Preset parameters
Group Name Default Minimum Maximum Comment
0 Start up
0-0 Acetate/Bicarbonate mode Acetate or Bicarbonate at start-up
Acetate/Bicarb. Bicarb. Acetate Bicarb.
0-1 Double/single needle mode Double or Single needle at start-up
DN/SN DN DN SN
0-3 Priming type Normal = Machine starts up in normal priming mode
Type NORMAL NORMAL AUTO Auto = Machine starts up in Auto priming mode
1 Configuration
1-0 Language Language selection
Language English Depending on prom version (standard or extended)
1-1 Arterial pressure Machine equipped with arterial pressure transducer
Arterial pressure Inst Inst Not Inst
1-2 Arterial guard Machine equipped with arterial pressure switch
Arterial guard Not Inst Inst Not Inst
1-3 Arterial pressure preset Not inst = Both the lower and the upper arterial limit will centralize
to the measured arterial value according to arterial window with
set in preset 2-15.
Arterial pressure preset Not Inst Not Inst Inst Inst = Only the lower limit will centralize to the measured arterial
value. The upper limit is possible to preset to anything between
-200mmHg and +200mmHg with preset 2-16..
1-4 Heater power Heater rod power
Heater Power(W) 3*580 1000, 1300, 1750, 1000 = 100V mains voltage
3*580 1300 = 115V mains voltage
1750 = 230V mains voltage
3*580 = 230V mains voltage (new heater)
1-5 pH cell Machine equipped with pH cell
pH Cell Not Inst Inst Not Inst
1-6 BiCart holder Machine equipped with BiCart holder
BiCart holder Inst Inst Not Inst
Page 1/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Group Name Default Minimum Maximum Comment
1-7 BiCart emptying Fluid or powder emptying of BiCart
BiCart emptying Fluid Fluid Powder If this preset is set to powder, the BiCart will be completely
emptied.
1-8 BiCart priming, extra flush Extra flush time during BiCart priming.
Flush (sec) DISABLED 10 600
1-9 Degassing pressure Degassing pressure (negative sign omitted)
Degass pressure -610 -300 -700 The value must be changed depending on altitude.
Page 2/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Not Inst = This must be confirmed when exiting the preset mode.
No back filtration warning will be given when setting a negative
TMP low limit during treatment.
1-28 Blood pump stop in priming Blood pump stops when machine detects blood
Blood pump stop in priming Not Inst Not Inst Inst When this preset is set to Inst, the blood pump will automatically
stop when blood is detected by the priming detector.
1-29 Click Knobs Click knobs installed
Page 3/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Click Knobs Inst Not Inst Inst
1-30 Installed panel type Panel type installed
Installed panel type 2 1 2 1 = old panel type
2 = new panel type
See GI 146 for more information
1-31 Blood pressure monitor BPM installed
BPM Not Inst Not Inst Inst
1-32 Profiling Profiling installed
Profile Not Inst Not Inst Inst
1-33 Inlet conductivity check False = No inlet cond
In cnd check False False True True = Max 2,5 mS/cm
True = It is possible to, under the control and confirmation of the
operator, run the machine with higher inlet conductivity. If the inlet
conductivity is above 2,5 mS/cm the error code C CFM 028 003
will occur.
1-34 Variable fluid flow Inst = Variable flow kit installed.
Var. flow Inst Inst Not Inst Inst Not Inst = Variable flow kit not installed.
1-35 Night light True = Possible to turn off the panel light by pressing mute button
for 3 seconds.
False False True False = Not possible to turn off the panel light.
1-36 No water consumption mode True = INVA close, pumps stop, signal to turn of WRO when
concentrate standby mode is activated.
No water consump False False True False = No water consumption during concentrate standby mode.
1-37 Concentrate standby Fch 0 = No automatic concentrate standby mode after Fch.
Standby Fch (Sec) 0 0 3600 1-3600 = Seconds after Fch until concentrate standby mode is
activated.
