Professional Documents
Culture Documents
INTRODUCTION
The Continuous Bubble Cap Tray Distillation Column is developed primarily for
teaching, training and research purposes. It is used for the separation of mixtures
at atmospheric pressure in a continuous operation. The unit has the following
special features, which are of particular value in training situations:
Direct heat input into the reboiler using steam and to be controlled by
reboiler temperature or pressure differential across the column.
Silver coated and vacuum jacketed bubble cap tray column minimizes
heat loss to the surroundings.
Sampling on each trays and on the bottom and top product to determine
the practical separation behaviour of the particular mixture.
The method of operation of a bubble cap tray distillation column for the
separation of liquid mixtures can be described in the following terms: The vapour
mixture rises from the reboiler, passes over the column trays and is fully
condensed in the condenser. The condensate runs down the column counter-
current to the vapour through the trays back into the reboiler. At each trays, the
rising vapour attempts to achieve equilibrium with the liquid. In the ideal case the
rising vapour should be in equilibrium with the liquid. In actual columns, complete
exchange does not always take place.
The efficiency of a bubble cap tray distillation column can be evaluated using the
number of theoretical plates required to achieve the same separation as the
actual column. As the purity of a distillate is dependent to a very large extent on
the number of theoretical plates, determination of the number of theoretical plates
in a column is very important.
-1-
Cooling
Water V17
V15
Cooling FI
TI
V18
17
Water 02
TI TI FI
16 V16 18 03
W2
KS-01
FI
04 TI
13
Vent V5 TI
sampl
TI 14
12
sampl
TI
11
sampl
TI
10
PdIC sampl
TI
01
09 K1 W3
sampl
W5 TIC
01
TI
08
TI
15
sampl
el. TI
07
FI sampl
01 TI
06 B5
sampl
TI
05
sampl
V4 TI
V6
TI 04
21 sampl
TI
03
TIC
02
TIA
20
V10 PIA
01
LS-
01
B2 B3
W4
B1
W1
B6
V3 PI TI FI
02 19 05
V19
V2 V11 Condensate
V20 CV01 Steam
V9 6 bar
V1 M
P1 B4
HS
1
V13
V12 V7
V14 V8
Figure 1: Process flow diagram for the continuous bubble cap tray distillation unit
-2-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
It is important that the user read and fully understand all the instructions and
precautions stated in the manufacturer's manuals supplied with the distillation unit
prior to operating. The following procedures will serve as a quick reference for
operating the distillation unit.
1. Cooling Water System – Laboratory tap water is used as the cooling media.
Hose connections are such that the cooling water enters the top condenser
(W2), distillate cooler (W3) and bottom product cooler (W4) in parallel. Adjust
each cooling water flow rates using valves V16, V17 and V18 to maximise
cooling and condensation. The temperature of the cooling water leaving the
condenser should not exceed 45°C.
2. Heating Control – The heat input or steam flow rate into the reboiler can be
controlled using PID controls by either monitoring the reboiler temperature
(TIC-02) or pressure differential across the column (PdIC-01). Select the
heating control mode to either TIC-02 or PdIC-01 on the control panel, and
adjust the setpoint on the controller. The setpoints can also be adjusted on
the DAS if the control switch is at the Remote position.
3. Pump Settings – The feed flow rate can be adjusted by changing the stroke
and frequency settings on the pump. Adjust the stroke length between 0 and
100 mm by turning the knob on the pump, and set the frequency between 0 to
130 strokes/minute.
4. Feed Preheater (W5) – The feed temperature entering the distillation column
can be controlled by adjusting the temperature setpoint of TIC-01 on the
control panel.
5. Reflux Divider – The reflux divider controller KS-01 has 3 positions with the
following settings:
To adjust the reflux ratio, set the reflux timer and distillate timer on the control
panel at KS-01. The reflux timer is represented by the time in seconds on the
display, while the distillate timer is represented by the time in seconds with an
additional blinking light at the lower right corner of the display. The reflux ratio
is equal to the ratio between the total reflux time and the total offtake time.
