You are on page 1of 6

THE JOURNAL OF ALTERNATIVE AND COMPLEMENTARY MEDICINE

Volume 15, Number 5, 2009, pp. 495–500


ª Mary Ann Liebert, Inc.
DOI: 10.1089=acm.2008.0331

Characteristics of Electrical Skin Resistance at Acupuncture


Points in Healthy Humans

Sybille Kramer, M.D.,1 Kathrin Winterhalter, M.D.,1 Gabriel Schober, M.D.,1 Ursula Becker, Dipl.-Stat.,1
Bernhard Wiegele, Dr.-ing,2 Dieter F. Kutz, Ph.D.,3 Florian P. Kolb, Ph.D.,3 Daniela Zaps,1
Philip M. Lang, M.D.,1 and Dominik Irnich, M.D., Ph.D.1

Abstract

Objectives: The aim of this study was to evaluate the phenomenon of electrical skin resistance (ESR) changes at
different acupuncture points (APs).
Setting: This single-blinded study was performed at the hospital of the University of Munich.
Design: Six common APs were measured (TE5, PC6, LU6, ST36, SP6, GB39) in 53 subjects. Subgroups were
formed with varying time intervals for follow-ups (1 minute, 1 hour, 1 week) and a varying grade of reduction of
the stratum corneum.
Methods: Electrical skin resistance measurements (ESRMs) were taken from a skin area of 66 cm using an array
consisting of 64 (88) electrodes. The electrodes corresponding to the AP were located and the ESRM results
were compared to those of the surrounding electrodes. The methodological setting made it possible to minimize
major influence factors on electrical skin impedance measurements.
Results: A total of 631 ESRMs was evaluated: In 62.8% of the measured APs, no significant ESR difference was
found. In 234 (37.2%) of the ESRMs, the ESR at the AP was significantly different from the surrounding skin area,
with 163 (25.9%) points showing a lower and 71 (11.3%) points showing a higher ESR. Reproducibility was
extremely high after 1 minute but was low after 1 hour and 1 week.
Conclusions: This study shows that electrical skin resistance at APs can either be lower or higher compared to
the surrounding area. The phenomenon is characterized by high short-term and low long-term reproducibility.
Therefore, we conclude that APs might possess specific transient electrical properties. However, as the majority
of the measured APs did not show a changed ESR, it cannot be concluded from our data that electrical skin
resistance measurements can be used for acupuncture point localization or diagnostic=therapeutic purposes.

Introduction diagnosis-and-treatment-systems are widely used in the


clinical practice today.3,4

A widespread theory claims that meridians and acu-


puncture points (APs) possess certain properties, such as
a lower electrical resistance or impedance (and consequently
Many investigators have tried to show a difference be-
tween the skin impedance of an AP and the impedance of its
surrounding area.5–9 A review published earlier this year
an increased electrical conductivity) compared to the sur- found as many as 320 citations dealing with this topic. Only
rounding tissue.1,2 Furthermore, it has been assumed that the 18 studies met their evaluation criteria and were generally
difference in the electrical skin resistance (ESR) may reflect described as of poor quality and limited by a small sample
upon the ability of the meridians and acupuncture points to size. All of them used different methods and their results
react to physiologic processes and pathologies in the body. were controversial.10 Even less scientific data exist to support
Based upon this theory, numerous AP location devices and the reliability of electrical skin resistance measurements

1
Multidisciplinary Pain Centre, Department of Anaesthesiology, University of Munich, Munich, Germany.
2
Fachbereich Elektrotechnik, FH Augsburg, Augsburg, Germany.
3
Department of Physiological Genomics, Institute of Physiology, University of Munich, Munich, Germany.

495
496 KRAMER ET AL.

