You are on page 1of 6

Journal of Orthopaedic Surgery 2001, 9(2): 71–76

Qualitative analysis of polyethylene wear in


a bipolar femoral prosthesis: A case report

Nobuhiro Kaku, Hiroshi Tsumura and Takehiko Torisu


Department of Orthopaedeic Surgery, Oita Medical University, Oita, Japan

INTRODUCTION

ABSTRACT Wear and the generation of ultra high molecular weight


polyethylene (UHMWPE) wear particles is now
We researched the qualitative changes in the considered the most important cause of long term
polyethylene of a bipolar head retrieved at revision failure in total hip arthroplasty. 1,2,10–13,16 Micron-size
occurring in vivo using optical microscopy and Fourier polyethylene wear particles enter the tissues
transform infrared spectroscopy. The bearing surface surrounding the joint, causing adverse cellular
of the outer head was smooth with no definable reactions which lead to osteolysis and loosening 18,24.
scratches delamination. However, the evidence of To limit the generation of wear debris, implants should
delamination of the beveled throat region and cracking have a low coefficient of friction and a resistance to
at the base of the leaf was noticed. The degree of third-body wear damage of the bearing surface.6,14–17
oxidation was different in each area of the polyethylene Also, the cellular reactions to the wear debris should
in the bearing component and that of the rim was be minimised.
shown to be higher than that of the bearing surface. Since Bateman, in a 1974 study, described the use
Comparing the rate of ketone and ester, ketone binding of the bipolar femoral head prosthesis in osteoarthritis
was frequently found around the prosthetic rim rather of the hip,3 some clinical results have been reported as
that at the bearing surface. However, the ratio of ester favorable.4,21 Messieh et al., however, described wear
at the surface was higher than that at the deep area of debris originating from the contact of the bipolar head
the same rim. We considered these results showed local with the femoral neck in 1994,22 and subsequently other
environmental effects in vivo play a significant role in reports have described acetabular osteolysis following
dictating the response. bipolar hip arthroplasty.7,19 Polyethylene wear and
subsequent osteolysis have also been recognized as
Key words: bipolar hip arthroplasty, polyethylene, wear debris, severe problems in bipolar hip arthroplasty.
oxidation, Fourier transform infrared spectroscopy The purpose of this paper is to investigate
qualitative changes in polyethylene in vivo using an

Address correspondence and reprint requests to: Dr Nobuhiro Kaku, Department of Orthopaedeic Surgery, Oita Medical
University, Hasama-machi, Oita 879–5593, Japan. E-mail: nobuhiro@oita-med.ac.jp
72 Nobuhiro Kaku et al. Journal of Orthopaedic Surgery

