You are on page 1of 6

THORACIC ANAESTHESIA

Bronchoscopic insertion Learning objectives


of double lumen After reading this article, you should be able to:

endotracheal tubes and C describe the anatomical difference between the right and left
main bronchi

bronchial blockers C

C
explain the differences between left and right DLTs
demonstrate safe placement of DLTs and BBs
C recall the key fibreoptic views required to confirm correct
Nicholas Heseltine positioning of DLTs
Andrew Knowles C describe the advantages and disadvantages between DLTs and
BBs
C outline some common problems encountered while siting DLTs
Abstract and the appropriate action to take
Double-lumen tubes (DLTs) should be placed using a fibreoptic bron-
choscope. This allows correct positioning of the bronchial lumen in the
chosen mainstem bronchus. It also ensures that the blue bronchial of DLTs was commonplace and associated with significant
cuff does not obstruct the side to be ventilated when it is inflated morbidity and mortality. Therefore familiarity with broncho-
under direct vision. Fibreoptic bronchoscopy facilitates correct posi- scopic views and corresponding anatomy is essential for safe
tioning of the ventilatory side slot of a right DLT over the right upper DLT placement.
lobe bronchus. The anaesthetist must know the fibreoptic tracheo- Bronchial blockers (BBs), using a single lumen endotracheal
bronchial anatomy to properly position left- and right-sided DLTs tube (SLT), were introduced in 1982 as an alternative means of
and should always reconfirm the position of a DLT with fibreoptic lung isolation. They require a fibreoptic bronchoscope (FOB) for
bronchoscopy after repositioning the patient. Maintaining orientation placement and can be more technically difficult to place.
(anterioreposterior) during fibreoptic bronchoscopy is crucial to posi- Therefore as for DLTs good anatomical knowledge and a degree
tioning a DLT, particularly after the patient has been turned to the of FOB dexterity is required for safe placement.
lateral position. A fibreoptic bronchoscope can also be used as a
guide to direct a DLT under direct vision into its correct position.
Anatomy
Keywords Bronchial blocker; double lumen tube; fibreoptic bron-
choscopy; one lung ventilation; vivaSight
The trachea averages 12 cm (range 10e13 cm) long in an adult
male, being slightly shorter in females. At its termination it bi-
Royal College of Anaesthetics CPD Matrix: 2A01, 3A01, 3G00 furcates to form the right and left main bronchi. The right main
bronchus is shorter (approximately 2 cm), steeper and of larger
caliber than the left main bronchus. The right upper lobe bron-
chus (RUB) divides first and the right bronchus continues as the
bronchus intermedius for a further 2e3 cm before splitting into
Background the right middle and lower bronchi.
Double lumen endotracheal tubes (DLTs) allow the anaesthetist The left main bronchus projects more in the horizontal plane,
to isolate the airways of the left and right lungs. Lung isolation is approximately 5 cm in males, is narrower and divides into the
may be used to prevent contralateral lung contamination during left upper and lower bronchi.
bronchial lavage as in pulmonary haemorrhage and infection or Multiple anatomical variants exist; a common example being
in case of bronchial disruption. Differential lung ventilation a tracheal RUB. Although characteristically picked up on pre-
strategies may be employed via a DLT in thoracic surgery to operative computed tomography (CT), endoscopic conforma-
access the lungs, aorta and spinal cord, in minimally invasive tion of the right upper lobe location is still important when
cardiac surgery and in upper gastrointestinal/oesophageal placing a DLT.
surgery.
Carlens introduced the first DLTs for bronchorespirometry in DLT design
the 1940s. Since then there have been numerous design im-
Most modern DLTs are based on the Robertshaw design of two
provements to reduce misplacement. Despite this, until the
D-shaped endotracheal tubes bonded together. These lumens are
advent of routine fibreoptic bronchoscopy, incorrect positioning
fixed together on their flat sides to maintain a circular external
surface. The bronchial lumen has blue markings and the tracheal
is white. Both end lumens have cuffs proximal to their ventila-
Nicholas Heseltine MB ChB MRes is an Anaesthetic Trainee at tion ports. They are single use only, made of polyvinyl chloride
Blackpool Victoria Hospital, Blackpool, UK. Conflicts of interest: none and without a carinal hook.
declared. Unlike other breathing tubes they are sized by their external
Andrew Knowles FRCA is a Consultant Cardiothoracic Anaesthetist diameter rather than internal diameter. Their internal diameter is
at Lancashire Cardiac Centre, Blackpool, UK. Conflicts of interest: in fact much smaller and so requires a narrow FOB (<4 mm)
none declared. during placement.

