You are on page 1of 7

The New Wave of

Ion-Selective
Thanks to recent advances, ISEs are
tackling trace-level measurements.

few years ago, leading scientists advised their


younger colleagues not to work in the field of
potentiometric sensors. Researchers viewed po-
tentiometry as a mature field; all the basic facts
were well understood, and no possibilities existed
for significant new achievements.
In fact, although ion-selective electrodes (ISEs) had been de-
scribed for more than 50 analytes, their practical use was limited
to only a few fields. They were especially well established in clin-
ical chemistry for the measurement of K+, Na+, Ca2+, and Cl–
concentrations (activities) in whole blood, serum, or plasma. The
two main limitations were insufficient detection limits and selec-
tivities. Indeed, most potentiometric sensors had detection limits
in the micromolar range, and interfering ions were discriminated
by, at most, a factor of 103–104. Such detection limits and selec-
tivities were not acceptable for trace-level measurements, that is,
for assessing submicromolar activities in the presence of large
concentrations of interfering ions. Applications such as monitor-
ing heavy metals in drinking water or blood seemed like hope-
less ideas.
This situation has dramatically changed in the past few years.
Today, it is feasible to construct ISEs for a number of analytes
with detection limits down to low parts-per-trillion levels (10–10
M). Accurate monitoring of heavy metals in unspiked environ-
mental samples has been demonstrated and compared with rou-
tine analytical methods. This breakthrough certainly came as a
surprise to most researchers working with chemical sensors and
Eric Bakker has fueled renewed interest in pursuing active research in this
Auburn University field. Unlike most other analytical methods for trace-level deter-
minations, these advances were not achieved by improving the
Ernö Pretsch instrumental technique but by studying the underlying chemi-
Swiss Federal Institute of cal principles and eliminating experimental biases. In this article,
Technology (ETH Zürich) we describe how the detection limits and selectivities of poten-

420 A A N A LY T I C A L C H E M I S T R Y / A U G U S T 1 , 2 0 0 2
Electrodes

A U G U S T 1 , 2 0 0 2 / A N A LY T I C A L C H E M I S T R Y 421 A
(a) emf (b) emf

