You are on page 1of 7

Systems Support Center

8080 Tristar Drive, Suite 120


Irving TX, 75063 469-713-5300

EMI Measurements, Test Receiver vs. Spectrum Analyzer

Introduction
Spectrum Analyzer tests set the
EMI measurements require a following parameters:
different approach than other types • Start / Stop Frequency
of general RF tests. In EMI, • Resolution Bandwidth Filter
engineers never quite know what and type (3 or 6 dB)
signals may be present. Since each • Detector(s)
new DUT (device under test) will be • Sweep Time
different, having the correct tools for • Video Bandwidth
characterization of the EMI signals is
of key importance. Also, knowing Most spectrum analyzers “couple” or
the strengths and weaknesses of the lock the resolution bandwidth, video
tools being used is vital. Two bandwidth and sweep time together
instruments are used for EMI testing, so that critical timing requirements
spectrum analyzers and test are not violated. These parameters
receivers. Each requires a different can be manually controlled however,
approach to the test, and each has and even forced into modes where
advantages and disadvantages. incorrect measurements are
This paper will explore and compare displayed.
both instruments.
Test Receivers follow a similar
Background setup:
• Start / Stop Frequency
Spectrum Analyzers and Test • Resolution Bandwidth Filter
Receivers are often lumped together and type (3 or 6 dB)
as the same thing. Although both
• Detector(s)
instruments measure the amplitude
• Measurement (dwell) Time
of signals in the frequency domain,
the units are not the same, and • Step size
require knowledge to extract correct
measurements. Each can be These settings are often listed in the
abused, and each can return specification or standard being used
incorrect measurement results. to qualify the product. Using a
“compliant” test receiver and
The differences between the two can calibrated test environment set
be explored by examining a common according to the standard, should
test setup (i.e. the parameters yield correct results.
necessary to configure each
instrument).
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

Spectrum Analyzers Explored its testing methods. This is


analogous to using a flat head
If a compliant EMI test receiver, screwdriver to drive a Philips type
properly configured, will yield correct head. It can be done, but there are
results, why use any other tool? “issues” to keep in mind.
There are several reasons to
consider using a spectrum analyzer Issue #1: Sub-ranging the span
for EMI type measurements. Most spectrum analyzers make 500
or 1000 measurements within the
Versatility is the main advantage of a span (stop frequency – start
spectrum analyzer. The same unit frequency = span). If that span is set
can be used for general transmitter wide, say from 30 MHz to 1 GHz like
and receiver measurements such as the typical CISPR test, the
adjacent channel power or noise measurement points will be made
figure. Most labs have one in house nearly 2 MHz apart. Obviously,
already, so it makes financial sense these settings will skip over
to use the same unit for EMI. emissions since the resolution is too
course for EMI applications. Sub-
Familiarity is another advantage of ranging the span into smaller chunks
spectrum analyzers. Most engineers allows the measurement points to fall
have been exposed to them, even if closer together, thus improving the
they are not RF specialized. resolution of the test. Sub-ranging
the span is an attempt to make the
There may have been a Speed spectrum analyzer mimic a test
advantage in the past if the operator receiver’s method. Since our goal is
knows EMI and the instrument to improve the resolution, if sub-
limitations. Newer test receivers ranging the span, use the following
erase this advantage in most cases. rule of thumb:

And finally, spectrum analyzers # measurement points = xxxx


“tailored” for EMI are the ½ RBW = max step size
Comfortable solution in the US step size X # points = subrange size
because that is what most engineers
are used to. For instance, CISPR specifies a 120
KHz resolution bandwidth filter for
Spectrum Analyzer for EMI the radiated emissions range (30
MHz – 1 GHz). A proper sub-range
Spectrum Analyzers are a versatile for testing this band will make the
tool, but one that must be used with measurement points fall a maximum
caution for EMI applications. of 60 KHz apart (½ RBW). Using a
Operators must be aware that for spectrum analyzer with 500
EMI, the spectrum analyzer is measurement points, and running
required to mimic a test receiver and the calculation yields:
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