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 142 Ó 2021 Published by Elsevier Ltd.
THORACIC ANAESTHESIA

Selecting the correct size of DLT  The bronchial lumen must not encroach on the lobar
bronchi.
Frequently 35e37 French (Fr) gauge is used for females and 39
 The bronchial cuff must not herniate into the trachea.
e41 Fr for males. In women less 5 ft 3 in (160 cm) the author uses
32 Fr, 35 Fr up to 5 ft 7 in (67 cm) and 37 Fr above this height. In
Left DLTs
men below 5 ft (165 cm), a 35 Fr is used, a 37 Fr up to 5 ft 8 in
(172 cm), a 39 Fr up to 6 ft (182 cm) and a 41 Fr above this Left DLTs are more commonly used than right DLTs for both left-
height. This can be adjusted with clinical examination for and right-sided surgery. This is because right DLTs are more
particularly long or short necks. For paediatric use a 28 Fr is difficult to position correctly due to the proximity of the RUB to
available.1 Preoperative estimation of size can be done by mea- the carina. Also left DLTs reliably permit either left or right OLV.2
surement of the tracheal and bronchial diameters on available A left DLT is contraindicated when its bronchial lumen is
imaging. However the most reliable assessment is to insert and likely to interfere with surgery. This occurs when the staple line
visualize the ‘fit’ with a FOB. crosses the left main bronchus during a left pneumonectomy or
The correct size DLT ultimately fulfills the following criteria: in the presence of proximal intra-luminal pathology.
 It passes through the larynx and trachea with minimal
resistance. Insertion of a left DLT
 It achieves one lung ventilation (OLV) with low bronchial Intubation of the trachea is performed with the DLT mounted on
cuff volumes (<3 cm). an introducer and the bronchial lumen orientated anteriorly. This

Technique for left DLT insertion

Tracheal intubation Minimal or no resistance

Remove introducer, rotate DLT


anticlockwise and advance to Length (cm) = (height (cm)/10) + 12
resistance

Y-connector

Auscultate, inspect and note EtCO2

Cuff volume should not exceed 3ml.


Clamp tracheal lumen and
Auscultate and inspect to

Unclamp tracheal lumen then


clamp the bronchial lumen contralateral lung isolation

Unclamp bronchial lumen

Perform FOB via tracheal lumen 1. Termination of the tracheal


lumen must be proximal to
the carina
2. Bronchial cuff is seen in the
left main bronchus with no
herniation
3. Position of the RUL is
Position patient, recheck DLT
site and then secure main bronchus

Figure 1

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 143 Ó 2021 Published by Elsevier Ltd.
THORACIC ANAESTHESIA

Figure 2

can be via direct laryngoscopy, videolaryngoscopy or guided by a The position of the DLT and the adequacy of OLV is assessed
FOB. Although an awake intubation over a FOB is possible, in by endoscopic and clinical examination. These can be performed
practice it is less well tolerated due to the size of the DLT. simultaneously or sequentially. This article describes a simulta-
After the trachea is intubated the introducer is removed. The neous technique.
DLT is advanced and simultaneously rotated anticlockwise to- First the bronchoscope is passed through the tracheal lumen of
ward the left main bronchus. The following equation can be used the DLT. The carina is observed with the tube passing into the left
to predict the length required: main bronchus with the bronchial cuff deflated but visible. Next the
cuff is inflated with 1 ml of air at a time under vision. The cuff should
(Height (cm)/10) þ12 ¼ Insertion length at teeth (cm) still be visible, not herniate into the trachea and have a volume less
than 3 ml. The right bronchial tree is now examined, noting the RUB
Therefore for a 170 cm adult the insertion length is approxi- with its three segmental bronchi (apical, posterior and anterior).
mately 29 cm. This unique feature acts as an anatomical landmark and confirms
Once the bronchus is intubated the tracheal cuff is inflated. the observed anatomy. Now the FOB is passed through the bron-
Auscultation, inspection and observation of the end tidal carbon chial lumen of the DLT to view the left upper and lower bronchi. The
dioxide trace confirm bilateral ventilation. tip of the bronchial lumen should not encroach upon these struc-
tures3 (Figures 1e3).
Following FOB examination the tracheal and bronchial lu-
mens are sequentially clamped and released enabling clinical
confirmation of OLV.

