TEST REPORT
For
Prepared For
Prepared By
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 1 of 36
TABLE OF CONTENTS
VERIFICATION OF CONFORMITY ................................................................................................................. 3
1. TEST SUMMARY.......................................................................................................................................... 4
2. SYSTEM DESCRIPTION ............................................................................................................................. 5
3. PRODUCT INFORMATION ......................................................................................................................... 6
4. SUPPORT EQUIPMENT .............................................................................................................................. 7
5. TEST FACILITY ............................................................ .............................................................................. 8
6. EN 60601-1-2 LINE CONDUCTED EMISSION TEST ................................................................................. 9
6.1. TEST EQUIPMENT OF CONDUCTED EMISSION TEST ......................................................................... 9
6.2. LIMITS OF LINE CONDUCTED EMISSION TEST ................................................................................... 9
6.3. BLOCK DIAGRAM OF TEST SETUP ........................................................................................................ 9
6.4. PROCEDURE OF LINE CONDUCTED EMISSION TEST .......................................................................10
6.5. TEST RESULT OF LINE CONDUCTED EMISSION TEST ..................................................................... 11
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 2 of 36
11.1. TEST EQUIPMENT ............................................................................................................................... 23
11.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 23
11.3. TEST PROCEDURE ............................................................................................................................. 24
11.4. PERFORMANCE & RESULT .............................................................................. .................................. 25
12. IEC 61000-4-4 TEST ................................................................................................................................ 26
12.1. TEST EQUIPMENT ............................................................................................................................... 26
12.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 26
12.3. TEST PROCEDURE ............................................................................................................................. 27
12.4. PERFORMANCE & RESULT ................................................................................................................ 27
13. IEC 61000-4-5 SURGE IMMUNITY TEST ............................................................................................... 28
13.1. TEST EQUIPMENT OF SURGE TEST ................................................................................................. 28
13.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 28
13.3. TEST PROCEDURE ............................................................................................................................. 29
13.4. PERFORMANCE & RESULT ................................................................................................................ 29
14. IEC 61000-4-6 TEST ................................................................................................................................ 30
14.1. TEST EQUIPMENT ............................................................................................................................... 30
14.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 30
14.3. TEST PROCEDURE ............................................................................................................................. 31
14.4. PERFORMANCE & RESULT ................................................................................................................ 31
15. IEC 61000-4-8 TEST ................................................................................................................................ 32
15.1. TEST EQUIPMENT ............................................................................................................................... 32
15.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 32
15.3. TEST PROCEDURE ............................................................................................................................. 33
15.4. PERFORMANCE & RESULT ................................................................................................................ 33
16. IEC 61000-4-11 TEST .............................................................................................................................. 34
16.1. TEST EQUIPMENT ............................................................................................................................... 34
16.2. BLOCK DIAGRAM OF TEST SETUP .................................................................................................... 34
16.3. TEST PROCEDURE ............................................................................................................................. 35
16.4. PERFORMANCE .................................................................................................................................. 35
APPENDIX 1 .................................................................................................................................................. 36
PHOTOGRAPHS OF EUT ............................................................................................................................. 36
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 3 of 36
VERIFICATION OF CONFORMITY
Equipment Under Test: Cryolipolysis Slimming Machine
Model Number: LB-C4B, LB-C4, LB-C5, LB-C6, LB-C2, LB-C2P, LB-CRYO, LB-IS4
Model Difference: The series models have same electrical structure as LB-C4B, except
for the different appearance and power.
West Zone No.5 Mauhwa Industry Park, No.1 Caida 3rd Street, Shunyi
District, Beijing, China
Deviation: None
The above equipment was tested by East Notice Certification Service Co., Ltd. for compliance with the
requirements set forth in MEDICAL DEVICES: GENERATL Directive and the Technical Standards mentioned
above. This said equipment in the configuration described in this report shows the maximum emission levels
emanating from equipment and the level of the immunity endurance of the equipment are within the
compliance requirements.
Sample and Standards in accordance with ANSI C63.4 and CISPR 11 / EN 55011
Should any objections to the test reports occurred, should submit it to the Company within ten days since the
issuing of the report, Fail to accept.
The test results of this report relate only to the tested Sample identified in this report.