Ion-selective Reference Primary ion I+


electrode electrode
Interfering
Inner reference ion J+
(Ag/AgCl) Gibb’s free energy of ion transfer from the aque-
Inner
reference Reference ous to the organic phase; and R, T, and F are the
(Ag/AgCl) electrolyte gas constant, the absolute temperature, and the
Diaphragm
Inner filling
Faraday constant, respectively.
Bridge
solution electrolyte Figure 2a illustrates this simple mechanism.
Ion-selective Diaphragm or In the case shown here, the composition of the
membrane Sample solution capillary membrane phase (aM(org) in Equation 1) remains
ai(DL) aj(BG) essentially unchanged as the concentration in the
log a
sample phase is altered, and a response accord-
FIGURE 1. How electrodes work. ing to the so-called Nernst equation is observed:
(a) Schematic of a potentiometric cell with an indicator and a reference electrode. (b) Calibration
curves for a primary cation (blue) and an interfering cation (red) of the same charge. The emf RT
shift between the two curves is a measure of the selectivity. For example, a shift of 178 mV E PB = K + 1naM(aq) (2)
signifies a 3-orders-of-magnitude preference of the analyte ion over the interference. The back-
zF
ground activity of interfering ions, aj(BG), defines the lowest possible potential (horizontal green
dotted line) that can be measured under these conditions. This potential represents the best
achievable detection limit of the primary ion ai(DL) in the presence of the background. This de- in which K is a constant for each M. If interfer-
tection limit is often not observed in practice because of transmembrane ion fluxes. ence from other sample ions is significant, the
membrane concentration of M+ changes, which
alters the observed emf (Equation 1). To obtain
tiometric sensors have been improved by many orders of magni- highly selective ISEs on the basis of polymeric membranes, se-
tude by simple chemical means. lective complexing agents (ionophores) have been synthesized
and doped into these membranes. Their purpose is to selective-
Basics ly incorporate only one type of ion into the membrane and to
Potentiometric sensors are essentially passive electrochemical de- suppress competition from interfering ions (3, 4). The synthe-
vices, in which changes in the electromotive force (emf) are mon- sis and application of novel ionophores were the primary fo-
itored under virtually zero current conditions. In so-called di- cuses of research in this field for many years (5).
rect potentiometry, the emf is ideally a function of the activity The relationships outlined earlier can be adequately described
of only one chosen sample ion so that it can be selectively as- with thermodynamic models. Indeed, the composition of the or-
sessed in the presence of other ions. Achieving sufficient selec- ganic-phase boundary layer can be related to the composition
tivity to measure one ion in the presence of others has long been of the adjacent sample layer on the basis of ion-exchange and elec-
the main focus of research in this field (1). The basic concept is trolyte coextraction equilibria between the two phases and com-
perhaps best illustrated with ISEs containing polymeric mem- plexation and ion-pair formation equilibria in the organic mem-
branes, which are assumed to behave as organic liquids of high brane phase (2). Today, the theoretical background has matured
viscosity. Today, this class of sensors is by far the most versatile to a stage that it can be used to optimize the membrane compo-
and widely established (2). sition and describe how ion sensors behave in mixed electrolyte
If this membrane contains a lipophilic cation exchanger, its solutions. The lower detection limit, for example, is reached in
countercation can freely exchange or equilibrate with the two theory when interfering ions displace 50% of the ions to be mea-
contacting aqueous solutions (Figure 1a). Assume that these sured from the membrane. As a consequence, one established
solutions contain the same hydrophilic electrolyte, M+A–, and experimental protocol to determine the selectivity of the sensor
M+ is the countercation of the ion exchanger in the membrane. toward an ion calls for the measurement of the lower detection
A chemical driving force that depends on the activity of this ion limit in the presence of a constant ionic background (6). If the
in both phases causes M+ to partition from the organic to the detection limit is low, the selectivity toward the background ion
aqueous phase. Because M+ carries a charge and no net current is good (Figure 1b).
can flow, this driving force is counterbalanced by the phase Unfortunately, the thermodynamic treatment of ISE function
boundary potential. It builds up spontaneously as M+ starts to has apparently failed for dilute solutions. The observed emf has
cross the interface and leads to interfacial charge separation. often been affected by sample stirring, for example, and selectiv-
The relationship between this interfacial potential, EPB, and the ities determined under such conditions have turned out to be
ion activities on either side of the phase boundary is grossly underestimated. It is now well established that the ob-
served detection limit is often not dictated by changes in the
RT k a (aq) membrane concentration of M+, as expected from the thermo-
E PB = 1n M M (1) (1) dynamic model alone. Rather, nonthermodynamic
zF aM(org)
processes may polarize the concentration of M+
in the contacting sample solution and give a
in which aM(aq) and aM(org) are the activities of the ion, higher value at the sensing surface than in
with charge z in the aqueous and organic phases, respec- the sample bulk. A thermodynamic treat-
tively; kM is a constant that incorporates the standard ment is still valid for the local interfacial

422 A A N A LY T I C A L C H E M I S T R Y / A U G U S T 1 , 2 0 0 2
processes, even though the sensed concentration at the phase layer near the membrane. Because ordinary ion concentrations
boundary may deviate from the concentration in the bulk. This is are on the order of 10–2 M in the membrane phase, a concen-
very surprising. After all, potentiometry is assumed to be a static tration polarization of only 1% in the membrane produces a con-
method and should give thermodynamic information about a centration difference of ~10–7 M between the surface layer and
sample. How can we locally polarize the concentra- the bulk of the sample. The analyte will have this concentra-
tion if no current is passed across the cell? tion in the aqueous layer adjacent to the membrane even
if the bulk contains virtually no sample ions. Because
Ion fluxes in potentiometry the phase boundary concentration determines the
Generally, the sample and the inner solu- emf, it is the fundamental reason the lower detec-
tion do not have identical compositions. tion limits of innumerable polymer membrane
Because ions can diffuse across the organ- electrodes were reported to be on the order of
ic membrane phase according to various 10– 6–10–7 M (3).
mechanisms, this concentration asymme- The same process has also dramatically biased
try induces a process toward a true equi- the quantification of the selectivities of many
librium. This process is so slow that it can membrane electrodes. To quantify selectivities cor-
be ignored if the solution concentrations rectly, the measurements must be done under con-
of the relevant species are higher than ~1 µM. ditions in which only one of the ions is potential-de-
However, if lower concentrations are involved, termining. If the membrane were highly selective, the
transmembrane ion fluxes may have a decisive in- emf of a concentrated solution of an interfering ion would
fluence on the response of ISEs. According to the equilib- be lower than the emf of a 10–6-M solution of the analyte ion
rium model (Figure 2a), no gradients arise in the system. This (Figure 1b). In other words, the small amount of leached primary
model is very close to reality, even if the composition of the inner ions, not the discriminated interfering ions, determined, in many
solution is different from that of the sample. The ion-exchange cases, the measured selectivities (6).
property of the hydrophobic membrane largely blocks the incor-
poration of the aqueous cation and its counterion into the organ- How can ion fluxes be reduced?
ic phase. Furthermore, the high selectivity largely hinders the The key to the recent spectacular improvements toward lower
exchange of the analyte ion with an interfering ion. The key to un- detection limits and better selectivities has been the reduction of
derstanding the processes at low sample concentrations is the word transmembrane ion fluxes (9, 10). As shown in Figure 2b, opti-
“largely”. A very small fraction of the primary ions is exchanged by
interfering ions already at concentrations that are orders of magni-
tude larger than the lower detection limit (Figure 1b). (a)
Ideal potentiometric response
For clarity, the situation is represented in Figure 2b with ex-
aggerated gradients (only the unstirred layer of the sample is Sample Membrane Inner solution
+ +
– – + – + – + +
shown). The exchange of a small portion of the sample ions + –
+– + + +– – –+

