You are on page 1of 4

Anti-TPO iFlash

Immunoassay Analyzer

REAGENTS
REF C89004G 2×50 Tests Reagent kit, 100 tests, 2 packs, 50 tests/pack

TPO antigen-coated paramagnetic


INTENDED USE R1 microparticles, 3.5 mL/pack, 0.05%
The iFlash-Anti-TPO assay is a paramagnetic particle ProClin 300.
chemiluminescent immunoassay (CLIA) for the
Acridinium-ester-labeled anti-TPO
quantitative determination of antibody to thyroid
R2 conjugate; 6.5 mL/pack; 0.05% ProClin
peroxidase (TPO) in human serum and plasma using the 300.
iFlash Immunoassay Analyzer.
Sample diluent, phosphate buffer, 10
SUMMARY AND EXPLANATION R3
mL/pack; 0.05% ProClin 300.
TPO is a membrane-bound glycoprotein enzyme with an
approximate mass of 107kD. The in vivo function is the Calibrator 1, 1 bottle, 1.0 mL, phosphate
iodination of tyrosine in the synthesis of T3 and T4. CAL1 buffer with protein stabilizers, 0.05%
ProClin 300.
Elevated serum titers of antibodies to TPO are found in
several forms of thyroiditis caused by autoimmunity. Calibrator 2, 1 bottle, 1.0 mL, Anti-TPO in
Autoimmune reactivity to TPO is believed to be polyclonal CAL2 phosphate buffer with protein stabilizers,
and heterogeneous in nature with a minimum of six 0.05% ProClin 300.
antigenic determinants being recognized, comprising both
conformational and linear epitopes. Calibrator 3, 1 bottle, 1.0 mL, Anti-TPO in
CAL3 phosphate buffer with protein stabilizers,
Anti-TPO antibodies are found often in conjunction with 0.05% ProClin 300.
anti-Tg in the majority of cases of Hashimoto’s thyroiditis,
Primary Myxedema, and Graves’ disease. High anti-TPO
MATERIALS REQUIRED (BUT NOT PROVIDED)
titers are found in up to 90% of patients with chronic
Hashimoto’s thyroiditis. In Graves’ disease, 70% of the REF C89999/C89959/C89949, iFlash Pre-Trigger
patients have an elevated titer. Anti-TPO antibodies are Solution: hydrogen peroxide solution.
also found in some other autoimmune disorders such as REF C89998/ C89958/ C89948, iFlash Trigger Solution:
type 1 diabetes mellitus or systemic lupus erythematosus. sodium hydroxide solution.
REF C89997, iFlash Wash Buffer: phosphate buffered
ASSAY PRINCIPLE
saline solution with 0.05% ProClin 300.
The iFlash-Anti- TPO assay is a sandwich immunoassay.
REF C80001, iFlash Wash Buffer (10×): phosphate
 Incubation: Total Anti- TPO in the sample, TPO coated buffered saline solution with 0.05% ProClin 300.
paramagnetic microparticles and Anti-TPO
REF C89996, reaction vessels.
acridinium-ester-labeled conjugate react to form a
sandwich complex. Controls: Commercial controls could be used.
 Wash: The unbound materials are washed away from WARNINGS AND PRECAUTIONS
the solid phase in a magnetic field.
IVD For in vitro diagnostic use
 Trigger of signal: The Pre-Trigger and Trigger
 The calibrators (CAL2 and CAL3) have been prepared
Solutions are added to the reaction mixture. The
exclusively from the blood of donors tested individually
resulting chemiluminescent reaction is measured as
and shown to be free from HBsAg, Anti-HCV, and
relative light units (RLUs).
anti-HIV-1/2 by approved methods.
 A direct relationship exists between the amount of total
 However, as no known test method can offer the
Anti-TPO in the sample and the RLUs detected by the
complete assurance that products derived from human
iFlash optical system.
sources will not transmit infection. Therefore, all
 Results are determined via a calibration curve, which humanized materials should be considered potentially
is instrument-specifically generated by 3-point infectious.
calibration and a master curve provided via the
 Exercise the normal precautions required for handling
reagent QR code.
all laboratory reagents.
 Disposal of all waste material should be in accordance
with local guidelines.
 Wear gloves when handling specimens or reagents.
 Clean and disinfect all spills of specimens or reagents
1/4 V2.0 English Ed.2018-02-01
Anti-TPO iFlash
Immunoassay Analyzer

