MANUAL
STILLETM MEDSTONE
Urology Imaging Table
The contents of this manual are the property of STILLE AB. Any reproduction in whole or in part is strictly
prohibited.
At the time of printing, this manual correctly described the device and its functions. However, as
modifications may have been carried out since the production of this manual, the system package may
contain one or more amendments to the manual. This manual including any such amendments must be
thoroughly read, before using the device. Images in the manual are for informational use and may differ
from actual product and/or model.
The English version of the User Manual and Technical Manual is the original instructions. Manuals in other
languages are a translation of the original instructions.
The following MEDSTONE™ models are covered by this manual MEDSTONE URO and GYN with 1-5
motors.
The STILLE MEDSTONE™ Mobile C-Arm Imaging Tables are CE marked in conformity with the essential
requirements, according to annex I, and classified, according to annex IX, as Class I (rule 12) of the
Medical Device Directive 93/42/EEC.
We declare the compliance of the medical device concerned with the Swedish Medical Devices Act
(1993:584) and the regulation LVFS 2003:11 of the Medical Products Agency. Hereby, the medical device
complies with the requirements of the Medical Device Directive 93/42/EEC, the amending Directive
2007/47/EC. The directive 2011/65/EU (RoHs2) and applicable harmonized standards.
The device does also comply with the International standards, IEC 60601-1, IEC/EN. CAN/CSA C22.2
#601601-1. Any modification to the device, not authorized by us, will invalidate this declaration.
The original signed CE declaration of conformity is a separate document can be obtained upon request.
Classifications
CE Medical device classification: Class 1
GMDN code: 40658
FDA Regulation Status: Class II Exempt 892.1980 (IXR)
FDA List No.: D291538
Information regarding STILLE Warranty Provisions can be found on STILLEs website, http://www.stille.se.
If you do not find the installation report please do contact QA@STILLE.SE for a copy.
It is recommended that all repairs and modifications are carry out by authorized personnel.
Manufacturer:
STILLE AB
Ekbacken 11
SE-644 30, Torshälla
Sweden
Introduction……………………………….A-1
Important note……………………………………A-1
Safety……..………………..………………….A-3
General…..………………………………………A-3
Safety continued…..……..………………A-4
Maintenance……………………………………A-4
Checks, test……………………………………… A-4
Installation, repair……………………………… A-4
Cleaning………………………………… A-5
10 %
The table is shipped in a heavy duty crate. Upon delivery please inspect for any sign of external damage.
Operational Warning
WARNING:
To avoid risk of electrical shock, this equipment must only be connected to a
supply main’s (outlet) with a protective ground (earth)
Before using the table for the first time, please make sure that the emergency stop switch has not been
pushed in during transportation or uncrating.
If the table movements do not function twist the emergency switch clockwise as indicated on the switch the
table movements should then function correctly
A-1
Introduction
Important note
Using the STILLE Medstone Tables presupposes that the operating personnel is thoroughly familiar
with the operating instructions.
• Safety (A-3)
Document Information
It is important that this document kept for the life of the equipment and transferred to any subsequent
holder or user of the equipment.
Text emphasis
WARNING!
All text labeled “WARNING” calls attention to potential risk to health or life
A-1
0047-Series mobile tables are used for imaging applications. The tables, depending on your option are
provided with several actuators and motorized column giving the tables good movement range.
F
C
Range of application
The intended use of the table is for examinations and imaging (‘C; Arm required)only. Medical staff
must supervise the equipment when it is in use near the patient.
A-2
Safety
General
CAUTION! This device is restricted to use by or on the order of a physician. The
device is intended for use in radiographic examinations under the guidance of trained
health care professionals.
Operating personnel must be familiar with the equipment and the instructions given in this
manual before using it.
A-3
Warning
Checks, test
Checks and tests that shall be performed with specified intervals are described in Function and
safety checks (A-6)
Installation, repair
Modifications of, or additions to the product must be made in accordance with the legal
regulations and generally accepted engineering standards.
The supplier cannot assume responsibility for the safety features and for the reliability and
performance of the equipment, if:
• components affecting safe operation of the unit are not replaced by original spare parts in
case of a malfunction.
On request, we will make technical documents for the product available to you at a charge that
will be invoiced. This does not imply an authorization to undertake repairs. We accept no
responsibility for repairs performed without our expressed written approval.
We suggest that you obtain a report indicating the nature and extent of the work performed
from the persons carrying out such work, and specify that the report shows any changes in
rated parameters or operating ranges, as well as date, the name of the firm and a signature.
A-4
Cleaning
CAUTION! Surfaces that are in contact with the patient shall be cleaned with a lint free
cloth dampened with a small amount of soapy water or cleaning solution.