Right DLTs
Right DLTs are used less frequently because of difficulty in siting
them and their propensity to block the RUB. However, they are
still indicated when the left main bronchus needs to remain clear
as when the staple line is crosses the left main bronchus
(pneumonectomy and lung transplant) or in the presence of
proximal intra-luminal pathology. An important contraindication
is a tracheal bronchus, where if one were sited only the right
middle and lower lobes would be ventilated.
The right DLT differs from the left in having, an extra venti-
lation port and an oblique shaped cuff (Figure 4). These modi-
Figure 3 A DLT in the left main bronchus with its cuff; not herniating fications facilitate positioning without obstructing the RUB.
into the trachea.

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 144 Ó 2021 Published by Elsevier Ltd.
THORACIC ANAESTHESIA

RUB is in the 3 o’clock position. If this manoeuvre does not bring


the RUB into view the whole DLT is slowly withdrawn while
being rotated until it is in view. After positioning the patient
bronchoscopy is repeated to check the DLT has not moved
(Figure 5).

Endoscopic-guided bronchial intubation


Alternatively, after tracheal intubation the FOB can be passed
through the bronchial lumen of the DLT and used to guide
the tube into position. This has the advantages of reduced
failure rate and also allows for simultaneous assessment of
tracheal and bronchial anatomy. Disadvantages are that it
may increase apnoea time and requires a two-handed
technique.

VivaSight-DL
Figure 4 A right DLT with its extra ventilatory side port.
The VivaSight-DL is a left DLT incorporating a fibreoptic scope at
the end of its tracheal lumen. It has an extra port to clean se-
Insertion of a right DLT cretions from the fibreoptic tip but is otherwise based on the
The insertion technique right DLTs is as for left-sided tubes Robertshaw design. It can reduce intubation and correct place-
except that the DLT must be rotated clockwise toward the right ment times, is only available as a left-sided tube and is dispos-
main bronchus and the extra ventilatory port aligned with the able (Figures 6 and 7).
orifice of the RUB. A concern in use is that since its camera is fixed a flexible FOB
After siting the right DLT the FOB is passed down the tracheal may sometimes be needed to confirm its placement thereby
lumen. The carina should be seen with the blue bronchial cuff in adding to cost. However, research suggests the requirement for
the right main bronchus. It is then inflated under vision to ensure another bronchoscope with the VivaSight-DL is infrequent and
the cuff does not herniate into the trachea. The FOB is then thereby enhancing its cost effectiveness.4
passed through the bronchial lumen and through the end port
into the bronchus intermedius. After viewing the right lower and Troubleshooting a DLT
middle bronchi the scope is withdrawn slowly into the DLT until
the blue colour around the bronchial cuff is seen. The tip of the Common problems encountered and possible solutions are
FOB is then anteverted to observe the extra ventilation port. If the detailed in Table 1.
DLT is in the correct position the RUB should be seen through
this. If it is not observed then the DLT is manipulated to bring it Bronchial blockers
into view. A trained assistant must deflate the bronchial and Bronchial blockers vary in design however can be summarized as
tracheal cuffs then the DLT is rotated, usually the orifice of the flexible narrow gauge cuffed devices (Figure 8). They are used

Figure 5

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 145 Ó 2021 Published by Elsevier Ltd.
THORACIC ANAESTHESIA

Figure 6 VivaSight-DL. Camera between tracheal and bronchial cuffs.

Figure 8 Left DLT and Uniblocker.

nasal and tracheostomies), in difficult airways and rapid


sequence inductions. The use of an SLT also reduces laryngeal
and tracheal trauma. Furthermore, because they are smaller and
Figure 7 VivaSight in situ view with integrated camera and screen. more flexible than DLTs they can be used in patients with
deformed anatomy and can block individual lobes (Table 3).
However, they have some disadvantages related to their
with a specific mount and a normal single lumen tube. The narrow lumens and flexibility. Because of their narrow lumens,
cuffed end requires placement under vision of a FOB prior to cuff suction, oxygen insufflation and CPAP are less effective or not
inflation and lung isolation. achievable. As a result, management of hypoxia during OLV
There are multiple designs in common practice, each with can be more difficult. Equally, once the lung is blocked its
specific features (see Table 2). collapse takes more time and because of greater flexibility BBs
are more prone to displacement. Likewise BBs, except for the
Bronchial blockers versus double lumen tubes EZ blocker, unless repositioned, cannot be used when alter-
The key advantage of BBs is that they can be used with SLTs. nating side of ventilation is required (e.g. thoracic
This means they can be used in patients already intubated (oral, sympathectomy).

