Professional Documents
Culture Documents
Patient Monitors
Release C.00
English
Ins
11 in.
Service Guide
Efficia CM Series
Patient Monitors
Service Guide
C.00
English
Notice
Proprietary Information
This document contains proprietary information, which is protected by copyright.
Copyright
Copyright © 2019 Koninklijke Philips N.V.
All Rights Reserved
Trademark Acknowledgments
Efficia is a registered trademark of Koninklijke Philips Electronics N.V. Other product
names may be trademarks of their respective owners.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover, MA 01810-1085
(978) 687-1501
Document Number
453564623591 Rev C
Warranty Disclaimer
The information contained in this document is subject to change without notice. Philips
Medical Systems makes no warranty of any kind with regard to this material, including,
but not limited to, the implied warranties or merchantability and fitness for a particular
purpose. Philips Medical Systems shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing, performance, or
use of this material.
Printing History
New editions of this document incorporate all material updated since the previous
edition. Update packages may be issued between editions and contain replacement and
additional pages to be merged by a revision date at the bottom of the page. Pages that
are rearranged due to changes on a previous page are not considered revised.
The documentation printing date and part number indicate its current edition. The
printing date changes when a new edition is printed. (Minor corrections and updates that
are incorporated at reprint do not cause the date to change.) The document part number
changes when extensive technical changes are incorporated.
Third Edition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . September 2019
Text Formatting
The following typographical conventions are used in this guide.
Decimal Points
Because the Efficia CM monitors use a period (.) to indicate a decimal point in decimal
numbers (for example, 10.0), all decimal numbers in this guide use a period as a decimal
point. Commas are not use as decimal points.
Caution A Caution calls attention to a condition or possible situation that could damage or
destroy the product or the user’s work.
Warning A Warning calls attention to a condition or possible situation that could cause injury
to the user, patient, or both.
Explanation of Symbols
The following symbols appear on the monitor and its packaging.
Note — The symbols that appear on your monitor depend on the monitor model and its
configured options.
1014 hPA
max Atmospheric pressure range for transport ISO 15223-1* 5.3.9
and storage
708 hPA
min
90%
Relative humidity range (non-condensing) ISO 15223-1* 5.3.8
for transport and storage
15%
40°C
max
Ambient temperature range ISO 15223-1* 5.3.5
(non-condensing) for transport and
-20°C
storage
min
* ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labeling and information to be
supplied - Part 1: General requirements
** ISO 7000 Graphical symbols for use on equipment - Registered symbols
NBP key NA
Record key** NA
Home Key** NA
SpO2 connector NA
NBP connector NA
IBP connectors NA
IBP I
IBP 2
Temperature connectors ISO 7000*** 0034
VGA Output NA
USB port NA
Ethernet port NA
2019-02-08
Ingress protection against vertically falling IEC 60529, Ed. 2.2 Degrees
IPX1 water drops of protection provided by
enclosures (IP Code),
section 6
* ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labeling and information to be
supplied - Part 1: General requirements
** IEC 60417 Graphical symbols for use on equipment
Internal Symbols
The following symbols are located inside the monitor:
-20°C
min
* ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labeling and information to be
supplied - Part 1: General requirements
** IEC 60417 Graphical symbols for use on equipment
Accessory Symbols
The following symbols are located on the accessory packaging or on the accessory itself.
CapnoTrak n/a
Contains or Presence of Phthalate
DEHP
Ingress protection against splashing water IEC 60529, Ed. 2.2 Degrees
IPX4 of protection provided by
enclosures (IP Code), section
6
Ingress protection against solid foreign IEC 60529, Ed. 2.2 Degrees
IP44 objects of 1.0 mm diameter and against of protection provided by
splashing water enclosures (IP Code), section
6
Declaration
The Efficia CM10, CM12, CM100, CM120, and CM150 monitors are Class IIb medical
devices and comply with the requirements of the Council Directive 93/42/EEC of 14 June
0123
1993 concerning medical devices and carry CE-marking accordingly.
The radio device used in the Efficia CM10, CM12, CM100, CM120, and CM150 monitors is
in compliance with the essential requirements and other relevant provisions of Directive
2014/53/EU (Radio Equipment Directive).
Authorized EU Representative
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
EC REP 71034 Böblingen
Germany
Australian Sponsor
Philips Electronics Australia Ltd.
65 Epping Road, North Ryde, NSW
Australia, 2113
Rx Only
Caution United States Federal Law restricts this device to sale by or on the order of a
physician.
Safety Standards
Applicable Standards
Parameter Specification
Protection Against Hazards of Ignition Equipment is not suitable for use in the presence of a
of Flammable Anesthetic Mixtures flammable anesthetic mixture with air or oxygen or
nitrous oxide, per IEC 60601-1.
2. Routine Maintenance
Recommended Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Routine Safety and Operational Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Cleaning and Disinfecting the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Shutting Down the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Maintaining the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
About the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Viewing Battery Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Reconditioning the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Replacing a Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Software Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Registering for Software Upgrades from Philips InCenter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Upgrading the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Upgrading the Efficia Global Rack and Modules Firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Updating the Serial Number and Key Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Clearing Notifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Inspecting the Wall Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Contents-1
Efficia CM Series Service Guide
Temperature Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Cardiac Output Test (C.O.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Ground Integrity Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Touch Leakage Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Patient Leakage Current With Mains Voltage Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Nurse Call Relay Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Network Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
4. Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
When You Cannot Correct a Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Viewing System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Diagnosing a Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Start-up and Power Sequences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Troubleshooting Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Power Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Display Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Alarm Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
NBP Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Temperature Measurement Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
SpO2 Measurement Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Sidestream etCO2 Measurement Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Mainstream etCO2 Measurement Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
ECG/Respiration Measurement Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
IBP Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Cardiac Output Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Navigation Wheel, Touch Screen, and Key Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Recorder Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Nurse Call Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Security Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Full Disclosure SD Card Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Error Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Running System Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Running the Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Testing the Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Testing the Navigation Wheel and Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Testing the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Calibrating the Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Testing the Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Testing the LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-18
Managing Component Usage Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Resetting Component Usage Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Managing Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Formatting the SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
A. Theory of Operation
Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Front End Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Communications (I/O) Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Front Panel Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Power Supply Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Contents-2
Efficia CM Series Service Guide
Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Wireless Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Measurement Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
NBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Philips SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
NellCor SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
Masimo SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
IBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Respironics CapnoTrak Sidestream CO2 Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Respironics LoFlo Sidestream CO2 Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
Oridion Sidestream CO2 Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
Mainstream CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
ECG and ECG Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
Cardiac Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
B. Electromagnetic Compatibility
Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Reducing Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Restrictions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Emissions and Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Guidance and Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Contents-3
Efficia CM Series Service Guide
Contents-4
Efficia CM Series Service Guide
1
Introduction
Intended Audience
This guide is intended for biomedical engineers or technicians responsible for
maintaining and troubleshooting Philips patient monitoring systems.
For repair procedures and replacement parts, see the Efficia CM Series Patient
Monitors Repair Guide, which is available on Philips InCenter. For information about
creating an InCenter account, see “Registering for Software Upgrades from Philips
InCenter” on page 2-9.
For detailed information about the operation of the monitor, see the Efficia CM Series
Patient Monitors Instructions for Use. For information on configuration of monitoring
features, see the Efficia CM Series Patient Monitors Installation and Configuration
Guide.
Navigation Controls
You can navigate by using the touch screen or the navigation wheel. The touchscreen
is standard on the CM100, CM120, and CM150 monitors and is optional on the CM10
and CM12 monitors. All Efficia CM monitors include a navigation wheel.
Introduction
Efficia CM Series Service Guide 1-1
Efficia CM Documentation
The following table describes how to use the navigation controls on all Efficia CM
monitors.
Select Touch Touch a button, menu, or list item on the touch screen to select it.
Non-touch Rotate the wheel to highlight an item and then press the wheel to
select it.
Enter data Touch Touch the item to display the numeric keypad or the keyboard.
using the Touch the values, and then touch OK to close the keypad or
keyboard keyboard.
Scroll Touch Touch a scroll bar arrow to scroll through the list one segment at a
time. Touch and hold the arrow to quickly scroll through the list.
(You can also use the wheel to scroll and select items in the list.)
