Professional Documents
Culture Documents
net/publication/254531576
CITATIONS READS
8 570
2 authors:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Lorraine Scott Boak on 05 February 2017.
Summary Introduction
Scale inhibitors (SIs) are used to control oilfield scale formation, Squeeze treatments using large volumes of SIs remain one of the
and the ability to analyze these species is very important such most widely used techniques to protect the entire oil-production
that SI concentrations as low as 0.5-ppm active need to be mea- system against scale formation. This is especially significant for
sured accurately. If phosphorus is present in the SI molecule, then barium sulfate scale because once it is formed, much lost time
inductively-coupled-plasma (ICP) -based methods may be used and production may incurre while removing the deposit mechani-
for analysis. However, the oil industry’s increasing requirement to cally or by chemical means. In squeeze treatments, the injected
be environmentally friendly means that polymeric “green” SIs are inhibitor must return to the production tubing at levels above the
now being used more, which raises issues concerning detection threshold concentration Ct over long periods of time (typically, 3
techniques (i.e., ICP vs. wet-chemical techniques). ICP detec- to 12 months). After a squeeze treatment, the produced brine must
tion for SIs is generally easier, but if it cannot be used, then at be monitored to determine the SI concentrations so that the timing
least time-saving improvements to wet-chemical techniques are for a resqueeze can be correctly identified. Because threshold con-
extremely beneficial. centrations are often in the region of 5 ppm or less, it is important
In this paper, analytical approaches are described that have to accurately monitor the returning inhibitor at these low levels in
been used recently to improve chemical SI assay, especially at low order to determine the approach of the end of the squeeze lifetime.
near-threshold levels (a few ppm active of SI). Progress is reported This monitoring allows preparation time for a new squeeze treat-
in five areas of SI analysis: ment to be deployed. In addition, the monitoring and collating
1. Assay of sulfonated copolymer (VS-Co) was not possible of these low SI concentrations allows modeling of the SI-return
by straightforward analysis without extensive dialysis and sample profile from which predictions of the future squeeze lifetimes can
preparation. However, calibrations and repeats of accuracy similar be obtained. The optimization of these squeeze lifetimes can sig-
to that of the C18 were found for VS-Co using amino-propyl (NH2) nificantly reduce financial expenditure in terms of both chemical
cartridges and the Hyamine method. cost and deferred oil production.
2. The Oasis® 2×4 method has been applied to SI analysis, and Previous work on SI assay has been reported by this group
this is able to assay all types of polymeric SIs in principle. This (Sorbie et al. 1992; Graham et al. 1993, 1995a, 1995b, 1996) on
method has been used to detect a VS-Co SI in a wide variety of the basis of earlier developments (Christian 1993; Murphy and
different brine salinities from distilled water (DW) to high-salin- Riley 1962; Walters 1989; McTeir et al. 1993; Pennington 1988;
ity formation waters (FW) (e.g., a Heron-type FW). Although Kan et al. 1991; Bedford et al. 1994). The group’s earlier work
achievable under these different conditions, there was a signifi- has examined many problems associated with SI analysis in both
cant decrease in the absorbance signals recorded with increasing synthetic brines and in field-produced waters. The problems faced
salinity that was not significantly improved by a higher-capacity by oilfield analysts ranged from interferences with the analytical
sorbent cartridge. procedure by other constituents within the produced brines, pre-
3. Various elements have been assayed in the oil phase using cipitation, particulates, and limitations of the analytical techniques.
the ICP method. Calibrations and accurate repeats within 5 to 10% However, despite these difficulties, a number of successful proce-
error were achieved. After solving compatibility issues, the con- dures were obtained for the analysis of many generically different
centration of an oil-tolerant SI was determined successfully using SIs such as phosphonates (PH), polyphosphino carboxylic acids
calibrations and accurate repeats over a range of 0 to 10 ppm and (PPCAs), polyacrylic acids (PAAs), polyvinyl sulfonates (PVSs),
0 to 2,500 ppm active SI. and many others. A range of techniques were established, from
4. A matrix-matching Hyamine technique has been developed wet-chemical to the use of ICP, because no single technique can
that allows any chloride-ion effects on the chelating process be used for all the generically different SI species. A summary
between the Hyamine and SI to be negated, allowing accurate of the various SI assay methods with estimates of their analytical
analysis of low-polymeric SI concentrations. accuracy is shown in Table 1. The detailed techniques developed
5. ICP and wet-chemical techniques have been able to accu- previously for oilfield SI application along with the structures of
rately detect a P-tagged (phosphorus-tagged) copolymer-type SI. most common SIs can be found in the FAST Laboratory Manual
The ability to apply two independent analytical methods to a given (Sorbie and Boak 2006).
species offers some important advantages when more than one SI Since these early years, these techniques have not only been
is deployed in a field system. In this work, excellent correlation used widely throughout the oil industry, but they have also been
is observed between the wet-chemical and ICP assay methods for adapted and developed to allow more-accurate results for the con-
this P-tagged SI. ventional SIs of that time.
