Test Procedure
The TEN test is performed ipsilaterally, meaning that the tone and the noise are presented in the same ear.
It can only be conducted with TDH39, DD45 and Insert earphones.
1. On the device press and hold the Tests button and use the black scroll wheel to select TEN:
Threshold-equalizing noise. The test is set to start with the stimulus for channel 1 and 2 directed to
the same ear. The tone is presented in channel 1 and TEN is presented in channel 2, with the
stimulus in channel 2 reversed (Rev) to have a continuous masking signal.
D-0113812 - A – 2016/06
Threshold Equalizing Noise (TEN) Page 2 of 3
If operating the audiometer from the PC the TEN noise will be available in channel 2 if licensed. To
enable the TEN noise press TEN (a). Ensure it is presented to the same ear as in channel 1 (b).
The decibel step size can be changed to 2dB (c).
A masked threshold is at least 10dB or more above the level of the TEN
AND
Masked threshold is at least 10dB above the non-masked threshold.
D-0113812 - A – 2016/06
Threshold Equalizing Noise (TEN) Page 3 of 3
The example below shows an example of a positive TEN test, with a masked threshold indicating a dead
region.
If dead regions are present, this may have important implications for fitting hearing aids and for predicting the
likely benefit of hearing aids. When a patient has a dead region, there may be little or no benefit from hearing
aid amplification for frequencies well inside the dead region (Moore, 2009).
References
Moore, B. C. J. (2001). "Dead regions in the cochlea: Diagnosis, perceptual consequences, and implications
for the fitting of hearing aids," Trends Amplif. 5, 1–34.
Moore, B. C. J. (2009). “Audiometer Implementation of the TEN (HL) Test for Diagnosing Cochlear Dead
Regions”.