Professional Documents
Culture Documents
1 Research Background
1.1.1 Traditional Ultrasound Imaging
Ultrasound (ultrasound) is a high-frequency mechanical wave. The vibration frequency of
the sound source is greater than 20KHz, which exceeds the upper limit of human hearing [1].
The frequency range of ultrasound used for medical imaging is generally 1-10MHz. Ultrasound
has the characteristics of high frequency, short wavelength, strong directivity, and small harm.
As four major medical imaging technologies, ultrasound has the advantages of no harm, no
radiation, and good real-time performance compared with the other three common imaging
modes, CT, magnetic resonance (MRI), and nuclear medicine imaging (PET, SPECT). It is
widely used clinically [2].
Traditional medical ultrasound imaging is based on pulse-echo technology. The idea was
first proposed by Richardson in 1912 [3] and has been verified in water depth detection. Later
research found that ultrasound tissue can be used to scan human tissues, and different organs and
tissues have specific acoustic characteristics. Due to the different acoustic characteristics, sound
waves are reflected and scattered at different levels on the interface. By receiving and processing
the echo signal reflected or scattered to the transducer, an image or waveform can be obtained
[4]. The intensity of the reflected and scattered signals is determined by the intensity of the
incident signal and the acoustic impedance between different tissues. In 1942, Dussik and
Firestone first applied ultrasound to medical imaging to diagnose brain diseases. This was the
earliest type A ultrasound imaging technology. In 1952, Howry and Wild used B-mode
ultrasound for biological tissue imaging, which was a milestone in the clinical application of
ultrasound [5-6]. Since then, with the improvement of the manufacturing process of piezoelectric
materials and ultrasound probes, the detection and application of ultrasound in medical imaging
have been further developed.
A typical B-mode (brightness mode) imaging process is shown in Figure 1.1. The
ultrasonic transducer uses the inverse piezoelectric effect to convert electrical signals into
acoustic signals and emit ultrasonic waves. In the propagation path, the acoustic impedance is
different from that of the surrounding tissue. At the target, backscatter occurs at the interface,
and the transducer receives these scattered signals and converts them into electrical signals. The
received signal also needs to undergo processing such as amplification, filtering, and focusing
before image reconstruction. The signal received by the transducer is very weak, usually at the
millivolt level, so the signal needs to be amplified first. Filtering is to keep signals near the
center frequency and remove interference signals such as noise. Focusing is to calculate the time
difference caused by the difference between the distances of different array elements from the
focus, and to delay the signal in time to align the received signals at the focus. The aligned
channel data is weighted or apodized, added and summed to obtain the amplitude line. amplitude
line, A-line). The amplitude line is processed by envelope detection to obtain a scan-line in the
image. A series of scan lines can provide two-dimensional cross-section information of the target
and be displayed on the computer screen Display as image. The propagation time of the sound
wave is multiplied by the sound velocity of the medium, and the distance of the sound wave
propagation can be obtained, that is, the position of the reflected wave on the image. In common
ultrasound equipment, the sound velocity of soft tissue is generally set to a constant 1540m / s.
The corresponding gray value of the reflected wave on the image is determined by the amplitude
of the echo.
Ultrasound tomography can scan the imaging object, and finally reconstruct a series of
coaxial two-dimensional images to obtain three-dimensional images. Most of the probes used in
ultrasound tomography use a ring or a ring-like shape. This type of probe can receive both
reflected and transmitted signals. The reflection signal refers to the signal received by the array
element adjacent to the transmission array element, and the transmission signal refers to the
signal received by the array element opposite to the transmission array element, as shown in
Figure 1.2. There are two imaging methods for ultrasound tomography, reflection imaging and
transmission imaging, where transmission imaging includes sound velocity imaging and
attenuation imaging.
The reflection imaging method of ultrasound tomography is similar to the traditional B-
ultrasound. The reflection signal generated when the ultrasound propagates to the tissue is used
to characterize the boundary. After the reflection signal is filtered, focused, and time-delayed, the
beam synthesis method is processed and finally obtained. Reflection image. The transmission
imaging method has a similar data acquisition process as the CT imaging that generates X-rays,
but the imaging principles are not exactly the same. CT scans the imaging object by emitting
multiple sets of parallel or fan-shaped X-rays. The X-ray path can be approximated as a straight
line, and algorithms such as filtered back projection are used to reconstruct the image. The path
of the ultrasonic wave generated in ultrasonic tomography is not a straight line, but reflection,
refraction, and scattering. Therefore, transmission imaging usually uses ultrasound transmission
signals for sound velocity or attenuation imaging [9]. The imaging methods of sound velocity
imaging and attenuation imaging are similar. There are two types of methods: waveform-based
and ray-tracing. The former is slower in calculation, but the imaging quality is better. The latter
is much faster than the former, but the image quality is worse than the former. [10-12].
