for clearance adjustment in the pulley.
Tweezers -
12-2
Chapter 13: Maintenance parts list
2. Parts layouts
Each column of the list indicates as follows.
Column Description
Sheet # Parts layouts for a unit are spread into several sheets (or pages). Refer to
the sheet indicating the sheet #. “#” indicates the item # in the sheet.
Code Inform us of the code # when ordering.
Item Inform us of the item when ordering.
Type Inform us of the type when ordering.
Q’ty Purchasing unit
e.g.) When described “100” in the “Q’ty”, place an order by 100 pieces.
The details of the parts layout are indicated by the parts classification.
13-1
Chapter 13: Maintenance parts list
B. IRU
C. DTR
D. WU
E. MIX
F. PP
13-2
Chapter 13: Maintenance parts list
G. SWU
H. OPTION(ISE)
13-3
Chapter 13: Maintenance parts list
B. PSU
13-4
Chapter 13: Maintenance parts list
C. CNU
D. IRU
E. DTR
13-5
Chapter 13: Maintenance parts list
F. WU
G. MIX
13-6
Chapter 13: Maintenance parts list
H. PT (SPT/RPT)
I. ASP
13-7
Chapter 13: Maintenance parts list
J. RCU
K. PP (SPP)
13-8
Chapter 13: Maintenance parts list
L. PP (SPPW/RPPW)
M. PP (RPPR)
N. PP (WPP)
13-9
Chapter 13: Maintenance parts list
O. PP (STK)
P. SWU
Q. OPTION (CLOG)
R. OPTION (DGU)
13-10
Chapter 13: Maintenance parts list
S. OPTION (ISE)
T. OPTION (GTK)
13-11
Chapter 13: Maintenance parts list
A. CHS
13-12
Chapter 13: Maintenance parts list
13-13
Chapter 13: Maintenance parts list
B. PSU
13-14
Chapter 13: Maintenance parts list
C. CNU
13-15
Chapter 13: Maintenance parts list
D. IRU
13-16
Chapter 13: Maintenance parts list
13-17
Chapter 13: Maintenance parts list
E. DTR
13-18
Chapter 13: Maintenance parts list
F. WU
13-19
Chapter 13: Maintenance parts list
G. MIX
13-20
Chapter 13: Maintenance parts list
H. PT (SPT, RPT)
13-21
Chapter 13: Maintenance parts list
I. ASP
13-22
Chapter 13: Maintenance parts list
13-23
Chapter 13: Maintenance parts list
J. RCU
13-24
Chapter 13: Maintenance parts list
13-25
Chapter 13: Maintenance parts list
K. PP (SPPS)
13-26
Chapter 13: Maintenance parts list
L. PP (SPPW, RPPW)
13-27
Chapter 13: Maintenance parts list
M. PP (RPPR)
13-28
Chapter 13: Maintenance parts list
N. PP (WPP)
13-29
Chapter 13: Maintenance parts list
O. PP (STK)
13-30
Chapter 13: Maintenance parts list
P. SWU
13-31
Chapter 13: Maintenance parts list
Q. OPTION(Clot)
13-32
Chapter 13: Maintenance parts list
R. OPTION(DGU)
13-33
Chapter 13: Maintenance parts list
S. OPTION(ISE)
13-34
Chapter 13: Maintenance parts list
13-35
Chapter 13: Maintenance parts list
OPTION(GTK)
13-36
Chapter 14: Electrical connections
14-1
Chapter 14: Electrical connections
14-2
Chapter 14: Electrical connections
14-3
Chapter 15: Fluidic system diagram
15-1
Chapter 15: Fluidic tubes
No. a b c d e
1050 1 +5
100 0
1051 1 +5
140 0
1052 1 +5
80 0
1053 2 +5 +5
130 0 60 0
1054 2 +5 +5
130 0 60 0
1055 3 +5 +5 +5
330 0 20 0 20 0
1056 3 +5 +5 +5
250 0 20 0 20 0
1057 4 +5
60 0
1058 5 +5 +5 +5
250 0 20 0 20 0
1059 5 +5 +5 +5
230 0 20 0 20 0
1060 5 +5 +5 +5
250 0 20 0 20 0
1061 5 +5 +5 +5
270 0 20 0 20 0
1071 6 +10 +5 +5
800 0 20 0 20 0
1072 6 +10 +5 +5
800 0 20 0 20 0
1073 6 +10 +5 +5
600 0 20 0 20 0
1074 6 +10 +5 +5
600 0 20 0 20 0
1075 7 +5 +5 +5
330 0 50 0 50 0
1076 8 +5 +5 +5
400 0 50 0 50 0
1077 8 +5 +5 +5
400 0 50 0 50 0
1078 7 +5 +5 +5
300 0 50 0 50 0
1079 8 +5 +5 +5
400 0 50 0 50 0
1080 8 +5 +5 +5
400 0 50 0 50 0
1081 5 +10 +5 +5
500 0 20 0 20 0
15-3
Chapter 15: Fluidic tubes
No. a b c d e
1082 5 +10 +5 +5
500 0 20 0 20 0
1083 9 +5
30 0
1084 16 +5 +5 +5
450 0 50 0 50 0
1085 13 +5 +5
120 0 55 0
1750 9 +5
10 0
1751 10 +10 +5 +5
500 0 20 0 20 0
1752 10 +10 +5 +5
500 0 20 0 20 0
1753 10 +10 +5 +5
500 0 20 0 20 0
1754 10 +10 +5 +5
500 0 20 0 20 0
1755 5 +10 +5 +5
800 0 20 0 20 0
1756 10 +10 +5 +5
600 0 20 0 20 0
1757 10 +10 +5 +5
600 0 20 0 20 0
1758 10 +10 +5 +5
600 0 20 0 20 0
1759 10 +10 +5 +5
600 0 20 0 20 0
1931 11 +5 +5
60 0 25 0
2031 12 +10 +5 +5 +10
800 0 50 0 50 0 310 0 250
2131 12 +10 +5 +5 +10
1000 0 50 0 50 0 310 0 350
2231 2 +5 +5
210 0 100 0
2232 2 +5 +5
150 0 70 0
2233 2 +5 +5
210 0 100 0
2234 2 +5 +5
150 0 70 0
2235 5 +5 +5 +5
360 0 20 0 20 0
2236 5 +5 +5 +5
480 0 20 0 20 0
2237 5 +5 +5 +5
490 0 20 0 20 0