(blue) with interfering ions (red) induces opposite gradients of + –– + + + + – + –
– + + – – –+ – +
the primary (blue lines) and interfering ions (red lines) in the – –
+ +– + +
– – –+
– + – + + –
membrane and in the contacting stagnant aqueous layer. These + + – – +
+ – – + – – +
gradients generate ion fluxes without net charge transport and
cause an elevated primary ion concentration at the sensor surface. (b)
Because this phase boundary concentration determines the emf, Near detection limit
these diffusion processes may cause a strong bias when trying to Stagnant layer Membrane Inner solution
+ + + – + – + +
measure lower detection limits and apparent selectivities. Such – – + –
+ + + +– – –+
elevated concentrations near the membrane surface were indeed – +
+ –– + + + + – + –
recently observed by scanning electrochemical microscopy (7 ). – + + – – –+ – +
– –
+ +– + +
– – –+
Fortunately, the phase boundary potential model described – + – +
+ – – –+ – + + – –+
above is also valid when the whole system is out of equilibrium + + –
because phase boundary processes are much faster than trans- FIGURE 2. Schematic of a sensing membrane with the sample
membrane ion fluxes. Because the thickness of the concentration and the inner solution for a symmetrical setup (a) in equilibrium
gradient in the stagnant aqueous layer and the membrane are and (b) in the steady state with zero-current ion fluxes.
comparable (Figure 2b), the ratio of the two gradients is rough- (a) The concentration of analyte ions (blue) in the membrane is defined by
ly equal to the ratio of the corresponding diffusion coefficients the amount of lipophilic anions, which cannot leave the membrane. (b) Inter-
(8). Diffusion coefficients in the membrane phase vary according fering ions (red) replace a small amount of analyte ions in the membrane on
the sample side. Because of the resulting gradients, analyte ions leach from
to polymer and plasticizer content but are ~1000-fold smaller
the membrane into the sample and induce an elevated concentration near
than those in water. Therefore, the concentration difference be- the membrane. At submicromolar sample activities (e.g., during trace-level
tween the two surface layers of the membrane is ~1000-fold measurements or selectivity determinations), these analyte ions can become
higher than between the bulk of the sample and in the aqueous potential-determining and cause gross errors.

