You are on page 1of 7

Designation: E 1316 – 07b

Standard Terminology for


Nondestructive Examinations1
This standard is issued under the fixed designation E 1316; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
INDEX OF TERMS
Section
A: Common NDT Terms
B: Acoustic Emission (AE) Terms

C: Electromagnetic Testing (ET) Terms


D: Gamma- and X-Radiologic Testing (RT) Terms

E: Leak Testing (LT) Terms


F: Liquid Penetrant Testing (PT) Terms

G: Magnetic Particle Testing (MT) Terms


H: Neutron Radiologic Testing (NRT) Terms

I: Ultrasonic Testing (UT) Terms


J: Infrared Testing (IRT) Terms
K: Holographic Testing (HT) Terms
L: Visual Testing (VT) Terms

1. Scope Radiologic Testing (RT) is often used to examine material to


1.1 This standard defines the terminology used in the detect internal discontinuities.)
standards prepared by the E07 Committee on Nondestructive 1.3 Section A defines terms that are common to multiple
Testing. These nondestructive testing (NDT) methods include: NDT methods, whereas, the subsequent sections define terms
acoustic emission, electromagnetic testing, gamma- and pertaining to specific NDT methods.
X-radiology, leak testing, liquid penetrant testing, magnetic 1.4 As shown on the chart below, when nondestructive
particle testing, neutron radiology and gauging, ultrasonic testing produces an indication, the indication is subject to
testing, and other technical methods. interpretation as false, nonrelevant or relevant. If it has been
1.2 Committee E07 recognizes that the terms examination, interpreted as relevant, the necessary subsequent evaluation
testing and inspection are commonly used as synonyms in will result in the decision to accept or reject the material. With
nondestructive testing. For uniformity and consistency in E07 the exception of accept and reject, which retain the meaning
nondestructive testing standards, Committee E07 encourages found in most dictionaries, all the words used in the chart are
the use of the term examination and its derivatives when defined in Section A.
describing the application of nondestructive test methods.
There are, however, appropriate exceptions when the term test
and its derivatives may be used to describe the application of
a nondestructive test, such as measurements which produce a
numeric result (for example, when using the leak testing
method to perform a leak test on a component, or an ultrasonic
measurement of velocity). Additionally, the term test should be
used when referring to the NDT method, that is, Radiologic
Testing (RT), Ultrasonic Testing (UT), and so forth. (Example:

1
This terminology is under the jurisdiction of Committee E07 on Nondestructive
Testing and is the direct responsibility of Subcommittee E07.92 on Editorial
Review.
Current edition approved July 15, 2007. Published July 2007. Originally
approved in 1989. Last previous edition approved in 2007 as E 1316 – 07a.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
E 1316 – 07b
2. Referenced Documents
2.1 ASTM Standards: 2
3. Significance and Use
NOTE 1—This standard defines the terminology used in the standards
prepared by Committtee E07 on Nondestructive Testing and published in 3.1 The terms found in this standard are intended to be used
the Annual Book of ASTM Standards, Volume 03.03. uniformly and consistently in all nondestructive testing stan-
dards. The purpose of this standard is to promote a clear
understanding and interpretation of the NDT standards in
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or which they are used.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. 4. Terminology

Section A: Common NDT Terms


The terms defined in Section A are the direct responsibility of Subcommittee E07.92, Editorial Review.