A scroll bar to the right of the list indicates the current location in
the list.
Touch an item to select it.
Non-touch Rotate the wheel to highlight a list and then press the wheel.
To scroll through the list, rotate the wheel.
To select items in the list, press the wheel.
A scroll bar to the right of the list indicates the current location in
the list.
Efficia CM Documentation
Efficia CM documentation includes:
• Efficia CM Series Patient Monitors Instructions for Use: Provides information for
day to day operation of the Efficia monitors. Also includes safety information,
monitor specifications, and compatible accessories.
Note — For information about purchasing additional copies of the Efficia CM Series
Patient Monitors Instructions for Use, contact the Philips Customer Care Center.
Introduction
1-2 Efficia CM Series Service Guide
Efficia CM Documentation
Introduction
Efficia CM Series Service Guide 1-3
Efficia CM Documentation
Introduction
1-4 Efficia CM Series Service Guide
2
Routine Maintenance
Recommended Frequency
Perform the maintenance procedures at the recommended frequency shown in the
following table. The maintenance procedures described in this guide should only be
performed by qualified service personnel.
Caution The frequency recommendations in the following table do not supersede local
requirements. Always perform locally required testing in addition to the testing in
this chapter.
Routine Maintenance
Efficia CM Series Service Guide 2-1
Routine Safety and Operational Checks
Caution • Ensure that plug-in modules and cables are properly inserted and latched before
returning a monitor to service.
• If the Gases module is installed, use caution when turning the water trap to
prevent injury by crushing or pressing fingers against the exhaust connector of
the module.
Routine Maintenance
2-2 Efficia CM Series Service Guide
Maintaining the Batteries
Step
Notes
• The Efficia CM patient monitors do not support recording if only one 3-cell
battery is installed.
• When two batteries are installed, Philips recommends using batteries from
the same manufacturer for optimal performance. When using mixed
batteries from different manufacturers, the battery reconditioning function
may not be optimal.
Note — For information about the battery status indicators, see the Efficia CM Series
Patient Monitors Instructions for Use.
Routine Maintenance
Efficia CM Series Service Guide 2-3
Maintaining the Batteries
Step
1 Open the Diagnostics Menu. See “Accessing the Diagnostics Menu and
Maintenance Options” on page 3-6.
Parameter Description
Cycle Count The number of cycle counts the battery has accrued
Routine Maintenance
2-4 Efficia CM Series Service Guide
Maintaining the Batteries
Parameter Description
Full Capacity The predicted capacity of the battery when it is fully
charged
Caution Do not disconnect the monitor from AC power or remove the battery during the
reconditioning process. If the monitor is disconnected, reconditioning stops and
must be restarted when the power is restored.
Step
1 Open the Diagnostics Menu. See “Accessing the Diagnostics Menu and
Maintenance Options” on page 3-6.
Routine Maintenance
Efficia CM Series Service Guide 2-5
Maintaining the Batteries
3 Select Recondition.
The monitor automatically discharges the battery or batteries.
6 When the second recharge cycle completes, the monitor displays the
following message:
Reconditioning Complete.
Note — If the battery does not recharge after four reconditioning cycles,
replace it.
Battery Replacement
Caution • Before you remove the battery, perform a monitor hard shutdown to ensure that
all patient and system data is saved in memory.
• Dispose of used batteries in an environmentally responsible manner. Do not
dispose of the battery in normal waste containers. Consult your hospital
administrator to find out about local arrangements.
Warning The risk of battery failure increases with age, when a battery remains in use past
300 charge-discharge cycles or 3 years. Such failures can result in overheating that
in rare cases can cause the battery to ignite.
Routine Maintenance
2-6 Efficia CM Series Service Guide
Maintaining the Batteries
The following table contains Efficia lithium ion battery part numbers.
3-cell 989803189981
9-cell 989803199221
Caution Store batteries at a room temperature of 25˚ C (77˚ F) or lower. Exposure to higher
temperatures may reduce battery life, damage the batteries, and degrade overall
monitor performance.
Step
1 Shut down the monitor. See “Shutting Down the Monitor” on page 2-2.
Slot
4 Hold one hand over the battery cover while you twist the screwdriver
slightly to release the battery cover from the case.
Warning
Twist the screwdriver slowly and gently to avoid expelling the battery
cover with enough force to cause personal injury.
Routine Maintenance
Efficia CM Series Service Guide 2-7
Maintaining the Batteries
Note — The monitor shown may be slightly different than your monitor.
Retaining Clip
6 Remove the battery from the monitor by pulling the attached black ribbon.
Ribbon
7 Insert the new battery and ensure the retaining clip is in place.
Note — Ensure that the end of the battery with the ribbon is facing out. The
battery is keyed so that it will fit only one way inside the monitor.
Note — Lithium-ion batteries are shipped with only a 30% charge. Charge battery
completely before first use of the patient monitor.
Routine Maintenance
2-8 Efficia CM Series Service Guide
Replacing a Fuse
Replacing a Fuse
To replace a fuse in the AC power connector:
Step
1 Shut down the monitor. See “Shutting Down the Monitor” on page 2-2.
2 Unlatch the power cord locking clip on the power entry module.
Remove the AC power cord from the rear of the monitor.
4 Use needle-nose pliers to pull the fuse holder out of the AC power
connector.
5 Replace the fuse or fuses, and then push the fuse holder back into the
power connector until it snaps into place.
Software Upgrades
Use the following procedure to upgrade the system software with a USB flash drive.
When you upgrade the software, the monitor saves all of the system settings, including
the Admin settings, the log data, and the patient record data.
Before you place an upgraded monitor back into service, Philips recommends that
your facility establishes a protocol to train users about any changes in the functionality
of the monitor resulting from the software upgrade. Upgrades with new features may
require review to determine if there are new default settings that need to change.
Routine Maintenance
Efficia CM Series Service Guide 2-9
Software Upgrades
Before you register, obtain the serial number of the monitor or monitors that you plan
to include in your registration. The serial number is located on the product
identification label on the rear of case of the monitor.
You must download the software to a USB flash drive with a capacity of 8GB or
greater. The USB flash drive must be formatted with file system FAT or FAT32. NTFS is
not supported.
To register:
Step
3 Under Software Updates, click Click here for Efficia CM patient monitor
account registration.
The Efficia InCenter Registration Form appears.
4 Enter your personal information and answer the questions, and then click
Submit.
After your information is processed, an email with temporary login
information is sent to the email address you entered. It may take 24 to 48
hours to receive the email.
Caution Before you upgrade the software, you can back up the system settings by exporting
the current configuration settings or by recording them on the worksheets. For
worksheets and more information, see the Installation and Configuration Guide
provided with your monitor.
Do not downgrade the monitor software to an earlier version, because it may cause
hardware incompatibility and loss of system settings and patient records. The
current monitor software version is displayed on the start-up screen and the System
Menu on the General tab.
When you upgrade the software:
• Charge the battery before upgrading the software.
• Never perform a software upgrade with the monitor connected to a patient.
• Disconnect any USB peripherals.
After the upgrade starts:
• Do not unplug the monitor.
• Do not remove the USB flash drive.
• Do not press any keys.
Routine Maintenance
2-10 Efficia CM Series Service Guide
Software Upgrades
Step
2 Insert the USB flash drive with the software upgrade into a USB port on
the monitor.
Note — The software upgrade folder must be located in the top directory
of the USB flash drive.
1 Select System ( ).
The System Menu appears.
7 After the monitor restarts for a second time, open the System Menu by
selecting the System Menu button
Routine Maintenance
Efficia CM Series Service Guide 2-11
Software Upgrades
Step
9 Click OK to acknowledge that the export was successful. If no USB flash drive was
installed, you will receive an error instructing you to check the USB stick.
Note — If you already exported an upgrade code to this flash drive, the
most recent code will be appended to the original file and an
acknowledgment message will not be displayed.
Routine Maintenance
2-12 Efficia CM Series Service Guide
Updating the Serial Number and Key Code
Step
1 Access the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
Serial number
Key code
Routine Maintenance
Efficia CM Series Service Guide 2-13
Clearing Notifications
Clearing Notifications
The Notifications pane contains the following notifications, Dashboard software
upgrade requests and Dashboard configuration update requests. If these requests are
not responded to, they must be cleared manually by using the Clear Data function. For
information about clearing data, see “To delete the logs:Clearing Patient Data” on
page 4-22.