Approved products and methods for cleaning and disinfecting of table and
stretcher pads
Wash with generous application of neutral soap suds and lukewarm water. Rinse with water and dry. Do
not immerse.
Or
Use a diluted bleach solution of 10-parts water to1-part bleach. Wipe dry with clean cloth. Do not immerse.
Bleach must be diluted to the 10:0 ratio.
DISINFECTING
Use Lysol I.C. Disinfectant Spray by National Laboratories, Moontvale, New Jersey, per manufacturer’s instructions
Reference: National Labortories Item # 95029-000
Or
Use a diluted bleach solution of 10-parts water to1-part bleach. Wipe dry with clean cloth. Do not immerse.
Bleach must be diluted to the 10:0 ratio.
The following Steris products are also approved for use in cleaning and disinfecting our pads:
A-5
Daily checks
When necessary, clean according to Cleaning (A-5).
Monthly checks
• Check the legibility of the label according to section Location of Identification label (page
11). If any label must be replaced, please notify the supplier customer service.
• Power up the stand and check all functions according to Technical Specifications (page
12).
Annual maintenance
The following maintenance is necessary to perform every 12 months to keep the equipment
safe and operational.
• Power up the stand and check all functions according to Technical Specifications (page
12).
A-6
LONGITUDINAL
TRAVEL
TRENDELENBURG
TILT
LIFT
LATERAL TRAVEL
Protective measures
Mechanical safety
It is the operator’s duty to ensure that any danger to the patient or third parties is precluded
before movements are released.
A-8
1. Preventative maintenance should be performed every six (6) months, checking the following items:
3. Actuator mounting bracket bolts are tight and that the actuator pin(s) retention spring clips are securely in
position
4. Test movements of all Actuators and Elevation Lifts to insure that they are moving freely and smoothly
5. On all table that have longitudinal travel (extend and retract) check that the Rollon bearing slides are moving
freely and smoothly
6. On 004-TM5 (5-movement) model versions, check to make sure that the lateral movement linear slide
bearing rails are clean and that the cross table movement is moving freely and smoothly
7. “Emergency Stop Switch”, check that it is functioning correctly, that all movements are disabled when
pressed
8. Hand Control, check that all push button movements are functioning correctly
9. (Optional) Foot Control Switch, check that all movements are functioning correctly and that the spring loaded
switches are not sticking in the “ON” positions
This unit contains the following safety-relevant parts, which should be checked by service personnel
according to the following:
• Tightening of bolts:
M6 6 Ft Lbs
M8 16 Ft Lbs
M12 60 Ft 30 Ft Lbs - Table Top
M4 21 In Lbs
All “Equipment and Accessories” parts are recyclable and should be disposed of in
a safe manner at end of their expected life cycle.
A-9
A-10
Technical specifications
General
400mm
Travel
150mm
Travel
± 15o
o
± 15
300mm
Travel
A-11
Technical specifications
External power supply
Operating environment
Classifications
A-12
A-13
A-14
A-15
NC NEW RELEASE
Lat Lat
Lat Lat Trav
Tren Elev Tren Elev Elev Tren Elev Tren Lat Tilt Lat Tilt Long Long Trav
Trav Trav Long Long Lat Tilt Lat Tilt
up Up Down Down Up up left Right Head Foot left Right
Right left Foot Head Right left Down Down
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
9 9
10 10 SCU Footswitch
Hand Control Pandent 50-186, EHA35-23N50N-000 (OPTIONAL) Control
Longitudinal Travel
10 9 8 6 50-191, 400mm Actuator
Head-Foot
Elevation
4 Down - up
50-184, SCU58-003230-000 50-192, Lift Drive
Motor Control Box Elevation
3 Down - up
120VAC
0
INCOMING
POWER
11 GND
APPROVED A 0047-010-0009 NC
CONFIDENTIAL AND PROPRIETARY INFORMATION
The document contains Confidential and Proprietary information
of Arcoma-na inc., and all design, manufacturing, reproduction, SCALE NONE SHEET 2 OF 2
use, and sale rights in such information are expessly reserved.
NC NEW RELEASE
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
9 9
10 10 SCU Footswitch
Hand Control Pandent 50-186, EHA35-23N50N-000 (OPTIONAL) Control
Longitudinal Travel
10 9 8 6 50-191, 400mm Actuator
Head-Foot
Elevation
4 Down - up
50-195, SCU96-003230-000 50-192, Lift Drive
Elevation
Motor Control Box 3 Down - up
120VAC
INCOMING 0
POWER
11 GND
A-16
CONFIDENTIAL AND PROPRIETARY INFORMATION APPROVED A 0047-010-0009 NC
The document contains Confidential and Proprietary information
of Arcoma-na inc., and all design, manufacturing, reproduction,
use, and sale rights in such information are expessly reserved. SCALE NONE SHEET 2 OF 2