acceptable quality level—the maximum percent defective or imperfection, n—a departure of a quality characteristic from
the maximum number of units defective per hundred units its intended condition.
that, for the purpose of sampling test, can be considered indication—the response or evidence from a nondestructive
satisfactory as a process average. examination.
calibration, instrument, n—the comparison of an instrument DISCUSSION—An indication is determined by interpretation to be
with, or the adjustment of an instrument to, a known relevant, non-relevant, or false.
reference(s) often traceable to the National Institute of
Standards and Technology (NIST). (See also standardiza- inspection, n—see preferred term examination.
tion, instrument.) interpretation—the determination of whether indications are
cognizant engineering organization—the company, govern- relevant or nonrelevant.
ment agency or other authority responsible for the design, or interpretation, n—the determination of whether indications
end use, of the material or component for which nondestruc- are relevant, nonrelevant, or false.
tive testing is required Nondestructive Evaluation—see Nondestructive Testing.
Nondestructive Examination—see Nondestructive Testing.
DISCUSSION—In addition to design personnel, the cognizant engineer-
Nondestructive Inspection—see Nondestructive Testing.
ing organization could include personnel from engineering, material
and process engineering, stress analysis, nondestructive testing, quality Nondestructive Testing (NDT), n—the development and
assurance and others, as appropriate. application of technical methods to examine materials or
components in ways that do not impair future usefulness and
defect, n—one or more flaws whose aggregate size, shape, serviceability in order to detect, locate, measure and evaluate
orientation, location, or properties do not meet specified flaws; to assess integrity, properties and composition; and to
acceptance criteria and are rejectable. measure geometrical characteristics.
discontinuity, n—a lack of continuity or cohesion; an inten- nonrelevant indication, n—an NDT indication that is caused
tional or unintentional interruption in the physical structure by a condition or type of discontinuity that is not rejectable.
or configuration of a material or component. False indications are non-relevant.
evaluation—determination of whether a relevant indication is reference standard, n—a material or object for which all
cause to accept or to reject a material or component. relevant chemical and physical characteristics are known and
examination, n—a procedure for determining a property (or measurable, used as a comparison for, or standardization of,
properties) or other conditions or characteristics of a material equipment or instruments used for nondestructive testing.
or component by direct or indirect means. (See also standardization, instrument.)
NOTE 2—Examples include utilization of X-rays or ultrasonic waves relevant indication, n—an NDT indication that is caused by a
for the purpose of determining (directly or by calculation) flaw content, condition or type of discontinuity that requires evaluation.
density, or (for ultrasound) modulus; or detection of flaws by induction of standard—(1) a physical reference used as a basis for com-
eddy currents, observing thermal behavior, AE response, or utilization of parison or calibration; (2) a concept that has been established
magnetic particles or liquid penetrants.
by authority, custom, or agreement to serve as a model or
false indication, n—an NDT indication that is interpreted to be rule in the measurement of quality or the establishment of a
caused by a condition other than a discontinuity or imper- practice or procedure.
fection. standardization, instrument, n—the adjustment of an NDT
flaw, n—an imperfection or discontinuity that may be detect- instrument using an appropriate reference standard, to obtain
able by nondestructive testing and is not necessarily reject- or establish a known and reproducible response. (This is
able. usually done prior to an examination, but can be carried out
flaw characterization, n—the process of quantifying the size, anytime there is concern about the examination or instru-
shape, orientation, location, growth, or other properties, of a ment response. (See also calibration, instrument.)
flaw based on NDT response. test, n—see preferred term examination.

2
E 1316 – 07b

Section B: Acoustic Emission

The terms defined in Section B are the direct responsibility of Subcommittee E07.04 on Acoustic Emission Method.

