You are on page 1of 33

ZR893

OXYGEN ANALYSER

OPERATOR’S INSTRUCTION
MANUAL
CONTENTS

PAGE

SECTION 1: INTRODUCTION 3

1.1 PREFACE 3
1.2 WARRANTY 4
1.3 TECHNICAL SPECIFICATIONS 5
1.4 INSTALLATION 6
1.5 SAMPLE SYSTEM 7

SECTION 2: USING THE INSTRUMENT 8

2.1 START-UP 8
2.2 INSTRUMENT OPERATION 9
2.3 OPERATORS MODE 11
2.4 SET-UP MODE 12
2.5 TABLE OF AVAILABLE SCREEN 13
2.6 SETTING UP THE PASSWORD 14

SECTION 3: PROGRAMMING THE INSTRUMENT 15

3.1 OPERATORS MODE 15


3.2 SET-UP MODE 17
3.3 CALIBRATION 24

SECTION 4: OPTIONAL FACILITIES 26

4.1 ZR893 CONNECTIONS 26


4.2 SETTING THE ANALOGUE OUTPUTS 27
4.3 SWITCHING BETWEEN 4-20mA OUTPUT 28
AND 0-20mA OUTPUT
4.4 ALARM OUTPUT OPTIONS 29
4.5 SETTING THE ALARMS 29
4.6 USING A PRINTER 30
4.7 PRINTING WITH THE ZR893 31

SECTION 5: SPARE PARTS & ACCESSORIES 32

5.1 SPARE PARTS 32

DRAWING: B 893 007


Systech Instruments Model ZR893 Instruction Manual

SECTION ONE: INTRODUCTION

1.1 PREFACE

The ZR893 continuous oxygen monitor is an electro-chemical oxygen analyser using a high
purity, high density stabilised zirconia ceramic.

The zirconia tube is held in its own heated furnace and the resulting voltage signal generated
by the zirconia is antilogged and displayed in direct oxygen concentration on a high visibility
LED display panel.

The instrument is capable of reading between 0.5 ppb to 100% and is autoranging. The
concentration is displayed in parts per billion, parts per million and percent.

The ZR893 oxygen monitor can exist in a number of standard variations and these are
covered in this manual.

The instrument has however been designed in such a way that sub-variations to customer’s
specific requirements may be provided. Where these have been included, the data sheet
has been appended.

We are always trying to improve our product, of which this manual is part and so we would
greatly appreciate any information you can give us of any difficulties you may encounter.

IMPORTANT

Please read this manual- at least part 1 before attempting to install or operate the equipment.

The sample gas is heated to 650oC and therefore no combustibles should be present
at a limit to cause an explosion.

The equipment should be electrically connected and grounded in accordance with good
standard practice and at the correct voltage. (Refer to the voltage tag label on the back of the
instrument).

No responsibility is accepted by Systech Instruments


for accidents resulting from improper use of this equipment.

___________________________________________________________________________

3 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

1.2 WARRANTY

This instrument is guaranteed for a period of one year from its delivery to the purchaser
covering faulty workmanship and replacement of defective parts. This assumes fair wear
and tear and usage specified on the data sheet. It does not cover routine calibration and
housekeeping. Systech are happy to quote for routine servicing of the equipment at the
customer's premises.

We maintain comprehensive after sales facilities and the instrument should be returned to
our factory for repairing or servicing if this is necessary.

The type and serial number should always be quoted together with full details of any fault.

Warranty repair or replacement will be made without charge noting the above comments.
Equipment must be shipped prepaid after return authorisation has been obtained.

All service and technical enquires are covered from our factory in Thame, Oxfordshire, where
we will endeavour to give a quick and helpful response to al enquiries.

The factory address details are:

Systech Instruments Ltd


17 Thame Park Business Centre
Wenman Road
Thame
Oxfordshire Tel: +44 (0) 1844 216838
OX9 3XA Tel: +44 (0) 1844 217220

For customers outside UK, your local agent is:

___________________________________________________________________________

4 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

1.3 TECHNICAL SPECIFICATIONS

RANGES: Oxygen: 0.5 ppb - 100%

RANGE SELECTION: Autoranging

DISPLAYS: 1 x 4 digit high visibility LED


12 digit alphanumeric dot matrix display

RESPONSE TIME: 90% of indication is 10 seconds

ACCURACY: ± 2% of the reading

REFERENCE GAS: Air (20.9% oxygen)

