You are on page 1of 6

RESEARCH AND EDUCATION

Influence of ceramic firing on marginal gap accuracy and


metal-ceramic bond strength of 3D-printed Co-Cr frameworks
Adolfo Di Fiore, DDS, PhD,a Gianpaolo Savio, MScEng, PhD,b Edoardo Stellini, DDS,c Paolo Vigolo, DMD, MSD,a
Carlo Monaco, DDS, MScPhD,d and Roberto Meneghello, MScEng, PhDe

ABSTRACT
Statement of problem. The marginal gap and ceramic bond strength of metal-ceramic restorations are important for success. However,
studies evaluating the marginal gap and ceramic bond strength of fixed partial dentures (FPDs) produced with 3D printing technologies
such as selective laser melting (SLM) are scarce.
Purpose. The purpose of this in vitro study was to investigate the marginal gap of cobalt-chromium (Co-Cr) alloy frameworks produced by
SLM technology before and after ceramic firing. Additionally, the metal-ceramic bond strength was evaluated with the Schwickerath
crack-initiation test according to the International Standards Organization (ISO) 9693-1:2012.
Material and methods. Conventional impressions were made, and the definitive cast of a patient requiring a 4-unit FPD was scanned. After
designing the FPD, the files were sent to a service center for the fabrication of a metal master model, 80 Co-Cr frameworks, and 80 flat
specimens (25×3×0.5 mm) with SLM technology. The marginal gap between frameworks and the abutment tooth of the metal master
model was nondestructively measured by using an optical coordinate-measuring machine. A total of 80 sets, consisting of 1 framework
and 1 flat specimen, were sent to 80 dental laboratory technicians for ceramic firing. Detailed instructions for correct manipulation of the
framework and flat specimen were provided. The marginal gap was remeasured, and the 3-point bend test was used to evaluate
metal-ceramic bond strength.
Results. Only 28 of the 80 dental technicians returned the specimens within a prespecified time and/or in adequate condition. The mean
±standard deviation marginal gap of the framework before ceramic firing was 25 ±9 mm and 34 ±12 mm after firing. The difference was
statistically significant (P=.001). The mean ±standard deviation 3-point bend strength was 33 ±9 MPa.
Conclusions. Ceramic firing affected the marginal gap; however, all Co-Cr frameworks had a marginal gap lower than 120 mm, which is
reported to be a clinically acceptable limit. Most of the specimens (80%) had a metal-ceramic bond strength value higher
than the 25-MPa ISO 9693 requirement. Five of 28 dental laboratory technicians were not able to comply with ceramic firing
instructions. (J Prosthet Dent 2019;-:---)

Marginal gap and bond strength between the ceramic biological complications such as secondary caries or
and the metal framework are important factors affecting periodontal inflammation or mechanical complications
the success and survival of metal-ceramic fixed partial such as veneer chipping.2,3,7
dentures (FPDs).1-3 A marginal gap of less than 120 mm Cobalt-chromium (Co-Cr) alloy is commonly used to
has been reported to be clinically acceptable.4-6 Poor fabricate FPD frameworks because of its excellent
prosthesis fabrication may increase marginal gap with mechanical properties, durability, and low cost.8

The authors have no commercial or financial dealings that may pose a conflict of interest or potential conflict of interest.
a
Adjunct Professor, Department of Neurosciences, School of Dentistry, University of Padova, Padova, Italy.
b
Assistant Professor, Departments of Civil, Environmental and Architectural Engineering, University of Padova, Padova, Italy.
c
Full Professor and Head of Dental Clinic and School of Dentistry, Department of Neurosciences, University of Padova, Padova, Italy.
d
Researcher and Assistant Professor, Division of Prosthodontics and Maxillofacial Rehabilitation, Department of Biomedical and Neuromotor Sciences (DIBINEM), Alma
Mater Studiorum - University of Bologna, Bologna, Italy.
e
Associate Professor, Departments of Management and Engineering, University of Padova, Padova, Italy.