using a suitable disinfectant. transfer only the clarified specimen without the lipemic
 iFlash Trigger solution contains sodium hydroxide material.
(NaOH) and should be avoided contact with eyes.  Ensure that residual fibrin and cellular matter have
been removed prior to analysis.
REAGENT HANDLING
 Use with caution in handling patient specimens to
 The reagents may not be used after the stated prevent cross-contamination.
expiration date.
 Do not use heat-inactivated samples.
 Avoid the formation of foam with all reagents.
 Ensure that the patient samples, calibrators and
 The reagents in the pack and calibrators are ready for controls are at ambient temperature (20–25°C) before
use measurement.
 Close the bottles of calibrator right after calibration and  Due to the possible evaporation, specimens and
store at 2–8°C. calibrators on the analyzers should be measured
 Do not pool reagents within a reagent kit or between within 2 hours.
reagent kits.
ASSAY PROCEDURE
 Prior to loading the iFlash-Anti-TPO reagent pack on
the system for the first time, suspend the  Refer to the system operating instruction or the online
microparticles by inverting the reagent pack slightly. help system for detailed information on preparing the
system.
 For further information on reagent handling
precautions during system operation, refer to the  The test-specific parameters stored in barcode on the
iFlash system operating instruction. reagent pack are read in. In case the barcode cannot
be read, enter the sequence numbers.
STORAGE AND STABILITY  Carry out calibration, if necessary.
Storage:  Place the calibrators CAL1, CAL2 and CAL3 in the
 Store at 2–8°C in an upright position. calibrator rack in the sample zone. Only keep
calibrators open during calibration.
 The kit may be used immediately after removal from
2-8°C storage.  Test application.
 Load samples (Use 20 μL of sample for each
Stability: determination in addition to the sample container and
 Unopened at 2–8°C: up to the stated expiration date. system dead volumes).
 Opened at 2–8°C: 28 days.  Click RUN, the iFlash System performs all the
 Store on-board: 28 days. functions automatically and calculates the results.

SPECIMEN COLLECTION AND PREPARATION CALIBRATION


 Serum or plasma (lithium heparin, sodium heparin  Traceability:This assay has been standardized against
potassium EDTA, and sodium citrate) are the Anti-TPO standard substance of China's food and drug
recommended samples. Other anticoagulants have verification research institute.
not been validated for use with the iFlash-Anti-TPO  Every iFlash-Anti-TPO reagent kit has a QR code label
assay. containing the specific information for calibration of the
 Ensure that serum specimens to form complete clot particular reagent lot.
prior to centrifugation.  To perform an iFlash-Anti-TPO calibration, test CAL1,
 Centrifuge the specimens. CAL2 and CAL3 in duplicate, and the predefined
 Store specimens at room temperature (20 to 25°C) for master curve is adapted to the analyzer.
no longer than 8 hours.  Once an iFlash-Anti-TPO calibration is accepted and
 If the testing will not be completed within 8 hours, stored, all subsequent samples may be tested without
refrigerate the samples at 2 to 8°C. further calibration unless:

 If the testing will not be completed within 14 days, or  After 28 days when using the same reagent lot.
for shipment of samples, freeze at -20°C or colder.  A reagent kit with a new lot number is used.
 Frozen specimens must be mixed thoroughly after  Controls are out of range.
thawing.  Required by pertinent regulations.
 The samples may be frozen for maximum 1 time.
 Centrifuge specimens with a lipid layer on the top, and
2/4 V2.0 English Ed.2018-02-01
Anti-TPO iFlash
Immunoassay Analyzer

MEASURING RANGE interchangeably.