Step
1 Verify that the wall mount is securely fastened to the wall and tighten the
screws.
2 Verify that the wall mount bracket is securely fastened to the monitor and
tighten the screws.
Routine Maintenance
2-14 Efficia CM Series Service Guide
3
Performance Verification Testing
Overview
This chapter includes information about periodic testing of the performance of the
monitor. If the monitor fails any test, it must be repaired before it is returned to use.
Note — The procedures in this chapter assume knowledge of basic monitor operation.
For details about using the monitor, see the Efficia CM Series Patient Monitors
Instructions for Use.
Recommended Frequency
Perform the test procedures at the recommended frequency outlined in the following
table.
Caution The frequency recommendations in the following table do not supersede local
requirements. Always perform locally required testing in addition to the testing
outlined in the table.
ECG Output Before first use, and then once every two
years.
LoFlo Sidestream CO2 accuracy Before first use, and then once yearly.
checks
CapnoTrak Sidestream CO2 Before first use, and then once yearly.
accuracy checks
Nurse call relay (if part of facility Before first use, and then once every two
protocols) years.
Safety
In accordance with IEC 60601-1
• Patient simulator
“NBP Test” on page 3-20 • A reference manometer (including hand
pump and valve), with an accuracy of 0.2%
• An expansion chamber (volume 250 ml ±
10%)
• Tubing
“IBP Test” on page 3-22 Patient simulator
“Temperature Test” on page 3-24 Patient simulator (with an accuracy of 0.1oC
or 0.2oF)
Test Recording
Authorized Philips personnel report test results to Philips to add to the product
development database. Hospital personnel, however, do not need to report results.
The following table describes what to record on the service record after you complete
the tests in this chapter.
NBP NBP:P/X1/X2/X3 or
NBP:F/X1/X2/X3
Notes
• The Admin tab in the System Menu is password-protected, and should only
be accessed by qualified service personnel.
• The procedures in this guide use the default Administrator password, 215. To
improve the security of the monitor settings, you can change the
administrator password after you install the monitor. For more information
about changing the password, see the Efficia CM Series Patient Monitors
Installation and Configuration Guide.
Caution The System Menu fills the entire monitor display when it is open and obscures any
active measurement data. When the Admin tab is open, it does not automatically
close if left unattended, but remains open until it is closed manually.
Step
1 Select System ( ).
The System Menu appears.
Step
Caution
The password-protected tabs remain unlocked for 1 minute after you
close them. This allows you to re-open the tab without having to
re-enter the password. Do not leave the monitor unattended during
the unlock time.
The following options are described in this guide. All other options are
described in the Efficia CM Series Patient Monitors Installation and
Configuration Guide.
• Demo Mode — Demo Mode allows you to demonstrate the monitor
without actually monitoring parameters.
• Upgrade Software — Opens the Upgrade Software menu. For more
information, see “Software Upgrades” on page 2-9.
• Diagnostics — Opens the Diagnostics Menu. Monitoring is suspended
while this menu is open. For more information, see “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
Note — The Diagnostics button is unavailable when the monitor is in
Demo mode.
• View Logs — Opens the View Log menu, which allows viewing and
management of system logs. For more information, see “Managing
Logs” on page 4-20
Caution When you open the Diagnostics Menu, monitoring is suspended and measurements
stop.
Notes
• The Charging LED does not show battery status while the Diagnostics Menu
is open.
• The Diagnostics Menu is not available when the monitor is in Demo mode.
Use the Diagnostics Menu to perform the system diagnostic and verification tests and
track component usage hours.
The Diagnostics Menu allows you to:
• Configure the hardware and software;
• View monitor and rack status and errors;
• Perform measurement and monitoring functions;
• View connected plug-ins’ status and errors;
• Enter the system serial number if the monitor, rack, or plug-in module boards
have been replaced.
Note — You cannot access the Diagnostics Menu when the monitor is in Demo mode.
Step
1 Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
6 Select OK.
The Diagnostics Menu displays additional options.
Warning Do not connect a patient to a monitor running in Demo mode. Values represented in
Demo mode do not represent measurements from a patient connected to the
monitor, and may lead to incorrect diagnoses.
Demo mode is used to demonstrate the monitor without connecting the monitor to an
actual person. Demo mode simulates all patient parameters and generates alarms
when alarm settings are exceeded.
Step
1 Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
4 Select OK.
The monitor enters Demo mode and clears all patient data. DEMO
appears on the top of the monitor display.
5 To exit Demo mode, press the On/Standby key to turn the monitor off.
The monitor clears all simulated patient data.
Visual Test
To perform the visual test:
Step
1 Inspect the system for obvious signs of damage; for example, cracks, cuts,
or breakage.
2 Check all external cables and accessories for damage; for example, cuts,
kinks, wrong connections.
Step
The expected result is that the monitor starts up and displays the main
screen. For detailed information about the start-up and power
sequences, see “Start-up and Power Sequences” on page 4-2
If the LEDs do not function as expected, see “Power Problems” on
page 4-5.
If the display does not function as expected, see “Power Problems” on
page 4-5 or “Display Problems” on page 4-5.
If you do not hear a startup tone, or the monitor displays the Speaker
Malfunc error message, see “Error Codes” on page 4-13.
Note — Philips employees record this value as PO:P or PO:F.
Alarms Test
The following test allows you to verify that the visual and audio alarms are working.
To perform this test, you need an NBP cuff and hose.
Step
1 With the monitor turned on, make sure that all alarms are enabled (the
monitor is not in Audio Pause or Audio Off mode).
2 Make sure the NBP alarm is enabled (the crossed bell icon does not
appear in the NBP numeric pane).
3 Connect the NBP hose to the NBP input connector, but do not place the
cuff on your arm.
5 Wait for the NBP module to cycle and check that an NBP error message is
displayed in the message area and an alarm tone sounds.
6 If you do not get the results in step 5, see “NBP Problems” on page 4-6.
Note — Philips employees record this value as AL:P or AL:F.
SpO2 Test
The following procedure tests the performance of the SpO2 measurement and alarm.
To perform this test, you need an adult SpO2 sensor.
To perform the SpO2 test:
Step
3 Verify that the SpO2 value displayed on the monitor is between 95% and
100%. If it is not, try the test again with a patient simulator.
4 If you still do not get the results in step 3, see “SpO2 Measurement
Problems” on page 4-7.
Note — Philips employees record this value as SpO2:P or SpO2:F.
Caution A functional tester cannot be used to assess the accuracy of a pulse oximeter
monitor. However, if there is independent demonstration that a particular
calibration curve is accurate for the combination of a pulse oximeter monitor and a
pulse oximeter sensor, then a functional tester can measure the contribution of a
monitor to the total error of a monitor/sensor system. The functional tester can
then measure how accurately a particular pulse oximeter monitor is reproducing
that calibration curve.
The CO2 Test menu also displays the following information about the Sidestream CO2
module:
• Hardware version
• Firmware version
• Last calibration date
• CO2 usage hours
• Hours to calibration
• Hours to service
Caution If the initial calibration is performed before 720 hours of use, the module resets to
require the next calibration after 1,200 hours instead of after 4,000 hours.
This test uses calibration equipment that you can order. Refer to the documentation
accompanying the calibration equipment for more details.
Warning A monitor that is not calibrated at the recommended intervals may provide
inaccurate results.
Dispose of empty or partially empty calibration gas containers in accordance with
the manufacturer’s instructions and local regulations.
Step
1 Ensure that the CO2 module is not in Standby mode (when the monitor is
in Display Waveform mode, Standby is displayed in the CO2 waveform
pane).
If necessary, turn on the CO2 module by opening the CO2 Menu and
setting CO2 On/Standby: to On.
2 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
4 Connect one end of the sampling line to the CO2 inlet on the monitor and
leave the other end unconnected.
6 Connect the other end of the sampling line to the gas controller
equipment inlet with a 5% gas concentration.
7 Open the valve on the gas controller equipment to allow 5% CO2 gas to
flow into the monitor.