Troubleshooting DLTs
Problem Underlying cause Possible solution

Resistance on passing DLT C Too large DLT C Re-attempt with smaller DLT
C Tracheal or laryngeal pathology C Visualize the larynx with
videolaryngoscope
C Visualize trachea and bronchus with bron-
choscope during bronchial intubation
High bronchial cuff volume or persistent C Too small DLT C Check position with bronchoscope
bronchial cuff leak C Incorrect placement C If position correct consider siting a larger
DLT
One-lung ventilation not achieved C Misplacement C Perform bronchoscopy to check DLT posi-
C Aberrant anatomy tion and assess for anatomical variants
One-lung ventilation prior to lumen clamping C Bronchus intubated with bronchial and C Withdraw DLT until bilateral ventilation
tracheal lumens occurs

Table 1

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 146 Ó 2021 Published by Elsevier Ltd.
THORACIC ANAESTHESIA

Common bronchial blocker types


Bronchial blocker Design features

Arndt BB is mounted on a guide wire with a distal


loop to snare the FOB and so aid its
placement. It has a 1.4 mm lumen
Cohen The tip is anteverted or retroverted with a
proximal wheel
EZ blocker The tip is Y shaped to cover the carina. Both
arms are cuffed to enable blocking of either
bronchus without repositioning
Uniblocker (Univent) A semi-rigid BB with a curved end to aid
placement. Its predecessor the Univent, a SLT,
incorporated this BB

Table 2
Figure 9 The Univent bronchial blocker is deflated and visible in the
left main bronchus.
Advantages and disadvantages of bronchial blockers
Advantages Disadvantages Tracheal bronchus
C Allows for SLT C Less effective suction A tracheal RUB is a relatively common variant (0.1e2%) offering
 easier rapid sequence C Needs bronchoscopy a unique DLT challenge. Management depends on which lung is
intubation C More difficult to manage to be deflated. If the left lung is to be deflated and the right
 less laryngeal and hypoxia ventilated then placing a single lumen tube above the aberrant
tracheal trauma C Unable to perform bron- bronchus and inserting a bronchial blocker in the left main
 less tracheal cuff choscopy past BB once bronchus is the best option. If the right lung is to be deflated and
rupture inflated the left ventilated then placing a left-sided DLT will normally
C Can be used with tra- C Unable to do alternate suffice. However, if the right lung does not deflate adequately
cheostomy or nasal SLT OLV without then another option is to use a Univent tube with an extra
C Lobe or segmental repositioning bronchial blocking device. Consequently two bronchial blockers
isolation C More easily displaced may be employed to occlude separately; the right main bronchus
C Can be used with and tracheal RUB. A
abnormal anatomy

Table 3 REFERENCES
1 Ng A, Swanevelder J. Hypoxaemia during one lung ventilation.
Cont Educ Anaesth Crit Care Pain 2010; 4: 117e22.
Insertion of a BB (Figure 9) 2 Slinger P. Principles of practice of thoracic anesthesia. Springer
First tracheal intubation is performed, ideally using a larger SLT Verlag New York, 2011.
(8.5/9) to allow for easy passage of the BB. Then the BB mount is 3 Foley K, Slinger P. Fibreoptic bronchoscopic positioning of double
connected and ventilation is resumed prior to passing it. lumen tubes. Anaesth Intensive Care Med 2011; 12: 554e7.
Depending on the type of BB differing techniques are used. The 4 Larsen S, Holm JH, Saure TN, Andersen C. A cost-effective anal-
Arndt is snared onto the tip of the scope and guided into position. ysis comparing the VivaSight doulbe-lumen tube and a conven-
Other types are semi-rigid so are guided under FOB vision. Once tional double-lumen tube in adult patients undergoing thoracic
in position the cuff is inflated ensuring there is no balloon surgery involving one-lung ventilation. PharmoEconomics e Open
herniation. 2020,Mar; 4: 159e69.

ANAESTHESIA AND INTENSIVE CARE MEDICINE 22:3 147 Ó 2021 Published by Elsevier Ltd.

You might also like