A U G U S T 1 , 2 0 0 2 / A N A LY T I C A L C H E M I S T R Y 423 A
emf
Ca2+ Cd2+ Pb2+

59.2 mV 59.2 mV 59.2 mV


by excessively rapid diffusion through the membrane and exces-
sively slow diffusion in the aqueous phase. Any measure that re-
–14 –10 – 6 –2 –10 –8 –6 –12 –8 –4
log aCa log aPb
duces ion fluxes in the membrane is, therefore, beneficial (14,
log aCd
15). For example, membranes are more rugged if the diffusion
emf ClO4– coefficients are reduced by increasing their polymer content; if
NH4+ K+
the ionophore is covalently bound to the polymer backbone; if
the total amount of ions (i.e., the concentration of the ion ex-
changer in the membrane) is reduced; if the thickness of the
membrane is increased; if its effective surface is reduced relative
59.2 mV 59.2 mV 59.2 mV to water by adding inert lipophilic particles; or if the stirring of
the aqueous phase is increased (e.g., by using a flow-through sys-
–9 –7 –5 –3 –9 –7 –5 –3 –4 –6 –8 tem or a wall-jet configuration) (15, 16). By combining two of
log aNH4 log aK log aCIO4
these measures (i.e., adjusting the inner solution and making the
FIGURE 3. Response functions of some ISEs with recently membrane less sensitive to concentration gradients), scientists
improved lower detection limits. have designed potentiometric sensors that are adequate for mon-
itoring heavy metals such as lead in drinking water (17 ).

mal detection limits and unbiased selectivity behavior would, in What’s going on now?
theory, be achieved by using an inner solution that exactly match- A series of ISEs have recently been developed with lower detec-
es the sample. Here, “matching” means that the composition of tion limits (Figure 3) that are far better than what was expected
the inner solution is designed in a way that the same amount of only a few years ago (18). These ISEs can now compete with the
analyte ion is replaced by an interfering ion on either side of the most sophisticated methods of trace analysis. More importantly
membrane. As shown earlier, diffusion of analyte ions from the and parallel to these developments, a sound theoretical under-
membrane to the sample generates less-than-optimal detection standing of the basic mechanisms has been achieved, which makes
limits and selectivities. On the other hand, if the inner solution it possible to optimize the electrodes very efficiently (12, 17, 19).
is designed so that too many analyte ions at the inner membrane Once their selectivity has been determined, the composition of
side are replaced by other ions, a flux of analyte is created from the inner solution can be predicted (12).
the sample to the membrane. Also, this process is harmful be- The most important practical applications seen today are heavy
cause during low sample activities, it depletes sample ions near metal analyses of drinking water, other environmental samples,
the membrane and leads to a nonideal response (11). and biological samples, which demand high selectivity, rugged-
Originally, the composition of the inner solution was adjust- ness, and short response and calibration times. For lead in drink-
ed by using a hydrophilic complexing agent such as EDTA to ing water, the detection limit should be at least 10 times lower
buffer the primary ion activity and a rather high concentration than the 15-ppb (7.2  10–8 M) action limit imposed by the U.S.
of the interfering ion (10). On the basis of the selectivity behav- Environmental Protection Agency (20). In Figure 4b, the action
ior of the membrane, researchers can predict the composition of limit is marked by the color transition between the green and red
the inner solution in order to replace the required portion (typ- zones of the plot. The detection limit of 7.2  10–9 M must be
ically <1%) of primary ions with interfering ions (12). For ions reached in the presence of 10–3–10–4-M levels of the abundant
such as alkali metal ions or many anions for which no appropriate interfering ions Na+, Ca2+, and H+.
complexing agents are known, an ion exchanger can be added to Screening of several existing ionophores determined that the
the inner solution to act as a nonselective ion buffer because of basket-shaped, thioamide-derivatized calix[4]arene was the most
the rather large capacity of the resin (13). appropriate host molecule for lead because of its excellent selec-
Because the inner solution is optimal for only one (typical) tivity (21). In fact, the characteristics of the resulting Pb2+ ISE ex-
sample composition, any concentration change in the sample will ceeded the earlier stated requirements, and alkali earth metal ions
induce an ion gradient in either direction. It is therefore also im- were found to be discriminated by as much as 14 orders of mag-
portant to make ion-selective membranes as insensitive as pos- nitude. This left some room for reducing the ion exchanger con-
sible against such gradients. It is astonishing at first glance that centration in the membrane, which resulted in a further reduction
problems like biased detection limits and selectivities are caused in ion flux and more rugged behavior, even though the selectivi-
ty was somewhat reduced. The detection limits in typical ionic
backgrounds for tap water were ~1 nM, and they were perfectly
Table 1. Direct potentiometric determinations in line with theoretical expectations. Because response times dur-
of Pb2+ in tap water (17 ). ing such low activities are rather long, calibration solutions in the
range 10–5–10–4 M were used, indicated by the colored points
ISE (µg/L) ICPMS (µg/L)
Unspiked tap water 0.71 ± 0.04 0.54 ± 0.07
in Figure 4b and coded with the same colors in Figure 4a. The
2.0 µg/L 2.61 ± 0.06 2.80 ± 0.04 5-point calibration by standard addition required ~10 min. It
10 µg/L 11.5 ± 0.3 11.30 ± 0.06 may seem dangerous to rely on calibration solutions that are
20 µg/L 22.2 ± 1.2 22.0 ± 0.2 1000–10,000-fold more concentrated than the samples. How-