acoustic emission (AE)—the class of phenomena whereby array, n—a group of two or more AE sensors positioned on a
transient elastic waves are generated by the rapid release of structure for the purposes of detecting and locating sources.
energy from localized sources within a material, or the The sources would normally be within the array.
transient waves so generated. Acoustic emission is the arrival time interval (Dtij)—see interval, arrival time.
recommended term for general use. Other terms that have attenuation, n—the decrease in AE amplitude per unit dis-
been used in AE literature include (1) stress wave emission, tance, normally expressed in dB per unit length.
(2) microseismic activity, and (3) emission or acoustic average signal level, n—the rectified, time averaged AE
emission with other qualifying modifiers. logarithmic signal, measured on the AE amplitude logarith-
acoustic emission channel—see channel, acoustic emission. mic scale and reported in dBae units (where 0 dBae refers to
acoustic emission count (emission count) (N)—see count, 1 µV at the preamplifier input).
acoustic emission. burst emission—see emission, burst.
acoustic emission count rate—see count rate, acoustic emis- channel, acoustic emission—an assembly of a sensor, pream-
sion (emission rate or count rate) (Ṅ). plifier or impedance matching transformer, filters secondary
acoustic emission event—see event, acoustic emission. amplifier or other instrumentation as needed, connecting
acoustic emission event energy—see energy, acoustic event. cables, and detector or processor.
acoustic emission sensor—see sensor, acoustic emission. NOTE 3—A channel for examining fiberglass reinforced plastic (FRP)
acoustic emission signal amplitude—see signal amplitude, may utilize more than one sensor with associated electronics. Channels
may be processed independently or in predetermined groups having
acoustic emission. similar sensitivity and frequency characteristics.
acoustic emission signal (emission signal)—see signal, acous-
tic emission. continuous emission—see emission, continuous.
acoustic emission signature (signature)—see signature, count, acoustic emission (emission count) (N)—the number
acoustic emission. of times the acoustic emission signal exceeds a preset
acoustic emission transducer—see sensor, acoustic emission. threshold during any selected portion of a test.
acoustic emission waveguide—see waveguide, acoustic emis- count, event (Ne)—the number obtained by counting each
sion. discerned acoustic emission event once.
count rate, acoustic emission (emission rate or count rate)
acousto-ultrasonics (AU)—a nondestructive examination
(Ṅ)—the time rate at which emission counts occur.
method that uses induced stress waves to detect and assess
diffuse defect states, damage conditions, and variations of count, ring-down—see count, acoustic emission, the preferred
mechanical properties of a test structure. The AU method term.
combines aspects of acoustic emission (AE) signal analysis couplant—a material used at the structure-to-sensor interface
with ultrasonic materials characterization techniques. to improve the transmission of acoustic energy across the
interface during acoustic emission monitoring.
adaptive location—source location by iterative use of simu-
cumulative (acoustic emission) amplitude distribution F(V)—
lated sources in combination with computed location.
see distribution, amplitude, cumulative.
AE activity, n—the presence of acoustic emission during a cumulative (acoustic emission) threshold crossing distribution
test. Ft(V)—see distribution, threshold crossing, cumulative.
AE amplitude—see dBAE. dBAE—a logarithmic measure of acoustic emission signal
AE rms, n—the rectified, time averaged AE signal, measured amplitude, referenced to 1 µV at the sensor, before amplifi-
on a linear scale and reported in volts. cation.
AE signal duration—the time between AE signal start and AE Signal peak amplitude ~dBAE! 5 ~dB1µV at sensor! 5 20 log10~A1/A0!
signal end. (1)
AE signal end—the recognized termination of an AE signal,
usually defined as the last crossing of the threshold by that
signal. where:
A0 = 1 µV at the sensor (before amplification), and
AE signal generator—a device which can repeatedly induce a A1 = peak voltage of the measured acoustic emission signal
specified transient signal into an AE instrument. (also before amplification).
AE signal rise time—the time between AE signal start and the
peak amplitude of that AE signal. Acoustic Emission Reference Scale:
AE signal start—the beginning of an AE signal as recognized dBAE Value Voltage at Sensor
by the system processor, usually defined by an amplitude 0 1 µV
excursion exceeding threshold. 20 10 µV