outer head that was retrieved at revision. We reported the IR spectrum between 1749–1691cm–1 and PE is the
a case of a female patient in whom the polyethylene C-C rocking band 739–708 cm–1 (Graph 1).
component failed following bipolar hip arthroplasty
with acetabular reaming. This patient had marked
acetabular osteolysis; however, neither osteolysis nor RESULTS
any radiolucent lines were apparent around the
femoral stem. The roentgenograms showed that the outer head had
moved medially, appearing similar to the protrusio
acetabli of rheumatoid arthritis. The extent of the clear
MATERIALS AND METHODS zone around the outer head widened gradually from
12 years after the primary operation and finally
This case involved a female patient who was 49 years reached a maximum of 5 mm. However, as there was
old at the time of the primary operation. She had no osteolysis or any clear zone around the stem, we
suffered from dysplasic hip osteoarthritis which was only replaced the bipolar outer head with allo-bone
treated with primary arthroplasty with the reaming grafting of the acetabular defect (Fig.1).
of the acetabulum and implantation of a bipolar hip
arthroplasty with an Aufranc-Turner cemented stem
with a 22.225 mm inner femoral head. When this
patient returned to the clinic with hip pain 15 years
after the primary operation, osteolysis around the outer
head was recognized. When the patient was 64 years
old, revision was performed and the outer head was
retrieved. Therefore, we have been able to follow this
patient for fifteen years following the primary
operation.
In order to research the qualitative changes in the
polyethylene occuring in vivo, we investigated the
shape, location and volume of the polyethylene wear
in the retrieved outer head. The wear analysis was
performed on this component to determine wear rate
penetration in millimeters per year. This calculation
was made utilizing mathematical calculations of
polyethylene thickness. The size of it was identified,
manufacturing blueprints were matched up to
determine the original polyethylene thickness at the
pole region, and the outer head was measured in the
same region to define the thickness difference. This
difference between the original and actual thickness
was used in conjunction with the implantation age of
the component to calculate the wear rate. Also, optical
microscopy and Fourier transform infrared (FT-IR)
spectroscopy was used to assess the oxidation of the
polyethylene. Thin sections approximately 0.025 mm
thick were cut with a microtome, parallel to the radial
direction, through the leaflets and bearing surface. The
sections were placed on a microscope slide with a cover
slip glued over the section in order to keep them flat.
The sections were then examined with transmitted
plane and polarized light microscopy. The same 0.025
mm sections used for the optical microscopy were used Figure 1 Case. The patient was female and her age at revision
to measure the IR spectra by transmission, placing the was 64 years, followed up for 15 years. The clear zone around
the outer head on the roentgenograms widened gradually
sections on a KBr plate on a Nicolet-Plan FT-IR twelve years later postoperatively and the outer head moved
microscope. The degree of oxidation was measured as to the supero-medial direction, so she felt hip pain at the same
the ratio C=O:PE, where C=O represents the area under time.
Vol. 9 No. 2, December 2001 Qualitative analysis of polyethylene wear in a bipolar femoral prosthesis 73

The outer head diameter was 44 mm. The bearing Features such as a black band could not be seen in
surface of the outer head was smooth, polished and the main socket load-bearing region with an optical
glistening with no definable scratches or delamination microscopy using reflected light (Fig. 4). With
(Fig. 2). The average diameter of the bearing socket transmitted light, this black band was visible
was 22.771 mm (normal diameter 22.377 mm). Average approximately 0.70 mm below the surface in many
diameter of the ‘throat’ was 20.015mm (normal regions of the polyethylene component, but not
value19.380 mm). The wear on the bearing surface was uniformly over the section. This is equivalent to the
0.197 mm. This represented an annual linear wear rate white band seen under reflected light microscopy (Fig.
of 0.0131 mm. 5). These optical effects were caused by micro fractures
produced in the brittle, highly crystalline regions. Some
leaflets exhibited variable degrees of recrystallization
as determined by a black subsurface band, with
relatively little on the inner surface but severe in the
area of the delamination on the beveled surface. In the
beveled region under polarized light, the development
of the recrystallization within the resin grains was
clearly apparent.

Figure 2 Visual findings of bearing surface was smooth and


glistening with no definable scratches.

Visible discoloration of the beveled polyethylene


rim was apparent in places. Evidence of delamination
of the beveled throat region and cracking at the base Figure 4 There was no visible evidence such as a black band
of the leaf was noticed. Very little abrasive wear in the main socket bearing region in optical microscopy and
occurred on the beveled surface of the leaflet (Fig. 3). reflected light.

Figure 3 Evidence of delamination in the beveled region and Figure 5 With transmitted light, a black band was visible
a crack were visible at the base of the leaf. No evidence for approximately 0.70 mm below the surface in many regions,
excessive wear of the beveled region prior to the delamination but not uniformly over the section. This was equivalent to the
damage. white band seen under reflected light.
74 Nobuhiro Kaku et al. Journal of Orthopaedic Surgery