SUITABLE GASES: All inert gases

AMBIENT TEMPERATURE: 0-40oC

POWER: 230/115V ±10%, 50/60Hz at 90VA

WEIGHT: 7.6 Kg

SIZE: BENCH MOUNT 449W X 133H X 318D (mm)


19" RACK MOUNT 483W X 133H X 318D (mm)

OPTIONS

SIGNAL OUTPUTS: Voltage: 0-100mV & 0-10V


Current: 4-20mA or 0-20mA
(All isolated)

ALARMS: Four, all with volt free c/o contacts fully


configurable over the complete oxygen range

COMMS I/F: Serial interface allowing connection


to computer or printer

PRINTER I/F: Parallel interface allowing print connection

___________________________________________________________________________

5 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

4.7 PRINTING RESULTS WITH THE ZR893

Manual Printing

The third column of the OPERATORS mode allows the instantaneous printout of the results
of an analysis to be printed out on any printer that is connected to the ZR893.

The printout shows the time and date of the printout and also the oxygen span and zero.

At the end of the printout it shows whether it was a timed or a manual printout.

Timed printing

The ZR893 can be programmed such that the instrument automatically prints out every few
minutes.

The automatic printout can be set as follows:

1) Enter the SET-UP mode. The timed printout set up can be found on the fifth row of
the SET-UP mode.

2) The dot matrix screen should display the message PRINT MINS and a number the
number relates to the number of minutes between printouts. To program in a new
number type in the number using the number pad and press ENTER to log it into
memory.

3) The timed printout is now set, to return to the OPERATORS mode press the ENTER
key.

4) If you wish to turn off the timed printout follow the steps above and on the PRINT MINS
screen type in zero and press ENTER. The instrument will not do a timed printout.

31 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

1.4 INSTALLATION

The ZR893 is operated from AC line power either 230V + 10% 50Hz or 115V + 10% 60Hz.
The voltage range to which the instrument is set is marked on the rear panel and should be
checked before power is connected.

The IEC standard cable supplied should be connected to a captive plug with an integral earth
connection and wired as follows:

Brown to Live
Blue to Neutral
Green/Yellow to Earth

CARE SHOULD BE TAKEN TO ENSURE THAT THE REAR OF THE INSTRUMENT IS


ALLOWED SUFFICIENT VENTILATION FOR HEAT DISSIPATION FROM THE
FURNACE.

___________________________________________________________________________

6 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

1.5 SAMPLE SYSTEM

The analyser requires little or no sample system with the exception of a pressure regulator to
reduce high pressure streams to below 7 barg or a sample pump for streams below 0.25
barg.

If the sample stream is dirty or wet, both solid material and moisture need to be removed
from the sample stream to ensure reliable and accurate results.

A suitable filter and knockout pot can be supplied by Systech.

Metal tubing is recommended for the sample lines, especially if the instrument is being used
for measuring trace levels.

Hard plastics, such as nylon, Teflon or PVC may be used for measurements in the high ppm
or percent regions.

The use of rubber or soft plastics is not recommended as they may leak and lead to
inaccurate measurements.

On permanent installations, where the process stream is situated at great distance from the
analyser, it is advisable to use the bypass loop flowmeter to provide an up to date sample at
the analyser.

___________________________________________________________________________

7 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

SECTION TWO: USING THE INSTRUMENT

2.1 START-UP

1. Ensure that the pressure of the sample line (and calibration gas if used) is between
0.25 and 7 bars gauge. If not, either a pump or reducing value should be fitted. When
installation requires these extra sample handling components, and the expected
oxygen concentration in the sample is less than 100ppm, these components should
be high integrity - leak tight.

Contact Systech for further information.

The sample outlet should not be taken to a point other than atmosphere pressure.

2. Connect the gas lines to the bulkhead connectors.

3. Check the flow through the instrument, the flow should be between 100-200cc/min.
The sample flow meter is located on the front panel of the instrument and the float
indicates the present flow through the cell.

4. The power switch is mounted on the IEC receptacle situated at the back of the
instrument. The switch should be depressed to the ON position. When this is done,
the instrument will light and the dot matrix display will show the message NOT
READY, since the instrument cannot produce accurate readings until the sensor is at
the proper operating temperature. Pressing ENTER at this stage will stop the NOT
READY flashing on the screen. The instrument takes 20 minutes to reach operating
temperature, although it is advisable to allow 35 minutes for the Instrument before any
analysis is done. If in doubt check the furnace temperature located in the
OPERATORS mode. However, if the instrument is measuring trace oxygen
levels, it is advisable to allow the instrument a couple of hours to stabilise to the
correct reading. When the furnace and zirconia cell are at equilibrium temperature,
and all the ‘pockets’ of high oxygen have been swept out of the system.

