Laboratory Department

Quantity Equipment Voltage for Volt-Ampere Amps Frequency


AC or DC Or Watts

1 Automated Enzyme Immunoassay 100-240 V 250VA 50/60HZ


System
1 Easylyte Ion Selective Electrode 100/115 V AC 0.8A/250 V 0.8A 50/60HZ
Analyzer
1 Vitalab Selectra E Chemistry Analyzer 110/240 V AC 320VA 50/60HZ
1 Cell Dyn 1800 Analyzer 100/120 VAC MAX POWER 50/60HZ
220/240 VAC 900 Watts
1 Clay Adams Centrifuge 115 V 1.5A 60 HZ

1 MDC Stat Dry 125 V AC 150 Watts 60 HZ

1 No. 100 Incubator 120 V AC 60 Watts 50/60HZ

1 Olympus Cx21 Microscope 100-120 V AC 0.42/0.25 A 50/60HZ


220-240 V AC
1 Stat Fax 303/plus 110-120 V AC 50 Watts 50/60HZ
220-240 V AC
1 Vortex Mixer 110 V AC 2.0 A 60HZ

1 Vitalab Scientific Fibron 1 100-240 V AC 55 VA 50/60HZ

1 Acros Refrigerator +-10% 127 V AC 2.2-2.5A 60HZ

1 Laser Jet Pro MFP M127 fn 110-127 V AC 3.5A 50/60HZ

1 Dell Selectra Pc 115/230 V AC 6.0/3.0A 50/60HZ


1 Avaya Desk top Phone 14 w
1 Dell XPS PC [You can do with 1 PC up] 100-240 V AC 8A/4A 50/60HZ

1 Acer Pc 100-127 V AC 6/3.15A 50/60HZ


220-240 V AC
1 Samsung Selectra Monitor 14 V DC 1.43A

1 ROC Monitor 100-240 V AC 1.5A 50/60HZ

1 Dell Monitor 100-240 V AC 0.6A 50/60HZ

1 LG Gold AC Unit 10,000 BTU 220 V AC 1000 W

Pharmacy Department
Quantity Equipment Voltage for Volt-Ampere Amps Frequency
AC or DC Or Watts

1 Dell Optiplex 3010 PC 90-264 V AC 250 w 5.0/2.5 A 47/53 Hz


1 Hp Lazerjet 2010 110 – 127 V AC 3.5 A 50/60 Hz.
1 Dymo Label Writer 450 Turbo 100 – 240 V AC 1.6 A 50/60 Hz.
1 POS Thermal Receipt Printer 12 V DC 2.6 A
1 Samsung SyncMaster Monitor 14 V DC 1.43 A
S19B150
1 Avaya Desk top Phone 14 w

1 Kenmore Mini Refrigerator 120 V AC 0.9 A 60 Hz


Reception & Cashier Department
Quantity Equipment Voltage for Volt-Ampere Amps Frequency
AC or DC Or Watts

2 Dell Insprion 20 100 -240 V AC 1.6 A

1 Avaya Desk top Phone 14 w

1 HP Lazer Jet Color MFP M277dw 110-127 V AC 4.5A 50/60Hz


1 POS Thermal Receipt Printer 12 V DC 2.6 A

2 Cash Drawer 12 V DC 2.6 A


1 Samsung LED 32 inch TV 110 – 120 V AC 28 W 60 Hz
1 Sanky Upright Show Case 120 V AC 2.9 A 60 Hz
1 Honey well Water Cooler 485 W 4.7 A
Clinical Consultation Department
Quantity Equipment Voltage for Volt-Ampere Amps Frequency
AC or DC Or Watts

1 Vasculight Elite Lumenis 120 V AC 20 A 50/60 Hz


1 Ritter Examination Lamp 120 V AC 30 A 60 Hz
1 Ritter Examination Bed 100 115 V AC 5A 60 Hz
1 Ritter Examination Bed 104 115 V AC 5A 60 Hz
1 Ultra baby Scale 9-12 V AC 100 mA
1 Ritter Examination Bed 100 115 V AC 5A 60 Hz
1 Stryker Cast Cutter 100 – 120 V AC 1.5 A 50/60 Hz
1 X-ray film Illuminator 110 V AC 1.40 A 60 Hz
1 Ritter Examination Bed 104 115 V AC 5A 60 Hz
1 Hamilton Examination Bed 115 V AC 10 A 50/60 Hz
1 LG AC Unit 10,000 BTU 220 V AC 1,800 W