Caution Because the monitor uses the current date and time as the calibration date and
time, ensure that the system date and time is set correctly.
Note — The monitor always displays the CO2 module’s operating (usage) time in the
CO2 Test menu.
Step
1 Connect one end of the sampling line (if not already connected) to the
CO2 inlet on the monitor and leave the other end unconnected.
3 Connect the other end of the sampling line to the gas controller
equipment inlet with a 5% gas concentration.
4 Open the valve on the gas controller equipment to allow 5% CO2 gas to
flow into the monitor.
1 Ensure that CO2 is not in Standby mode (when the monitor is in Display
Waveform mode, Standby is displayed in the CO2 waveform pane).
If necessary, turn on the CO2 module by opening the CO2 Menu and
setting CO2 On/Standby: to On.
2 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
4 Attach a water filter assembly and large airway adapter to the sampling
inlet of the module.
Sensor warming up. is displayed.
Wait for warm-up to complete.
6 When the Zero button is enabled, select it and wait for the zero to
complete.
If the zero fails (“Zero Error” or “Zero Required” is present after the zero):
• Ensure the “Module Not Ready to Zero” status is not present.
• Ensure that the sampling accessory is away from any sources of CO2.
• Repeat the zero process. If the zero fails again, contact service
7 Connect the gas mixture to the adapter and set the flow rate to 2.0 l/min.
8 Allow one minute for the gas mixture to stabilize and observe the CO2
value. The expected value is 5% ± 0.3%.
12 Select the Stop Accuracy Check button at any time to stop the test.
Note — Philips employees record this value as SCO2:P or SCO2:F.
The CapnoTrak module’s flow rate and cell pressure accuracy should be verified every
12 months by direct measurement using a calibrated flow meter.
Note — The test must be performed under load. Use the CapnoTrak Adult Airway
Adapter for proper test verification.
2 Attach a water filter assembly and adult airway adapter to the sampling
inlet of the CapnoTrak module.
5 Verify that the flow rate is 50 ml/min ± 10 ml/min. If the measured flow
rate is outside the specified limits, remove the CapnoTrak module from
use and contact service. A mass flow meter is typically used to verify the
flow rate (1.5% full scale accuracy at 100 ml/min).
4 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
6 Read the cell pressure value from the screen. Verify that the pressure is
between -20 to -30 mmHg.
7 If the pressures are not within range, refer the module for service.
1 Ensure that CO2 is not in Standby mode (when the monitor is in Display
Waveform mode, Standby is displayed in the CO2 waveform pane).
If necessary, turn on the CO2 module by opening the CO2 Menu and
setting CO2 On/Standby: to On.
2 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
5 Verify that the Amb. Pressure (mmHg) and Gas Temperature (°C) values are
accurate.
If the values are incorrect, enter the correct values.
6 Set the gas compensation settings of the LoFlo CO2 Module to the
verification gas mixture.
7 Zero the LoFlo CO2 Module on the airway adapter being used in this test.
9 Allow 10 seconds for the gas mixture to stabilize and observe the CO2
value. The expected value is 5% ± 0.3%.
11 The accuracy check is now complete. Remember to set the LoFlo module
settings for units and gas composition back to their previous settings.
12 Select the Stop Accuracy Check button at any time to stop the test.
Note — Philips employees record this value as SCO2:P or SCO2:F.
The LoFlo module’s flow rate accuracy should be verified every 12 months by direct
measurement using a calibrated flow meter.
Note — The test must be performed under load. Use the LoFlo Adult On-Airway
Adapter Kit for proper test verification.
2 Connect the RS232 output of the power supply to the Efficia monitor.
3 Attach the LoFlo Adult On-Airway adapter kit to the sample receptacle of
the LoFlo module. A click will be heard when properly inserted.
6 Verify that the flow rate is 50 ml/min ± 10 ml/min. If the measured flow
rate is outside the specified limits, remove the LoFlo module from use and
contact service.
1 Ensure that CO2 is not in Standby mode (when the monitor is in Display
Waveform mode, Standby is displayed in the CO2 waveform pane).
If necessary, turn on the CO2 module by opening the CO2 Menu and
setting CO2 On/Standby: to On.
2 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
6 When the Zero button is enabled, select it and wait for the zero to
complete.
7 Connect the gas mixture to the adapter and set the flow rate to 2.0 – 5.0
l/min.
11 Select the Stop Accuracy Check button at any time to stop the test.
Note — Philips employees record this value as MCO2:P or MCO2:F.
Step
1 Connect the ECG patient cable to the ECG connector on the side of the
monitor.
3 Connect the other end of the ECG patient cable to the patient simulator.
4 Verify that the displayed ECG wave and HR value (± 2 bpm) correlate to
the values you selected on the patient simulator. For an HR of 80 bpm, the
displayed value on the patient monitor should be between 78 – 82 bpm.
5 Redo the test, selecting several heart rates on the patient simulator: 160,
80, 100.
Step
NBP Test
These tests check the performance of the non-invasive blood pressure measurement:
• NBP accuracy
• NBP calibration procedure (if required)
• NBP pneumatic leakage
To perform these tests, you need:
• A reference manometer (includes hand pump and valve), accuracy 0.2% of
reading
• An expansion chamber (volume 250 ml 10%)
• Appropriate tubing
NBP Accuracy
Note — Philips recommends that you do not test NBP accuracy with a simulator.
Step
1 Connect the manometer and the pump with tubing to the NBP connector
on the monitor.
5 Select the Start Static Pressure Test button to start the test.
Step
7 Wait 10 seconds for the pressure to stabilize. Note the pressure displayed
in the NBP Test menu. The value should be 280 mmHg 3 mmHg.
Note — Philips employees record this value as X1.
8 Squeeze the manometer pump to apply a pressure of 150 mmHg to the
monitor.
9 Wait 10 seconds for the pressure to stabilize. Note the pressure displayed
in the NBP Test menu. The value should be 150 mmHg 3 mmHg.
Note — Philips employees record this value as X2.
10 Select the Stop Static Pressure Test button to stop the test.
Step
4 Select the Set Pressure Value field, and then select the value to match
that displayed on the manometer.
Step
8 To verify the calibration, check the accuracy of the NBP. See “NBP
Accuracy” on page 3-20.
9 If you do not get the expected results after several tries, see
Chapter 4, “Troubleshooting.”
Step
1 In the NBP Test menu, select the Start Static Pressure Test button.
3 Wait 10 seconds for the pressure to stabilize. Note the pressure value in
the NBP Test menu.
Note — Philips employees record this value as P1.
4 Wait 60 seconds for the pressure to stabilize. Note the pressure value in
the NBP Test menu.
Note — Philips employees record this value as P2.
5 Calculate and document the leakage test value. The leakage test value
should be less than or equal to 6 mmHg.
Note — Philips employees record this value as X3 (where X3 = P1 - P2).
6 Select the Stop Static Pressure Test to stop the process.
7 If the leakage test value exceeds 6 mmHg, check the test setup cuff and
tubing, and then test again. If the test still fails, check the pneumatic
tubing inside the monitor.
IBP Test
This test checks the performance of the invasive blood pressure measurement.
To perform this test, you need a patient simulator such as the Foss Systems BP28.
To perform the IBP test:
Step
2 Open the Blood Pressure menu and make sure that the Display IBP check
box is selected
Make sure that the IBP Label option selected is the same as the IBP
measurement type selected on the patient simulator.
3 Ensure that the IBP waveform displays in one of the sectors on the
monitor. ? appears in the numeric IBP pane.
5 Select the IBP waveform and then press the wheel to open the <Pressure>
Waveform Menu.
6 Select the IBP Zero button and then press the wheel to begin zeroing the
pressure.
The <Pressure> Waveform Menu closes. The message, IBP Zero - In
Progress, appears in the IBP Waveform pane.
7 Configure the patient simulator as P (static) = 200 mmHg.
Note — To configure the patient simulator to 200 mmHg, use the blood
pressure menu on the monitor to set 0 – 300 on the scale menu.
10 Wait for the displayed waveform and make sure that the IBP waveform
corresponds to a steady pressure of 200 mmHg ± 5 mmHg.