Checked By _______________________
Authorized By ______________________
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 4 of 36
1. TEST SUMMARY
Test Test Method Complies Level Result
Emission Tests
Radiated Emission (3m semi-anechoic
CISPR 11 Group1, Class B PASS
chamber) 30MHz to 1000MHz
Conducted Emission on AC 150kHz to
CISPR 11 PASS
30MHz
Harmonic EN 61000-3-2:2014 PASS
Flicker EN 61000-3-3:2013 PASS
Immunity Tests
Immunity to electrostatic discharges Contact: ±8kV
IEC 61000-4-2:2008 PASS
±2kV ±4kV ±6kV ±8kV ±15kV Air: ±2,4,6,8,15kV
Immunity to RF electromagnetic fields
80MHz to 2700MHz 10V/m (rms) 10V/m, 80% Am at 1kHz
380MHz to 390MHz 27V/m (rms) PM at 18Hz
430MHz to 470MHz 28V/m (rms) FM ± 5 kHz deviation at 1kHz
704MHz to 787MHz 9V/m (rms) IEC 61000-4-3:2010 PM at 217 Hz PASS
800MHz to 960MHz 28V/m (rms) PM at 18Hz
1700MHz to1990MHz 28V/m (rms) PM at 217 Hz
2400MHz to 2570MHz 28V/m (rms) PM at 217 Hz
5100MHz to 5800MHz 9V/m (rms) PM at 217 Hz
+2 kV for power supply lines
Electrical Fast Transient Immunity IEC 61000-4-4:2012 PASS
+1 kV for input/output lines.
+1 kV differential mode
Surge Immunity IEC 61000-4-5:2014 PASS
+2 kV common mode
3 Vrms
Conducted Disturbances Immunity IEC 61000-4-6:2013 PASS
150 kHz to 80 MHz
Power Frequency Magnetic Immunity IEC 61000-4-8:2009 30 A/m - 50 or 60 Hz PASS
<5 % UT (>95 % dip in UT)
For 0,5 cycle 40 % UT
(60 % dip in UT) for 5 cycles
Voltage Dips, Interruptions Immunity IEC 61000-4-11:2004 70 % UT (30 % dip in UT) for 25 PASS
cycles
<5 % UT (>95 % dip in UT) for 5
sec
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 5 of 36
2. SYSTEM DESCRIPTION
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 6 of 36
3. PRODUCT INFORMATION
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 7 of 36
4. SUPPORT EQUIPMENT
Device Type Manufacturer Model Name Serial No. Data Cable Power Cable
-- -- -- -- -- --
**Note: All the above equipment/cables were placed in worse case positions to maximize emission signals
during emission test.
Grounding: Grounding was in accordance with the manufacturer’s requirements and conditions for the
intended use.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 8 of 36
5. TEST FACILITY
Location: 1/F, Haohui Commercial Building, Zhuji Street, Dongpu Town, Tianhe District,
Guangzhou City, China
Description: There is one 3m semi-anechoic an area test sites and two line conducted labs for
final test. The Open Area Test Sites and the Line Conducted labs are constructed
and calibrated to meet the FCC requirements in documents ANSI C63.4 and CISPR
11/EN 55011 requirements.
Site Filing: The site description is on file with the Federal Communications Commission, 7435
Oakland Mills Road, Columbia, MD 21046.
Instrument Tolerance: All measuring equipment is in accord with ANSI C63.4 and CISPR 22 requirements
that meet industry regulatory agency and accreditation agency requirement.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 9 of 36
500kHz-5MHz 56 46
5MHz-30MHz 60 50
**Note: 1. The lower limit shall apply at the transition frequency.
2.The limit decreases linearly with the logarithm of the frequency in the range 0.15 MHz to 0.50 MHz
Shielded Room
Filter 0.4M
EUT
To Measuring Equipment
0.8M
LISN Power
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 10 of 36
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 11 of 36
30
20
10
0
150 KHz 30 MHz
RESULT: PASS
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 12 of 36
30
20
10
0
150 KHz 30 MHz
RESULT: PASS
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 13 of 36
3M
Amplifier
EUT 1-4M
0.8M Spectrum
Analyzer
The magnetic component shall be measured by means of a loop antenna as described in EN 60601-1-2. The
lighting equipment shall be placed in the centre of the antenna, and the position is not critical.