In China, X-ray photography for breast cancer screening is not as popular as in Western
countries. The main reason is that East Asian women generally have relatively small high-
density breast soft tissues, and X-ray photography needs to squeeze the breast to form a
projection image. Sensitivity to high-density breast soft tissue is only 50-68%, while for high-
density breast soft tissue lesions, X-ray sensitivity will decrease to 45% [20], and it will also
appear in 50% in distinguishing benign and malignant tumors. False negative conclusion.
Routine ultrasound examination is convenient, easy, painless, radiation-free, and low-cost, and
can quickly obtain the results. Studies have shown that combined ultrasound and X-ray
examination can increase the accuracy of breast cancer diagnosis [21-22]. For breast examination
in adolescents, pregnant and lactating women, ultrasound is the imaging method of choice.
However, the quality of conventional ultrasound images is poor, and the false detection rate is
high [23]. The examination process depends on the experience of the doctor, and the
repeatability is poor. At present, ultrasound is mainly used as an auxiliary method in the
diagnosis of breast cancer.
Other imaging methods that can be used for breast cancer examinations include CT and magnetic
resonance imaging, each with its own advantages and disadvantages. CT also produces X-rays,
with a potential risk of excessive radioactivity, and is not useful for distinguishing soft tissues
Sensitive, less used for breast examinations. Magnetic resonance imaging is radiation-free, but
because its examination takes a long time, the testing costs and equipment are expensive, it is
mainly used for further examination of women at high risk for breast cancer, and is not suitable
for routine screening. In addition, breast cancer diagnosis and treatment guidelines and
specifications issued by the Chinese Anti-Cancer Association clearly indicate that near-infrared
scanning, nuclide scanning, and catheter lavage are not supported as screening methods for
breast cancer [24].
In recent years, ultrasound tomography for the early detection of breast cancer has
gradually become a research hotspot [25]. In addition to the advantages of traditional B-mode
ultrasound, for dense breast tissue, ultrasound tomography can find small tumors that cannot be
detected by X-ray photography, and also has a good ability to distinguish malignant tumors from
benign tumors [13]. Correctly identifying the type of tumor can significantly reduce the large
number of biopsy biopsies in current medical clinics, reduce the pain of patients, and reduce the
cost of diagnosis. At the same time, ultrasound tomography can also perform a section scan of
the entire breast to reconstruct a three-dimensional breast image.
At present, the most influential ultrasound CT system in the world is the SoftVue system
developed by the team of Professor Neb Duric of Karmanos Cancer Institute (KCI) of Wayne
State University [43-49]. As shown in Figure 1.4, this system uses a circular array probe to
transmit and receive signals. When the ultrasonic signal is transmitted, it can pass through the
breast to the other side of the circular array, so it can receive reflected and transmitted ultrasonic
signals at the same time. The circular array probe contains 2048 array elements, which can be
moved up and down to obtain data on various sections of the breast, and finally a three-
dimensional image is reconstructed [50]. Generally, there are 30 to 40 tomographic slices per
patient, the resolution of the reflection image is 0.7mm, and it takes 15 seconds to scan each slice
[51]. Nicole Ruiter, Data Processing and Electronics Center, Karlsruhe institute of technology
(KIT), Germany The USCT (UltraSound Computed Tomography) system developed by
Professor [52-55] uses a semi-ellipsoidal probe. As shown in Figure 1.5, 628 transmitting
elements and 1413 receiving elements are distributed inside the semi-ellipsoid. The spherical
probe can also be rotated and translated in the Z direction to collect data at different positions,
and the collected data can be directly reconstructed into a three-dimensional image. In addition,
Professor Andre's team at the University of California, San Diego developed the CVUS
(Computer volumetric ultrasound) system [56-58]. The probe of the system includes a
transmitting module, a receiving module, and three reflecting modules, with a total of 1536 array
elements, as shown in the figure. As shown in 1.6. The transmitting module is used to transmit
ultrasonic signals. The receiving module is opposite to the transmitting module and used to
receive transmitted signals. The three reflection modules on the side are used to receive reflected
signals. After the transmitting module transmits a signal once, other modules perform Data
acquisition, and then the probe needs to be rotated 2 degrees to retransmit the signal and collect
data. A total of 360 degrees of rotation are required to obtain all reflection data and transmission
data for the entire section, and use these data for reflection imaging and transmission imaging
[59-61].