This study updates and adds to the set of analytical methods and However for more than 15 years now, the oil industry is
procedures reported for SI analysis almost 20 years ago (Graham et much more environmentally aware. One of the consequences is
al. 1993, 1995a, 1995b, 1996; Sorbie et al. 1992) and are described that more-environmentally-friendly SIs have been developed and
in our Flow Assurance and Scale Team (FAST) laboratory proce- applied. These are known as green scale inhibitors (GSIs), and they
dures manual (Sorbie and Boak 2006). do not pollute the waters around platforms and the pipeline sys-
tems through bioaccumulation and nondegradation (Inches et al.
2006). GSIs are also being applied in chemical squeeze treatments.
Although these GSIs are more environmentally friendly, they bring
their own set of problems to the analysis world because they do
Copyright © 2010 Society of Petroleum Engineers
not contain phosphorus, making analysis by ICP nonapplicable. In
This paper (SPE 130401) was accepted for presentation at the SPE International some cases, it may be possible that a P-tag can be attached to the
Conference on Oilfield Scale, Aberdeen, UK, 26–27 May 2010, and revised for publication.
Original manuscript received 8 March 2010. Revised manuscript received 22 June 10. GSI, which for now appears to be in limited concentration, making
Paper peer approved 26 July 10. low-level SI analysis by ICP potentially possible but extremely
Phosphonate (PH) ICP 0.1 0.1 < 0.2 < 0.2 Very accurate detection in various brines.
Possibly matrix independent.
Phosphino poly- Hyamine 0.1 0.1 0.2 0.4 Accurate detection in various brines.
carboxylate (PPCA) Errors larger in very contaminated brines.
Phosphino poly- ICP 0.1 0.1 0.1 0.1 Very accurate detection in various brines.
carboxylate (PPCA) Possibly matrix independent
Poly acrylate (PAA) Hyamine 0.1 0.1 0.2 0.4 Accurate detection in various brines.
Errors larger in very contaminated brines.
Poly vinyl sulphonate Pinacyanol 0.2 0.5 0.8 1.0 Detection in low-salinity brines.
(PVS)** Blue Repeat analysis essential for accuracy.
Pretreatment to isolate inhibitor assists
detection in high-salinity brines.
Background response very variable—daily
calibration and repeat analysis essential.
Poly vinyl sulphonate Hyamine 0.2 0.3 0.2 0.5 Accurate detection in many brines.
(PVS)** Pretreatment essential to isolate inhibitor.
PVS/PAA copolymer** Hyamine 0.2 0.3 0.3 0.5 Accurate detection in many brines.
Pretreatment essential to isolate inhibitor.
PVS/PAA copolymer** Pinacyanol 0.2 0.5 0.8 1.0 Detection in low-salinity brines.
Blue Repeat analysis essential for accuracy.
Pretreatment to isolate inhibitor assists
detection in high-salinity brines.
Background response very variable—daily
calibration and repeat analysis essential.