The test object was operated at its upper limit of its rated voltage and its rated frequency. The induced current
in the loop antenna is measured by means of a current probe (1V/A) and the CISPR measuring receiver. By
means of a coaxial switch the three field directions can be measured in sequence. Each value shall fulfill the
requirements given.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 14 of 36
7.5. TEST RESULTS OF RADIATION EMISSION
50
40
30
20
10
0
30M 50M 70M 100M 200M 300M 500M 700M 1G
Frequency [Hz]
RESULT: PASS
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 15 of 36
Radiated Emission Measurement
50
40
30
20
10
0
30M 50M 70M 100M 200M 300M 500M 700M 1G
Frequency [Hz]
RESULT: PASS
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 16 of 36
0.8M
Note:
1. The EUT was tested with the equipment configured to its rated current.
2. The measurements were carried out under steady conditions. When a piece of EUT is brought into
operation or is taken out of operation, manually or automatically, harmonic currents and power are not
taken into account at first 10s following the switching event. EUT shall not be in standby mode for more
than 10% of any observation period.
3. Harmonics of the fundamental current were measured using a digital power meter with an analogue
output and frequency analyser which was integrated in the harmonic & flicker test system.
For each harmonic order, measure the 1,5 s smoothed r.m.s. harmonic current in each DFT time window and
calculate the arithmetic average of the measured values from the DFT time windows, over the entire
observation period. Each harmonic order, all 1.5 s smoothed r.m.s. harmonic current values and the average
values for the individual harmonic currents, taken over the entire test observation period shall be less than or
equal to the applicable limits.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 17 of 36
8.3. LIMITS OF HARMONIC CURRENT
Limits for Class A Equipment
Harmonics Order n Max. permissible harmonic current (A)
Odd harmonics
3 2.3
5 1.14
7 0.77
9 0.40
11 0.33
13 0.21
15≤n≤39 0.15×15/n
Even harmonics
2 1.08
4 0.43
6 0.30
8≤n≤40 0.23×8/n
NOTE:
1. According to section 5 of EN 61000-3-2:2014, the EUT is Class A equipment.
2. The above limits are for all applications having an active input power>75W. No limits apply for
equipment with an active input power up to and including 75W.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 18 of 36
8.4. RESULT
Test Specification
Test Frequency: 50Hz Test Voltage: 230Vac
Waveform: Sine Test Time: 2.5min
Classification: Class A Test result: PASS
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 19 of 36
0.8M
Note:
1. The test supply voltage (open-circuit voltage) was the rated voltage of the EUT. The test voltage was
maintained within ±2 % of the nominal value. The frequency was 50 Hz ±0.5 %.
2. The voltage fluctuations and flicker were measured at the supply terminals of the EUT.
3. The observation period, Tp, for the assessment of flicker values by flicker measurement, flicker
simulation, or analytical method was:
— for Pst, Tp = 10 min;
— for Plt, Tp = 2 h.
The observation period included that part of the whole operation cycle in which the EUT produces the
most unfavourable sequence of voltage changes.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 20 of 36
9.3. RESULT
Flicker Test Summary per EN/IEC 61000-3-3 (Run time)
0.75
Pst
0.50
0.25
0.00
14:37:19
0.50
Plt
0.25
14:37:19
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 21 of 36
Standard require : B
Tester : Sam Liu
Temperature : 24℃
Humidity : 54%
VCP
EUT &
Support Units Support
>1M Units
HCP
Wooden Table
0.8M
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 22 of 36
The following test condition was followed during the tests.
Note: As per the A2 to IEC 61000-4-2, a bleed resistor cable is connected between the EUT and HCP during
the test.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance
or loss of function is allowed below a performance level specified by the manufacturer, when
the apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be
restored by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 23 of 36
7x4x3 Chamber
3M
1.5M
0.8M
Signal
Power Amp EUT Monitoring using a
Generator
camera
PC Controller to Controller
SG. & PA as well as Control Room
forward power
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 24 of 36
11.3. TEST PROCEDURE
The EUT was located at the edge of supporting table keep 3 meter away from transmitting antenna, it just the
calibrated square area of field uniformity. The support units were located outside of the uniformity area, but
the cable(s) connected with EUT were exposed to the calibrated field as per IEC 61000-4-3.
EUT worked with resistance load, and make sure EUT worked normally.
Setting the testing parameters of RS test software per IEC 61000-4-3.