In China, USCT related research mainly focuses on imaging algorithms [62-65] and
industrial detection [66-69], and the development of a complete ultrasound CT imaging system
for clinical use is currently only developed by the team of Professor Ding Mingyue of Huazhong
University of Science and Technology. USCT imaging system [70]. As shown in Figure 1.7, the
probe has a total of 1024 array elements. Each array element transmits signals in sequence, all
array elements receive signals at the same time, and finally all the collected data is processed for
imaging. The system is currently in the verification phase.
Multi-beam synthesis refers to the technology of forming multiple received sound beams
from a received signal after transmitting an ultrasonic pulse. Multi-beam synthesis technology
uses multi-line reception of the transmit beam to generate an image with an extended focus
range. This method can effectively reduce speckle noise in the image and improve the signal-to-
noise ratio. This paper uses this technique for USCT imaging, and makes a theoretical analysis of
how to choose the beam deflection angle on a circular array. Selecting a suitable beam deflection
angle can effectively reduce the main lobe width, suppress the grating lobe, and obtain better
imaging results. The structure of this paper is as follows:
Chapter One Introduction. First introduce the traditional ultrasound imaging process and
ultrasound tomography process; second, introduce the background of breast cancer, the
advantages of ultrasound tomography for the early detection and diagnosis of breast cancer; then
introduce the current research level and status at home and abroad; finally It explains the
research significance and content arrangement of this thesis.
Chapter 2 ultrasound tomosynthesis imaging. This chapter first introduces the time-delay
superposition method and some methods commonly used in traditional ultrasound imaging to
improve the contrast and resolution of the image, including receiving data using a dynamic
aperture.
The time gain compensation method is used to increase the ultrasonic echo intensity in the far-
field region and the apodization weighting method is used to weight the transmit and receive
channels. Subsequently, the related information of the ultrasonic tomographic ring array is
introduced. Then the reflection data acquisition and processing methods and The reflection
reconstruction algorithm performs reflection reconstruction on the collected data and analyzes
the reconstruction effect. Finally, this chapter summarizes this chapter.
Chapter 3 Coherent factor-based adaptive ultrasound tomography. This chapter first introduces
the definition of coherent factor classes. Second, it applies coherent factor classes to the
reconstruction of USCT reflection images. Through the collected body film and experimental
data from volunteers, it verifies that the coherent factor classes are in resolution, sidelobe level,
and noise. Suppression, contrast, and other imaging effects, and pointed out the problems of the
method and discussed the balance between the sensitivity and robustness of the method by
adjusting the parameter range of the method; Perform space-time smoothing to further improve
the quality of the image; then, this paper also points out that the use of coherent factor methods
still has the problem of high speckle noise level. To solve this problem, a VMD algorithm is used
to perform modal decomposition of the data. The decomposed data is recombined and combined
with coherence factor imaging. It can be seen through experiments that this method has a good
effect on improving contrast and suppressing noise; finally, this chapter summarizes it.
Chapter 4 is the method of minimum variance adaptive ultrasound tomography. This chapter first
introduces the definition of the MV beamforming method. Second, it applies the MV
beamforming method to reflection reconstruction. The experimental data of the body film
verifies the application of the MV adaptive beamforming method in resolution, sidelobe level,
noise suppression, Contrast and other aspects of imaging effects, also pointed out the problems
of the method and discussed the balance between the sensitivity and robustness of the method by
adjusting the parameter range of the method; finally, this chapter summarizes .
Chapter 5: Filtering and multi-beam synthesis for ultrasound tomography. This chapter first
introduces the method of spatial filtering of image data after reflection reconstruction. Filtering
can effectively reduce the large variance of the background caused by speckle fluctuations and
significantly improve the contrast. Secondly, the principle of multi-beam synthesis and its use in
ultrasound are introduced. In tomography, choosing the proper beam deflection angle can
effectively suppress speckle noise and clutter, and at the same time remove artifacts. The
imaging results have been verified through body film experiments, human upper arm, calf, and
breast experiments; the chapter concludes with this chapter.
Chapter 6 summarizes and looks forward. The main research contents of this paper are reviewed,
the research ideas are listed, the main innovation points are pointed out, the advantages and
disadvantages of the proposed method are also discussed, and the outlook for future work is
given.