3
E 1316 – 07b
40 100 µV overload level and the minimum signal level (usually fixed
60 1 mV by one or more of the noise levels, low-level distortion,
80 10 mV
100 100 mV
interference, or resolution level) in a system or sensor.
effective velocity, n—velocity calculated on the basis of arrival
DISCUSSION—In the case of sensors with integral preamplifiers, the
times and propagation distances determined by artificial AE
AO reference is before internal amplification.
generation; used for computed location.
dead time—any interval during data acquisition when the emission, burst—a qualitative description of the discrete
instrument or system is unable to accept new data for any signal related to an individual emission event occurring
reason. within the material.
differential (acoustic emission) amplitude distribution F(V)—
NOTE 4—Use of the term burst emission is recommended only for
see distribution, differential (acoustic emission) ampli- describing the qualitative appearance of emission signals. Fig. 1 shows an
tude f(V). oscilloscope trace of burst emission signals on a background of continuous
differential (acoustic emission) threshold crossing distribution emission.
ft(V)—see distribution, differential (acoustic emission)
emission, continuous—a qualitative description of the sus-
threshold crossing.
tained signal level produced by rapidly occurring acoustic
distribution, amplitude, cumulative (acoustic emission)
emission from structural sources, leaks, or both.
F(V)—the number of acoustic emission events with signals
that exceed an arbitrary amplitude as a function of amplitude NOTE 5— Use of the term continuous emission is recommended only
V. for describing the qualitative appearance of emission signals. Fig. 2 and
distribution, threshold crossing, cumulative (acoustic emis- Fig. 3 show oscilloscope traces of continuous emission signals at two
different sweep rates.
sion) Ft (V)—the number of times the acoustic emission
signal exceeds an arbitrary threshold as a function of the energy, acoustic emission event—the total elastic energy
threshold voltage (V). released by an emission event.
distribution, differential (acoustic emission) amplitude energy, acoustic emission signal—the energy contained in a
f(V)—the number of acoustic emission events with signal detected acoustic emission burst signal, with units usually
amplitudes between amplitudes of V and V + DV as a reported in joules and values which can be expressed in
function of the amplitude V. f(V) is the absolute value of the logarithmic form (dB, decibels).
derivative of the cumulative amplitude distribution F(V). evaluation threshold—a threshold value used for analysis of
distribution, differential (acoustic emission) threshold the examination data. Data may be recorded with a system
crossing ft (V)—the number of times the acoustic emission examination threshold lower than the evaluation threshold.
signal waveform has a peak between thresholds V and V + D For analysis purposes, dependence of measured data on the
V as a function of the threshold V. ft(V) is the absolute value system examination threshold must be taken into consider-
of the derivative of the cumulative threshold crossing ation.
distribution Ft (V). event, acoustic emission (emission event)—a local material
distribution, logarithmic (acoustic emission) amplitude change giving rise to acoustic emission.
g(V)—the number of acoustic emission events with signal event count (Ne)—see count, event.
amplitudes between V and aV (where a is a constant event count rate (Ṅe)—see rate, event count.
multiplier) as a function of the amplitude. This is a variant of examination area—that portion of a structure being monitored
the differential amplitude distribution, appropriate for loga- with acoustic emission.
rithmically windowed data. examination region—that portion of the test article evaluated
dynamic range—the difference, in decibels, between the using acoustic emission technology.

FIG. 1 Burst Emission on a Continuous Emission Background. (Sweep Rate—5 ms/cm.)

4
E 1316 – 07b

FIG. 2 Continuous Emission. (Sweep Rate—5 ms/cm.)

FIG. 3 Continuous Emission. (Sweep Rate—0.1 ms/cm.)

Felicity effect—the presence of acoustic emission, detectable instrumentation dead time—see dead time, instrumenta-
at a fixed predetermined sensitivity level at stress levels tion.
below those previously applied. interval, arrival time (Dtij)—the time interval between the
Felicity effect—the presence of detectable acoustic emission at detected arrivals of an acoustic emission wave at the ith and
a fixed predetermined sensitivity level at stress levels below jth sensors of a sensor array.
those previously applied. Kaiser effect—the absence of detectable acoustic emission at
Felicity ratio—the ratio of the stress at which the Felicity a fixed sensitivity level, until previously applied stress levels
effect occurs to the previously applied maximum stress. are exceeded.
NOTE 6—The fixed sensitivity level will usually be the same as was location accuracy, n—a value determined by comparison of
used for the previous loading or test. the actual position of an AE source (or simulated AE source)
first hit location—a zone location method defined by which a to the computed location.
channel among a group of channels first detects the signal. location, cluster, n—a location technique based upon a speci-
floating threshold—any threshold with amplitude established fied amount of AE activity located within a specified length
by a time average measure of the input signal. or area, for example: 5 events within 12 linear inches or 12
hit—the detection and measurement of an AE signal on a square inches.
channel. location, computed, n—a source location method based on