___________________________________________________________________________

8 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

2.2 INSTRUMENT OPERATION

The ZR893 is controlled entirely through an alphanumeric display and keypad.

There are two basic modes of operation. The first is the OPERATORS mode and this is the
normal everyday running mode. When the instrument is in this mode, the oxygen
concentration is displayed on the dot matrix screen.

When the instrument is first turned on and it has heated to the correct operating temperature,
the micro-processor automatically leaves the user in this mode.

The other mode of operation is the SET-UP mode and this has access via a password which
allows limited access in this mode to stop unauthorised tampering.

These modes access grid information and the display may be thought of as a small window
which can scan over a much larger page of information. The window can be moved up or
down, left or right. When the window is at the edge of the page, a message warns that you
cannot go further. (TOP LIMIT, BOTTOM LIMIT, LEFT LIMIT, RIGHT LIMIT).

DISPLAY 7 8 9

4 5 6

1 2 3

F1 SET-UP CE ENTER 0 . -

KEYBOARD LAYOUT

___________________________________________________________________________

9 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

The Arrows

The arrows on the keyboard allow a user to move through both the OPERATORS mode and
the SET-UP mode.

The ENTER button

The ENTER button is for switching between modes. So to enter the SET-UP mode press the
ENTER button and the instrument will ask for a password. Enter the password and press
enter twice to go into the SET-UP mode.

To leave the SET-UP mode press the ENTER button again and you will be returned to the
OPERATORS mode.

Note: On receiving the instrument it will not have a password in memory and
pressing ENTER twice will allow entry to the SET-UP mode.

For more information on setting password please see Section 2.6.

The CE button

The CE button will clear any typing errors that have been made whilst trying to type a number
in allowing the correct number to be input.

The ENTER/TEST button

This button will log into memory any number that has been typed using the number pad.

The DATA button

This button allows us future expansion, and at the moment this button is not in use and
nothing will happen if it is pressed.

The LED display

This shows the instantaneous reading of oxygen, and the display will also show whether the
sample is in the percent, ppm, or ppb range.

___________________________________________________________________________

10 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

2.3 OPERATORS MODE

There are six windows accessible in the operation mode. These are:

i) Oxygen value
ii) Cell temperature
iii) ENTER for printout
iv) Eprom version
v) Serial number
vi) Instrument I.D.

The first window gives the instantaneous values for the oxygen.

The cell temperature indicates the operating temperature of the internal zirconia furnace.
The normal operating temperature is 650oC.

ENTER for a printout is to download the instantaneous value for oxygen to a parallel or serial
printer.

The last three are for reference only, but EPROM version and serial number should be
quoted with any query to Systech.

___________________________________________________________________________

11 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

2.4 SET-UP MODE

The SET-UP mode allows an engineer to change certain settings on the instrument for
calibrating, setting analogue outputs etc.

To access the SET-UP mode, press the enter key on the keypad and the message ENTER
PASSWORD will be displayed. The instrument is shipped with no password and if no
password has been entered, then the SET-UP mode can be accessed by pressing enter
twice. (Section 2.6)

To leave the SET-UP mode press ENTER key again and the instrument returns to the
OPERATORS MODE.

___________________________________________________________________________

12 of 32

Version: 2.3 Instruction Manual Systech Instruments Ltd


Updated: 02/11/04 Thame
U.K.
2.5 TABLE OF AVAILABLE SCREENS
OPERATORS MODE
O2 CELL PRESS ENTER FOR EPROM SERIAL INST
VALUE TEMP PRINTOUT VERSION NUMBER I.D.