THE JOURNAL OF PROSTHETIC DENTISTRY 1


2 Volume - Issue -

would be found between the marginal gap of Co-Cr alloy


Clinical implications frameworks obtained before and after ceramic firing. The
The marginal gap of frameworks fabricated by using second null hypothesis was that the ceramic bond
strength would not be influenced by the workflow of the
the selective laser melting technology before and
dental technician.
after ceramic firing was clinically acceptable;
however, not all dental technicians were able to
MATERIAL AND METHODS
follow ceramic firing instructions with regard to
metal-ceramic bond strength requirements. The definitive cast of a 4-unit FPD was scanned by using
a laboratory scanner (Zfx Evolution; Zfx GmbH), and the
digitized data were transferred to a dental CAD software
Conventional lost-wax technology is problematic for Co- program (exocad; exocad GmbH) for the design of the
Cr FPD frameworks,9 with difficulty in achieving FPDs. Standard tessellation language (STL) files of the
adequate accuracy.10,11 Computer-aided design and model and FPDs were then sent to a manufacturing
computer-aided manufacturing (CAD-CAM) technolo- center for review and the fabrication of a Co-Cr master
gies have replaced many traditional prosthetic proced- model (Fig. 1). Eighty Co-Cr frameworks (EOS M270;
ures, including the casting process, with benefits that EOS GmbH Electro Optical Systems), made of EOS
include avoidance of human error.12,13 Cobalt-Chrome SP2 powder and 80 stereolithographic
Both subtractive machining and additive 3D-printed models (EOS Formiga P100; EOS GmbH
manufacturing have been used for framework fabrication, Electro Optical Systems) were made from the EOS
and a combination of additive and subtractive processes polyamide PA 2200 powder (Fig. 2). Eighty plate speci-
has been described.14 Among different additive mens (25×3×0.5 mm),32,34 conforming to ISO 9693,34
manufacturing technologies, selective laser melting (SLM) were fabricated by using the same SLM system and
has become popular in prosthetic dentistry.15 SLM sys- powder (Fig. 3).
tems, based on powder-bed technology, fabricate metal The master model, 80 frameworks, and 80 plate
components with adequate accuracy of about 20 mm from specimens were sent to the laboratories of the Depart-
a 3D CAD cast of the framework by using a laser beam to ment of Civil, Environmental, and Architectural Engi-
melt thin layers of Co-Cr metal powder.15,16 neering for preprocessing assessment. The marginal gap
The marginal gap of Co-Cr FPD frameworks depends between the frameworks and the abutment tooth of the
not only on the accuracy of the fabrication method but master model was measured by using an optical coor-
also on subsequent production steps, including ceramic dinate-measuring machine (OCMM; OGP SmartScope
firing,17,18 which could deform the framework and thus Flash CNC 300; Quality Vision International Inc) with a
increase the marginal gap.1,19,20 In addition to a clinically multiprobing system equipped with a contact mechanical
acceptable marginal gap, adequate bond strength be- probe, an optical video probe, and a laser probe. The
tween the ceramic and the metal framework is neces- measuring system had a maximum allowable error of
sary.21,22 The marginal gap for metal-ceramic crowns has E1(z)=2.5+5L/1000 mm, E2(xy)=1.8+5L/1000 mm, and
been reported to range from 26 to 138 mm depending on E3(xyz)=2.8+5L/1000 mm (with L, in millimeters,
whether these are fabricated by a traditional or digital equating to the distance measured), according to ISO
workflow.23 The failure rate from veneer chipping has 10360 standards.35,36
been reported to range from 2.3% to 8%.24 Information Measurements were made with the framework placed
on the marginal gap of Co-Cr metal-ceramic FPDs pro- on the master model and seated with an elastic by
duced by the SLM technology before and after ceramic following an automated measuring procedure requiring
firing is limited. A few studies have investigated the effect no operator intervention. The marginal gap was assessed
of ceramic firing on the marginal gap of Co-Cr from optical video probe grayscale images at ×50
crowns,10,25,26 with most studies using the silicone and ×400 magnifications (Fig. 4). For each framework, a
replica technique.27,28 The ceramic bond strength29-33 of total of 30 marginal gap measurements were made, 5
Co-Cr frameworks fabricated by different workflows has each on the buccal, mesial, and lingual sides of the canine
also been reported. and of the molar. Each marginal gap value was obtained
The purpose of this in vitro study was to have as a red (centroid) area (Fig. 4B) detected by using an
different dental technicians investigate the marginal gap image analysis algorithm (Rhinoceros 5.0 Beta; Robert
before and after the ceramic firing of Co-Cr alloy McNeel & Associates), and its centroid location was
frameworks by SLM. Additionally, ceramic bond strength identified; then, the minimum radial distance from the
was measured by using the Schwickerath crack-initiation centroid location to the perimeter of the area was
test (3-point bend test), according to ISO 9693-1.34 The calculated, and the marginal gap was defined as twice the
first null hypothesis was that no significant difference value of the radial distance. The repeatability of the

