NM Cameras: Head Holder
NM Cameras: Head Holder
Direction 5405481-1EN
Rev. 1
NM Cameras
Head Holder
User’s Guide
©2011 by GE Healthcare
©2011 by GE Healthcare. All Rights Reserved.
Acquisition and/or processing software and the related documentation are confidential and proprietary information of GE Healthcare. Only licensees
of GE Healthcare have a right to use the information contained herein. Only licensees specifically granted copy and/or transfer rights have the right
to copy and/or transfer the information. Any unauthorized use, disclosure, assignment, transfer or reproduction of this confidential information will be
prosecuted to the full extent of the Law.
Disclaimer
GE Healthcare shall not be liable nor obligated in any manner in respect of bodily injury and/or property damage arising from the use of this system
if such use is not in strict compliance with instructions and safety precautions contained in the relevant operating manuals and in all supplements
thereto, in all product labels, and according to all terms of warranty and sale of this software, nor if any change not authorized by GE Healthcare is
made to the software contained herein.
Warning
User provided programs or scripts are NOT validated nor warranted by GE Healthcare. The use of data obtained using such user provided programs
or scripts is the sole responsibility of the party using such programs or protocols.
Users exchanging files and media should beware of the risk of software viruses.
Authorized CE Representative:
GE Medical Systems SCS
283, Rue de la Minière
78530 BUC France
IMPORTANT
This document is an extension of the document set supplied with the camera, which contains all pertinent
operational, safety, regulatory and other information. The reader must read the documents in the set prior to
reading the information herein and prior to using the accessory.
The images in this manual are for demonstration only.
There may be minor differences that do not affect functionality.
Discovery NM/CT 670 images are used to demonstrate head holder usage; these images are equally relevant for
the other cameras.
In This Guide
Overview: Brain Scanning Using the Head Holder, p.5
Head Holder Kit (P/N 5418183), p.6
Setting Up For Scanning With the Head Holder, p.6, including:
Preliminary Configuration for D630 and D670, p.7
Tomo Scan Acquisition Parameters For Head Holder Scanning, p.8
Patient Positioning With Head Holder for NM & Hybrid Brain Scans, p.10
CT-Only Scanning on D670, p.15
CAUTION
Exercise extreme caution when moving the detectors.
When the head holder is attached, it protrudes into the detectors' motion path.
Be extremely careful when moving the detectors so as to avoid contact with the head holder and the patient.
Minimize automatic motion and prefer manual operation.
Avoid switching detector modes while the head holder is attached (except when moving to CT mode for CT-
only scans, see CT-Only Scanning on D670, p.15).
Use the head holder only for head and neck scans with patient in supine position.
Performing chest scans with the patient positioned on the head holder will generate cumulative attenuation from
the head holder and pallet.
Excessive load can break the head holder and cause injury.
The maximum permissible load on the neck and head support is 34 kg (75 lbs).
NOTE
There are two sets of adhesive strips
Use the strips that allow you to center
the head holder on your patient table
Cushion G
Forehead straps H I
C D E F G H I
B
C
CAUTION
If Body Contour is enabled, it is possible that
during the scan the detector housing will collide
E
with the patient or the head holder. F
c. D670 Only:
Select Scan on extender only F
This enables scanning off the table.
3. Proceed to Patient Positioning With Head Holder for NM & Hybrid Brain Scans, p.10
Patient Positioning With Head Holder for NM & Hybrid Brain Scans
1. Verify that you have completed the preliminary steps in Setting Up For Scanning With the
Head Holder, p.6.
2. Select detector mode-H and press <SET>.
Verify that the table is fully retracted and at minimal elevation.
3. Load the selected protocol and apply the scan.
4. Using the hand-held controller:
Move the detectors out fully
Rotate the detectors to 90° (so that both detectors are positioned vertically)
5. Remove the mattress from the bottom of the pallet.
permissible
• Placed symmetric
Holder Placement relative to the sides
CAUTION
• Secured
load
holder on
in place
on the head
safety warnings
label on
ally
the pallet,
ION
the
line on the
line on the label C adhered to the
UT nt
and
CA ceme
r Pla
and general
Head
lde
system
Ho
ad
Align the
5418443-EN
He
instru supplied
entatio
Head Hold
tant safety
Head Holde
n
warnin Operator
r.
er to this
pallet.
gs and
the
general
line
and
5418179
-EN
The lateral edges B of the head
holder are parallel and symmetrical
relative to the edges of the pallet. C
CAUTION
the head holder onto the pallet, Align the Head Holder to this line
centering the holder on the pallet. For important safety warnings and general
usage instructions, review the Operator
documentation supplied with the system and
with the Head Holder.
5418179-EN
H I
NM Cameras – Head Holder – User’s Guide
5405481-1EN, Rev. 1, ©2011 GE Healthcare 12
Patient Positioning With Head Holder for NM & Hybrid Brain Scans
IMPORTANT
This requirement is very important for brain imaging when J K
using a fan-beam collimator.
Longitudinal position: Using the persistence image, ensure that the target anatomy is within the
FOV. (Adjust the pan as needed, depending on the zoom and the longitudinal position.)
NOTE
In specific cases, such as a patient with a short neck, in order to achieve full visibility of the patient's brain in the FOV, use
the hand-held controller to carefully move the pallet towards the gantry (as allowed by patient's shoulders and the detector
housing).
12.Perform a clearance check by rotating the detectors 360° and ensuring that there is no
contact with the patient’s face, neck or shoulders, or with the head holder. If necessary,
adjust the radius of rotation and re-confirm clearance.
CAUTION
The clearance check step is mandatory for ensuring patient safety.
13.Rotate the detectors back to the desired start angle (typically 0°) and press <GO>.
NOTE
To perform a CT-only scan with the NM detectors in L-mode under the table:
1. Select L-mode and press <SET>.
2. Verify that the table is fully retracted and at minimal elevation.
3. Attach the head holder as described in steps 5 to 7 of Patient Positioning With Head Holder for NM & Hybrid Brain Scans, p.10.
4. Load the patient and attach one or both of the head holder straps in order to immobilize the patient's head.
5. Using the hand-held controller:
Move the table up to ~89.0
Rotate the detectors to 180° (so that both detectors are below the table)
6. Select CT Mode A, then press <SET> to move the table to the CT position.
Monitor the patient while the table moves to CT position to ensure safe travel into the bore (the detectors will stay in L mode).