Page 4/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
1-38 Concentrate standby priming 0 = No automatic concentrate standby mode after priming.
Standby prim (Sec) 0 0 3600 1-3600 = Seconds after priming until concentrate standby mode is
activated.
1-40 WRO 300 H installed Inst = Possible to perform an integrated heat with a WRO 300 H
WRO 300 H inst Not Inst Not Inst Inst Not Inst = Not possible to perform an integrated heat with a WRO
300 H
1-41 Installed RTC type The real time clock on the protective driver board.
RTC Type 2 1 2 1 = Old type of RTC, Prot. driver board K16646006 and below.
2 = New type of RTC Prot. driver board K16646008 and above.
Page 5/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Page 6/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
2-4 Heparin stop limit Automatic stop time for heparin pump (in minutes before end of
treatment)
Heparin stoptime 00:20 00:00 09:59
2-5 Temperature Set temperature value and operator LL and HL (low/high level)
Temp 35.0, 37.0, 39.0 33.0 40.0
2-6 pH pH low and high limits
pH 6.00, 8.00 5.00 9.00
2-7 TMP limits TMP start-up limits
TMP -100, 400 -200 500
2-8 TMPwindow TMP window size during treatment (diffusion)
Width (mmHg) 200 10 300
TMP low part 25 20 80 Part (%) of the TMP window below the set point (diffusion)
2-9 TMP window isol. UF TMP window size during treatment (isolated UF)
Width (mmHg) 200 10 300
TMP isol low part 75 20 80 Part (%) of the TMP window below the set point (isolated UF)
Page 7/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
limits (mmHg) 0 -200 250 Only active if preset 1-3 is set to Inst.
2-17 Na set acetate (mmol/l)
Acetate Na set 130.0, 140.0, 160.0 130.0 160.0 Sodium value (acetate mode) and operator LL and HL (low/high
level)
2-18 Ac conc. acetate Active Acetate bank during start-up
Ac conc. select ALT1 ALT1 ALT3
2-19 Na set bicarbonate (mmol/l) Sodium value (bicarbonate mode) and operator LL and HL
(low/high level)
Bicarb Na set 130.0, 140.0, 160.0 130.0 160.0
2-20 HCO3 set bicarbonate (mmol/l) Bicarbonate value and operator LL and HL (low/high level)
Bicarb HCO3 set 20.0, 34.0, 40.0 20.0 40.0
2-21 Bc conc. bicarbonate Active Bicarbonate bank during start-up
Bc conc. select ALT1 ALT1 ALT3
2-22 Concentrate pump windows
2-22-0 Acetate pump B window (%) Maximum allowed pump speed deviation (acetate)
Ac pumpB window % 10 0 30
2-22-1 Bicarb. pump A window (%) Maximum allowed pump speed deviation (A-concentrate)
Bc pumpA window % 10 0 30
2-22-2 Bicarb. pump B window (%) Maximum allowed pump speed deviation (Bicarb-concentrate)
Bc pumpB window % 20 0 40
2-22-3 Composition deviation ( % ) Maximum pump speed ratio
Comp. dev. % 15 0 30 Maximum difference between pump A and pump B
2-23 UF volume and time/Diffusion
UF vol Diff. (l) 0.00 0.00 10.00 UF volume, start-up value
Time Diff. 04:00 00:00 09:59 Treatment time, start-up value
2-24 UF volume and time/Seq. -isol.