THE JOURNAL OF PROSTHETIC DENTISTRY Di Fiore et al


- 2019 3

Figure 1. Master cast for marginal gap analysis. Figure 2. SLM Co-Cr framework before ceramic firing. SLM, selective
laser melting.
1.1 ±0.1

8 ±0.1
0.5 ±0.05

25 ±1

Figure 3. Specimen configuration for metal-ceramic bond


characterization (mm).

measurement method is shown in Table 1. The Chau- Figure 4. Marginal gap before ceramic firing. A, original
venet criterion was used to filter the data to avoid mar- magnification ×50. B, original magnification ×400.
ginal gap measurement errors from excessive local calibrated in the range of ±300 N, applying a crosshead
standard deviation.37 speed of 1.5 mm/min according to ISO standard 9693.34
Marginal gap measurements were obtained before The ceramic debond strength (sb ) was calculated ac-
and after ceramic firing, with the influence on framework cording to the following formula: sb = FFAIL k, where
deformation being obtained from the difference between FFAIL is the fracture force in N and kk is the correction
the 2 measurements. For metal-ceramic bond strength coefficient, which depends on the mechanical charac-
characterization, the specimens were measured by using teristics and thickness of the specimens. kk was ascer-
digital calipers (ABSOLUTE Digimatic Caliper Series 551; tained using the algorithm given in ISO standard 9693.34
Mitutoyo Europe GmbH) with a 0.01-mm resolution, The marginal gap (mm) was considered as the statistical
ensuring the specimens conformed to the dimensional unit. The primary variable was the deformation of the
tolerances of the ISO standard 9693 (Fig. 3).34 Eighty sets framework before and after ceramic firing. Descriptive
(each comprising 1 framework and 1 flat specimen) were and comparative statistics were analyzed, and the results
sent to 80 different private, independent dental techni- were submitted to the Shapiro-Wilk test, where P=.0002
cians for ceramic firing together with a letter providing was considered not normally distributed. The Wilcoxon
detailed instructions on the appropriate handling of the matched-pairs signed-rank test (1-tailed) was then used
framework and specimens. The sets were collected from to compare groups (a=.05). A statistical software program
the dental laboratories within 2 weeks (Figs. 5, 6), and (SPSS v16.0; SPSS Inc) was used for the analysis.
the marginal gaps were measured with the same auto-
mated method used before ceramic firing. The 3-point RESULTS
bend test (Fig. 7) was used to measure ceramic bond
strength in a universal testing machine (Acumen 3; MTS As only 28 of 80 dental technicians returned the speci-
Systems Corp) equipped with a 4-column load cell and mens by the deadline and/or in adequate condition, 28