Table 1 presents the FT-IR spectroscopy data. The DISCUSSION


degree of oxidation was different in each area of the
polyethylene-bearing component. The degree of The mean annual wear rate with the 22 mm inner head
oxidation of the rim was shown to be higher than that in total hip arthroplasty (THA) was reported by
of the bearing surface (Fig. 6) (Table1). Comparing the Charnley and Halley in 1975 to be 0.09 mm, 8 while
ratio of ketone and ester, the ketone binding was Wrobleski found this to be 0.096 mm in 1993.26 In 1998,
frequently found around the prosthetic rim rather than Ingemar et al. studied wear and migration rates in a
at the bearing surface. However, the ratio of ester at series of 102 hip replacements randomised to have
the surface was higher than that at the deep area of either a cemented polyethylene Charnley socket or a
the same rim. The distribution of the radiation-induced porous uncemented Harris-Galante type–1 cup using
oxidation is of interest. In this case almost no oxidation radiostereometric analysis.15 The mean annual wear
was detectable in the lower region of the insert. All of rate was found to be 0.09 mm in the Charnley sockets
the regions, which appeared clear, exhibit an oxidation and 0.1 mm in the Harris-Galante socket. Chun-Hsiung
ratio of between 0.18 and 0.28. The main crystallization measured linear polyethylene wear in uncemented
changes occurred at the resin boundaries, as illustrated Osteonics components using digital and special
by the darkened boundaries. The area below the clear software of their own design in 1998.9 The mean annual
surface layer, i.e. rom 0.70 to 1.40 mm, had a maximum wear was 0.15 mm, which increased to 0.23 mm in hips
oxidation ratio of 0.33–0.41 and is where the showing evidence of osteolysis.
delamination may occur. For bipolar prosthesis, Bose et al. reported a case
that had osteolysis of the acetabulum and in which
the mean annual wear rate of the retrieved outer head
was 0.17 mm.5 Using radiographs showing the inner
articulation of the bipolar hip prosthesis, Kusada and
Kuroki also reported an 0.17 mm wear rate, and that
there was no difference between the annual wear rate
of hips with and without osteolysis.20 In this case, the
mean annual wear rate of the bearing socket was 0.0131
mm. We think the wear results here are quite low.
However, this result cannot be compared directly with
other reports which used roentgenograms to assess
liner wear rates and the mechanical-stress was not
usually of uneven distributions in bipolar hip
arthroplasy unlike THA. We found smooth surfaces
within the socket region of this outer head which
differed from the roughness often found in the
Figure 6 The number shows the area of the polyethylene acetabular cups of THA.25 We think that one reason is
insert. the reduction of friction with the dual bearing design.
Another reason is the relative absence of third body

Table 1
The C=O rate of place numbers 4,5,6 was higher than that of other place numbers
and ketone band rate of numbers 5 and 6 was higher than other places

Area C=O PE C=O/PE Keton/Ester

1 0.288 11.148 0.03 ND


2 1.858 10.124 0.18 0.5
3 3.159 11.273 0.28 0.7
4 2.437 13.337 0.18 4.3
5 5.228 12.808 0.41 3.1
6 4.187 12.658 0.33 1.3
Vol. 9 No. 2, December 2001 Qualitative analysis of polyethylene wear in a bipolar femoral prosthesis 75