(ESRMs).6,11–13 Apart from deficiencies in study design and The objective of this trial was to determine whether and
statistical analysis, the main reason for the lack of scientific to what degree differences in ESR at APs actually do exist.
evidence can be found in the vulnerability to various inter-
ferences of the measuring procedure itself.
Factors that can interfere with ESRMs include environ- Materials and Methods
mental disturbances (temperature, relative air humidity), Participants
physiologic differences (body temperature, blood circulation,
skin humidity, skin surface topography, skin thickness), or Fifty-three (53) volunteers (18–38 years, mean age 26.3
technical issues (size, angle, pressure, and duration of elec- years, 24 male, 29 female) were recruited on the campus of
trode application and measurement).6,14 the Ludwig-Maximilians University, Munich, Germany. All
Most of numerous devices for ESRM consist of one elec- volunteers signed an informed consent form before partici-
trode (of varying size) that is placed at the tip of a pen-shaped pation. The study was approved by the local Ethics Com-
device, which is pressed manually against the skin by the mittee of the University of Munich. Inclusion criteria were
examiner.9,13,15–17 It is known that variation of pressure, general good health, age 18–40 years, and the ability to
angle, or duration of the measurement can significantly in- understand and sign an informed consent form. Exclusion
fluence the results.8,9 Furthermore, repeatedly pressing the criteria were pregnancy, current medication, acute or chronic
electrode tip into the skin or scratching over the skin surface medical problems of any kind, skin irritations, and previous
may result in hyperemia and stratum corneum abrasions that treatments with acupuncture or transcutaneous electrical
increase skin conductivity and lower the electrical resistance. nerve stimulation.
As the stratum corneum is claimed to form the major part of
ESR,18 small abrasions of its surface could lead to a reduced Setting and procedures
layer and thus a reduced resistance at the damaged area. Even
The study was carried out in the Multidisciplinary Pain
though this might considerably change skin resistance values,
Unit of the Ludwig-Maximilians University, Munich, Ger-
it has never been investigated whether this has a relevant
many in a standardized setting. Ambient room temperature
impact on ESRMs of APs.
was 22.58C  1.38C (mean  standard deviation), outside tem-
Moreover, controversy exists about the reproducibility of
side temperature 9.58C  8.68C ( 3.1, 23.3), relative air hu-
ESMRs delivered by the usual measuring devices.8,11 There-
midity outside was 29.7%  7.6%, mean barometric pressure
fore, a comprehensive investigation of the phenomenon of
was 1027.2  7.2 hPa, and body temperature was 36.28C 
skin resistance should take into account the following:
0.28C. Subjects were advised to rest for 15 minutes be-
1. Reproducible measuring process fore measurements were taken. If necessary, the skin was
2. Minimization of disturbing factors shaved in the area of the APs on the evening before mea-
3. Evaluation of the influence of the stratum corneum layer surements.

FIG. 1. Study design: Test persons were divided into five subgroups with varying applications of tape stripping (TS) and
different time intervals for follow-up measurements for evaluation of reproducibility. Follow-up measurements were not