12 If you still do not get the expected results, see “IBP Problems” on
page 4-10.
Note — Philips employees record this value as IBP(1 or 2):P or IBP(1 or 2):F.
Step
1 Connect the patient simulator to the ECG connector on the monitor and
the defibrillator to the ECG output on the monitor.
Temperature Test
To perform this test, you need a patient simulator (with 0.1oC or 0.2oF). Perform the
following test for each temperature connector in use.
To test the performance of the temperature measurement:
Step
Step
1 Connect the patient simulator to the C.O. connector on the side of the
monitor.
6 Verify that the Computation Constant is the same setting as the simulator.
10 Connect one side or the service tool to the injectate receptacle of the C.O.
interface cable and the other side to the catheter cable receptacle.
Safety Tests
Use the following safety test procedures to verify safe installation or service of the
monitor. The setups and the acceptable ranges of values used for these tests are
derived from local and international standards, but may not be equivalent. These tests
are not a substitute for local safety testing where it is required for an installation or a
service event. If you use the Metron Safety tester, perform the tests in accordance with
your local regulations. For example, in Europe, use IEC 60601-1 and in the United
States, use ES 60601-1. The Metron Report should print results with the names listed
below, together with other data.
Note — Safety tests meet the standards of, and are performed in accordance with
IEC 60601-1, Clause 8 (EN60601-1).
Figure 3-3 S(2) Part 2: Touch Leakage Current, Single Fault (Open Earth)
Medical Electrical
Medical
S5
This measures leakage current of exposed metal parts of Instrument under Test (IUT)
within a system with Protective Earth (PE) normal condition (S4 = closed) and between
parts of the system within the patient environment; normal and reversed polarity using
S2 with S5 closed/open for each in turn of “Other instrument” making up the medical
electrical system.
Expected Test Results:
Normal condition maximum leakage current x1 100A
Medical Electrical
S5
This measures leakage current of exposed metal parts of Instrument under Test (IUT)
within a system with Protective Earth (PE) open circuit (S4 = open) and between parts
of the system within the patient environment; normal and reversed polarity using S2 in
turn for all other equipment that is connected to the medical instrument under test. S5
closed/open for each in-turn of "Other Instrument" making up the medical electrical
system.
Expected Test Results:
Single Fault maximum leakage current x2 500A (IEC 60601-1)
Performance Verification Testing
3-28 Efficia CM Series Service Guide
Nurse Call Relay Test
Note — To test the patient leakage current of the Mainstream CO2 connector, if
present, connect the cable to the monitor. Wrap the sensor with foil and
connect it to the safety test equipment. Then test the Mainstream CO2 as an
applied part.
Tip Sleeve
Ring
Step
1 Plug the phono connector into the Nurse Call connector on the back of
the monitor.
3 If the nurse call function does not work as expected, see “Nurse Call
Problems” on page 4-12
Note — Philips employees record this value as NC:P or NC:F.
Network Test
If your Efficia monitor is equipped with a wireless LAN module, perform the following
procedure to test the module.
Step
1 Open the Diagnostics Menu. See “Accessing the Diagnostics Menu and
Maintenance Options” on page 3-6.
3 Select Diagnostics. The status of the wireless module will be checked, and
information displayed.
Use the information in this chapter to diagnose and correct monitor problems. This
chapter describes how to troubleshoot a monitor that is not operating correctly. To
perform any recommended repairs, see the repair guide for your monitor.
The repair guide is available on Philips InCenter. For information about creating an
InCenter account, see “Registering for Software Upgrades from Philips InCenter” on
page 2-9.
Overview
You can repair the monitor in either of two ways:
• Return the monitor to a Philips authorized service center for repair.
• Purchase spare parts from a Philips authorized service center.
Note — If you have created a custom password for your institution, you must change
the password to the Philips default (215) before you return the monitor for any
reason.
Troubleshooting
Efficia CM Series Service Guide 4-1
Diagnosing a Problem
Diagnosing a Problem
Before you begin to troubleshoot a problem or open the monitor for repair, check the
following basics:
1. Is the power switch turned on?
2. Is the battery adequately charged?
3. Is the AC power cord connected to the monitor and plugged into a functional AC
outlet?
4. Is the monitor's display functioning?
5. Are the LEDs on the front of the monitor lit as you expect?
Notes
• It may take several seconds for the AC Power LED to light or turn off after
the AC power cord has been connected or disconnected.
• It may take several seconds for the battery status icon to update and
appear.
Troubleshooting
4-2 Efficia CM Series Service Guide
Start-up and Power Sequences
For the monitor to start correctly, it must be powered correctly as indicated by the lit
LEDs on the front panel.
Connect the AC power cord or The color bar is displayed and then the LCD turns
insert a battery or batteries (or off.
both) into an unpowered The Charging LED turns on (when a battery is
monitor. present).
Press the On/Standby key. A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
The Date/Time Menu appears to allow the
configuring of date and time.
After you acknowledge the Date/Time Menu, the
main screen appears.
Press the On/Standby key The color bar is displayed and then the LCD turns
when the monitor (with AC off.
power and/or battery power) is The Charging LED turns on (when a battery is
off. installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
If configured, the Admit Patient Menu appears.
After you acknowledge the Admit Patient Menu,
the main screen appears.
Software reset (system or user). The color bar is displayed and then the LCD turns
off.
The Charging LED turns on (when a battery is
installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
The main screen appears.
Plug AC power cord into a The color bar is displayed and then the LCD turns
monitor that is off (battery off.
only). The Charging LED turns on (when a battery is
Press the On/Standby key. installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
If configured, the Admit Patient Menu appears.
After you acknowledge the Admit Patient Menu,
the main screen appears.
Troubleshooting
Efficia CM Series Service Guide 4-3
Troubleshooting Tables
Press the On/Standby key The Philips splash screen appears and the LCD
when the monitor is on (with AC turns off.
power and/or battery power).
Select Shut Down on the After you select OK, the Philips splash screen
General tab in the System Menu. appears while data is being saved. The LCD turns
off.
Troubleshooting Tables
Use the following tables to diagnose and fix monitor problems. The tables describe a
monitor problem by symptom, list a possible cause, and suggest actions. Check the
first possible cause listed, and then perform the associated action to repair the
problem. Perform all actions in the order that they appear in a table. If you cannot
resolve the problem by following the actions in the table, contact your service provider
for repair.
Monitor problems are categorized as follows:
• Power
• Display
• Alarms
• Measurements, including:
– NBP
– Temperature
– SpO2
– CO2
– ECG/Respiration
– IBP
– C.O.
• Navigation wheel, touch screen, and keys
• Recorder
• Nurse Call
• Security
• SD Card
If using the following troubleshooting tables does not solve the problem, see “Running
System Diagnostics” on page 4-13.
Troubleshooting
4-4 Efficia CM Series Service Guide
Troubleshooting Tables
Power Problems
Display Problems
Alarm Problems
Troubleshooting
Efficia CM Series Service Guide 4-5
Troubleshooting Tables
NBP Problems
Troubleshooting
4-6 Efficia CM Series Service Guide
Troubleshooting Tables
Troubleshooting
Efficia CM Series Service Guide 4-7
Troubleshooting Tables
Troubleshooting
4-8 Efficia CM Series Service Guide
Troubleshooting Tables
Troubleshooting
Efficia CM Series Service Guide 4-9
Troubleshooting Tables
IBP Problems
Troubleshooting
4-10 Efficia CM Series Service Guide
Troubleshooting Tables
Troubleshooting
Efficia CM Series Service Guide 4-11
Troubleshooting Tables
Recorder Problems
Notes
• The Unsupported Battery
Config message will not be
triggered if a 9-cell battery
is installed.
• In a CM150, the message
will be triggered if 3-cell
batteries are installed.
Troubleshooting
4-12 Efficia CM Series Service Guide
Error Codes
Security Problems
Error Codes
When the monitor detects an error condition (depending on the type of error), an error
message appears on the screen or an error code is written to the Error Log. An error
code is an indication that the monitor detected an error in operation and service may
be required. An error code is a hexidecimal number that indicates the nature of the
error.