5
E 1316 – 07b
algorithmic analysis of the difference in arrival times among at the processing speed before the system must interrupt data
sensors. collection to clear buffers or otherwise prepare for accepting
additional data.
NOTE 7—Several approaches to computed location are used, including
linear location, planar location, three dimensional location, and adaptive processing speed—the sustained rate (hits/s), as a function of
location. the parameter set and number of active channels, at which
(a) linear location, n—one dimensional source location requiring two or AE signals can be continuously processed by a system
more channels. without interruption for data transport.
(b) planar location, n—two dimensional source location requiring three
rate, event count (Ṅe)—the time rate of the event count.
or more channels.
(c) 3D location, n— three dimensional source location requiring five or rearm delay time—see time, rearm delay.
more channels. ring-down count—see count, acoustic emission, the pre-
(d) adaptive location, n—source location by iterative use of simulated ferred term.
sources in combination with computed location.
sensor, acoustic emission—a detection device, generally pi-
location, continuous AE signal, n—a method of location ezoelectric, that transforms the particle motion produced by
based on continuous AE signals, as opposed to hit or an elastic wave into an electrical signal.
difference in arrival time location methods. signal, acoustic emission (emission signal)—an electrical
signal obtained by detection of one or more acoustic
NOTE 8—This type of location is commonly used in leak location due
to the presence of continuous emission. Some common types of continu-
emission events.
ous signal location methods include signal attenuation and correlation signal amplitude, acoustic emission—the peak voltage of the
analysis methods. largest excursion attained by the signal waveform from an
(a) signal attenuation-based source location, n—a source location emission event.
method that relies on the attenuation versus distance phenomenon of AE signal overload level—that level above which operation
signals. By monitoring the AE signal magnitudes of the continuous signal
at various points along the object, the source can be determined based on
ceases to be satisfactory as a result of signal distortion,
the highest magnitude or by interpolation or extrapolation of multiple overheating, or damage.
readings. signal overload point—the maximum input signal amplitude
(b) correlation-based source location, n—a source location method that at which the ratio of output to input is observed to remain
compares the changing AE signal levels (usually waveform based ampli- within a prescribed linear operating range.
tude analysis) at two or more points surrounding the source and deter-
mines the time displacement of these signals. The time displacement data
signal strength—the measured area of the rectified AE signal
can be used with conventional hit based location techniques to arrive at a with units proportional to volt-sec.
solution for the source site.
DISCUSSION—The proportionality constant is specified by the AE
location, source, n—any of several methods of evaluating AE instrument manufacturer.
data to determine the position on the structure from which signature, acoustic emission (signature)—a characteristic set
the AE originated. Several approaches to source location are of reproducible attributes of acoustic emission signals asso-
used, including zone location, computed location, and con- ciated with a specific test article as observed with a particular
tinuous location. instrumentation system under specified test conditions.
location, zone, n—any of several techniques for determining
stimulation—the application of a stimulus such as force,
the general region of an acoustic emission source (for
pressure, heat, and so forth, to a test article to cause
example, total AE counts, energy, hits, and so forth).
activation of acoustic emission sources.
NOTE 9—Several approaches to zone location are used, including system examination threshold—the electronic instrument
independent channel zone location, first hit zone location, and arrival threshold (see evaluation threshold) which data will be
sequence zone location.
detected.
(a) independent channel zone location, n—a zone location technique
that compares the gross amount of activity from each channel. transducers, acoustic emission—see sensor, acoustic emis-
(b) first-hit zone location, n—a zone location technique that compares sion.
only activity from the channel first detecting the AE event. verification, AE system (performance check, AE system)—
(c) arrival sequence zone location, n—a zone location technique that the process of testing an AE system to assure conformance to
compares the order of arrival among sensors.
a specified level of performance or measurement accuracy.
logarithmic (acoustic emission) amplitude distribution g(V)— (This is usually carried out prior to, during and/or after an
see distribution, logarithmic (acoustic emission) ampli- AE examination with the AE system connected to the
tude. examination object, using a simulated or artificial acoustic
overload recovery time—an interval of nonlinear operation of emission source.)
an instrument caused by a signal with amplitude in excess of voltage threshold—a voltage level on an electronic compara-
the instrument’s linear operating range. tor such that signals with amplitudes larger than this level
performance check, AE system—see verification, AE system. will be recognized. The voltage threshold may be user
pressure, design—pressure used in design to determine the adjustable, fixed, or automatic floating.
required minimum thickness and minimum mechanical waveguide, acoustic emission—a device that couples elastic
properties. energy from a structure or other test object to a remotely
processing capacity—the number of hits that can be processed mounted sensor during AE monitoring. An example of an

6
E 1316 – 07b
acoustic emission waveguide would be a solid wire of rod
that is coupled at one end to a monitored structure, and to a
sensor at the other end.