424 A A N A LY T I C A L C H E M I S T R Y / A U G U S T 1 , 2 0 0 2
ever, thanks to excellent reproducibility of the response slopes, the original papers, if the biasing effect of ion fluxes is elimi-
the observed uncertainties were acceptably small (Table 1). nated. Very probably, such improvements will be found for many
Figure 4b shows a typical measurement of untreated tap water other available systems.
at its native pH 8, indicated as the lower blue point on the cali- A sound theoretical basis of the involved zero-current trans-
bration curve. Clearly, the assessed level of lead (2  10–10 M) is port processes has been developed and the main predictions have
extremely low. The lead concentration of the same sample deter- been experimentally verified. A palette of measures is available
mined with inductively coupled plasma MS (ICPMS), indicated as that can be used to optimize membranes. The theory can be di-
a filled circle on the calibration curve of Figure 4b, is an order of rectly applied to predict the required composition of an inner
magnitude higher than the one obtained with the potentio- electrolyte on the basis of the selectivity behavior of the mem-
metric sensor. This is not surprising if one realizes that brane and the composition of the sample. Improving the detec-
potentiometric measurements assess free ion activities, tion limits of future electrodes will now be more routine. Lastly,
not total ion concentrations. At elevated pH, most the practical use of recently developed electrodes for trace-level
of the lead ions are complexed, mainly by carbonate, determination has been demonstrated. On the basis of this, we
and <10% remains in the free, uncomplexed form. It expect that a series of potentiometric sensors with submicromo-
is this form that the potentiometric sensor observes, lar detection limits for use in environmental and clinical analyses
hence the lower value. If the sample is buffered to pH 4.0, of trace metals will be described soon. Although the possibili-
where lead is mostly in its free form, the values obtained by po- ties are promising, clinical trace metal determinations by poten-
tentiometry and ICPMS are indeed nearly identical (Figure 4b). tiometry have not yet been the focus of research.
The unique characteristics of potentiometric sensors that allow A very small external current can be used to compensate for
speciation measurements are further illustrated in Figure 4c. A zero-current ion fluxes (16, 22). This method is attractive because
spiked tap water sample was adjusted to various pH values, and it is easier to alter the required current than the compositions of
the potentiometrically assessed lead values (open circles) agree
well with theoretical expectations on the basis of two carbonate
levels in the sample (dotted lines). Again, it can be seen that (a) 29.6 mV (b) Tap water
buffering to pH 4 transforms all the lead complexes into their
free form, and potentiometric and atomic spectrometric methods

emf
yield comparable results under these conditions (17 ).
emf

10 –9 M
10 –9 M
To compare total assessed lead levels determined by poten- pH 4
29.6 mV
tiometry and ICPMS, various samples were collected and buffered pH 8 ICPMS
to pH 4.0 before measurement. Table 1 shows the observed lead 5 x 10 –7 M 5 x 10 –7 M
concentrations for native and spiked samples and demonstrates
excellent correlation and acceptably small standard deviations for 0 5 10 15 20 25 –10 –9 –8 –7 –6 –5 –4
the potentiometric data. The standard deviations were obtained Time (min) log [Pb2+]
from measurements with three different sensors on the same sam-
ple. The determination of lead represents a prototype study and (c) 100
indicates that a procedure for environmental monitoring of other
% Uncomplexed lead