purchasing instruments, since this is


500 points a key area.
120 KHz / 2 = 60 KHz
60 kHz x 500 points = 30 MHz span Issue #3: Resolution Bandwidths
General purpose spectrum analyzers
In this case, a proper test method is may not have the proper resolution
to test the first 30 MHz span (30 – 60 bandwidth filters for EMI
MHz), then the next (60-120 MHz), applications. Often “personalities”
and so on until you arrive at 1 GHz. for the EMI bandwidths must be
This process is highly time purchased as an option. EMI type
consuming. If your spectrum RBW filters are narrower, rolling off
analyzer tests 1000 points, your at 6 dB per decade rather than 3 dB.
subranges will be 60 MHz rather Common CISPR RBW filters are 200
than 30 MHz wide. Also note that Hz, 9 KHz, 120 KHz and 1 MHz.
some devices may require a smaller MIL-STD 461 calls out 6 dB RBW
resolution than ½ RBW (e.g. 1/3 filters between 10 Hz – 10 MHz.
RBW), requiring even more Using the incorrect filter can
subranges. influence both amplitude and
frequency values returned by the
Issue #2: QP and AVE detectors instrument. Very few spectrum
Most EMI applications specify two analyzers have the 6dB filters for
types of detectors that may not be MIL-STD 461, so if you test to MIL-
present on generic spectrum STD, check for the correct
analyzer units. The first is called the bandwidths.
Quasi-Peak or QP detector. This
detector attempts to quantify the Issue #4: Power vs Fieldstrength
“annoyance factor” of a signal by General spectrum analyzers
weighing its repetition rate in addition measure either power or voltage on
to its frequency and amplitude. The the amplitude axis. Most EMI
QP detector must also conform to standards and limit lines are
the CISPR requirements. Pre- specified in fieldstrength at some
compliant QP detectors are not distance from the DUT. In order to
acceptable for those who will “self- arrive at a measurement that is
certify” compliance, or for test labs comparable to a limit line, and one
offering compliance services. The that eliminates the differences
second detector to look for is called between test environments, the
a CISPR-AVE detector. The spectrum analyzer must allow for
differences between a typical AVE “correction” of system transducers.
detector and a CISPR AVE detector When measuring radiated emissions,
have to do with the averaging time the effect of each particular antenna
constants used, which is beyond the (antenna factor) and cable (cable
scope of this paper. Be watchful for loss) must be normalized through the
both of these detectors when use of files that give a correction
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

value at some frequency. When even though you may not be looking
transducer correction is active, the Y at a signal on the screen, the RF
axis of the EMI spectrum analyzer front end is still measuring the total
will read dBuV / meter (fieldstrength) signal power over the entire
rather than voltage or power. Since frequency spectrum. Now, consider
limit lines for radiated emissions are the scenario when a wide band
specified in fieldstrength, a limit line signal is of sufficient amplitude to
can now be displayed on the screen cause an RF overload condition -
for direct pass / fail evaluations of usually seen on the SA as spectral
the measurement results. lines on the display. These “signals”
Conducted emissions are specified will appear as though they come
in dBuV, or a pure voltage, but a from the DUT, when in fact they are
correction value for cables and the induced by the overload condition.
tranducer (usually a line impedance An additional filter can be placed in
stabilization network or LISN) is still the RF section to help prevent this
used to eliminate the differences condition. Called a preselector, this
between various equipment and filter cuts out any signals not being
manufacturers. Insist on an displayed on the screen (with some
instrument that has this capability. limitations). This is critical for
conducted EMI measurements when
Issue #5: Measurement dwell time a very high amplitude signal may
The QP, average and CISPR-AVE prevent measurements on the rest of
detectors have dwell time the spectrum. What about using the
requirements in order to output RF attenuator? That may help, but
correct results. For the QP, a one in some cases it may attenuate
second dwell time is required in wanted signals so much that they
order to fully charge and discharge can no longer be displayed. Plus,
the filters in the detector. Both setting the attenuation is a manual
average detectors require 100 ms or process on spectrum analyzers, one
more. Some spectrum analyzers more thing to complicate the testing
now offer “zero-span”. This process. Preselector filters are
capability is another “work around” to seldom found on spectrum
make the spectrum analyzer work analyzers, but are a valuable option
like a test receiver and provide this if available. Automatic attenuators
dwell feature. can’t work fast enough for a
spectrum analyzer. Manual RF
Issue #6: Preselection & overloads attenuators are available but they
Some of the signals encountered in require constant adjustment during
EMI testing are high amplitude and the test process and add room for
can cover a wide frequency band. inconsistency.
Spectrum analyzers measure
everything that falls in the pass band
of the RF front end. In other words,
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