If an error occurs during monitoring, an audible alarm also sounds. Press the Alarm
Silence key to end the audible alarm. For detailed information on error codes, see the
Efficia CM Series Patient Monitors Repair Guide.
Note — Accessing the Maintenance options requires a password that is different from
the Admin Password. Only trained biomedical engineers or support persons
should access the Maintenance options and perform maintenance and repair.
Troubleshooting
Efficia CM Series Service Guide 4-13
Running the Self Test
Note — When the monitor is powered up from Standby mode, the monitor performs the
front end tests in addition to the start-up sequence. When the monitor is
powered up from Deep Sleep mode, the monitor performs the complete
memory test and front end tests.
Step
Step
Troubleshooting
4-14 Efficia CM Series Service Guide
Testing the Navigation Wheel and Keys
Step
Step
Troubleshooting
Efficia CM Series Service Guide 4-15
Testing the Display
Step
1 In the Diagnostics Menu, select the Display Test button to start the test.
The display test draws a sequence of patterns on the monitor screen.
Each pattern displays for five seconds. When the last pattern is drawn,
the sequence starts again from the first pattern. The following describes
the test patterns displayed:
Caution Use your finger to perform the test. Do not use a sharp object that could scratch the
screen.
Troubleshooting
4-16 Efficia CM Series Service Guide
Testing the Speaker
Step
1 Open the Maintenance options in the Diagnostics Menu. See “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
3
Caution
Touch and release each target and wait for the target to move to the
next position before touching it. Be careful not to touch any target more
than once, because it may adversely effect the calibration.
Touch and release the center of the target cross hair ( ) that appears.
The target moves to the next position, which is the top right corner.
4 Repeat step 3 for the remaining target cross hairs. The target moves
clockwise to each corner and then to the center.
After the center target is touched and released, a confirmation window
appears.
5 Select OK.
Caution
If the confirmation screen does not appear within a few seconds after
the center calibration target disappears, you must restart the monitor
and recalibrate the touchscreen.
Note — The CM10 and CM12 models have one speaker. The CM100, CM120, and CM150
have two speakers.
Troubleshooting
Efficia CM Series Service Guide 4-17
Testing the LEDs
Step
1 In the Diagnostics menu, select the Audio Test button to start the test.
A window appears confirming that the monitor is entering the audio test
and displays the text: Listen for three short tones. If your monitor has two
speakers, the three tones will repeat.
2 Verify that three short tones sound for each speaker.
When the test completes, the window closes.
Step
1 In the Diagnostics menu, select the LED Test button to start the test.
Troubleshooting
4-18 Efficia CM Series Service Guide
Managing Component Usage Parameters
Parameter Description
NBP Cycle Count The number of NBP measurements taken by the monitor.
LCD Usage Hours The number of hours that the LCD screen has been on.
Note — CO2 usage information is displayed in the CO2 Test menu, accessed through
the Diagnostics Menu. See Chapter 3, “Performance Verification Testing.”
Caution Always reset a component usage parameter after replacing the associated part.
Step
Troubleshooting
Efficia CM Series Service Guide 4-19
Managing Logs
Managing Logs
The monitor collects data in the logs described in the following table.
Caution To avoid loss of data, export the log contents before the log becomes full.
Log Description
Warning
The alarm logs are retrievable after a controlled
shutdown. Manually performing the shutdown
procedure or allowing the battery to completely
discharge are considered controlled shutdowns.
However, the loss of AC power without battery backup
will result in the loss of the alarm logs.
Troubleshooting
4-20 Efficia CM Series Service Guide
Managing Logs
Log Description
Network Log The following events are logged:
• LAN cable connect and disconnect
• WLAN connect and disconnect
• Central connect and disconnect
• Dashboard connect and disconnect
• EMR server connect and disconnect
• SNTP server connect and disconnect
The log contains up to 500 entries. The oldest entry is
deleted when full capacity is reached.
The exported log is named NetworkActivity.csv.
For detailed information about errors and error codes,
see “Error Codes” on page 4-13.
Error Log The following events are logged:
• All hardware malfunctions
• Watch dog resetting
• Loss of monitoring
The log contains up to 500 entries. The oldest entry is
deleted when full capacity is reached.
The exported log is named Error.csv.
The log entries include a timestamp that indicates when the error occurred. You can
view, export, or delete the logs.
Monitor, rack and plug-in module HW Error Logs are automatically stored in internal
memory. A summary report can be downloaded to a USB flash drive.
To view the logs:
Step
1 Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
4 To view the entire list of errors, select and drag the scrollbar.
1 Insert a compatible USB flash drive into any USB port on the monitor.
2 Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
Troubleshooting
Efficia CM Series Service Guide 4-21
Managing Logs
1 Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
Patient data and a summary are automatically stored in internal memory. A summary
report can be downloaded to a USB flash drive.
If you return a monitor for service, you must first clear the patient data. The Clear Data
function resets the user settings to the factory defaults and clears the following:
• All patient data
• All profiles and profile settings
• All logs
• All cloneable settings, except the AC line frequency
• Monitor name
The following settings are not cleared:
• Monitor serial number
• Key code
• LAN/WLAN MAC Address
• Touchscreen, temperature, IBP, and C.O. calibration data
• Language
• NBP cycle count
• LCD usage time
• Time zone
• AC line frequency
• RTC calibration data
• Country code
Troubleshooting
4-22 Efficia CM Series Service Guide
Formatting the SD Card
Step
Step
Troubleshooting
Efficia CM Series Service Guide 4-23
Formatting the SD Card
Troubleshooting
4-24 Efficia CM Series Service Guide
A
Theory of Operation
The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are indicated for
use by healthcare professionals whenever there is a need for monitoring the
physiological parameters of patients.The Efficia CM10, CM12, CM100, CM120 and
CM150 patient monitors are intended for monitoring, analysis, recording, and alarming
of multiple physiological parameters in healthcare environments. Additionally, the
monitors may be used in transport situations within a healthcare facility.The monitor is
powered from either the external AC input or its internal battery or batteries.
The Efficia monitors are available with 10-inch, 12-inch, and 15-inch screen sizes. Each
model has purchasable options available.
The following features, some optional, are available:
• 3-lead, 5-lead, and 10-lead ECG lead sets
• ECG analog output
• Dual IBP
• Cardiac output
• SpO2 (Philips, NellCor or Masimo)
• Impedance respiration
• Respiration — RRa
• Respiration —awRR
• NBP
• DualTemperature
• etCO2 (Mainstream or Sidestream)
• Recorder
• Touch screen
• Nurse call output
• LED backlight LCD display
• Li-ion battery
• Recorder
• Internal radio for connectivity to EHR system
• External ports: USB, LAN, VGA
In addition to monitoring and displaying the status of physiological parameters, the
monitor performs various microprocessor-programmed analytical functions, including:
• Creating visual and audible alarm signals when alarm settings are violated
• Creating and displaying technical alarms when conditions are detected that
would degrade or prevent valid measurements
• Creating and displaying trend waveforms or tabular data
• Providing input to an optional recorder for printout of current or trend waveforms
or tabular data
Theory of Operation
Efficia CM Series Service Guide A-1
Communications Board
Speaker(s) LAN
USB
USB
Main Board
Nurse Call
Wireless Module SD card VGA
SpO
Front End Assembly
2
Microprocessor Front Panel Assembly
Front End Interface
Patient Connectors
Battery Connector
Monitor Keypad
Front Panel
Board
Temp (2) Temp (2) Battery
(1 or 2)
Power management regulator,
Battery charger,
Touchscreen controller
Theory of Operation
2
Patient Isolation Module
Block Diagram
NBP
Module AC Power Connector
and Fuses
Block Diagram
A-2
Block Diagram
Main Board
The main board contains the following components:
• Microprocessor
• Flash memory
• DDR SDRAM
• LAN controller
• Power management regulator
• Front end interface
• Battery charger
• Touchscreen controller
• VGA DAC
• Audio
• SD card connector
The nurse call alarm contacts are available through a 3.5 mm miniature insulated phone
jack on the back of the monitor. The alarm contacts close (open) for any alarm condition
detected by the monitor.