Imaging – Diagnostics & General Ward

Quantity Equipment Voltage for Volt-Ampere Amps Frequency


AC or DC Or Watts

1 ECG – 12 C 100 – 240 V AC 0.9 – 0.4 A 50/60 Hz


1 Voluson 730 Pro USG 115-120 V AC 920 VA 50/60 Hz
1 Sharp TV 110-220 V AC 62 W 50/60 Hz
1 Patient Hospital Bed 115 V AC 1–2A 60 Hz
1 LG AC Unit 10,000 BTU 220 V AC 1,000 W
1 Sharp TV 120 V AC 53 W 60 Hz
1 Patient Hospital Bed 115 V AC 1–2A 60 Hz
1 LG AC Unit 12,000 BTU 220 V AC 2,200 W
Observation Unit
1 Patient Hospital Bed 115 V AC 1–2A 60 Hz
1 LG AC Unit 12,000 BTU 220 V AC 2,200 W
1 Microwave 120 V AC 1050 W 60 Hz
1 Radio 110 – 120 V AC 50/60 Hz
Normal Delivery
1 LG AC Unit 12,000 BTU 220 V AC 2200 W 60 Hz
1 Welch Allyn Procedure Light 120 V AC 0.7 A 60 Hz
1 Arron Electro Surgical Catheter 100 – 240 V AC 4A 50/60 Hz
1250
[Maybe this belongs in OT. In OT
it does need power all times]
1 Ritter 307 Examination Bed 115 V AC 5A 60 Hz
1 Cadance Fetal Cardiac Monitor 100 – 240 V AC 60 VA 50Hz/60Hz
Operating Theatre & Equipment Storage room
1 Incubator Isollete VH568-100 120 V AC 635 W 60 Hz
Series Air Shield
1 Suction Pump 115 V AC 0.68 A 60 Hz
1 Thoracic Drainage 115 – 230 V AC 50/60 Hz
2 Amsco SQ 240 Surgical Light 100 V AC 4 A Single & 50/60 Hz
8 A Dual
Controller
220-250 V AC 2 A Single &
4A
Dual
Controller
Equipment in Storage room [Highlighted are equipment used during a surgery]
2 Amsco Surgical Table 3080 100 -240 V AC 5 – 2.5 A 50/60 Hz
1 Narkomed B2 Anaesthesia Unit 90-130 V AC 5A 50/60 Hz
200-260 V AC
1 Spacelabs Anaesthesia Patient 115 V AC 2A 50/60 Hz
Vital signs Monitor
1 Ohmeda Ventilator 100-120 V AC 0.5 A 50/60 Hz
220-240 V AC 0.25 A
1 Ohmeda Eletrical Box 6 outlets 100-120 V AC 10 A 50/60 Hz
for Ventilator and Monitor Use
1 Captomac Ultima *Spelled 115 V AC 100 VA 870 mA 50/60 Hz
Capnomac Ultima
1 Electrical Pressure Steam 120 V AC 8.75 A 50/60 Hz
Sterilizer [This is a sterilizing
machine. It is never in the OT it
is for strorage. It sterilizes
equipment] technically it is used
at the end of the day.
1 Video Copier Mitsubishi 220-240 V AC 1.0 A 50 Hz
1 Ultrasound Equipment Toshiba 230 V AC 4.5 A 50/60 Hz
1 X-Ray film illuminator 118 V AC 2.1 A 60 Hz
1 Amsco 2013 Eagle Steam/Vac Controls: 120 V AC 2.5 A 1-phase 50/60 hz
Sterilizer Generator Heaters:
208 V AC 83.2 A 3-phase 50/60 hz

440/480 V AC 31/37A 3-phase 50/60 hz

240 V AC 72.2A 3-phase 50/60 hz

1 Seimens Mobilette 11 Elema AB 96 kV 30 kW 10 ms

102 kV 20 kW 100 ms

450mA
(maximum)
Lighting Fixtures for Entire Compound
74 Single light Bulbs 110-125 V AC 3W 50/60 hz

13 Large Single florescent bulbs 110-125 V AC 20 W 50/60 hz

8 Large Double Florescent bulbs 110-125 V AC 20 W 50/60 hz

Ceiling and Standing Fans


10 Ceiling Fans

4 Standing Lasco Fans


X-ray Equipment
1 400 – 480 V AC 50-60 hz
2007 SIEMENS AXIOM ARISTOS
MX-NISR8409
DIGITAL SYSTEM 3 Phase
DUAL DETECTOR
2016  TUBE  OPTI TOP  150-40-
80HC-100
3D -OVERHEAD TUBE CRANE
AXIOM TABLE High-
TOMO frequency, 50
Recommended High- or 80 kW
units (X-RAY frequency, 50 (POLYDOROS
GENERATORS) or 80 kW LX)
(POLYDOROS
LX)

NOTE FOR OT AND EQUIPMENT STORAGE. IN THEATRE THE HIGHLIGHTED IN YELLOW ARE THE EQUIPMENT YOU NEED IN A POWER OUTAGE.
THE RED HIGHLIGHTED HOWEVER ARE DIFFERENT. USUALLY STERILIZING OCCURS IN A CSU (CENTRAL STERILIZING UNIT) WHICH IS A
COMPLETELY SEPARATE ROOM. LOOKS LIKE THEIR STORAGE ROOM WILL BE THEIR STERILIZING ROOM AS WELL. IN WHICH CASE IT DOES NEED
POWER IN A BLACKOUT.

Special note fro KC

: An operating theatre has an air handling unit that is on 24/7. During day time the standard has to be 20-25 air changes x hour. At night this
goes to a setback mode to 10-12 air changes x hour.