UF vol Isol. UF 0.00 0.00 10.00 UF volume (isolated UF), start-up value
Time Isol. UF 00:00 00:00 09:59 Treatment time (isolated UF), start-up value
2-25 UF stop timeout Attention delay time for stopped ultrafiltration
Timeout 00:05 00:01 01:00
2-26 UF rate
2-26-0 Minimum UF rate 0.10 0.00 4.00 Start-up value for minimum UF rate
Page 8/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
UF rate (l/hour) 0.00, 4.00 0.00 4.00 Minimum allowed UF rate, high and low limits
2-26-1 Maximum UF rate Maximum allowed UF rate
UF rate (l/hour) 4.00 0.00 5.00
2-26-2 Priming UF rate Start-up value for priming UF rate
UF rate (l/hour) 0.10 0.00 4.00
2-26-3 UF gain coefficient Maximum allowed deviation between actual and calculated UF rate
(UF rate limit = “UF Gain Max gain” x calculated UF rate)
UF Gain Max gain 1.20 1.00 2.00
2-27 DialysisFlow Limit for low flow attention. If the machine is equipped with
Limits (ml/min) 425 200 800 adjustable flow this value will be set to the set value –75 ml/min
automatically.
2-28 Max conductivity difference Maximum allowed conductivity difference
Limits (mS/cm) 0.3 0 1.0 Between cond A and cond B during FCH
2-29 Max temperature limits Maximum allowed temperature limits
Temp limits 33,40 33 40
2-30 Max pH limits Maximum allowed pH limits
pH limits 5.00, 9.00 5.00 9.00
2-31 Max venous pressure limits Maximum allowed venous pressure limits
Limits (mmHg) -100, 500 -100 500
2-32 Max arterial pressure limits Maximum allowed arterial pressure limits
Limits (mmHg) -700, 750 -700 750
2-33 Variable fluid flow value 500 300 700 Start up value
Fluid flow 300, 700 300 700 Minimum fluid flow allowed, Maximum fluid flow allowed
3 Override times
3-0 AirDetOverrideTime Air detector override time (0 = inactivated)
AirDet Time (s) 0 0 120 In alarm situations it can be possible to override the alarm actions
for a limited time.
Page 9/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
4 Auto priming
4-0 Auto priming install Auto priming feature active
Auto-priming Not Inst Inst Not Inst
4-1 Auto priming default ALT Active auto-priming bank during start-up
Default Alt ALT1 ALT1 ALT3
4-2 Autopriming ALT 1
4-2-0 Label Label in ALT1
Label DIA 1
4-2-1 Fill flow Blood pump speed during fill of dialyzer
FillFlow (ml/min) 100 0 500 This is the initial flow rate when
filling the dialyzer and blood lines
4-2-2 Fill time Blood pump running time during fill of dialyzer
Fill time (s) 180 0 3600 This is the time for filling the dialyzer
and blood lines completely.
4-2-3 Pressure test build time Time-out if venous pressure high limit not reached
Pr. build time (s) 0 0 30 The venous clamp closes. This is the time for
building up the venous pressure alarm limit.
Page 10/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
4-2-4 Pressure test drop time Pressure drop measuring time (only active if 4-2-3 > 0)
Pr. drop time (s) 0 0 60 The measurement time for the pressure
drop when performing the leakage test
of the dialyzer.
4-2-5 Pressure test drop Pressure drop limit (only active if 4-2-3 >0)
Pr. drop (mmHg) 0 0 1000 The acceptable pressure drop when per-
forming the leakage test of the dialyzer.
4-2-6 Priming flow Blood pump speed during priming of dialyzer
PrimFlow (ml/min) 200 0 500 This is the flow rate during the priming
phase as well as during the flushing phases.
4-2-7 Priming time Blood pump running time during priming of dialyzer
Priming time (s) 120 0 3600 This is the required time for the priming
phase.
4-2-8 Cycle time Venous clamp toggle time
Cycle time (s) 3 0 60 This time specifies the opening and closing
time of the venous clamp during the priming
phase.
4-2-9 Level time Time during fill of venous line
Level time (s) 0 0 60 This is the time necessary to fill up the
venous drip chamber to a correct level.
4-2-10 Flush time 1 Time during flush of dialyzer, part 1
Flushing time1 (s) 120 0 3600 During this time, the remaining air in
the dialyzer and blood lines will be expelled.
4-2-11 Flush time 2 Time during flush of dialyzer, part 2
Flushing time2 (s) 120 0 3600 If a second flush is requested, select this time.