Di Fiore et al THE JOURNAL OF PROSTHETIC DENTISTRY


4 Volume - Issue -

Table 1. Repeatability of marginal gap (MG) analysis methodology


Side of FPDs Location MG Mean (mm) Standard Deviation (mm)
Buccal 1 23.6 1.4
2 20.2 0.3
3 22 0.9
4 8 1.1
5 7.5 1.1
6 33 0.6
7 41.2 0.8
8 45.7 0.3
9 43.5 0.7
10 42.4 0.4
Mesial 1 26 1.9
2 15.2 6.7
3 31.1 3
4 37.9 0.9
5 37.9 0.9
Palatal 1 43.6 0.8
2 49.1 0.6 Figure 6. SLM Co-Cr specimens after ceramic firing. SLM, selective laser
3 50.1 1 melting.
4 55.8 0.2
5 46.1 0.6
6 17.6 1.4
7 6 0.9
8 11.6 1.6
9 20.6 0.8
10 18.5 1.4
Distal 1 37.6 3
2 48.8 6.9
3 62.9 3.2
4 59.4 2.3
5 43.1 1.3

Figure 7. Three-point bend test apparatus with SLM Co-Cr specimen


after failure. SLM, selective laser melting.

DISCUSSION
The present study examined the influence of ceramic
firing on the marginal gap of 28 FPDs and the ceramic
bond strength of 28 specimens of cobalt-chromium alloy
specimens fabricated by SLM. Significant differences
emerged between the marginal gap values obtained
Figure 5. SLM Co-Cr framework after ceramic firing. SLM, selective laser
before and after ceramic firing, so the first null hypothesis
melting. was rejected. The second null hypothesis was also re-
jected because not all plate specimens met ISO 9693
standard requirements for sb >25 MPa.34 All the marginal
frameworks and specimens were analyzed after ceramic gap values were still clinically acceptable, however. The
firing. The mean ±standard deviation (SD) marginal gap maximum clinically acceptable marginal gap has been
of the framework was 25 ±9 mm, maximum=53 mm and reported to be 120 mm,4-6 but a marginal gap of around
minimum=13 mm, before ceramic firing, and 34 ±12 mm, 50 to 80 mm is recommended to avoid mechanical and
maximum=63 mm and minimum=20 mm, after firing. The biological complications.
difference in the marginal gap between before and after Few studies have examined the marginal gap before
ceramic firing was 9 ±8.6 mm. The Wilcoxon rank-sum and after ceramic firing of FPDs produced using
test showed that this difference was statistically signifi- SLM.6,10,25 Zeng et al25 analyzed the marginal gap of
cant (P<.001). The mean ±SD of the 3-point bend test dental Co-Cr alloy copings fabricated by SLM after
was 33 ±9 MPa, maximum=50 MPa and minimum=14 repeated firing and reported values of 36 mm, 37 mm, 38
MPa. Eighteen specimens did not meet ISO 9693 stan- mm, and 38 mm after the first, third, fifth, and seventh
dard requirements for sb >25 MPa.34 firings. The authors concluded that repeated firing had

