Professional Documents
Culture Documents
PerkinElmer, Inc.
710 Bridgeport Avenue
Shelton, CT 06484-4794, U.S.A.
Internet: http://www.perkinelmer.com
email: info@perkinelmer.com
Trademarks:
Registered names, trademarks, etc. used in this document, even when not specifically
marked as such, are protected by law.
PerkinElmer is a register trademark of PerkinElmer, Inc.
Notice:
The information in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer
makes no warranty of any kind with regard to this document, including, but not
limited to, the implied warranties of merchantability and fitness for a particular
purpose. PerkinElmer shall not be liable for errors contained herein for incidental
consequential damages in connection with furnishing, performance or use of this
material.
Release Information:
Manual Part No. Release Release Date
09932003 G May 2015
1
CONTENTS
SYMBOLS AND CONVENTIONS USED IN THIS
INSTRUCTION SHEET ..................................................................... 2
NOTES, CAUTIONS, AND WARNINGS ................................................... 2
THE PINAACLE SERIES METAL BODY NEBULIZER ............. 3
WARNING
The metal body nebulizer can be used for most solutions containing
< 5 % acid. It has excellent performance (precision and signal-to-noise
ratio) and it is less prone to clogging. The needle assembly and venturi
can be changed to produce a Platinum alloy and corrosion resistant
version. See Figure 1 for a close up of this nebulizer.
1 2
00570957 Capillary Assembly with long nebulizer capillary tubing for use with
the autosampler (nebulizer capillary tubing comes attached since the
capillary assembly is made of plastic)
03031920 Venturi
6
03033124 Venturi
4 3
NOTE: Use Apiezon®-L grease Make sure that no grease gets onto
the tip of the capillary.
2. Apply a small amount of Apiezon grease on all the inner
O-rings (insert, venturi and the needle assembly). The
nebulizer parts will seal better.
3. Insert the re-assembled nebulizer into the end cap in the
unlocked position.
4. Rotate the arm 90 degrees clockwise into the locked position.
5. Push the locking clamp all the way in to lock the nebulizer in
place.
WARNING
1. Turn off the spectrometer and turn off all gas supplies.
17
2. Lift the end-cap spring clamp and gently pull the nebulizer
out of the burner end cap.
3. After you have taken the nebulizer apart, clean the
components as follows:
4. Clean the capillary assembly using a mild laboratory
detergent solution, or a solvent such as isopropanol, preferably
in an ultrasonic bath.
5. Clean the nebulizer body and any other parts that appear
dirty in a mild laboratory detergent solution, or a solvent
such as isopropanol, preferably in an ultrasonic bath.
6. Rinse all the cleaned components thoroughly with deionized
water. If the venturi or the tip of the needle assembly looks
worn, then that part should be replaced.
7. Inspect the all O-rings on the nebulizer body and clean
them if necessary in mild laboratory detergent solution,
followed by a thorough rinsing with deionized water. If
they look damaged, replace them.
2 3 7 8 10
5
1 6 9
4 11
NOTE: If not already done, inspect all o-rings for damage and
replace them if necessary. Lightly lubricate the surface of the inner
O-rings with a minimum of the supplied Apiezon®-L grease. Do not
use Apiezon grease of unknown source and quality as it may contain
trace metals. Make sure that no grease gets onto the tip of the
capillary.
1. Replace the insert (with an o-ring placed on it) in the body of the
nebulizer if it was removed. The insert is in the body of the
nebulizer and is not seen in the above picture.
2. Replace the locking nut if it was removed from the nebulizer body.
3. Slide the spring back onto the capillary assembly. Place the Teflon
washer on the other end of the capillary assembly.
5. If the fine adjustment nut is screwed all the way in, the orientation
of the bevel on the needle assembly can be checked, by looking at
the other end of the nebulizer.
20
6. Replace the venturi (with o-ring) into the other end of the
nebulizer. The side with the o-ring goes in first. Twist the
venturi to make a good seal.
7. Replace the cap, then place the o-ring which seals the
nebulizer to the end cap, on the outside of the venturi.
If you are located outside of the U.S., call your local PerkinElmer
sales or service representative.
22
ATOMIC ABSORPTION
PerkinElmer, Inc.
710 Bridgeport Avenue
Shelton, CT 06484-4794, U.S.A.
Internet: http://www.perkinelmer.com
email: info@perkinelmer.com