Professional Documents
Culture Documents
SCHOOL OF MEDICINE
• HUMAN HISTOLOGY
• HISTOLOGICAL TECHNIQUES
- Protein fibers
vCollagen (most abundant of all protein fibres)
vReticular
vElastic
- Ground substance
• Both formaldehyde and glutaraldehyde react with the amine groups (NH2)
of tissue proteins, preventing their degradation
• In order to cut very thin sections, tissues must be infiltrated after fixation
with embedding material that imparts a rigid consistency to the tissue
• Paraffin is used routinely for light microscopy, resins for both light and
electron microscopy
• GAGs and many acidic glycoproteins do not undergo the PAS reaction, but
because of their high content of anionic carboxyl and sulfate groups, they
show a strong electrostatic interaction with alcian blue and other basic stains
• One grouping is based on what interacts with the sample to generate the
image, i.e., light or photons (optical microscopes), or electrons (electron
microscopes)
1. Optical microscopes
Ø Conventional bright-field microscope
Ø Fluorescence microscope
Ø Phase-contrast microscope
Ø Differential interference microscope
Ø Polarizing microscope
Ø Confocal microscope
2. Electron microscopes
- Transmission electron microscope (TEM)
- Scanning electron microscope (SEM)
• Objective lens:
- Enlarges and projects the image of the object in the direction of the eyepiece
- Usually you will find 3 or 4 objective lenses on a microscope
- They almost always consist of 4X, 10X, 40X and 100X powers
- When coupled with a 10X (most common) eyepiece lens, we get total magnifications
of 40X (4X times 10X), 100X , 400X and 1000X
• Resolving power is the smallest distance between two particles at which they can
be seen as separate objects
• Likewise, two structures such as mitochondria will be seen as only one object if
they are separated by less than 0.2 μm
2/13/20 Dr. Mukape Mukape 33
Bright-Field Microscopy
• The quality of the image—its clarity and richness of detail—depends on the
microscope’s resolving power
• Resolving power of a microscope depends mainly on the quality of its objective lens
• Eyepiece lens enlarges only the image obtained by the objective; it does not
improve resolution
• For this reason, when objectives of different magnifications are compared, those
providing higher magnification also have higher resolving power
• Diaphragm or Iris:
- Many microscopes have a rotating disk under the stage
- This diaphragm has different sized holes and is used to vary the intensity and size of
the cone of light that is projected upward into the slide
- There is no set rule regarding which setting to use for a particular power
- Rather, the setting is a function of the transparency of the specimen, the degree of
contrast you desire and the particular objective lens in use
2/13/20 Dr. Mukape Mukape 37
Bright-Field Microscopy
• How to Focus The Microscope:
- The proper way to focus a microscope is to start with the lowest power
objective lens first and while looking from the side, crank the lens down as
close to the specimen as possible without touching it
- Now, look through the eyepiece lens and focus upward only until the image is
sharp
- If you can't get it in focus, repeat the process again
- Once the image is sharp with the low power lens, you should be able to simply
click in the next power lens and do minor adjustments with the focus knob
- If your microscope has a fine focus adjustment, turning it a bit should be all
that's necessary
- Continue with subsequent objective lenses and fine focus each time
2/13/20 Dr. Mukape Mukape 38
Bright-Field Microscopy
• Digital cameras highly sensitive to light enhance the power of the bright-field and
other light microscopes by allowing the capture of images suitable for quantitative
analysis and immediate printing
• The frontiers of light microscopy have been redefined by the use of digital cameras,
and image-enhancement programs (eg, to improve contrast) allow objects that may
not be directly visible through the eyepieces to be analysed on the video screen
• Such systems are also useful for studying living cells for long periods of time because
they use low-intensity light that avoids damaging the cells with heat from more
intense illumination
• Software developed for image analysis allows rapid measurements and quantitative
study of microscopic structures
2/13/20 Dr. Mukape Mukape 39
Fluorence Microscopy
• When certain cellular substances are irradiated by light of a proper wavelength,
they emit light with a longer wavelength—a phenomenon called fluorescence
• For this method, the microscope has a strong UV light source and special filters that
select rays of different wavelengths emitted by the substances
• Fluorescent compounds with affinity for specific cell macromolecules may be used
as fluorescent stains
2/13/20 Dr. Mukape Mukape 40
Fluorence Microscopy
• When observed in the fluorescence microscope, these nucleic acids emit slightly
different fluorescence, allowing them to be localized separately in cells
• Other compounds such as DAPI and Hoechst stain specifically bind DNA and are
used to stain cell nuclei, emitting a characteristic blue fluorescence under UV
• Cellular detail is normally difficult to see in unstained tissues because all parts
of the specimen have roughly similar optical densities
• These changes are used by the phase-contrast system to cause the structures to
appear lighter or darker in relation to each other
• Because they allow the examination of cells without fixation or staining, phase-
contrast microscopes are prominent tools in all cell culture laboratories
• Stray (excess) light reduces contrast within the image and compromises the
resolving power of the objective lens
• Confocal microscopy avoids these problems and achieves high resolution and
sharp focus by using a:
1. Small point of high-intensity light, often from a laser
2. Plate with a pinhole aperture in front of the image detector
• When normal light passes through a polarizing filter, it exits vibrating in only one
direction
• If a second filter is placed in the microscope above the first one, with its main axis
perpendicular to the first filter, no light passes through
• If, however, tissue structures containing oriented macromolecules are located between
the two polarizing filters, their repetitive structure rotates the axis of the light emerging
from the polarizer and they appear as bright structures against a dark background
• The wavelength in the electron beam is much shorter than that of light,
allowing a 1000-fold increase in resolution
• The voltage difference between cathode and anode can be varied between
roughly 60 and 120 kV, producing electron beams of different wavelengths
• Electrons transmitted through the specimen reach the objective lens, which
forms a focused, magnified image that is then magnified further through
other lenses and captured on a viewing screen