2461 3 +5 +5 +5
260 0 20 0 20 0
2531 13 +5 +5
150 0 70 0
15-4
Chapter 15: Fluidic tubes
No. a b c d e
3030 5 +10 +5 +5
700 0 20 0 20 0
3031 14 +5
100 0
3032 15 +10 +5 +5
600 0 50 0 150 0
3033 15 +10 +5 +5
600 0 50 0 150 0
3034 15 +5 +5 +5
350 0 50 0 150 0
3035 15 +10 +5 +5
1160 0 50 0 150 0
3121 11 +5 +5
150 0 70 0
3122 11 +5 +5
110 0 50 0
3123 11 +5 +5
70 0 30 0
3131 14 +5
60 0
3132 14 +5
30 0
3201 1 +10
1000 0
3202 14 +10
1000 0
3901 16 +5 +5 +5
250 0 50 0 50 0
4021 15 +10 +5 +5
700 0 50 0 150 0
4022 16 +5 +5 +5
400 0 50 0 50 0
4023 14 +5
50 0
4024 14 +5
150 0
4025 2 +5 +5
150 0 70 0
4026 13 +5 +5
150 0 70 0
4031 16 +10 +5 +5
1160 0 50 0 50 0
4131 17 +5 +5
80 0 35 0
15-5
Chapter 15: Fluidic tubes
15-6
Chapter 15: Fluidic tubes
15-7
Chapter 15: Fluidic tubes
15-8
Chapter 15: Fluidic tubes
15-9
Chapter 15: Fluidic tubes
15-10
Chapter 15: Fluidic tubes
15-11
Chapter 16: Sensor list
16-1
Chapter 16: Sensor list
16-2
Chapter 17: Test points
Figure 17.1-1
When the PCB incorporated in the system, it is difficult to check each of the
voltage on the PCB.
17-1
Chapter 17: Test points
Figure 17.1-2
17-2
Chapter 17: Test points
Figure 17.1-3
Figure 17.1-4
17-3
Chapter 17: Test points
Figure 17.1-5
17-4
Chapter 17: Test points
Figure 17.1-6
When the PCB incorporated in the system, it is difficult to check each of the
voltage on the PCB.
17-5
Chapter 18: Barcode (BCR) adjustment
Are label size and the pasting position under Check that the labels are pasted in the proper
B. Barcode Specifications
required.
NW7、CODE 128(A/B/C)
Both of the setting in the operational PC and The setting for check digits (with/without) is
the actual labels are consistent under the available only for the Sample BCR
□
check-digit setting (with or without). specifications.
18-1
Chapter 18: Barcode (BCR) adjustment
Is the lid closed when reading labels? If any ambient light break through on the
Are barcode labels are facing outside? Adjust the label position at center of the
No contamination or condensation on the If the RCU lid is kept open or opened and
glass BCR window of the RCU? closed frequently, condensation may occur.
□
18-2
Chapter 18: Barcode (BCR) adjustment
Figure 18.2-1
Figure 18.2-2
3. Click the [OK] button. Reagent BCR starts reading the barcode label.
18-3
Chapter 18: Barcode (BCR) adjustment
4. When 20 scans successfully completed and all of the scan results are
recognized equal, the pop-up message appears including the scanned data.
When one or more out of 20 scans failed or any of the scan results are not
recognized equal, the following pop-up message appears including all 20
scanned data.
Figure 18.2-3
18-4
Chapter 18: Barcode (BCR) adjustment
Figure 18.2-4
18-5
Chapter 18: Barcode (BCR) adjustment
Figure 18.2-5
2. Click the [Sample Barcode Adjustment] button. The following pop-up message
appears. Place the labeled reagent bottle on the position #40.
Figure 18.2-6
3. Click the [OK] button. Sample BCR starts reading the barcode label.
18-6
Chapter 18: Barcode (BCR) adjustment
4. When 20 scans successfully completed and all of the scan results are
recognized equal, the pop-up message appears including the scanned data.
When one or more out of 20 scans failed or any of the scan results are not
recognized equal, the following pop-up message appears including all 20
scanned data.
Figure 18.2-7
Figure 18.2-8
Figure 18.2-9
18-8
Chapter 18: Barcode (BCR) adjustment
Reagent BCR label including the blank space needs to be positioned within
9mm to 77mm (9mm+68mm) area from the bottom.
Sample BCR label including the blank space needs to be positioned within
12.5mm to 85.5mm (12.5mm+73mm) area from the bottom.
Figure 18.3-1
For the reagent bottle, paste the label within the angle α=1º.
For the sample tube, paste the label within the angle β=5º.
Figure 18.3-2
18-9
Chapter 19: LAN connection check
Ethernet: 10/100/1000Base-T
LAN cable: crossover cable
TCP/IP setting
Operational PC: 172.17.31.201
Security setting
Firewall: Off
Automatic updates: Off
Malware protection: Off
Figure 19.2-1
19-1
Chapter 19: LAN connection check
Figure 19.2-2
2. After pinging the command, the following results are displayed on the screen.
Figure 19.2-3
19-2