used for evaluation of electrical skin resistance changes. APs, acupuncture points.
ELECTRICAL SKIN RESISTANCE AT ACUPUNCTURE POINTS 497

Volunteers were divided into five subgroups with 4–6 male Instruments
and 4–6 female participants in each subgroup (Fig. 1). In each
An impedance meter array was specially designed for the
subgroup there was a variation in time intervals between
purpose of this study.20 It consists of a field of 64 electrodes
follow-ups and grade of reduction of the stratum corneum by
on a flexible plastic foil surface of 66 cm with a distance of
tape stripping (TS). Follow-up measurements were per-
8 mm between the centers of each electrode (Fig. 2A).
formed at short (group I: 1 minute, four follow-ups), medium
Voltage was measured through a voltmeter between each
(group II: 1 hour, one follow-up), and long (groups III, IV, and
of the electrodes and a reference electrode, which was held in
V: 1 week, one follow-up) time intervals. The array was left on
the participants’ hands during measurements. Electrical skin
the skin between ESRMs with 1-minute intervals. Measure-
resistance was indicated in arbitrary units for each of the
ments with 1-week intervals were performed at the same time
measuring electrodes. All the electrodes were scanned re-
of the day as the first measurement in order to avoid variation
peatedly one after another (average scanning time 15 ms).
due to changes in the circadian rhythm. Reduction of the
This method of fast scanning allowed an almost simultaneous
stratum corneum was done before the start of ESRMs and
measurement of the whole array within 1 second. We used
varied by the number of applications of TS: groups I and III:
direct current (DC, max flow 20 mA, mean voltage 4.8 V).
no TS; groups II and IV: 3TS; group V: 5TS.
These technical details correspond approximately to other
commercially available devices. The array has been approved
Tape stripping by the German Technical Inspection Authority for security.
The stratum corneum has been shown to form the major
Skin resistance measurements
part of the ESR, and with its reduction the ESR is known to
decrease.18 Therefore, we wanted to ensure that our results First the array was placed on the skin area in which the
were not caused by irregularities in the thickness of its layer. acupuncture point is known to be found. A pad of medical gel
A common way to reduce the stratum corneum layer in was placed upon the array to distribute fixation pressure
dermatological research is the procedure of tape stripping evenly over the whole array. Array and gel pad were then
(TS). In this study, it was performed according to the pro- fixed with a blood pressure cuff (pressure 10 mm Hg). The
tocol described by Jacobi et al.19 An adhesive tape (‘‘tesa measuring process was started and terminated after 30
film’’ 5529, 19 mm, Beiersdorf, Hamburg, Germany) was seconds. The resistance values at the end were used for eval-
pressed against the skin and then removed again with cor- uation. The position of the array was marked on the skin and
neocytes stuck to it. Five repetitions of TS have been shown the array was removed (Fig. 2A). Afterward, the acupuncture
to lead to a removal of approximately 25% of the stratum points were localized and marked by an acupuncturist, who
corneum and a near total removal of substances that had was blinded regarding the results of the measurement (Fig.
penetrated into the horny layer.19 2B). An array template was used to match the array electrodes