 1 - 1,000 IU/mL  Samples containing an apparent Anti-TPO level as
high as 124,561 IU/mL did not exhibit a hook effect in
QUALITY CONTROL the iFlash-Anti-TPO assay.
Quality control materials should be run as single  The assay is unaffected by icterus (bilirubin < 10
determinations at least once every 24 hours when the test mg/dL), hemolysis (Hb < 500 mg/dL), lipemia
is in use, once per reagent kit and after every calibration. (Intralipid < 1,800 mg/dL) and total serum protein (< 10
Include commercially available quality control materials g/dL).
that cover at least two levels of analyte. Follow
 No interference was observed from rheumatoid factors
manufacturer’s instructions for reconstitution and storage.
up to a concentration of 2,000 IU/mL.
Each laboratory should establish mean values and
acceptable ranges to assure proper performance. Quality  No interference was observed from anti-nuclear
control results that do not fall within acceptable ranges antibodies up to a concentration of 500 AU/mL.
may indicate invalid test results.  No interference was observed from HAMA up to a
concentration of 600 ng/mL.
RESULT
PERFORMANCE CHARACTERISTICS
Calculation:
Below are the representative performance data, and the
The iFlash system automatically calculates the analyte results obtained in individual laboratories may differ.
concentration of each sample. The results are given in
IU/mL Precision
The precision of iFlash-Anti-TPO was determined using
Interpretation of Results
Anti-TPO reagents, samples and controls. Two serum
A study of with iFlash-Anti-TPO assay on samples from samples, consisting low and median concentration of
357 apparently healthy males of various age groups Anti-TPO were assayed.
yielded the following result:
th
The within run precision was determined by testing each
< 35 IU/mL (95 percentile) sample in replicates of 10 (n = 10), and calculating percent
It is recommended that each laboratory establish its own coefficient of variation (%CV). The results of the study are
expected reference range for the specific population. shown below:
Sample Mean (IU/mL) SD %CV
LIMITATIONS
 The iFlash-Anti-TPO assay is limited to the 1 40.24 2.15 5.34%
determination of Anti-TPO in human serum or plasma 2 397.12 18.86 4.75%
(lithium heparin, sodium heparin, potassium EDTA,
The between run precision was determined by testing
and sodium citrate). It has not been validated for use
each sample in duplicate, two separate runs daily for 20
with other types of plasma.
days (n = 80), and calculating percent coefficient of
 The use of serum separator (gel) blood collection variation (%CV). The results of the study are shown
tubes has been validated for use with this assay. below:
However, it is not possible to survey all manufacturers
Sample Mean (IU/mL) SD %CV
or tube types.
1 1.68
 The upper limit of the measuring range of this assay is 40.63 4.13%
1,000 IU/mL. Over-range samples may be diluted with 2 399.8 18.98 4.75%
wash buffer (1:20 recommended) and re-tested to
obtain an estimate of the actual concentration. Analytical Sensitivity
 If the results are inconsistent with clinical evidence, The detection limit representing the lowest measurable
additional testing is suggested to confirm the result. analyte level is 1 IU/mL, which can be distinguished from
zero. It is calculated as the value lying two standard
 For diagnostic purposes, the results should be
deviations above that of the lowest standard of the master
interpreted in light of the total clinical presentation of
curve (standard 1 + 2 SD, n = 20).
the patient, including symptoms, clinical history results.
 Specimens from heparinized patients may be partially Method Comparison
coagulated and erroneous results could occur due to A comparison of the iFlash-Anti-TPO assay (y) with a
the presence of fibrin. commercially available Anti-TPO assay (x) using clinical
 The results from an alternative assays (i.e. EIA or RIA) samples was performed, and the curve is fitted with Linear
may not be equivalent and cannot be used regression)
3/4 V2.0 English Ed.2018-02-01
Anti-TPO iFlash
Immunoassay Analyzer

y = 1.007x + 42.35 ANNEX A:


r = 0.9945 Explanation of abbreviation
Sample concentration: 1 – 1,000 IU/mL Abbreviation Explanation
Number of samples measured: 88

REFERENCES Product No.


1. Trotter WR, Belyavin G, Waddams A. Precipitating
and complement-fixing antibodies in Hashimoto’s Calibrator
disease. Proc Royal Soc Med 1957; 50: 961-2.
2. DeGroot LJ, Niepomniszcze H. Biosynthesis of thyroid Reagent
hormone: basic and clinical aspects. Metabolism 1977;
26: 665-718.
Number of tests
3. Mariotti S, Caturegli P, Piccolo P, Barbesino G and
Pinchera A. Antithyroid peroxidase autoantibodies in
thyroid diseases. J Clin Endocrinol Metab 1990; 71: Manufactured by
661-9.
4. Toda S, Koike N, Sugihara H. Cellular integration of EU Representative
thyrocytes and thyroid folliculogenesis: a perspective
for thyroid tissue regeneration and engineering.
EC Declaration of Conformity
Endocrine Journal 2001; 48 (4): 407-425.
5. Feldt-Rasmussen, U. Analytical and clinical
performance goals for testing autoantibodies to Caution
thyroperoxidase, thyroglobulin and thyrotropin
receptor. Clinical Chemistry1996; 42:1, 160-3. Instructions for use
6. Feldt-Rasmussen U, Høier-Madsen M, Bech K,
In vitro diagnostic medical
Blichert-Toft M, Anti-thyroid peroxidase antibodies in
thyroid disorders and non-thyroid autoimmune device
diseases. Autoimmunity 1991; 9: 245–253.
Lot No.

SHENZHEN YHLO BIOTECH CO., LTD. Date of manufacture


1st-4th Floor, No.5 Building, Lishan Industrial
Area, Xinghai Road, Nanshan District, Expiry date
Shenzhen 518054, P.R. China

Biohazard Symbol
Wellkang Ltd (www.CE-marking.eu)

Suite B, 29 Harley St., London W1G 9QR, UK Pictograms for Caution

Pictograms for Hazardous to the


aquatic environment

4/4 V2.0 English Ed.2018-02-01

You might also like