* Graham et al. (1995b)
** Graham et al. (1993)
difficult. Therefore, the industry may have to resort to wet-chemi- However, C18 cartridges have never been successful with sul-
cal techniques or other instrumentation such as high-pressure fonated products. This relates to the bonding that occurs between
liquid chromatography (HPLC) (Dionex). In either case, major the SI and the silica sorbent within these cartridges. The SI is
improvements are needed to allow low-level SI assay in produced applied at pH = 2 with C18 cartridges. This ensures that the SI
brine to be acheived, at less than 0.5 ppm active in some cases. is fully nondissociated, allowing the SI to be retained on the C18
Some of the current techniques, such as membrane or car- cartridge through H-bonding or weak interactions. This allows
tridge dialysis, to remove the interfering brine salts before using separation of the SI from the brine salts that flow through the
Hyamine 1622 for generic species such as PVS and VS-Co need cartridges and are discarded. This cannot be achieved for sul-
to be made less laborious, or an alternative technique needs to be fonated products because they remain extensively dissociated at
found. Recently, an alternative method for the detection of sul- pH = 2 because they possess such low pKa values. Therefore,
fonated SI products using amino-propyl cartridges was reported by the sulfonated SI cannot be retained on the C18 by reverse phase
researchers (Jones et al. 2007), and this method has been applied hydrogen bonding. Hence, in late 2006, investigations into alter-
here. Following this approach, a number of other solid-phase- native cartridges, such as the amino-propyl (NH2) cartridges, for
extraction (SPE) methods for polymeric-SI analysis were studied, sulfonated products commenced. This work coincided with other
allowing the development of the OASIS® 2×4 method described researchers’ examinations (Jones et al. 2007) presented at the
in this paper. National Association of Corrosion Engineers (NACE) 2007 con-
ference describing their techniques for sulfonated products using
Experimental Procedures these same NH2 cartridges.
The final SI assay experimental procedures developed in the course FAST’s investigations examined these cartridges using the
of this work are presented in detail at the end of this paper in normal C18 procedure. As stated previously, these NH2 cartridges
Appendix A, some of which are also detailed in the FAST labora- are more polar and basic in nature. Hence, ionic interactions could
tory manual (Sorbie and Boak 2006). The various brine composi- occur between the sulfonated SI and the NH2 cartridge at low pH
tions used thoughout the following studies are also detailed in allowing the sulfonated SI to be retained on the cartridge, separat-
Appendix A. ing it from the salt ions (mainly the Cl– ions) that interfere with
the ionization of the Hyamine—quaternary ammonium chloride—
Presentation of Data and Results reducing the turbidity signal. The normal 5% sodium citrate wash,
C18 and NH2 (Amino-Propyl) Cartridges and Hyamine. C18 however, did not appear strong enough to release the sulfonated SI
cartridges have been used routinely for many years for Hyamine from the cartridge to allow its detection. After trying a number of
analysis of PPCA and other SIs containing the carboxylic acid different solutions, 0.1N NaOH was found to be of a sufficiently
functional group. Fig. 1 and Table 2 illustrate the calibrations and high pH value (pH 10–12) to release the sulfonated product from
repeats that can be obtained for PPCA, polymaleic acid (PMA), the NH2 cartridge to allow its detection. An initial calibration (and
and polyaspartate using C18 cartridges. Note that the calibration repeats) in DW was achieved for VS-Co. This led to trials of both
and repeats for PMA had to be performed simultaneously for the VS-Co and PVS in seawater (SW) where successful calibrations
results to be successful. and repeats were also achieved, as shown in Fig. 2 and Table 3.
Fig. 1—Examples of calibration graphs that can be obtained following the C18 Hyamine procedure for (a) PPCA, (b) PMA, and
(c) polyaspartate.
STANDARD REPEATS FOR PPCA, PMA, AND STANDARD REPEATS FOR VS-Co in DW/SW and PVS/SW
POLYASPARTATE USING C18 HYAMINE USING NH2 HYAMINE
OASIS® 2×4 Method. Oasis cartridges are not thought to have Therefore, all further experiments were performed on the WAX
been used previously for polymeric-SI assay within the oil indus- cartridge. Since there is great flexibility with these cartridges,
try despite having been featured very prominently throughout due to of the high stability of the sorbent material, a number of
the pharmaceutical industry for many years. As a result, a major variables were examined:
method development was undertaken to establish a method that • Effect of pH on the SI being retained on the cartridge
could detect both conventional SIs and GSIs because these car- • Volume of reagents (i.e., of acid, alkali, DW wash, sodium
tridges have two main advantages: (1) the robustness of the citrate, Hyamine)
sorbent material to aid separation from interferences and (2) the • Volume of SI sample to achieve greatest signal, which in
high throughput of 20 tests simultaneously that can be achieved. turn should provide better repeatability, eventually requiring an
A brief explanation of how this SPE method works using the vari- introduction vessel for larger sample volumes to avoid errors from
ous types of cartridge (MCX, MAX, WCX, and WAX) is presented pipetting
in Appendix A, Section A2. • Type of acid used to pretreat sorbent before SI wash step to
An initial screening of all the cartridges (MCX, MAX, WCX, lock SI onto cartridge
and WAX) was performed on the basis of the C18 Hyamine • Type/concentration of alkali used to release SI from sorbent
method. This showed that the main cartridge of interest for oilfield before its detection by Hyamine method
SIs appeared to be WAX, although under certain circumstances, By optimizing these factors to allow the greatest and clean-
a signal was detected for PPCA and PCA on the MCX cartridge. est separation of the SI from the interfering SW salts, a method
(a) NH2 Hyamine for VS-Co/DW (b) NH2 Hyamine for VS-Co/SW (c) NH2 Hyamine for PVS/SW
10 8 8
y = 173.17x 3 –101.69x 2 + 35.969x– y = 984.51x 3 –308.76x2 + 56.556x– y = 109x 3 –48.577x 2 + 30.434x–0.1
[SI], active ppm
[SI], active ppm
8
[SI], active ppm
2 2 2
0 0 0
0.000 0.100 0.200 0.300 0.400 0 0.05 0.1 0.15 0.2 0 0.1 0.2 0.3
Absorbance (500nm) Absorbance (500nm) Absorbance (500nm)
Fig. 2—Calibration graphs obtained following the NH2 Hyamine procedure for (a) VS-Co in DW, (b) VS-Co in SW, and (c) PVS in SW.