The alarm contacts are the three switching signals associated with a single pole, double
throw relay:
• Normally open contact (NO)
• Common contact
• Normally closed contact (NC)
Contact Rating:
NO = 1A @ < 25 VAC
Common contact = 1A @ < 60 VDC
Isolation of 1.5 KV
Theory of Operation
Efficia CM Series Service Guide A-3
Block Diagram
The monitor has one external standard VGA port to transmit the displayed monitor
images to an external display.
Navigation Wheel
The navigation wheel is a rotating, push-switch wheel. The associated wheel circuitry
generates a pulse when pressed and generates a digitally encoded pair of quadrature
signals whose relative magnitudes and polarities represent the angular position of the
wheel. These outputs are connected to the microprocessor where they are interpreted
as required for the functions involved. Successive angular positions determine the
direction of wheel rotation. In addition to the functions performed in conjunction with
the keypad, the wheel operates in conjunction with the display to select menus and
lists of parameter variables.
Power Management
Theory of Operation
A-4 Efficia CM Series Service Guide
Measurement Modules
Speaker
The monitor is equipped with one or two speakers. The speaker is driven by an
amplifier and Audio CODEC that interfaces to the microprocessor. The speaker
provides the monitor’s audio during alarm conditions. The microprocessor outputs
different audio tones specified by different alarm priorities and conditions.
Note — The CM10 and CM12 models have one speaker. The CM100, CM120, and CM150
have two speakers.
Recorder
The recorder contains the following components:
• Digital thermal head array assembly
• Printing width
– 48 mm on 58 mm paper
– Resolution:
Time axis = 16 dots/mm at the speed of 25mm/sec, 8 dots/mm at 50
mm/sec
Voltage axis = 8 dots/mm
• Motor (paper drive) and driver circuitry
• Paper out/door open detector
• Selectable print speed
– 6.25, 12.5, 25, 50 mm/s
Print head control logic is implemented on the recorder board. The host
microprocessor sends data to the recorder through a local serial bus.
Wireless Module
The wireless module includes a radio board and an internal PCB antenna. The module
provides wireless connectivity for the monitor. The radio board interfaces with the
main board through a 60-pin connector. The antenna interfaces with the radio board
through a Hirose U.FL connector. The module supports all 802.11a, 802.11b, and 802.11g
data rates and automatically adjusts data rates and operational modes based on
various environmental factors. It also supports MCS 0-7 rates for 802.11n.
Measurement Modules
The monitor includes the measurement modules described in this section, some of
which are optional.
NBP Module
The NBP assembly contain these elements:
• Pump
• Valves (2-valve system of dump and safety valve)
Theory of Operation
Efficia CM Series Service Guide A-5
Measurement Modules
• Manifold
• Filter
• Pressure measurement and control circuitry
Pressure data is converted to digital format and conveyed to the processor. The NBP
pump uses an oscillometric method that employs stepwise pressure deflation. Pump
software eliminates most ambient noise and motion interference. Applications are
neonatal, pediatric and adult patients. The blood pressure range is as follows:
• Systolic, 30 – 270 mmHg
• Diastolic, 10 – 245 mmHg
• Mean, 20 – 255 mmHg.
Accuracy is ±5 mmHg, with standard deviation no greater than 8 mmHg.
NBP processing uses an oscillometric technique to provide needed measurements at
selected intervals. This technique uses an inflatable sphygmomanometer cuff similar to
those used by clinicians in routine measurements. A motorized pump inflates the cuff
to approximately 160 mmHg (adult mode) initially, at which point the pressure
effectively stops the flow of blood. Then, under monitor control, the pressure in the
cuff is gradually reduced, while a pressure transducer detects the pressure and
transmits the parameter signal to the NBP input circuitry. As the pressure is reduced,
blood flows in the previously occluded artery, and changes the measurements made
by the transducer. The point at which oscillation increases sharply is defined as
systolic pressure. As the cuff continues to deflate, the oscillation amplitude increases
to a maximum, and then decreases. The peak oscillation amplitude is defined as the
mean arterial pressure. The point at which the system detects a rapid decrease in
oscillation is defined as the diastolic pressure.
Philips SpO2
The SpO2 module measures oxygen saturation in the blood using an optical
spectrophotometry technique. It is based on the differences in the absorption of red
and Infrared light by oxygenated and deoxygenated Hemoglobin.
A pulse oximeter passes red and infrared light into an arteriolar bed and measures
changes in light absorption during the pulsatile cycle. The light sources are red and
infrared light emitting diodes (LEDs). A photo diode provides detection.
To identify the oxygen saturation of arterial hemoglobin, the monitor uses the pulsatile
nature of arterial flow. During systole, a new pulse of arterial blood enters the vacuolar
bed, and both blood volume and light absorption increase. During diastole, blood
volume and light absorption reach their lowest point. The measurement is based on
the difference between maximum and minimum absorption, focusing on the pulsatile
arterial blood. In addition to the oximetry function, the measurement can also provide
the heart rate.
NellCor SpO2
The monitoring system uses pulse oximetry to measure functional oxygen saturation in
the blood. Pulse oximetry works by applying a NellCor™ sensor to a pulsating arteriolar
vascular bed, such as a finger or toe. The sensor contains a dual light source and a
photodetector.
Bone, tissue, pigmentation, and venous vessels normally absorb a constant amount of
light over time. The arteriolar bed normally pulsates and absorbs variable amounts of
Theory of Operation
A-6 Efficia CM Series Service Guide
Measurement Modules
light during the pulsations. The ratio of light absorbed is translated into a measurement
of functional oxygen saturation (SpO2).
Ambient conditions, sensor application, and patient conditions can influence the
ability of the monitoring system to accurately measure SpO2.
Pulse oximetry is based on two principles: oxyhemoglobin and deoxyhemoglobin
differ in their absorption of red and infrared light (measured using spectrophotometry),
and the volume of arterial blood in tissue (and hence, light absorption by that blood)
changes during the pulse (registered using plethysmography). A monitoring system
determines SpO2 by passing red and infrared light into an arteriolar bed and measuring
changes in light absorption during the pulsatile cycle. Red and infrared low-voltage
light-emitting diodes (LED) in the sensor serve as light sources; a photo diode serves
as the photo detector.
Since oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of
red and infrared light absorbed by blood is related to hemoglobin oxygen saturation.
The monitoring system uses the pulsatile nature of arterial flow to identify the oxygen
saturation of arterial hemoglobin. During systole, a new pulse of arterial blood enters
the vascular bed, and blood volume and light absorption increase. During diastole,
blood volume and light absorption reach their lowest point. The monitoring system
bases its SpO2 measurements on the difference between maximum and minimum
absorption (measurements at systole and diastole). By doing so, it focuses on light
absorption by pulsatile arterial blood, eliminating the effects of nonpulsatile absorbers
such as tissue, bone, and venous blood.
Masimo SpO2
The Masimo SpO2 module contains a Masimo® SET board with a rainbow option that
uses a multi-wavelength sensor to distinguish between oxygenated blood,
deoxygenated blood, oxidized blood, and blood plasma.
Note — The Masimo features available depend on the purchased options for the
monitor. The following description may not apply to all monitors.
The module performs pulse oximetry by using a sensor with various light-emitting
diodes (LEDs) that pass light through the measurement site to a diode (detector).
Signal data is obtained by passing various visible and infrared lights (LEDs, 500 nm to
1400 nm) through a capillary bed (for example, a fingertip, a hand, a foot) and
measuring changes in light absorption during the blood pulsatile cycle. This
information may be useful to clinicians. The maximum radiant power of the strongest
light is rated at 25 mW. The detector receives the light, converts it into an electronic
signal, and then sends it to the Masimo rainbow board for calculation.
The Masimo rainbow board uses proprietary algorithms to calculate the patient's
functional oxygen saturation (SpO2 [%]), total hemoglobin concentration (SpHb [g/dL]),
and pulse rate (PR). The SpHb measurements rely on a multi-wavelength calibration
equation to quantify the concentration of total hemoglobin in arterial blood.