-3-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
EXPERIMENTAL PROCEDURES
1. Ensure all valves are closed except V6, V11 and the ventilation valves.
2. Turn on the power for the control panel.
3. Prepare a 50 to 60 L mixture containing ethanol and filtered water at the
desired composition. Charge the unit as follows:
a) Open valve V3 and V12.
b) Open the quick closure cap on reboiler B1.
c) Pour all the ethanol-water mixture into reboiler B1 and observe the
overflow of liquid into the bottom product vessel B4 and pump P1.
d) Switch on feed pump P1 and adjust the stroke length and frequency.
e) Allow the mixture from vessel B4 to be pumped into the feed vessel B2
through valve V3.
f) Stop the pump as soon as the bottom product vessel B4 is almost empty.
g) Close valves V3 and V12.
4. Set the reflux divider KS-01 to total reflux. This is to prevent any distillate from
escaping when starting up the distillation column.
5. Let the cooling water flow into the condenser (W2) and product coolers (W3 &
W4) by opening valves V15, V16, V17 and V18.
6. Set the heating control mode to TIC-02 and adjust the temperature setpoint to
100°C, which is the boiling point of the heavy key component, water.
7. Drain any condensate in the steam line by opening valve V20.
8. Open the steam inlet valve V19 and let the flow of steam through the control
valve CV01 and coil heat exchanger W1. Allow the liquid in the reboiler to boil.
As the liquid boils, the liquid level in the reboiler will start to drop.
9. Observe the vapor rising from the boiling liquid into the distillation column. As
the vapor reaches the condenser, it will condense and flow back into the
column as liquid. Mass transfer will take place between this liquid and vapor
phase in the column.
10. Check that the condenser has sufficient cooling water flow to minimize escape
of vapor into the vent.
11. Let the distillation column run until a stable condition is observed. The stable
condition is reached when all temperature indicators on the column give a
constant reading with an allowable ± 0.2°C fluctuation.
-4-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
3. Stop the steam input to the coil heat exchanger (W1) by adjusting the setpoint
at TIC-02 to 0°C and PdiC-01 to 0 mbar. Monitor the temperature drop in the
reboiler. Wait until it has cooled to room temperature.
4. Close the cooling water valves V15, V16, V17 and V18.
5. If necessary, drain all liquid from the reboiler B1 and product vessels B3 and
B4. The distillate and bottom product can also be recycled back into the feed
vessel B2 using pump P1 and opening valves V7 and V12.
6. Turn off the power for the control panel.
Sampling Procedures
Samples can be taken from the trays, distillate and the bottom product for
analysis:
1. Tray sampling:
a) Carefully insert a syringe needle through the gaskets in the sampling port.
b) Collect enough sample in the syringe and withdraw the needle.
2. Distillate sampling:
a) Drain any remaining liquid in the sampling valve V5.
b) Place a vial below valve V5.
c) Set the reflux divider to total distillate for a few seconds.
d) Open valve V5, collect enough sample and close the valve.
e) Switch the reflux divider back to the original setting.
3. Bottom sampling:
a) Drain the stagnated liquid in the sample line above valve V9.
b) Place a vial below valve V9.
c) Open valve V9, collect enough sample and close the valve. Be careful of
the hot liquid while collecting a sample.
d) Cool the sample by immersing the vial in cold water.
Note: Always re-cap the vials quickly after collecting samples to avoid
excessive evaporation of ethanol before analysis.
-5-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
OBJECTIVE:
To determine the number of theoretical plates at total reflux.
PROCEDURES:
1. Perform the general start-up procedures as described in Section 3.1.
2. Once the distillation column has reached stable condition, record the trays,
top and bottom temperature.
3. Collect samples from the trays, distillate and bottom product (Section 3.4).
4. Measure the refractive index for all samples and determine their
compositions.
-6-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
OBJECTIVE:
To operate a continuous distillation column at constant reflux.
To carry out mass balance calculations on the distillation column.
To examine the effects of reflux ratio and feed conditions on the number of
theoretical plates in a continuous distillation.
PROCEDURES:
1. Perform the general start-up procedures as described in Section 3.1.
2. Set the reflux ratio to 1 (e.g. 10 sec reflux & 10 sec distillate) at the reflux
divider controller (KS-01).
3. Switch on the feed pump P1 and adjust the pump stroke to 80 mm and
frequency to 40 strokes/min in order to obtain a flow rate of approx. 10 L/hr.