FIG. 2. Electrical skin resistance measurements (ESRMs): (A) ESRM array consisting of 64 electrodes on a 66-cm square
plastic foil. (B) Localization of the acupuncture point (AP) was performed after the ESR. (C) The localization of the AP is
matched with the electrodes on the array. Ink marks are used for the exact placement of the template. (D) Screenshot of one
exemplary ESRM: ESR was indicated by colors and arbitrary units (shown here in black and white). The ink cross marks the
localization of the AP and the ink square shows the electrodes defined as AP.
498 KRAMER ET AL.

and marked APs (Fig. 2C). A square of 33 electrodes sur-


rounding the detected electrode was assumed to correspond
to the AP. The ESR values of these nine electrodes were
compared to the other 55 electrodes of the array (Fig. 2D).

Acupuncture points
The following APs were selected: TE5, PC6, LU6, ST36, SP6,
and GB39. They were chosen for reasons of clinical relevance as
well as their location on the forearm or lower leg, which allows
fixation of the array with a steady pressure. The location of the
APs was determined according to standards of TCM.21 FIG. 3. Results of electrical skin resistance measurements
(ESRMs): 631 ESRMs were evaluated. One hundred and
Statistical analysis sixty-three (163) acupuncture points (APs) showed a signif-
icantly lower and 72 APs a significantly higher ESR than the
Microsoft Excel and SPSS were used for statistical analysis. surrounding areas.
Correlation of follow-up measurements was estimated by
calculating the Spearman correlation coefficient. For corre-
lations, the pattern of the ESR distribution was considered to changed ESR at the AP in comparison to the ESR of the sur-
be more important than the absolute ESR values, as the rounding skin area. Two hundred and thirty-four (234)
question examined in this study was measurable changes in (37.2%) of the measured APs showed an ESR that was sig-
ESR and their connection to the AP location. Correlation was nificantly different, with 163 (25.9%) APs showing a lower
estimated in group I with five repeat measurements with and 71 (11.3%) APs showing a higher skin resistance than the
intervals of 1 minute, in group II with one follow-up mea- surrounding skin areas (Fig. 3). So, in contrast to the theory
surement after 1 hour, and in groups III, IV, and V with a that the ESR at acupuncture points is lower than the ESR of
follow-up after 1 week. Correlation coefficients were calcu- the surrounding areas, we found that electrical skin resistance
lated between each pair of successive measurements (1st and at APs can also be significantly higher. ESR values in general
2nd, 2nd and 3rd, 3rd and 4th, 4th and 5th) and between the varied broadly between approximately no impedance (»high
1st and 5th measurement. conductance) and approximately no conductance at all (»high
For the evaluation of ESR changes at APs, only first mea- impedance), even within one array measurement.
surements of a series were used. Differences in the ESR were
evaluated employing the Wilcoxon rank sum test. This is a Acupuncture points
nonparametric test transforming the absolute resistance
Not every acupuncture point (TE5, PC6, LU6, ST36, SP6,
values into ranks. The resulting ranks of the pooled sample
GB39) showed the same percentage of measurements, with
are used for the calculation of the test statistic, which indi-
higher and lower ESR within-acupuncture-point-analysis
cates if the distribution of the ESR between two groups of
showing a distribution of 7.7%–15.0% with a significantly
electrodes (AP versus surrounding) is significantly different.
higher skin ESR and 19.6%–31.8% of points with a signifi-
We decided to define an area of 33 electrodes around the
cantly lower ESR (Table 1).
located point as the region of interest (Fig. 2D) and compare
these 9 electrodes to the rest of the electrodes on the array.
Reproducibility
Electrodes on the array corresponding to other close-by APs
were excluded from evaluation. ESRMs were excluded, if the Reproducibility of measurements was tested using the
AP was located too far on the edge of the measured area, and Spearman correlation coefficient (see Statistical analysis sec-
the lesser number of electrodes could therefore influence tion above). In group I (5 repeated measurements, intervals 1
validity of the evaluation. P-values of <0.05 were considered minute), out of 583 correlation coefficients only 34 (5.8%)
as statistically significant. False discovery rate was used as were <0.85 and 236 (40.5%) were >0.95, indicating highly
an adjustment for multiple testing. The significance level repeatable results. After 1 hour (group II) and after 1 week
between the groups with different repetitions of TS and the (groups III, IV, and V), there was no clear correlation be-
different APs was determined using w2 test. tween first and second measurement, with only 1 of 106
correlation coefficients >0.85, 36 (33%)<0.2, and 17 (15.6%)
Results showing negative correlation in group II and only one of 340
correlation coefficients >0.85, 154 <0.2 and 72 showing a
The aim of this study was to measure the ESR at acu- negative correlation in group III, IV and V. These results
puncture points and to compare changes in ESR to the lo- indicate that there is a high short-term reliability and a low
cation of the acupuncture point (AP). More than 1450 ESRMs reliability of the measurements with medium- and long-term
were performed in a collective of 53 healthy volunteers fol- follow-up.
lowing high-quality standards regarding methodology and
statistical evaluation. Influence of stratum corneum
To examine the influence of the reduction of the stratum
Changes in ESR
corneum layer on ESRMs, we measured ESRMs after zero,
We evaluated a total of 631 measurements at six different three, and five TS: In groups I and III (no tape stripping) 273
acupuncture points: 396 (62.8%) showed no significantly ESRMs were evaluated: 60 APs (21.9%) showed a significantly
ELECTRICAL SKIN RESISTANCE AT ACUPUNCTURE POINTS 499