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 25 of 36
11.4. PERFORMANCE & RESULT
Criteria A: The apparatus continues to operate as intended. No degradation of performance or loss of
function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance
or loss of function is allowed below a performance level specified by the manufacturer, when
the apparatus is used as intended. In some cases the performance level may be replaced by
a permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be
restored by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 26 of 36
Controller Computer
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 27 of 36
Test conditions:
Repetition Frequency: 100 kHz
Tr/Th: 5/50ns
Burst Duration: 15ms
Burst Period: 300ms
Inject Line Voltage kV Inject Method Result (Pass/Fail)
PE+L+N + /- 2 Direct Pass
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance
or loss of function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be
restored by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 28 of 36
AC Source
Surge Immunity
EUT &
Test
Support Units
0.8M
Controller Computer
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 29 of 36
13.3. TEST PROCEDURE
The EUT and support units were located on a wooden table 0.8 m away from ground floor.
EUT worked with resistance load, and make sure EUT worked normally.
Recording the test result as shown in following table.
Test conditions:
Voltage Waveform : 2µs x 50µs
Current Waveform : 8µs x 20µs
Polarity : Positive/Negative
0 0 0
Phase angle : 0 , 90 , 270
Number of Test : 5
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance or
loss of function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be restored
by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 30 of 36
PC
Controller
0.1M<L<0.3M Power
10cm isolation supporter Amplifier
EUT &
Support
Units
CDN
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 31 of 36
14.3. TEST PROCEDURE
The EUT and support units were located at a ground reference plane with the interposition of a 0.1 m
thickness insulating support and the CDN was located on GRP directly.
EUT worked with resistance load, and make sure EUT worked normally.
Related peripherals work during the test.
Setting the testing parameters of CS test software per IEC 61000-4-6.
Recording the test result in following table.
Test conditions:
Frequency Range: 0.15MHz-80MHz
Frequency Step: 1% of fundamental
Dwell Time: 1 sec
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance or
loss of function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be restored
by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 32 of 36
AC Source Dips/Interruption
and Variations EUT &
Simulator Support Units
0.8M
Controller Computer
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 33 of 36
15.3 TEST PROCEDURE
The EUT shall be subjected to the test magnetic field by using the induction coil of standard dimensions (1m
x1m). The induction coil shall then be rotated by 90¡ã in order to expose the EUT to the test field with different
orientations.
Test Conditions:
Meet Performance
Frequency Polarity Level Observation
Criteria
50 Hz and 60Hz X 30A/m Normal B
50 Hz and 60Hz Y 30A/m Normal B
50 Hz and 60Hz Z 30A/m Normal B
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance or
loss of function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be restored
by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 34 of 36
Voltage Dips
Cycles Sync Angle (degrees)
% UT
>95 0.5* 0,45,90,135,180,225,270,315
>95 1 0
30 25 (50Hz); 30 (60Hz) 0
Voltage Interruption
Cycles Sync Angle [degrees]
% UT
>95 250 (50Hz),300 (60Hz) 0
AC Source Dips/Interruption
and Variations EUT &
Simulator Support Units
0.8M
Controller Computer
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 35 of 36
16.3. TEST PROCEDURE
The EUT and support units were located on a wooden table, 0.8 m away from ground floor. EUT worked with
resistance load, and make sure EUT worked normally.
Setting the parameter of tests and then perform the test software of test simulator.
Conditions changes to occur at 0 degree crossover point of the voltage waveform.
Recording the test result in test record form.
Test conditions:
The duration with a sequence of three dips/interruptions with interval of 10 s minimum
(Between each test event)
Voltage Dips:
Voltage Interruptions:
16.4. PERFORMANCE
Criteria A: The apparatus continues to operate as intended. No degradation of performance or loss of
function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of performance or
loss of function is allowed below a performance level specified by the manufacturer, when the
apparatus is used as intended. In some cases the performance level may be replaced by a
permissible loss of performance. During the test, degradation of performance is however
allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or can be restored
by the operation of controls.
PASS FAIL
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com
Report No.: ENC1908218GZ64E1
Page 36 of 36
APPENDIX 1
PHOTOGRAPHS OF EUT
The results shown in this test report refer only to the sample(s) tested unless otherwise stated and the sample(s) are retained for 30 days only. The document is
issued by ENC, this document cannot be reproduced except in full with our prior written permission. The document is available on request and the brief
information for its validation can be assessable and confirmed at http://www.enc-lab.com.
1/F, Haohui Commercial Building, Zhuji Street, Tel:+86-020-2331 4234 Fax:+86-020-8256 8534
East Notice Certification Dongpu Town, Tianhe District, Guangzhou City E-mail: enc@ enc-lab.com Http:// www.enc-lab.com