SET-UP MODE (ACCESSED BY PASSWORD)


*1 ANALOGUE
ANALOGUE OUTPUT
OUTPUT HIGH O2
LOW O2

*2 ALARM 1 ALARM 1 ALARM 2 ALARM 2 ALARM 2 ALARM 3 ALARM 3 ALARM 3 ALARM 4 ALARM 4 ALARM 4
ALARM 1 HI/LO SET O2 OR HI/LO SET O2 OR HI/LO SET O2 OR HI/LO SET
O2 OR VALUE FAULT VALUE FAULT VALUE FAULT VALUE
FAULT

O2 ZERO
ADJUST

*3 O2 O2 CAL =
O2 AUTO AUTOCAL
CAL MINS
(EN-DIS)

*3 SEQ CAL ENTER HOLD =


SEQ CAL MANUAL/ FOR SEQ ON/OFF
HOURS = AUTO CAL

ENTER ENTER CLRS


PRINTS DATA LOG
ENGR LOG

PRINT MINS

TIME SET DATE SET PRINTER


SET
*1 Screen depends on set up - refer to Section 3.
*2 Note that if the alarms option has not been purchased then the alarms row will not be present.
*3 Note that if the autocal option has not been purchased the two autocal rows will not be present.
Systech Instruments Model ZR893 Instruction Manual

2.6 SETTING UP THE PASSWORD

To allow limited access to the SET-UP mode, a password can be programmed into the
instrument. By pressing the ENTER key the message ENTER PASSWORD will be
displayed, press the ENTER key once and the message NEW PASSWORD? will be
displayed, then a password can be entered using the keypad. The password can be any
number between 10 and 9999999. Once the password has been typed in, press ENTER to
log it into memory.

Now, when entering the SET-UP mode the microprocessor will ask you to ENTER
PASSWORD. Type in your password and press ENTER twice.

NB: You are strongly advised to keep a note of your chosen password in a safe place.
Losing the password can be very time consuming and expensive to rectify.

The set-up mode accesses a 12 x 8 page, the six rows are as follows:

(i) Analogue output calibration row, see section on analogue outputs.


(ii) Alarm set-up row, see section on optional alarms.
(iii) Calibration row, see calibration section.
(iv) Auto calibration row, see section on optional autocal.
(v) Auto calibration row.
(vi) Data logging row.
(vii) Automatic printout, see section on printing.
(viii) Time, date and printer configuration row, see time and date setting section.

14 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

SECTION THREE: PROGRAMMING THE INSTRUMENT

3.1 OPERATORS MODE

Column 1

O2 Value

The instrument automatically leaves you in this window when the internal furnace
reaches its operating temperature. This value represents the value of the oxygen content.

Column 2

Cell Temperature

This value is the instantaneous temperature of the internal zirconia furnace.

Column 3

ENTER for printout

Press enter for a printout to a serial or parallel printer connected to the RS232 or parallel
socket respectively. Protocol for the printer is detailed in the PRINTER SECTION of the
manual.

Column 4

Eprom
Version

This shows the version of the eprom contained within the instrument.
Please quote this with any enquiry to Systech.

15 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

Column 5

Serial Number

This shows the instrument serial number.


Please quote the instruments serial number with any enquiry to Systech.

Column 6

Instrument
I.D

This shows the instrument I.D.


This is used when more than one instrument is being used on the premises. When the
instrument prints out the instrument I.D. is printed with the results so that the instrument
with which a batch is tested is recorded on the printout.

16 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

3.2 SET-UP MODE

(Press enter key and enter PASSWORD to enter this mode, see section 2.5)

ROW 1 ANALOGUE OUTPUT CONFIGURATION ROW (OPTION)

SINGLE OUTPUT

ROW 1 COLUMN 1

O2 LO

Note: These screens will be available even if the option has not been purchased but there
will be no output on the rear of the instrument.

This allows setting of the bottom of the scale for the analogue output for oxygen. This
can be set up such that a chart recorder will only read only over a certain oxygen range.

The required value can be entered using the keyboard and pressing ENTER to log it into
memory.

ROW 1 COLUMN 2

ANALOGUES (SINGLE)

Pressing the enter button toggles the display to read analogues (multi) - leave in the
single mode.

ROW 1 COLUMN 3

O2 HI

As detailed above, this sets the top of the scale for the analogue output.

Note: The output range may be selected with a live zero such as 5.0ppm - 8.0ppm
oxygen or 95% - 100% oxygen for example.

17 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

MULTI OUTPUT

The instrument is able to give multi analogue output ranges (up to eight) and together with
the range ID (identification) contact closure on the back panel can give the operators the
choice of one single output or a number of outputs.

Pressing the enter button in the second window will change the display to read analogues
(multi). Press the right arrow to now show the O2 HI(1) = XX. Pressing the up and down
arrow will change the index of the output and using the keyboard different values of
analogue output may be preset (up to eight). Volt free contact closures on the back panel
(see Drg. B893 007 at the rear of the manual) may be used to show the operator/plant
computer what range is represented by the analogue outputs.