wear of the bearing surfaces due to the locking system in order to generate OH-free radicals from the hydro-
between the inner head and the bearing insert, as well molecular fluid. Additionally, the adjoining fluid
as the narrow entrance in a bipolar hip prosthesis. moves rapidly due to the squeezing of fluid as neck-
Additionally, the insert wear in the socket region cup impingement continues, so the vibration which
occurred in a surface region where there was minimum occurs encourages crysyallization in polyethylene.
oxidation and no excessive recrysyallization, where These factors — oxidation, crystallization and
0.70 mm of clear region remained. So the amount of mechanical stress — may lead to polyethylene debris
wear was minor and the bearing surface looked production.
polished, so no direct relationship was found between In view of the distribution of ketone and ester,
the wear in the inner articulation and occurrence of ketone binding was frequently found in the rim
osteolysis in this study. compared with the bearing surface. However , the ester
In bipolar hip prosthesis, not only the bearing rate was high at the surface of the rim that had contact
surface, but also the polyethylene rim may produce with the joint fluid. This means that the polyethylene
wear debris. Some investigators have reported that a rim is also in an unstable situation, which easily
large amount of debris was due to severe rim abrasion oxidizes at the molecular level. The ester bindings
by neck-cup contact and that this seemed to cause usually exist at the end of a polyethylene chain after
osteolysis.23 In addition, we found discoloration and oxidation, so there should be many short chains at the
extensive delamination occurred in the beveled region surface of the rim; the cause is recrystallization due to
on some leaflets. The excessive crystallization that vibration of the surface. This situation weakens the
occurred approximately 0.70 mm below the insert polyethylene and can cause occurrence of polyethylene
surface, we think, was responsible for the delamination debris.
on some of the leaflets with the infusion of fluid and
resultant discoloration. This was the result of
recrystallization in the resin, which was able to take CONCLUSION
place by the increased chain mobility resulting from
the radiation-induced oxidation. Considering that the radiation during sterilization was
The degree of oxidation was different within each uniform over such a small object as a bipolar implant,
area in a polyethylene insert, and that of the rim, which the variation in the degree of radiation induced
had been in contact with joint fluid, was shown to be recrystallization that varies with location indicates
higher than that of bearing surfaces. It was our local environmental effects play a significant role in
understanding that the degree of oxidation at the point dictating the response. In bipolar prosthesis of the hip,
of femoral neck contact was higher than the other areas the majority of polyethylene wear debris is generated
around the rim. from the rim and not the bearing surface. We believe
It is possible that supersonic waves are generated the cause of polyethylene wear debris generated from
in the joint fluid with the regular fluctuations of the rim is closely related to oxidation and
pressure that occur with load bearing and movement. recrystallization of the material, which may be caused
With this in mind, cavitation, where many small by cavitation, vibration and the creation of many short
bubbles break out in the fluid during the decreasing polyethylene chain segments due to joint fluid effects
the pressure phase would be generated. 27 These as well as neck-cup impingement.
bubbles expand with decreasing pressure and when
their diameter reaches more than about 0.1 mm, they
are crushed on the next increasing pressure phase. This ACKNOWLEDGEMENTS
phenomenon is called ‘cavitation collapse’ and the heat
(about a few thousands degree Centigrade) increases The authors wish to thank Dr Donald F Gibbons,
locally at that time. The effect of heat makes a situation Corporate Scientist, 3M Biosciences Laboratory, for his
where oxidation occurs more easily at the same place technical support and advice.

REFERENCES

1. Bal BS, Vandelune D, Gurba DM, Jasty M, Harris WH. Polyethylene wear in cases using femoral stems of similar geometry,
but different metals, porous layer, and modularity. J Arthroplasty 1998, 13:492–9.
2. Barrack RL, Folgueras A, Munn B, Tvetden D, Sharkey P. Pelvic lysis and polyethylene wear at 5–8 years in an uncemented
total hip. Clin Orthop 1997, 335:211–7.
76 Nobuhiro Kaku et al. Journal of Orthopaedic Surgery

