Saritorious Combics 1

Move the knob (behind the indicator) to the left side, as seen from behind.

 Switch Combics Off, and On


 Press tare (-T-) with in second, after the Combics is switched on.

Appl Appl will appear

Press function to toggle until setup

Setup

Press Tare to select WP1


WP1

After pressing Tare, you will be prompted to enter


202122 code, enter 202122

After entering code press <T>, u will go back to WP-1 or setup


depending on the timing of code whether it appeared before
the Wp1 or after the Wp1

WP1 Press Tare to select WP1

Press Tare to Seect Intercal


InteCal

Press Tare to select Param 1


Param 1

Toggle by pressing function till you reach

Max.Ext.W Max.Ext.W
Press T after Cal/adj is appeared

Cal/adj

Press again Tare (t) to insert the desired


Insert the value of the
weight
weight

After entering the weight press again tare T,


cal/adj will be displayed (put the weight on
Cal/adj platform when this appears) Press and hod T
for some time

Error, that will plus minus some thing will


Error (tollerence) be shown, Press and hold ISO test to
calibrate

Cal.esur will be shown, press Tare to end


Cal.Euser

Real weight after


calibration will be shown