Note that when multi analogue outputs are used the analogue low output is fixed at 0ppm.
It is not possible to have suppressed zero.

18 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

ROW 2 ALARM CONFIGURATION ROW

ROW 2 COLUMN 1

ALARM 1 O2 OR
FAULT

Note: This row will not be available if the alarms option has not been purchased.

This screen allows the user to set whether the alarm is to be initiated on oxygen or when
a fault occurs. Pressing the ENTER button toggles between the two.

See section 4.4.

ROW 2 COLUMN 2

ALARM 1 HI or LO

This allows the user to program the alarm to go off on either higher than a certain value or
lower than a certain value which can be set in the next window.

ROW 2 COLUMN 3

ALARM 1 VPM %

An exact value can be set for the alarm in vpm or %.

Note: When vpm displayed on the dot matrix screen typing 10 will set the alarm at
10vpm, to change the range to percent, 10000 must be typed in to change from 1vpm to
1% and vice versa so to change from 1% to 1vpm 0.0001 must be typed in.

The ZR893 analyser allows the setting of three more alarms which are set as detailed for
Alarm 1 see section 4.4.

19 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

ROW 3 CALIBRATION ROW

ROW 3 COLUMN 1

02 ZR ADJ

This allows calibration of the oxygen zero value and a new value can be input using the
keyboard.

Pressing ENTER whilst on this screen will automatically reset the instrument to 20.9%.
See section 3.3.

ROW 4 CALIBRATION ROW

ROW 4 COLUMN 1 (OPTION)

02 AUTO CAL =
[DIS/EN]

This screen allows the oxygen auto calibrate option to be enabled (EN) or disabled (DIS).
Pressing the ENTER button toggles between the two.

ROW 4 COLUMN 2

02 AUTO CAL MINS =

This screen allows you to enter the time that the Auto Calibrate solenoid is switched on
and after this time the O2 calibrate will be reset.

ROW 4 COLUMN 3

02 CAL =

This allows you to input the O2 calibration value that will be set after the time period set in
window Row 5 Column 3.

20 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

ROW 5 SEQUENCE CAL ROW (OPTION)

ROW 5 COLUMN 1

SEQ CAL HRS =

This screen allows you to enter the time period between calibrations when Row 5
Column 2 is set to [timed].

ROW 5 COLUMN 2

SEQ CAL =

This screen allows you to set the calibration sequence to [manual] or [timed]. Pressing
the ENTER button will toggle between the two. As soon as you toggle to [timed] the
period set in Row 5 Column 1 will begin counting down.

ROW 5 COLUMN 3

ENTER ---> SEQ


CAL

If Row 5 Column 2 is set to manual. Press ENTER will start the calibrate sequence.

Also if Row 5 Column 2 is set to manual the calibration sequence can be started by a
customer contact closure across two connections on the back panel (see Drg. B893
007) at rear of manual.

ROW 5 COLUMN 4

HOLD = OFF/ON

Pressing enter toggles between off and on.

21 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

When performing a manual or a timed, unattended autocalibration you may want to


suppress analogue outputs and alarms from operating which may be linked into the plant
operation. The instrument may be programmed to suppress alarms and outputs and will
hold for a further five minutes after the autocal has finished in order for the instrument to
equilibrate back to the plant gas.

Note

While doing an autocalibrate the operator display will show A/Cal O2 = XXX and after the
autocal has finished it will show O2 = XXX for five minutes, the asterix indication it has
just finished a cycle and may be in equilibration.

A 'calibration in sequence' contact closure is located on the rear terminal strip - see
Drg. B893 007.

ROW 6 THE ENGINEERS LOG

ROW 6 COLUMN 1

ENTER PRINTS
ENGR LOG

Pressing ENTER whilst on this screen will send a printout of the engineers log to a printer
if the printer option has been purchased.

ROW 6 COLUMN 2

ENTER CLRS
ENGR LOG

Pressing ENTER whilst on this screen will clear the ZR893 memory of the data stored
from the engineers log.

Note that once data has been cleared from memory it is not possible to retrieve it.

22 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

ROW 7 PRINTOUT CONFIGURATION

ROW 7 COLUMN 1

PRINT MIN

The number displayed on this screen is the number of minutes between the instrument
printing out automatically. To change the number of minutes between printouts type in a
new figure on the instrument keyboard and press ENTER. For no printout this screen
should show a zero.