THE JOURNAL OF PROSTHETIC DENTISTRY Di Fiore et al


- 2019 5

no significant influence on the marginal gap but failed to nonstandardized ceramic firing revealed the different
mention any measurements obtained before ceramic workflow of the dental technicians, and the forces
firing. Results consistent with the present study were applied in the 3-point bending test were not represen-
reported by Kaleli and Sarac10: the marginal gap tative of clinical conditions. However, distortion of the
increased after porcelain application (from 49 mm before marginal gap can be caused by many factors other than
firing to 59 mm), but the gap was still clinically acceptable. ceramic firing, such as alloy type and margin design.1,19,20
The authors also reported that the marginal gap was Overall, using the OCMM could be a more standardized
narrower for coping fabricated by laser sintering than for procedure. All SLM Co-Cr frameworks showed clinically
coping prepared by milling or the conventional lost-wax acceptable marginal gaps. Although the gaps increased
method. However, Kocaa g lu et al6 concluded that
g ao after ceramic firing, they still did not cause clinically
applying the veneering ceramic influenced the marginal relevant distortions. In vivo studies are needed to assess
gap of coping prepared by SLM (from 72.7 mm before to the performance of frameworks fabricated by the SLM
94.7 mm after firing). technology.
Katosoulis et al26 examined the fit of complete-arch
FPDs fabricated by milling before and after veneering, CONCLUSIONS
with findings similar to those of the present study,
Based on the findings of this in vitro study, the following
although different technologies were used in the 2 studies.
conclusions were drawn:
The ceramic firing had no effect on the fit, and all the
values of marginal gaps were clinically acceptable. The 1. Ceramic firing affected the marginal gap and the
median vertical gap was of 19 mm for unveneered and 20 metal-ceramic bond strength.
mm for veneered FPDs. The present data identified a 9-mm 2. All Co-Cr frameworks had a marginal gap narrower
mean difference in marginal gaps obtained before and than 120 mm, which is considered the clinically
after ceramic firing and narrower marginal gaps compared acceptable limit.
with other studies.10,25 This difference may be attributable 3. Only 23 of 28 specimens had a metal-ceramic bond
to the methods used to measure the marginal gap. Most strength value higher than the 25 MPa specified by
authors used a scanning electron microscope (SEM)10,25,27 ISO 9693, as 5 of 28 dental technicians were unable
or the silicone replica technique6,28; 1 study used micro- to follow the ceramic firing instructions.
computed tomography,27 and OCMM was only used in
the present study. OCMM ensures a maximum error about
REFERENCES
10 times lower than that of the optical microscope
(5 mm).11,14,35,36 One of the main advantages of using 1. Joias RM, Tango RN, Junho de Araujo JE, Junho de Araujo MA, Ferreira
Anzaloni Saavedra Gde S, Paes-Junior TJ, et al. Shear bond strength of a
OCMM, moreover, is that, unlike the SEM or silicone ceramic to Co-Cr alloys. J Prosthet Dent 2008;99:54-9.