Table 1. Within-Acupuncture-Point-Analysis of Electrical Skin Resistance Measurements (ESRMs)

Acupuncture
Point
ESR
Properties TE5 GB39 LU6 ST36 SP6 PC6

ESRMs (total) 107 108 107 103 101 104


ESR $ 21 (19.6%)* 28 (25.9%) 34 (31.8%)* 29 (28.2%) 21 (20.8%) 32 (30.8%)
ESR # 16 (15.0%) 11 (10.2%) 9 (8.4%) 13 (12.6%) 12 (11.9%) 8 (7.7%)

The acupuncture point (AP) LU6 showed the highest percentage (31.8%) and the TE5 showed the lowest percentage of ESRMs with
significantly lower ESR. This difference was statistically significant (*p < 0.05). The highest percentage (15%) of APs with significantly higher
ESR was found in TE5. Three hundred and ninety-seven (397) APs showed no significant differences in ESR.

lower ESR, whereas 38 APs (13.9%) had a higher resistance. In 8 mm between the centers of the electrodes. This is a possi-
group V (5TS) 24 (20.7%) acupuncture points of 116 showed bility we cannot exclude. It raises general questions in
lower and 14 (12.1%) higher skin impedance. The percentage measuring ESR: How big is an AP and the skin area in which
of APs with significantly lower ESR was highest in groups II we can expect ESR changes due to the acupuncture point?
and IV (3TS) with 32.6%. In the same group, the percentage The location and the size of an AP in the literature can differ
of APs with higher ESR was lowest (7.9%). up to more than 1 cm2. We hypothesized that ESR changes at
AP may not strictly be limited to an area too small to be
Discussion registered by our array. An array with smaller electrodes and
a smaller distance between the electrodes would provide a
In this study, we performed ESRMs on six common acu- higher resolution of the measured area, but would also
puncture points in 53 subjects with a multi-electrode device. necessarily lead to a smaller screening area. We decided to
Because the absolute skin resistance values can vary strongly start with a larger screening area even if this meant that the
inter- and intra-individually, we evaluated each ESRM resolution of our array was lower.
individually in respect to the ESR of the skin area at the AP
compared to the ESR of the surrounding skin area. Then the
results were summarized and the percentage of APs with ESR changes in the course of time
ESR changes was calculated: In a total sample size of 631, we Our reproducibility testing shows a high reproducibility
found 62.8% of the tested AP to have no significantly chan- for short time intervals between the measurements. Follow-
ged skin resistance. A significantly changed ESR at the AP ups had a low reproducibility. These findings are in ac-
was shown in 32.7% (26.9% decreasing, 11.3% increasing cordance with those of other investigators.6,17 One simple
ESR). These results are partly in line with previous high- explanation might be a change of environmental conditions.
quality studies, most of which seem to show a trend for However, especially for the reproducibility testing after 1
negative results to this topic (review article10,22), but it still hour, we cannot exclude artifacts caused by the procedure of
must be admitted that there is some evidence for a physio- TS. However, the reproducibility after 1 week was low in
logic basis of the phenomenon of ESR changes at an AP. One every group irrespective of whether TS was performed be-
example for this is the study by Pearson et al.22 While their fore. The observation that the values of skin impedance
overall results showed no statistically different skin imped- measurements can shift and vary widely inter- and intra-
ance at AP compared to control points, several of their hy- individually has also been described before.6,8,15,17 So in our
pothesis tests did. Thus, perhaps the theory of electrical skin opinion, the likeliest explanation is that electrical skin im-
resistance changes at APs is nothing to be generally proved pedance shifts physiologically like transepidermal water loss
or rejected, but something that can be measured at certain or skin hydration.23
points in certain persons at a certain time. Nevertheless, we
can only agree with Pearson et al. in their conclusion that
caution is warranted in using and interpreting results from Increase or decrease in ESR
electrodermal screening devices. A few studies reported increasing ESR at AP24 or a general
increase of skin impedance.22 We found a certain percentage
Technical issues of measurements of each AP to have a significantly higher
skin resistance. To our knowledge, we are the first ones to
ESRMs are technically complex.14 We believe that we describe this finding. If ESR shifts in the course of time, there
overcame most disturbances in skin impedance testing by is no reason why it should not be increased at certain points
our evaluation: The array was constructed to ensure equal and certain locations.
technical conditions for all electrodes. It was completely
measured within 1 second. Within this time interval, it is
Stratum corneum
unlikely that environmental conditions had changed. At any
rate, by comparing the electrodes of each measurement to Former investigators described that the major component
themselves, most disturbance factors should be eliminated of ESR is generated by the stratum corneum layer.16,18,25
by statistical analysis. We therefore hypothesized that changes in the thickness
One limitation might be that in the negative cases we (e.g., small abrasions) could significantly influence ESRMs.
simply failed to measure the AP itself due to the distance of Our results show that the group with a medium reduction of
500 KRAMER ET AL.