(e) Calibration for VS-Co in (f) 0–2 ppm Calibration (g) Calibration for VS-Co in (h) 0–2ppm Calibration
Forties FW at initial pH2, WAX for VS-Co in Forties FW Heron FW at initial pH2, WAX for VS-Co in Heron FW
2.5 2.5
10 12 y= -31768x2 + 624.31x -
3 2
y= -2091.5x2 + 134.12x -
y= 2435.5x -680.18x + 0.132 10 2 0.6008
8 87.924x + 0.0176 2 R2 = 0.9997
R2 = 1
Absorbance (500 nm) Absorbance (500 nm) Absorbance (500 nm) Absorbance (500 nm)
Fig. 3—Calibration graphs obtained for VS-Co initially in various salinity brines: DW, NSSW, Forties FW, and Heron FW; (a, c,
e, g) Full 0–8 or 0–10 ppm active range achieved and (b, d, f, h) 0–2 ppm active calibration using the same absorbance values
improves accuracy at low concentrations, mainly 1 ppm active.
was developed that has the potential to detect both conventional an FW-type brine system (i.e., containing a high quantity of salts),
SIs and GSIs, although the major focus after initial examinations an increase in the absorbance signals of SIs, especially at the high-
was on the polymeric VS-Co-type inhibitor. This VS-Co inhibitor end of the calibration range, would be observed. However, this
was initially examined in DW and SW, but this analysis was later was not the case. A slight increase was observed for the low end
extended to include higher-salinity FWs such as a medium-scaling of the calibration range but no significant increases for the high
Forties-type FW and an even-higher-salinity Heron-type FW. The end, implying that all the SI was “sticking” to the previously used
calibrations and repeats for the VS-Co SI in a variety of different lower-capacity sorbent, and there was no competition between the
salinity brines—DW, North Sea seawater (NSSW), Forties FW, and salts and SI.
Heron FW—are shown in Fig. 3 and Table 4. It was observed that
if the calibration range was decreased, then more-accurate repeats ICP Analysis in the Oil Phase. Aqueous ICP methods have
were recorded for the lower concentrations. been used routinely for many years. However, with nonaqueous
It was also observed that as the salinity of the brine system SI treatments now being widely applied for scale protection, the
increased, the absorbance signal decreased. This was thought to detection of various elements within an oil matrix needs to be
relate to sorbent site competition on the cartridge between the brine examined. The main difference between ICP aqueous and oil anal-
salts and the SI during the SI lock-on in Step 3 of the procedure. ysis is the range of settings required in order to light the plasma,
It was hoped that by using a higher-sorbent-capacity cartridge for as highlighted in Appendix A. Once this has been achieved, then
TABLE 4—EXAMPLE OF THE R COEFFICIENT AND STANDARD REPEATS FOR THE 10 -mL
2
ANALYSIS OF THE SIs, VS-Co IN THE VARIOUS SALINITY BRINES: DW, NSSW, FORTIES
FW, AND HERON FW; (a) FULL CALIBRATION REPEATS, (b) 0–2ppm CALIBRATION
REPEATS
Calibration R
2
0.9995 0.9999 0.9992 0.9963
Repeats
0ppm 0.109 –0.049 0.105 0.303
0.5ppm 0.336 0.147 0.400 0.303
1ppm 0.661 0.538 0.906 0.849
4ppm 3.15 3.92 4.02 4.20
6ppm 5.77 5.90 5.17 5.85
(b) 0–2 Calibration DW NSSW Forties FW Heron FW
Calibration R
2
1.0000 0.9989 0.9972 0.9997
Repeats
0ppm 0.109 –0.098 –0.024 –0.008
0.5ppm 0.357 0.274 0.397 –0.008
1ppm 0.823 0.934 1.049 0.986
Note: The absorbance for the 0ppm repeat was similar to the absorbance of the 0ppm in the calibration however
depending on the curve fit the determined 0ppm concentration could be higher/lower than expected.