replica method, it allows the marginal gap to be measured 2. Bader JD, Rozier RG, McFall WT, Ramsey DL. Effect of crown margins on
periodontal conditions in regularly attending patients. J Prosthet Dent
without operator error. 1991;65:75-9.
To achieve high clinical survival rates of metal- 3. Felton DA, Kanoy BE, Bayne SC, Wirthman GP. Effect of in vivo crown
margin discrepancies on periodontal health. J Prosthet Dent 1991;65:357-64.
ceramic FPDs and clinically acceptable marginal gaps, it is 4. McLean JW, von Fraunhofer JA. The estimation of cement film thickness by
essential to ensure an excellent bond between the an in vivo technique. Br Dent J 1971;131:107-11.
5. Örtorp A, Jönssön D, Mohsen A, Von Steyern PV. The fit of cobalt-chromium
ceramic and the substructure. The mean ceramic bond three unit fixed dental prostheses fabricated with four different techniques: a
strength of the specimens in the present study was 33 ±9 comparative in vitro study. Dent Mater 2011;27:356-63.
6. Kocaagaoglu H, Kılınç HI, Albayrak H, Kara M. In vitro evaluation of mar-
MPa, but not all specimens satisfied the ISO 9693 stan- ginal, axial, and occlusal discrepancies in metal ceramic restorations produced
dard requirement of sb >25 MPa.34 Six of 28 dental with new technologies. J Prosthet Dent 2016;116:368-74.
7. Nicolaisen MH, Bahrami G, Finlay S, Isidor F. Comparison of fatigue resis-
technicians were unable to return the frameworks with tance and failure modes between metal-ceramic and all-ceramic crowns by
ceramic applied. cyclic loading in water. J Dent 2014;42:1613-20.
8. Lee DH, Lee BJ, Kim SH, Lee KB. Shear bond strength of porcelain to a new
The bond strength of ceramic to Co-Cr dental alloy millable alloy and a conventional castable alloy. J Prosthet Dent 2015;113:
obtained with SLM has been reported to be 33 ±3 MPa 329-35.
9. Willer J, Rossbach A, Weber HP. Computer-assisted milling of dental res-
by Li et al,29 46.8 ±5.1 MPa by Wang et al,30 and 55 ±3 torations using a new CAD/CAM data acquisition system. J Prosthet Dent
MPa by Wu et al.31 These authors all concluded that the 1998;80:346-53.
10. Kaleli N, Saraç D. Influence of porcelain firing and cementation on the
ceramic bond strength did not depend on the marginal adaptation of metal-ceramic restorations prepared by different
manufacturing method.29-33 The mean bond strength in methods. J Prosthet Dent 2017;117:656-61.
11. Paniz G, Stellini E, Meneghello R, Cerardi A, Gobbato EA, Bressan E. The
the present study was lower probably because the precision of fit of cast and milled full-arch implant-supported restorations. Int
ceramic firing workflow was not standardized (the J Oral Maxillofac Implants 2013;28:687-93.
12. Park JK, Kim HY, Kim WC, Kim JH. Evaluation of the fit of metal ceramic
specimens were sent to different dental technicians to restorations fabricated with a pre-sintered soft alloy. J Prosthet Dent
simulate actual clinical conditions). 2016;116:916-23.
13. Tamac E, Toksavul S, Toman M. Clinical marginal and internal adaptation of
Limitations of this study include the use of only one CAD/CAM milling, laser sintering, and cast metal ceramic crowns. J Prosthet
technology to produce the frameworks. The subsequent Dent 2014;112:909-13.