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of
total hemoglobin (SpHb) and carboxyhemoglobin concentration (SpCO) in arterial
blood. It relies on the same principles of pulse oximetry to make its SpHb and SpCO
measurement. The measurement is taken by a sensor capable of measuring SpHb,
usually on the fingertip for adult and pediatric patients. The sensor connects directly to
the Pulse CO-Oximeter or with a patient cable. The sensor collects signal data from the
Theory of Operation
Efficia CM Series Service Guide A-7
Measurement Modules
patient and sends it to the module. The calculated data is measurement of total
hemoglobin concentration and a percentage value for the SpCO, which reflect blood
levels of carbon monoxide bound to hemoglobin.
The specified PR accuracy is the root-mean-square (RMS) difference between the
measured values and the reference values. The Masimo sensors have been validated
for pulse rate accuracy for the range of 25 – 240 bpm in bench top testing against an
electronic pulse simulator.
Rainbow acoustic monitoring continuously measures a patient's respiration rate based
on airflow sounds generated in the upper airway. Respiratory sounds include sounds
related to respiration such as breath sounds (during inspiration and expiration),
adventitious sounds, cough sounds, snoring sounds, sneezing sounds, and sounds
from the respiratory muscles. The acoustic sensor translates airflow sounds generated
in the upper airway to an electrical signal that can be processed to produce a
respiration rate, RRa, measured as breaths per minute.
IBP
The invasive blood pressure (IBP) measurement produces real-time waveforms and
numeric values for systolic, diastolic, and mean arterial blood pressure. IBP
measurements use an A/D converter. The A/D converter interfaces to the
microprocessor through an SPI connection in the front end interface.
An isolation transformer and opto-coupler provide the required isolation between
isolated and grounded sections of the monitor. Software algorithms filter typical
artifacts like respiratory variation and pressure changes caused by flushing the line or
drawing blood samples.
Temperature
The temperature circuitry processes a signal from a probe containing a thermistor
component whose resistance is temperature-dependent. The thermistors follow the
YSI 400 resistance vs temperature curve. These resistance temperature detectors
(RTDs) measure temperatures because of the physical principle of the positive
temperature coefficient of electrical resistance of metals. The advantage is stable
outputs for long periods of time, ease of recalibration, and accurate readings over
relatively narrow temperature spans. The monitor’s input circuitry conditions and
processes the signal to drive the numeric display.
Theory of Operation
A-8 Efficia CM Series Service Guide
Measurement Modules
Theory of Operation
Efficia CM Series Service Guide A-9
Measurement Modules
Mainstream CO2
The Mainstream CO2 sensor is used for the continuous measurement of CO2 and
respiratory rate. The sensor measures CO2 with the infrared absorption technique. The
infrared absorption technique is based on the principle that CO2 molecules absorb
infrared (IR) light energy of specific wavelengths and the amount of energy absorbed is
directly related to the CO2 concentration. When an IR beam is passed through a gas
sample containing CO2, the electronic signal from the photo-detector, which measures
the remaining light energy, is measured. This signal is then compared to the energy of
the IR source and adjusted to accurately reflect the CO2 concentration in the sample.
The Mainstream CO2 sensor's response to a known concentration of CO2 is stored at
the factory in the sensor's memory. A reference channel accounts for optical changes
in the sensor, allowing the system to remain in calibration without user intervention.
Cardiac Output
The Cardiac Output (C.O.) measurement invasively measures cardiac output and other
hemodynamic parameters using the thermodilution technique. Thermodilution
determines the flow rate of a system by introducing a cold solution into the system
and measuring the resulting drop in temperature at a downstream site. The C.O.
measurement circuit measures the temperature of the blood and an injectate with
thermistors in the CO sensor set.
In the right heart thermodilution method, a fluid of known volume and temperature is
injected into the right atrium through the proximal port of a pulmonary artery (PA)
(Swan-Ganz) catheter. The injectate bolus mixes with the blood in the right ventricle
and the change in blood temperature is measured with a thermistor at the distal end of
the catheter in the pulmonary artery.
The temperature change is displayed as a curve in the C.O. window, and the monitor
calculates the C.O. value from this curve. The C.O. value is inversely proportional to the
area under the curve. As cardiac output varies continuously, a series of measurements
must be carried out to achieve a reliable C.O. average value. Always use the average of
multiple thermodilution measurements for therapy decisions.
Theory of Operation
A-10 Efficia CM Series Service Guide
B
Electromagnetic Compatibility
This appendix lists the tests and compliance levels that make the Efficia CM patient
monitors suitable for use in the specified electromagnetic environment according to
IEC 60601-1-2:2014, 4th Edition Standard.
Warnings • Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
• The Efficia CM patient monitors should not be used adjacent to or stacked with
other equipment. This should be avoided because it could result in improper
operation. If adjacent or stacked use is necessary, the monitor should be
observed to verify normal operation in the configuration in which it is used.
Electromagnetic Compatibility
Efficia CM Series Service Guide B-1
•
Does the interference occur only when in close proximity to certain medical electrical
equipment?
• Do parameter measurement values change dramatically when the AC line cord is
unplugged?
Once the source is located, attempt to attenuate the interference by distancing the
product from the source as much as possible. If assistance is needed, contact your
technical support team.
The EMC standards state that manufacturers of patient-coupled equipment must specify
immunity levels for their systems. See Table B-2 to Table B-6 for this detailed immunity
information.
Table B-1
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions Group 1 Efficia CM Patient Monitors may contain an internal radio which
CISPR 11 emits RF energy. Including these radio modules, total RF
emissions are very low and not likely to cause any interference
in nearby electronic equipment.
Electromagnetic Compatibility
B-2 Efficia CM Series Service Guide
Table B-1
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions Class A The Efficia CM monitors are suitable for use in a Hospital
CISPR 11 Environment, whose power is not directly connected to the
public low-voltage power supply network.
Harmonic emissions Class A
IEC 61000-3-2
Note — The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Proximity fields Refer to Table B-3 Refer to Table B-3 Typical Hospital Environment
from RF wireless
communications
equipment
IEC 61000-4-3
Electromagnetic Compatibility
Efficia CM Series Service Guide B-3
Table B-3 Proximity fields from RF wireless communications equipment
Test IMMUNITY
Band a) Modulation b) Distance
Frequency Service a) Modulation b) TEST LEVEL
(MHz) (W) (m)
(MHz) (V/m)
Notes
If necessary to achieve the Immunity Test Level, the distance between the
transmitting antenna and the Efficia CM patient monitors may be reduced to 1 m.
The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the up link frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case
Electromagnetic Compatibility
B-4 Efficia CM Series Service Guide
Table B-4 Input A.C. Power Port
Electromagnetic
Immunity Test Immunity Test Level Compliance Level
Environment - Guidance
Conducted disturbances 3V 3V
induced by RF fields 0.15 MHz – 80 MHz 0.15 MHz – 80 MHz
IEC 61000-4-6 6 V in ISM and amateur 6 V in ISM and amateur
radio bands between radio bands between
0.15 MHz and 80 MHz 0.15 MHz and 80 MHz
Electromagnetic
Immunity Test Immunity Test Level Compliance Level
Environment - Guidance
IEC 61000-4-4
Conducted disturbances 3 V, 0.15 MHz – 80 MHz 3 V, 0.15 MHz – 80 MHz Typical Hospital
induced by RF fields Environment
Electromagnetic Compatibility
Efficia CM Series Service Guide B-5
Table B-6 Patient Coupling Port
Electromagnetic
Immunity Test Immunity Test Level Compliance Level
Environment - Guidance
Conducted disturbances 3 V, 0.15 MHz – 80 MHz 3 V, 0.15 MHz – 80 MHz Typical Hospital
induced by RF fields Environment
IEC 61000-4-6 6 V in ISM and amateur 6 V in ISM and amateur
radio bands between radio bands between
0.15 MHz and 80 MHz 0.15 MHz and 80 MHz
Electromagnetic Compatibility
B-6 Efficia CM Series Service Guide
C
Replacement Parts and Assembly Drawings
Assembly Drawings
The drawings show the parts of an Efficia CM 120 monitor. Your monitor may be slightly different.
Figure C-1 Rear View
3 10
11 7
4
12
8
1
2
5
13
14
15
Printed in USA
453564623591, Revision C
*453564623591*
*C*