Installation

International Service Department


BC-5300

Version: 1.3, Update Date: 2009-5-7


Installation

1. Installation requirements
2. Unpacking
3. Removal and installation
4. Connections
5. Start up and adjustment
6. Shut down
7. Upgrade the software

Version: 1.3, Update Date: 2009-5-7


Installation requirements-Tools

Flat screw driver Scissors 7um particles

Controls / Calibrator/Tissue/Gloves/Other necessary tools

Version: 1.3, Update Date: 2009-5-7


Installation requirements-Space

Space Requirement:
1. Between right or left panel and the wall ≥ 100 cm
2. Between back panel and the wall ≥ 50 cm
3. Make sure there are enough space for you to put the
reagents bottle.

Version: 1.3, Update Date: 2009-5-7


Installation requirements-Power

Power Requirement:
1. Voltage: A.C.100V-240V
2. Input power: ≤300 VA
3. Frequency: 50/60 Hz
4. Make sure the analyzer is properly grounded.

Version: 1.3, Update Date: 2009-5-7


Installation requirements-Environment
Environment Requirement:
1. Optimal operating temperature: 15℃-30℃
2. Optimal operating humidity: 30%-85%
3. Atmospheric pressure: 70KPa-106KPa (corresponding
altitude range: -400-3000 meters, serving as a reference
only);
4. The environment should be as free as possible from dust,
mechanical vibrations, pollution, loud noises and
electrical interference;
5. Free from electrical interference.

Version: 1.3, Update Date: 2009-5-7


Installation requirements- PC Configuration
PC configuration requirements:
1. Memory: 256M or above
2. IE version: 6.0 or higher
3. Hardisk:4G or above
4. Resolution: 1024 x 768 or higher
5. OS: Win2000 Professional+SP4(English version), Win XP
Home/XP Professional+SP2(English version), Windows vista
ultimate 32, Windows vista home basic 32
Notice: Need to install 2 network boards on the computer if
you would like to connect the LIS with the BC-5300.

Version: 1.3, Update Date: 2009-5-7


Unpacking-main unit
Top wood

Packing tapes

Version: 1.3, Update Date: 2009-5-7


Unpacking-main unit
Accessory kit

Paper protective
corners

Version: 1.3, Update Date: 2009-5-7


Unpacking-main unit
Remove the foam

Version: 1.3, Update Date: 2009-5-7


Unpacking-main unit
Remove the packing bag

Version: 1.3, Update Date: 2009-5-7


Unpacking-checking the packing list

Check the packing list to see if everything is delivered.

Version: 1.3, Update Date: 2009-5-7


Removal and installation
-- remove the clamp and tie

Clamp

Plastic
cable tie

Version: 1.3, Update Date: 2009-5-7


Removal and installation
-- install the software in PC

Please install the software just under the


default path, or you must make sure the data
address do not contain any other character
besides English.

Version: 1.3, Update Date: 2009-5-7


Removal and installation
-- install the software in PC

Install software Finish

Version: 1.3, Update Date: 2009-5-7


Removal and installation--IP setting of PC

Set IP of PC :10.0.0.1

Version: 1.3, Update Date: 2009-5-7


Connections--cable

Use this network cable to connect Connect to power source.


the analyzer with computer

Version: 1.3, Update Date: 2009-5-7


Connections- reagents

Reagents, identified by
color when connecting

Waste

Version: 1.3, Update Date: 2009-5-7


Connections- reagents

Install the
supporting board for
diluent.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Username and password

Login name and password:

Service engineer:
Name: service
Password: Se s700
Then click „OK‟ to go on.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Initializing

The machine starts


initializing.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Initializing

You can choose to initialize the


liquid system or not. But we
recommend not to abort the
initializing.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Reagent settings

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Prime reagents

If you have passed liquid initialization, please access Menu-Service-


Maintenance-Replace reagent to replace reagent.
Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Cleaning

Access to Clean-clean fluidics to clean the liquid system

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Other setting

Set the printers that might be used

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-HGB gain

Check if HGB blank


voltage is 4.5 or not. If
not, please adjust HGB
gain to be about 4.5V.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Background test

Access Work list.


Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Background test

Select work mode: click new, and choose “WB” “CBC+DIFF”, then click
“Save”.
Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Background test

Press start key, then it will do background test.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Background test

Click special info


Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Background test
Diff pulse data
length should be
less than 200.
If not, you may
need to clean the
DIFF bath, flow cell,
and related tubing.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Particle test

1. Prepare particle: press “Dliuent”, and add diluent 3 times then add one
drop 7um particle.
2. Then select work mode: WB CBC+DIFF and use the prepared particle
to run as sample.
Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Particle test

Click the particle test result in “Table” and click “Optical”


Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Particle test

Input the target value of FS SS peak value, then click calculate,


record the new gain.
Version: 1.3, Update Date: 2009-5-7
Start up and adjustment-Particle test
7um Standard particle target value

Param Before version 1.4 After version 1.4

FS 26 3 18 2

SS 117 5 104 6

SS 0.1Max Width ≤18 ≤15

FS 0.1Max Width ≤13 ≤6.5

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Particle test
Access Menu-
Setup-Gain,
please first
record the
original gain.
Then input the
new gain to FS
and SS, and save.

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Quality control

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Quality control

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Calibration

Version: 1.3, Update Date: 2009-5-7


Start up and adjustment-Calibration

Version: 1.3, Update Date: 2009-5-7


Shut down

Press “shut
down”

Version: 1.3, Update Date: 2009-5-7


Maintenance Menu-Replace reagent

Version: 1.3, Update Date: 2009-5-7


Maintenance Menu-Clean

Version: 1.3, Update Date: 2009-5-7


Maintenance Menu-Maintain

Version: 1.3, Update Date: 2009-5-7


Maintenance Menu-Maintain the whole device

Version: 1.3, Update Date: 2009-5-7


Self-test Menu-Motors

Version: 1.3, Update Date: 2009-5-7


Self-test Menu-Valves

Version: 1.3, Update Date: 2009-5-7


Self-test Menu-Others

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- check the version first

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- upgrade the IPU

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- upload the file into main unit

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- upgrade main unit three times

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- confirm the corresponding files

Version: 1.3, Update Date: 2009-5-7


Software upgrade procedure:
-- succeed and check the version again

Version: 1.3, Update Date: 2009-5-7


Thank you!

Version: 1.3, Update Date: 2009-5-7