Professional Documents
Culture Documents
Immunoassay Analyzer
REAGENTS
REF C89004G 2×50 Tests Reagent kit, 100 tests, 2 packs, 50 tests/pack
using a suitable disinfectant. transfer only the clarified specimen without the lipemic
iFlash Trigger solution contains sodium hydroxide material.
(NaOH) and should be avoided contact with eyes. Ensure that residual fibrin and cellular matter have
been removed prior to analysis.
REAGENT HANDLING
Use with caution in handling patient specimens to
The reagents may not be used after the stated prevent cross-contamination.
expiration date.
Do not use heat-inactivated samples.
Avoid the formation of foam with all reagents.
Ensure that the patient samples, calibrators and
The reagents in the pack and calibrators are ready for controls are at ambient temperature (20–25°C) before
use measurement.
Close the bottles of calibrator right after calibration and Due to the possible evaporation, specimens and
store at 2–8°C. calibrators on the analyzers should be measured
Do not pool reagents within a reagent kit or between within 2 hours.
reagent kits.
ASSAY PROCEDURE
Prior to loading the iFlash-Anti-TPO reagent pack on
the system for the first time, suspend the Refer to the system operating instruction or the online
microparticles by inverting the reagent pack slightly. help system for detailed information on preparing the
system.
For further information on reagent handling
precautions during system operation, refer to the The test-specific parameters stored in barcode on the
iFlash system operating instruction. reagent pack are read in. In case the barcode cannot
be read, enter the sequence numbers.
STORAGE AND STABILITY Carry out calibration, if necessary.
Storage: Place the calibrators CAL1, CAL2 and CAL3 in the
Store at 2–8°C in an upright position. calibrator rack in the sample zone. Only keep
calibrators open during calibration.
The kit may be used immediately after removal from
2-8°C storage. Test application.
Load samples (Use 20 μL of sample for each
Stability: determination in addition to the sample container and
Unopened at 2–8°C: up to the stated expiration date. system dead volumes).
Opened at 2–8°C: 28 days. Click RUN, the iFlash System performs all the
Store on-board: 28 days. functions automatically and calculates the results.
If the testing will not be completed within 14 days, or After 28 days when using the same reagent lot.
for shipment of samples, freeze at -20°C or colder. A reagent kit with a new lot number is used.
Frozen specimens must be mixed thoroughly after Controls are out of range.
thawing. Required by pertinent regulations.
The samples may be frozen for maximum 1 time.
Centrifuge specimens with a lipid layer on the top, and
2/4 V2.0 English Ed.2018-02-01
Anti-TPO iFlash
Immunoassay Analyzer
Biohazard Symbol
Wellkang Ltd (www.CE-marking.eu)