80
heavy metal ions for which ISEs with sufficient selectivities are
available (e.g., Cd2+, Ag+, Cu2+) is likely to be developed soon. 60
In the context of practical applications, it is important to re- 40
member that ISEs are quite unique in their capability to assess
free ion activities. This is in contrast to the information obtained 20
by other methods: Atomic spectrometry gives total concentra- 0
tions, and voltammetry yields concentrations of labile species 3 4 5 6 7 8 9
and complexes. Depending on the problem at hand, one or the pH
other value may be of primary interest. For speciation purpos-
es, a comparison of several methods or potentiometric measure- FIGURE 4. Lead in drinking water.
ments under different conditions may be useful. (a) Time responses during a five-point calibration of two different samples
containing the indicated concentration of lead ions. (b) Lead in native tap
What holds the future? water at pH 8, after buffering to pH 4, and five calibration points with the
same activities, as in (a). The black-filled circle represents the concentration
It’s been almost four years since the first papers on improving de-
determined by ICPMS analysis. The observed slope of 29.6 mV/decade is
tection limits were published. The successes achieved since have expected, according to the Nernst equation (Equation 2) for a divalent ion
been rather impressive. A series of ISEs have been described, (ln 10  RT/2F = 29.6 mV). (c) Potentiometric responses of a drinking water
showing how the lower detection limits of any polymeric mem- sample spiked with 10 ppb of lead at different pH values. Dotted lines are
the theoretically expected fractions of uncomplexed lead, taking into account
brane electrode of sufficient selectivity can be improved. The se- complexation equilibria with the specified levels of CO2– 3 (two assumed con-
lectivities of a series of reinvestigated electrodes have been im- centrations), HCO3–, OH–, SO42–, NO3–, HSO4– , and Cl–. Carbonate is the domi-
proved by many orders of magnitude over those determined in nating complexing ion.