methods can be set up for the ions of interest. The oil methods testing the analysis method. No major differences in accuracy or
are very similar to the aqueous methods, with any differences repeatability were noted between the 0- to 2,500-ppm active cali-
mainly relating to an element’s wavelength and background cor- brations and repeats, performed for all the 10 to 50% EGMBE to
rection points. The analysis can then be performed, and successful white spirit ratios. Hence, in order to choose one ratio, the deterio-
calibrations were achieved. The accuracy and repeatability of the ration in the signal of the solutions was examined after 1 week. It
results obtained for the ions Ca2+, Mg2+, Fe3+, S–, P–, Y+, and Li+ was observed that as long as a new 50 ppm standard was prepared
are shown in Table 5. in a manner similar to that of the previous one (from the original
The first column of results for each ion shows the original raw 500 ppm diluted standard and not from the very high original
ICP data and the next shows the determined drift-corrected ion stock), the repeatability of the 50 ppm standard was acceptable.
concentration. Because these two sets of concentrations are very However, no major differences were recorded for the calibration
close in value (both to themselves and the known concentration), intensity values from the original or the 1-week-later analysis,
the analysis was deemed successful. Therefore, having achieved although both the 10 and 50% EGMBE/SI solutions were observed
both accurate and repeatable analysis for several ions, it was of to have the least variation. These two ratios of EGMBE/SI to white
great importance to be able to determine SI concentrations within spirit were then carried forward to a low-calibration range of 0 to
an oil sample. 10 ppm active SI, which were successful.
Initially, incompatibility issues were observed for an oil-toler- The ICP raw data recorded for the 10% EGMBE/SI: 90% white
ant SI in white spirit. However, after a number of investigations spirit ratio was initially more accurate. However, once the data had
examining different solvent packages and different ways of dilut- been drift corrected, then data from either EGBME/SI to white
ing the oil-tolerant SI, it was decided to look at different ratios of spirit ratio were acceptable to use. However, it was decided that
ethylene glycol monobutyl ether (EGMBE) (or 2-butoxyethanol), the 10% EGMBE/SI: 90% white spirit was the most suitable ratio
containing the SI to white spirit. It was found that ratios of 10 to to be used for this oil-tolerant SI, as shown in Table 6.
50% EGMBE/SI to white spirit gave clear solutions, while 60% This work shows that it is possible to analyze for ions of interest
EGMBE/SI and upward gave cloudy or precipitated solutions, in an oil matrix, such as white spirit. This includes an oil-tolerant
which could not be used. Because the 10 to 50% EGMBE/SI to SI, although it required a constant matrix of 10% EGMBE/SI:
white spirit solutions were clear, these ratios could be used for 90% white spirit and for all the standards to be made fresh at the
same time or for any to be made by exactly the same procedure a 0-to-10 ppm active SI DW calibration range, then initially 0-to-
as when made for the first time. This oil-phase analytical method 100 ppm SW calibration standards would be prepared. These SW
will be applied to a nonaqueous phase coreflood during the oil- calibration standards would then require the same ×10 DW dilution
flowback period. as the samples (i.e., 500 s and 4,500 d for 5 mL), and hence both
the sample and calibration standards will have equivalent chloride-
Easy Hyamine: Matrix-Matching Technique. This method ion concentration. Thus, the sample and calibration standards will
development was carried out because of the difficulties of deter- undergo similar interferences from these ions, allowing accurate
mining low polymeric-SI concentration from a restricted sample analysis to be performed (i.e., matrix matching).
volume. The procedure for the actual detection is exactly the Fig. 4 shows the successful calibration graph achieved for this
same as the normal easy Hyamine in Appendix A, Section A4 of ×10 dilution matrix-matching technique.
this paper. However, it is the steps taken before the analysis that This process was also successfully performed for ×5, ×2.5,
require some thought. The final procedure developed avoids the and ×2 dilution factors for the matrix-matching technique from
negative effects of the remaining chloride ion on the chelating SW to DW. With no dilution at all in DW (i.e., at full SW chlo-
process between the Hyamine and SI. For example, an SI stock ride-ion composition, approximately 19,700 ppm), a calibration
would be prepared in SW, as would happen for any experimental cannot be achieved. No dilutions in DW between 0 and 2 have
procedure, which would then be diluted to give a 25 ppm active been performed.