4-2-12 Priming UF Rate Auto-priming UF rate
UF Rate (l/hour) 1.00 0.00 4.00
4-2-13 Venous line time Time during flush of venous line (only active if 4-2-9 > 0)
Ven line time (s) 0 0 120 The time when the venous clamp is open.
In order to expell any air bubbles in the
venous line.
Page 11/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
4-2-14 Early start of FM Inst = Dialysis fluid must be achieved before autopriming
can start.
Early FM start Not Inst Inst Not Inst Not Inst = Autopriming can start without dialysis fluid.
4-2-15 FM active Inst = Dialysis fluid will automatically enter the filter during
first flush.
FM active Inst Inst Not Inst Not Inst = Operator must press the Bypass key when the
dialysis fluid should enter the filter.
4-2-16 Heparin set flow (ml/h) Heparin flow during auto priming
Flow (ml/hour) 0.0 0.0 10.0
4-3 Auto priming ALT 2
See 4-2
4-4 Auto priming ALT 3
See 4-2
5 Rinse back
5-0 Rinse back install Rinse back feature active
Rinse back Not Inst Inst Not Inst
5-1 Rinseback flow (ml/min) Blood pump speed during rinse back
Rinsebck flow 100 25 500
5-2 Rinseback volume Accumulated blood volume during rinse back
Rinsebck vol (l) 0.50 0.00 5.00
5-3 Rinseback cycle time Venous clamp toggle time
Cycle time (s) 2 1 10
5-4 Rinseback fluid bypass Fluid flow in bypass during rinse back
Rinseb. Fluidbyp Not Inst Inst Not Inst
6 Disinfection
6-0 Chemical disinfection
6-0-0 Chem ALT1 Chemical disinfection bank 1
Page 12/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Page 13/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Page 14/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Page 15/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Delay minutes (min) 0 0 1440 1 – 1440 = Time in minutes to start up in Auto rinse after shut
down from Central Chem. (Cental chem. and overnight disinfection
allowed must be set to True)
6-1 Heat disinfection
6-1-0 Heating (ALT1) Heat disinfection (water)
6-1-0-0 Heating temperature Disinfection temperature for ALT1, ALT2 and ALT3
Heating temp 93 70 95 Height above sea level: Water boiling temperature:
0 100ºC
500 98.1ºC
1000 96.2ºC
1500 94.4ºC
2000 92.6ºC
2500 90.7ºC
6-1-0-1 Heating cycles Number of heat cycles. 1cycle = 1 minute
Heating cycles 12 5 255
6-1-0-2 Heating extra cool off Cool off phase at end of disinfection
Cool off (sec) DISABLED 10 600
6-1-1 Heat decalcification (ALT2) Heat disinfection (liquid citric solution)
6-1-1-0 Heat decalc. label
Label Citric 20% Label for heat disinfection bank 2
6-1-1-1 Heat decalc. concentrate Concentrated or Working (mixed) solution
Concentrate Conc Conc Work sol
6-1-1-2 Heat decalc. mixing ratio Mixing ratio of the concentrated solution (only active if 6-1-1-1 =
Conc and 6-1-1-9 = 0)
Mixing ratio 1+9 1+9 1+39
6-1-1-3 Heat decalc. dwell time Heat disinfection dwell time. If dwell time FOREVER and logical
OFF, then machine will switch off after fill-up phase
Dwell time 00:10 00:00 FOREVER
6-1-1-4 Heat decalc. heating cycles Number of heat cycles. 1cycle = 1 minute
Heating cycles 12 12 255
6-1-1-5 Heat decalc. extra cool off Cool off phase at end of disinfection
Cool off (sec) DISABLED 10 600
6-1-1-6 Heat dealc. cond. check Conductivity limits for citric acid solution
Page 16/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Cond limits 0.0, 50.0 0.0 50.0
6-1-1-7 Post chem. rinse time Rinse time for Post Chem program. This preset will be set to 30
min (as before) when updating to P3.00 if not changed manually!