4. Once the distillation column has reached stable condition, record the trays,
top and bottom temperature.
5. Collect samples from the trays, distillate and bottom product (Section 3.4).
6. Measure the refractive index for all samples and determine their
compositions.
7. Measure the distillate and bottom product flow rate as follows:
a) Close the valve (V6 or V11) below the measuring vessel (B5 or B6) and
start the timer.
b) Collect the distillate or bottom product in the vessel for a period of time.
c) Record the volume of the liquid collected at that specific period of time to
determine the flow rate.
d) Re-open the valve (V6 or V11).
8. Repeat the experiment (steps 2 to 7) with a different value of reflux ratio.
9. Repeat the experiment (steps 3 to 7) with a different feed temperature by
switching on the feed preheater W5 and set the temperature to 60°C.
-7-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
SAFETY CONSIDERATIONS
1. The unit must be operated under the supervision of an authorized staff who
has been properly trained to handle the unit.
2. All operating instructions supplied with the unit must be carefully read and
understood before attempting to operate the unit.
3. Feed stock which severely affect PTFE, PFA and borosilicate glass are not to
be used.
4. The system should not be subjected to shock, sudden impact, vibration,
additional load, or permanent external action of aggressive vapours.
5. Always check and rectify any leak.
6. Do not touch the hot components of the unit.
7. Be extremely careful when handling hazardous, flammable or polluting
materials.
8. Restore the system to operating conditions after any repair job.
9. Make sure the system is sufficiently ventilated when working at atmospheric
pressure.
10. Do not exceed the maximum cooling pressure of 2 bar(g) for the top condenser
(W2) and product coolers (W3,W4).
11. Only properly trained staff shall be allowed to carry out any servicing.
Manufacturer's manual must always be observed.
12. Before any servicing, shut down the whole operation and let the system to cool
down and be properly ventilated.
13. Do not use any coarse or abrasive cleaners on glass components.
14. Leaking couplings should be carefully retightened. Replace any gaskets or
seals if necessary.
REFERENCES
APPENDIX A
-8-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
0 10
1.0 9.0
2.0 8.0
3.0 7.0
4.0 6.0
5.0 5.0
6.0 4.0
7.0 3.0
8.0 2.0
8.5 1.5
9.0 1.0
9.5 0.5
10.0 0.0
-9-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
APPENDIX B
Temperature
Tray Refractive index Mole fraction
(°C)
Distillate
1
2
3
4
5
6
7
8
9
10
Bottom
-10-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
APPENDIX C.1
B
Bottom N4 N3 N2 N1 Distillate
B
Bottom N4 N3 N2 N1 Distillate
3. The number of actual stages are shown by the number of steps drawn on the diagram, which is
the same as the number of trays found on the distillation column. An extra step is observed due
to the extra equilibrium stage in the reboiler.
yn yn 1
4. Murphree tray efficiency =
y n* yn 1 yn*
yn
yn+1
-11-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
APPENDIX C.2
B
Bottom N4 N3 F N2 N1 Distillate
-12-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
B
Bottom N4 N3 F N2 N1 Distillate
B
Bottom N4 N3 F N2 N1 Distillate
B
Bottom N4 N3 F N2 N1 Distillate
-13-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
6. The number of actual stages are shown by the number of steps drawn on the diagram, which is
the same as the number of trays found on the distillation column. An extra step is observed due
to the extra equilibrium stage in the reboiler.
yn yn 1
7. Murphree tray efficiency =
y n* yn 1 yn*
yn
yn+1
-14-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
APPENDIX D
65.21 (25°C)
Vapour heat capacity (J/mol.K)
73.15 (78°C) 34.02 (100°C)
-15-
CONTINUOUS BUBBLE CAP TRAY DISTILLATION UNIT
-16-
X-Y Equilibrium Diagram for Ethanol-Water System at 1 atm
1.0000
0.9000
0.8000
0.7000
Vapor mol fraction
0.6000
0.5000
0.4000
0.3000
0.2000
0.1000
0.0000
0.0000 0.1000 0.2000 0.3000 0.4000 0.5000 0.6000 0.7000 0.8000 0.9000 1.0000
Liquid mol fraction
-17-
Continuous Bubble Cap Distillation
-18-
Continuous Bubble Cap Distillation
-19-