Section C: Electromagnetic Testing (ET) Terms


The terms defined in Section C are the direct responsibility of Subcommittee E07.07 on Electromagnetic Methods.

absolute coil—a coil (or coils) that respond(s) to the total comparative readout—the signal output of comparator coils.
detected electric or magnetic properties, or both, of a part or (See also comparator coils.)
section of the part without comparison to another section of comparative system—a system that uses coil assemblies and
the part or to another part. associated electronics to detect any electric or magnetic
absolute measurements—measurements made without a di- condition, or both, that is not common to the specimen and
rect reference using an absolute coil in contrast to differential the standard (see comparator coils).
and comparative measurements. (See also absolute coil). comparator coils—two or more coils electrically connected in
absolute readout—the signal output of an absolute coil. (See series opposition but arranged so that there is no mutual
also absolute coil.) induction (coupling) between them such that any electric or
absolute system—a system that uses a coil assembly and magnetic condition, or both, that is not common to the
associated electronics to measure the total electromagnetic specimen and the standard, will produce an unbalance in the
properties of a part without direct comparison to another system and thereby yield an indication.
section of the part or to another part (see absolute coil.) conductivity—the intrinsic property of a particular material to
acceptance level—a level above or below which specimens carry electric current; it is commonly expressed in percent
are acceptable in contrast to rejection level. IACS (International Annealed Copper Standard) or MS/m
acceptance limits—levels used in electromagnetic sorting (MegaSiemens/metre).
which establish the group into which the material under coupling—two electric circuits are said to be coupled to each
examination belongs. other when they have an impedance in common so that a
amplitude distortion—same as harmonic distortion. current in one causes a voltage in the other.
amplitude response—that property of an examination system cut-off level—same as rejection level.
whereby the amplitude of the detected signal is measured defect resolution—a property of an examination system that
without regard to phase. (See also harmonic analysis and enables the separation of indications due to defects in a
phase analysis.) sample that are located in proximity to each other.
annular coil clearance—the mean radial distance between depth of penetration—the depth at which the magnetic field
adjacent coil assembly and part surface in electromagnetic strength or intensity of induced eddy currents has decreased
encircling coil examination. to 37 % of its surface value. The depth of penetration
annular coils—see encircling coils. depends upon the coil size, the frequency of the signal, and
the conductivity and permeability of the material. It is
artificial discontinuity—reference discontinuities, such as
related to the coil size at low frequencies and is equal to the
holes, grooves, or notches, that are introduced into a refer-
skin depth at high frequencies. Related synonymous terms
ence standard to provide accurately reproducible sensitivity
are standard depth of penetration and skin depth. (See also
levels for electromagnetic test equipment.
skin effect.)
band pass filter—a wave filter having a single transmission
diamagnetic material—a material whose relative permeabil-
band; neither of the cut-off frequencies being zero or infinity.
ity is less than unity.
bobbin coil—see ID coil.
bucking coils—same as differential coils. NOTE 10—The intrinsic induction Bi is oppositely directed to the
circumferential coils—see encircling coils. applied magnetizing force H.
coil, absolute—see absolute coil. differential coils—two or more coils electrically connected in
coil, reference—see reference coil. series opposition such that any electric or magnetic condi-
coil size—the dimension of a coil, for example, length or tion, or both, that is not common to the areas of a specimen
diameter. being electromagnetically examined will produce an unbal-
coil spacing—the axial distance between two encircling coils ance in the system and thereby yield an indication.
of a differential system. differential measurements—measurements made in which
coil, test—the section of the probe or coil assembly that excites the imbalance in the system is measured using differential
or detects, or both, the electromagnetic field in the material coils in contrast to absolute and comparative measurements.
under examination. (See also differential coils.)
comparative measurements—measurements made in which differential readout—the signal output of differential coils.
the unbalance in the system is measured using comparator (See also differential coils.)
coils in contrast to differential and absolute measurements. differential signal—an output signal that is proportional to the
(See also comparator coils.) rate of change of the input signal.

You might also like