the stratum corneum layer showed the highest percentage of 9. Barlea N-M, Sibianu H, Ciupa RV. Electrical detection of
APs with significantly lower ESR. There is no study with acupuncture points. Acta Eletrothenica Napocensis 2000;41:
a similar objective to compare to our results. Thus, one 16–19.
explanation might be artifacts caused by the procedure of TS, 10. Ahn AC, Colbert AP, Anderson BJ, et al. Electrical properties
or maybe the medium grade of stratum corneum reduction of acupuncture points and meridians: A systematic review.
achieved by threefold TS provides the best balance of a Bioelectromagnetics 2008;28:245–256.
smooth surface with minimum damage to the layer. How- 11. Colbert AP, Hammerschlag R, Aickin M, McNames J. Re-
ever, as it is well known that the electrode size influences the liability of the Prognos electrodermal device for measure-
depth of the tissue layer that is being measured (the smaller ments of electrical skin resistance at acupuncture points.
J Altern Complement Med 2004;10:610–616.
the deeper14), future trials should conduct some pilot
12. Treugut H, Gorner C, Ludtke R, et al. Reliability of energetic
measurements to find out the optimum number of TS for the
terminal point diagnosis according to Mandel [in German].
electrode size used.
Forsch Komplementarmed 1998;5:224–229.
13. Falk CX, Birch S, Avants SK, et al. Preliminary results of a
Conclusions new method for locating auricular acupuncture points.
Acupunct Electrother Res 2000;25:165–177.
This study shows that ESR at APs can either be lower or— 14. Ahn AC, Martinsen OG. Electrical characterization of acu-
in contrast to common belief—higher compared to the puncture points: Technical issues and challenges. J Altern
surrounding area and that this phenomenon shows a high Complement Med 2007;13:817–824.
short-term and a low long-term reproducibility. Therefore, 15. Colbert AP, Hayes M, Aickin M, Hammerschlag R. Physiolo-
we conclude that APs might possess specific transient elec- gic variability of electrical skin resistance measurements at the
trical properties. However, as the majority of the measured ting acupuncture points. Med Acupuncture 2006;17:12–19.
APs did not show a changed ESR and taking into account 16. McCarroll GD, Rowley BA. An investigation of the existence
that reproducibility was given only in the short term, it of electrically located acupuncture points. IEEE Trans
cannot be concluded from our data that ESR measurements Biomed Eng 1979;26:177–181.
can be used for localization of acupuncture points or diag- 17. Semizzi M, Senna G, Crivellaro M, et al. A double-blind,
nosis or therapy. placebo-controlled study on the diagnostic accuracy of an
electrodermal test in allergic subjects. Clin Exp Allergy 2002;
32:928–932.
Acknowledgments
18. Yamamoto T, Yamamoto Y. Electrical properties of the epi-
This study was supported by the German Medical Acu- dermal stratum corneum. Med Biol Eng 1976;14:151–158.
puncture Association (DÄGfA). 19. Jacobi U, Meykadeh N, Sterry W, Lademann J. Effect of the
vehicle on the amount of stratum corneum removed by tape
stripping. J Dtsch Dermatol Ges 2003;1:884–889.
Disclosure Statement 20. Wiegele B, Schober G, Kuder J, et al. A new sensor technique
No competing financial interests exist. for measurements of electrical potential profiles of human
skin at acupuncture points. Forsch Komplement Med 2006;
13:227–232.
References 21. Cheng X. Chinese Acupuncture and Moxibustion. Beijing:
1. Reichmanis M, Marino AA, Becker RO. D.C. skin conduc- Foreign Languages Press, 1987.
tance variation at acupuncture loci. Am J Chin Med (Gard 22. Pearson S, Colbert AP, McNames J, et al. Electrical skin
City NY) 1976;4:69–72. impedance at acupuncture points. J Altern Complement
2. Reichmanis M, Marino AA, Becker RO. Electrical correlates Med 2007;13:409–418.
of acupuncture points. IEEE Trans Biomed Eng 1975;22:533– 23. Cravello B, Ferri A. Relationships between skin properties and
535. environmental parameters. Skin Res Technol 2008;14:180–186.
3. Voll R. Twenty years of electroacupuncture diagnosis in 24. Ahn AC, Wu J, Badger GJ, et al. Electrical impedance along
Germany: A progress report. Am J Acupuncture 1975;3:7–17. connective tissue planes associated with acupuncture me-
4. Lam FMK, Tsuei JJ, Zhao Z. Bioenergetic regulatory mea- ridians. BMC Complement Altern Med 2005;5:10.
surement instruments and devices. Am J Acu 1988;16:345– 25. Lawler JC, Davis MJ, Grifith EC. Electrical characteristics of
349. the skin: The impedance of the surface sheath and deep
5. Hyvarinen J, Karlsson M. Low-resistance skin points that tissues. J Invest Dermatol 1960;34:301–308.
may coincide with acupuncture loci. Med Biol 1977;55:88–94.
6. Cho SH, Chun SI. The basal electrical skin resistance of
acupuncture points in normal subjects. Yonsei Med J 1994; Address reprint requests to:
35:464–474. Dominik Irnich, M.D., Ph.D.
7. Kwok G, Cohen M, Cosic I. Mapping acupuncture points Interdisziplinäre Schmerzambulanz Innenstadt der LMU
using multi channel device. Australas Phys Eng Sci Med Pettenkoferstr. 8a
1998;21:68–72. 80336 Munich
8. Mayer-Gindner A, Lek-Uthai A, Abdallah O, Bolz A. Newly Germany
explored electrical properties of normal skin and special skin
sites. Biomed Tech 2004;49:117–124. E-mail: dominik.irnich@med.uni-muenchen.de

You might also like