In summary, spectrum analyzers can


be used for EMI, but various work- Frequency start: 30 MHz
around methods must be used in Frequency stop = 1 GHz
order for the results to be accurate. RBW = 120 kHz (6 dB)
Obviously, there are many Step size = 60 KHz (1/2 RBW)
opportunities for error. Incorrect Activate transducer set
sub-ranging, violation of dwell time Overlay limit line
requirements, use of incorrect RBW
filters, overloads, and many other As you can see, these parameters
issues force us to conclude that the are called out in the standard,
spectrum analyzer may not be the entered directly as a test parameter,
best tool for the job. Look at the and now the operator needs only to
tedious nature of sub-ranging alone. press start. No sub-ranging, no
Even a high-end spectrum analyzer zero-spanning, no mental
with 1000 points forces a 60 MHz conversions from voltage to
sub-range. Seventeen ranges must fieldstrength. Furthermore, with
be tested between 30 MHz and 1 automatic attenuators and
GHz. The results of each range preselection, it is possible for a
must be collated, corrected for compliant test to run without
transducer values, checked for intervention. There are obviously
overload conditions, and finally reasons for an operator to interrupt
compared to a limit line for a pass / the test, but the possibility does exist
fail decision. Software helps, but it is for a fully automatic test to yield
one more complexity and one more accurate results.
cost. Also, software may be tied to
only one spectrum analyzer or In order to explore test receivers a
vendor. Fortunately, there is a better bit more, some key features are
way. called out in the next section that
illustrate why a test receiver is a
better tool for EMI testing.
Test Receivers Explored
Feature #1: Measurement Points
CISPR, EN, FCC, MIL-STD and all A scanning receiver will measure
the other standards bodies many tens of thousands of points if
recommend using a test receiver for needed. For instance, in the
EMI applications for all of the above previous test setup, the test receiver
reasons. With a test receiver, what will measure:
you set is what is measured, and
these settings are listed right in the 1000 MHz - 30 MHz = 970 MHz
test standard. A typical test receiver 970 Mhz / 60 kHz = 16,170 pts
setup illustrates the point. Using the
same basic case used before, the This level of accuracy is much
keystokes are: improved over a spectrum analyzer’s
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

500 or 1000 points. Tests with an expert, EMI novices may be


100,000 measurement points are not saved from collecting incorrect data,
uncommon in some MIL-STD and worse, making judgments on
applications. bad data.

Feature #2: Tune & Dwell Test Receiver disadvantages


The second key feature is the tune
and dwell capability native to a Similar to spectrum analyzers, there
scanning receiver. Dwell time is set are also disadvantages to test
as a keystroke parameter, and receivers. Scanning receivers are
violations to dwell time requirements the right tool for the job mainly
of the various detectors are often because they were built with EMI in
“flagged” as incorrect. mind. Specialized tools often cost
more, and have limited use outside
Feature #3: EMI Specificity of the area of specialization.
This key feature rolls up all the
“issues” discussed during the Many proponents of spectrum
spectrum analyzer exploration, analyzers will suggest that receivers
namely the availability of EMI are slower than analyzers. On the
specific features such as 6 dB RBW face, this is true, but if apples are
filters, transducer sets, limit lines and compared to apples, there is no time
preselector filters. penalty associated with using a
receiver. The type of detector and
Feature #4: Automatic Control its associated dwell time impose the
Scanning receivers can also take true timing limitation. Also the
advantage of automatic control of number of measurements that must
certain parameters. Remember at be made to ensure proper signal
the beginning when it was stated that capture will be the same, regardless
certain modes could be forced on a of which unit is being used. A true
spectrum analyzer that will yield side-by-side comparison of a
incorrect results. This is also true for receiver scan and the sub-ranged
receivers, but much less likely since spectrum analyzer sweep will show a
the receiver will “take care of itself” if time advantage to the receiver, not
left in auto-control mode. Automatic the spectrum analyzer.
control of RF attenuation,
preselection filtering, preamplification Summary
settings, RBW settings, and the step
size is common. Auto-control of In the end both tools may be used
these settings can keep the user out for EMI measurements. There may
of trouble in most cases. The auto- be overriding factors that guide your
control of these parameters will choice that were not discussed. One
follow what is called out in the factor might be the requirements
standards. Since no one starts as imposed by an accreditation body
Systems Support Center
8080 Tristar Drive, Suite 120
Irving TX, 75063 469-713-5300

like A2LA or NVLAPP. These bodies


and military / government agencies
may disallow generic spectrum
analyzer measurements for EMI.
The number of “work-arounds” is
high, and the room for error is great.
Before purchasing, check for these
factors first.

Look also into the expertise level of


the engineers and technicians that
will be conducting these tests. What
if the measurements are wrong?
Will this force a board redesign,
device recall, or rocket launch delay?
The money value of incorrect
measurement will almost always
overcome any additional money
spent acquiring the best tool for the
job.

And if all else fails, look for an


instrument that has both, spectrum
analyzer and EMI test receiver.
These units cost less than buying
both units separately, will give
flexibility to test in both modes, and
can be used for many purposes
other than EMI. There are some out
there, and most of them can be
found at www.rohde-schwarz.com.

You might also like