Rinse Time 00:11 00:11 09:55
6-1-1-8 Post chem. rinse time UFD UFD Installed
Rinse Time UFD 00:21 00:21 09:59
6-1-1-9 Heat decalc. 0 = Not activated (6-1-1-2 is used)
Conc percent (%) 0 0 100 1-100 = The percent marked on the container, machine is
calculating the ratio automatically. (6-1-1-2 is NOT used)
6-1-2 Heat decalc. CleanCart (ALT3) Heat disinfection (CleanCart)
6-1-2-0 Heat CleanCart dwell time Heat disinfection dwell time. If dwell time FOREVER and logical
OFF, then machine will switch off after fill-up phase
Dwell Time 00:10 00:00 FOREVER
6-1-2-1 Heat CleanCart heating cycles Number of heat cycles. 1cycle = 1 minute
Heating cycles 12 12 255
6-1-2-2 Heat CleanCart extra cool off Cool off phase at end of disinfection
Cool off (sec) DISABLED 10 600
6-1-2-3 Post chem. rinse time Rinse time for Post Chem program. This preset will be set to 30
min (as before) when updating to P3.00 if not changed manually!
Rinse Time 00:13 00:11 09:55
6-1-2-5 Post chem. rinse time UFD UFD Installed
Rinse Time UFD 00:21 00:21 09:59
6-1-3 Active Heat ALT Active heat disinfection bank during start-up
Heat Dis Alt ALT1 ALT1 ALT3
6-1-4 Pump control values
6-1-4-0 Conc. pump flush recirc. Speed Conc. Pump speed during dwell
A+B pump speed (%) 10 10 100
6-1-4-1 Flush recirc. HPG pressure Suction pump regulation during dwell
Pressure (mmHg) -100 -200 +200
6-2 Rinse / Decalcification
6-2-0 Rinse/Drain (ALT1) Rinse, bank 1 (water)
6-2-0-0 Rinse/Drain total time Total rinse time
Page 17/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Total time 00:10 00:10 09:59
6-3 Remove CleanCart attention Attention to remind operator to remove CleanCart after priming.
If FALSE then CleanCart mounted during all rinse/disinfection
program
Remove CleanCrt False True False
6-4 Disinf/Decalc attn. and lim.
6-4-0 Time between two disinfections Time, in hours, between disinfections, concerns calendar time.
Time Hour 120 0 500 The time flag is stored in Saferam C and cleared by both the heat
programs and the chemical programs.
6-4-1 Time between two decalcificat. Time, in hours, between decalcifications, concerns time in
Time Hour 0 0 500 Bicarbonate mode. The time flag is stored in Saferam C and
cleared by either CleanCart C or liquid citric.
To clear the flag by CleanCart C, there is a conductivity measuring
during CleanCart heat (<15mS/cm).
6-4-2 Time between two cleanings Time, in hours, between CleanCart A / hypochlorite cleanings
Time Hour 50 0 200 concerns total time in fluid mode (Bicarbonate or Acetate).
The time flag is stored in Saferam C and cleared by either
Cleancart A or hypochloride.
To clear the flag by CleanCart A, there is a conductivity measuring
during CleanCart heat (>15mS/cm for approximately 2sec).
To clear the flag by hypochloride it is of importance that
hypochloride is placed in bank 2 since in this case it is
the bank that clears the flag.