A U G U S T 1 , 2 0 0 2 / A N A LY T I C A L C H E M I S T R Y 425 A
the membrane and the inner solution. metric, and optical sensors based on molecular recognition prin-
However, determining the required cur- ciples. Ernö Pretsch is a professor at the Swiss Federal Institute
rent for an unknown sample is not a sim- of Technology (ETH). His more recent research interests focus
ple task, and further research is needed on the improvement of lower detection limits for ISEs and the
to show the practical applicability of this development of computer programs for spectrum estimation
approach. Another related research en- and automatic spectra interpretation. Address correspondence
deavor is the development of ISEs without about this article to Pretsch at pretsch@org.chem.ethz.ch.
inner solution to eliminate one major source of
transmembrane ion fluxes (23). This promises to References
eventually eliminate steady-state ion fluxes completely because only (1) Buck, R. P.; Lindner, E. Anal. Chem. 2001, 73, 88 A–97 A.
two phases would be involved (Figure 2). If successful, it should (2) Bakker, E.; Bühlmann, P.; Pretsch, E. Chem. Rev. 1997, 97, 3083–3132.
further improve detection limits by many orders of magnitude. (3) Bühlmann, P.; Pretsch, E.; Bakker, E. Chem. Rev. 1998, 98, 1593–1687.
Sensor miniaturization will be necessary for achieving low de- (4) Bakker, E.; Bühlmann, P.; Pretsch, E. Electroanalysis 1999, 11, 915–933.
tection limits because an ordinary membrane contains ~10–7 moles (5) Umezawa, Y. Handbook of Ion-Selective Electrodes: Selectivity Coefficients;
of primary ions, which corresponds to a picomolar concentra- CRC Press: Boca Raton, FL, 1990.
tion in 10,000 L of sample. If a sensing membrane is placed di- (6) Bakker, E.; Pretsch, E.; Bühlmann, P. Anal. Chem. 2000, 72, 1127–1133.
rectly on a metal, there are two main concerns: how to guarantee (7) Gyurcsányi, R. E.; et al. Anal. Chem. 2001, 73, 2104–2111.
a stable (redox) potential between the membrane and the metal (8) Sokalski, T.; Zwickl, T.; Bakker, E.; Pretsch, E. Anal. Chem. 1999, 71,
(23–26) and how to prevent the spontaneous formation of a 1204–1209.
water film between the two (27, 28). Active research on different (9) Mathison, S.; Bakker, E. Anal. Chem. 1998, 70, 303–309.
possible solutions is being pursued, so the development of low- (10) Sokalski, T.; Ceresa, A.; Zwickl, T.; Pretsch, E. J. Am. Chem. Soc. 1997, 119,
detection-limit, solid-contact electrodes is likely. 11,347–11,348.
Another promising field of research that will benefit from (11) Sokalski, T.; Ceresa, A.; Fibbioli, M.; Zwickl, T.; Bakker, E.; Pretsch, E.
these developments is the design of optical sensors with small Anal. Chem. 1999, 71, 1210–1214.
sensing volumes (29), whose response mechanism is closely re- (12) Ion, A. C.; Bakker, E.; Pretsch, E. Anal. Chim. Acta 2001, 440, 71–79.
lated to that of potentiometric sensors (2). However, steady-state (13) Qin, W.; Zwickl, T.; Pretsch, E. Anal. Chem. 2000, 72, 3236–3240.
ion fluxes are not anticipated here because these sensors rely on (14) Zwickl, T.; Sokalski, T.; Pretsch, E. Electroanalysis 1999, 11, 673–680.
a two-phase equilibration process. Therefore, low detection lim- (15) Ceresa, A.; Sokalski, T.; Pretsch, E. J. Electroanal. Chem. 2001, 501, 70–76.
its should be more easily obtained than with classical ISEs (30). (16) Lindner, E.; Gyurcsányi, R. E.; Buck, R. P. Electroanalysis 1999, 11,
For most techniques, lower detection limits are either set by 695–702.
interferences or by the physical limitations of the instruments. (17) Ceresa, A.; Bakker, E.; Günther, D.; Hattendorf, B.; Pretsch, E. Anal. Chem.
The ultimate low detection limit of potentiometry is still not 2001, 73, 343–351.
yet known. Highly selective systems exist and, in principle, they (18) Bakker, E.; Pretsch, E. Trends Anal. Chem. 2001, 20, 11–19.
should be able to achieve such extremely low detection limits (19) Morf, W. E.; Badertscher, M.; Zwickl, T.; de Rooij, N. F.; Pretsch, E.
that they do not seem realistic. It is expected, but not yet shown J. Phys. Chem. B 1999, 103, 11,346–11,356.
experimentally, that the small currents on the order of 100 fA (20) Lead and Copper Rule Minor Revisions; EPA 815-F-899-010; U.S. Environ-
required for zero-current potentiometric measurements will set mental Protection Agency, 1999, http://www.epa.gov/safewater/
the limits, and these might be somewhere between picomolar standard/leadfs.html.
and femtomolar activities. (21) Malinowska, E.; Brzózka, Z.; Kasiura, K.; Egberink, R. J. M.;
Compared with other methods of trace analysis, potentiome- Reinhoudt, D. N. Anal. Chim. Acta 1994, 298, 253–258.
try is an extremely inexpensive technique. Because it can com- (22) Pergel, E.; Gyurcsányi, R. E.; Tóth, K.; Lindner, E. Anal. Chem. 2001, 73,
pete with the most sophisticated methods, it is very 4249–4253.
likely that we will encounter many practical applica- (23) Cattrall, R. W.; Freiser, H. Anal. Chem. 1971, 43, 1905–1906.
tions in the future. Potentiometric sensors are (24) Bobacka, J.; Lindfors, T.; McCarrick, M.; Ivaska, A.; Lewenstam, A.
known to only minimally perturb the sample, Anal. Chem. 1995, 67, 3819–3823.
compared with most other analytical methods, (25) Cadogan, A.; Gao, Z.; Lewenstam, A.; Ivaska, A.; Diamond, D. Anal. Chem.
and they should be especially well suited for in- 1992, 64, 2496–2501.
tegration into optimized fluidic and microfluidic (26) Fibbioli, M.; et al. Chem. Commun. 2000, No. 5, 339–340.
systems for a variety of applications. (27) Fogt, E. J.; Untereker, D. F.; Norenberg, M. S.; Meyerhoff, M. E.
Anal. Chem. 1985, 57, 1995–1998.
We are grateful for financial support of our research from the Na- (28) Fibbioli, M.; Morf, W. E.; Badertscher, M.; de Rooij, N. F.; Pretsch, E.
tional Institutes of Health (grant R01-GM59716), the Swiss National Science Electroanalysis 2000, 12, 1286–1292.
Foundation, and Orion Research. (29) Tsagkatakis, I.; Peper, S.; Retter, R.; Bell, M.; Bakker, E. Anal. Chem. 2001,
73, 6083–6087.
Eric Bakker is a professor at Auburn University. His research interests (30) Lerchi, M.; Orsini, F.; Cimerman, Z.; Pretsch, E.; Chowdhury, D. A.;
include fundamental and applied aspects of potentiometric, voltam- Kamata, S. Anal. Chem. 1996, 68, 3210–3214.

426 A A N A LY T I C A L C H E M I S T R Y / A U G U S T 1 , 2 0 0 2

You might also like