ROW 8 REAL TIME CLOCK AND PRINTER CONFIGURATION

ROW 8 COLUMN 1

Time

This sets the time and can be changed via the keyboard. Type in the new time and press
ENTER to log into the memory.

ROW 8 COLUMN 2

Date

This allows the date to be entered via the keyboard. Type in new date and press ENTER
to log into the memory.

ROW 8 COLUMN 3

PRINTER

Note: Ignore this screen if the print option has not been purchased.

This window allows the output to a printer to be set either for a parallel or a serial printer.
To switch between the two options press the ENTER key. SEE SECTION 4.6

23 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

3.3 CALIBRATION

WITHOUT AUTOCALIBRATE OPTION

Calibration should be carried out using a gas which is in the same range as the gas that
is being analysed.

For applications in the percentage range the calibration maybe carried out using
atmospheric air.

CALIBRATION WITH ATMOSPHERIC AIR

1. Set up the instrument such that atmospheric air is circulating through the
instrument.

2. Enter the SET-UP mode by pressing the ENTER button on the front panel of the
instrument.

3. Once in the SET-UP mode the third row is the calibration row. The dot matrix
screen should display O2 ZR ADJ and a number when in this row.

4. By pressing the ENTER button in this window the instrument will automatically
reset itself to 20.9 %.

5. Calibration with atmospheric air is now complete press the ENTER button to
return to the OPERATORS mode.

CALIBRATION WITH CERTIFIED GAS

1. Connect gas bottle to the sample inlet at the correct pressure.

2. Allow to equilibrate.

3. Enter the SET-UP mode as above.

4. Go to the window O2 SP ADJ = XXX.

5. Enter the value into the keyboard that is printed on the certificate with the bottle.
The analyser will only permit small offset values, so if in difficulty, first do an air
calibration, then a certified gas calibration and if still in difficulties call Systech.

24 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

SECTION FOUR: OPTIONAL FACILITIES

4.1 ZR893 CONNECTIONS

The ZR893 instrument offers various optional facilities which are shown below.

All the connections to the ZR893 are made in the rear of the instrument in 5 12 way
connection blocks.

Block T5 is the analogue outputs used for connecting a chart recorder or computer to the
ZR893.

Block T4 is the digital inputs. This allows the ZR893 to do special applications and remote
calibration.

Block T3 is the digital outputs. This allows the ZR893 to do special applications and
stream switching.

Block T7 are the alarm relays for connection of audible or visual devices.

Block T2 is the serial communications.

Block T8 & T9 is the analogue output range identification relays when the unit is in multi
output mode.

Block T10 is the expansion outputs as well as the fault relay and 'CAL ON' status relay.

26 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

WITH AUTOCALIBRATE OPTION

Refer to Section 3.2, Row 4 and 5. Air or certified mixture may be piped permanently to
the rear of the analyser and the analyser when so programmed may, once a week, day,
month on a timed basis check and adjust itself. This check may also be performed
manually by keyboard entry or by contact closure on the rear terminal strip.

25 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

4.2 SETTING THE ANALOGUE OUTPUTS (OPTION)

The analogue outputs provided are voltage (0-10V & 0-100mV) and current (4-20mA or
0-20mA), all isolated and configurable over the complete oxygen, carbon dioxide and nitrogen
ranges.

All the settings for the analogue outputs are found in the first row of the SET-UP MODE.

The first window contains the oxygen low setting and this can be entered in percent or vpm
using the keypad. The next window to the right contains the oxygen high settings and this is
set in the same way.

The analogue output may be programmed over any range, 0-1ppm, 0-2ppm, 0-24.6ppm,
even though the display of the analyser may at times be outside this range.

Some users may want to programme the analogue output over a number of ranges, i.e.
0-1ppm, 0-5ppm, 0-10ppm, 0-1000ppm and in conjunction with a series of contact closures
on the rear panel will know which range the analyser is operating.

See the programming section for further information.

27 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

4.3 SWITCHING BETWEEN 4-20mA OUTPUT AND 0-20mA OUTPUT

1. Ensure that the POWER IS TURNED OFF before attempting to remove the
instrument cover.

2. Remove the 4 screws retaining the top cover of the instrument.

3. Locate the 2 red jumper switches which can be found to the rear of the main printed
circuit board in front of the 50-way ribbon cable. (See diagram below). The board is
labelled 4-20mA.