SI/DW stock in order to create an easy Hyamine calibration If all the calibration graphs for both easy Hyamine and matrix-
range of 0 to 10 ppm active SI/DW (i.e., easy Hyamine relies matching Hyamine at various dilutions are plotted together, it is
on the diluting out of any salts that may interfere with the SI obvious that the presence of chloride ions can have a significant
detection). effect on the absorbance recorded, as shown in Fig. 5.
However, it was found that as the SI/SW concentration became For example, if a sample gave an absorbance value of approxi-
lower in the samples, requiring a smaller dilution in DW (e.g., a ×2 mately 0.24, then depending on which calibration had been used
dilution—2,500 d and 2,500 s on the diluter for a 5-mL DW test during the analysis, the determined ppm value of the sample could
solution), then a greater chloride ion concentration is present with range between approximately 2.7 and 6 ppm active SI. Therefore,
respect to the target 0- to 10-ppm active SI DW calibration stan- it is very important to matrix match the background of the sample
dards. The greater concentration of chloride ions present interfere and calibration.
more significantly with the chelating process between the Hyamine Validation of the Matrix-Matching Technique. Having achieved
and the SI. Therefore, the turbidity absorbance recorded for that successful calibrations using the matrix-matching technique for
concentration of SI is incorrect. Matrix matching the sample and ×10, ×5, ×2.5, and ×2 dilutions, the procedure was validated by
calibration standards, for example using the same dilution factor examining the repeatability and robustness of the process. Table 7
from an SW background to a DW background for both, gives presents the average SI concentration of three repeats at each of
equivalent chloride-ion concentration within both. This means these known concentrations using the matrix-matching technique
that any chloride-ion effects on the chelating process between the for different dilutions.
Hyamine and SI can be negated, allowing accurate analysis to be Using this matrix-matching technique allowed samples to be
performed for the low-SI-concentration samples. For example, if a analyzed down to 2 to 3 ppm with confidence (approximately 0.022
sample was thought to require a ×10 dilution to give an SI within absorbance units ≈ 1 ppm on calibration).
8 2 10
R = 0.999 x10 DW
6 8 x5 DW
6 x2.5 DW
4
4 x2 DW
2
2
0
0
0.0 0.2 0.4 0.6 0.8 1.0
Absorbance (500nm) 0.000 0.200 0.400 0.600 0.800 1.000
Absorbance (500nm)
Fig. 4—The successful matrix-matching calibration from 0- to
100-ppm SW standards to give a 0 to 10 ppm active DW range Fig. 5—Comparison of all calibration types—Easy Hyamine and
for a ×10 DW dilution. Matrix-Matching Hyamine at ×10, ×5, ×2.5 and ×2 dilutions.
SI Repeats
Active ppm x10 x5 x2.5 x2
To ensure that repeatable results could be achieved for the same lines, one which examines the lower end of the calibration range (0,
sample but using different matrix-matching calibrations, a 5 ppm 5, 50 ppm active) and the other the high end (0, 50, 500, 2,500 ppm
active SI SW solution was prepared and analyzed by both ×2.5 and active). The results observed show that all the SI species examined
×2 DW calibrations. The results are shown in Table 8. can be detected by ICP, including the new P-tagged copolymer,
By analysing the same 5-ppm active SI SW solution by both over a range of 0 to 2,500 ppm. The same set of standard solutions
×2.5 and ×2 matrix-matched DW calibrations, the achieved con- was used for both method types.
centration was reasonably accurate taking into consideration that This ICP process was repeated for a low-end calibration of 0, 5,
fairly low absorbance numbers (approximately 0.060 equivalent and 10 ppm active to compare the level of quantification between
to 2.5 or 2 ppm on the ×2 or ×2.5 DW calibration) were being the P-tagged copolymer and the PPCA. This is important because
recorded. Hence, this work provides an initial evaluation and vali- very-low-level polymer analysis is now required for returning
dation of a technique that will be very useful for future analysis. squeeze-treatment samples in the field as more-environmentally-
friendly-polymer-type species are being used.