Di Fiore et al THE JOURNAL OF PROSTHETIC DENTISTRY


6 Volume - Issue -

14. Ciocca L, Meneghello R, Monaco C, Savio G, Scheda L, Gatto MR, et al. 29. Li J, Chen C, Liao J, Liu L, Ye X, Lin S, et al. Bond strengths of porcelain to
In vitro assessment of the accuracy of digital impressions prepared using a cobalt-chromium alloys made by casting, milling, and selective laser melting.
single system for full-arch restorations on implants. Int J Comput Assist J Prosthet Dent 2017;118:69-75.
Radiol Surg 2018;13:1097-108. 30. Wang H, Feng Q, Li N, Xu S. Evaluation of metal-ceramic bond character-
15. Azari A, Nikzad S. The evolution of rapid prototyping in dentistry: a review. istics of three dental Co-Cr alloys prepared with different fabrication tech-
Rapid Prototyping J 2009;15:216-25. niques. J Prosthet Dent 2016;116:916-23.
16. Koutsoukis T, Zinelis S, Eliades G, Alwazzan K, Rifaiy MA, Al Jabbari YS. 31. Wu L, Zhu H, Gai X, Wang Y. Evaluation of the mechanical properties and
Selective laser melting technique of Co-Cr dental alloys: a review of structure porcelain bond strength of cobalt-chromium dental alloy fabricated by se-
and properties and comparative analysis with other available techniques. lective laser melting. J Prosthet Dent 2014;111:51-5.
J Prosthodont 2015;24:303-12. 32. Kaleli N, Saraç D. Comparison of porcelain bond strength of different metal
17. Gemalmaz D, Alkumru HN. Marginal fit changes during porcelain cycles. frameworks prepared by using conventional and recently introduced fabri-
J Prosthet Dent 1995;73:49-54. cation methods. J Prosthet Dent 2017;118:76-82.
18. Shiratsuchi H, Komine F, Kakehashi Y, Matsumura H. Influence of finish line 33. Ren XW, Zeng L, Wei ZM, Xin XZ, Wei B. Effects of multiple firings on metal-
design on marginal adaptation of electroformed metal-ceramic crowns. ceramic bond strength of Co-Cr alloy fabricated by selective laser melting.
J Prosthet Dent 2006;95:237-42. J Prosthet Dent 2016;115:109-14.
19. Cho SH, Nagy WW, Goodman JT, Solomon E, Koike M. The effect of mul- 34. International Organization for Standardization. ISO 9693-1. Dentistry
tiple firings on the marginal integrity of pressable ceramic single crowns. compability testing. Part 1: Metal-ceramic systems. Geneva: International
J Prosthet Dent 2012;107:17-23. Organization for Standardization; 2012. ISO Store Order: OP-184149 (Date:
20. Fonseca JC, Henriques GEP, Sobrinho LC, De Goes MF. Stress-relieving and 2017-06-09). Available at: http://www.iso.org/iso/home.html.
porcelain firing cycle influence on marginal fit of commercially pure titanium 35. Di Fiore A, Meneghello R, Savio G, Sivolella S, Katsoulis J, Stellini E.
and titanium-aluminum-vanadium copings. Dent Mater 2003;19:686-91. In vitro implant impression accuracy using a new photopolymerizing
21. Stannard JG, Marks L, Kanchanatawewat K. Effect of multiple firing on the SDR splinting material. Clin Implant Dent Relat Res 2015;17(Suppl 2):
bond strength of selected matched porcelain-fused-to-metal combinations. 721-9.
J Prosthet Dent 1990;63:627-9. 36. Savio G, Meneghello R, Concheri G. Optical properties of spectacle
22. Ozcan M. Fracture reasons in ceramic-fused-to-metal restorations. J Oral lenses computed by surfaces differential quantities. Adv Sci Lett 2013;19:
Rehabil 2003;30:265-9. 595-600.
23. Witkowski S, Komine F, Gerds T. Marginal accuracy of titanium copings 37. Thompson WR. On a criterion for the rejection of observations and the
fabricated by casting and CAD/CAM techniques. J Prosthet Dent 2006;96: distribution of the ratio of deviation to sample standard deviation. Ann Math
47-52. Stat 1935;6:214-9.
24. Libby G, Arcuri MR, LaVelle WE, Hebl L. Longevity of fixed partial dentures.
J Prosthet Dent 1997;78:127-31.
Corresponding author:
25. Zeng L, Zhang Y, Liu Z, Wei B. Effects of repeated firing on the marginal
accuracy of Co-Cr copings fabricated by selective laser melting. J Prosthet Dr Adolfo Di Fiore
Dent 2015;113:135-9. Department of Neurosciences
26. Katsoulis J, Mericske-Stern R, Enkling N, Katsoulis K, Blatz MB. In vitro Dental School
precision of fit of computer-aided designed and computer-aided manufac- University of Padova, via Giustiniani 2
tured titanium screw-retained fixed dental prostheses before and after 35100, Padova
ceramic veneering. Clin Oral Implants Res 2015;26:44-9. ITALY
27. Kim EH, Lee DH, Kwon SM, Kwon TY. A microcomputed tomography Email: adolfo.difiore@unipd.it
evaluation of the marginal fit of cobalt-chromium alloy copings fabricated by
Acknowledgments
new manufacturing techniques and alloy systems. J Prosthet Dent 2017;117:
The authors thank 3D Fast for technical support.
393-9.
28. Huang Z, Zhang L, Zhu J, Zhang X. Clinical marginal and internal fit of metal
ceramic crowns fabricated with a selective laser melting technology. Copyright © 2019 by the Editorial Council for The Journal of Prosthetic Dentistry.
J Prosthet Dent 2015;113:623-7. https://doi.org/10.1016/j.prosdent.2019.08.001

THE JOURNAL OF PROSTHETIC DENTISTRY Di Fiore et al

You might also like