Change Notification 1093 BM800 SW 2.9.3
Change Notification 1093 BM800 SW 2.9.3
Description of change
We released SW 2.9.3 for Alfa, M-series and Exigo during September, below changes have been
implemented:
• Bug fix regarding intermittent indication 103 and 309 during exit standby resolved.
• Complete removal of all control results from the control memory when the corresponding control
assay is deleted. ( Only for US human market )
Consequences
• All BM800 instruments can be upgraded
• SW 2.9.3 can be download from each support web page after login
• Software upgrade from 2.7.1 and below will erase control and patient memory
Traceability
BM800 Human:
Implemented in serial number: 24131, 24153, 24240 - 24241, 24244, 24245, 24255 - 24257, 24325, 24394,
24396 - 24398, 24428, 24745 and above regarding 20 000 series
BM800 VET:
Implemented in serial number: 51882 and above regarding 50 000 series
Implemented in serial number: 52753 and above regarding 52 000 series