6-5 AUTO OFF after Rinse/Drain Time in minutes before the machine automatically turns off with
blinking Rinse/Drain after a disinfection program
Time Minute Disabled Disabled,1 100
7 Single needle
7-0 Arterial phase time Arterial phase in single needle mode (start-up value)
Art time S/N (s) 10.0 2.0 20.0
7-1 Venous phase time Venous phase in single needle mode (start-up value)
Ven time S/N (s) 10.0 2.0 20.0
Page 18/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
7-2 Stroke volume limit Stroke volume limit in single needle mode
Stroke Vol (ml) 0 0 200
8 Acetate concentrates There are three options to enter the values for each Acetate
alternative,
Preset using gram/litre, molecules
Preset using mmol/litre, molecules
Preset using mmol/litre, ions
Page 19/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
- KCl Potassium Chloride
- MgCl2 Magnesium Chloride
- CaCl2 Calcium Chloride
Page 20/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
There are three options to enter the values for each Bicarbonate
alternative,
Preset using gram/litre, molecules
Preset using mmol/litre, molecules
Preset using mmol/litre, ions
Page 21/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
- B:CH3COONa Sodium Acetate
- H2O with CH3COONa Acetic Acid, hydrated
Page 22/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
9-0-1-0 BiCart 1 (gram/l, molecule)
See 9-0-0-0 "Bicarbonate 1..."
9-0-1-1 BiCart 1 (mmol/l, molecule)
See 9-0-0-1 "Bicarbonate 1..."
9-0-1-2 BiCart 1 (mmol/l, ions)
See 9-0-0-2 "Bicarbonate 1..."
9-1 Bicarbonate alt. 2
See 9-0 "Bicarbonate alt.1"
9-2 Bicarbonate alt. 3
See 9-0 "Bicarbonate alt.1"
10 BPM Settings
10-0 BPM systolic pressure limits Lower and upper alarm limit for Systolic blood pressure (mmHg)
See General Information 136 for more information. Can also be
set by the operator during treatment.
BPM sys limits 100/150 60 250
10-1 BPM diastolic pressure limits Lower and upper alarm limit for Diastolic blood pressure (mmHg)
See General Information 136 for more information. Can also be
set by the operator during treatment.
BPM dia limits 60/100 40 200
10-2 BPM mean pressure limits Lower and upper alarm limit for Mean blood pressure (mmHg) See
General Information 136 for more information. Can also be set by
the operator during treatment.
BPM mean limits 75/115 45 235
10-3 BPM pulse limits Lower and upper alarm limit for Pulse rate. See General
Information 136 for more information. Can also be set by the
operator during treatment.
BPM pulse 40/150 40 200
Page 23/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
10-4 BPM measurement interval Measurement interval for BPM measurements during treatment if
the BPM is in auto mode. Can also be set by the operator during
treatment.
BPM meas interval 30 5 60
10-5 BPM alarm mode BPM alarm on/off. Enable /Disable BPM alarms according to 10-
0,
10-1, 10-2 and 10-3. Can also be set by the operator during
treatment.
BPM alarm OFF OFF ON
10-6 BPM initial cuff pressure Initial cuff pressure
BPM init. cuff pr 180 120 250
10-7 BPM inflation pressure mode Preset = Inflation pressure set by the preset 10-6, is used.
BPM infl pr mode Measured Preset Measured Measured = First measurement: inflation pressure is set to initial
value.
Thereafter: inflation pressure = last systolic BP value + α
(α = 50mmHg)
Note! Preset 10-6 not used.
11 Profiling
11-0 Model 1
11-0-0 Na+ Acetate profiling type Na+ profiling type in Acetate
Na+ prof type OFF OFF LINEAR Profiling curve for Na+ in Acetate mode. Can be either OFF or
LINEAR. Can also be set by the operator during treatment.
11-0-1 Na+ profiling type Na+ profiling type in Bicarbonate
Na+ prof type OFF OFF LINEAR Profiling curve for Na+ in Bicarbonate mode. Can be either OFF or
LINEAR. Both Na+ and HCO3- profiling can be active in
Bicarbonate mode. Can also be set by the operator during
treatment.
11-0-2 HCO3- profiling type HCO3- profiling type
HCO3- prof type OFF OFF LINEAR Profiling curve for HCO3-. Both Na+ and HCO3- profiling can be
active in Bicarbonate mode. Can be either OFF or LINEAR. Can
also be set by the operator during treatment.
Page 24/25
AK 95 / AK 95 S,
PROGRAM VERSION 12.20
Preset parameters
Page 25/25