Summary clinical study report

Study title:

Retrospective clinical trial to evaluate clinical performance of BIOCREDIT


COVID-19 Ag using clinical specimens in human nasopharynx obtained from
symptomatic patients confirmed by RT-PCR.

Product name: BIOCREDIT COVID-19 Ag

Clinical trial centers: 1. Seoul Asan Hospital

2. Eunpyung St. Mary’s Hospital

RapiGEN INC.

RQP-702-F09(‘18.04.24) REV.02 RapiGEN


RapiGEN INC.
16, LS-ro 91 beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do 14119, Republic of Korea

Tel: +82-31-427-4677 Fax: +82-31-427-4678 Website: www.rapigen-inc.com

1. Type of specimen

Type of Number of Number of Comments


Clinical Centers
Specimen specimens patients

UTM
Catholic University of Korea, Positive: 3 RT-PCR
(nasopharyngeal Patient: 3
Eunpyung St. Mary's Hospital Negative: 2 confirmed
collection)
for positive
Positive: 12 Patient: 4
Seoul Asan Hospital UTM specimens
Negative: 0 Patient: 0

2. Trial results
1) Positive specimens
Catholic University of Korea, BIOCREDIT
PCR data
Eunpyung St. Mary’s Hospital COVID-19 Ag
Date of Date of
Patient Specimen Day after
Symptom specimen CT value Results Ag Results
No. type onset
onset collection

1 No data No data UTM* No data E:16.76, R:17.20 Pos. ± Pos.

Positive
2 No data No data UTM No data E:15.36, R:15.58 Pos. + Pos.
Sample
3ea

3 No data No data UTM No data E:18.18, R:18.55 Pos. + Pos.

4 No data No data UTM No data >40 Neg. - Neg.

Negative
Sample
2ea

5 No data No data UTM No data >40 Neg. - Neg.

RQP-702-F09(‘18.04.24) REV.02 RapiGEN


RapiGEN INC.
16, LS-ro 91 beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do 14119, Republic of Korea

Tel: +82-31-427-4677 Fax: +82-31-427-4678 Website: www.rapigen-inc.com

BIOCREDIT
Seoul Asan Hospital PCR data
COVID-19 Ag
Date of Date of
Patient Specimen Day after
symptom Specimen CT value Results Ag Results
No. type onset
onset collection

27 Feb. E:22.61, R:24.88


No data UTM 1 Pos. w+ Pos.
2020 N:25.18

29 Feb. E:32.5, R:33.09


A001 No data UTM 3 Pos. w+ Pos.
2020 N:35.24

02 Mar. E:31.95, R:34.49


No data UTM 5 Pos. - Neg.
2020 N:35.95

08 Mar. E:19.94, R:21.93


No data UTM 1 Pos. w+ Pos.
2020 N:35.95

10 Mar. E:22.57, R:24.53


A002 No data UTM 3 Pos. w+ Pos.
2020 N:26.79

Positive
Sample 14 Mar. E:32.6, R:34.83
No data UTM 7 Pos. - Neg.
12ea 2020 N:36.59

13 Mar. E:21.07, R:22.96


No data UTM 1 Pos. - Neg.
2020 N:24.29

16 Mar. E:9.63, R:12.81


A003 No data UTM 4 Pos. + Pos.
2020 N:13.51

17 Mar. E:18.46, R:19.95


No data UTM 5 Pos. w+ Pos.
2020 N:23.36

28 Mar. E:20.35, R:21.93


No data UTM 1 Pos. w+ Pos.
2020 N:23.99

30 Mar. E:27.95, R:29.81


A004 No data UTM 3 Pos. w+ Pos.
2020 N:30.99

03 Apr. E:29.99, R:31.69


No data UTM 7 Pos w+ Pos.
2020 N:32.26

RQP-702-F09(‘18.04.24) REV.02 RapiGEN


RapiGEN INC.
16, LS-ro 91 beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do 14119, Republic of Korea

Tel: +82-31-427-4677 Fax: +82-31-427-4678 Website: www.rapigen-inc.com

3. Conclusion
1) 15 positive specimens which were confirmed by RT-PCR are tested with BIOCREDIT
COVID-19 Ag kit, and 12 specimens are identified as positive as well.

Relative Sensitivity (%) = 100 x (No. of specimens with positive results /


No. of positive specimens tested by RT-PCR)

2) 2 negative specimens which were confirmed by RT-PCR are tested with BIOCREDIT
COVID-19 Ag kit, and all the specimens are identified as negative.

Relative Specificity (%) = 100 x (No. of specimens with negative results /


No. of negative specimens tested by RT-PCR)

BIOCREDIT RT-PCT
Sensitivity Specificity
COVID-19 Ag
Positive Negative (%) (%)

Positive 12 0
Ag 80 100
Negative 3 2
(12/15) (2/2)
Total 15 2

3) UTM specimen test has a specimen dilution effect comparing to direct nasopharyngeal
swab collection method. Consequently, sensitivity becomes lower than original swab
specimen when BIOCREDIT COVID-19 Ag is tested using UTM specimen.

[Comment]

(A001, 08Mar2020) PCR CT Value (E:31.95, R:34.49, N:35.95) & (A002, 14Mar2020) PCR CT Value (E:32.60, R:34.83, N:36.59)

In these two cases, PCR CT Value show near offset(decision) value (34~36). According as the offset value setting, this result
can be decided by "NEGATIVE" also.

Including these two cases, Sensitivity is 12/15, 80%. Except them (as invalid result), 12/13, 92.3%.

RQP-702-F09(‘18.04.24) REV.02 RapiGEN