ICP and Wet-Chemical Analysis of a New P-Tagged Copolymer. The results in Table 9 highlight that the detection of the P-tagged
Recently, one of the FAST sponsors supplied the research group copolymer is very similar to the PPCA by ICP at these low concen-
with a new P-tagged polymer that they claimed had sufficient P-tag- trations. Again, both inhibitors can be detected by each ICP method
ging to be easily detectable with ICP but was low enough for it to equally well, with the level of quantification being around 0.5 ppm
be classed as environmentally friendly. In terms of inhibitor assay, active for both SIs, although lower concentrations may be achievable.
FAST’s interest in this material was that it may be one of only a The signal intensities for both methods indicate that the P-
few polymeric products (e.g., PPCA, other P-tagged polymers) that tagged copolymer contains more P content within its structure than
could be detectable by both wet-chemical and ICP techniques. PPCA. Although this does not make a significant difference in the
ICP Detection: Initial examinations in this study investigated actual levels recorded here for each SI in NSSW (Table 9), it may
the ICP detection of this new P-tagged copolymer in NSSW along- make a difference in detecting the P-tagged copolymer compared
side more commonly used SIs such as di-ethylene triamine penta to the PPCA at lower concentrations or in more-severe brines—for
methylene phosphonic acide (DETPMP), poly-phosphino methyl- instance, if the samples were in more-severely-scaling brine or there
ated poly-amine (PMPA), and PPCA that can also be detected by were interferences in the produced water that contained the SI. This
ICP. Two ICP methods were employed—the PHOS method and higher signal may also be helpful if ICP detection cannot be used for
the P POLYMER method. The major differences between the two some reason and wet-chemical techniques need to be adopted.
methods are the wavelengths used and the fact that the PHOS Wet-Chemical Techniques and Analysis of ICP Solutions.
method uses a peak height measurement where as the P POLY- In addition to using ICP, a number of wet-chemical techniques
MER uses a Gaussian peak area. In each method, there are two were investigated to see if they could not only detect the P-tagged
TABLE 9—RESULTS FOR LOW-END CALIBRATIONS, 0–10ppm ACTIVE, FOR THE SIs P-TAGGED COPOLYMER AND PPCA
copolymer but, by using the same ICP standard solutions, could determined concentrations are of the correct magnitude, except at
determine corresponding values for these standards by each wet- 50 and 500 ppm active for the Easy Hyamine. This Easy Hyamine
chemical technique. result at the lower concentrations was expected because the dilu-
Successful calibrations and repeats were achieved for amino- tions required from these concentrations to within the calibration
propyl cartridge Hyamine, Oasis Hyamine, and Easy Hyamine. range mean that there is still an influence of the salt ions, mainly
Again a lower range of calibration improves the low-end accuracy chloride ions, in the solution on the Hyamine process, although
of the repeats approximately 0.5 and 1 ppm. All the procedures for they are being diluted out in DW. This observation relates to the
each wet-chemical technique and any brine compositions used are matrix-matching technique.
detailed in Appendix A. If the wet-chemical analysis for the P-tagged copolymer is then
To verify that the range of wet-chemical techniques will give compared with the corresponding ICP determined values (Fig. 6), it
concentrations similar to those of the ICP analysis, the previously can be said that the PHOS ICP methods are a slightly better match
used ICP solutions and additional independent 1,000 and 5,000 (shown by the linear relationships) with the wet-chemical analysis,
ppm active solutions were analyzed by each of the wet-chemi- while the Oasis technique appears to be rather low in its determina-
cal techniques. The corresponding values determined for these tions at the high values of 2,500 and 5,000 ppm active SI.
P-tagged copolymer ICP solutions by wet-chemical analysis are
highlighted in Table 10. Summary and Conclusions
Between the wet-chemical techniques, there is variation in the A survey of some recent developments in the analysis of SI species
measured concentrations of these ICP solutions, although all of the has been presented with an emphasis on their assay at threshold
3000
Wet Chemical Analysis [SI],
y = 0.9648x + 5.0102
Active ppm
3000
Wet Chemical Analysis [SI],
y = 0.8874x + 11.787
Active ppm
2000 2000 2
2
R = 0.9995 R = 0.9994
1500 1500 y = 0.9389x + 42.454
y = 0.9434x + 40.156
2
2
R = 0.9979 R = 0.9976
1000 1000
0 Oasis Oasis
0
0 1000 2000 3000 Easy 0 1000 2000 3000 Easy
ICP [SI], Active ppm ICP [SI], Active ppm
(c) (d)
Fig. 6—A Comparison of the ICP-determined values with the wet-chemical determination for the P-tagged copolymer in NSSW,
(a) PHOS 177.440 nm, (b) PHOS 214.914 nm, (c) P POLYMER 177.440 nm low, and (d) P POLYMER 177.440 nm high.