3. Bateman JE. Single-assembly total hip prosthesis: preliminary report. Orthop Digest 1974, 2:15.
4. Bateman JE, Berenji AR, Bayne O, Greyson ND. Long-term results of bipolar arthroplasty in osteoarthritis of the hip.Clin
Orthop 1990, 251:54.
5. Bose WJ, Miller GJ, Petty W. Osteolysis of the acetabulum associated with a bipolar Hemiarthroplasty:A late complication.
J Bone Joint Surg 1995, 77A:1733–5.
6. Bragdon CR, Jasty M, Kawate K, McGrory BJ, Elder JR, Lowenstein J, Harris WH. Wear of retrieved cemented polyethylene
acetabula with alumina femoral heads. J Arthroplasty 1997,12:119–25.
7. Calton TF, Fehring TK, Griffin WL, McCoy TH. Failure of the polyethylene after bipolar hemiarthroplasty of the hip: A report
of five cases. J Bone Joint Surg 1998, 80A:420–3.
8. Charnley J, Halley DK. Rate of wear in total hip replacement. Clin Orthop 1975,112:170–9.
9. Chun-Hsiung S,Po-Cheng L,Jian-Horng C , Ching-Lung T, Li-Fan C,J.Shih-Shyn W, Walter HC. Measurement of polyethylene
wear in cementless total hip arthroplasty. J Bone Joint Surg 1997, 79-B:361–5.
10. Devane PA, Robinson EJ, Bourne RB, Rorabeck CH, Nayak NN, Horne JH. Measurement of polyethylene wear in acetabular
components inserted with and without cement. J Bone Joint Surg 1997, 79A:682–9.
11. Fisher J. Wear of polyethylene in artificial hip joints: superolateral wear of the acetabulum. J Bone Joint Surg 1998, 80B:190–
1.
12. Goyal S, Moss MC, Breusch SJ. Venous outflow obstruction 8 years following Ring total hip arthroplasty: A rare complication
of acetabular loosening and polyethylene wear. Acta Orthop Scand 1997, 68:67–9.
13. Hall RM. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-
eight components retrieved at autopsy or revision operations. J Bone Joint Surg 1998, 80A:764–5.
14. Hozack WJ, Mesa JJ, Carey C, Rothman RH. Relationship between polyethylene wear, pelvic osteolysis, and clinical
symptomatology in patients with cementless acetabular components: A framework for decision making. J Arthroplasty
1996,11:769–72.
15. Ingemer O, Ake SC, Jack B. Wear in uncemented porous and cemented polyethylene sockets. J Bone Joint Surg1994, 76-
B:263–6.
16. Jasty M, Goetz DD, Bragdon CR, Lee KR, Hanson AE, Elder JR, Harris WH. Wear of polyethylene acetabular components in
total hip arthroplasty:An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations.
J Bone Joint Surg 1997, 79A:349–58.
17. Joshi AB, Markovic L, Ilchmann T. Polyethylene wear and calcar osteolysis. Am J Orthop 1999, 28:45–8.
18. Kadoya Y, Kobayashi A,Ohashi H. Wear and osteolysis in total joint replacement.Acta Orthop 1998, 278:1–16.
19. Kobayashi S, Takaoka K, Tsukada A, Ueno M. Polyethylene wear from femoral bipolar neck-cup impingement as a cause of
femoral prosthetic loosening. Arch Orthop Trauma Surg 1998, 117:390–1.
20. Kusaba A, Kuroki Y. Wear of bipolar hip prostheses. J Arthroplasty 1998,13:668–73.
21. McConville OR, Bowman AJ, Kilfoyle RM. Bipolar hemiarthroplasty in degenerative arthritis of the hip: 100 consecutive
cases. Clin Orthop 1990, 251:67.
22. Messieh M, Mattingly DA, Turner RH, Scott R, Fox J, Slater J. Wear debris from bipolar femoral neck-cup impingement: A
cause of femoral stem loosening. J Arthroplasty 1994, 9:89–93.
23. Nakata K, Ohzono K, Masuhara K. Acetabular osteolysis and migration in bipolar arthroplasty of the hip: 5 to 13-year
follow-up study. J Bone Joint Surg 1997, 79B:258–64.
24. Sabokbar A, Fujikawa Y, Neale S. Human arthroplasty derived macropharges differentiate into osteoclastic bone resorbing
cells. Ann Rheum Dis 1997, 56:414–20.
25. Wang A, Polineni VK, Stark C, Dumbleton JH. Effect of femoral head surface roughness on the wear of ultrahigh molecular
weight polyethylene acetabular cups. J Arthroplasty 1998,13:615–20.
26. Wroblewski BM, Fleming PA, Siney PD. Charnley. Low-friction arthroplasty of the hip.Long-term results. Clin.Orthop 1993,
292:191–201.
27. Wroblewski BM. Wear of the high-density polyethylene socket in total hip arthroplasty and its role in endosteal cavitation.
Proc Instn Mech Engrs 1997, 211:109–18.

You might also like