Humidity measurement in Hydrogen Peroxide


sterilization process in Life Science industry
Advantages of using Vaisala

 Vaisala HUMICAP® humidity sensors


can measure the VHP saturation rate
directly (0-100%)
Humidity output can be used to control
the VHP generator
Catalytic filter available for saturated
VHP to protect the sensor
Vaisala sensor warming to prevent
sensor and filter from getting wet
Vaisala chemical purge feature to
optimize long term sensor performance
Reliable measurement instruments
with global customer support

Maintaining contamination free areas is a must in life science industries and critical
environments are also sterilized regularly. Some of these areas include small volume chambers
like isolators and incubators and less often larger spaces like cleanrooms and processing lines.
Vaporized hydrogen peroxide (VHP) is an effective agent for decontamination/sterilization when
high enough VHP concentration and long enough exposure time are applied. Excessively high
VHP concentration can result in unwanted saturation and surfaces wet with liquid hydrogen
peroxide. Humidity measurement in VHP sterilization is an essential part of the procedure to
control the VHP saturation level.
Humidity influences the VHP saturation level
Humidity levels have a direct correlation to temperatures at which the air containing VHP will saturate. When
saturation conditions occur, condensate, including liquid hydrogen peroxide, will form on surfaces in the
decontaminated space. The decontaminated space may be dried prior the VHP exposure to reach high enough
VHP concentration without saturation. The change of gas saturation rate is due to the molecular structure of
hydrogen peroxide (H2O2) which is similar to water vapor (H2O).The mixture of these two gases determines the
saturation vapor pressure of the gas mixture. Hydrogen peroxide has relatively low saturation pressure which will
increase the possibility of saturation of the air containing humidity.
VHP compliant Humidity measurement technology from Vaisala
Vaisala HUMICAP® Humidity measurement technology with thin film capacitive polymer sensors is capable of
humidity measurements prior and during VHP exposure and in high and low VHP concentrations. The main
advantage of the HUMICAP® sensor is that it is compatible with VHP even without any specific protection. In
applications where saturation occurs a Vaisala protective catalytic filter is recommended.
Measure VHP saturation level
The ability to measure humidity directly in mixtures of VHP and air allows the user
to follow the true saturation level of the gas mixture. Hydrogen peroxide and water
are both polar molecules that change the polymer sensor capacitance. Because the
Vaisala HUMICAP® directly measures the true VHP saturation rate (0-100%), it
can be used to control the sterilization process. Saturation level is highly valuable Forms of Water and Hydrogen
information especially when wet surfaces must be avoided. peroxide molecules
Seven good reasons to choose Vaisala HUMICAP® for the VHP application
1. Durability for direct Vaisala HUMICAP® sensors that 6. Vaisala sensor
VHP exposure are permanently installed to warming to keep sensor
measure normal operating
Vaisala HUMICAP® sensor is conditions need no additional and filter dry in
compatible with direct VHP protection beyond the catalytic saturation
exposure even at high filter. In a full saturation the wet surfaces
concentrations and hundreds of are not only demanding for the
VHP cycles up to saturation. The 4. Catalytic filter to read materials but also make the
long term performance is very good plain humidity in the humidity sensor wet and slow. The
at concentrations higher than VHP gas humidity reading will stay at 100%
typically used for sterilization. value until the sensor and filter
However, for the applications If the pure humidity in VHP gas is
needed, a normal way is to use a have dried. Drying can take a long
where full saturation is possible a time depending on the flow and
Vaisala catalytic filter is Vaisala catalytic filter that
saturation rate of the surrounding
recommended. prevents the VHP from reaching
the humidity sensor, but allows air. To respond to real changes in
2. Catalytic VHP filter for other gas components to be the measured environment the
measured. When the humidity HUMICAP® sensor is available
saturated conditions with a sensor warming feature.
sensor is protected from VHP it
When VHP concentrations are high Sensor warming keeps the sensor
responds to humidity only. The
and condensation is expected, the surface and filter temperature
influence of the catalytic filter is
Vaisala catalytic filter is used to about +1 %RH at 300 pmm VHP higher than the gas saturation point
generate additional protection for or +3 %RH at 900 ppm VHP (at (dewpoint) and this helps to avoid
the humidity sensor. The catalytic 23°C). saturation on the sensor and filter.
filter is based on the generally Sensor warming is a proven Vaisala
known method where the H2O2 is 5. Vaisala chemical technology that is used widely
split to H2O and O2. The saturating when a risk of a wet sensor must be
purge to complete the avoided.
VHP up to wet surfaces improves
the decontamination effect but may
long term stability
damage materials that are exposed The chemical purge technology
7. Reliable humidity
to these conditions regularly. In helps to correct the possible measurement
this environment, sensor protection sensor long term drift and to The HUMICAP® thin film
with a catalytic filter is extend the required calibration capacitive sensor technology was
recommended to protect the sensor interval if VHP is measured invented and patented by Vaisala in
and to extend the calibration without catalytic filter. Chemical the early 1970's. Since then Vaisala
interval. purge automatically heats up the has been leading the way in
sensor in set intervals, evaporating humidity measurement in
3. Measure true the unwanted excess chemicals applications with the most difficult
saturation of the humid out of sensor polymer. This demands. The latest HUMICAP®
VHP gas returns the sensor performance 180R humidity sensor version
back to same level as used in represents the highest humidity
The similarity of the water vapor normal chemical free air.
and vaporized hydrogen peroxide measurement performance for
Utilizing the chemical purge with complex chemical environments.
allows the capacitive HUMICAP® HUMICAP® sensor the long term
sensor to read the true saturation In VHP applications it performs
stability for measurement with excellent durability and long
rate of a humid VHP gas. This is accuracy in VHP application will
important in VHP applications term stability. This excellence
be excellent. combined with the Vaisala
where the gas concentration should
be high enough to kill the living chemical purge and sensor
organisms, but should not saturate. warming technologies makes the
This measured value shows how measurement realible and an
close the true saturation value is. If unique choice for the VHP
the saturation value is not required application.
the catalytic VHP filter can be used
in all VHP measurements. The VHP catalytic filter
in a humidity probe
Ref. NNNNNNNNN-X ©Vaisala 2011
For more information, visit
This material is subject to copyright protection, with all copyrights
www.vaisala.com or contact
us at sales@vaisala.com retained by Vaisala and its individual partners. All rights reserved.
Any logos and/or product names are trademarks of Vaisala or
its individual partners. The reproduction, transfer, distribution or
storage of information contained in this brochure in any form
Page 2 / date / name / ©Vaisala
without the prior written consent of Vaisala is strictly prohibited.