19. Calculate the concentration of the repeat tests at their leave for 40 minutes. A time interval of 1 minute is suggested for
known concentrations (i.e., 0, 0.5, 1, 4, and 6 ppm) to determine the addition of Hyamine to each test tube to allow for analysis time
the method repeatability using the previously constructed calibra- on the spectrophotometer.
tion graph. 5. After 40 minutes, pour the solution into a 2-cm cell and
20. Repeat the procedure for samples, and determine the concen- measure the absorbance at 500 nm using a spectrophotometer.
tration of chemical in the solution using their recorded absorbance 6. Construct a calibration graph from the recorded standard
values and the previously constructed calibration graph. Remember solution absorbance values.
to include blanks and repeat standards within the analysis run. 7. Perform repeats at known concentrations (i.e., 2, 5, and 10
In summary, the WAX protocol is in Table A-1. ppm) to determine the method repeatability using the previously
constructed calibration graph.
A3: Oil Analysis by Inductively Coupled Plasma. Table A-2 8. Repeat the procedure for samples and determine the con-
displays the gas settings required to achieve a lit plasma for aque- centration of chemical in the solution using their recorded
ous and oil ICP analysis. absorbance values and the previously constructed calibration
graph. Remember to include blanks and repeat standards within
A4: Easy Hyamine. In this procedure, the interferences from brine the analysis run.
salts are diluted out. The initial stock has to be of a high enough
value that when a dilution in DW is performed from the experiment A5: Brine Compositions. The brine compositions used throughout
SW stock, little interference is observed from the brine salts on this study are given in Table A-3.
the addition of the sodium citrate and Hyamine for the detection
of the SI concentration present. This is checked by ensuring that
a calibration and good repeatability can be achieved for each SI Ken Sorbie is a professor of petroleum engineering at Heriot-Watt
species being evaluated. The advantage of this method is that little University, Edinburgh, Scotland, UK. His research interests include
pretreatment of the sample, other than a dilution in DW, is required. the modeling of multiphase flow in porous media, particularly
Therefore, it is a quick method to use. on the fundamentals of two- and three-phase flow at the pore
If the inhibitor samples to be analyzed are low-salinity brines level and on the effects of wettability. He is also active in the
study of chemical and porous media transport aspects of min-
(e.g., SW) or if the SI stock/samples are of a sufficiently high eral-scale formation and prevention in oil fields. Sorbie holds a BS
concentration, then dialysis is not required and a large dilution in degree in chemistry from Strathclyde University, Scotland, and a
DW is sufficient to reduce the salinity. PhD degree in theoretical chemistry/applied mathematics from
The analytical procedure is therefore as follows: Sussex University. Sorbie received the SCA 2004 Distinguished
1. Dilute the inhibitor stock/sample that is in SW or FW brine Achievement Award and the SPE IOR Pioneer Award in 2008.
to a concentration that is appropriate for the test type and the fol- Lorraine Boak is a research fellow and also laboratory manager
lowing analysis (i.e., 25 ppm or 100 ppm active, in DW). with the Flow Assurance and Scale Team (FAST) at The Institute
2. From this lower DW stock, make 5-mL standards at con- of Petroleum Engineering, Heriot-Watt University, Edinburgh,
centrations 0, 0.5, 1, 2, 3, 4, 5, 6, 8, and 10 ppm active in DW, in Scotland. She joined the FAST group in 1993 as a research tech-
nician. Her research examines the mechanisms/kinetics of sur-
order to obtain a calibration curve. face nucleation and growth of barium sulphate, the effects of
3. Add 1 mL of 5% sodium citrate solution to each standard desulphation, and analytical techniques. Boak has presented a
solution. number of papers/posters at both international and national
4. Add 1 mL of 5,000 ppm (as supplied) Hyamine 1622 solution conferences. She holds a BSc degree in chemistry and a MPhil
to each standard solution. Stopper the test tube, shake, and then degree by research from Heriot-Watt University.