4. The current output will normally have been set up as 4-20mA and by moving the two
red jumper switches to the left will give 0-20mA output.

50 WAY CONNECTOR

RED JUMPER SWITCHES


RESISTORS

YELLOW
POT

TOP HALF OF MAIN PRINTED CIRCUIT BOARD

The location of the jumper switches on the main printed circuit board

28 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

4.4 ALARM OUTPUT OPTIONS

If the instrument has been fitted with the alarms option, they may be used to provide a
visual/audible warning of when a reading has gone over a pre-defined level.

Refer to the programming section.

4.5 SETTING THE ALARMS

Enter the SET-UP mode by pressing the ENTER key, then type in your password or press
ENTER twice if no password has been input.

The alarm set-up windows are situated in the second row of the SET-UP mode.

The first three columns deal with ALARM 1, the next three columns ALARM 2 etc so on. The
instrument has 4 alarms when fitted.

The first column allows ALARM 1 to be set as either O2 or FAULTS and this is done by
pressing the ENTER key until the correct one is shown on the dot matrix screen.

The instrument faults that will initiate the FAULT alarm are:

i) thermocouple open circuit (a temperature control fault)


ii) high temperature
iii) low temperature

The high and low temperature faults relate to the internal zirconia furnace. Normally this
furnace will run at 650oC but if it is too high or low this may effect the reading.

The second column allows ALARM 1 to be set as a high alarm or a low alarm, and the
preference for HI/LO is set by pressing enter until the required one is displayed on the dot
matrix screen.

The third column sets the actual value in ppm or % and this is entered by the keypad. Press
enter to log it into memory. So, for example, setting ALARM O2, HI, 20% means that the
alarm will warn when O2 goes higher than 20%. The other alarms are set in the same
manner.

N.B. To set the alarm to 1 vpm from 1% 0.0001 must be typed in via the keypad and enter
pressed, and visa versa to change the alarm setting from 1ppm to 1% 10000 must be
entered via the keyboard.

There is a master fault alarm - refer to Drg. B 893 007.

29 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

4.6 USING A PRINTER

When the instrument is fitted with the print option it comes with both a RS232 serial port and
a parallel printer port. This interfaces are both situated at the back of the instrument next to
the IEC receptacle. Once the printer is plugged into one of the instruments ports it must be
configured such that:
Baud 9600
Bits 8
Parity DISABLED
Handshaking DTR
Line Feed ON.

this applies to both serial and parallel printers.


The cabling (RxTx Connections) for the parallel cabling should be pin to pin.
i.e RxTx Conns: 1 - 1
2 - 2
3 - 3
4 - 4
5 - 5 etc.

The serial cable requires that the 5 & 20 pins should be crossed as displayed below.

25 way D type 25 way D type


female connector male connector

1 1
2 2
3 3
4 4
7 7
5 5
20 20

9 core screened unused


wires to be tied back

The Serial printer lead connections

Note: The ZR893 instrument assumes a 80 column printer or greater as standard. If a printer
with less than 80 columns is used some of the printout will be printed on the line below.
Communication Software

Communication software is available, for more information please contact Systech.

30 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.
Systech Instruments Model ZR893 Instruction Manual

SECTION FIVE: SPARE PARTS & ACCESSORIES

5.1 SPARE PARTS

PART NUMBER DESCRIPTION

089 027 Assy Zirconia Probe

089 045 Assy Furnace 110V MkII

089 046 Assy Furnace 240V MkII

089 047 Assy Furnace Zirconia 240V MkII

089 048 Assy Furnace Zirconia 110V MkII

089 049 Assy Furnace Zirconia Special 240V MkII

089 050 Assy Furnace Zirconia Special 110V MkII

100 459 Fuse 5A (Power)

100 272 Control Valve

893 053 CPU pcb ZR893 Mk II All Options

893 030 CPU pcb ZR893 Mk II No Options

099-61244 Alphanumeric Display

893 034 LED pcb Display ZR893 Mk II

894 036 Keyswitch pcb

101 043 Flowmeter

101 037 Transformer

Important Note

The serial number of the instrument for which the spare parts are required must be quoted
on all orders.

32 of 32
Version: 2.3 Instruction Manual Systech Instruments Ltd
Updated: 02/11/04 Thame
U.K.

You might also like